You are on page 1of 7

Blackbody Comparator for Thermocouple Calibration

M. Ojanen, O. M. Hahtela, and M. Heinonen

Centre for Metrology and Accreditation (MIKES), P.O.Box 9, FI-02151 Espoo, Finland

Abstract. MIKES is developing a measurement set-up for calibrating thermocouples in the temperature range 960 °C –
1500 °C. The calibration method is based on direct comparison of thermocouples and radiation thermometers. We have
designed a graphite blackbody comparator cell, which is operated in a horizontal single-zone tube furnace. The cell
includes two blackbody cavities for radiation temperature measurements. The cavities have openings on opposite sides
of the cell, allowing simultaneous measurement with two radiation thermometers. The design of the comparator allows
three thermocouples to be calibrated simultaneously. The thermocouples to be calibrated are inserted in thermometer
wells around one of the measurement cavities. We characterize the blackbody comparator in terms of repeatability,
temperature distribution and emissivity. Finally, we validate the uncertainty analysis by comparing calibration results
obtained for type B and S thermocouples to the calibration results reported by Technical Research Institute of Sweden
(SP), and MIKES. The agreement in the temperature range 1000 °C – 1500 °C is within 0.90 °C, the average deviation
being 0.17 °C.
Keywords: Thermocouple Calibration, Radiation Thermometer, Variable Temperature Blackbody

INTRODUCTION characterization of the blackbody comparator. Then,


we compare the radiation and contact temperature
Thermocouples are widely used in industrial measurements using the blackbody comparator. We
applications in contact temperature measurements present the uncertainty budget for the comparison
above 1100 °C. The International Temperature Scale measurements. Finally, we present the conclusions and
of 1990 (ITS-90) is determined by Planck’s radiation future prospects of this work.
law above the freezing point of Ag (961.78 °C), and
does not include fixed points above the freezing point DESCRIPTION OF THE BLACKBODY
of Cu (1084.62 °C) [1]. Currently, thermocouple COMPARATOR
calibrations at higher temperatures are often carried
out at Pd melting point (1554 °C) using wire bridge The material used for the blackbody comparator is
method. Improved methods for calibration need to be graphite. Its layout is presented in Fig. 1. The length of
developed, as the wire bridge method requires the whole comparator block is 340 mm and the
destruction of the thermocouple junction, and because diameter is 50 mm. The comparator contains three
the calibration uncertainty is typically large. thermometer wells for three thermocouples to be
Two alternative methods for improved calibration inserted and calibrated simultaneously, and two
have been studied. Eutectic fixed points [2-8] would identical cavities on opposite sides of the comparator.
provide a calibration facility similar to the currently The cavities can be used for simultaneous
used ITS-90 temperature fixed points. Another method measurements with an ITS-90 calibrated pyrometer
is based on direct comparison between radiation and absolute calibrated filter radiometer. The cavities
thermometer and thermocouple measurement using are cylindrical, and their diameter is 15 mm. The
comparator blackbodies [9-11]. These blackbodies diameter of the cavity aperture is 10 mm. The length
allow direct and simultaneous comparison of of the cavities is 165 mm, leaving a10 mm thick wall
thermocouples and radiation thermometers. They can between the cavities. The bottoms of the cavities are
be applied directly in interpolation of the temperatures conical, with an angle of 118 °.
between the fixed points. These two methods are The thermometer wells are placed at 120 ° intervals
compared e.g. in [12]. around one comparator cavity, as shown in Fig. 2. The
At MIKES, we have designed a blackbody depth of the thermometer wells is 170 mm. The
comparator for thermocouple calibration in the thermometer shields made of Alsint are permanently
temperature range 1000 °C – 1500 °C. The layout of installed, only the thermocouples need to be inserted
the comparator allows calibration of three when carrying out measurements. The shields can be
thermocouples simultaneously against a pyrometer and easily replaced, if necessary.
/ or a filter radiometer. First, we present the layout and

Temperature: Its Measurement and Control in Science and Industry, Volume 8


AIP Conf. Proc. 1552, 468-473 (2013); doi: 10.1063/1.4821391
© 2013 AIP Publishing LLC 978-0-7354-1178-4/$30.00

468
170 mm

10 mm 50 mm
15 mm 118 ° 165 mm

340 mm
FIGURE 1. Layout of the blackbody comparator.

