Technical Installation Sheet : Steam Sterilizer Plusteam 4 - Double Door Right Model (Block 3F)

ESS * EL CA*
TURQUOISE BIOMEDICAL TECHNOLOGIES

+260
+1200

B 1165 B

1930
50 40 50 50

CW DW CA
DR

±0.00
A A

CW DW CA
DR

Section A Section B
445
374
190 100

430
470 430
525
CA 570 795
DR
795 CA* Wall
10
1163 DW 1163 Double Side
CW Bonding Tape
Sterilizer

EL EL
Stainless Steel Trim

140 140
1165 1165

Loading Side (Front) Loading Side (Front)


(Lower side view) (Upper side view)

The copyright on this drawing and all enclosures is reserved to Turkuaz Biyomedikal Teknolojiler at any time. This documentation is personal and entrusted to the consignee. Do not copy this drawing or parts thereof without prior
written approval by Turkuaz Biyomedikal Teknolojiler. Also it is not allowed to make it available to or to hand it over to third parties. TBT has various designs and several technical variants. There may be differences between the
same product types and various versions. This document is intended for the use of Distributors/Dealers/Customers of TBT Inc. Reproduction in whole or in part by any other party is prohibited. Specifications are given for reference
only and are subject to change without any previous notice. Accessory/Product images are for illustrative purposes only. rev. EN 003-2023 © 2015, TBT Inc. All rights reserved Drawing by GT.MOL

Kazım Ozalp Mah. Hafta Sok. No:23/2 info@tbtmedical.com.tr T. +90 850 840 8 828
06610 Cankaya / ANKARA / TURKEY www.tbtmedical.com.tr F. +90 312 911 25 28
Technical Installation Sheet : Steam Sterilizer Plusteam 4 - Double Door (Block 3F)

TURQUOISE BIOMEDICAL TECHNOLOGIES

Capacity Overall dimensions (mm) Weight (kgs.) Floor Load (N/m²) c

4 STU Height 1930 Net 910 4.20 x 10 ⁵


Witdh 1165 Gross 940
Depth 1163 Operating a 1044
Sampling b 960
a Stands for total device weight with a 100% full generator, including loading trolley and full capacity of STU baskets loaded as per EN 285 norms.
b Stands for total device weight with a 100% full generator.
c Ceiling load is calculated as per 4 rutile feet available and can be reduced by installing sterilizer directly on a frame ground as surface area will increase.

Connection Area
Connection size

Nominal value
Specifications
Symbol

Notes
CW Cold water <15oC, <2,0 mmol 1/1 from water softener Muff ¾" 3 - 5 bar Ground The top point of external valve must be max.
50 mm from ground level.
DW Reverse osmosis treated water acc. to EN 285 Muff ¾" 3 - 5 bar Ground The top point of external valve must be max.
50 mm from ground level.
DR Metal pipe or heat resistant plastic pipe up to 150°C Muff 1" Ground The top point of connection must be max. 40
(fixed) mm from ground level. For each device, a
separate drain must be installed. After 5
meters, lines can be combined with a 2"pipe or
suitable piping depending on device quantity.
Drain lines of multiple devices must not collide
to prevent backflow. Drain line should be
connected to main drainpipe with fewest
elbow connections possible to reduce the flow
resistance. Increased resistance in flow reduces
vacuum efficiency hence affects device
performance.
CA Compressed air, oil free 10 mm 5 - 7 bar Ground Quick Air Fitting Connector.
CA* * Optional Ceiling connection Top
EL 3L + N + PE, 380 / 400 V, 50 Hz (Standard) 5 x 16 mm Top Customers supply : Fuse ; 80A - Peak ; 30 kW
Main Electrical Panel : As per below chart.
Panel distance is max. 10 m from device and
with separate panels (fuses) for each device.
Notes : Cable must be 1,2 m above the ground
level. Device must have a separate grounding
line.
AL Air outflow from service room: heat flow to be 1.5 kW Top Customers HVAC System
dissipated, at T = 30 °C, ∆T=10 °C 450mn³/h

ESS House / External Steam Supply Connection Muff ½" min. Ground This connection is only available with optional
(Optional) 4 bar features "Ref. No: 512037 - Apparatus for
Steam quality should be according to EN 285 Connection to in Build Hospital Steam - No
Steam Generator" or "Ref. No: 512038 -
Double Selection Connection for Steam
Connection - Generator + Hospital Steam".

The customer shall provide all supply connection lines with a built in Ball valve (parallel to the floor ). All connection assemblies between the
sterilizer and supply / exhaust lines (customers supply) are designed with flexible, pressure resistant tubes made of 316L material: in that way,
mechanical tolerances of +/- 25 mm in any direction can be easily compensated which results in short assembly times. Only exception is the
External steam supply " ESS " which is realized in a rigid manner. When the exhaust line will be neither connected nor led outside the tubing
ends up in a bend oriented upwards.

The copyright on this drawing and all enclosures is reserved to Turkuaz Biyomedikal Teknolojiler at any time. This documentation is personal and entrusted to the consignee. Do not copy this drawing or parts thereof without prior
written approval by Turkuaz Biyomedikal Teknolojiler. Also it is not allowed to make it available to or to hand it over to third parties. TBT has various designs and several technical variants. There may be differences between the
same product types and various versions. This document is intended for the use of Distributors/Dealers/Customers of TBT Inc. Reproduction in whole or in part by any other party is prohibited. Specifications are given for reference