You are on page 1of 7

Acta vet. scand. 1991,32,123-129.

Comparison of Different pH Measurement


Methods in Meat
By O. Mdki-Petiiys, H. Korkeala, T. Alanko and O. Sorvettula

Veterinary Department, Ministry of Agriculture and Forestry, Vuorikatu , and


Department of Food and Environmental Hygiene, College of Veterinary Medicine, Helsinki, and
Technical Research Centre of Finland , Computing Service, Espoo, Finland.

Miiki-Petiys, 0., H. Korkeala, T. Alanko and O. Sorvettula: Comparison of dif-


ferent pH measurement methods in meat. Acta vet. scand. 1991, 32, 123-129. -
The accuracy of portable pH meters and the nitrazine yellow method was com-
pared with the reference method by determining the pH of 74 beef and 96 pork
muscles. The pH was measured directly from the muscle. The results showed sta-
tistically significant differences (p < 0.001) between the different electrometric
combinations. Combinations of portable pH meters with puncture electrodes gave
systematically higher pH values than the reference method. These differences were
not very large but they may be of practical significance. The use of a piercing cover
on the electrode to help the insertion of the electrode into the meat is not recom-
mended, since it may cause a rise in pH values. Electrometric methods were found
to be more precise than the nitrazine yellow method . On the basis of these findings
there still is a need of further harmonization of the methods used for pH measure-
ment of meat.

meat quality ; pH determ ination ; electrometric methods ; portable pH


meters ; nit razine ye ll ow method.

Introduction There are several works in which the au-


The pH measurement of meat is widely used thors have compared the precision of the
to assess the shelf life and quality of the car- different methods used in the determination
cass in meat inspection and in the meat in- of the pH of meat (Prost 1955, Hofmann
dustry for determining the different uses of 1968, Bager & Petersen 1983, Dran sfield et
the carcass. Dark cutting beef poses a speci- al. 1983, Korkeala et al. 1986). Korkeala et
fic problem . The high ultimate pH of DFD al. (1986) found that the differences between
meat boosts bacterial growth and shortens the different electrodes used appeared to be
storage life considerably. The determination greater than those due to the treatment of
of pH is essential in order to avoid the the meat samples. They used the direct
packaging of DFD meat and keep it off the puncture method, meat-water mixtures and
fresh-meat market. This means rapid , con- muscle homogenates . Prost (1955) and van
secutive and accurate measurements on the Gi/s & van Logtestijn (1965) considered elec-
killing line. trometric determinations better than the
The pH value of meat is also used in the in- various colorimetric and indicator methods.
ternational meat trade (Anon. 1982a). The On the other hand Honikel & Fischer (1977)
increase and liberation of the meat trade showed a very close relationship between
also implies harmonization and rehabilita- pH values obtained with special indicator
tion of the methods used. test strips and with glass electrode . There is

Acta vel. scand. vol. 32 no. I - 199 1


124 O. Mdki-Petdys et at.

still a need for more reliable evaluation of Knick 651 Portamess pH meter with
the different methods. combined glass electrode Ingold 404-T
The purpose of the present work is to eva- with piercing cover, referred to below as
luate the reliability of methods used in prac- "method A".
tice, since the pH measurement of meat be- WTW pH 90 pH meter (Wissenschaft-
fore cutting and packaging is becoming more lich-Techn ische Werkstiitten GmbH, Fe-
and more important. Two portable pH me- deral Republic of Germany) with com-
ters with puncture electrodes, which can be bined glass electrode Ingold LoT
used on the killing line and which are used 406-M3, referred to below as "method
in Finland, were therefore compared to the B".
combination Korkeala et at. (1986) found to
give the average picture of the pH among the The pH of the samples was also measured
7 electrometric methods they used. In addi- with an aqueous solution of the nitrazine
tion, the colour indicator nitrazine yellow yellow indicator (0.0 I %) test (Anon. 1955,
method was compared with the electrome- Schonberg & Zietzschmann 1958).
tric determinations. The study used beef and Each pH measurement was read and recor-
pork carcasses with low and high pH. ded to the nearest 0,0 I pH units electro-
metrically and to 0, I pH units with nitra-
Materials and methods. zine yellow. The meat used for the measure-
Sampling ment was free of visible fat and connective
Thirty-seven beef muscle samples from tissue.
M . triceps brachii caput longum (MT) and The samples were measured chilled. The
M. adductor (MA) and 48 pork muscle measurements were carried out in the labo-
samples from MT and MA were taken about ratory (22"C) at least twice with all electro-
65 h (range 24-180 h) after slaughter . The des and once with nitrazine yellow. The
pH of the samples was measured immedi- electrodes were cleaned (Anon. 1974) after
ately after sampling . each measurement and the pH meter cali-
bration was checked according to the in-
Mea surement of the pH structions of the manufacturer with buffer
The pH of the samples was measured direct- solutions with pH 4 and 7 (Dr. W. Ingold
ly from the muscle with penetration electro- AG) at regular intervals . The temperature of
des (Anon. 1974, Anon. 1982). the buffer solutions was SOc.
The Knick 742 Microprocessor pH meter
(Knick Elektronische Messgerate Gmbh & Statistical methods
Co, Berlin) with a Knick 6929 Thermocom- The ordinary pH scale, rather than the hy-
pensator probe and with a combined glass drogen ion concentration scale (Hs-scale)
electrode Ingold 404-T (Dr. Ingold AG, Zu- sometimes recommended in the literature
rich, Switzerland) was used as the reference (cf. Murphy 1982), was used as in Korkeala
method. The ability of this method to deter- et al. (1986).
mine pH is described by Korkeala et at. The arithmetric mean of the 2 replicate pH-
(1986). values was used as the basic unit for all
The other methods used were combinations other calculations except in the case of nitra-
of portable pH meters and electrodes, as fol- zine yellow, where only I measurement was
lows: available.

