You are on page 1of 9

Providing Ventilation to the Newborn Infant

in the Delivery Room


Justin B. Josephsen, MD,* Marya L. Strand, MD, MS*
*Department of Pediatrics, Saint Louis University, St. Louis, MO

Education Gap
Ventilating the lungs of the newborn is the most important element of
neonatal resuscitation. Recognizing and delivering effective ventilation
remains a critical skill for all attendants at a newborn delivery.

Objectives After completing this article, readers should be able to:

1. Recognize the critical role of ventilation in the delivery room.


2. Identify the signs of effective ventilation.
3. Understand devices that deliver positive pressure ventilation.

INTRODUCTION

The Neonatal Resuscitation Program (NRP) emphasizes the importance of ven-


tilation in the resuscitation of compromised newborns. (1) The purpose of this re-
view is to explain the physiologic necessity of ventilation in newborn resuscitation
and to describe the different methods of providing ventilation in the delivery room.
We will also discuss the assessment of ventilation to ensure efficacy.

AUTHOR DISCLOSURE Drs Josephsen and CARDIORESPIRATORY TRANSITION AT BIRTH


Strand have disclosed no financial
relationships relevant to this article. This
Blood flow during fetal life is focused on perfusion of the brain and major organs
commentary does not contain a discussion of with oxygenated blood and avoidance of fluid-filled lungs (Fig 1A). This blood flow
an unapproved/investigative use of is dependent on oxygenated umbilical venous blood from the placenta.
a commercial product/device.
Normal cardiopulmonary transition of the newborn from intra- to extrauterine life
ABBREVIATIONS requires a rapid and complex process of fetal lung fluid clearance, expansion of the al-
BPD bronchopulmonary dysplasia veolar spaces with air, and redirection of cardiac output toward the lungs. All of these
CPAP continuous positive airway pressure events take place independent of intervention in most newborns at the time of delivery.
ETT endotracheal tube (2) Failure of any of these events can cause delays in transition or progress to asphyxia.
FiO2 fraction of inspired oxygen
Fetal lung fluid produced by the pulmonary epithelium fills the pulmonary
NRP Neonatal Resuscitation Program
PEEP positive end-expiratory pressure
spaces during gestation. The pressure of fluid within air spaces and resistance
PIP peak inspiratory pressure to outflow through the glottis facilitates the growth and development of lung
PPV positive pressure ventilation parenchyma. Rapid fluid clearance at birth likely occurs via several mechanisms,

e658 NeoReviews
Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
Figure 1. A. Fetal circulation path. Only a small amount of blood travels to the lungs. There is no gas exchange in the lung. Blood returning to the right
side of the heart from the umbilical vein has the highest oxygen saturation. B. Transitional circulation path. The infant breathes, pulmonary resistance
decreases and blood travels to the lungs. Gas exchange occurs in the lungs. Blood returning to the left side of the heart from the lungs has the highest
oxygen saturation. (From Textbook of Neonatal Resuscitation, 7th ed., 2016. Reproduced with permission from the American Academy of Pediatrics.)

