You are on page 1of 11

CONCISE CLINICAL REVIEW

Reverse Triggering during Controlled Ventilation


From Physiology to Clinical Management
Antenor Rodrigues1, Irene Telias1,2,3, L. Felipe Damiani4, and Laurent Brochard1,2
1
Keenan Centre for Biomedical Research, Li Ka Shing Knowledge Institute, Unity Health Toronto, Toronto, Ontario, Canada;
2
Interdepartmental Division of Critical Care Medicine, University of Toronto, Toronto, Ontario, Canada; 3Division of Respirology,
Department of Medicine, University Health Network and Sinai Health System, Toronto, Ontario, Canada; and 4Departamento Ciencias
de la Salud, Carrera de Kinesiologıa, Facultad de Medicina, Pontificia Universidad Catolica de Chile, Santiago, Chile

Abstract asynchronous efforts. Diagnosing reverse triggering might be


challenging and can easily be missed. Inspection of ventilator
Reverse triggering dyssynchrony is a frequent phenomenon waveforms or more sophisticated methods, such as the electrical
recently recognized in sedated critically ill patients under activity of the diaphragm or esophageal pressure, can be used for
controlled ventilation. It occurs in at least 30–55% of these diagnosis. The occurrence of reverse triggering might have
patients and often occurs in the transition from fully passive to clinical consequences. On the basis of physiological data, reverse
assisted mechanical ventilation. During reverse triggering, patient triggering might be beneficial or injurious for the diaphragm and
inspiratory efforts start after the passive insufflation by the lung, depending on the magnitude of the inspiratory effort.
mechanical breaths. The most often referred mechanism is the Reverse triggering can cause breath-stacking and loss of
entrainment of the patient’s intrinsic respiratory rhythm from the protective lung ventilation when triggering a second cycle. Little
brainstem respiratory centers to periodic mechanical insufflations is known about how to manage patients with reverse triggering;
from the ventilator. However, reverse triggering might also occur however, available evidence can guide management on the basis
because of local reflexes without involving the respiratory rhythm of physiological principles.
generator in the brainstem. Reverse triggering is observed during
the acute phase of the disease, when patients may be susceptible Keywords: reverse triggering; mechanical ventilation;
to potential deleterious consequences of injurious or dyssynchrony; monitoring

Definition instead of the patient triggering the there is a lack of strong clinical evidence on
ventilator. It often occurs with regularity, the how to manage this asynchronous breathing
Reverse triggering is a frequent and patient being entrained by the ventilator (i.e., pattern.
intriguing phenomenon frequently observed reverse triggering) (Figure 1) (1). It can have
in sedated mechanically ventilated patients different consequences (Table 1), such as
under controlled ventilation, such as assist- breath stacking and loss of protective lung Prevalence During Controlled
control ventilation with only mandatory ventilation when triggering a second Ventilation
breaths or synchronized intermittent mandatory breath (2–4), but it may also offer
ventilation with mandatory breaths. The some protective effects, such as being The prevalence of reverse triggering in
apparent pattern suggests that the ventilator protective against diaphragm disuse atrophy mechanically ventilated patients (i.e., the
triggers respiratory efforts from the patient (4). Recognized only recently in the ICU (1), proportion of patients within the

(Received in original form August 5, 2022; accepted in final form December 5, 2022)
Author Contributions: All authors contributed to the conception of the work. All authors contributed to drafting the manuscript or critically revised
it for important intellectual content, approved the final version of the manuscript, and took responsibility for the integrity of the data.
Correspondence and requests for reprints should be addressed to Antenor Rodrigues, Ph.D., St. Michael’s Hospital, Room 4-709,
36 Queens Street E, Toronto, M5B 1W8, Ontario, Canada. E-mail: antenor.rodrigues@unityhealth.to.
This article has an online supplement, which is accessible from this issue’s table of contents at www.atsjournals.org.
CME will be available for this article at https://shop.thoracic.org/collections/cme-moc/ethos-format-type-journal.
Am J Respir Crit Care Med Vol 207, Iss 5, pp 533–543, Mar 1, 2023
Copyright © 2023 by the American Thoracic Society
Originally Published in Press as DOI: 10.1164/rccm.202208-1477CI on December 5, 2022
Internet address: www:atsjournals:org

Concise Clinical Review 533


CONCISE CLINICAL REVIEW

Figure 1. Signals to identify reverse triggering. (A) Patient on pressure-control ventilation triggering every mechanical insufflation. (B) Patient on
pressure-control ventilation with midcycle reverse triggering at a 1:1 entrainment ratio; reverse triggering occurring after the fifth mechanical
insufflation triggered an additional mechanical insufflation before complete exhalation and caused breath stacking (horizontal black bracket).
(C) Patient on volume-control ventilation with midcycle reverse triggering at a 1:2 entrainment ratio. The gray area in B and C shows the
difference between the onset of the mechanical insufflation and the onset of the electric activity of the diaphragm (EAdi) and esophageal
pressure (Pes) activity (respectively). a: Absence of drop in the pressure at the onset of the mechanical insufflation, suggesting an absence of
effort at the onset of the mechanical insufflation. b: Absence of an increase in the EAdi signal at the onset of the mechanical insufflation,
indicating the absence of effort at the onset of the mechanical insufflation. c: Drop in the pressure–time curve during the inspiratory phase of the
mechanical ventilator because of reverse triggering-associated effort occurring during pressure-control ventilation. d: Increase in the flow–time
curve during the inspiratory phase of the mechanical ventilator because of a reverse triggering effort occurring during pressure-control
ventilation. e: Deformation of the flow–time curve during the expiratory phase of the ventilator because of a reverse triggering-associated effort

534 American Journal of Respiratory and Critical Care Medicine Volume 207 Number 5 | March 1 2023
CONCISE CLINICAL REVIEW

Table 1. Criteria Associated with Potentially Favorable or Harmful Effects of Reverse Triggering and Potential Approaches for
Management

Reverse Triggering Potential Approach for


Characteristic Ventilation Mode Potentially Favorable Effect Potentially Harmful Management

Timing: during Pressure control Mitigate diaphragm disuse *Distending Modify set inspiratory time,
inspiration atrophy. transpulmonary tidal volume, and rate.†
pressure
*Increasing tidal volume Assess for the presence
of IRR and consider
transition to
spontaneous mode.
Timing: during Volume control Mitigate diaphragm disuse *Pendelluft Modify set inspiratory time,
inspiration atrophy. tidal volume, and rate.†
*Increasing regional Assess for the presence
distension of IRR and consider
transition to
spontaneous mode.
Timing: during late Pressure or volume Postinspiratory diaphragm Pendelluft Modify set inspiratory time,
inspiration and/or control activity can reduce lung tidal volume, and rate.†
expiration collapse and improve Eccentric diaphragm Assess for the presence
oxygenation. contractions of IRR and consider
Breath stacking transition to
spontaneous mode.
Degree of effort Pressure or volume Small efforts can mitigate Large efforts can Assess/manage inspiratory
control diaphragm disuse atrophy. contribute to drive stimuli (e.g.,
diaphragm injury concentration of CO2).
Assess for the presence
of IRR and consider
transition to
spontaneous mode.
Frequency Pressure or volume Patients with a higher rate of — Assess for the presence
control reverse triggering had of IRR and consider
better oxygenation and transition to
were more likely to spontaneous mode.
progress to an assisted
mode or be extubated
within the following 24 h.

