You are on page 1of 6

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D4972 − 19

Standard Test Methods for


pH of Soils1
This standard is issued under the fixed designation D4972; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope* 1.4 All measured and calculated values shall conform to the
1.1 These test methods cover the measurement of the pH of guidelines for significant digits and rounding established in
soils that will pass the 2.00 mm (No. 10) sieve. Such Practice D6026.
measurements are used in the agricultural, environmental, 1.4.1 The procedures used to specify how data are collected/
geotechnical, and natural resources fields. This measurement recorded and calculated in the standard are regarded as the
determines the degree of acidity or alkalinity in soil materials industry standard. In addition, they are representative of the
suspended in water and a 0.01 M calcium chloride solution. significant digits that generally should be retained. The proce-
Measurements in both liquids are necessary to fully define the dures used do not consider material variation, purpose for
soil’s pH. This variable is useful in determining the solubility obtaining the data, special purpose studies, or any consider-
of soil minerals and the mobility of ions in the soil and ations for the user’s objectives; and it is common practice to
assessing the viability of the soil-plant environment. A more increase or reduce significant digits of reported data to be
detailed discussion of the usefulness of this parameter is given commensurate with these considerations. It is beyond the scope
in Refs (1-6)2. of these test methods to consider significant digits used in
analysis methods for engineering data.
1.2 Two methods for measuring the pH of soils are pro- 1.5 This standard does not purport to address all of the
vided. The method to be used shall be specified by the safety concerns, if any, associated with its use. It is the
requesting authority. When no method is specified, Method A responsibility of the user of this standard to establish appro-
shall be used. The pH is determined in test water and a calcium priate safety, health, and environmental practices and deter-
chloride solution for both methods. mine the applicability of regulatory limitations prior to use.
1.2.1 Method A—The pH is measured using a potentiometer 1.6 This international standard was developed in accor-
having a pH sensitive electrode system. This method can be dance with internationally recognized principles on standard-
used for any application and must be used when the application ization established in the Decision on Principles for the
warrants a higher level of resolution. Development of International Standards, Guides and Recom-
1.2.2 Method B—The pH is measured using pH sensitive mendations issued by the World Trade Organization Technical
paper. This method can be used for any application, however, Barriers to Trade (TBT) Committee.
because paper typically has a lower resolution, it provides an
approximate estimate of the pH of the soil and should not be 2. Referenced Documents
used when the application requires a higher level of resolution 2.1 ASTM Standards:3
(Note 1). C670 Practice for Preparing Precision and Bias Statements
NOTE 1—For example, paper with a sensitivity to the nearest 1 pH unit for Test Methods for Construction Materials
placed into a buffer solution of 4 should indicate a pH of 4, however, it D653 Terminology Relating to Soil, Rock, and Contained
would not indicate if the pH is 4.449 or 3.449. Fluids
1.3 Units—The values stated in SI units are to be regarded D1193 Specification for Reagent Water
as standard. No other units of measurement are included in this D2487 Practice for Classification of Soils for Engineering
standard. Purposes (Unified Soil Classification System)
D2488 Practice for Description and Identification of Soils
(Visual-Manual Procedures)
1
D3740 Practice for Minimum Requirements for Agencies
These test methods are under the jurisdiction of ASTM Committee D18 on Soil
and Rock and is the direct responsibility of Subcommittee D18.22 on Media for Engaged in Testing and/or Inspection of Soil and Rock as
Plant Growth.
Current edition approved May 1, 2019. Published May 2019. Originally
3
approved in 1989. Last previous edition approved in 2018 as D4972 – 18. DOI: For referenced ASTM standards, visit the ASTM website, www.astm.org, or
10.1520/D4972-19. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
2
The boldface numbers in parentheses refer to a list of references at the end of Standards volume information, refer to the standard’s Document Summary page on
this standard. the ASTM website.

*A Summary of Changes section appears at the end of this standard


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D4972 − 19
Used in Engineering Design and Construction 6. Interferences
D4220/D4220M Practices for Preserving and Transporting 6.1 When using Method A, a pH probe has possible inter-
Soil Samples ferences due to a suspension effect or sedimentation potential.
D4753 Guide for Evaluating, Selecting, and Specifying Bal- Refs (5) and (6) provide a more detailed discussion of the effect
ances and Standard Masses for Use in Soil, Rock, and of this mechanism.
Construction Materials Testing
D6026 Practice for Using Significant Digits in Geotechnical 6.2 This effect is the main reason Test Method G51 cannot
Data be used for general measurement of pH outside of that for
D6913 Test Methods for Particle-Size Distribution (Grada- corrosion analysis. Test Method G51 measures pH, an aqueous
tion) of Soils Using Sieve Analysis parameter, without adding any aqueous phase to the soil. This
lack of aqueous phase results in excessive soil particle-pH
G51 Test Method for Measuring pH of Soil for Use in
probe contact that overestimates the activity of the hydrogen
Corrosion Testing
ions in solution and is therefore unacceptable for general soil
analysis.
3. Terminology
6.3 The suspension effect can be mitigated by careful
3.1 Definitions: attention to inserting the probe into the aqueous phase of the
3.1.1 For definitions of common technical terms used in this slurry. The aqueous phase consists of the partially settled
standard, refer to Terminology D653. suspension.

