GE DIRECTION 5604324, REVISION 18 8-3-6 System Software - Installation Procedure ‘Two separate methods of software installation is offered in LOGIQ P7/P8/P9IP10. A). Software Load from DVD(R1/R2) or USB media (R3/R4/R4.5) B) Software Re-Load from Repository LoGia P7/P8/PS/P10 ‘SERVICE MANUAL ‘Situation > Handerive is new + Reload from Repository falls + Wish to perform ‘Clean in ~ Recinsalaton for application or operating system issues + Perform Software Load from DVD or USE media Reler to section 82-7 t Perform Software Reload Refer to section 8-3-8 Chapter 8 - Replacement Procedures 8:31 GE LOGIQ P7/PB/PSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 8-3-7 Software Installation from DVD 7-1 System Software - Installation Procedure (R1.x.x) 1) Prepare the Base System Software DVD. 2) Power ON the scanner and press "DEL"(If AN key is installed) or press left bottom key of trackball (lf there’s no AN Key) to enter BIOS Ad NOTICE there's no AN key, please use trackball fo move menu in BIOS mode and click left bottom key of trackball to select menu in BIOS and double-click to set or change the value in BIOS. Right bottom key of trackball is ESC’ in BIOS. Left bottom key of trackball to tnter B08 and selet nu en set Right bottom key of trackball or change the value in BIOS 3) Type “maru’ for password, 8:32 Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 4) Go to "Boot" in BIOS menu and change the setting value from “SATA 1 Drive" to "SATA 2 Drive" in "4st Boot Device’, change the setting value from "SATA 2 Drive" to “SATA 1 Drive" in ‘2nd Boot Device’ 5) Goto "Save & Exit’ in BIOS menu and select “Save Changes and Exit’. Then, the system will e- boot. 6) Press DVD-Drive eject button. 7) Place the Base System Software DVD on the tray. 8) Wait few moments for DVD drive to read DVD contents. Chapter 8 - Replacement Procedures 8:33 GE LOGIQ P7/PIPSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 98) The notification screen appears, and then hit the Enter key. NOTE: Use simulator A/N keyboard if customer's system does not have extemal A/N keyboard 10) Select either [1]"Load the complete disk’, or [2] "Load the bootable C: and repository Z: partition only" and hit the Enter key. 3-34 ‘Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 11) A confirmation window appears, and then hit the Enter key. 12) Approx. 40 minutes later, loading the software will be completed. 13) Remove Base System Software DVD, and type "exit" to reboot the system. 14) At the first system reboot after loading the software, the OS will reboot the system automatically. 18) After the reboot, the installation of system (Windows Embedded Standard will be fully completed. 16) Go to BIOS mode (Refer to the procedure 2) 3) above.) Chapter 8 - Replacement Procedures 8:35 GE LOGIQ P7/PIPSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 17) Go to “Save & Exit" in BIOS menu and select ‘Restore Defaults" 18) Go to “Boot’ in BIOS menu and check that ‘1st Boot Device’ is set to“SATA 1” and ‘2nd Boot Device’ is set to “SATA 2" 8:36 ‘Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 19) Go to "Save & Exit’ in BIOS menu and select “Save Changes and Exit”. Then, the system will boot. Chapter 8 - Replacement Procedures 8:37 GE LOGIQ P7/PB/PSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL ‘System Software - Installation Procedure (R2x.x) 41) Prepare the Base System Software DVD. 2) Power ON the scanner and press “DEL"(If AN key is installed) or press left bottom key of trackball (if there's no AN Key) to enter BIOS Ad. NOTICE Hthere's no AN key, please use trackball to move menu in BIOS mode and clic left bottom key of trackball to select menu in BIOS and double-click to set or change the value in BIOS. Right bottom key of trackball is ESC’ in BIOS. Left bottom key of trackball to enter BIOS and select menu and set or change the value in BIOS Right bottom key of trackball for ‘ESC’ in BIOS 3) Type ‘maru' for password. 8:38 Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 4) Go to "Boot" in BIOS menu and change the setting value from “SATA 1 Drive" to "SATA 2 Drive" in "4st Boot Device’, change the setting value from “SATA 2 Drive" to “SATA 1 Drive" in ‘2nd Boot Device’. 5) Goto "Save & Exit’ in BIOS menu and select “Save Changes and Exit”. Then, the system will reboot. 6) Press DVD-Drive eject button. 