Definition of abbreviation: IRR = intrinsic respiratory rate.


*Should be increased.

The main concept is not to mandate a rate superior to the patient’s intrinsic rate.

population studied presenting any amount Prevalence of Reverse Triggering 55% (Table 2) (5–12). Yet, these observations
of reverse triggered breaths during the Dyssynchrony in Patients under may have underestimated the phenomenon,
observed period) is influenced by the Controlled Ventilation given the short duration of the recordings.
duration of the observation, the diagnostic The overall prevalence of reverse triggering Most of these studies were developed in
method, and the ICU population (3, 5–12) dyssynchrony is high in sedated patients under patients undergoing lung-protective ventilation,
(Table 2). controlled ventilation, at least between 30% and with low tidal volumes (i.e., around 6 ml/kg

Figure 1. (Continued ). that persisted throughout the expiratory phase during both (B) pressure- and (C) volume-control ventilation. f: Variation in
the plateau pressure on the pressure–time curve of the ventilator because of a reverse triggering during volume-control ventilation. g: Positive
change in the Pes signal at the onset of a mechanical insufflation, indicating passive insufflation (note this change can be subtle or not visible
in some cases [e.g., because of cardiac artifacts]). The horizontal solid line indicates the expiratory time is equal to that set on the ventilator,
indicating the mechanical insufflation was triggered by the ventilator. The horizontal dashed line indicates that the expiratory time was shorter
than that set on the ventilator, and the mechanical insufflation was triggered by the patient’s inspiratory effort associated with reverse
triggering. The vertical dashed line indicates the onset of the EAdi activity that occurred before the onset of the mechanical insufflation,
indicating the mechanical insufflation was triggered by the patient. *Expiratory flow–time curve of the ventilator in the absence of a patient
inspiratory effort. †Plateau pressure on the pressure–time curve during volume-control ventilation in the absence of a patient inspiratory
effort. ‡Drop in the pressure–time curve of the ventilator at the onset of the mechanical insufflation, indicating a patient inspiratory effort
was present at the onset of the mechanical insufflation and triggered the mechanical insufflation. EAdi = electric activity of the diaphragm;
PAW = airway pressure.

Concise Clinical Review 535


536
Table 2. Summary of the Characteristics of Studies Reporting the Prevalence of Reverse Triggering in Patients under Controlled Ventilation

Duration of Recordings
Prevalence Prevalence Each per D Recorded D
of RT of Mode of Recording, per Patient, per Patient, Method of Degree of
Study n Dyssynchrony RT Breaths Population Ventilation min n n Detecting RT Sedation

Pham, et al. 109 53% of NR Hypoxemic Pressure or 20 to 40 2 to 3 1 to 7 Visual inspection of NR


the patients respiratory failure volume C ventilator
(PF ,200 mm Hg) waveforms

Rodriguez, et al. 100 50% of Median ARDS Volume C 30 1 1 Automated algorithm RASS 25 to 24
the patients (IQR, 25% to 75%) (PF ,300 mm Hg) on the basis
4.8 (0.03 to 4.3) of ventilator
per min per patient waveforms
or 17.7%
(0.95% to 49.5%)
of the patient’s
mechanical insufflations
Su, et al. 56 30% of 4% of all mechanical ARDS (PF ,300 mm Hg) PRVC, or NR 1 1 Visual inspection NR
the patients insufflations recorded pressure C of Campbell
in all patients diagram with
Pes signals

Baedorf Kassis, 55 45% of 12.1% of all mechanical ARDS (PF ,300 mm Hg) PRVC, or 4.5 to 8.5 1 1 Visual inspection of NR
et al. the patients insufflations recorded pressure C Campbell diagram
in patients with Pes signals
presenting RT
Sklar, et al. 55 33% of NR General ICU patients Volume or 60 1 1 Automated algorithm SAS 1 to 4
the patients in the 24 h preceding pressure C on the basis of
on controlled EAdi resumption ventilator
ventilation waveforms
and EAdi signals

Mellado Artigas, 39 90% of Ranged from General ICU patients Volume or 60 1 1 Automated algorithm SAS 1 to 4
et al. the patients 0.1 to 75% of in the first 24 h of pressure C on the basis of
each patient's mechanical ventilation ventilator
mechanical waveforms
insufflations and EAdi signals
(median, 8%)
Shimatani, et al. 100 41.6% of 2.4% of all mechanical Pediatric patients SIMV PC-PS 30 1 1 Visual inspection SBS 21 to 21
the patients insufflations with ARDS of ventilator
recorded in waveforms and
all patients Pes signals

Bourenne, et al. 10* 33% of NR Moderate-to-severe Volume C 60 1 1 Visual inspection of NR


the patients ARDS PF ,150 ventilator
waveforms and
Pes signal

Definition of abbreviations: ARDS = acute respiratory distress syndrome; C = controlled ventilation; EAdi = electric activity of the diaphragm; IQR = interquartile range; NR = not
reported; Pes = esophageal pressure; PF = PaO2/FIO2 ratio; PRVC = pressure-regulated volume control; RASS = richmond agitation sedation scale; RT = reverse triggering;
SBS = state behavioral scale; SIMV PC-PS = synchronized intermittent mandatory ventilation pressure control–pressure support.
*Patients not on neuromuscular-blocking agents.
CONCISE CLINICAL REVIEW

American Journal of Respiratory and Critical Care Medicine Volume 207 Number 5 | March 1 2023
CONCISE CLINICAL REVIEW

Figure 2. Schematic representation of the potential feedback loop system associated with reverse triggering. The ventilator is set in assist
control and triggers a mandatory mechanical insufflation. The mechanical insufflation provides a stimulus (dashed black arrow) that can
stimulate one or multiple sensors located in the respiratory muscles (depicted in red color [e.g., muscle spindles]), lungs (depicted in light pink
[e.g., lung stretching receptors]), or cardiovascular system (not depicted in the figure [carotid bodies]). These sensors send feedback to the
central pattern generator (CPG) (solid red arrow). The feedback stimulates the activity of the CPG, which activates the respiratory muscles
(e.g., the diaphragm via phrenic nerve activity) to contract (red dashed arrow). The respiratory muscle contraction, therefore, starts after the onset
of the mechanical insufflation characterizing the reverse-triggered breath (solid black arrow). See text for more details. (A) Mechanical ventilator set
on controlled ventilation. (B) Brain, where the CPG and respiratory centers are located. (C) Ribcage, including the diaphragm and intercostal
muscles, lungs, and carotid bodies. (D) Mechanical ventilator showing reverse-triggered breaths. For definition of abbreviations, see Figure 1.