4. Summary of Test Method 7. Apparatus


4.1 This test method is used to determine the pH of a soil. 7.1 Use one of the following devices to measure the pH.
Two similar test specimens are obtained from a representative 7.1.1 pH Meter—Potentiometer equipped with an electrode
sample that has been air-dried and screened over a 2.00 mm system with a readability to the nearest 0.1 pH unit and an
(No. 10) sieve. From the material passing the 2.00 mm (No. 10) accuracy of 60.1 pH unit or better. Follow the manufacturer’s
sieve, one specimen is mixed with test water and one is mixed instructions for the pH meter used. A silver/silver chloride
with a calcium chloride solution. They are then thoroughly electrode system or similar is also acceptable.
stirred and both are allowed to stand for one hour prior to 7.1.2 pH Paper—pH-sensitive paper having a typical range
testing using Method A or B. from 1 to 12 with sensitivity to the nearest 0.5 pH unit or better.
The range of the paper may vary based on the typical pH of the
5. Significance and Use materials being tested. It is acceptable to use paper ranging
between other values as long as the range is large enough to
5.1 The pH of the soil is a useful variable in determining the bracket the expected pH result.
solubility of soil minerals, the mobility of ions in the soil, and
assessing the viability of the soil-plant environment. 7.2 Balance—Balances shall conform to the requirements of
Specification D4753.
5.2 pH measurements are made in both test water and a 7.2.1 To determine the mass of the specimen, the balance
calcium chloride solution because the calcium displaces some shall have readability without estimation of 0.1 g or better. The
of the exchangeable aluminum. The low ionic strength coun- capacity of this balance will need to exceed the mass of the
ters the dilution effect on the exchange equilibrium by setting container plus soil. In general, a balance with a minimum
the salt concentration of the solution closer to that expected in capacity of 300 g is sufficient.
the soil solution. The pH values obtained in the solution of
7.3 Separation Sieve—2.00 mm (No. 10) sieve used to
calcium chloride are slightly lower than those measured in
separate the coarse material from the representative sample.
water due to the release of more aluminum ions which then
This sieve is subjected to rough operation and shall not be used
hydrolyses. Therefore, both measurements are needed to fully for quantitative testing.
define the character of the soil’s pH.
7.4 Specimen-Mixing Container—A glass beaker or equiva-
5.3 For the purpose of these test methods, the test specimens lent inert container with enough capacity to hold the specimen
are sieved through a 2.00 mm (No. 10) sieve. Measurements on and the test water. Usually, a 100 mL to 250 mL glass beaker
soils or soil fractions having particle sizes larger than 2.0 mm is large enough.
by these test methods may be invalid. If soil or soil fractions
with particles larger than 2.0 mm are used, it must be stated in 7.5 Drying Oven (Optional)—Vented, thermostatically con-
the report since the results may be significantly different. trolled oven capable of maintaining uniform temperatures of
105 6 5°C and 130 6 5°C as needed throughout the drying
NOTE 2—The quality of the result produced by this standard is
dependent on the competence of the personnel performing it, and the chamber. These requirements typically require the use of a
suitability of the equipment and facilities used. Agencies that meet the forced-draft oven.
criteria of Practice D3740 are generally considered capable of competent
7.6 Volumetric Flask or Cylinder—Two, glass or plastic;
and objective testing/sampling/inspection/etc. Users of this standard are
cautioned that compliance with Practice D3740 does not in itself assure one each with capacities of 1 L and 2 L. Required for making
reliable results. Reliable results depend on many factors; Practice D3740 calcium chloride hydrate solutions, otherwise, they are op-
provides a means of evaluating some of those factors. tional.