including fetal postural changes, cellular sodium channel a period of secondary apnea. Secondary apnea is followed by
pumps, and transpulmonary pressure from inspiration after bradycardia and eventual hypotension. (7)
delivery. (3) Infants with apnea are unable to contribute to These early animal studies also showed that ventilation
the transpulmonary pressure (which can range from –28 to of the lungs is sufficient, by itself, for resuscitation of the as-
–105 cm H2O pressure), (4) which can significantly com- phyxiated newborn. (6)(7)(8)(9) Only newborns with critically
promise fluid clearance from the air spaces. low mean arterial pressure were unable to be resuscitated with
Once the fetal lung fluid is cleared and air fills the alveoli, ventilation alone. It is important to note that effective ventilation
the pulmonary arterial pressure drops precipitously to (with adequate expansion of the pulmonary air spaces and
increase blood flow from the heart to the lungs. With this perfusion of the lungs) is the required resuscitative measure.
change in cardiac output and transition of gas exchange to Chest compressions alone do not revive asphyxiated newborns,
the pulmonary bed, the left ventricular preload shifts from and are useful only as an adjunct to adequate ventilation. (9)
the umbilical venous return from the placenta to pulmonary
venous return (Fig 1B). The systemic vascular resistance Assessment of Ventilation Efficacy
increases simultaneously to result in left-to-right flow An increasing heart rate response is the most reliable indi-
through the ductus arteriosus and the beginning of ductal cator of adequate ventilation. (10) Upon initiation of positive
closure. pressure ventilation (PPV) during a neonatal resuscitation,
Lung expansion and establishment of pulmonary functional the assessment must include an indication of heart rate
residual capacity and respiration is vital to the normal transi- response. (1) Other ancillary indicators of effective ventilation
tion of the newborn. When the umbilical cord is clamped after in the delivery room include visible chest wall rise, exhalation
delivery, the venous return from the placenta abruptly ceases. If of carbon dioxide, and respiratory function monitoring.
the lungs have not yet expanded before the clamping of the Rise of the chest wall during PPV is a quick and easily
umbilical cord, the preload to the left ventricle is compromised assessed marker of effective ventilation. The accuracy of the
and cardiac output falls. With the resultant hypoxemia, brady- human perception of chest rise has been questioned, however.
cardia ensues with further respiratory failure and asphyxia. (11) Visual impediments, such as equipment or other person-
The newborn’s physiologic response to asphyxia was nel, practitioner positioning, and abdominal movement instead
established in animal experiments in the 1960s. (5)(6)(7)(8) of chest wall movement, can impede proper interpretation.
The sequence of responses to asphyxia is the same for Studies have reported that accurate assessment of chest wall
newborn animals and humans. Primary apnea occurs when rise in the preterm population is particularly difficult. (12)(13)
the pulmonary expansion and perfusion are compromised, Colorimetric carbon dioxide monitoring has been stud-
usually from the fluid-filled alveoli. Gasping then precedes ied as a proxy for adequacy of gas exchange during PPV.

Vol. 18 No. 11 NOVEMBER 2017 e659


Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
Figure 2. The 6 ventilation corrective steps: MR. SOPA (From Textbook of Neonatal Resuscitation, 7th ed., 2016. Reproduced with permission from the
American Academy of Pediatrics.)

Placement of the carbon dioxide monitor in line with the of troubleshooting ventilation, in the face of nonresponse
PPV device and mask can reflect the inspiratory/expiratory of the newborn, should be completed in 15 to 30 seconds,
cycle of PPV. A cycling color change during exhalation exclusive of alternative airway placement. A video demon-
indicates the presence of a patent airway. (14) A change in strating the troubleshooting can be found by clicking on
color on the monitor, when used, is also observed before the the Video link below or on the free NRP mobile application,
response in heart rate, indicating that efficacy of ventilation available at https://itunes.apple.com/us/app/nrp-app-neonatal-
precedes a heart rate response. (15) resuscitation/id1125320382?mt=8 or https://play.google.com/store/
Higher technology devices to monitor respiratory func- apps/details?id=com.nrp.nrp.
tion have also been used to assess ventilatory efficacy in the
delivery room. These monitoring devices allow for inspira-
tory and expiratory tidal volume measurements, gas flows,
and mask leak and pressures. (13) The use of such high-
technology monitoring allows practitioners to adjust mask
position, view tidal volume being delivered, and more fully
understand the physiology of the infant during resuscita-
tion. Concern remains, however, about how much informa-
tion a health care practitioner can integrate and properly use
in a crisis such as a delivery room resuscitation. (16)

Troubleshooting Ineffective Ventilation Video: Click here to view the video.


A lack of heart rate response to PPV requires investigation.
The mnemonic MR. SOPA was introduced by the NRP (Fig 2). VENTILATION TECHNIQUES
(1) This mnemonic addresses the fact that most of the difficulty
with effective PPV results from mask leak or position as well PPV administration is currently recommended for newborns
as airway patency. (13) If repositioning the airway and adjusting (preterm and term) who have poor respiratory effort, apnea,
the mask to eliminate leak do not result in lung expansion with or bradycardia. Positive end-expiratory pressure (PEEP) to
improvement in heart rate, there may be a blockage to airway maintain functional residual capacity is postulated to be
patency. Suctioning with a bulb or catheter may be indicated. beneficial when administered with PPV, though there are
Opening the mouth of the newborn decreases resistance to few studies available to support this speculation. Two recent
airflow compared with the nares. If PPV remains ineffective, the studies showed more rapid improvement in heart rate and no
newborn may require increased pressure to establish lung reduction in mortality, need for chest compressions, or need
inflation and functional residual capacity. (17) If the newborn for delivery room intubation when PPV was applied with
does not respond to increased inspiratory pressure, an a device supplying PEEP. (18)(19) In one of these studies, the
alternative airway may be required. This entire process highest need for supplemental oxygen to reach saturation