predicted body weight) during controlled Reflexes from the phrenic, intercostal and Stimuli and Potential Mechanisms of
ventilation, and most patients were under vagal nerve, spine, and suprapontine Respiratory Entrainment to Periodic
continuous sedation (3, 5–9). structures can contribute to reverse Insufflation
triggering as part of respiratory entrainment Central pattern generators (CPGs) are
Rate of Reverse-triggered Breaths or independent mechanisms (1, 12–15). networks of neurons underlying the
In patients with reverse triggering production of central commands that control
identified, the rate of reverse-triggered Respiratory Entrainment stereotyped, rhythmic motor patterns such as
breaths varied and has been reported in Respiratory entrainment refers to the breathing (21, 22). The respiratory CPG is
different ways. The overall rate of reverse establishment of a fixed repetitive temporal composed of neurons in the brainstem that
triggering breaths in sedated patients under relationship between the patient’s respiratory connect with respiratory regions of the
controlled ventilation can vary from 0.1% rhythm with external input. For example, the medulla and pons, exerting their motor
to 75% of all patient’s mechanical respiratory rhythm can be entrained to the control mostly via phrenic motoneurons
insufflations (Table 2) (3, 5, 7, 8, 12). rhythm of the limbs’ movements during innervating the diaphragm (21, 23). Within
Studies analyzing long recording exercise (16, 17). Respiratory entrainment to the CPG, neurons in the pre-Botzinger
periods (e.g., 24 hours/day) from intubation mechanical ventilation (1) results in the complex have intrinsic pacemaker properties
until patients recover spontaneous breathing patient’s breathing efforts following the primarily responsible for rhythm generation.
might be necessary to confirm the real ventilator’s rhythm during controlled They receive inputs from the Botzinger
prevalence. Possibly, the majority of patients ventilation with a regular time interval (phase complex, postinspiratory complex, and
under sedation and lung-protective delay angle). It has been described in humans lateral parafacial nucleus, allowing breathing
controlled ventilation develop reverse during quiet wakefulness and sleep (18), under to coordinate with functions such as
triggering at some point during the course of anesthesia (19), and after a double lung speaking, singing, and swallowing (21–23).
mechanical ventilation. transplant (20). When respiratory entrainment The CPG normally receives inputs from
occurs, the patient’s rate of breathing effort various sources that modulate the rate of the
Pathophysiological Mechanisms involuntary follows the rate of the external respiratory center’s discharge and, ultimately,
The pathophysiological mechanisms periodic stimulus. This interaction between the the timing of effort (24). A potential input for
involved during reverse triggering are not patient and the ventilator can occur with respiratory entrainment to mechanical
fully understood. Respiratory entrainment is different patterns of coupling (see later and ventilation is the periodic oscillations of
the most often referred mechanism (Figure 2). Video E1 in the online supplement). arterial PCO2 (24). The onset of the inspiratory

Concise Clinical Review 537


CONCISE CLINICAL REVIEW

Figure 3. Assessment of intrinsic respiratory rate drive and effort in patients with reverse triggering. (A) Patient on volume-control ventilation
with midcycle reverse triggering at a 1:2 entrainment ratio and an intrinsic rate present during the occlusion. (B) Patient on volume-control
ventilation with reverse triggering without a clear ratio of entrainment and no visible intrinsic rate during the occlusion. The solid horizontal line
shows the time elapsed between the onset of two inspiratory efforts (drop in esophageal pressure) during the pause (duration of one respiratory
cycle). The patient’s intrinsic respiratory rate can thus be estimated as 12 breaths per minute (4.99 sec/60 sec). The horizontal dashed line
shows the time elapsed between the onset of two inspiratory efforts during control ventilation. Long dash dot horizontal lines show the duration
of the end-expiratory pause. Black arrows show the absence of plateau pressure, which indicates the presence of reverse triggering.
DPes = pressure swing in the esophageal pressure waveform during the end-expiratory pause; DPocc = pressure swing in the pressure–time
waveform during the end-expiratory pause; P0.1 = drop in the airway pressure time waveform during the first 100 ms of the end-expiratory
occlusion maneuver; Paw = airway pressure; Pes = esophageal pressure.

activity coincided with the time of the highest hypothesized to contribute to respiratory support), or by changing the ventilator
degree of PCO2 during these periodic entrainment (30, 31). During controlled parameters (e.g., inspiratory time, respiratory
oscillations (24). During controlled ventilation, the rib cage is periodically rate, and tidal volume) (1, 12–15). All these
ventilation, oscillations in the degree of PCO2 displaced. Activation of intercostal muscle factors can make the apparent pattern look
occur as a consequence of the ventilator’s spindles sends afferent signals to the CPG that irregular. Respiratory entrainment can also
periodic insufflation and exhalation (25, 26). inhibit phrenic nerve activity and modify the occur in an apparently chaotic pattern (33)
Thus, PCO2 oscillations occurring during onset timing of the inspiratory effort (30, 31). with a lack of fixed temporal relationship with
controlled ventilation can induce respiratory Stimuli from higher neural centers may mechanical insufflation (1, 13–15). However,
entrainment by phase-locking the onset of the also contribute to respiratory entrainment in the phase-angle (see below) that expresses the
neural inspiratory time to the peak PCO2 critically ill patients, like periodic noise arising difference between patient effort compared
degree achieved during each of these from the ventilator, as happens with music. with the start of the mechanical insufflation
oscillations (i.e., after the onset of the The neural centers processing the auditory remains relatively constant across breaths,
mechanical insufflation) (24). stimuli send input to the respiratory CPG lacking the natural variability of physiological
Lung inflation through vagal afferents with a specific frequency (i.e., the rhythm of breathing (1).
may also mediate respiratory entrainment via music). If the magnitude of the auditory
the Hering-Breuer reflex (20, 27, 28). For stimulus is stronger than that of other stimuli, Alternative Mechanisms (Reflex
instance, a mechanical insufflation occurring respiratory entrainment can occur (32). Contraction without Entrainment of
during the patient’s neural expiratory time the Respiratory CPG)
would activate vagal afferents evoking the Patterns of Respiratory Entrainment Reverse triggering has been reported in two
Hering-Breuer expiratory prolonging reflex. During Controlled Ventilation brain-dead patients without brainstem
Early animal models suggested vagal Respiratory entrainment can occur in a phase- function (34) and in patients with a lack
afferents influencing the respiratory CPG locked rhythmic pattern with the rhythm of of inspiratory effort during long end-
were important for respiratory entrainment the mechanical insufflations at different expiratory occlusions (i.e., no intrinsic
to occur as entrainment disappeared after integral ratios (1, 13–15). It frequently occurs activity from the CPG) (Figure 3). These
vagotomy (27). Vagal afferents facilitate in a 1:1 (one inspiratory muscle contraction results suggest that reverse triggering can
phase-locking in animals (28) and humans after each mechanical insufflation) or 1:2 (one also occur without entrainment of the
during sleep (20). Later, however, it was inspiratory muscle contraction after every respiratory rhythm generator but as a
shown that vagal afferents are not mandatory other mechanical insufflation) ratio (1, 13–15). consequence of local reflexes, such as a
for respiratory entrainment to mechanical The integral ratio can be, however, modified spinal reflex contraction of the respiratory
ventilation because it has been reported in by factors such as incomplete exhalation, muscle during chest wall inflation (35).
vagotomized dogs (24) and double-lung hyperinflation, variable effort, breath stacking,
transplant patients (20, 29). ventilator modes without a constant pattern Clinical Predisposing Factors
Stretching of intercostal muscles with rib for breath delivery and cycling (e.g.,
cage displacement and activation of the synchronized intermittent mandatory Predisposing factors that influence the
intercostal-to-phrenic reflex is also ventilation pressure control–pressure occurrence of reverse triggering and

