Professional Documents
Culture Documents
GS200 Parameter Settings 20210415 V4.0
GS200 Parameter Settings 20210415 V4.0
Reaction Trend: Includes ascending (ASC) and descending (DESC). Only 6 items (ALT, AST, Urea,
TBIL-vanadate, DBIL- vanadate and CO2 ) are descending reactions among the 20 common clinical test
items, while the rest are ascending reactions.
Sample/Reagent Volume: According to experience, the recommended test volume (Sample and
Reagent) is listed below. The sample / reagent volume can be reduced or increased with the same
proportion of volume mention in the reagent manual. Please keep the total reaction solution volume
within the range of 160ul-400ul.
Sample 2-48ul
Reagent 1 180-350ul
Reagent 2 20-200ul
Reaction time: For a single reagent item, reaction time starts from the time after the sample is added
(Reaction starts) to the end of reaction. For a two-reagent item, reaction time starts from the time
after R2 is added (Reaction starts) to the end of reaction.
End 1 End 2
End 1 = 10 + (Reaction time/ 36) (Round it up to the next higher integer if the result is not an integer)
End 2 = End 1 + 1/ 2/ 3
Example: In the reagent manual, reaction time is 200s. Time converts into cycle: 200s/ 36=5.5, round
it up to 6.
End 2 = End 1 + 1/ 2/ 3= 17 / 18 / 19
Reaction cycle is set 38 〜 40 or 16 〜 18 (Or set to 16-17, in this case it will get the
mean value of point 16 and 17 as the end point value)
1.2 Two-reagent:
Blank Cycle = 16 〜 18
Reaction Cycle = 38 〜 40
or
End 1 = 19 + (Reaction time/ 36) (Round it up to the next higher integer if the result is not an integer)
End 2 = End 1 + 1/ 2/ 3
Example: In the reagent manual, reaction time is 200s. Time converts into cycle: 200s/ 36=5.
End 2 = End 1 + 1/ 2/ 3= 25 / 26 / 27
Reaction cycle is set 38 〜 40 or 24 〜 26 (Or set to 24-25, in this case it will get the mean
value of point 24 and 25 as the end point value)
2 Kinetic:
Example: In the reagent manual, delay time is 120s, monitoring time is 200s. Time converts into cycle: 120/
36=3.3, round it up to 4. 200s/ 36=5.5, round it down to 5.
2.2 Two-reagent:
Blank Cycle = 0 〜 0 leave it zero, no need to set the blank cycle
Reaction Cycle:
Begin point = 19 + (Delay time/ 36)
End point = 19 + (Delay time/ 36)+(Monitoring time/ 36)
(Round ‘Delay time’ up to the next higher integer if the result is not an integer, round ‘Monitoring time’ down
to the pre lower integer if the result is not an integer)
Example: In the reagent manual, delay time is 90s, monitoring time is 180s. Time converts into cycle: 90/
36=2.5, round it up to 3. 180s/ 36=5.
Example: In the reagent manual, delay time is 180s, monitoring time is 200s. Time converts into cycle: 180/
36=5. 200s/ 36=5.5, round it down to 5.
3.2 Two-reagent:
Blank Cycle = 0 〜 0 leave it zero, no need to set the blank cycle
Reaction Cycle:
Begin point = 10 + (Delay time/ 36)
End point = 10 + (Delay time/ 36) + (Monitoring time/ 36)
(Round ‘Delay time’ up to the next higher integer if the result is not an integer, round ‘Monitoring time’ down
to the pre lower integer if the result is not an integer)
Example: In the reagent manual, delay time is 120s, monitoring time is 180s. Time converts into cycle: 120/
36=3.3, round it up to 4. 180s/ 36=5.
Appendix