e660 NeoReviews
Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
targets was reduced when using a device that can supply hemorrhage. (24)(25)(26) Furthermore, similar respiratory
PEEP, though the clinical significance may be small (highest outcomes of former extreme premature infants at 8 years of
oxygen needed 46% vs 50%). (18) This study also showed age were similar in an early cohort (1991–1992) that used
a reduced maximum positive inspiratory pressure needed to CPAP less frequently, compared with a later cohort (2005),
provide ventilation to the infant in the device that supplied which used CPAP more frequently. (27) Given this, CPAP
PEEP. The current NRP and International Liaison Committee may be preferred over intubation and mechanical ventila-
on Resuscitation guidelines recommend approximately 5 cm tion in spontaneously breathing premature infants. (10)
H2O of PEEP when resuscitating preterm infants. (10)
The optimal length of the inspiratory time (i-time) and VENTILATION DEVICES
the inspiratory-to-expiratory (I:E) ratio are unknown. The
The current NRP guidelines suggest that PPV can be effectively
current 2015 NRP guidelines do not offer guidance on
administered with any of 3 devices: a T-piece resuscitator, flow-
a recommended i-time, but do suggest an I:E ratio of 1:2,
inflating bag, or a self-inflating bag (Fig 3). (10)(18)(19) Though
using the “Breathe, Two, Three” sequence. (1) Using the
any of these is considered acceptable, each device has relative
recommended respiratory rate of 40 to 60 breaths/min, this
strengths and weaknesses. The choice should be based on the
would suggest an i-time of 0.25 to 0.33 seconds.
skills of the resuscitator, the availability of compressed gasses,
A sustained initial lung inflation of more than 5 seconds
and the type of ventilation assistance needed (PPV or CPAP).
has been postulated to be beneficial. Three randomized
controlled studies examining 404 premature newborns T-Piece Resuscitator
showed that this approach decreased the need for intubation The T-piece resuscitator (Fig 3, last panel) requires compressed
in the first 72 hours after birth. (20)(21)(22) However, these air and oxygen sources with a blender. When PPV is needed,
studies did not find a difference in either bronchopulmo- the device provides reliable PEEP, and when PPV is not
nary dysplasia (BPD) or decreased mortality in these co- needed, it also provides reliable CPAP. It delivers a consistent
horts. Because of this, routine use of sustained inflation for peak inspiratory pressure (PIP) for each breath. A consistent
more than 5 seconds is not currently recommended, though PIP allows for a more consistent tidal volume to be delivered
in individual circumstances, it might be used. (23) Larger, if lung compliance remains stable, reducing the potential
randomized controlled trials are currently ongoing, which lung injury from higher pressures and tidal volumes. (28)
may change future practice recommendations. (29) Positive pressure is delivered by occluding the PEEP valve
Extremely premature infants with spontaneous breathing at the end of the device. Increasing or decreasing the PIP or
and infants in respiratory distress may benefit from contin- PEEP delivered to the neonate is achieved by manually adjust-
uous positive airway pressure (CPAP) without the need for ing the respective dials. Though this device may be easier to
PPV. Compared with PPV and intubation, using CPAP for use and even preferred by some experienced and inexperi-
initial ventilation support resulted in reduced delivery room enced providers, (30) PIP adjustment has been shown to occur
intubations for spontaneously breathing infants, but did not more slowly than with other ventilation devices. (31) This
result in a decrease in death, BPD, or intraventricular device can deliver a fraction of inspired oxygen (FiO2) of 1.0.

Figure 3. Three types of devices are commonly used for positive pressure ventilation: a self-inflating bag, a flow-inflating bag, or a T-piece resuscitator.
(From Textbook of Neonatal Resuscitation, 7th ed, 2016. Reproduced with permission from the American Academy of Pediatrics.)