538 American Journal of Respiratory and Critical Care Medicine Volume 207 Number 5 | March 1 2023
CONCISE CLINICAL REVIEW

entrainment are related to characteristics of volume and reverse triggering is complex as deflection in the ventilator airway
the external stimuli (i.e., periodic insufflation there might be an interplay between tidal pressure–time curve (Figures 1B and 1C); a
as dictated by mechanical ventilator settings) volume and CO2 depending on the set positive change in the Pes signal at the onset
or to patient intrinsic factors that modify ventilator rate and the patient’s intrinsic of the mechanical insufflation (Figure 1C);
susceptibility to entrainment or facilitate the respiratory rate. In these circumstances, the absence of an increase in the EAdi signal
occurrence of local reflexes (e.g., respiratory particularly when patients are sedated with at the onset of the mechanical insufflation
drive, sleep stages, and sedation). opioids (see below) and have low intrinsic (Figure 1B); and the expiratory time of the
The tidal volume and respiratory rate respiratory rates, reverse triggering may previous breath is the expiratory time preset
set on the ventilator modulate the intensity occur even with larger tidal volumes. on the ventilator in controlled ventilation
and the characteristics of the stimulus- Sedation is associated with reverse (Figures 1B and 1C) (1, 7, 9).
inducing respiratory entrainment. Overall, triggering. Reverse triggering is typically
there is a range of rates within which the found during the transition from deep Ventilator Waveforms
respiratory rhythm can be entrained, related sedation to the patient resuming During the ventilator’s expiratory time, the
to the intrinsic rate of the CPG (18, 20). spontaneous ventilation (1, 7, 8). Moreover, patient effort can usually be detected by the
Reducing the ventilator rate can modify the specific drugs can increase the susceptibility deformation of the expiratory flow–time
entrainment ratio or abolish reverse for reverse triggering, although this is not curve that lacks its normal exponential decay,
triggering if the set rate falls below the fully elucidated. For instance, higher doses of and a peak expiratory flow reduced
patient’s intrinsic rate and the patient starts opioids but not benzodiazepines were compared with breaths without reverse
triggering the ventilator (8, 18, 20, 36). Set associated with increased susceptibility to triggering (Figures 1B and 1C) (1, 7–9).
tidal volume may also play an important role, reverse triggering in adult patients with During the inspiratory time of volume-
and most of the recent observations ARDS (8). control ventilation, the airway pressure curve
described concerned patients with protective Sleep can also change the susceptibility shows a deformation induced by the effort. If
ventilation and tidal volumes around 6 ml/kg to entrainment in healthy humans (18), a short inspiratory pause is set at the end of
predicted body weight. In Damiani’s possibly because of the loss of modulation the ventilator’s inspiratory phase, patient
experimental work in animals with lung from forebrain influences to perceived effort will cause an apparent decrease in
injury, reverse triggering was induced by respiratory sensations (e.g., lung inflation by the plateau pressure during the pause
reducing tidal volume (4). In adult patients mechanical insufflation) during sleep (18). (Figure 1C) (1, 7–9).
with acute respiratory distress syndrome The loss of such influences would reduce the During the inspiratory time of pressure-
(ARDS), reverse triggering was capacity of the respiratory CPG to modify control ventilation, there might be a subtle
independently associated with lower tidal the respiratory rhythm and its susceptibility
drop in the airway pressure and an increase
volumes set on the ventilator (5). Cause or to phase-locking with the rhythm of the
in the flow–time curves, but often difficult to
more likely a consequence, reverse triggering mechanical insufflations (18). Reverse
detect (Figure 1B) (1, 7–9).
was associated with larger tidal volumes in triggering dyssynchrony is often associated
pediatric patients with ARDS under pressure with the transition between deep-to-light
Reference Techniques: EAdi and Pes
controlled ventilation (12). In anesthetized sedation that a similar mechanism can
EAdi and Pes can detect reverse triggering
humans, larger tidal volumes reduce explain (8).
(i.e., patient effort) (Figures 1B and 1C)
respiratory entrainment by modifying the
onset time and duration of the triggered (7, 39). If the EAdi signal increases after the
breath and either changing the entrainment’s Diagnosis onset of a mechanical insufflation triggered
integral rate or abolishing it (19). One by the ventilator, it indicates reverse
hypothesis is that greater lung insufflation General Principles triggering (Figure 1B). Because ECG artifacts
could activate the Hering-Breuer reflex, Different approaches can be used to identify in the EAdi signal can lead to false positives
shorten the neural inspiratory time reverse triggering at the bedside, including (40), using different cutoffs for the peak EAdi
associated with the triggered breath or ventilator waveforms, displayed on most (e.g., >3 μv) may help improve accuracy (7).
prolong the neural expiratory time (19). ventilators (8), and more sophisticated An inspiratory muscle contraction will
Greater lung insufflation can also change methods, such as the electric activity of the generate a negative change in pleural
arterial CO2, and it is possible that the high diaphragm (EAdi) (7) and esophageal pressure and Pes (41). A negative swing in
occurrence of reverse triggering in recent pressure (Pes) (1, 3, 5, 9). Pes after an initial positive deflection at the
studies could be associated with more To classify a breath as reverse triggering, onset of a passive mechanical insufflation
frequent use of slightly elevated it is necessary to first determine if the patient triggered by the ventilator can detect reverse
concentrations of CO2 (37). CO2 is a classical or the ventilator initiated the mechanical triggering (Figure 1C) (1, 9). Campbell
stimulus of the respiratory drive and insufflation. Nowadays, most ventilators diagrams constructed with the Pes signal
decreases susceptibility for respiratory indicate on their screens whether the have also been used to detect reverse
entrainment in animal models (38). mechanical insufflation was triggered by the triggering (3, 5).
However, mild increases in the respiratory ventilator or the patient, which facilitates
drive because of increasing CO2 the identification process at the bedside. Electrical Impedance Tomography
concentrations were not able to affect To determine whether the mechanical Reverse triggering has been described with
respiratory entrainment in humans (18). In insufflation was triggered by the ventilator, electrical impedance tomography giving a
summary, the relationship between tidal one can also check the following: absence of secondary rise in ventilation (2).

