You are on page 1of 1

1 2 3 4

BRIGHTNESS CONTROL WITH REMOTE PROGRAM.

CONTROLS:
R5 (INTERNAL): SETS MINIMUM CONTROL RANGE (3V @nominal line).
R10 (REAR PANEL): SETS MAXIMUM CONTOL RANGE (12.3V @nominal J1
12V OUT
line).
R16, PART OF MICROSCOPE, IS THE USER CONTROL FOR BRIGHTNESS. 5
D D
THEORY:
Q3,4 COMPOSE A ZERO CROSSING DETECTOR WHICH RESETS TIMING J1
6V OUT
CAPACITOR C2 AT THE BEGINNING OF EVERY AC CYCLE.
C2 CHARGES AT A RATE SET BY R4,5 IN A SAWTOOTH SHAPE. 4
WHEN THE THE VOLTAGE ON C2 IS GREATER THAN THE DC VOLTAGE
Q2 EMITTER, TRIAC Q5 IS TURNED ON FOR THE REMAINDER OF THE 5
HALF CYCLE. Q1
VOLTAGE AT Q2 EMITTER IS SET BY THE VOLTAGE DIVIDER PN2907 Gry
COMPOSED OF R8,10,11,16.
C3 M1 LAMP1
R1 4.7uF METER
1K
VR1 Wht
D1 LM7805
Vin 5V R9 2

GND
1N4007 R4 100 J1
C J2B 1K C
R7 R8 3
L N PLUG AC MALE 1K 220
G C1
Q5
470uF
R5 R10
1K 2N4410 100
R3 SC142B
0.5 AMP THERMAL CB

15K Q2 R12
1K
R6
JMP
BRIGHTNESS
CB1 D5 D4 J1 CONTROL
1N4148 1N4148 2N4410 R11
S1 R14
Q3 C2 2
330
5.6K 4.7uF
R16
2N4410 D3 100 OHM
(RED) R15 Q4
1N4148 J1
SW DPST TR1 GRN 1K
4
B 1 B
PART OF
6.3V
'SCOPE
ORG 3
115V

10.5V MINI PHONE JACK


J3
RED 1
R13
#1027JM
47 OHMS
2W

Title LUDL ELECTRONIC PRODUCTS Ltd


A NOTE: ALL VOLTAGE MEASURMENTS ARE MADE WITH A TRUE RMS READING METER. SLIT LAMP POWER SUPPLY 171 BRADY AV. HAWTHORNE, N.Y. 10532 A
TEL 914/769-6111 FAX 914/769.4759
Size Number Revision
310458 A
A
Date: 11-May-2022 Sheet 1of 1
File: N:\PROTEL\PSUPPLY\SLITLAMP.SCH Drawn By: RO
1 2 3 4

You might also like