TomoTherapy’s
1,400 m2
Middleton facility. www.tomotherapy.com
Outline
Varian Elekta
Sequential Tomotherapy
The First Form of IMRT
X-Ray Beam
Simulated
Annealing
Binary Multileaf Courtesy: NOMOS
Optimization
Binary MLC Leaves Collimator Corporation
Individual leaf
controls opening
MIMiC
Multileaf
Intensity
Modulating
Collimator
From Bruce Curran, NOMOS
68 Gy
60 Gy
50 Gy
30 Gy
Bi-Lateral 3D-CRT
From Bruce Curran, NOMOS
85Gy
70Gy
60Gy
50Gy
30Gy
100
80
Volume (%)
GTV_Corvus
60
GTV_Peacock
GTV_3DCRT
40 GTV_Bilateral
20
0
0 20 40 60 80 100
Dose (Gy)
Chiasm_Corvus
100
Chiasm_Peacock
Chiasm_3DCRT
80 Chiasm_Bilateral
Volume (%)
60
40
20
0
0 20 40 60 80 100 120
Dose (Gy)
Stem_Corvus
100 Stem_Peacock
Stem_3DCRT
80 Stem_Bilateral
Volume (%)
60
40
20
0
0 20 40 60 80 100
Dose (Gy)
100 Parotid_Peacock
Parotid _3DCRT
Parotid_Bilateral
80
Volume (%)
60
40
20
0
0 20 40 60 80 100
Dose (Gy)
In this case
the primary
goal was to
avoid the
parotid and
spinal cord.
With the same
amount of
avoidance,
the 45 angle
delivery
(tomotherapy)
provided a
more
homogeneous
delivery.
Re-Engineering Radiotherapy
CT Couch
Helical
Fan Beam
Helical
Scanning
Animation of Helical Delivery
64 Leaf Binary MLC
Close-Up
Movie Clips of the MLC Being Tested
Close-Up
UW Clinical Helical Tomotherapy Unit
Siemens
RF
Siemens System
Linac
GE CT GE
Detector Gantry
40 cm
Profile Along Length of a
1 cm Wide Fan Beam
Transverse Profile
120
100
80
Dose (%)
60
40
20
0
-240 -180 -120 -60 0 60 120 180 240
Distance (m m )
Monte Carlo Model of the Treatment Head
Target
Electron Stopper
Electron Beam
from the Accelerator Beam Hardener
Monitor Chamber
Primary
Collimator
and Jaws
Photon Spectrum
1.0
Tomotherapy (6MV)
2100 Clinac (6MV, Stanford)
0.8
Spectrum [%]
0.6
0.4
0.2
0.040
0-5 cm
0.035 5-10 cm
0-5 cm 10-15 cm
0.030 5-10 cm 15-20 cm
0.01
10-15 cm
Photon spectrum
Photon spectrum
0.025 15-20 cm
0.020
0.015
1E-3
0.010
0.005
0.000
0 1 2 3 4 5 6 7 0 1 2 3 4 5 6
Tomotherapy Conventional RT
Designed for Image-Guided Designed for Conventional
IMRT RT Processes
Single Photon Energy Dual Mode Multiple Energy
98 % 70 % 50 % 40 % 20 %
Tumor
Spinal Cord
Tomotherapy Plan to Treat Lung Cancer
80 % 70 % 50 % 40 % 20 %
Tumor
Spinal Cord
Tomotherapy Decreases
Mean Dose to Spinal Cord
25
25
Conventional 3D
3D plans
Tomotherapy
20 20
Tomotherapy
Dose (Gy)
15
10
5 5
0 0
P1 P2 P3 P4 P5 Mean
1 2 3 Patient 4 5 Mean
Case
Prostate Carcinoma
Dose Rate Cumulative Dose
25
20
% of volume
exceeding 15
stated dose >60 Gy
10 >70 Gy
0
Conventional 3D Tomotherapy
Conventional 3D Tomotherapy
IMC nodes
Breast Carcinoma Case: Contr.
breast Tumor
ROI slice 47
Isodose Lines
100%
90%
80%
50%
Targets
GTV
Hypoxic
Proliferative
Tomotherapy for GBM Treatments
Dose Rate Cumulative Dose
0 to 50 % 50 to 75% 75 to 85 % 85 to 95 % 95 to 100%
Mesothelioma
ROI slice 27
MV CT Delivery
+ Image Fusion Modification
Why CT Before Delivery is Necessary
CT Images
Acquired Daily
Original CTV
Courtesy Di Yan
and Marcel Van Herk
Interfraction Movement of the Breast
Electronic
portal image
of the breast
treated using a
compensator.
Each frame is a
different day.
0.2% 4.3%
-2.4% -54.1%
5.2%
-4.8%
11.7%
-81.0%
51.2%
1.50 mm
1.25 mm
1.00 mm
0.8 mm
MVCT Calibration Curve
Image
Density
Head Pelvis
Rando Phantom MVCT
Slice-by-Slice Traversal Through Volume
MV CT Delivery
+ Image Fusion Modification
Delivery Modification
• There are two potential ways to adjust
the patient setup:
– Adjust the patient.
– Adjust the beams.
• It may be easier and more reliable to
move the beams to the patient.
• It may be possible in tomotherapy to
alter the leaf delivery pattern.
Optimized
50 %
80 %
95 %
Delivery
Modified
50 %
80 %
95 %
1.42 cm offset
1cm x-offset
1cm y-offset
3-D Imaging Adaptive Radiotherapy
Deformable Dose
Dose Reconstruction
Registration
Helical
Optimized Tomotherapy
Planning
MV CT Delivery
+ Image Fusion Modification
Dose Reconstruction
L. eye Target
R. eye
Brain
The reconstructed
dose reveals that the
dose distribution is
rotated toward the left
eye and underdosing
the target volume.
Optimized
Reconstructed
3-D Imaging Adaptive Radiotherapy
Deformable Dose
Dose Reconstruction
Registration
Helical
Optimized Tomotherapy
Planning
MV CT Delivery
+ Image Fusion Modification
Deformable Dose Registration
MV CT Delivery
+ Image Fusion Modification
Image of Regret Adapted Dose Distribution
0 to 3 mm(%) 3 to 6 mm(%)
6 to 9 mm(%) 9 to 12 mm(%) Dose to deliver to correct regret.
Incorrect +Perfect Incorrect +Corrected
Non - Adapted Adapted