Concise Clinical Review 539


CONCISE CLINICAL REVIEW

Automated Techniques Coefficient of variation ð%Þ (e.g., increasing the tidal volume) or
Algorithms to identify reverse triggering standard deviation of the phase angle reinforcing sedation to decrease respiratory
were developed on the basis of the criteria 5 : drive. As described above, it is often easy to
mean phase angle
described above. On the basis of only identify if the first mechanical insufflation is
ventilator waveforms (8, 9) or using EAdi It has been shown that during reverse triggered by the patient or by the ventilator
(7), they showed good accuracy, sensitivity, triggering with respiratory entrainment, (Figure 1). When available, EAdi or Pes can
and specificity in detecting reverse a lower coefficient of variation (e.g., less provide further information (Figure 1) (15).
triggering in relatively large data sets (7–9). than 15%) of its phase angle than the Breath stacking can occur as a consequence
This highlights the feasibility of identifying normal variability of spontaneous of reverse triggering, but the first mechanical
reverse triggering on the basis of signals breathing is expected (43). Factors can insufflation is triggered by the ventilator,
readily available at the bedside and the modify the entrainment ratio and may lead followed by an inspiratory effort, late and
potential for future largescale to a greater coefficient of variation strong enough to trigger a second
implementation of these automated (1, 12–15). mechanical insufflation (Figure 1B) (1, 3, 15).
methods in ICUs worldwide, which will Phenotypes. Because the onset of the The differentiation between breath stacking
likely provide support to clinicians and for reverse-triggered breath can occur either because of double triggering or reverse
future studies to improve reverse triggering during the inspiratory or expiratory time of triggering will impact the strategy for
detection and management (7–9). the ventilator, different phenotypes have managing ventilation and sedation. If breath
been suggested (3). Early reverse triggering stacking occurs because of reverse triggering,
Expiratory Hold with early relaxation, when the maximum adjusting the ventilatory settings may have
Using an end-expiratory hold has been effort and relaxation occur during the little effect. The clinician may instead choose
suggested for differentiating patients with ventilator inspiratory time; early reverse to reduce or interrupt sedation to let the
reverse triggering from those with triggering with late relaxation, when the patient control breathing frequency
spontaneous efforts during controlled relaxation and termination of the reverse (see below).
ventilation (42, 43). However, inspiratory triggering occur during the ventilator
effort frequently occurs during the expiratory expiratory time; midcycle reverse triggering,
hold in patients with reverse triggering when reverse triggering starts during the Clinical Implication of
(see Figure 3) and depicts the intrinsic ventilator inspiratory time of the ventilator, Reverse Triggering
respiratory rate of the patient lower than that but the maximal effort is reached during
set in the ventilator. the ventilator expiratory time; late reverse Reverse triggering may have various effects
triggering, when reverse triggering occurs on lung and diaphragm structure and
Characterization of Reverse completely during the ventilator expiratory function (Table 1). Its consequences might
Triggering phase; and reverse triggering leading to depend on the frequency, phenotype, and
Phase angle. The phase-locking of the double-cycling. These different phenotypes degree of breathing effort. When reverse
inspiratory effort to the mechanical were associated with different physiological triggering is observed at the bedside, a
insufflation is quantified by the phase angle, consequences, such as various degrees of careful assessment and interpretation should
which is the difference between the onset of eccentric contraction (3). be made.
the mechanical insufflation and the onset of
the inspiratory effort (Figures 1A and 1B) (1): Lung Consequences
Reverse triggering may be harmful to the
Phase angle ð Þ lungs because of multiple mechanisms.
Inspiratory effort onset time 2 mechanical insufflation onset time Superimposing a spontaneous effort to a
5 360 ,
mechanical ventilator cycle duration ventilator breath has an additive effect on the
distending transpulmonary pressure and the
in which the mechanical ventilator cycle Double breaths and breath stacking. resulting tidal volume depending on the
duration is the time elapsed between two If strong enough, the inspiratory effort mode. In pressure-control modes, tidal
mandatory mechanical insufflations. associated with the reverse triggering can volume is variable and is influenced by the
A phase angle of zero means the cause the ventilator to trigger a second patient’s breathing efforts. Depending on the
simultaneous activity of the inspiratory mechanical insufflation, resulting in breath phenotype, reverse triggering can increase
effort and the mechanical insufflation; a stacking (Figure 1B) (1, 3, 15). It is important the average tidal volume and inspiratory
negative phase angle means the inspiratory to differentiate breath stacking caused by transpulmonary pressure (3). If deep and
effort began before the mechanical reverse triggering from double triggering. long enough, the patient’s inspiratory effort
insufflation; and a positive phase angle Double triggering is usually a consequence of will persist beyond the end of the mechanical
means that the inspiratory effort began short cycling and high inspiratory drive. It insufflation, leading to double cycling, with
after the mechanical insufflation (i.e., occurs when the same patient effort triggers a potentially the sum of two consecutive tidal
reverse triggering) (1). first mechanical insufflation, continues after volumes. During volume-controlled
In addition, the coefficient of variation the ventilator cycles off, and is strong enough ventilation, pendelluft (i.e., inspiratory
between the timing of the reverse-triggered to trigger a second mechanical insufflation. transfer of gas from one part of the lung to
breaths is typically used to determine its Managing this dyssynchrony requires another, usually nondependent to dependent
variability (1): readjusting the ventilator when possible lung [2]) may occur during reverse triggering

540 American Journal of Respiratory and Critical Care Medicine Volume 207 Number 5 | March 1 2023
CONCISE CLINICAL REVIEW