2
D4972 − 19
7.7 Thermometric Device—A thermometric device capable 10. Sampling and Test Specimens
of measuring the temperature range within which the test is 10.1 General—These test methods do not address, in any
being performed readable to 0.5°C or better and having an detail, procurement of the sample. It is assumed the sample is
accuracy of at least 60.5°C. obtained using appropriate methods and is representative of the
7.8 Miscellaneous Items—Items such as a wash/rinse bottle soil under evaluation. However, the testing agency shall
(squirt bottle), rubber scraper, spatula, plastic or glass contain- preserve all samples in accordance with Practice D4220/
ers to hold reagents, and stirring rod may be useful. D4220M, Group B, except if the as-received sample does not
meet those requirements. In which case, the water content of
8. Reagents the material does not have to be maintained.
8.1 Purity of Reagents—Reagent grade chemicals should be 10.2 Sampling—Obtain a representative sample from the
used in all tests. Unless otherwise indicated, it is intended that bulk sample. Follow guidance given in Test Method D6913,
all reagents should conform to the specifications of the Annex A2, for specimen splitting/reducing. The moist mass of
Committee on Analytical Reagents of the American Chemical the representative sample must be large enough such that, after
Society, where such specifications are available.4 Other grades air drying and screening over the 2.00 mm (No. 10) sieve, the
may be used, provided it is first ascertained that the reagent is air-dried mass will be large enough to yield two test specimens.
of sufficient purity to permit its use without lessening the 10.2.1 Place the representative sample on a tray, pan, or flat
accuracy of the determination. surface to facilitate uniform air-drying. Be careful not to lose
fines during this process. After drying, if the sample appears to
8.2 Test Water—Type III water in accordance with Specifi-
have particles larger than 2.0 mm, the sample must be screened
cation D1193 or distilled/deionized water with a pH between
over the 2.00 mm (No. 10) sieve. If there are no coarse
6.5 and 7.5.
particles present, screening is not necessary. See 5.3 for soils
8.3 Buffer Solutions—Buffer solutions are commercially having coarse particles and tested using these methods.
available and providing they meet the purity of reagents as 10.2.2 Air drying is necessary to accomplish sieving and to
described above, they are an acceptable alternative to making control the amount of water present at the time of measure-
buffer solutions. The following NIST traceable buffer solutions ment.
are typically needed: pH of 4.0, 7.0, and 10.0. When making
10.3 Test Specimens—Since the soil is in the air-dried state,
buffer solutions, refer to Annex A1 for the necessary chemicals
segregation and loss of fines are common problems and care
and instructions.
should be taken to make sure the test specimens accurately
8.4 Calcium Chloride Hydrate Stock Solution (CaCl2· represent the sample. From the soil passing the 2.00 mm (No.
2H2O)—Using the 1 L volumetric flask or cylinder, dissolve 10) sieve, obtain two similar specimens for testing. Typically,
147 g of CaCl2· 2H2O in test water and allow it to come to a mass of 10 6 0.1 g is adequate, however, 20 6 0.1 g or 40
room temperature. Then, dilute by adding test water to the 1 L 6 0.1 g specimens may be necessary to allow for the pH
mark on the volumetric flask or cylinder. Mix the solution electrode to be fully submerged.
thoroughly. The molarity of this solution is 1.0 M (Note 3). 10.3.1 For both methods, determine and record the mass of
8.5 Calcium Chloride Hydrate Solution (CaCl2· 2H2O) the test specimens to the nearest 0.1 g.
Using the 2 L volumetric flask or cylinder, add 20.0 mL of the
calcium chloride hydrate stock solution and test water by 11. Standardization
adding test water to the 2 L mark on the volumetric flask or 11.1 Before each use standardize the pH meter using the
cylinder. Mix the solution thoroughly. The pH of this solution buffer solutions in 8.3 or Annex A1 following the manufactur-
should be between 5 and 7 and the molarity is 0.01 M. er’s instructions of the meter. Select buffers so that the
NOTE 3—It is acceptable to purchase the calcium chloride solutions expected pH value of the tested soil is bracketed. Typically
provided they have the same molarity and pH as described above. buffers of 4, 7, and 10 meet this requirement for most soils;
however, in some applications buffers with higher or lower pH
9. Hazards values are needed. Follow the manufacturer’s instructions to
9.1 These solutions may cause eye and skin irritation. Care adjust the pH meter accordingly. Record the date, time, and
should be taken to avoid contact with skin and eyes and name or initials of person completing the standardization. If
solutions should not be ingested. The use of latex or rubber other data in association with the standardization is
gloves, and possibly safety glasses, when working with the determined, record this information (Note 4).
solutions is recommended. Wipe up spills immediately. Rinse 11.1.1 After every 10 specimens or at the end of testing,
skin coming into contact with the solutions with water quickly. whichever occurs first, verify the pH meter’s standardization
by inserting the probe into the pH 7.0 buffer solution. The
value should be between 6.9 and 7.1. Read and record the pH
4
value to the nearest 0.1 pH unit. If the value falls outside of this
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not range, the pH meter must be re-standardized and the previous
listed by the American Chemical Society, see Analar Standards for Laboratory specimens must be retested to verify the results.
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville, 11.2 Check the pH sensitive paper by inserting a small
MD. portion of the paper into the pH 4.0, 7.0, and 10.0 buffer