Measuring equipment
Thermocouples

Two type S and two type B thermocouples were


used in characterization of the blackbody comparator.
The type S thermocouples are calibrated at MIKES
10 mm
with Cu, Ag, Al (660.323 °C), Zn (419.527 °C) and Sn
(231.928 °C) fixed point cells. An interpolation
8 mm polynomial is derived to calculate the deviation from
the IEC60584 standard. The expanded (k = 2)
calibration uncertainty is 0.5 °C. It consists of
50 mm components emerging from the uncertainty of the
FIGURE 2. Front view of the blackbody comparator. fixed point temperature realizations, uncertainty of the
thermal voltage measurement, repeatability of the
The blackbody comparator is installed in a single- results and uncertainty of the interpolation polynomial.
zone tube furnace with an operation range up to The type B thermocouples are calibrated at the Cu
1500 °C. The diameter of the furnace tube is 55 mm, point at MIKES, and at Au (1064.18 °C) and Pd
and the length is 1160 mm. The tube is open at both melting point at the Technical Research Institute of
ends, allowing simultaneous measurements of both Sweden (SP). As for the type S thermocouples, an
cavities. A specially designed insulator block made of interpolation polynomial is fitted between the fixed
ceramic insulator material is placed in the furnace tube calibration points. The expanded (k = 2) uncertainty
in front of the blackbody. The design helps to keep the of the type B thermocouple calibration is 1 °C at
thermometer shields in their place. An argon gas flow temperatures below 1200 °C and 2 °C in the
is applied at both ends of the blackbody comparator to temperature range 1200 °C to 1500 °C.
protect the graphite from oxidation. During the measurements, the reference junctions
of the thermocouples are immersed in an ice-water
bath. The thermal voltages are measured with a
CHARACTERIZATION OF THE calibrated high-accuracy nanovoltmeter.
BLACKBODY COMPARATOR
The characterization of the blackbody comparator Radiation thermometer
includes studies in repeatability, temperature
homogeneity among the thermometer wells, axial and The radiation thermometer used in the
longitudinal gradients and cavity emissivity. measurements was the MIKES standard radiation
thermometer, of type IKE LP3. The central
wavelength of the band pass filter used in this work is
900 nm. The radiation thermometer was calibrated
according to ITS-90 at Ag fixed point prior to the

469
measurements. The effective wavelength of the TABLE 1. Uncertainty budget for radiation thermometer
pyrometer, 899.32 nm, was determined with Ag and calibration at Ag fixed point.
Cu fixed point measurements, and applying Wien’s Uncertainty component Contribution in
approximation [13]. The standard uncertainty of the temperature / °C
effective wavelength determination was 0.01 nm. The Signal measurement 0.003
Emissivity 0.038
standard uncertainty for calibration at Ag fixed point Fixed point repeatability 0.001
was 0.049 °C. The uncertainty for the radiation Impurities in fixed point 0.010
thermometer calibration is calculated according to Alignment 0.029
[14]. The contributions of uncertainty components in Temperature drop 0.005
temperature are presented in Table 1. Combined standard uncertainty 0.049

reference junction temperature was 0.12 °C, which


Measurement uncertainty corresponds to an uncertainty of 0.05 °C in the
thermocouple calibration temperatures. For the
The uncertainty budget for thermocouple pyrometer measurements, the standard deviations were
calibration with radiation thermometer and blackbody between 0.03 °C and 0.11 °C. The fluctuation of the
comparator is presented in Table 2. furnace temperature is included in the thermocouple
The repeatability due to removal / insertion of a and radiation thermometer signal measurement
thermocouple was studied simultaneously by components.
switching the thermocouples between the different Uncertainty due to longitudinal temperature
thermometer wells at 1000 °C. One thermocouple was gradient and alignment was studied with both
kept constantly at its place in order to compensate for thermocouples and pyrometer. Thermocouple signals
furnace temperature fluctuation. The maximum were compared with each other, when one
temperature difference was 0.41 °C. Switching a thermocouple was kept at the bottom of one
thermocouple from one thermometer shield to another thermometer well, while the other was glided out.
required the reference junction to be temporarily With this method, the uncertainty due to gradient at
removed from the ice-water bath. 5 mm distance from the bottom of the thermometer
To study the temperature distribution between well was estimated to be 0.13 °C at 1000 °C. The
different thermometer wells, the thermocouples were 5 mm distance corresponds to the distance between the
compared to each other. The comparisons were carried thermometer well and cavity bottoms, as presented in
out at temperatures of 1000 °C and 1500 °C. The Fig.1. The gradient study with pyrometer was carried
maximum deviation in the thermocouple readings was out by focusing it at different distances in the cavity.
found to be 0.21 °C, with standard deviation of This allowed also the effect of pyrometer alignment to
0.06 °C. be studied. The uncertainty due to gradient and
The axial temperature gradient was estimated with different focalizations was found to be 0.3 °C at
the radiation temperature measurements carried out at 1000 °C.
the same side as the thermocouples. The effective emissivity of the cavities was
Standard deviation of a single thermocouple signal modeled using STEEP3 software. It was predicted to
during the measurement was between 0.03 °C and be 0.9994 ± 0.0001 at 900 nm.
0.04 °C. The standard uncertainty of the voltmeter was
0.33 PV, which corresponds to an uncertainty of
0.03°C in temperature. The uncertainty of the