without any increase in tidal volume and Clinical Outcomes after attempts to optimize ventilator settings,
could generate regional hyperinflation. The impact of reverse triggering on patient- metabolic derangements, and pain and
Pendelluft during reverse triggering can centered outcomes is unknown. anxiety were implemented (52).
induce regional overstretch in dependent Many breathing efforts during reverse
lung regions compared with that observed in triggering are probably not excessive. In a Assessing the Presence of Intrinsic
the same region during passive controlled clinical study, muscular efforts (muscular Respiratory Rate in Patients with
ventilation at the same tidal volume (2). pressure) during reverse triggering had a Reverse Triggering
On the opposite, the potential lung median (interquartile range, 25–75%) of 8.7 Looking for the presence of intrinsic
benefit from the (postinspiratory) expiratory (5.6–9.9) cmH2O (9); the median peak EAdi respiratory rate in patients with reverse
activity of the diaphragm was recently was 1.7 (0.8–4.3) μV (7); and the rate of triggering (Figure 3) can be the first step to
demonstrated in a pig model of acute double-cycling was very low (,0.1–3% of help the clinician decide whether the patient
hypoxemic respiratory failure (AHRF). reverse-triggered breaths) (7, 8). has the potential to be transitioned to a
Diaphragmatic activity during the ventilator Evidence also supports that a positive spontaneous mode of ventilation. A trial can
expiratory phase delayed and reduced the impact of reverse triggering in the early be made to transition the patient to a
expiratory collapse and increased lung phases of hypoxemic respiratory failure is spontaneous mode of ventilation and
aeration, compared with mechanical possible. In a prospective cohort study, sedation further reduced or stopped.
ventilation with muscle paralysis and absence patients with a higher rate of reverse
of diaphragm activity (44). The diaphragm triggering and low breathing effort (average Changing Ventilator Settings
can exert an expiratory braking effect EAdi, 1.7 μV) had better oxygenation and Reducing respiratory rate can, in some cases,
helping to preserve the end-expiratory lung were more likely to progress to an assisted abolish reverse triggering. If the respiratory
volume (44). mode or be extubated within the following rate is set below the patient’s intrinsic
24 hours as compared with patients with a respiratory rate, the patient can start
Diaphragm Consequences
low rate or no reverse-triggered breaths (7). triggering the ventilator (8, 36). Decreasing
Diaphragm activity during reverse triggering Moreover, reverse triggering has been the rate may increase CO2, which may also
associated with reduced hospital mortality contribute to stopping the reverse triggering
often occurs during the ventilator expiratory
(hazard ratio, 0.65; 95% confidence interval, process. Increasing respiratory rate seems to
phase (e.g., midcycle reverse triggering),
0.57–0.73), indirectly suggesting that reverse have no (or a less clear) effect (8, 36). It can
resulting in an eccentric contraction. The
triggering may be a marker of favorable be more complex if the patient has no
impact of diaphragm eccentric contractions
outcomes (8). intrinsic respiratory rate and reverse
on its function and structure could go in
triggering occurs purely because of
opposite directions and is associated with the
respiratory entrainment. In this scenario,
degree of breathing effort (4). Possible Management reverse triggering is likely to occur over a
Reverse triggering may harm the
range of ventilator respiratory rates, which
diaphragm when contractions repeatedly There is no direct evidence about the best can vary between patients and, for example,
occur during the ventilator’s expiratory strategy to manage reverse triggering. If be influenced by the integrity of (e.g., vagal
phase, concomitantly with lung volume reverse triggering is not associated with and/or phrenic) afferences to the CPG, the
reduction and inspiratory muscle potentially harmful consequences (e.g., intensity of the respiratory drive, ventilator
lengthening in the context of excessive double cycling, breath stacking, and/or settings, and wakefulness and/or sleep stages
breathing efforts (4). Animal models strong inspiratory efforts), perhaps no action (20, 28, 38, 54).
demonstrated that the combination of is required, and it can be used as a sign that Increasing the inspiratory time set on
reverse triggering and high inspiratory effort the patient may be close to being ready to the ventilator can reduce the number of
or induced eccentric contractions might be trigger the ventilator. If reverse triggering is breath-stacking events in patients who are
associated with impaired diaphragm associated with these potentially harmful mechanically ventilated (48). It should
structure and force-generating capacity consequences, modifying ventilator settings be differentiated if breath stacking is caused
(4, 45). The potentially harmful effect of or transitioning the patient to a spontaneous by reverse triggering or double triggering.
these eccentric contractions relates to a lack ventilation mode should be considered. Even if breath stacking caused by reverse-
of homogeneity in sarcomeres stretching Unfortunately, a frequent approach triggered breaths is abolished, reverse
(asymmetric lengthening), triggering a clinicians adopt to manage patient–ventilator triggering can still occur.
cascade of events possibly leading to dyssynchrony is increasing the depth of In patients with ARDS, reverse
inflammatory reaction and reactive oxygen sedation and, in some cases, paralysis with triggering was associated with small set tidal
species production (46, 47). neuromuscular blocking (48, 49). Paralysis volumes (8). Increasing tidal volume on the
Conversely, reverse triggering at low or and sedation are associated with diaphragm ventilator during volume-control ventilation
moderate efforts could help protecting inactivity, muscle dysfunction (including the or increasing inspiratory time or driving
diaphragm function compared with passive diaphragm), longer ICU length of stay, and pressure during pressure-control ventilation
ventilation. Experimental data demonstrated potentially increased mortality (6, 50–53). will induce larger tidal volumes, which may
that reverse triggering resulting in low Increasing sedation to treat dyssynchronies, shorten the duration of the inspiratory effort
breathing effort preserved diaphragm including reverse triggering, is only associated with the reverse-triggered breath
function after 3 hours compared with passive appropriate when dyssynchrony results in and possibly mitigate breath-stacking or
mechanical ventilation (4). excessive respiratory drive and effort and strong efforts.

Concise Clinical Review 541


CONCISE CLINICAL REVIEW

Assessing Breathing Effort in Patients adequate degree of ventilation, lung injury is factors influence the degree of effort
with Reverse Triggering still significant, or develops excessive associated with reverse-triggered breaths is an
Assessing the magnitude of the effort inspiratory effort, there is an increased risk of important question. Knowing whether the
associated with the reverse-triggered breath lung and (load-induced) diaphragm injury, same factors influencing the degree of effort in
will help guide the management of reverse and other management strategies might be patients triggering the ventilator (e.g., PCO2 )
triggering because the consequences to the needed (58). could be useful in developing management
lungs and diaphragm are mediated by the Reducing the sedation can also facilitate strategies. Finally, clinical and specifically
effort associated with the reverse-triggered the resumption of spontaneous breathing in designed studies to assess the impact of
breath (2, 4). patients with reverse triggering and accelerate reverse triggering on lung and diaphragm
The magnitude of the inspiratory effort the transition to a spontaneous ventilation function and injury are also required.
associated with the reverse-triggered breath mode. If reverse triggering is associated with
can be estimated during an end-expiratory excessive inspiratory efforts or double cycling,
occlusion maneuver by calculating the airway which does not improve after changes in Conclusions
pressure swing in the pressure waveform (42, ventilator settings and the patient cannot be
55, 56) (Figure 3A) in patients who have an safely transitioned to a spontaneous mode of Reverse triggering is highly prevalent in
intrinsic rate. Esophageal pressure swings can ventilation, paralysis can be a last resource sedated patients under mechanical ventilation
also be used during the reverse-triggered option to mitigate possible harmful in a controlled mode. It can be either
breath or during the end-expiratory occlusion consequences to the lungs and diaphragm. beneficial or harmful for patients depending
maneuver (Figure 3) (52, 57). There are no on the magnitude of the effort associated with
reference values to determine whether a it and its consequences on the interaction
patient’s inspiratory drive is adequate on the Priorities for Research between the patient and the ventilator. The
basis of only EAdi signals (39, 52, 57). Further work is necessary to better clinician at the bedside should be able to
characterize reverse triggering prevalence, perform the diagnosis, understand the
Transitioning Patients to a risk factors increasing its susceptibility, and pathophysiological mechanisms and
Spontaneous Mode of Ventilation to elucidate when reverse triggering could be predisposing factors associated with reverse
If the patient has an intrinsic respiratory rate beneficial or harmful for patients. triggering, and use potential tools for its
and clinically improving, a trial to transition Prospective studies investigating the effects management. Reverse triggering should not
the patient to a spontaneous ventilation of changes in ventilator settings, such as rate be regarded as always harmful to the patient,
mode (e.g., pressure support, (proportional and tidal volume, are necessary to support and in some cases, it can be interpreted as a
assist ventilation) PAV1, or (neurally the development of management strategies sign of readiness to resume spontaneous
adjusted ventilatory assist) NAVA) can be that can be tested in randomized trials. ventilation. Careful assessment of ventilator
performed, ideally concomitantly to reducing Studies with recordings of ventilator waveforms at the bedside can provide the
sedation. If the patient can sustain adequate waveforms 24 hours per day over multiple clinician with valuable information to decide
degrees of ventilation and effort, the consecutive days are also useful to provide a which management strategies (if any) might
transition is likely beneficial, as it can clearer picture of the prevalence and be necessary. 䊏
mitigate diaphragm atrophy without harmful distribution (i.e., the presence of clusters
effects on the lungs and diaphragm (52). (59)) of reverse triggering in patients who are Author disclosures are available with the
However, if the patient cannot maintain mechanically ventilated. Moreover, which text of this article at www.atsjournals.org.

