GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 1 With Xtream
CT Scanner System

Cardiac Function, SmartScore Pro and


Advantage Sim.
• Compact system design of the
16
LightSpeed allows for installation in only
28m² (302 square feet).
• Productivity features designed for the CT
Technologist: In-Room Start, Remote
Gantry Tilt, Breathing Lights with countdown
timer, Gantry Controls mounted on all four
corners of the gantry and an Integrated IV
Pole at the foot of the table.
• 1700 mm scannable range for full body
scans.
• 254 GB Disk (system, image, scan disks)
2
stores up to 250,000 512 images and 2881
scan data files.
• Cardiac acquisition with Segment, Burst
and Burst Plus enable you to scan patients
with heart rates from 40 to 110 BPM, and
provide for temporal resolution up to 65
ms.
• Color Coding for Kids, winner of a
Introduction TM
• microVoxel imaging improves 3D and National Heroes Award from the Emergency
reformatted 2D resolution through the Medical Services for Children, provides
16
The LightSpeed CT Scanner with optimum choice of sub-millimeter slice pediatric scan protocols based on the
TM TM
Xtream technology is the next step in thickness and reconstructed voxel size. Broselow-Luten Pediatric System. This
GE’s CT Continuum and its award- Color Coding system is incorporated into the
• Unmatched Volume Image Quality
winning LightSpeed CT platform. The protocol selection on the operator’s console
16 specification in all planes (sagittal, coronal
LightSpeed CT Scanner routinely and is designed to facilitate pediatric
and axial): 15.4 lp/cm @ 0% MTF @ 0.5
acquires 16 sub-millimeter slices in one emergency care and reduce medical errors.
sec. Rotation
rotation, at a scan speed of 0.5 seconds. TM
TM • VariViewer : provides an interactive axial
As with the very first LightSpeed product • HiLight Matrix II Detector is designed to
16 review mode that can change the slice
introduction, the LightSpeed utilizes deliver consistent image quality with its
TM thickness reconstruction instantaneously,
Designed for Six Sigma (DFSS ) 21,888 individual elements: 16 rows of
resulting in the ability to “scan thin and view
methodology in all aspects of 0.625mm thickness and 8 rows of 1.25mm
thick.”
manufacturing and engineering. thickness
TM
16
LightSpeed is available both as a new • Xtream technology, a revolutionary new
system or as a cost-effective upgrade to workflow design, breaks through existing Clinical Performance
an existing LightSpeed system. limits on speed, image quality and flexibility
to provide an optimized workflow solution Chest / Abdomen / Pelvis
from acquisition to final report.
TM
• Lowest dose acquisition using SmartmA, Coverage 600 mm
an automatic modulation technique and DLP Rotation 0.5 sec.
Primary Benefits display that tells the operator the dose Mode 16 x 0.625 mm
• Routine use of sub-millimeter slices before the scan starts. Pitch 1.75:1
without image noise or coverage TM mA 420
• HyperPlane reconstruction algorithm
compromise. mAs 210
tackles the inherent issues with fast, high
• Small lesion and small vessels pitch helical scanning by reducing helical Speed 35 mm/s
assessments – pancreas, liver or circle artifacts and dose, and optimizing the slice Scan Time 17.1 seconds
of willis, renal arteries, coronary arteries profile.
and peripheral vascular arteries. TM
• CrossBeam reconstruction algorithm
• The 6.3 MHU Performix X-Ray Tube solves the technical challenge of cone beam
demonstrates the highest reliability and artifacts.
delivers the power you need for image
• Large breadth of Advanced Software
quality. The Performix Tube now comes
Applications: Advanced Lung Analysis, CT
standard with a one-year, non pro-rated
Colonography, Advanced Vessel Analysis,
warranty.
Brain and Body Perfusion, Cardiac Analysis,

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 2 With Xtream
CT Scanner System
16
• Prospective or retrospective image • SmartPrep, standard on LightSpeed ,
Peripheral Run-Off reconstruction of 0.625 mm, 1.25 mm, provides software for real-time monitoring of
2.5 mm, 3.75 mm, 5 mm, 7.5 mm or 10 contrast enhancement.
mm images from inherently thinner
Coverage 1,400 mm • * Direct3D, makes 3D routine and instant
acquisition data sets;
Rotation 0.5 sec. by building the 3D model during axial image
Mode 16 x 1.25 mm • For both helical and axial acquisitions. reconstruction, reducing the need for user
Pitch 1.75:1 • Image decomposition to: interaction.
mA 380 • Retrospective reconstruction of thin • * SmartScorePro, provides EKG-gated
mAs 190 images from data sets where thicker hardware and software on the Advantage
Speed 70 mm/s images were initially reconstructed; Workstation for coronary artery calcium
Scan Time 20 seconds scoring.
• Facilitates more detailed image analysis
without need for re-scans;
• * CardiacSnapShot, provides software
• Improves 3D visualization; and hardware to perform EKG-gated
High Resolution Chest • For both helical and axial acquisitions reconstructions of the heart in SnapShot
Segment mode.
• Up to 2,881 multi-slice 1-sec scan rotations
Coverage 200 mm stored on disk. Facilitates retrospective - SnapShot Segment is a single sector
Rotation 0.5 sec. image decomposition during off hours. protocol using information from one heart
Mode 16 x 0.625 mm cycle to generate an image with temporal
• Eight fundamental scan modes to simplify resolution of 250ms.
Pitch 1.75:1
the inherently complex nature of multi-slice
mA 440 - SnapShot Burst is a multi-sector protocol
helical scanning.
mAs 220 using up to two sectors from two different
Speed 35 mm/s • Standard set of clinically- proven heart cycles to produce an image with
Scan Time 5.7 seconds protocols derived from leading luminary temporal resolution of 125ms.
sites around the world. Up to 8,460 - SnapShot Burst Plus is a multi-sector
The net result is that in many cases, helical protocols can be edited, modified and stored protocol using up to four sectors of data
16
scans on the Lightspeed are up to 12 times on the system. from four different heart cycles to produce
faster than 4-slice CT systems. With the • Remote Gantry Tilt from the Operator’s images with temporal resolution to 65ms.
16
Lightspeed , users can routinely use a 0.5 Console to increase exam speed, including • CTDIw, DLP (Dose Length Product), and
second scan speed and 0.5625:1, 0.9375:1, built-in safety features to prevent accidental
1.375:1, and 1.75:1 helical pitches. This Dose Efficiency display during scan
contact of the gantry with the patient. prescription provides patient dose
added performance, at the same image
quality, may allow you to: reduce contrast • Built-in patient breathing lights and digital informatbon to the operator.
media, perform better thin-slice CT counter provides a goal-oriented approach
to coach the patients in holding their breath
angiography exams, use thinner slices for
during an exam.
Interactive CT Technology
most exams, and perform longer helical exams
without tube cooling delays. • SmartStart gantry-mounted start scan
16
The LightSpeed CT Scanner supplies
16 button and countdown display, facilitates exceptional computer and image processing
The LightSpeed delivers this level of power that significantly enhances clinical
single-technologist operation by allowing
performance through a balanced design productivity, building upon the strength of the
start of scan at the gantry, with a visual
that focuses on image quality, coverage, LightSpeed family - true interactive CT
reminder of time until X-ray initiation.
and exam speed. technology.
• SmartHelical GE proprietary, non-linear
interpolation algorithms, balance slice Interactive CT embodies a variety of design
Main Features profile, helical pitch, image noise and choices all striving to enhance operator and
• Full 360° rotation speed of 0.5 sec. provides required technique. department productivity. A truly interactive CT
for improved temporal resolution and image • SmartTrack advanced hardware and system will:
quality in areas such as cardiac imaging; software for X-ray beam tracking minimizes • Provide a user interface beyond “intuitive”
• Faster scan times enable shorter breath patient dose. to become purely “natural” - from the
holds and more comfortable exams for • SmartTools software automates every screens to the console hardware itself;
patients; exam task in order to increase throughput. • Allow two users to review cases side-by-
• Faster scan times reduce the occurrence of TM
• SmartBeam hardware and software side, with minimal interference;
re-scans by lowering the possibility of optimizes X-ray beam filtration • Supply a truly multi-tasking environment
patient motion during the exam; independently for body and head where even advanced image processing
• Routine scanning with a slice thickness of applications. can take place quickly and simultaneously
0.625mm - optimizes lesion detection and • HiRes Chest software provides a single- with other processes underway;
facilitates the use of thinner images for slice, high-resolution scan mode that • Operate with a very high degree of
volume presentations. minimizes dose for this critical application. automation, yet allow patient-specific
• Reduced partial volume artifacts:

