ISSN 1693-1831
Abstract: Curcumin is a lipophilic compound which suffers from the poor bioavailability after oral
administration. Increasing its dissolution rate can be a successful strategy to improve the bioavailability.
Along with the formulation developments, a rapid and simple analytical method to determine curcumin
concentrations in the dissolution medium is required. The aim of this study was to develop and to
validate an analytical method based on thin layer chromatography (TLC) to determine curcumin
concentrations in a dissolution medium containing 0.5% w/v sodium lauryl sulfate (SLS) and 20 mM
sodium phosphate buffer. A polyvinylpyrrolidone K30 based solid dispersion of Curcuma longa extract
and its corresponding physical mixture were dissolved in a medium containing 0.5% w/v SLS and
20 mM sodium phosphate buffer (pH 6.0). Dissolution samples were spotted on a normal TLC plate
and eluents of various compositions were evaluated. The retardation factor (Rf), resolution (Rs), and
asymmetry factor (As0.05) of the optimized method were determined. Using the optimized eluent,
proper separation of curcumin peak was achieved with an Rf of 0.50, Rs of 2.62 and As 0.05of 0.87.
Linearity (5-30 µg/mL) was demonstrated by r value of 0.9965. The TLC method provided precision
with RSD ≤3.50 and accuracy with recovery value of 94-105%.
Keywords: Curcuma longa, curcumin, solid dispersion, dissolution, thin layer chromatography.
Kata kunci: Curcuma longa, kurkumin, dispersi padat, disolusi, kromatografi lapis tipis.
* Penulis korespondensi,Hp. 082135993494
e-mail: dewisetya03@yahoo.com
148 SETYANINGSIH ET AL. Jurnal Ilmu Kefarmasian Indonesia
susceptible to its instability for a longer time. e.g. they method was first employed to analyze pure curcumin/
suggested that when using buffers and ethyl acetate curcuminoids dissolved in methanol. Next, the method
for extraction, polymerization of curcumin can occur was applied to dissolution samples containing 0.5%
at the interface of the two phases(21). w/v SLS and 20 mM sodium phosphate buffer. As
Thin layer chromatography (TLC) has been widely controls, a solution of C. longa extract and SLS in
used particularly in phytochemistry. Compared to mixture of methanol and water at a volume ratio of
HPLC, this method is relatively efficient and cheaper. 92/8 and a solution of C. longa extract and sodium
TLC provides an opportunity to simultaneously phosphate in mixture of methanol at a volume
analyse a large number of samples. The separation ratio of 80/20 was analyzed. The weight ratios of
of curcuminoids by TLC/high performance thin curcumin/SLS and curcumin/sodium phosphate in
layer chromatography (HPTLC) has been discussed these solutions were adjusted to 1/31 and 1/19.5,
in several publications(22-25). However, none of the respectively, which is identical to the ratios found in
studies determined curcumin concentrations in the dissolution samples when all curcumin would have
presence of the other two curcuminoids in aqueous been dissolved. The separation was performed on
dissolution media. In order to tackle poor curcumin a precoated silica gel aluminium plate 60F254 (TLC
bioavailability, the availability of an efficient and plate) which was activated at 90 °C for 5 minutes prior
validated method for curcumin quantification is crucial to analysis. In brief, 10 µL of the dissolution samples
during early screening of formulations, which can were spotted in the form of bands of width 6 mm with a
serve to optimize the dissolution behavior. 100 µL CAMAG microliter syringe on the TLC plate.
In the current study, a TLC method for curcumin On the same plate, 10 µL of a 15 µg/mL curcumin
quantification from samples collected directly from standard solution was spotted. Camag Linomat V
the dissolution experiment, thus containing SLS (CAMAG, Muttenz, Switzerland) connected to
and sodium phosphate buffer, was developed and nitrogen gas was used for the spot application at a
validated. An established TLC method for curcumin constant rate of 30 nL/s, with pre-dosage volume of
quantification(24) was used as a starting point. The 0.2 µL, and distance between the bands were 10 mm.
presence of dissolution media containing SLS and The settings were maintained throughout the analysis.