7) Place the Base System Software DVD on the tray. 8) Wait few moments for DVD drive to read DVD contents. Chapter 8 - Replacement Procedures 8:39 GE LOGIQ P7/PB/PSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 9) The notification screen appears, and then hit the Enter key. SOS ee Sn nn THIS PROCEDURE HAY RESULT IM COMPLETE PATIENT DATA LOSS USED CORRECTLY? PLEASE READ THE OPTION BELOM CAREFULLY \ PROCEEDING. This process ts MOT REVERSIBLE and should WOT b Martoat 0 NOT power off the spsten 1 Soepleteds Te will take. Less. than 15.1 Chis process is stopped for x a ain to ‘completion or else the system stopped once If you want to pro 14 with this process press the ‘Enter’ key to Continue with option i ‘CTRL-C" now to cb ha’ upRan tron the DVBRON dive end Press “CTRL aoe ae ae am and oft the austen to restart. it withou Se toe Sout dish drive's current contents NOTE: Use the trackball and left set key if the customer's system does not have an extemal A/N keyboard. 10) Select either [A] ‘Load the complete disk’, or [B] ‘Load the 'C partition only” and hit the Enter key. 8 lect ONE of the shioatagie ‘Base 258 selact ONE of the folloving options f stam Software Leed Inage’ onto gore ayatons, ott Nove using (UP EDOWM} CENTER] to se! t "Load the 'C* partition only ~ Patten monly = Patient dete is NOT Tost «Exit to the :\ PROMPT . 8-40 Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 11) A confirmation window appears, and then hit the Enter key. 412) Approx. 30-40 minutes later, loading the software will be completed 13) Remove Base System Software DVD, and hit the Enter key to reboot the system, 14) At the first system reboot after loading the software, the OS will reboot the system automatically. 45) After the reboot, the installation of system (Windows Embedded Standard 7) willbe fully completed. 16) Go to BIOS mode (Refer to the procedure 2.) 3.) above.) Chapter 8 - Replacement Procedures 841 GE LOGIQ P7/PIPSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 17) Go to “Save & Exit" in BIOS menu and select "Restore Defaults’ 18) Go to “Boot' in BIOS menu and check that “1st Boot Device’ is set to “SATA 1” and ‘2nd Boot Device’ is set to “SATA2", 842 ‘Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 19) Go to "Save & Exit’ in BIOS menu and select “Save Changes and Exit”. Then, the system will boot. Chapter 8 - Replacement Procedures 843 GE LOGIQ P7/PB/PSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL ‘Systm Software - Installation Procedure (R3.x.x or later) 8.3-7-3-1 Set the system boot order (BIOS setting) 1.) Prepare the THUMB DRIVER R3.x.x, Rd.x.x oF R4.5.x.x (USB media) including Base Image Load ‘Windows 10, 2.) Insert USB media to the system USB port Figure 8-26 System USB port 3.) Power on the scanner 4.) Press “DEL” (IFAN key is installed) or press left bottom key of trackball If there’s no AN Key) to center BIOS 5.) Enter the “maru’ in Enter Password Figure 8-27 Enter Password 8-44 Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 6.) In“Boot” page, Change "Boot Option #1” to USB memory which is inserted (R3.x.x) or change “1st Boot Device" to USB Harddisk (R4.xx or later). NOTE: The name of USB memory can be displayed different depend on the inserted USB memory (RX). Figure 8-29 “Boot” page 1 (R4.x.x or later) Chapter 8 - Replacement Procedures 845 GE LOGIQ P7/PIPSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 7.) After changing the “Boot Option#1” (R3.x.x) or “1st Boot Device" (R4.x.x), save the BIOS set and ‘exit by pressing 'F10' key. Figure 8-30 “Boot” page 2 (R3.xx) 8-46 ‘Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 8-3-7-3.2 Base and Application Software installation + Base Image Software Load Procedure 1.) The system will reboot automatically. 2.) Wait few moments for USB drive to read USB contents. 3,) The notification screen appears like below. 4) Select [1] Load the complete disk (FULL), and Press [OK] button using the trackball key. NOTE: If there is no physical keyboard option, use the external keyboard or on-screen menu. Figure 8-32 The notification screen 4 5.) confirmation window appears, and then Press [Yes] button using the trackball key. Figure 8-33 The notification screen 2 Chapter 8 - Replacement Procedures 847 GE LOGIQ P7/PIPSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 6.) The Installation progress screen appears like below. Figure 8-34. The notification sero: Ay WARNING Do not interrupt the software loading at any time. 7.) Approximately 20~ 30 minutes (R3.x.x) or 10~ 15 minutes (R4.x.x or later) later, loading the software will be completed. 