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 3 With Xtream
CT Scanner System

changes to be easily made, with virtually no • Enhanced multi-tasking allows operators to • Operator console convenient to locate in
restrictions; review more than one exam simultaneously, suite.
independently - even with AutoView and • Computer, image processor and image
• Be as self-teaching as possible, and have
AutoFilm on/ reconstruction hardware completely
on-line access to tutorials as needed.
• BrightBox™ dedicated controls for integrated in base of console - no
One key element of this design is to combine image next, prior, manual paging and separate computer cabinet to site;
some of the best features from several trackball W/L helps make two person • Split table top allows unrestricted patient
product families into one state-of-the art CT image review practical; viewing while still supporting 2 large color
16
system. For example, the LightSpeed CT • Up to four 512x512 images from four monitors;
Scanner combines: different exams can be viewed on a large • Front and back work surfaces can be set
• SmartTools productivity software to 1024 color display. during installation within a range of
automate every step of the examination, • SmartmA™ User Interface - automatically vertical heights that help accommodate a
critical for ensuring the highest productivity optimizes mA to maintain constant image variety of siting requirements - especially
and throughput possible with the noise when collimation/detector surrounding the height of the console
16
LightSpeed CT Scanner; configuration, scan mode, scan rotation relative to the window into the scanning
speed, table speed, or image thickness suite.
• Large screen interface for controlling scan
acquisition easily, with virtually everything at changes. It takes the guesswork out of • Work surface accommodates a large variety
a single glance; setting scan technique when changing of working conditions and individual
parameters (note: user must select initial operator preferences.
• Excellent simultaneity and multi-tasking Noise Index as well as maximum mA
performance; setting). Noise index enable 100% Image • Wide work surface has room for papers,
Quality reproducibity from one patient to patient charts, logbooks or other
• Completely protocol-driven scan control peripheral devices;
with a dramatic reduction in number of another and from one user to another.
screens; • Wide area also helps make two person
• Protocol Pro™ protocol manager - provides operation more comfortable and
• Highly flexible editing tools that allow easy operator control of automated features (like practical;
tailoring of the exam to the patient; AutoFilm, AutoStore, and AutoTransfer) on
per patient basis. • All console components (monitors,
• Large, 1024 color display; keyboard, mouse and BrightBox) are
• Patient demographics and exam protocols free-standing and can be easily moved
• Leading edge, real-time image processing can be preprogrammed in advance of on the front work surface;
(MPR, MPVR, Volume Viewer*). patient arrival through the Schedule Patient • All surfaces made of high-impact resin
feature. that hides marks and are easy to clean.
In summary, primary benefits of Interactive
16
CT on the LightSpeed Scanner: • A preprogrammed selection of AutoView • Direct network connection means a multi-
and Image Review Layouts allow simple suite Ethernet card is not required for a
• A natural scan control user interface customization of the image presentation to gateway out of the suite - saving costs and
• Dramatic reduction in the number of match the anatomical area of interest - simplifying installation.
screens; only 2 screens to set up first without the complexity of free-form
scan and 1 screen for real-time “windows”. • Learning Solutions provides an on-line
monitoring while scanning; Operator manual detailing system operation
• ImageWorks™ provides instant access to via a multi-media CD-ROM player integrated
• Easier and more flexible protocols advanced image processing features such into the front of the operator console.
• Much more flexible and intuitive graphic as MIP, MPR, MPVR, Volume Viewer*, Learning Solutions can also be accessed on
prescription process with a 1024 Advanced Vessel Analysis*, CT Perfusion a stand alone PC providing flexibility and
Localizer; 2* and DentaScan*. productivity for on demand learning of
• View/Edit Wizard™ intuitively adjusts system operation
dependent parameters automatically in • No waiting for network transfer or image
response to operator-initiated changes retrieval from archive media.
and highlights them for quick review; also • Background filming allows use of the full
Advanced Tube & Generator
alerts the operator to incompatible screen for AutoView and image Technology
dependencies requiring operator review/processing without interruption when
intervention; auto or batch filming. Special “one touch”
• DynaPlan Plus™ full screen display controls provide on-screen viewing of • The Performix X-ray tube, with 6.3 MHU of
illustrates scan status pictorially, with camera progress during AutoFilm without storage and 53.2 kW operation, provides
real-time feedback. disrupting other image processes in increased helical performance with greater
progress. patient throughput and virtually no tube
• Large color screen
• ProView™ visualization algorithms cooling. Advanced technology in the
• Extensive use of picture icons and color Performix tube includes a metal-ceramic
available to enhance anatomical structures
cues enhance ease of use; frame for long life, a high-speed bearing for
without additional image reconstruction
• Large on-screen controls and attractive time. sub-second scanning, a high-efficiency
color palette provide comfortable viewing motor to accelerate the large anode and
over extended periods.

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 4 With Xtream
CT Scanner System

efficient cooling for high throughput and Helical Multi-slice Modes: Prospective Multiple-
superior helical performance.
• Wide range of technique factors (10 to 440
Thickness Reconstruction:
Simplified scan prescriptions and easy-to-use
mA, in 5 mA increments) gives the operator 16
default protocols make the LightSpeed fast
and physician flexibility to tailor technique to For all helical scan modes, the operator can
and efficient in patient set up.
specific needs, optimizing patient dose, and choose to reconstruct images prospectively in
providing the power needed to perform a Multi-slice acquisitions and short intergroup any of the defined nominal slice thicknesses.
broad spectrum of axial and helical delays significantly reduce potential mis-
examinations. In addition to the initial reconstructed slice
registration between scans by increasing the
thickness, the operator has the option to
number of scans possible in a patient breath
prospectively specify additional images to be
hold. Contrast agents may be better utilized
System Specifications as well due to significantly faster scan
reconstructed from a single raw data set.
These images can be reconstructed at any of
acquisition.
Scan Modes: the defined nominal slice thicknesses
Helical protocols are almost identical to available for a given table speed and scan
16
The LightSpeed CT system can perform “classical” axial scan protocols. At the mode.
virtually any clinical application due to its wide beginning of a study, the operator selects the
16
variety of scan modes. With the LightSpeed , This effectively facilitates later, more detailed
type of exam with the anatomical programmer,
body CT studies are easier to perform and image analysis without additional patient
and indicates the desired scan range - either
more productive than ever before. scans and subsequent dose and image
manually, or from a Scout.
registration concerns.
Helical: After completing the prescribed exam, the
system remains ready to continue with
• Continuous 360° scanning with table additional helical scans or a set of axial scans.
8-SLICE HELICAL MODES
incrementation and no interscan delay.
The operator may reconstruct helical scans
• Scans can be acquired in a wide variety of prospectively with up to 90% overlap, and Table Speed (mm/rotation)
speeds. retrospectively, at any arbitrary table location Slice Pitch
in 0.1 mm increments. Thick 0.625:1 0.875:1 1.35:1 1.675:1
Axial: -ness
(mm)
• Up to 16 contiguous axial slices acquired Helical Multi-slice Modes 1.25 6.25 8.75 13.5 16.75
simultaneously with each 360° rotation, with
the time between scans set by the user- The complex nature of helical multi-slice
selected interscan delay (ISD) or intergroup 2.5 6.25 8.75 13.5 16.75
scanning has been simplified by grouping
delay (IGD). all critical acquisition parameters within 8 12.5 17.5 27.0 33.5
• Scans may be easily clustered in groups to basic scan modes, all optimized for image 3.75 6.25 8.75 13.5 16.75
allow multiple scans in a single breath hold. quality and speed. For eight-slice
12.5 17.5 27.0 33.5
acquisition: 0.625: 1; 0.875: 1; 1.35: 1 and
• Minimum scan-to-scan cycle time of only 1.5 1.675: 1. For sixteenslice acquisition: 5 6.25 8.75 13.5 16.75
sec with table moves of ≤ 10 mm (any scan 0.5625: 1; 0.9375: 1; 1.375: 1 and 1.75: 1. 12.5 17.5 27.0 33.5
time). These clinically derived multi-slice scan
modes offer a wide range of selections 7.5 6.25 8.75 13.5 16.75
Scout™: that carefully balance acquisition speed, 12.5 17.5 27.0 33.5
image thickness, artifact level and 10 6.25 8.75 13.5 16.75
• Single radiographic plane for scan retrospective image reconstruction
localization and graphical prescription of flexibility. 12.5 17.5 27.0 33.5
prospective reconstruction;
This simplified user interface guides the
• Extended range matches helical scannable user in the choice of scan parameters.
range. The user selects a pitch mode, a desired
image slice thickness and table travel per
Helical Scans rotation. The user interface also displays
the resulting choice of retrospective image
thicknesses available for each choice of
Slip ring technology has advanced axial acquisition parameters.
scanning by enabling scans with zero
interscan delay and simultaneous table The 16 slice helical acquisition modes
movement. provide table speeds from 5.625
mm/rotation up to 35 mm per rotation,
enabling scan speeds that are up to 12
times faster than 4-slice helical scanners.