sodium phosphate buffer in the dissolution samples, The TLC plates were developed in ascending order in a
however, reduced the separation efficiency of closed CAMAG 20 x 20 cm glass flat bottom chamber
curcumin from the other two curcuminoids. Therefore, which was pre-saturated with mobile phase for at least
the aim of this study was to develop and validate a 30 minutes. To maintain the chamber saturated with
TLC based method for the determination of curcumin vapour of the mobile phase, the wall of inner side
concentraion in a dissolution medium containing 0.5% of the chamber was lined with filter paper. The TLC
w/v of SLS in 20 mM sodium phosphate buffer. plate, which was spotted with samples, was developed
to a distance of 10 cm from the spotting zone. The
MATERIAL AND METHOD developed plate was dried at room temperature after
which the TLC bands were visually inspected using
MATERIAL. Curcumin standard was purchased from UV light (366 nm; CAMAG) to evaluate the peak
Nacalai, Japan. C.longa extract (98.3% curcuminoids separation. The spots densities were densitometrically
as mentioned in the Certificate of Analysis) was scanned sequentially using a CAMAG TLC Scanner
kindly given by PT. Phytochemindo Reksa, Bogor, 3 at 424 nm. The scanning speed was 20 mm/s with
Indonesia. The extract contained 67.9% curcumin the slit dimension of 8 mm x 0.2 mm. All the TLC
as quantified in our lab using a validated HPLC analysis was carried out under ambient conditions,
method(26). Polyvinylpyrrolidone K30 (PVP K30) i.e. 20±2 °C and a relative humidity of 55%. The data
was a gift from Zenith Pharmaceuticals, Semarang, were collected and the obtained peaks were integrated
Indonesia. De-ionized water was prepared in our with filter factor of Savitsky-Golay 7 using winCATS
laboratory. SLS was obtained from Merck, Darmstadt, software V1.4.4.
Germany. Precoated silica gel aluminium plate 60F254 Mobile Phase Optimization. In the TLC method
(TLC plate) (20x20 cm and a thickness of 0.2 mm; described previously(24), a mobile phase consisting
were purchased from Merck, Darmstadt, Germany. of chloroform/ethanol/water (at a volume ratio of
All solvents were analytical grade and obtained from 50:1.92:0.08) was used. Although the TLC system was
Merck, Darmstadt, Germany. suitable for solutions of pure curcumin/curcuminoids
METHOD. TLC Instrument and Conditions. in methanol, it failed for curcumin determination in
The chromatographic separation was based on the aqueous solutions containing 0.5% w/v SLS and 20
method described by Tonnesen and Karlson(24). The mM sodium phosphate buffer (pH 6.0). To be able
150 SETYANINGSIH ET AL. Jurnal Ilmu Kefarmasian Indonesia
and PVP K30 was used as a control. The PM was an appropriate separation of the curcuminoids in
prepared by gently mixing C. longa extract with PVP dissolution samples of solid dispersions or physical
K30 in a mortar following sieving using a mesh size mixtures containing C.longa extract. The mobile phase
60. The curcumin content in PM formulation was composition employed by Tonnesen and Karlson(24),
20 wt-%. The powder was kept in a desiccator for at which was chloroform/ethanol/water at a volume
least one day. ratio of 50:1.92:0.08, was used as a starting point.
For dissolution studies, gelatine capsules of size The dissolution samples were spotted without prior
00 were filled with approximately 400 mg of either extraction of the curcumin.