Vol. 18 No. 11 NOVEMBER 2017 e661


Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
Flow-inflating Bag endotracheal tube (ETT) sizes of 2.5 to 3.5 mm, depending
The flow-inflating bag (Fig 3, middle panel) also requires on estimated infant size. The 4.0-mm ETT is no longer
continuous compressed air and oxygen sources along with recommended for use in the delivery room. (1)
a blender. It can deliver both reliable PEEP and CPAP, but
the delivery of effective PIP is operator-dependent. Positive Laryngeal Mask
pressure is generated by squeezing the attached bag, which is The laryngeal mask device is an advanced airway that is inserted
inflated with the compressed gas. To ventilate with a higher in the mouth and rests on the hypopharynx, over the supra-
PIP, the provider must squeeze the bag with more force. This glottic structures, to provide ventilation through an isolated
device may require more technical skill to operate compe- trachea. This can be used as an alternative to endotracheal
tently. Similar to the T-piece resuscitator, the flow-inflating intubation, and appears to provide more effective ventilation
bag is capable of delivering an FiO2 of 1.0 and can more than bag-mask ventilation alone. (37)(38) The laryngeal mask
reliably provide a sustained breath of at least 5 seconds. (32) can be placed quickly with success, which might provide an
advantage over endotracheal intubation when skilled providers
Self-inflating Bag of intubation are not present. (39)(40) This device can provide
The self-inflating bag (Fig 3, first panel) is the only device that effective ventilation for infants of at least 34 weeks’ gestation with
can deliver PPV without an external source of compressed gas a weight greater than 2,000 g. It is not well-studied in smaller or
to ventilate the lungs. Supplemental oxygen can be provided more premature infants. It is also not well-studied when used
when the bag is attached to an oxygen source. Because of air with coordinated neonatal chest compressions or for admin-
mixing in the self-inflating bag (pop-off valve, expiratory valve, istration of emergency medications, such as epinephrine.
etc), oxygen concentrations delivered to the infant cannot be
precisely known. By supplying an FiO2 of 1.0 at 5 L/min flow VIGNETTE
with NRP-recommended PIP and respiratory rates, more The video shows a newborn resuscitation in the delivery
than 80% oxygen may be delivered to the infant without room. Multiple, high-resolution cameras are able to record
a reservoir. With the addition of a reservoir, higher oxygen the infant, personnel, and instrument panels during the
concentrations can be achieved. (33) stabilization. A 605-g infant was born at 24 6/7 weeks’
With the self-inflating bag, PEEP may be delivered with the gestation following unrelenting preterm labor. The neonate
addition of a PEEP valve to the device, but it may not be received 60 seconds of delayed cord clamping. The heart
effective or reliable. (19)(34)(35)(36) This valve is typically rate remained below 100 beats/min until corrective mea-
spring-loaded, and is affixed to the expiratory end of the device, sures for ventilation were taken appropriately.
adjacent to the expiratory valve. During exhalation, the expira-
tory valve opens as a result of the pressure differential between
the high pressure at the end of inspiration and the lower
atmospheric pressure. The adjustable spring of the PEEP valve
compresses the expiratory valve, increasing expiratory resis-
tance, and thus, maintaining pressure during the exhalation
phase. As the intrathoracic and extrathoracic pressures equal-
ize, the resulting pressure differential at the end of exhalation
continues to fall until another breath is given through the self-
inflating device. Because the end-expiratory pressure will fall to
zero in the absence of ongoing PPV, this device cannot be used
to provide CPAP, even when a PEEP valve is present.

ADVANCED AIRWAYS

Endotracheal intubation is indicated if effective ventilation


cannot be provided with a standard face mask, if prolonged
ventilation is needed, if chest compressions are performed, or
in specific clinical circumstances (ie, congenital diaphrag-
matic hernia). The NRP currently recommends uncuffed Video: Click here to view the video.

e662 NeoReviews
Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
Key Events in Video Resuscitation (NRP), 7th ed. Elk Grove Village, IL: American
Academy of Pediatrics; 2016
1. Team briefing
2. Perlman JM, Risser R. Cardiopulmonary resuscitation in the
2. Mask and airway repositioning
delivery room: associated clinical events. Arch Pediatr Adolesc Med.
3. Increase in positive airway pressure 1995;149(1):20–25
4. Unsuccessful intubation 3. Hooper SB, Te Pas AB, Kitchen MJ. Respiratory transition in the
5. Improvement in heart rate and oxygen saturations after newborn: a three-phase process. Arch Dis Child Fetal Neonatal Ed.
correction of incorrect respiratory rate and a longer I:E ratio 2016;101(3):F266–F271 10.1136/archdischild-2013-305704