470
TABLE 2. Uncertainty budget for thermocouple calibration against pyrometer.
Temperature / °C
1000 1100 1200 1300 1400 1500
Repeatability 0.12 0.12 0.12 0.12 0.12 0.12
Temperature distribution 0.06 0.06 0.06 0.06 0.06 0.06
Axial temperature gradient 0.17 0.17 0.17 0.17 0.17 0.17
Thermocouple signal measurement 0.03 0.03 0.03 0.04 0.04 0.04
Voltmeter calibration 0.03 0.03 0.03 0.03 0.03 0.03
Reference junction temperature 0.05 0.05 0.05 0.05 0.05 0.05
Pyrometer calibration 0.06 0.08 0.09 0.10 0.11 0.13
Linearity 0.04 0.05 0.05 0.06 0.07 0.08
Pyrometer stability 0.00 0.01 0.02 0.03 0.04 0.05
Pyrometer signal measurement 0.06 0.05 0.03 0.11 0.03 0.03
Longitudinal temperature gradient 0.30 0.30 0.35 0.35 0.40 0.40
Emissivity 0.01 0.01 0.01 0.01 0.01 0.01
Size-of-source effect 0.04 0.05 0.05 0.06 0.07 0.08
Gradient at cavity bottom 0.13 0.13 0.20 0.20 0.30 0.30
Combined standard uncertainty 0.41 0.41 0.47 0.49 0.57 0.57
Expanded uncertainty (k = 2) 0.82 0.82 0.95 0.98 1.14 1.14

during the measurements, except when changing the


Comparison radiation temperature measurement cavity or the
thermocouples. The radiation thermometer was
aligned separately for each temperature. The results
The blackbody comparator was validated by are presented in Fig. 3.
comparing the calibration results obtained with the As can be seen in Fig. 3., the temperatures
system for the thermocouples to earlier results measured with thermocouples are within the
obtained with calibrations at MIKES and SP. The IKE measurement uncertainties of the reference
LP3 was used as a reference in the comparator. The temperature. The average deviation is 0.17 °C. The
measurements were carried out at temperatures from maximum deviation is 0.90 °C, measured with
1000 °C to 1500 °C with 100 °C intervals. Type S thermocouple B114 at 1300 °C. The results when
thermocouples were used in the measurements at measuring at the same / opposite side with the
temperatures below 1100 °C. The measurements with insertion of thermocouples agree within 0.01 °C for
radiation thermometer were made on both cavities; at S18 and 0.34 °C for S21.
the same and opposite sides with the thermocouples. The uncertainty budget for the comparison between
Measurements with both methods at the same side thermocouple and radiation thermometer
require attention in order not to block the field-of-view measurements is presented in Table 3.
of the radiation thermometer with the thermometer
cables. The thermocouples were kept on their places

TABLE 3. Uncertainty budget for comparison of thermocouples and pyrometer.


Temperature / °C
1000 1100 1200 1300 1400 1500
Thermocouple calibration 0.25 0.25 0.50 1.00 1.00 1.00
Thermocouple stability 0.08 0.08 0.13 0.13 0.13 0.13
Calibration against pyrometer 0.41 0.41 0.47 0.49 0.57 0.57
Combined standard uncertainty 0.48 0.49 0.70 1.12 1.16 1.16
Expanded uncertainty (k = 2) 0.96 0.98 1.40 2.24 2.32 2.32

471
2.5
Difference T TC - T, °C

1.5

0.5

-0.5

B114 B115
-1.5
S18 same side S21 same side
S18 opposite side S21 opposite side
U, k = 2 -U
-2.5
900 1000 1100 1200 1300 1400 1500 1600
Temperature, °C
FIGURE 3. Differences between thermocouple (TTC) and radiation thermometer measurements (T). Labels B114, B115, S18
and S21 refer to different thermocouples used in this study. Open symbols represent the radiation temperature measurements at
the opposite side, and the closed symbols represent the measurements at the same side with the thermocouples. The dashed lines
represent the expanded uncertainty of the comparison (k = 2).