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 5 With Xtream
CT Scanner System

Scan Fields of View: Biopsy: Simplified prescription for single


or multiple scans around an arbitrary table
— 25 cm for adult head position aids biopsy studies.
16-SLICE HELICAL MODES
— 25, 50 cm for body Advanced Artifact Reduction (AAR)
Table Speed (mm/rotation) Filter significantly reduces streaking
Slice
— 25 cm for pediatric head artifacts when highly absorbent objects
Pitch
Thick- are in the field of view – ie: large
ness 0.5625 0.9375 1.375 1. 75:1
Helical Scan Enhancements shoulders, screws.
:1 :1 :1
(mm)

0.625 5.625 9.375 13.75 17.5


Full simultaneity allows complete image Helical Image Reconstruction
display, processing and analysis, as well
1.25 5.625 9.375 13.75 17.5 as image archival and filming, concurrent Reconstruction Algorithms: Soft Tissue,
11.25 18.75 27.5 35 with scanning and reconstruction -- even Standard, Detail, Bone, Bone Plus, Lung
when acquiring helical images in a multi- and Edge
2.5 5.625 9.375 13.75 17.5 slice mode.
11.25 18.75 27.5 35 Reconstruction Matrix: 512
Confirm Rx to X-Rays on: < 15 sec. for
3.75 5.625 9.375 13.75 17.5
any state of tube and gantry; < 10 sec. Display Matrix: 1024.
11.25 18.75 27.5 35 with the gantry rotating
5 5.625 9.375 13.75 17.5 Display FOV: Freely variable center/off-
Anatomical Programmer: a ten-region center, prospective/retrospective target
11.25 18.75 27.5 35 anatomical selector allows quick and easy selection.
7.5 11.25 18.75 27.5 35 access to 90 user-programmable
protocols per region. Separate selector CT Number Scale: -1024 to 3071 HU
10 11.25 18.75 27.5 35
for adult and pediatric exams. There are
four selection tabs to select: GE, User, Helical Reconstruction Times:
Service and Most Recent Patient.
Copy/Paste is supported for easy  Reconstruction time as fast 6
Helical Scan Parameters modification and copying of protocols. images per second

Ten user-defined regions. Each region  Typical 0.167 sec. image-to-image


Scan Speeds: Full 360° rotational scans in recon in normal 16 slice recon mode.
has one default protocol displayed with the
0.5, 0.6, 0.7, 0.8, 0.9, 1.0 seconds.
anatomical selector for very fast access to
most commonly used protocols  Maximum image-to-image cycle
Scan Technique:
time is ± 10% for prospective and
— kVp: 80, 100, 120, 140 Protocols include preset scan time, kVp, retrospective image-to-image display.
mA, scan mode, slice thickness and This applies for 512 matrix; any
— mA: 10 to 440 in 5 mA increments spacing, table speed, scan FOV, display display FOV; in AutoView (all
FOV and center, recon algorithm and layouts); with concurrent filming and
— Power: 0.8 to 53.2 kW special image acquisition and processing image archival for all scan modes.
options
— Focal Spot Selection:  Iterative bone processing
Any scan parameters may be edited for increases time by 250 milliseconds.
Small spot for up to 24 kW each scan or all scans - either before or Iterative bone processing, which is
during an exam. The number of scans always enabled for adult head
Large spot for greater than 24 kW may also be easily changed. scanning, reduces image artifacts in
head scans stemming from X-ray
Single Acquisition: 120 second scan AutoScan™: Fully automates longitudinal beam hardening effects.
maximum. table movement and start of each scan.
Minimum DFOV: 9.6 cm
Helical Tilt: helical acquisition is possible AutoVoice™: 3 preset (English) and 17
with the gantry tilted to a maximum of 30 user-defined messages automatically Minimum Pixel Size: 0.19 mm
degree, in half degree increments. deliver patient breathing instructions;
especially useful for multiple helical Queued Recon: Requests will be
Multiple Acquisition Maximum Scan Time: scanning. processed continuously and
Multiple scans may be acquired in one simultaneously with other processes on
series to produce up to 3,000 contiguous Trauma Patient: Allows patient scans and the system including scanning.
helical images. Up to 2,881 seconds image display/analysis without entering Prospective recon will be prioritized over
helical coverage is possible in multiple patient data before scanning. retrospective recon.
series.
Priority Recon Queuing: One touch
Minimum Inter-Group Delay (IGD): 5 selection marks most recent rotation for
seconds between adjacent helical scans

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 6 With Xtream
CT Scanner System

next available recon. Available during or Total Scan Total Elapsed Max
after scanning. Time (sec.) Time (sec.) mA
Helical Scan Protocols
Images annotated to indicate continuous 150 155 100-200
scan acquisition with table incrementation: All protocols assume 120 kVp scans under 200 205 80-150
typical clinical conditions. 450 465 50-70
HE (helical) + Pitch
Single Helical Scans:
Table speed Helical Scan Image Quality
Prospective Multiple Reconstruction Scan Maximum Volumetric Helical Scan Image Quality
(PMR): Up to 3 sets of reconstructions can Time mA
be pre-programmed as part of the scan 16
With LightSpeed being a sub-millimeter
10 sec. 440
protocol prior to acquisition. The operator isotropic CT scanner, it is now possible to
20 390-440
can select different start/end location, slice specify coronal and sagittal image quality:
30 330-380
thickness, interval, reconstruction
algorithms and display fields of view for 40 290-350
1. Visual Measurement:
each reconstruction. This frees the 50 260-320
operator from sitting at the console and 60 230-310 Reformatted resolution is demonstrated on
directly contributes to increased 70 210-300 the Nuclear Associates AAPM High
productivity. 80 180-280 Contrast Resolution Insert #76-413.
90 170-270
Prospective Recon: Operator may initiate 100 160-260 0.5mm
full recons at any table location in 110 150-250
increments of 1/10 the image thickness; 120 140-230
image thickness remains constant.
Multiple Helical Scans
Retrospective Recon: Operator may
initiate full recons at any table location in (contiguous helical coverage with 5-
0.1 mm increments; image thickness second IGD):
remains constant.
Scan No. Max
Retrospective Image Decomposition: Time IGD Scans mA
The operator has the option to 10 sec. 5 sec. 3 340-420
retrospectively decompose the original 4 300-400
raw data set and reconstruct additional 5 270-390 0.4mm
images at any of the defined nominal 6 240-360
image thickness available for a given table 20 sec. 5 sec. 2 300-360 An Effective 0.5 mm voxel size is clearly
speed and scan mode. 3 240-330 seen in coronal and sagittal views.
4 180-310
30 sec. 5 sec. 2 230-310 2. Statistic Measurement:
3 180-280 Volumetric MTF is computed from the X,Y
4 140-240 and Z axis of a 0.2mm bead using sagittal
40 sec. 5 sec. 2 190-290 and coronal images of the 8” Catphan,
3 140-230 MTF module with bead source #CTP445.
4 110-190
50 sec. 5 sec. 2 170-270
3 120-200
4 90-160 Detail Algorithm
60 sec. 5 sec. 2 140-240
MTF (z-axis) is 12.4 lp/cm @ 10%, 15
3 100-170
lp/cm @ 0% which is a 0.333 mm limiting
resolution
Contiguous Helical Coverage
In Plane Helical Scan Image Quality
Multiple helical scans can be performed in
succession with only 5-second delays For details of scan technique please refer
between helical scans, providing up to to the technical reference manual.
2,881 contiguous rotations (with up to
1,500 images in one series). High-Contrast Spatial Resolution