SD or PM. The dissolution studies were carried out on
a SOTAX AT-7 dissolution tester apparatus II (Sotax
AG, Basel). The medium consisted of 0.5% w/v SLS curcumin curcumin
medium used in these experiments (aqueous solution chromatographic separation of curcuminoids from
containing 0.5% SLS and 20 mM sodium phosphate our dissolution samples using chloroform/ethanol/
buffer) which was different from the one used by water at 50:1.92:0.08 mL as mobile phase can be
Tonnesen and Karlson (methanol). When water in the ascribed to the dissimilarity of the used solvent
mobile phase was replaced by 0.08 mL of 1% acetic (matrix) of this dissolution samples with the one
acid, the peak separation of curcuminoids from the used by Tonnesen and Karlson (1986). Since the
dissolution sample of SD C.longa extract was slightly dissolution medium contained 0.5% w/v SLS and 20
improved (data not shown). However, when water was mM sodium phosphate buffer, it can be calculated
replaced by 2% acetic acid at various volumes (i.e. that 500 mL medium contains 2.5 g SLS and 1.56 g
0.12 - 0.18 mL), the quality of peak separation in the sodium phosphate buffer. As approximately 400 mg
chromatograms was much better (Figure 2B-D). The of C. longa extract powder with 20 wt-% drug load
best peak separation was achieved using 0.15 mL of was used in the dissolution study, maximally 80 mg
2% acetic acid (see Figure 2C). Using this mobile of curcumin dissolved in the bulk medium. Thus, the
phase, which thus consisted of chloroform/ethanol/2% minimal weight ratio of SLS and sodium phosphate
acetic acid of 50:1.92:0.15 mL, the shape of the buffer to curcumin in sample were 31:1 and 19.5:1,
curcumin peak was sharper and less diffused than respectively, implying that the samples spotted onto
in the other chromatograms. The resulted Rs value a TLC plate contained a substantial amount of SLS
was 2.52, which is considered to be acceptable(30). and sodium phosphate buffer. Contrary to that, in
Further increase of the amount of 2% acetic acid to the study of Tonnesen and Karlson (1986), only the
0.18 mL did not provide any better peak separation as curcuminoids were present in spotting zone as the
shown in Figure 2D. Thus, further experiments were solvent, methanol, was fully evaporated. Therefore,
conducted by using chloroform/ethanol/2% acetic acid we conducted additional TLC analysis to clarify the
of 50:1.92:0.15 mL as mobile phase. effect of the sample matrix.
Separation of compounds by a TLC system The Influence of Dissolution Matrix on The
is influenced by the combination of 1) the mobile Separation of Curcuminoids Peaks. To investigate
phase composition that transports the compounds the effect of 0.5 w/v SLS and 20 mM sodium phosphate
to be separated and 2) the physical and chemical buffer on the peak separation of curcuminoids, TLC
interaction between the compound with the stationary analysis of C. longa extract dissolved in either 0.5%
phase(30). Too strong interaction of compounds with w/v SLS in methanol or 20 mM sodium phosphate
the stationary phase relative to the eluent can cause buffer in methanol was carried out. Due to their limited
a tailing effect, which may lead to a poor peaks solubility in methanol, SLS or sodium phosphate
separation. Tonnesen and Karlson (1986) studied buffer were first dissolved in small amount of water
the interaction of curcumin with (soluble) silicic and then added to an excess of methanol. The volume
acid which was used to mimic the (solid) stationary ratios of water/methanol required to prepared clear
phase, silica gel, as both contain silanol groups. They solutions of 0.5% SLS and 20 mM sodium phosphate
found that the 1-3 diketone groups of curcuminoids buffer were 8/92 and 20/80, respectively.
can interact with silicic acid: using UV-Vis spectra, Two mobile phase systems which consisted of
a bathochromic shift and an increase of Ɛ value was chloroform/ethanol/water of 50:1.92:0.08 mL and
observed when a silicic acid solution was added to chloroform/ethanol/2% acetic acid of 50:1.92:0.15
1-3 diketone compounds including curcuminoids, mL were used to develop the TLC plate. The samples
indicating intermolecular H bonding. contained SLS-curcumin and sodium phosphate-
Although intermolecular interaction of curcumin at identical ratios to those found in the
curcuminoids at the 1-3 diketone sites with silanol dissolution samples when the formulations were fully
of the stationary phase is important, a successful dissolved.
chromatographic separation of curcuminoids also In Figure 3, it is demonstrated that elution of
depends on their interaction with the mobile phase. the sample containing SLS-curcumin and sodium
There is a difference in polarity among curcuminoids phosphate-curcumin, with two mobile phase
compound which causes a specific interaction to systems, generated different chromatogram profiles.
a certain mobile phase system and leads to peaks Chloroform/ethanol/water elution resulted in a poor
separation. In addition to that, it is noted that peaks peak separation of curcumin/curcuminoids (Figure
separation is also influenced by other component(s), 3A, B). In contrast, similar C. longa extract sample
which might present in the TLC spotting zone e.g. the eluted with chloroform/ethanol/2% acetic acid
sample matrix(34). (50:1.92:0.15 mL) resulted in an excellent separation
Therefore, we hypothesized that the poor of the three curcuminoids peaks (Figure 3C,D).