Acta vel. scand. vol. 32 no. \ - 199 \


pH measurement methods in meat 125

Table I . Means and standard deviations of pH values of muscle samples measured


with different methods.
Methoda
Mu scle
samples N Re ference A B Nitraz .

All 168 5.84 ± 0.40 6.11 ± 0.38 6.03 ± 0.37


Beef 74 5.74 ± 0.39 6.00 ± 0.35 5.94 ± 0.36
Pork 94 5.92 ± 0.39 6.20 ± 0.39 6.10 ± 0.36
MTb 84 5.93 ± 0.41 6.22 ± 0.38 6.11 ± 0.38
MAC 84 5.74 ± 0.37 6.01 ± 0.35 5.94 ± 0.34
pH < 6.00 118 5.62 ± 0.19 5.91 ± 0.18 5.82 ± 0.16
ph > 6.00 50 6.35 ± 0.29 6.59 ± 0.28 6.50 ± 0.28 6.5 ± 0.2
a See Materials and Methods.
b MT = Musculus triceps brachii caput longum.
c MA = Musculus adductor.

Statistical com parison of the 3 pH -measure- of the F-tests for the equality of means are
ment methods was performed using repe - shown in Table 2. Results showing statisti-
ated-measures-model analysis of variance cally significant ditTerences between the refe-
(Winer 1971). The ditTerences between the rence method and other methods are also
reference method and methods A and B presented in Table 2.
were tested by pre -defined contrasts in the
repeated meas ures analysis of variance. An Ta b Ie 2 . Significance of differences between pH
measurement methods and the reference method.
ad hoc cut -otT ru le was used to divide the
The F-values refer to pre-defined contrasts in re-
samples into 2 groups - the " high" pH- peated measures analysisof variance.
group and the "low" pH-group - as follows:
If the average over the 3 pH-measurement Methoda F-value

methods of a muscle sample was above 6.0 p < 0.0001


Method A 365.35
it was included in the group " pH above Method B 346.17 p < 0.0001
6.00", if the average was lower than or equal Nitrazine
to 6.00 it was included in the group "pH yellowmethod 8.52 p < 0.01
6.00 or below". Analysis of variance was a See Materialsand Methods.
then performed for the second group sepa-
rately, with nitrazine yellow as a fourth me- Table 3 . Classification of samples into pH clas-
thod incl uded in the "pH above 6.0" samp- ses by different methods(in per cent).
les. pH
All calculations were performed using the
Method'' > 6.00 > 6.20 > 6.50
SAS statistical package on a microcomputer.
Reference method 27.4 20.2 10.7
Results Method A 50.6 32.7 16.7
The means and standard deviations of the Method B 40.5 27.4 14.9
pH-measurement methods for the 4 types of Nitrazine
yellow method 39.9 28.6 13.1
methods and the ditTerent muscle sample
groups are presented in Table I . The results a See Materials and Methods.

Acta vet. scand . vol. 32 no.1 - 1991


126 O. Mdki- Petdys et al.

0.8


0.6
••
.. .... . .
P
•••• •
H
•• .. ,: • • •

d
I
0.4
.... ....
.. ·r·. ,
.... .-

. . . . ...
'It
'\1: . . . •
••
,. • •


• •
••
••
... .-....
..·a· . '. •\ • • • •
I 'It.
• • • •
... . .
I
e 0.2 •• • ••
r • •
e • • • • • •
n
0.0 •• • • • •
c
• •
e • • • •

- 02

- 0.4
5.2 5.4 5.6 5.8 8.0 8.2 6.4 6.6 6.8 7.0
pH measured by the reference method

Fi gure 1. Scatter diagram of pH difference between method A and the reference method .