(The following chart assumes 120-second On GE Performance phantom:


helical with 5-sec IGD)
Standard Algorithm

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 7 With Xtream
CT Scanner System

• 0.584 mm limiting resolution • Nominal Thickness: 0.625, 1.25 mm

• 4.0 lp/cm @ 50% MTF Axial Scans


• 6.5 lp/cm @ 10% MTF Multi-slice acquisitions and short interscan Axial Scan Parameters
delays significantly reduce potential missed
• 8.5 lp/cm @ 0% MTF registration between scans by increasing the Scan Time:
number of scans possible in a patient breath • 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 2.0, 3.0 and 4.0
Hi-Res Algorithm (Edge). hold. Contrast agents may be better utilized second full scans (360° acquisition)
as well due to significantly faster scans.
• 0.324 mm limiting resolution
Scan Technique:
• 8.5 lp/cm @ 50% MTF Axial Multi-slice Prescription • kVp: 80, 100, 120, 140

• 13.0 lp/cm @ 10% MTF • mA: 10 to 440, in 5 mA increments


Simplified scan prescriptions and easy-to-use
default protocols make the Lightspeed CT
16 • Power: 0.8 to 53.2 kW
• 15.4 lp/cm @ 0% MTF
Scanner fast and efficient in patient set-up. • Focal Spot Selection:
Line pair values decrease with larger focal Axial protocols are nearly identical to helical
scan protocols. • Small spot for up to 24 kW
spot (by 5% with Standard and by 7% with
• Large spot for greater than 24 kW
Hi-Res); limiting resolution is unaffected.

Low-Contrast Detectability
Axial Multi-slice Modes Scan Plane Geometry:

16
• +/- 30° gantry tilt, in 0.5° increments
The LightSpeed acquires axial scans in
On 8 inch (20 cm) CATPHAN phantom: sets of 8 or 16 contiguous images in one 360° • Longitudinal positioning in 0.01 mm per slice
5 mm @ .3% at 13.3 mGy rotation. increment. Gantry display in 0.5 mm
3 mm @ .3% at 37.2 mGy increments.
For each rotation of the gantry, the
Noise:
16
LightSpeed collects 16 rows of scan Interscan Delay (ISD):
data. There are five reconstruction modes • Minimum ISD with table movements of
On either an AAPM water phantom or GE available for creating images from the 0 - 10 mm: 1.0 sec.
Quality Assurance phantom: multi-slice scan data (1i, 2i, 4i, 8i, and 16i).
By using 1i, 2i, 4i, and 8i reconstruction • Minimum ISD with table movements of more
0.32% +/- 0.03% at 28.5 mGy (2.85 Rad) modes, scan data can be combined prior than 10 mm and up to 20 mm: 1.3 sec
to image reconstruction to create slices • User-selectable.
with reduced partial-volume artifacts. This
is particularly useful for posterior-fossa Inter Group Delay (IGD):
CTDI: imaging.
• Minimum IGD is the same as minimum ISD;
On CTDI Head and Body Dose Reference also user-selectable.
Phantoms:
1i Mode: Scan-to-Scan Cycle:
CTDIvol expressed in mGy/100 mAs for
• Produces 1 image per rotation • Minimum scan-to-scan cycle of 1.5 seconds
IEC pitch 1: (normalized to a pitch of 1).
• Nominal Thickness: 1.25, 5, 10 mm possible for 0.5 sec. scan speed with
Head 20.9 mGy/100 mAs minimum ISD's.
Body 10.6 mGy/100 mAs 2i Mode:
Scan Fields of View:
• Produces 2 images per rotation
• 25 cm for adult head
• Nominal Thickness: 0.625, 2.5, 5, 7.5, 10
mm • 25, 50 cm for body
• 25 cm for pediatric head
4i Mode:
• Produces 4 images per rotation Scan with no table incrementation,
contiguous image location, or skipped
• Nominal Thickness: 2.5, 3.75, 5 mm image location are possible. Overlapped
axial scans are not possible.
8i Mode:
• Produces 8 images per rotation
Axial Image Reconstruction
• Nominal Thickness: 1.25, 2.5 mm
Reconstruction Algorithms: Soft Tissue,
16i Mode:
Standard, Detail, Bone, Bone Plus, Lung and
• Produces 16 images per rotation Edge.

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 8 With Xtream
CT Scanner System

Reconstruction Matrix: 512 Prospective Perscription


Slice Recon # Scans Acquisition
Display Matrix: 1024. Scan Mode Thickness Slice Thicknesses mA (Clusters) Time
16i - 0.625mm 440 15-45 (3-11) 0:41-2:49
Display FOV: Freely variable center/off-center,
prospective/retrospective target selection. 8i - 1.25mm 400 25-55 (5-13) 1:13-3:21
16 row 0.625
4i - 2.5mm 340 35-75 (7-18) 1:45-4:41
CT Number Scale: -1024 to 3071 HU 2i - 5.0mm
320 45-85 (9-20) 2:17-5:13
16i - 1.25mm
Axial Image Reconstruction: 300 50-95 (10-22) 2:33-5:45
8i - 2.5mm
16row 1.25 280 60-110 (12-25) 3:05-6:33
• Reconstruction time as fast as 6 images 4i - 5.0mm
per second. 2i - 10mm 260 75-125 (13-28) 3:21-7:21
8i - 1.25mm 240 85-145 (14-31) 3:37-8:09
• Typical 0.167 sec. image-to-image recon in
16 slice recon mode. 4i - 2.5mm 220 100-170 (16-34) 4:09-8:57
8 row 1.25
2i - 5.0mm
200 115-185 (18-38) 4:41-10:01
• Maximum image-to-image cycle time is 1i - 10mm
8i - 2.5mm 180 135-250 (22-43) 5:45-11:21
± 10% for prospective and retrospective
image-to-image display. This applies for 8row 2.5 4i - 5.0mm
512 matrix; any display FOV; in 2i - 10mm Axial Scan Image Quality
AutoView (all layouts); with concurrent
filming and image archival for all scan
modes. Axial Scan Protocols For details of scan technique parameters,
please refer to the technical reference
• Iterative bone processing increases time by manual.
250 milliseconds. All protocols assume 120 kVp scans under
typical clinical conditions. High Contrast Spatial Resolution:
Prospective Multiple Reconstruction Standard Scans:
(PMR): Up to 3 sets of reconstructions can On GE Performance phantom:
be pre-programmed as part of the scan
protocol prior to acquisition. The operator Standard Algorithm
can select different reconstruction Scan Acquisition
algorithms and display fields of view for Time ISD mA Scans Time • 0.584 mm limiting resolution
each reconstruction. This frees the 1 sec. 1 sec. 440 18-45 0:35-1:27
operator from sitting at the console and • 4.0 lp/cm @ 50% MTF
1 1 400 24-55 0:47-1:49
directly contributes to increased 1 1 360 32-68 1:03-2:15
productivity. • 6.5 lp/cm @ 10% MTF
1 1 340 37-76 1:13-2:31
The operator has the option to reconstruct 1 1 320 43-86 1:25-2:49 • 8.5 lp/cm @ 0% MTF
the original raw data set at any of the 1 1 300 50-97 1:37-3:11 Hi-Res Algorithm
defined nominal slice thicknesses.
1 1 280 58-110 1:45-3:37
Reconstructions can be prescribed down 1 1 260 66-122 1:55-3:59 • 0.324 mm limiting resolution
to 1/16 the original acquisition image 1 1 240 74-135 2:07-4:21
thickness for images acquired in the 1i • 8.5 lp/cm @ 50% MTF
scan mode, down to 1/8 the original 1 1 220 84-152 2:21-4:49
1 1 200 94-168 2:37-5:21 • 13.0 lp/cm @ 10% MTF
thickness for 2i mode, and down to 1/4 the
original thickness for 4i mode. 2 1 200 37-77 1:13-2:33 • 15.4 lp/cm @ 0% MTF
2 1 180 42-86 1:23-2:51
Similarly, additional reconstruction
Line pair values decrease with larger focal
supports partial-volume artifact reduction
spot (by 5% with Standard and by 7% with
by reconstructing images with 4, 8, or 16
• Cluster Scans (All cluster protocols assume Hi-Res); limiting resolution is unaffected.
times the acquisition image thickness.
9-second clusters of five slices, 1-second
scans with 1-second interscan delays and 7 Low-Contrast Detectability
These reconstruction features effectively
facilitate later, more detailed image seconds between clusters).
On 8 inch (20 cm) CATPHAN phantom:
analysis without additional patient scans
and subsequent dose and image 5mm @ .3% at 13.3 mGy
registration concerns. 3mm @ .3% at 37.2 mGy
The following table illustrates the
retrospective reconstruction image
thicknesses available for each acquisition
thickness and mode:

