Service Note
CONFIDENTIAL
SOFTWARE BUG
MODEL: MT8801B
SERIAL: MT40463 through MT64169
We have found software bug on Main Software. The phenomenon of the problem is the
result of standard (spectrum analyzer method) adjacent channel power measurement
includes too much error when Tx Ref level is changed by 1dB step.
The bug has already fixed and the latest products will not have the sane problem.
Please update the software when you receive the above MT8801B from your
customer for repair and/or calibration.
The software update needs a new version ROM and a floppy disk. (See the attached
procedure.)
We will provide them for you and send them later.
The number of shipped instruments which require the update are below.
Office Qty
ACUS(East and West) 52
ACUK(include ERC) 21
ACSE 35
No calibration necessary after the software update.
CONFIDENTIAL
STEP ATTENUATORS
MG3633A,MG3641A/42A,MG3660A,MG3670A/B/C,MG3671A,MS2650/60,
MT8801A/B are using step attenuators and the attenuators should be ordered with listed
model name and description. Exchange price is not available for these attenuators.
Attenuators List
Model Name Description Drawing Number Remarks
UN84-7204D-12 SW/ATT 34K117707K MG3633A
UN85-7301D-12 SW/ATT 34K117707L MG3633A
339H38476 STEP ATTENUATOR 339H38476 MG3641A/42A
339H40998C HR S-ATT 339H40998C MG3660A
UN84-7204C-12 SW/ATT 34K117707G MG3670A/B/C
UN85-7301V-12 SW/ATT 34K117707J MG3670A/B/C
339H40998 HR S-ATT 339H40998 MS2650/60
339H40998C HR S-ATT 339H40998C MT8801A/B
Confidential ADVISORY
September 22 / 2000
1.Phenomenon
If the burst clock signal is inputted in PDC/PHS Bit Error Rate measurement,
it will become "Sync Loss Error".
Clock no Clock
Burst clock signal
Software bug
1.Phenomenon
Software bug
1.Phenomenon
◇ Measuring Object of ACeS set to Continuous mode then measuring system turn to GSM.
MT8801 becomes unable to connect to Mobile Station.
◇ The result of “Judge : Timing advance” on the All measurement screen is not renewed.
Software bug
1.Phenomenon
◇ It will become an error on Call Processing function when a message is checked by Layer3.
◇ Measurement stops when a channel is changed on the Occupied Band width screen.
◇ RX User Cal Factor value cannot be held when Main / AUX switch is changed.
◇ If the control channel of an analog is set to 313, Mobil Station(MS) will not recognize
Base Station(BS).
Rev. no. : 3.03 Updating Date : Jul. 9 / 1999
Software bugs
1 . Phenomenon
◇ The version number was different with the MT8801B and MT8801C.
It unified it from the version now.
◇ The data that we read with GP-IB and the screen data of Adjacent Channel Power
sometimes differ at the Carrier power OFF.
( There is case the GP-IB data is 0.01nW small then screen data.)
◇ The data that we read with GP-IB and the screen data of Frequency error
sometimes clearly differ.
Example
Screen data : -1.2 ppm
GP-IB data : -1.0 ppm
Rev. 4.02 FD FD 0 Updating date : 29 Sep. 2000
◇ The mantissa of the Bit Error Rate measurement data always becomes zero
that we obtained with the GP-IB Query command(BERRATE?).
Example
Screen data : 2.57×10¯³
GP-IB data : 0.00×10¯³
Software bugs
1 . Phenomenon
◇ The data that we read with GP-IB and the screen data of Adjacent Channel Power
sometimes differ at the Carrier power OFF.
( There is case the GP-IB data is 0.01nW small then screen data.)
◇ The data that we read with GP-IB and the screen data of Frequency error
sometimes clearly differ.
Example
Screen data : -1.2 ppm
GP-IB data : -1.0 ppm
◇ The mantissa of the Bit Error Rate measurement data always becomes zero
that we obtained with the GP-IB Query command(BERRATE?).
Example
Screen data : 2.57×10¯³
GP-IB data : 0.00×10¯³