References 6. Sklar MC, Madotto F, Jonkman A, Rauseo M, Soliman I, Damiani LF, et al.
Duration of diaphragmatic inactivity after endotracheal intubation of
1. Akoumianaki E, Lyazidi A, Rey N, Matamis D, Perez-Martinez N, Giraud R, critically ill patients. Crit Care 2021;25:26.
et al. Mechanical ventilation-induced reverse-triggered breaths: a 7. Mellado Artigas R, Damiani LF, Piraino T, Pham T, Chen L, Rauseo M,
frequently unrecognized form of neuromechanical coupling. Chest 2013; et al. Reverse triggering dyssynchrony 24 h after initiation of mechanical
143:927–938. ventilation. Anesthesiology 2021;134:760–769.
2. Yoshida T, Nakamura MAM, Morais CCA, Amato MBP, Kavanagh BP. 8. Rodriguez PO, Tiribelli N, Fredes S, Gogniat E, Plotnikow G,
Reverse triggering causes an injurious inflation pattern during Ceballos IF, et al. Prevalence of reverse triggering in early ARDS:
mechanical ventilation. Am J Respir Crit Care Med 2018;198: results from a multicenter observational study. Chest 2021;159:
1096–1099. 186–195.
3. Baedorf Kassis E, Su HK, Graham AR, Novack V, Loring SH, Talmor DS. 9. Pham T, Montanya J, Telias I, Piraino T, Magrans R, Coudroy R, et al.;
Reverse trigger phenotypes in acute respiratory distress syndrome. Am BEARDS study investigators. Automated detection and quantification
J Respir Crit Care Med 2021;203:67–77. of reverse triggering effort under mechanical ventilation. Crit Care 2021;
4. Damiani LF, Engelberts D, Bastia L, Osada K, Katira BH, Otulakowski G, 25:60.
et al. Impact of reverse triggering dyssynchrony during lung-protective 10. Rodriguez PO, Tiribelli N, Gogniat E, Plotnikow GA, Fredes S,
ventilation on diaphragm function: an experimental model. Am J Respir Fernandez Ceballos I, et al.; GRAAVEplus (Grupo Argentino de
Crit Care Med 2022;205:663–673. estudio de Asincronıas en la VEntilacio  n mecanica y temas
5. Su HK, Loring SH, Talmor D, Baedorf Kassis E. Reverse triggering with relacionados a los cuidados crıticos). Automatic detection of
breath stacking during mechanical ventilation results in large tidal reverse-triggering related asynchronies during mechanical
volumes and transpulmonary pressure swings. Intensive Care Med ventilation in ARDS patients using flow and pressure signals. J Clin
2019;45:1161–1162. Monit Comput 2020;34:1239–1246.

542 American Journal of Respiratory and Critical Care Medicine Volume 207 Number 5 | March 1 2023
CONCISE CLINICAL REVIEW