3
D4972 − 19
solutions and compare each portion to the color guide provided 13.2.4.3 Thermometric device identification number.
by the manufacturer of the pH paper. This check is typically 13.2.4.4 Specimen mixing container identification number.
undocumented. 13.2.5 Description and classification of the soil in accor-
NOTE 4—Some pH meters output a linearity or other data to indicate the dance with Practice D2488 or when Atterberg limit data are
meter is working properly. available, Practice D2487.
12. Procedure 13.2.6 Describe any problems that were encountered.
12.1 Preparation—Place one of the test specimens in a 13.3 Record as a minimum the following test specimen
specimen mixing container and add the same amount of test data:
water in mL as the specimen mass in g to make a 1 to 1 slurry. 13.3.1 Method used, A or B.
Place the other test specimen into a specimen mixing container 13.3.2 The pH of the soil in test water and in the calcium
and add the same amount of calcium chloride solution in mL as chloride solution.
the specimen mass in g to make a 1 to 1 slurry (Note 5). Mix 13.3.3 The temperature of the slurry at the time of
both slurries thoroughly and allow them to stand for 1 h. The measurement, as applicable.
slurry should be at approximately room temperature (15 to 13.3.4 If testing was performed on soil that contained coarse
25°C) at the time of pH measurement. In cases where the test particles as discussed in 5.3, include a statement indicating the
specimen readily absorbs the water such that no aqueous phase results may be significantly different.
exists, it is permissible to make a 1 to 2 slurry. This situation
can occur with clay soils that swell. 14. Precision and Bias
12.2 Method A—After 1 h, verify the temperature of the 14.1 Precision:
slurry is at room temperature then, fully submerge the pH 14.1.1 Within-Laboratory Precision—The within laboratory
electrode into the aqueous part of the slurry. Be careful not to standard deviations for Method A are 0.031 (pH units) for the
insert the electrode into the solids. Read and record the pH of water mixture and 0.1395 (pH units) for the calcium chloride
the slurry to the nearest 0.1 pH unit. If the pH meter does not mixture. Therefore, results of two properly conducted tests in
automatically compensate for temperature, determine and re- the same or different laboratories should not differ by more
cord it to the nearest 0.5°C or better. than 0.0655 (pH units) for the water mixture and 0.389 pH units
for the calcium chloride mixture. The within-laboratory stan-
12.3 Method B—After 1 h, determine and record the tem- dard deviations for Method B are 0.189 (pH units) for the water
perature of the slurry to the nearest 0.5°C then, carefully insert mixture and 0.2125 (pH units) for the calcium chloride mixture.
a strip of the pH sensitive paper into the aqueous part of the Therefore, results of two properly conducted tests in the same
slurry. Be careful not to insert it into the solids. Read and or different laboratories should not differ by more than 0.535
record the pH of the slurry to the nearest 0.5 pH unit. (pH units) for the water mixture and 0.60 pH units for the
NOTE 5—For example, if the mass of the test specimen is 20.1 g, then
20.1 mL of test water or calcium chloride solution is added to the test
calcium chloride mixture.
specimen to make a 1 to 1 ratio. In rare instances and usually only for 14.1.1.1 The precision of Method A presented was deter-
clays that, can and do, absorb significant amounts of water relative to their mined by the National Technical Center of the United States
mass, a 1 to 5 ratio may be used when a 1 to 2 ratio does not yield an Department of Agriculture. In their evaluation they used 174
adequate amount of aqueous phase. replicates for the water mixture and 32 replicates in testing the
13. Report: Test Data Sheet(s)/Form(s) calcium chloride mixture.
14.1.1.2 The precision of Method B presented was deter-
13.1 The methodology used to specify how data are re- mined by the United States Army Environmental Hygiene
corded on the test data sheet(s)/form(s), as given below, is Agency. In their evaluation they used 25 replicates in testing
covered in 1.4 and in Practice D6026. each mixture.
13.2 Record as a minimum the following general informa- 14.1.2 Between-Laboratory Precision—The between-
tion (data): laboratory standard deviation has not been determined for
13.2.1 Identification of the soil being tested, such as the either method. Subcommittee D18.22 is actively seeking data
project identification, boring number, sample number, and to evaluate the between laboratory precision of this test
depth. method.
13.2.2 Test number, if any, testing dates and the initials of 14.2 Bias—There is no accepted reference value for this test
the person(s) who performed the test. method; therefore, bias cannot be determined.
13.2.3 Describe any material that was excluded from the
specimen and any special selection and preparation process. 15. Keywords
13.2.4 The following apparatus identification used during
the test: 15.1 acidity; alkalinity; pH; reaction; soil
13.2.4.1 Balance identification number.
13.2.4.2 pH meter and electrode identification number
(Method A only). 5
These data satisfy the 1S and D2S requirements outlined in Practice C670.