0.8

0.6
P
H

..
• •

, .............' ...... .. ..
d 0."
• • • •
,..
I : •••,e. •• • • •
I • • • •
I
,,,. ,:• ••• • .•' .•


,
e 02
r • • e. •
•• •• • • •
•• •• •
• •• • .
e
n
0.0
•• • • •• • • •
c
• • •
e • •
-02


- 0.4 •
52 5.4 5.6 5.8 6.0 62 6.4 6.6 6.8 7.0
pH measured by the reference method

F igu re 2. Scatter diagram of pH difference between method B and the reference method .

Acta vet. scand. vol. 32 no. 1 - 1991


pH measurement methods in meat 127

Table 3 and Figures I and 2 illustrate the vantageous. Comparing method B and refe-
magnitude and nature of these differences . In rence method, we find that method B too
Table 3 the percentage of muscle samples gives higher pH values than the reference
classified into particular pH-groups by a method (Fig. 2). The mean of the difference
method is reported. In Fig. I the difference was 0.19. We also find that the difference
between method A and the reference method does not depend on the pH of the sample.
and in Fig. 2 the difference between the refe- Differences like these may have a practical
rence method and method B is plotted meaning.
against the pH-level of the sample measured Errors in the pH determination of meat may
by the reference method. Figs. I and 2 show also be harmful in meat inspection. The esti-
both that the differences are rather systema- mation of the pH of meat is of value in the
tically positi ve and that they do not vary judgement of borderline cases, particularly
markedly along the pH-scale. of emergency-slaughtered animals, since it
The correlation coefficient between nitra- indicates whether or not the meat will pos-
zine yellow and the reference method was sess adequate durability (Gracey 1986). In
0.66. addition the pH determination is important
in the interpretation of the boiling test. Kor
Discussion keala et al. (1988) found that in beef the
All the F-tests given in Table 2 reveal that odour scores remain steady for samples with
there are statistically significant differences a pH value under 6,2 and start to increase
between the reference method and the other rapidly in higher values .
pH-measurements methods. In chosing carcasses for different uses on the
The distribution of the samples into diffe- basis of the pH value of meat, are the mea-
rent pH classes seems to depend on the me- surements to be made on the killing line or
thod used (Table 3). In the case of method A in the chilling room. Some authors have
the proportion of samples with high pH va- stressed that the most accurate results are
lues are greater than by the other methods. obtained by the electrometric method using
Comparing method A and the reference minced meat, juice or homogenates (H of
method, which gave the highest and lowest mann 1988). Korkeala et al. (\ 986) reported
mean of the pH (Table I), we find that me- the differences between electrodes to be
thod A gives systematically higher pH va- greater than the differences due to the pre-
lues than the reference method (Fig. I). The sentation of the meat samples. However,
maximum value of the difference between only direct puncture measurements are use-
method A and the reference method was ful on the killing line, where it is necessary
0.65 and the mean of the difference was to make a large number of measurements in
0.27. This difference could be due to the rapid sequence. Portable pH meters are the
piercing cover used on the electrode. Even if only way to measure the pH of meat electro-
the 3 diaphragms of the electrode were in metrically on the killing line or in the chil-
free contact with the meat , the cover might ling room.
diminish the strength of the contact. Al- Although nitrazine yellow has a high corre-
though with puncture measurements there is lation with the reference method (0.66) it is
always the risk of breaking the electrode not suitable for common use in meat tech-
shaft on insertion, the use of a piercing cover nology . Van Gi/s & van Logtestijn (1967)
on the electrode nevertheless seems disad- also stressed that nitrazine yellow is not

