Professional Documents
Culture Documents
O02
Revision 1.0
01.09.2021
3.0 QC Point
3.1 Output: Post-production activities
4.0 Reference
4.1 Inspection Work Instruction (PR-QC-PI03)
4.2 Sampling table (PR-QC-PD01)
4.3 Dispense order (PR-PR-PR02)
4.4 Item dispense (PR-WH-PR02)
5.0 Type:
5.1 In-House.
• Verify the colour and status referring to the Table 2 (Reagent colour
Colour and
and status)
status
• Check the size as follows:
Filling Weighted one vial/bottle with requested volume as standard to check the
size of filled vial/bottle by vision. The standard volume should follow
Reagent size as per requested in item dispend, for example 5mL volume The standard
vial needs to change weekly.
Note: The acceptance criteria: The volume level of filled vial should
even/same as the standard volume level.
Closing Closing cap • Closing the cap and invert the vial to check that the vial is closed well
Page 1 of 4
QC.O02
Revision 1.0
01.09.2021
8.3. QC Officer: Conduct biochemical inspection. The tests involved in biochemical inspection as per listed
below:
A. Inspection using controls
B. Inspection using febrile antiserum factors.
Page 2 of 4
QC.O02
Revision 1.0
01.09.2021
4. Mix the reagent and the control using a clean stirring stick over an area with a diameter of
approximately 20-40mm
5. Rotate the slide for respective times as Table 3 below:
6. Read the results immediately by observing any visible agglutination in good indirect light Record the
time of agglutination and read the result based on the acceptance criteria as Table 4 below.
7. Record the result on Febrile Inspection Form (QC.09F) and QC Sampling Inspection Form (PR-QC-
PR01). Attach them to production batch.
Note: For controls product, record the results on General Inspection Form (QC.02F) and QC
Sampling Inspection Form (PR-QC-PR01). Attach the two forms with related production batch.
B) Inspection using febrile antiserum factors (applicable for salmonella reagent only)
1. Add one drop of febrile on four circles of the white glass slide
2. Add the factors on each circle respectively.
Page 3 of 4
QC.O02
Revision 1.0
01.09.2021
3. Mix the reagent and the control using a clean stirring stick over an area with a diameter of approximately
20-40mm
4. Rotate the slide for 1 min.
5. Read the results immediately by observing any visible agglutination in good indirect light. Record the
time of agglutination and read the results based on the acceptance criteria as Table 5 below.
Page 4 of 4