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 9 With Xtream
CT Scanner System

Orientation: AP, RLAT, PA, LLAT (preset); Exam Rx:


Noise: or any angle from 0° - 359° (manually
selected). The Exam Rx desktop environment provides
On either an AAPM water phantom or GE
the clinical tools necessary for comfortable,
Quality Assurance phantom: Axial scan prescription lines indicate scan efficient control of patient studies.
location to nearest 1 mm table position.
0.32% +/- 0.03% at 29.3 mGy (2.93 Rad)
These tools include patient scheduling and
Scouts longer than 1,000 mm are auto data entry, exam protocol selection, protocol
CTDI: minified to fit the display. viewing and editing, scan data acquisition,
On CTDI Head and Body Dose Reference image reconstruction, image display and
Phantoms: User Interface routine analysis, AutoFilm or manual filming,
AutoStore and AutoTransfer.
CTDI expressed in mGy/100 mAs: 16
The Lightspeed Operator Console utilizes a
computer workstation with the following user ImageWorks:
Head 16.8 mGy/100 mAs center interface features:
17.0 mGy/100 mAs surface • Two 20-inch monitors ImageWorks is a desktop environment
designed to take advantage of the
• Scan/recon monitor for scan and recon 16
Lightspeed CT Scanner System computer
Body 5.50 mGy/100 mAs center control with no image display
10.0 mGy/100 mAs surface and image processor.
• Image monitor for image display,
CTDI100 expressed in mGy/100 mAs: analysis, processing, and management Standard features include archive, network
• Each monitor provides a 1280 x 1024 and manual film control, as well as some
Head 18.7 mGy/100 mAs center high resolution, flicker-free display advanced image processing such as multi-
17.9 mGy/100 mAs surface planar reformatting (MPR), multi-projection
• Optional 18” LCD Monitors. volume rendering (MPVR), and MR image
Body 5.4 mGy/100 mAs center
11.2 mGy/100 mAs surface • Scan control keyboard assembly with display. It also has optional add-on packages
intercom speaker, microphone and volume for Volume Viewer*, CT Perfusion2* and
CTDIvol expressed in mGy/100 mAs: controls DentaScan*.

Head 18.4 mGy/100 mAs • Three button mouse with mouse pad The ImageWorks desktop also provides a
Body 9.3 mGy/100 mAs • BrightBox (trackball assembly) gateway for DICOM 3.0 image transactions,
either through a local area network, or via
• Two wide work surfaces DICOM-formatted MOD media.
Scout Scans
All these devices are free-standing and can be
easily moved to accommodate a large variety Learning Solutions:
ScoutView™ scans provide excellent
of working conditions and individual operator
detail for anatomical localization in 16
The LightSpeed provides an on-screen, on-
preferences.
conjunction with scan prescription. line operator manual via a multi-media CD-
Split table top allows unrestricted patient ROM player integrated into the front of the
Scan locations may be prescribed at the
viewing while still supporting 2 monitors. Each operator’s console. Learning Solutions is also
operator console either graphically (via
work surface can be adjusted at installation to viewable on a stand-alone PC providing
mouse), or explicitly (keyboard entry) from
help accommodate a variety of siting flexibility and productivity for on-demand
a Scout scan.
requirements. learning of system operation.
Prescription of scans with multiple gantry
angles are also available on a single
Scout.
Desktop Overview
The user interface utilizes the paradigm of Exam Rx
Scout Scan Parameters managed work environments for a more
intuitive clinical workflow.
Aperture: 1.25 mm effective aperture Virtually all clinical operations are managed
through three “virtual desktops” or applications Patient Scheduling
Table speed: 100 mm/sec. managers: Exam Rx, ImageWorks and
Learning Solutions. Operators can effortlessly Patient demographics and exam protocols can
Scan Technique:
move back and forth between these be pre-programmed in advance of patient
• kVp: 80, 100, 120, 140 environments simply by clicking on an icon.
TM
arrival by selecting Schedule Patient from the
• mA: 10 to 440 in 5 mA increments Xtream technology enhances multi-tasking scan/recon monitor. This productivity
• Power: 0.8 to 53.2 kW architecture and maintains simultaneously all enhancement allows entry of all or some of a
processes so no work is lost or disrupted as patient’s demographic data, as well as pre-
desktops are switched. selection of the exam protocol.

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 10 With Xtream
CT Scanner System

This feature is available any time a patient display the 1024 Localizer each time a new • AutoVoice: Preset (English) and user-
exam is not currently underway. series is requested. recorded messages automatically deliver
patient breathing instructions, especially
This feature uses the same interface as New • Default protocols also include preset scan
useful for multiple or multi-pass helical
Patient selection for simplified, consistent time, kVp, mA, slice thickness, scan mode,
scans
programming. table speed, image interval, gantry tilt, scan
field-of-view, display field-of-view and • Full Simultaneity allows scan and recon to
Patient information can be easily recalled to center, recon types, and breath timing work concurrently with image display,
set up an immediate exam via List/Select parameters. processing and analysis (including
Scheduled Patient on the scan/recon monitor. computationally intensive features such as
• Any scan parameter can be edited for each
Pre-programmed patient exams can also be MPR, MPVR and 3D*/MIP) while still
scan or all scans either before or during an
recalled from the New Patient screen running image archival, filming and
exam. Scans can be easily added or
automatically by entering the patient ID networking processes.
removed from the prescription.
number.
• Scan/recon control uses only 2 screens to
set up first scan - New Patient and Protocol Dose Computation & Display
Patient Data Entry: View/Edit.
CTDIvol (CTDI volume), DLP (Dose Length
Patient data can be entered as part of New Product), and Dose Efficiency computation
Patient set-up, or can be recalled from the list
Protocol View/Edit: and display during scan prescription provides
of pre-scheduled patients. patient dose information to the operator.
• A single, full screen View/Edit table allows
fast and easy examination and modification CTDIvol is a dose index defined by IEC 601-2-
Trauma Patient ID allows patient scans and
of exam parameters before scanning begins 44. This index is computed automatically by
image display/analysis without entering patient 16
the Lightspeed CT System and reported on
data before scanning. • Exam parameters can be changed for just the Exam Rx screen. CTDIvol is a single
one scan, or for all scans in a series number consisting of 2/3 of the CTDI100
Exam Protocol Selection: • When used in conjunction with the 1024 peripheral dose plus 1/3 of the CTDI100 central
Localizer, changes made in the View/Edit dose that is divided by the helical or axial pitch
table that affect the number of scans, image factor.
One of the main contributions of the
16 interval, starting/ending locations, tilt, or
LightSpeed CT Scanner System to CTDI100 is a dose index based upon CTDI
display FOV are automatically shown on the
department productivity is its simplified exam dose measurements over a 100 mm volume,
1024 Localizer
set-up. as defined in IEC 601-2-44.
• Any changes made directly on the 1024
• Exam parameter set-up has been greatly Dose Length Product (DLP) is given in
Localizer display using the mouse and the
simplified through the exclusive use of mGy*cm and is computed and displayed for
on-screen cursor controls are also reflected
protocols each group prior to the scan. Additionally, an
automatically in the View/Edit table
• Protocols can be easily selected in one of accumulated DLP is displayed for the entire
• View/Edit Wizard intuitively adjusts exam, as the exam prescription progresses.
three convenient ways:
dependent parameters automatically in The final exam accumulated DLP provides a
• A large, graphical Anatomical response to operator-initiated changes and convenient measure for maintaining patient or
Programmer located on the New Patient highlights them for quick review. It also procedure dose management statistics.
screen alerts the operator to incompatible
• A default list of the “top 10” most dependencies requiring operator Dose Efficiency is automatically
commonly used protocols located near intervention. computed and displayed on the Exam Rx
the anatomical programmer screen. The dose efficiency is a measure
• Tab card groupings for Timing, Recon and
• A numerical entry of how much of the Z-axis X-ray beam is
Filming help organize the large number of
used by the system, as defined in IEC
• Two Anatomical Programmers - one for parameters available within each protocol.
601-2-44.
adults and one for pediatrics - provide quick • As many as 8,460 protocols can be stored
and easy access to 8,460 user- on the Operator Console.
programmable protocols (total). Each AutoView Layouts:
programmer has ten anatomical regions with
90 protocols for each region Scan Data Acquisition: • Eight powerful AutoView layouts provide
exceptional flexibility in tailoring the 1,024
• Default protocols have been expanded • Full-screen DynaPlan Plus illustrates scan image display to the user or the application
through Protocol Pro - a “behind the scenes” status graphically, with real-time feedback at hand - without the complexity of free-form
protocol manager - that allows preselection while the exam is underway. Scans, “windows.”
of automated features like AutoVoice, programmed delays (prep, breathing, inter-
AutoFilm, AutoStore and AutoTransfer on a • AutoView Layouts include:
group), and even AutoVoice
per-exam basis. announcements are clearly shown before • 1024 AutoView image
• Protocol Pro also provides preselection of and during scanning. • 768 AutoView image (matches the image
two different window/level settings per • AutoScan: Fully automates longitudinal size shown on the HiSpeed Advantage
image for AutoFilm and can automatically table movement and the start of each scan 2.X Series OC monitor)