Acta vel. scand, vol. 32 no. I - 1991


128 O. Mdki-petdys et al .

suitable for scientific investigation. The me- Muscles. Nord. Vet.-Med. 1983, 35, 86-
thod may, however, in some cases be useful 90.
in field conditions where electrometrical Dransfield E, Casey JC, Boccard R, Touraille C,
methods are not available. Buchter I, Hood DE, Joseph RL, Schon I, Ca-
Our results show that the combinations of steels M, Cosentino E, Tinbergen BJ: Compa-
rison of Chemical Composition of Meat Deter-
different electrodes and pH meters by which
mined at Eight Laboratories. Meat Sci. 1983,
the measurement is carried out behave dif-
8,79-92.
ferently. However, we cannot say with cer- van Gi/s J HJ , van Logtestijn JG: Die pH-Mes-
tainty which of the methods gives the best sung des Fleisches-Hilfsmethode zur Beurteil-
picture of the real pH of the meat (Korkeala ung des Fleisches von Schlachttieren. (Mea-
et al . 1986). It therefore seems important to surement of pH in meat - and aid in the
attempt further research of the reliability of judgement of meat of slaughtered animals).
methods usable in practice. The use of a Arch. Lebensmittelhyg. 1967,18,49-56.
piercing cover on the electrode should be Gracey JF: Meat Hygiene. Bailliere Tindall, Lon-
avoided. When the pH value is used in the don 1986, 238.
Hofmann K: Untersuchung der Zuverlassigkeit
evaluation of meat quality, the recommen-
der pH-Wert-Messung an Fleisch mit Hilfe
dations for the pH determination should be
der Glaselektrode. (Investigations into the reli-
included, due to great differences in results ability of pH-value measurement in meat
obtained by different combinations of elec- using a glass electrode). F1eischwirtsch. 1968,
trodes and pH meters. 48, 317-322.
Hofmann K: pH, a quality criterion for meat.
References F1eischwirtsch. 1988,68,67-70.
Anonymous: Join FAOIWHO expert committee Honikel KO, Fischer C: Vergleichende Messung
on meat hygiene. WId. Hlth. Org. tech. Rep. des pH-Werts mit Glaselektrode und spezial-
Ser. No. 99, World Health Organisation, Ge- indikatorstabchen bei Rindfleisch. (Compara-
neva 1955. tive pH measurements in muscles of beef using
Anonymous: Meat and meat products. Measure- a glass electrode and special indicator test
ment of pH (reference method) ISO 2917, strips). Fleischwirtsch. 1977,5 7,2175-2177.
International Organisation for Standardisation, Korkeala H, Mdki-Petdys 0 , Alanko T, Sorvet-
Geneva 1974. tula 0 : Determination of pH in meat. Meat
A nonymous: Mindestanforderungen an Schlacht-, Sci. 1986,18,121-132.
Zerlege- und Verarbeitungsbetriebe sowie Korkeala H, Alanko T, Mdki-Petdys 0 , Sorvet-
Klihlhauser, die Fleisch oder Fleischwaren tula 0 : The effect of the pH of meat on the
nach der Schweiz exportieren. (Minimum re- boiling test. Acta vet. scand. 1988, 29, 15-22.
quirements for the approval of slaughterhou- Murphy MR : Analyzing and Presenting pH Data.
ses, cutting and processing plants, and cold J. Dairy Sci. 1982,65, 161-163.
stores which export meat or meat products to Prost E: Studies on Determinations of pH of
Switzerland). Bundesamt filr Veterinarwesen meat. Medycyna Wet. 1955,11,142-147.
(Federal Veterinary Office), Liebefeld 1982a. Schonberg F, Zietzschman MO: Die Ausfilhrung
An onymou s: Lihan pH:n mittaaminen. (Determi- der tierarztlichen Fleischuntersuchung. (The
nation of pH in meat). Circular No. 201, Vete- Practice of Veterinary Meat Inspection). Ver-
rinary Department, Ministry of Agriculture lag Paul Parey, Berlin 1958, 160.
and Forestry, Helsinki 1982b. Winer BJ: Statistical principles in experimental
Bager F, Petersen GV: The Relative Precision of design. 2nd edn. McGraw-Hili, New York
Different Methods of Measuring pH in Carcass 1971 ,261-308.

Acta vet. scand. vol. 32 no. \ - 199\


pH measurement methods in meat 129

Sammanfattning tiskt hogre pfl-varden an referensmetoden. Dessa


Jiimfbrelse mel/an olika pH miitn ingsmet oder skillnader var inte stora men de kan ha praktisk
a v kott. betydelse. Anvandningen av penetrerande hylsa
Noggrannheten av portabla pH matare och nitra- elektroden fbr att underlatta inforandet av
zingult jamfordes med en referensmetod genom att elektroden i kottet ar inte att rekommendera,
bestarnma pl-l-vardet av muskelprover 74 not emedan den fororsakar en forhojning av prl-var-
och 96 svin. pl-l-vardet mattes direkt fcln mu- det. De elektrometriska metodema befanns vara
skeIn. Resultatet visar statistiskt markbara skill- exaktare an nitrazingult. Resultaten av undersok-
nader (p < 0.001) mellan de olika elektrometriska ningen utvisar att det fortfarande finns ett behov
kombinationema. Kombinationer av portabla pH att harmoniera metodema som anvands for att
rnatare med penetrerande elektroder gay systema- mata kottets pH .

(R eceived February 7, 1990. accepted February 9, 1990).

Reprints may be requested from: H. Korkeala, College of Veterinary Medicine,


Department of Food and Environmental Hygiene, P. O . Box 6,
SF-00551 Helsinki, Finland.

Acta vet. scand. vol. 32 no. 1 • 1991

You might also like