Professional Documents
Culture Documents
My Ethics of Eye
My Ethics of Eye
Validation
Muhammad Saqlain Tahir Ph.D.
Validation
A well-designed experimental
matrix (Written protocol)
Step-by-step methodology (STP)
Test samples of good quality
(should meet the quality as per
defined specifications)
Peak purity (preferably >99%
purity)
Equipments & analytical
instruments should be in
calibrated state.
Scope
Analytical Category II
Performance Category Category Category
Characteristics I Quantitative Limit Tests III IV
Accuracy Yes Yes a a No
Precision Yes Yes No Yes No
Specificity Yes Yes Yes a Yes
DL No No Yes a No
QL No Yes No a No
Linearity Yes Yes No a No
Range Yes Yes a a No
a May be required, depending on the nature of the specific test
Methodology
Definition: Types:
Closeness of agreement (degree of Measurement / Injection
scatter) between a series of Repeatability (System Precision)
measurements obtained from multiple
Method Repeatability (Method
sampling of the same homogeneous Precision)
sample under the prescribed
conditions Intermediate Precision - precision
under different laboratory conditions
Discussion:
(within laboratory variation, as on
Repeatability:- precision under same different days, or with different
operating conditions (within laboratory analysts, or equipments within the
over a short interval of time using the same laboratory).
same analyst with same equipment). Reproducibility – precision between
Repeatability is also termed intra-assay
laboratories
precision.
Precision
System Repeatability
Injection No.
Peak Area
1 1121533
2 1121262
3 1120831
4 1121177
5 1115016
6 1120936
Mean 1120125.8
SD 2515.6
% RSD 0.22
Precision - Repeatability
Method Repeatability
Sample No.
Assay (%)
1 98.82
2 99.20
3 99.41
4 98.51
5 98.75
6 98.68
Mean 98.89
SD 0.34
% RSD 0.34
Precision - Intermediate
Acceptance Criteria:
% RSD
Procedure System Precision Method Precision IMP
API DP API DP API DP
Assay 1.0/2.0 1.0/2.0* 1.0 2.0 1.0 3.0
Dissolution NA 1.0/2.0* NA 6.0 NA 5.0
Impurities 5.0* 5.0* 10.0 10.0 10.0 15.0
Acceptance Criteria:
Assay:-
Recovery should be between 98% to 102% (Depends upon the strength)
Dissolution:-
95% to 105%
Impurities:-
if, Specification is ≤ 0.2% : 85% to 115%
if, Specification is > 0.2% : 90% to 110%
At LOQ level : 80% to 120%
A simple logic behind this performance characteristic is whether the procedure is
capable of estimating a true value or not.
Linearity
Definition:
Linearity:- The Linearity of an analytical procedure is its ability (within a
given range) to obtain test results that are directly proportional to the
concentration of analyte in the sample.
Range:- The interval between the upper and lower level that have been
demonstrated to be determined with precision, accuracy and linearity using
this method as written.
Discussion:
In order to determine the quantity of any analyte present in unknown sample,
some kind of relation ship (mathematical/empirical) between concentration
and response is essential.
Linearity
Procedure:
Prepare a series of solutions (not less than five is recommended) with standard / reference
samples in the specified concentration range and analyze them as per method
Assay:-
80% to 120% of test concentration
CU:-
70% to 130% of test concentration
Dissolution:-
± 20% of expected release (Q) for immediate release
0 to 120% (for extended release)
Impurities:-
LOQ to 200% of specification
Linearity
Evaluation:
Slope:- indicates sensitivity of the method
Intercept:- indicates response for no analyte (interference)
Correlation Coefficient:- indicates the relation ship chosen is correct
Acceptance Criteria:
Correlation Coefficient should be not less than 0.999 for assay, CU,
dissolution test methods and 0.99 for impurities test method
Linearity
Parameter Results
% RSD (Area) 0.22
% RSD (Contents) 0.49
% RSD (Cumulative) 0.42
Individual Recovery 98.63 – 100.19%
Average Recovery 99.21 – 99.59%
Correlation coefficient 0.9999
y - Intercept -13045.0
Slope of regression line 11254.7
Linest y = 11254.7x – 13045.0
LOD & LOQ
Limit of Detection:
Lowest amount of analyte in a sample which can be detected but not
necessarily quantitated, under the stated experimental conditions (LOD)
Limit of Quantitation:
LOQ: Lowest amount of analyte in a sample which can be quantitatively
determined with suitable precision and accuracy (LOQ)
LOD & LOQ
Approaches: Procedure:
Different approaches suggested by ICH, Derived from standard preparations
USP & EP. prepared for linearity studies (five
levels)
Derived from the residual standard
deviation of regression line (Sy/x) Evaluation:
Derived from the residual standard Detection limit (DL) co-efficient 3.3
deviation of y-intercept of
Quantitation limit (QL) co-efficient
regression lines (Sy)
10
Signal to noise
Acceptance Criteria:
RSD of six replicate injections is ≤ 10%
for LOQ and between ˃ 10% ≤ 33.0%
for LOD
Solution Stability
Discussion: Procedure:
It is often essential that solution Prepare test sample as per procedure and
(standards, test samples) be stable analyze at initial and at different time
enough to allow for delays covering intervals by keeping the sample at room
instrument breakdowns / overnight temperature (25°C) / refrigerator condition
analysis. (2 – 8°C)
A minimum of 12 Hrs, 18 Hrs or 24 Hrs is Evaluation:
routinely recommended for
chromatographic methods for which % difference from initial response to
vialed solutions may remain on an specified interval for analyte / each
autosampler at ambient temperatures impurity.
due to various delays Acceptance Criteria:
A simple logic behind this study is to % difference is not more than
determine the period of time, a solution can
be held before analysis without Assay/CU: 2.0
compromising accuracy.
Dissolution 3.0
Impurities 10.0
Robustness
There are no official guidelines on the sequence of If stability and robustness data is not available with
validation experiments and the optimal sequence method development data
can be depending on the method itself. Based on
experience, for HPLC method the following ✓ Stability
sequence has been proven to be useful for time
management. ✓ Robustness
Generally method validation report should have Summary of report (overall view of
validation exercise, any critical issues,
Objective and scope of the method recommendations etc., for the application
Molecule details: IUPAC Name, CAS No. of method)
Molecular Formula, Molecular Weight and Instrument out puts, which should
Molecular Structure etc. represent critical method parameters
Detailed list of chemicals, reagents, Specificity and LOD – for Identification Test
reference standards (Generally, photographs)
Listing of equipment and its functional and Selectivity / Specificity data (discriminating
performance requirements chromatogram, peak purity data, blank and
Methodology followed placebo chromatograms and stressed
samples chromatograms)
Validation data: parameter wise – Linearity graphs
procedure, results, conclusion etc.
Summary of data (results in brief – Resolution and related system suitability
parameter wise) chromatograms
Any out puts which are significant
Revalidation
Thank you