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 11 With Xtream
CT Scanner System

• 512 AutoView image + 512 Localizer Image Access: • Measure Angle


Scout with cut lines automatically • Grid On/Off
showing the location of the AutoView • Point and click interface along with a
image on the Scout • Hide Graphics
pictorial directory (browser) allows for easy
• Two 512 AutoView images (same image selection by exam, series or image • Erase
but at different window/level settings) + • Screen Save
512 Localizer Scout with cut lines • MIROI (Multiple Image ROI)
automatically showing the location of the Routine Image Display:
• Report Pixels
AutoView images on the Scout
• Image display features provided within
• 512 AutoView image + 512 AutoFilm Exam Rx:
image Display Preferences:
• Zoom/Roam
• Last two 512 AutoView images
• Explicit Magnify • Display settings available to tailor the overall
• Last four 512 AutoView images display (settings apply to all images in all
• Flip/Rotate
• AutoLink which links the current series to exams):
a view port • ProView
• Annotation Levels
• Display Normal
• Basic image review features such as • Inverse Video
window/level, magnification and flip/rotate • List/Select
• Next/Prior Each View Port
are available for AutoView images. • Ellipse ROI
• Next/Prior Series Binding
• Any window not used for AutoView is • Measure Distance
• Continuous Report Cursor
available to independent, simultaneous • Grid On/Off
review of other exams. • Cross Reference
• Special BrightBox, a three-button trackball • User Annotation
Auto Image Management:
device, provides independent control of • Exam/Series Page
image next, prior, manual paging and The Exam Rx work environment conveniently
trackball window/level for any review images • Hide Graphics provides for selection of AutoFilm, AutoStore
in focus. This helps make two person • Erase (to local or remote MOD), and AutoTransfer
operation practical. • Screen Save (across a network).
• Regardless of the AutoView Layout used, • Gray Scale Enhancement An AutoFilm Composer provides a simple
AutoFilm viewing is available anytime via an programming interface for automated filming
on-image selection - without disrupting other • ProView visualization algorithms are
available to enhance anatomical structures set-up.
image processes in progress. Background
filming allows full use of the image display without additional reconstruction time:
Batch Filming is accomplished through a
monitor for AutoView and image • Four Selections for enhancement of high single keystroke which automatically prints an
review/processing without interruption contrast objects where fine detail is entire series at a time.
during AutoFilm. required without aliasing (such as lungs)
• Three Selections for modifying perceived
levels of noise and low contrast Manual Image Filming:
Image Review Layouts: discrimination
• On-screen filming is available for any analog
• Five flexible Image Review Layouts are • Three ways are provided to adjust window/ or digital camera using a 3M-952 protocol.
provided for those applications where level of images in focus in order to meet a
greater than 512 image display may be variety of clinical work environments and • Images may be individually filmed manually
desired and AutoView is not required. user preferences: via “drag and drop” to the on-screen Film
Composer.
• Image Review Layouts include: • Six user-programmable keys on the scan
control keyboard (F6 - F11), plus one key • Print Series permits automatic printing of an
Note: uses short notation for screen options for returning to prior setting (F5) entire series with one keystroke.
• 1024 single image display • On-image through middle mouse button • Page filming permits creation of an entire
• 768 single image display • BrightBox trackball film with one keystroke.
• Two 512 image display, horizontal format • Multiple image formatting allows filming of
• Two 512 image display, vertical format multiple images in a single film frame.
Routine Measurements:
• Four 512 image display • Film formats supported are 1:1, 2:1, 4:1, 6:1,
• Each image display window can be further • Image measurement features provided 8:1, 9:1, 12:1, 15:1, 16:1, 20:1, 24:1 and 35-
subdivided into four more images, within Exam Rx: mm slide
increasing the total number of images that • Box ROI 16
can be displayed at once to 16. Important note: The Lightspeed CT Scanner
• Ellipse ROI comes standard with a DICOM Print Interface
• BrightBox image control is also available for • Trace ROI configurable for multiple DICOM Print
Image Review Layouts. • Measure Distance destinations. Connections with cameras that

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 12 With Xtream
CT Scanner System

do not support DICOM Print may require a (Volume Rendering, Volume Analysis and scroll or flip and rotate to a user-defined
filming interface (purchased separately). Navigator) that includes : image set.

To save further filming cost, the Operator • Dynamic Volume Review™ for Fast • Save State stores user-selected image
Console can directly print to a postscript Screening orientation and window/level with each data
printer such as the GE Color Printer available set.
• Curved Volume Of Interest
as an option. • Window/Level values may be:
• Protocol Management and Loading
• Preset to provide six on-screen instant
• Review Layout Presets
ImageWorks window/level settings
• Multiple VR Objects Merge • Set independently for up to 16 images on
ImageWorks software is designed to take the screen
16
• Pseudo Surface Shading Mode
advantage of the Lightspeed CT Scanner’s • User-modified in discrete or variable
computer and image processor. This desktop • Predefined Cut Planes steps
environment includes image management and • Volume Rendered Navigator views • Adjusted real-time on-image by holding
networking. down the middle mouse button and
• VR Preset save/recall
moving the mouse
Because some of the image analysis and • 3D Rendered Lumen View
display features of ImageWorks replicate • Cine mode provides paging in up to 4 view
those in Exam Rx, the next section describes • Automatic Path Tracking ports of up to 128 previously-stored CT or
only features that are incremental or MR images at full selected display frame
• Path Bridging (in case of occlusions)
significantly different. rate. For more than 128 images, display
• SmartCursor™ for Easy Navigation frame rate may be reduced.