6. Successful intubation with a stable heart rate 4. Vyas H, Field D, Milner AD, Hopkin IE. Determinants of the first
inspiratory volume and functional residual capacity at birth. Pediatr
The infant was discharged from the hospital in room air Pulmonol. 1986;2(4):189–193

near term postmenstrual age. All individuals have provided 5. Adamsons K Jr, Behrman R, Dawes GS, James LS, Koford C.
Resuscitation by positive pressure ventilation and tris-
written permission for educational use of this video.
hydroxymethylaminomethane of rhesus monkeys asphyxiated at
birth. J Pediatr. 1964;65(6):807–818
CONCLUSION 6. Campbell A, Cross K, Dawes G, Hyman A. A comparison of air
and O2, in a hyperbaric chamber or by positive pressure
The establishment of effective ventilation is the most critical ventilation, in the resuscitation of newborn rabbits. J Pediatr.
aspect of neonatal resuscitation. In an infant with apnea or 1966;68(2):153–163
7. Cross KW. Resuscitation of the asphyxiated infant. Br Med Bull.
bradycardia, this can be achieved using bag-mask ventila-
1966;22(1):73–78
tion. All the available devices for bag-mask ventilation are
8. Dawes GS, Jacobson HN, Mott JC, Shelley HJ, Stafford A. The
effective, but the provider must understand the relative treatment of asphyxiated, mature foetal lambs and rhesus monkeys
strengths and limitations of each device. with intravenous glucose and sodium carbonate. J Physiol.
1963;169:167–184 10.1113/jphysiol.1963.sp007248
A rising heart rate is the most reliable marker of effective
9. Godfrey S. Respiratory and cardiovascular changes during asphyxia
ventilation. If this cannot be achieved with bag-mask ven-
and resuscitation of foetal and newborn rabbits. Q J Exp Physiol
tilation, ventilation corrective steps (MR. SOPA) must be Cogn Med Sci. 1968;53(2):97–118
taken, and if the infant does not respond, alternative airway 10. Wyckoff MH, Aziz K, Escobedo MB, et al. Part 13: Neonatal
placement is imperative. When endotracheal intubation is Resuscitation: 2015 American Heart Association Guidelines update
for cardiopulmonary resuscitation and emergency cardiovascular
not successful, or if the providers are not skilled in ETT
care. Pediatrics. 2015;136(suppl 2):S196–S218. doi:10.1542/
placement, a laryngeal mask should be strongly considered peds.2015-3373G.
as an alternative. 11. Tracy M, Downe L, Holberton J. How safe is intermittent positive pressure
ventilation in preterm babies ventilated from delivery to newborn
intensive care unit? Arch Dis Child Fetal Neonatal Ed. 2004;89(1):F84–F87
10.1136/fn.89.1.F84
12. Poulton DA, Schmölzer GM, Morley CJ, Davis PG. Assessment
American Board of Pediatrics of chest rise during mask ventilation of preterm infants in the
Neonatal-Perinatal Content delivery room. Resuscitation. 2011;82(2):175–179 10.1016/j.
resuscitation.2010.10.012
Specifications 13. Schmölzer GM, Kamlin OCOF, O’Donnell CPF, Dawson JA,
• Know the proper approach to airway management in the Morley CJ, Davis PG. Assessment of tidal volume and gas leak
delivery room. during mask ventilation of preterm infants in the delivery room.
• Know the indications for assisted ventilation, including Arch Dis Child Fetal Neonatal Ed. 2010;95(6):F393–F397 10.1136/
continuous positive airway pressure, immediately after birth and adc.2009.174003
how to assess its effectiveness. 14. Leone TA, Lange A, Rich W, Finer NN. Disposable colorimetric
carbon dioxide detector use as an indicator of a patent airway during
• Understand how to use self-inflating and flow-inflating bags
noninvasive mask ventilation. Pediatrics. 2006;118(1):e202–e204
or T-piece resuscitators to provide assisted ventilation
10.1542/peds.2005-2493
immediately after birth.
15. Blank D, Rich W, Leone T, Garey D, Finer N. Pedi-cap color change
precedes a significant increase in heart rate during neonatal
resuscitation. Resuscitation. 2014;85(11):1568–1572 10.1016/j.
resuscitation.2014.08.027
16. Gawron VJ, Drury CG, Fairbanks RJ, Berger RC. Medical error and
References human factors engineering: where are we now? Am J Med Qual.
2006;21(1):57–67 10.1177/1062860605283932
1. American Academy of Pediatrics and American Heart Association. 17. te Pas AB, Davis PG, Hooper SB, Morley CJ. From liquid to air: breathing
In: Weiner GM, Zaichkin J, Kattwinkel J, eds. Textbook of Neonatal after birth. J Pediatr. 2008;152(5):607–611 10.1016/j.jpeds.2007.10.041