4
D4972 − 19
ANNEX

A1. MAKING BUFFER SOLUTIONS

(Mandatory Information)

A1.1 General—Commercially available buffer solutions can Stir the solution thoroughly. The pH of this solution should be
be used to standardize and verify a pH meter; however, buffer 6.9 at 20°C. The molarity is 0.025 M. The effect of temperature
solutions can be made. The following sections provide the is as follows:
necessary information and chemicals to make pH solutions of Temperature (°C) pH
4, 7, and 10. For all buffers, record the temperature to the 0 7.0
nearest 0.5°C. 10 6.9
20 6.9
A1.1.1 pH 4—Potassium Phthalate Buffer Solution (10.21 30 6.8
g/L)—Dry the potassium phthalate (C8H4K2O4) in the oven for 40 6.8

1 hour at 105 6 5°C, then dissolve 10.21 g of it in test water A1.1.3 pH 10—Carbonate Buffer Solution—Dry both salts
and dilute to 1 L by adding test water to the 1 L mark on the in the oven for 2 h at 130 6 5°C before use, then dissolve 2.10
volumetric flask or cylinder. Stir the solution thoroughly. The g of sodium bicarbonate (NaHCO3) and 2.65 g sodium
pH of this solution should be 4.0 at 20°C. Protect the solution carbonate (Na2CO3) in test water and dilute to 1 L by adding
against evaporation and against contamination with molds. test water to the 1 L mark on the volumetric flask or cylinder.
Replace the solution when mold is noticed. There is no
Stir the solution thoroughly. The pH of this solution should be
temperature correction needed if the temperature of the solu-
10.1 at 20°C. The molarity is 0.025 M. The effect of tempera-
tion is between 5°C and 37°C. The molarity of this buffer is
ture is as follows:
0.05 M.
Temperature (°C) pH
A1.1.2 pH 7—Potassium Phosphate Buffer Solution—Dry 0 10.3
both salts in the oven for 2 h at 130 6 5°C before use, then 10 10.2
20 10.1
dissolve 3.40 g of monopotassium phosphate (KH2PO4) and 30 10.0
3.55 g of KH2HPO4 in test water and dilute to 1 L by adding 40 9.9
test water to the 1 L mark on the volumetric flask or cylinder.

REFERENCES

(1) Garrels, R. M., and Christ, C. L., Solutions, Minerals, and Equilibria, (4) Davies, J. T., and Rideal, E. K., Interfacial Phenomena, Academic,
Freeman Cooper, San Francisco, CA, 1965. New York, NY, 1963.
(2) Greenland, D. J., and Hayes, M. H. B., The Chemistry of Soil (5) Hunter, R. J., Zeta Potential in Colloid Science, Academic, New York,
Processes, Wiley, New York, NY, 1981. NY, 1981.
(3) Sposito, G., The Thermodynamics of Soil Solutions, Clarendon, (6) Perrin, D. D., and Dempsey, B., Buffers for pH and Metal Ion Control,
Oxford, United Kindom, 1981. Chapman and Hall, London, United Kindom, 1974.

SUMMARY OF CHANGES

Committee D18 has identified the location of selected changes to this standard since the last issue
(D4972 – 18) that may impact the use of this standard. (May 1, 2019)

(1) Included ratios other than 1 to 1 in 12.1 and Note 5. (2) Clarified aqueous phase as it relates to suspension effect.

Committee D18 has identified the location of selected changes to this standard since the last issue
(D4972 – 13) that may impact the use of this standard. (July 1, 2018)

(1) Updated the Scope section to include descriptions of the (4) Moved making buffer solutions to annex and included
test methods. information about commercially available buffers; updated
(2) Updated Summary of Test Method section. water requirements.
(3) Added missing apparatus to apparatus section. (5) Added Hazards and Sampling and Test specimen sections.

5
D4972 − 19
(6) Clarified the standardization of the pH meter. (8) Updated the Report section.
(7) Reworded procedure section for clarity.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

You might also like