Image Analysis • Synchronized Reformatted Views • Cine mode also provides temporal, spatial
or manual playback loops.
• Cut visualization mode
• Multi-Projection Volume • Text Page
Reconstruction (MPVR): Quick and easy
way to generate volumetric images for CT
angiography without thresholding data or *Advanced Vessel Analysis Image Annotation
removing unwanted anatomy. An entire Advanced Vessel Analysis is the ultimate
• Image annotation and cursor are shadowed
volume is used to generate images in any tool to assess and quantify vascular
to permit ease in reading.
plane, creating real-time frames of structures, including stenosis analysis,
reference at the same time; stent planning procedures, post stenting or

• Clinical utility is extended via two


vascular surgery follow-up. Image Management
additional modes: • protocol driven tools to perform quick,
• Images may be stored and retrieved via
flexible and accurate quantitative analysis of
Magnetic Optical Drive (MOD) media using
− MIPS - enhances contrast and improves vascular anatomy
DICOM 3.0 format. This allows interchange
visualization of calcifications
• provides maximum, minimum and mean with other imaging systems supporting
intraluminal diameter measurements DICOM 3.0 MOD media. Not all vendors
− Average - generates 2D radiographic
implementation of DICOM 3.0 are identical,
images; • provides cross-sectional areas of true
so please check with the manufacturer for
orthogonal sections of the aortoiliac systems
• Multi-planar Reformation (MPR): compatibility.
at selected anatomical points
Provides real-time assessment of anatomy • Off-line retrieval of all image files. Images
in offaxis planes. Sagittal, coronal, oblique • clinical benefits include: stenosis sizing, pre-
may be viewed as soon as they are restored
and curved planar reformations available; and post- surgical assessment, stent
from MOD.
planning
• Batch reformatting can also be defined • Image retrieval time is approximately 30
• Measurements in % stenosis or mm of
and executed for later viewing if desired; images 512 per second.
stenosis, and measurement of length and
dimension of stenosis.
• Image Addition and Subtraction: Includes
image addition of more than two images at VariViewerTM:
a time;
• VariViewer is an interactive axial review
Image Display mode that can change the slice thickness
reconstruction instantaneously.
• Magnifying Glass allows quick 2X mag
window that can be moved over an image. • The user selects the volume to be
*Volume Viewer
specifically analyzed and chooses the slice
• Image Scroll moves an image within its’ own thickness to be displayed as axial images.
Volume Viewer is an innovative and window.
powerfull suite of productivity enhanvers The user can then save a number of recon
• Groupings allow application of window/level images sets, viewing a large number of
values, magnification/minification, image slices for pure axial review and filming.

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 13 With Xtream
CT Scanner System

• Batch film can typically enable you to reduce DICOM Conformance Laser Alignment Lights:
filming images by 50%, thickening the slab
• Define both internal and external scan
from 1.25 to 2.5mm with no information loss. Standards: planes to ± 1 mm accuracy.
• DICOM 3.0 Storage Service Class • Operate over full range of gantry tilt;
Image Networking: • Service Class User (SCU) for image activated any time during exam (with tube
send stationary).
Exams can be selected and moved from • Service Class Provider (SCP) for image • Coronal light remains perpendicular to axial
MR Signa Advantage, CT HiSpeed receive light as gantry tilts.
Advantage, Advantage Independent
Console, Advantage Windows Workstation • DICOM 3.0 Query/Retrieve Service Class Visual readout is easy to read from the table
16
to the Lightspeed CT Scanner’s Operator • DICOM 3.0 MOD Media Service Class on side or from the operator console.
Console. This is performed via the 1.2- and 2.3-GB MOD media
AdvantageNet (GenesisNet) point-to-point Gantry tilt controls are located on the side of
protocol. This protocol supports send, • DICOM 3.0 Storage Commitment Class the gantry.
receive and pull/query (no broadcast). Push
16
NOTE: Because the Lightspeed creates • DICOM 3.0 Modality Worklist*
images in pure DICOM format, exams can Table
not be moved from the Lightspeed CT
16
• DICOM 3.0 Modality Performed Procedure
Scanner System to MR Signa Advantage, Step* • Single table, cantilever design with wide
CT HiSpeed Advantage or the Advantage height range
• DICOM 3.0 Print
Independent Console systems. • Vertical Range: 51.6 cm to 99.1 cm
Exams can also be selected and moved Filming Protocol: • Vertical Scannable Range: 80.6 cm to 99.1
16
between the Lightspeed CT Scanner cm
System and any imaging system • 3M-952 Standard • Elevation Speeds: 5 mm/sec and 40
supporting the DICOM 3.0 protocol for
mm/sec
network send, receive and pull/query. GE
systems that support this interoperability System Components • Horizontal Range: 170 cm
include CT HiSpeed CT/i, X/i, NX/i, QX/i,
• Horizontal Scannable Range: 170 cm
MR Signa LX, and Advantage Windows
1.2/2.x/3.x/4.x. Gantry: metal-free (axial) and 160 cm metal-free
(helical & Scout)
Image transfer time using either Advanced slip ring design continuously rotates • Horizontal Speed: Up to 100 mm/sec
AdvantageNet or DICOM 3.0 protocols is generator, tube, detector and data acquisition
approximately 0.1 second per 512 image • Table automatically recenters on scan plane
system around the patient. with changes in vertical position (after
on a 100baseT Network.
• Aperture: 70 cm setting internal landmark with alignment
lights on)
• Tilt: ± 30°
Computer Based Training • Table Load Capacity:
This innovative feature provides on- • Tilt Speed: 1°/sec.
• 180 kg (400 lb) with +/- 0.25 mm
screen, on-line support of Exam Rx and • Focus to Detector: 95 cm positional accuracy
ImageWorks functions via a multi-media
CDROM player, integrated into a tower on • Focus to Isocenter: 54 cm • 205 kg (450 lb) maximum allowed with
operator console for easy access. Use of normal operation and +/- 1 mm positional
• Maximum SFOV: 50 cm
rich graphics and text to provide truly accuracy
useful system help. • Rotational Speeds: 360° in 0.5, 0.6, 0.7 0.8,
• Controls on gantry for elevation and cradle
0.9, 1.0, 2.0, 3.0 and 4.0 sec.
incrementation. Foot pedals on both sides of
A well-indexed table of contents helps • Remote Tilt from Operator’s Console. table for fast elevation. Cradle position
speed access to information. The CD can controlled from OC for prescribed scans.
be viewed on the scanner or on a PC. • Integrated breathing lights and countdown
timer. • IV Pole integrated at the foot-end of the
table prevents IV lines from becoming
• Integrated start scan button with countdown crossed and tangled, and ensures that the
to X-ray on. lines stay securely in place on the patient.
Industry Standards Scan plane toward front of gantry for improved
16
• The LightSpeed CT Scanner System
positioning access. X-Ray Tube
complies with a wide variety of industry Biopsy and interventional studies have been
standards to facilitate more rapid adoption facilitated through a more streamlined gantry Performix ADV Metal-Ceramic Tube Unit.
of features and performance shroud, and bilateral table/gantry controls and Design optimized for exams requiring a
improvements as the computing and gantry display that maximize maneuverability large number of scans without tube
medical imaging industry evolves. while working next to the gantry. cooling.

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 14 With Xtream
CT Scanner System

• Heat Storage Capacity: 6.3 MHU • 16 x 1.25 mm (uses all 24 rows) Peripherals:
• Heat Dissipation: • 16 x 0.625 mm (uses center 16 rows )
• Anode (max) 840 KHU/min Total of 254 GB system:
70% geometric efficiency; 98% absorption
• Casing (cont) 300 KHU/min efficiency. • Main system (host) disk drive:
• Tube Unit: 6.9 kW continuous for 10 • High Performance Drive
minutes
Data Acquisition System • 36 GB, 3.5 inch form factor
• Dual Focal Spots: • 15,000 RPM
• Small Focal Spot: 12,288 available input channels. • Ultra320 SCSI interface
0.7 mm (W) x 0.6 mm (L) nominal • Assigned to applications software and
value 1,640 Hz maximum sample rate.
image files
(IEC 336/93)
Effective analog to digital conversion range • 2 system disk drives (Image Disk)
0.9 mm (W) x 0.7 mm (L) traditional greater than two million to one.
• High Performance Drive
methodology
• 73 GB, 3.5 inch form factor each.
• Large Focal Spot: Scan/Control Unit • 15,000 RPM
0.9 mm (W) x 0.9 mm (L) nominal • Ultra320 SCSI interface
value Located in base of Operator Console.
• Assigned to image files only
(IEC 336/93)
- 250,000 uncompressed 512 images
1.2 mm (W) x 1.2 mm (L) • Scan data disk drive:
traditional methodology Host Computer • High Performance Drive
• Maximum Power: 53.2 kW • Dual SMP 2.66.GHz Intel Xeon processors • 36 GB, 3.5 inch form factor
• Beam collimated to 55° fan angle.
with 512KB L2 cache. • UltraSCSI interface
• Intel Hyper-threading technology. • Scan data disk drive:
Average time to replace tube: < 10 hours
• 2GB DDR266 Dual Channel Memory with a • High Performance Drive
throughput of 4.2GB/sec • 36 GB, 3.5 inch form factor
High Voltage Generation • UltraSCSI interface
• High-frequency on-board generator. Image Processor:
Continuous operation during scan. • Standard MOD drive:
• Nvidia Quadro4 980XGL AGP 8X graphics
• 53.2 kW output power. with 128MB Memory • Magnetic Optical Disk Drive
• kVp: 80, 100, 120, 140 • Erasable, rewritable media
• Graphics Processor Unit (GPU) Clock
300Mhz • 2.3 GB, 3.5 inch form factor
• mA: 10 to 440 mA, 5 mA increments
• Graphics Memory Clock 325Mhz • Assigned to DICOM 3.0 format image
Maximum mA for each kVp selection: file.
• Stores 4,700 lossless JPEG compressed
512 image files per side
Image Reconstruction Engine • Off-line retrieval of image. Images may
KVp Max mA
80 400 (GRE) be viewed as soon as they are restored
100 420 from MOD
120 440 • Custom-designed special purpose
140 380 CT Image Generator • DVD Ram:
• Custom CT back projection hardware • 4.7 GB per side, 5.25" half height form
accelerates 2D & 3D back projection.
HiLight Matrix II Detector: • Intel Hyper-threading Technology.
factor
• Transfer rate 2.7MB/sec
21,888 individual elements composed by: 8 • 32-bit floating point data format • Assigned to scan data file and protocol
rows of 1.25 mm thickness and 16 rows of • 2GB DDR226 ECC Dual Channel file storage/retrieval.
0.625 mm thickness, each containing 888 Memory Standard (4.2 GB/sec).
active patient elements; 24 reference
elements. • Color monitors:
Software Architecture: • 21 inch diagonal width
4 typical modes of data output: • 1280 x 1024 dot resolution
• Software architecture based on industry
• 8 x 1.25 mm (uses center 16 rows) standards and client-server design • Non-interlaced, flicker-free presentation
• 8 x 2.5 mm (uses all 24 rows) • 76 kHz Horizontal deflection frequency