11. Talmor D, Sarge T, Malhotra A, O’Donnell CR, Ritz R, Lisbon A, et al. 38. Eldridge FL, Paydarfar D, Wagner PG, Dowell RT. Phase resetting of
Mechanical ventilation guided by esophageal pressure in acute lung respiratory rhythm: effect of changing respiratory “drive”. Am J Physiol
injury. N Engl J Med 2008;359:2095–2104. 1989;257:R271–R277.
12. Shimatani T, Yoon B, Kyogoku M, Kyo M, Ohshimo S, Newth CJL, et al. 39. Vaporidi K, Akoumianaki E, Telias I, Goligher EC, Brochard L,
Frequency and risk factors for reverse triggering in pediatric acute Georgopoulos D; Pathophysiology and Clinical Implications.
respiratory distress syndrome during synchronized intermittent Respiratory drive in critically ill patients. Am J Respir Crit Care Med
mandatory ventilation. Ann Am Thorac Soc 2021;18:820–829. 2020;201:20–32.
13. Telias I, Beitler JR. Reverse triggering, the rhythm dyssynchrony: potential 40. Jonkman AH, Roesthuis LH, de Boer EC, de Vries HJ, Girbes ARJ,
implications for lung and diaphragm protection. Am J Respir Crit Care van der Hoeven JG, et al. Inadequate assessment of patient-ventilator
Med 2021;203:5–6. interaction due to suboptimal diaphragm electrical activity signal
14. Sassoon CS, Mancebo J. The double-edged sword of reverse triggering: filtering. Am J Respir Crit Care Med 2020;202:141–144.
impact on the diaphragm. Am J Respir Crit Care Med 2022;205:606–608. 41. Telias I, Brochard L, Goligher EC. Is my patient’s respiratory drive (too)
15. Pham T, Telias I, Piraino T, Yoshida T, Brochard LJ. Asynchrony high? Intensive Care Med 2018;44:1936–1939.
consequences and management. Crit Care Clin 2018;34:325–341. 42. Dianti J, Bertoni M, Goligher EC. Monitoring patient-ventilator interaction
16. Potts JT, Rybak IA, Paton JF. Respiratory rhythm entrainment by somatic by an end-expiratory occlusion maneuver. Intensive Care Med 2020;
afferent stimulation. J Neurosci 2005;25:1965–1978. 46:2338–2341.
17. Kohl J, Koller EA, Ja €ger M. Relation between pedalling- and breathing 43. de Vries HJ, Jonkman AH, Tuinman PR, Girbes ARJ, Heunks LMA.
rhythm. Eur J Appl Physiol Occup Physiol 1981;47:223–237. Respiratory entrainment and reverse triggering in a mechanically
18. Simon PM, Zurob AS, Wies WM, Leiter JC, Hubmayr RD. Entrainment of ventilated patient. Ann Am Thorac Soc 2019;16:499–505.
respiration in humans by periodic lung inflations. Effect of state and 44. Pellegrini M, Hedenstierna G, Roneus A, Segelsjo € M, Larsson A,
CO(2). Am J Respir Crit Care Med 1999;160:950–960. Perchiazzi G. The diaphragm acts as a brake during expiration to
19. Graves C, Glass L, Laporta D, Meloche R, Grassino A. Respiratory prevent lung collapse. Am J Respir Crit Care Med 2017;195:
phase locking during mechanical ventilation in anesthetized human 1608–1616.
subjects. Am J Physiol 1986;250:R902–R909. 45. Gea J, Zhu E, Ga ldiz JB, Comtois N, Salazkin I, Fiz JA, et al. [Functional
20. Simon PM, Habel AM, Daubenspeck JA, Leiter JC. Vagal feedback in the consequences of eccentric contractions of the diaphragm]. Arch
entrainment of respiration to mechanical ventilation in sleeping Bronconeumol 2009;45:68–74.
humans. J Appl Physiol (1985) 2000;89:760–769. 46. Hody S, Croisier JL, Bury T, Rogister B, Leprince P. Eccentric muscle
21. Richter DW, Smith JC. Respiratory rhythm generation in vivo. Physiology
contractions: risks and benefits. Front Physiol 2019;10:536.
(Bethesda) 2014;29:58–71.
47. Morgan DL. New insights into the behavior of muscle during active
22. Steuer I, Guertin PA. Central pattern generators in the brainstem and
lengthening. Biophys J 1990;57:209–221.
spinal cord: an overview of basic principles, similarities and differences.
48. Chanques G, Kress JP, Pohlman A, Patel S, Poston J, Jaber S, et al.
Rev Neurosci 2019;30:107–164.
Impact of ventilator adjustment and sedation-analgesia practices on
23. Long S, Duffin J. The neuronal determinants of respiratory rhythm. Prog
severe asynchrony in patients ventilated in assist-control mode. Crit
Neurobiol 1986;27:101–182.
24. Takahashi E, Tejima K, Yamakoshi K. Entrainment of respiratory rhythm Care Med 2013;41:2177–2187.
to respiratory oscillations of arterial PCO2 in vagotomized dogs. J Appl 49. Bailey JM. Management of patient-ventilator asynchrony. Anesthesiology
Physiol (1985) 1992;73:1052–1057. 2021;134:629–636.
25. Collier DJ, Nickol AH, Milledge JS, van Ruiten HJA, Collier CJ, 50. Shehabi Y, Chan L, Kadiman S, Alias A, Ismail WN, Tan MA, et al.;
Swenson ER, et al. Alveolar PCO2 oscillations and ventilation at sea Sedation Practice in Intensive Care Evaluation (SPICE) Study Group
level and at high altitude. J Appl Physiol (1985) 2008;104:404–415. investigators. Sedation depth and long-term mortality in mechanically
26. Rackley CR. Monitoring during mechanical ventilation. Respir Care 2020; ventilated critically ill adults: a prospective longitudinal multicentre
65:832–846. cohort study. Intensive Care Med 2013;39:910–918.
27. Petrillo GA, Glass L, Trippenbach T. Phase locking of the respiratory 51. Shehabi Y, Bellomo R, Reade MC, Bailey M, Bass F, Howe B, et al.;
rhythm in cats to a mechanical ventilator. Can J Physiol Pharmacol Sedation Practice in Intensive Care Evaluation (SPICE) Study
1983;61:599–607. Investigators; ANZICS Clinical Trials Group. Early intensive care
28. Muzzin S, Trippenbach T, Baconnier P, Benchetrit G. Entrainment of the sedation predicts long-term mortality in ventilated critically ill patients.
respiratory rhythm by periodic lung inflation during vagal cooling. Am J Respir Crit Care Med 2012;186:724–731.
Respir Physiol 1989;75:157–172. 52. Goligher EC, Dres M, Patel BK, Sahetya SK, Beitler JR, Telias I, et al.
29. Roze H, Repusseau B, Thumerel M, Demant X, Blanchard E, Jougon J. Lung- and diaphragm-protective ventilation. Am J Respir Crit Care Med
Ventilation of denervated transplanted lung at risk for overdistention by 2020;202:950–961.
reverse triggering and breath stacking. Br J Anaesth 2022;129:e1–e4. 53. Dianti J, Angriman F, Ferreyro BL, Sklar MC, Brochard L, Ferguson ND,
30. Remmers JE. Inhibition of inspiratory activity by intercostal muscle et al. Association of mortality with neuromuscular blockade differs
afferents. Respir Physiol 1970;10:358–383. according to baseline diaphragm thickness. Am J Respir Crit Care Med
31. Remmers JE, Marttila I. Action of intercostal muscle afferents on the 2020;202:1717–1720.
respiratory rhythm of anesthetized cats. Respir Physiol 1975;24:31–41. 54. Muzzin S, Baconnier P, Benchetrit G. Entrainment of respiratory rhythm
32. Haas F, Distenfeld S, Axen K. Effects of perceived musical rhythm on by periodic lung inflation: effect of airflow rate and duration. Am J
respiratory pattern. J Appl Physiol (1985) 1986;61:1185–1191. Physiol 1992;263:R292–R300.
33. Guevara MR, Glass L. Phase locking, period doubling bifurcations and 55. Bertoni M, Telias I, Urner M, Long M, Del Sorbo L, Fan E, et al. A novel
chaos in a mathematical model of a periodically driven oscillator: a non-invasive method to detect excessively high respiratory effort and
theory for the entrainment of biological oscillators and the generation of dynamic transpulmonary driving pressure during mechanical
cardiac dysrhythmias. J Math Biol 1982;14:1–23. ventilation. Crit Care 2019;23:346.
34. Delisle S, Charbonney E, Albert M, Ouellet P, Marsolais P, Rigollot M, 56. de Vries H, Jonkman A, Shi ZH, Spoelstra-de Man A, Heunks L.
et al. Patient-ventilator asynchrony due to reverse triggering occurring Assessing breathing effort in mechanical ventilation: physiology and
in brain-dead patients: clinical implications and physiological meaning. clinical implications. Ann Transl Med 2018;6:387.
Am J Respir Crit Care Med 2016;194:1166–1168. 57. Telias I, Spadaro S. Techniques to monitor respiratory drive and
35. Chen BY, Vasilakos K, Boisteanu D, Garma L, Derenne JP, Whitelaw WA. inspiratory effort. Curr Opin Crit Care 2020;26:3–10.
Linkage of hiccup with heartbeat. J Appl Physiol (1985) 2000;88:2159–2165. 58. Telias I, Junhasavasdikul D, Rittayamai N, Piquilloud L, Chen L,
36. Shimatani T, Yoon B, Hotz JC, Khemani RG. Reply: not all breaths that Ferguson ND, et al. Airway occlusion pressure as an estimate of
follow a ventilator cycle are reverse triggering. Ann Am Thorac Soc respiratory drive and inspiratory effort during assisted ventilation. Am J
2021;18:1264–1266. Respir Crit Care Med 2020;201:1086–1098.
37. Lellouche F, Delorme M, Brochard L. Impact of respiratory rate and 59. Vaporidi K, Babalis D, Chytas A, Lilitsis E, Kondili E, Amargianitakis V,
dead space in the current era of lung protective mechanical ventilation. et al. Clusters of ineffective efforts during mechanical ventilation: impact
Chest 2020;158:45–47. on outcome. Intensive Care Med 2017;43:184–191.

Concise Clinical Review 543

You might also like