RZAEFLEROARAT
Mey Xingaoyi Medical Equipment Co,, Ltd
555 Yeshan Road, Yuyao, Zhejiang, China 315400
Tel: +86.574-62720809 Fax: +86-574-62730908
MAb: SuperScan—3.0 38} UHI ARE RT ARR
Appendix: Specification 8. Parameters of SuperScan—3.0 Superconducting Magnetic Resonance
Imaging System (MRI)
BUR AR. Magnet System
4S AU Ri a 1 | Whole body superconducting stainless steel magnet | 1
Sid BR: 3.0T Field strength: 3.07
BARR OR Ab ab te) Length of Magnet Geithout cover): <
600mm, 160mm
BIKER RAEZOOMM Magnet bore: 900mm
SSE ERM A ESpes
829.20 RMS SE Magnetic field homogeneity:
i) (V-RMS measurement)
28emDSV
18KKW » 2
Working scope of RF amplifier: 127.7MHz +
0.5MH=
Type of RF transmitting coil: Orthogonal cage
multisource transmitting
Type of RF amplifier cooling: water cooling
‘Type of RF amplifier: Impulse
HWHRBBRES Parallel acquisition technology platform
AEPRARCRD 1 | Phased array head & neck combined coil 1
J RR 1 | Phased array body coil 1
AERA PAB 1 | Phased array knee coil 1
ARPAAPARB 1 | Phased array shoulder coil 1
FAA SR 1_| Phased array multifunctional soft coil 1
HA AMR Gradient System
BARRA AP ALA | 1 | Gradient coil cooling: Hollow inside cooling | 1
HRBURA AKER
HALA AERA Gradient controlling technology: Full time digital
RAR AM CARA ROL) Macinwm single axial gradient field strength
2 AbmT/m {non—vaild value 245mT/m
ARH RAGE A RAIS | | Masimum single axial slew rate: {nonvaild
200T/m/s value}: 200T/ra/s
KART AM Bee RAL || Mac, Single axial gradient field strength & Mas.
AMBRE MPP TAMA | | Single axial slew rate can be reached
4 sivaultancously in one same sequence
RAG RRO Mac. Duty cycle: 100%
RK FOVS 48cm Max. FOV: Sem
BARTHEL CK 0 RAE | | Min, Gradient rise time (rise from 0 to its Max.
RAMA) 8GB
Solid stare hard dish: >960GB
1 | Displayer size:
Displayer resolution: 1920 » 1200
Number of displayers: >2pes
Maz. reconstruction matrix of scanning images:
102 x 1024
Image reconstruction speed (256 » 256 matrie, full
ctMey
RZAEFLEROARAT
Xingaoyi Medical Equipment Co,, Ltd
555 Yeshan Road, Yuyao, Zhejiang, China 315400
Tel: +86.574-62720809 Fax: +86-574-62730908
& FOV, 2D #9t £4&) 3000
IE
ooh 12 48 BAR AB (256 « 256
4E5F, AS ALR) > 7500000 #8
FOV, 2D Fourier transform): >3000images/sec
Max. numbers of patient images be stored (256
% 256 matrix, non—compressed images): >
7,500,000 images
PERI EIR,
RUA ART HE He ARE
See RE
PIA THEM AR
Ath
SABE AR AALS 200KG
RERRPERRARAS
200mm/see
BUR OY IN HY Ro BA a
fa, TR BRAM Dio
BERRA SZ 0em
BERETA: Cah
ERRATA: AR CTR
Scanning Table 8. Scanning Environment
1 | Two-way patient—doctor talking system
1 | Patient calling alarm system
1 | Adjustable ventilation & lighting system inside
patient bore
Maz. Loading of patient table: >200kgs
Mar. Moving—in speed horizontally: >
200mm/see
Operating butions on both sides of front magnet
cover, able to control scanning and movement of
patient table
Length of patient table: >210em
a) Way of patient table: Electric power—driven
PERL TGA Movement direction of patient table: 2D (up &
PARE LG ATE down, in & out)
SeRERA OE Moving scope in vertical: 25em
ERE AI Mowing scope in horizontal: 2180em
PORREA Patient table emergent stop coailable
RAB RRA Manual movement of patient table available
AREA RAE Laser positioning system
1 | Foorentering scanning mode available
Coit storage cabinet provided
BHR ‘Scanning Parameters
BLK FOV248em
3) FOV
SH 42 RA