Vol 14, 2016 Jurnal Ilmu Kefarmasian Indonesia 153
curcumin curcumin
A B
curcumin curcumin
the sample from the time sampling of 120 minutes was construct a calibration curve, solutions of curcumin
analyzed (Figure 5C). In Figure 5C, chromatogram standard were prepared by spiking of the blank sample
shows that the curcumin peak is clearly separated from (dissolution medium) with curcumin stock solution
the adjacent peak. The Rf value of curcumin was found (950 µg/mL in methanol) to achieve concentrations of
to be 0.50 (Figure 5C) which is in agreement with the 5, 10, 15, 20, 25, 30, and 35 µg/mL. The calibration
Rf value of curcumin standard spiked in blank sample samples were analyzed by the developed TLC method.
(Figure 5B). The value of As0.05 was 0.87 and the Rs Calibration curve was obtained by plotting peak
value appeared to be 2.63 showing an excellent peak area as a function of the concentration (Figure 7).
separation of curcumin from the other curcuminoids. 25000
Two peaks instead of three typical C.longa peaks
appeared in the chromatogram presented in the Figure 20000
% curcumin dissolution
Requirements, acceptance criteria for RSD are
concentration dependent, e.g.at categorized level of 60
for the 10 µg/mL can be below 7.3%. As shown in Figure 8. Dissolution profile of SD C. longa extract and its
corresponding PM formulation. Data is presented as mean
Table 2, the RSD for intraday analysis was less than
and SD of n=3.
3.50% and for interday was less than 2.62%. Thus,
RSDs obtained for the intraday and interday analysis
of all investigated concentrations were below 5.3% As can be seen in Figure 8, curcumin dissolved
which indicates that precision of the method fulfils faster from the SD formulation than from the PM
the requirements. formulation. Further assessment of dissolution
Accuracy. Accuracy is a measure of closeness performance was made by calculating the percentage of
between the measured value and the true value. Table dissolution efficiency (DE). DE provides information
2 shows the recovery values of the accuracy analysis for comparing dissolution behavior of different
for each concentration level (10, 20, 30 µg/mL). The formulations. The DE values were calculated up to
intraday and interday accuracy analysis resulted in 120 minutes and it was found that the DE value was
mean recovery of 94.00-102.60% and 97.54-105.00%, almost 4 times higher for the SD formulation (58.03 ±
respectively. According to the AOAC, the requirement 8.49%) than for the PM formulation (15.21 ± 1.60%)
for the mean recovery at a concentration of 100 (p<0.05).
ppm is 90-107%, while at a lower concentration, a The mechanisms of dissolution rate enhancement
broader range is acceptable(36). Thus, both the intraday of solid dispersions have been discussed elsewhere.
and interday recovery found in this study meet the Increasing specific surface area by decreasing particle
requirement of accuracy. size of drugs and improved wetting as result of the
intimate contact with hydrophilic carrier are the
Table 2. Intra and interday precision and accuracy of important keys in the dissolution enhancement of
curcumin (n=3). solid dispersions(37, 38). When the intimate contact is
Nominal Intraday Interday accompanied with a molecular interaction between
drug and carrier, the enhanced dissolution rate is
concentration Found Precision Accuracy Found Precision Accuracy further facilitated. It is well known that PVP interacts
(µg/mL) (µg/mL) (%RSD) (%recovery) (µg/mL) (%RSD) (%recovery) with many different lipophilic drugs on a molecular
10 9.40 3.50 94.00 9.75 2.62 97.54 level. E.g. Srinarong et al. reported that the interaction
between PVP K30 with a lipophilic drug (diazepam)
20 20.52 3.03 102.60 21.00 0.38 105.00
contributed significantly to the high dissolution rate
30 29.30 1.16 97.67 30.12 2.08 100.42 of drug. However, when they used other carriers
where the interaction was absent, dissolution was
LOD and LOQ. Detection limit is necessary to slower(39). Thus, although not investigated in this study,
be established especially during evaluation in early interaction of PVP K30 with curcumin on a molecular
sampling. The LOD and LOQ value calculation based level in the SD might have occurred, which might have
on standard deviation of response and slope was 1.3 contributed to the enhanced dissolution rate.
µg/mL and 4.0 µg/mL, respectively.
Application of Validated Method on The CONCLUSION
Estimation of Curcumin Dissolution. The validated
TLC method was applied for the determination of In conclusion, the TLC method developed in this study
curcumin from dissolution samples without prior is applicable to determine curcumin concentrations
separation e.g. sample extraction procedures from from dissolution samples containing buffer and
the dissolution medium. The dissolution profiles of SLS with acceptable precision (RSD), accuracy
curcumin (from the source of C.longa extract), from (recovery) and linearity. Although not subjected to
PVP K30 based SD and PM are shown in Figure 8.