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 15 With Xtream
CT Scanner System

• 72 Hz Vertical deflection frequency • Easy-to-use calcium scoring software Temperature Rate of Change: 3°C/hour max.
• Sync on green with integrated patient report generation
Relative Humidity Rate of Change: 5%
• LCD Color Monitor*: 18” diagonal width • Snapshot Option (B7710LS) RH/hour max.
• Scan control keyboard (English language) • CT Perfusion 2(B7700PL)
Power Requirements:
assembly with intercom speaker, • Smart Step (B7868WL) – SW
microphone and volume controls. The only facility input to the system is a 380 to
• Smart Step HW (B7868SM & B7868SN)
• Global modem to allow InSite connectivity 480 V nominal, 3 phase Delta or Wye, 50/60
• Advanced Vessel Analysis (B7868LA) Hz, 90 kVA service, 20 kVA average power;
• 3-Button Mouse main disconnect to be located within 5 feet
• CT Colonoscopy (B7868LC)
• 3-Button Trackball (1.5 m) of the PDU. The facility must also
• CardIQ2 (B7710LR) provide a protective disconnect device with
low voltage, low energy local and multi-point
• Remote 21 inch Color Monitor (B7530RC)
Image Networking: remote capability, in the line feeders to the
• Exide UPS (B7999PS) PDU.
• Standard auto-configuring rd
100BaseT/10BaseT Ethernet (UTP • Metal Free Table Interface for 3 Party Complete, detailed specifications of all power
connection) Attachments (B7999PM) requirements are available upon request. For
16
• LCD Monitor (B7858LC) most installations, the Lightspeed CT
• Direct network connection; multi-suite Scanner System does not require any power
ethernet card not required for gateway out of • GE Color Printer (E7014LA) conditioning equipment to be used in
suite conjunction with the PDU. Regulators are not
• Table Tray
• Protocols supported: recommended for use with this system. For
• DICOM 3.0 network send (one IP those sites with known large power line
address at a time) and receive,
Siting Requirements transients, a suppresser filter for the system
pull/query, and storage commitment computer and peripherals may be useful. In
Minimum Suite Size: general, suppresser filters are not
push;
recommended.
• AdvantageNet (GenesisNet) point-to- Overall Suite:
point send, receive and pull/query (no Compatible UPS options are available if
broadcast); 15.5 feet (4.74 m) wide x 19.5 feet (5.96 m) desired.
• InSite point-to-point; deep x 9 feet (2.75 m) high.
Cooling Requirements:
• TCP/IP (for system administration) Recommended Suite Size:
The cooling requirements do not include
Overall Suite:
Compatible Options cooling for the room lighting, personnel or
non-CT equipment present. Cooling
• A DICOM Print Interface is standard on the 19.5 feet (5.96 m) wide x 24.5 feet (7.49 m) requirements are listed by subsystem to allow
system. deep x 9 feet (2.75 m) high. planning for each room of the CT suite.
• English Keyboard (B7800KE) Exam Room: 13 feet (3.97 m) wide x 19.5 Cooling requirements are given for minimum,
feet (5.96 m) deep. recommended and growth allowance
• French UIF & Keyboard (Z62401F)
scenarios.
• German UIF & Keyboard (Z62401G) Control Room: 11 feet (3.36 m) wide x 12.5
feet (3.82 m) deep. • The minimum cooling figures assume
• Italian UIF & Keyboard (Z62401H) patient throughput of 3 patients per hour and
• Spanish UIF & Keyboard (Z62401S) Equipment Room: Not required. Power 75 scan rotations per patient.
Distribution Unit heat and noise output under
• Portuguese UIF & Keyboard (Z62401P) all operating conditions with all available • The recommended cooling requirements
options is low enough to allow inclusion in the assume patient throughput limited by the
• ConnectPro HIS/RIS Interface with tube cooling algorithm.
Performed Procedure Step (PPS) M4 patient scanning suite with the table and
compatibility (B7500LN) gantry. • The suite cooling can be sized for future
developments by using the growth
• Bar Code Reader (B7540RB) Temperature and Humidity: allowance figures. This cooling will
• Volume Viewer (B7540LD) accommodate more patients per hour and/or
Exam and Control Rooms: 60°-75°F (15°- potential future system enhancements.
• DentaScan (B7540LS) 26°C) at 30%-60% relative humidity (non-
condensing).
• SmartScorePro (B7850PL) – Gating SW
• Prospective or retrospective EKG-gated Equipment Room: If a separate equipment
cardiac imaging in a single breath-hold room is used to house the PDU, the allowable
• 250-300 msec temporal resolution temperature range is 60°-84°F (15°-29°C) at
30%-60% relative humidity (non-condensing).

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 16 With Xtream
CT Scanner System

Minimum Allowance
Subsystem Watts BTU/hr System Components Dimensions
Warranty
Gantry 5,440 18,600**
mm inches
Table 200 700 The published Company warranty in effect on
PDU 1,000 3,400 Gantry: the date of shipment shall apply. The
Company reserves the right to make changes.
Operator 2,165 7,400
Console Width: 2,230 87.6
All specifications are subject to change.
Optional Laser 800 2,730 Depth: 1,007 39.6
Camera
Height: 1,887 74.1
Regulatory Compliance
Recommended cooling values should not be
used for calculating system input power Table: This product is designed to comply with
requirements. applicable standards under the Radiation
Width: 610 24 Control for Health and Safety Act of 1968.
** Recommended Allowance: 7,150/24,400
Growth Allowance: 9,200/31,400 Length: 5,050 199 Laser alignment devices contained within this
product are appropriately labeled according to
Height: 1,120 44 the requirements of the Center for Devices
and Radiological Health.
Power Distribution Unit:
This product is a CE-compliant device which
Width: 762 30 satisfies regulations regarding Electro-
Magnetic Compatibility (EMC) and Electro-
Depth: 585 23 Magnetic Interference (EMI), pursuant to IEC-
601.
Height: 1,270 50

Operator Console:

Width: 1,238 48.7

Depth: 1,288 50.7

Height: 1,257 49.5

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 17 With Xtream
CT Scanner System

* Option

Downloaded from www.Manualslib.com manuals search engine


g
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:

LightSpeed16
Tokyo, Japan - Fax: 81 425 85 5490
Product Data Sheet – rev 4.2 Hong Kong - Fax: 852 2559 3588
July 4th 2003 GE Medical Systems-Europe
Page 18 With Xtream
CT Scanner System

Minimum Room Layout

Typical Room Layout

* Option