Vol. 18 No. 11 NOVEMBER 2017 e663


Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
18. Szyld E, Aguilar A, Musante GA, et al; Delivery Room Ventilation 29. Jobe AH. Mechanisms of lung injury and bronchopulmonary
Devices Trial Group. Comparison of devices for newborn ventilation dysplasia. Am J Perinatol. 2016;33(11):1076–1078 10.1055/s-0036-
in the delivery room. J Pediatr. 2014;165(2):234–239.e3 10.1016/j. 1586107
jpeds.2014.02.035 30. Hussey SG, Ryan CA, Murphy BP. Comparison of three manual
19. Dawson JA, Schmölzer GM, Kamlin COF, et al. Oxygenation with T- ventilation devices using an intubated mannequin. Arch Dis
piece versus self-inflating bag for ventilation of extremely preterm Child Fetal Neonatal Ed. 2004;89(6):F490–F493 10.1136/
infants at birth: A randomized controlled trial. J Pediatr. 2011;158 adc.2003.047712
(6):912–918.e1-2. doi:10.1016/j.jpeds.2010.12.003. 31. Hartung JC, Dold SK, Thio M, tePas A, Schmalisch G, Roehr CC.
Time to adjust to changes in ventilation settings varies significantly
20. Harling AE, Beresford MW, Vince GS, et al. Does sustained lung
between different T-piece resuscitators, self-inflating bags, and
inflation at resuscitation reduce lung injury in the preterm infant?
manometer equipped self-inflating bags. Am J Perinatol. 2014;31
Arch Dis Child Fetal Neonatal Ed. 2005;90(5):F406–F410.
(6):505–512 10.1055/s-0033-1354562
doi:10.1136/adc.2004.059303.
32. Bennett S, Finer NN, Rich W, Vaucher Y. A comparison of three
21. Lindner W, Högel J, Pohlandt F. Sustained pressure-controlled neonatal resuscitation devices. Resuscitation. 2005;67(1):113–118
inflation or intermittent mandatory ventilation in preterm infants in 10.1016/j.resuscitation.2005.02.016
the delivery room? A randomized, controlled trial on initial respiratory
33. Thió M, Bhatia R, Dawson JA, Davis PG. Oxygen delivery using
support via nasopharyngeal tube. Acta Paediatr. 2005;94(3):303–309
neonatal self-inflating resuscitation bags without a reservoir. Arch
10.1080/08035250410023647 Dis Child Fetal Neonatal Ed. 2010;95(5):F315–F319 10.1136/
22. Lista G, Boni L, Scopesi F, et al; SLI Trial Investigators. Sustained adc.2009.166462
lung inflation at birth for preterm infants: a randomized clinical 34. Morley CJ, Dawson JA, Stewart MJ, Hussain F, Davis PG. The effect
trial. Pediatrics. 2015;135(2):e457–e464 of a PEEP valve on a Laerdal neonatal self-inflating resuscitation
23. Perlman JM, Wyllie J, Kattwinkel J, et al. Part 7: Neonatal Resuscitation bag. J Paediatr Child Health. 2010;46(1-2):51–56 10.1111/j.1440-
2015 International Consensus of Cardiopulmonary Resuscitation 1754.2009.01617.x
and Emergency Cardiovascular Care Science with Treatment 35. Kelm M, Proquitté H, Schmalisch G, Roehr CC. Reliability of two
Recommendations. Pediatrics. 2015;132:S204–S241. doi:10.1161/. common PEEP-generating devices used in neonatal resuscitation.
24. Finer NN, Carlo WA, Walsh MC, et al; SUPPORT Study Group of Klin Padiatr. 2009;221(7):415–418 10.1055/s-0029-1233493
the Eunice Kennedy Shriver NICHD Neonatal Research Network. 36. Hartung JC, Wilitzki S, Thio-Lluch M, te Pas AB, Schmalisch G,
Early CPAP versus surfactant in extremely preterm infants. N Engl J Roehr CC. Reliability of single-use PEEP-valves attached to self-
Med. 2010;362(21):1970–1979 10.1056/NEJMoa0911783 inflating bags during manual ventilation of neonates: an in
vitro study. PLoS One. 2016;11(2):e0150224 10.1371/journal.
25. Morley CJ, Davis PG, Doyle LW, Brion LP, Hascoet JM, Carlin JB;
pone.0150224
COIN Trial Investigators. Nasal CPAP or intubation at birth for very
preterm infants. N Engl J Med. 2008;358(7):700–708 37. Zhu X-Y, Lin B-C, Zhang Q-S, Ye HM, Yu RJ. A prospective
evaluation of the efficacy of the laryngeal mask airway
26. Dunn MS, Kaempf J, de Klerk A, et al; Vermont Oxford Network during neonatal resuscitation. Resuscitation. 2011;82
DRM Study Group. Randomized trial comparing 3 approaches to (11):1405–1409 10.1016/j.resuscitation.2011.06.010
the initial respiratory management of preterm neonates. Pediatrics.
38. Yang C, Zhu X, Lin W, et al. Randomized, controlled trial comparing
2011;128(5):e1069–e1076 10.1542/peds.2010-3848
laryngeal mask versus endotracheal intubation during neonatal
27. Doyle LW, Carse E, Adams A-M, Ranganathan S, Opie G, Cheong resuscitation: a secondary publication. BMC Pediatr. 2016;16:17.
JLY; Victorian Infant Collaborative Study Group. Ventilation in doi:10.1186/s12887-016-0553-6.
extremely preterm infants and respiratory function at 8 years. N 39. Wanous AA, Wey A, Rudser KD, Roberts KD. Feasibility of laryngeal
Engl J Med. 2017;377(4):329–337 10.1056/NEJMoa1700827 mask airway device placement in neonates. Neonatology. 2017;111
28. Roehr CC, Kelm M, Fischer HS, Bührer C, Schmalisch G, Proquitté (3):222–227 10.1159/000450691
H. Manual ventilation devices in neonatal resuscitation: tidal 40. Reinhart DJ, Simmons G. Comparison of placement of the laryngeal
volume and positive pressure-provision. Resuscitation. 2010;81 mask airway with endotracheal tube by paramedics and respiratory
(2):202–205 10.1016/j.resuscitation.2009.10.008 therapists. Ann Emerg Med. 1994;24(2):260–263