156 SETYANINGSIH ET AL. Jurnal Ilmu Kefarmasian Indonesia
1985.180(5):402-4.
22. Ansari MJ, Ahmad S, Kohli K, Ali J, Khar RK. 30. Spangenberg B, Poole CF, Weins C. Quantitative
Stability-indicating HPTLC determination of thin layer chromatography: A practical survay.
curcumin in bulk drug and pharmaceutical Heidelberg, Dordrecht, London, New York:
formulations. J Pharm Biomed Springer;2010.
Anal. 2005.39(1-2):132-8. 31. Rashmin P, Mrunali P, Nitin D, Nidhi D, Bharat P.
23. Pathania V, Gupta AP, Singh B. Improved HPTLC HPTLC method development and validation:
method for determination of curcuminoids from strategy to minimize methodological failures. J
Curcuma longa. J Liq Chromatogr Relat Technol. Food Drug Anal. 2012.20(4):794-804.
2006-2014.29(6):877-87. 32. Ferenzi-Fodor K, Vegh Z, Nagi-Turak A.
24. Tonnesen H, Karlsen J. Studies on curcumin Validation and quality assurance of planar
and curcuminoids VII. Chromatographic chromatographic procedures in pharmaceutical
separation and quantitative analysis of curcumin analysis. J AOAC Int. 2001.84(4):1265-76.
and related compounds. Z Lebensm Unters 33. Khan KA. The concept of dissolution efficiency.
Forsch. 1986.182(3):215-8. J Pharm Pharmacol. 1975.27(1):48-9.
25. Zhang JS, Guan J, Yang FQ, Liu HG, Cheng XJ, 34. Wall PE. Thin-layer chromatography a practical
Li SP. Qualitative and quantitative analysis of four and modern approaches. Cambridge, UK: The
species of curcuma rhizomes using twice Royal Society of Chemistry; 2005.
development thin layer chromatography. J Pharm 35. Ferenzi-Fodor K, Renger B, Vegh Z. The frustated
Biomed Anal. 2008.48(3):1024-8. r e v i e w e r- r e c u r r e n t f a i l u r e s i n m a n u s c r i p t
26. Setyaningsih D, Murti YB, Martono S, Hinrichs describing validation of quantitative TLC/HPTLC
WLJ, Hertiani T, Fudholi A. A novel reversed procedures of analysis of pharmaceutical. JPC.
phase high performance liquid chromatography 2010.23(3):173-9.
method to accurately determine low concentration 36. AOAC. Official Methods of Analysis of AOAC
of curcumin in rat plasma. International Interntional. In: George W. Latimer J, editor.
Journal of Pharmaceutical and Clinical Research. Guidelines for Standard Method Performance
2016.8(5Suppl):377-86. Requirements. 19th ed. Gaithersburg, Maryland,
27. Ermer J. Performance parameters, calculations USA: AOAC International; 2012. p. Appendix
and tests. In: Ermer J, Miller JHM, editors. Method F, p9.
validation in pharmaceutical analysis. Weinheim: 37. Leuner C, Dressman J. Improving drug solubility
Willey-VCH Verlag GmbH & Co. KGaA; 2005. for oral delivery using solid dispersions. Eur J
p. 80-99. Pharm Biopharm. 2000.50(1):47-60.
28. Gustavo Gonzalez A, Angeles Herrador M. 38. Craig DQ. The mechanisms of drug release from
A practical guide to analytical method solid dispersions in water-soluble polymers. Int J
validation, including measurement uncertainty Pharm. 2002.231(2):131-44.
and accuracy profiles. Trends in Analytical 39. Srinarong P, Kouwen S, Visser MR, Hinrichs WL,
Chemistry. 2007.26(3):227-38. Frijlink HW. Effect of drug-carrier interaction on
29. ICH. ICH Harmonised tripartite guideline. the dissolution behavior of solid dispersion tablets.
Validation of analytical procedures: Text and P h a r m D e v Te c h n o l . 2 0 0 9 . 1 5 ( 5 ) : 4 6 0 - 8 .
methodology ed: Geneva; 2005.