8) After completing, remove USB media from the system USB port 9.) Press [OK] button using the trackball key to reboot the system. 10) Enter BIOS mode with the password "maru" and change "1st Boot Device" to NVMe Storage. (R4xx) Figure 8-35 “Boot” page 3 ( (R4.x.x or later) 848 ‘Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGia P7/P8/PS/P10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 11) After change the “1st Boot Device’, save the BIOS set and exit by pressing ‘F10’ key. (R4x.x.) 12) At the system reboot after loading the software, the OS will reboot the system automatically, 13) After reboot, the installation will be fully completed. NOTE: — Do NOT touch the system during this process. Activating the keyboard, mouse or front panel could corrupt the installation. While the script is running, several window or dialog boxes appear on the screen. At times it may look like the system is unresponsive. Chapter 8 - Replacement Procedures 8-49 GE LOGIQ P7/PB/PSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 837-33 Application Software Load Procedure 11) After the reboot again, the application software installation will start automatically. 2.) About few minutes, the following options will appear. Select your product. Click “Start Installation” button. Locia Please selec: your product ‘Wrong product selection nay leed system matunctions! a z sit stort Installation Figure 8-36 Click “Start Installation” 3.) Don't press any key and wait unti the software installation is done, re Figure 8-37 Installing application 4.) After the installation completes, the system will be rebooted automatically, 5.) SW License window may appear, then enter correct Option String and click "OK" NOTE: The HW/d shown in the picture is an example. Figure 8-38 SW License window 8-50 Section 8-3 - System Software - Installation/Upgrade Procedure LoGiQ P7/PBIP9IP10 GE ‘SERVICE MANUAL DIRECTION 5604324, REVISION 18 837-34 Base and Application Software installation for LOGIQ P7/P9 R2.5 to R3 upgraded system 4.) If LOGIQ P7/P9 R3.0.11 System Software or above is being installed to the R2.5 to R3 upgrade 2.5 to R3 upgrade system, follow the below instructions. 2) Insert the SW thumb drive to the USB port. 3) Reboot the system 4) Press “DEL (If AN key is installed) or press left bottom key of trackball there's no AN Key) to enter BIOS. 5) Enter the “maru’ in Enter Password. 6.) Select the install USB in Boot Option #1 of Boot menu. Figure 8-39 7.) Click "Restore User Defaults’ and then click "Yes". Figure 8-40 Chapter 8 - Replacement Procedures 851 GE LOGIQ P7/PIPSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 8.) Click “Save Changes" and then click “Yes. Figure 8-41 9) MUST force the shutdown by pushing the power button for § seconds. 10) After rebooting by forcing the shutdown, reinstall R3.0.11 or upgrade higher version by referring section System Software - Installation Procedure (R3.x.x oF later) on page 8-44 3-52 ‘Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 83-74 Application Software Installation Proparation (R1.x.x or R2xx) 41) Very the system boot up to Windows Screen 2). Prepare the Application Software DVD. 3) Press DVD-Drive eject button, and place the Application Software DVD on the tray. 4) Click Windows Explorer icon on the taskbar. ©! Service Br inlaws Explorer Ceo, 1 fies Ctrerty 20 1 owns aa i | 3 note TE owamens detent psa | Sis ian 6) Double click "LoadSoftware. bat”. Software Installation will begin. lie Name Otemedted | Type See «Fle Curenyon the Disc (2) Wace avane22 AM Flt Dlonticinereat ALA/A0670 PM _ Wink Boch ile 5 Chapter 8 - Replacement Procedures 8-53 GE LOGIQ P7/PB/PSIP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL Application Software Installation (R1.x.x or R2x.x) 1). Application Software Installation process will take approximately 20 minutes 2) Select your product, Click “Start Installation” button. 3) After the installation completes, the system will be rebooted automatically NOTE: Do not remove Application Software DVD from drive. (R1.x.x or R2.x.x) 4) After rebooting the system, the “Start Application” window appears. Remove the DVD from the tray (R1.xx or R2.xx), then click "Start" button ‘Start Application 8-54 Section 8-3 - System Software - Installation/Upgrade Procedure GE LoGiQ P7/PBIP9IP10 DIRECTION 5604324, REVISION 18 ‘SERVICE MANUAL 5) If SWLLicense window appears, then enter correct Option String and click “OK” Chapter 8 - Replacement Procedures 8-55

You might also like