e664 NeoReviews
Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017
Providing Ventilation to the Newborn Infant in the Delivery Room
Justin B. Josephsen and Marya L. Strand
NeoReviews 2017;18;e658
DOI: 10.1542/neo.18-11-e658

Updated Information & including high resolution figures, can be found at:
Services http://neoreviews.aappublications.org/content/18/11/e658
References This article cites 39 articles, 10 of which you can access for free at:
http://neoreviews.aappublications.org/content/18/11/e658#BIBL
Subspecialty Collections This article, along with others on similar topics, appears in the
following collection(s):
Pediatric Drug Labeling Update
http://classic.neoreviews.aappublications.org/cgi/collection/pediatric
_drug_labeling_update
Permissions & Licensing Information about reproducing this article in parts (figures, tables) or
in its entirety can be found online at:
http://classic.neoreviews.aappublications.org/site/misc/Permissions.x
html
Reprints Information about ordering reprints can be found online:
http://classic.neoreviews.aappublications.org/site/misc/reprints.xhtml

Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017


Providing Ventilation to the Newborn Infant in the Delivery Room
Justin B. Josephsen and Marya L. Strand
NeoReviews 2017;18;e658
DOI: 10.1542/neo.18-11-e658

The online version of this article, along with updated information and services, is
located on the World Wide Web at:
http://neoreviews.aappublications.org/content/18/11/e658

Neoreviews is the official journal of the American Academy of Pediatrics. A monthly publication,
it has been published continuously since . Neoreviews is owned, published, and trademarked by
the American Academy of Pediatrics, 141 Northwest Point Boulevard, Elk Grove Village, Illinois,
60007. Copyright © 2017 by the American Academy of Pediatrics. All rights reserved. Online
ISSN: 1526-9906.

Downloaded from http://neoreviews.aappublications.org/ by guest on November 8, 2017

You might also like