agreement in the temperature range 1000 °C to


1500 °C was within 0.9 °C, the average deviation
CONCLUSIONS AND DISCUSSION being 0.17 °C. These results are well within the
estimated measurement uncertainties.
In this article, we presented a graphite blackbody We conclude that the novel blackbody comparator
comparator for thermocouple calibration for the is suitable for calibration of thermocouples in the
temperature range 960 °C to 1500 °C. The design of temperature range 960 °C to 1500 °C.
the comparator allows three thermocouples to be The blackbody comparator has been designed for
calibrated simultaneously, using two radiation simultaneous measurement of ITS-90 calibrated
thermometers. We characterized the blackbody in pyrometer and absolute calibrated filter radiometer. In
terms of temperature distribution, thermal gradients, the future, irradiance mode filter radiometers with a
and emissivity. We compared the thermocouple defining aperture in front of the blackbody can be used
measurements with each other to study the temperature in radiation temperature measurements at the opposite
homogeneity between different thermocouple wells side with the thermocouples.
and insertion / removal of a thermocouple. The To further study and improve the calibration
gradients were studied with both thermocouples and uncertainty, measurements and thermal modeling of
pyrometer. The estimated standard uncertainties for both axial and longitudinal temperature gradients will
the calibration with standard radiation thermometer of be carried out. As the gradients are considered
MIKES and the blackbody comparator are between dominant in the uncertainty evaluation of the radiation
0.41 °C and 0.57 °C in the temperature range 1000 ° C temperature measurements, these investigations will
to 1500 °C. provide valuable information related to the overall
To validate the blackbody comparator, we uncertainty budget. The longitudinal gradient effects
compared thermocouple measurements traceable could also be reduced by using a 3-zone furnace [9-
through SP, Sweden, with radiation thermometer 11].
measurements directly traceable to ITS-90. The

472
ACKNOWLEDGEMENTS 8. Elliott, C. J., Pearce, J. V., Failleau, G., Deuzé, T.,
Briaudeau, S., Sadli, M. and Machin, G., Metrologia 49,
88-94 (2012).
The authors would like to thank Mr. Jorma 9. Astrua, M., Battuello, M. and Girard, F., Meas. Sci.
Manninen and Mr. Hannu Räsänen for technical Technol. 17, 2186-2190 (2006).
assistance, and Mr. Arttu Ollikainen for the 10. Kim, Y.-G., Kim, B. H. and Yang, I., Metrologia 47,
manufacture of the comparator blackbody. 239-247 (2010).
11. Martin, M. J., Zarco, M. and Del Campo, D., Int. J.
REFERENCES Thermophys. 32, 383-395 (2011).
12. Astrua, M., Battuello, M. and Girard, F., Int. J.
Thermophys. 29, 190-198 (2008).
1. Preston-Thomas, H., Metrologia 27, 3-10 (1990). 13. Ojanen, M., Anhalt, K., Hartmann, J., Schiller, S.,
2. Ogura, H., Izuchi, M. and Arai, M., Int. J. Thermophys. Weckström, T., Kärhä, P., Heinonen, M. and Ikonen, E.,
29, 210-221 (2008). Measurement 43, 183-189 (2010).
3. Pearce, J. V., Ogura, H., Izuchi, M. and Machin, G., 14. Fischer, J., Battuello, M., Sadli, M., Ballico, M., Park, S-
Metrologia 26, 473-479 (2009). N., Saunders, P., Zundong, Y., Carol Johnson, B., van
4. Kim, Y.-G., Yang, I., Kwon, S. Y. and Gam, K. S., der Ham, E., Sakuma, F., Machin, G., Fox, N., Li, W.,
Metrologia 43, 67-70 (2006). Ugur, S. and Matveyev, M.,” Uncertainty Budgets for
5. Edler, F. and Baratto, A. C., Metrologia 42, 201-207
Realization of ITS̺90 by Radiation Thermometry”, in
(2005).
Temperature: Its Measurement and Control in Science
6. Edler, F. and Baratto, A. C., Metrologia 43, 501-507
and Industry 7, AIP Conference proceedings 684, 2002,
(2006).
New York: American Institute of Physics, pp. 631-638.
7. Edler, F., Morice, R., Ogura, H. and Pearce, J.,
Metrologia 47, 90-95 (2010).

473
Copyright of AIP Conference Proceedings is the property of American Institute of Physics
and its content may not be copied or emailed to multiple sites or posted to a listserv without
the copyright holder's express written permission. However, users may print, download, or
email articles for individual use.

You might also like