Pressure Relief Valve Removal

NOTICE: TV-100 model configuration may vary. Contact Bio-Med Devices, Inc technical
support if your model differs from the configuration below for proper service and
maintenance.

EQUIPMENT REQUIRED: 2mm hex


TESTING REQUIRED: N/A
Follow rear and front case removal procedures (Refer to 3.1 � 3.2)
Remove 4 flat head screws securing the Pressure Relief Plate using 2mm hex wrench.
Relief valves will now be accessible. Make note of the valve locations for later
installation.
Top: Vacuum Relief Valve (PRT5532)
Middle: Pressure Relief Valve (PRT5538)
Bottom: Two Way Relief Valve (PRT5537)
Remove relief valves.