You are on page 1of 53

General TB830D, FLXA402T

Specifications Surface Scattering Light Turbidity Analyzer

GS 12E04B40-01EN

n General
To keep good quality in industrial or drinking water and
rivers or streams, “Turbidity” is one of the important
measurement parameters to control water quality.
Turbidity analyzer is more than ever demanded
not only for water treatment, but currently for
measurement/detection of suspended substances in
industrial waste or of turbidity at chemical processing
facility.
Since their sales began in 1959, Yokogawa
turbidimeters have been continuously developed and
improved using various measurement principles suited
for various applications. With its many achievements,
Yokogawa has earned its customers’ confidence.
Surface Scattering Light Turbidity Detector TB830D
and Liquid Analyzer FLXA402T, is a next-generation
analyzer, adopting the surface scattered light method,
which is less susceptible to contamination by non-
contact with the sample, realizing high efficiency
in facility operations with our latest digital sensor
technology.
FLXA402T is a converter designed, being tailored
from the well-received FLXA402, to provide multiple
detector measurement. In addition to conventional
analog output, Modbus TCP, Modbus RTU/RS485 are
available.
The technology of predictive maintenance prevents
accidental shutdown of the factory and reduces OPEX.
TB830D has drastically improved maintainability and
minimized the maintenance time.

n Features
TB830D FLXA402T
■ Surface scattering type measurement method ■ Active diagnosis advice service
■ Long life of 3 years by LED light source.  Detects disconnection of lens heater
■ High sensitivity, High precision  Condition of light source, dryness level of detector
 Realized good linearity, repeatability  Input element failure, calibration failure, various
■ Measuring range: 0-2 FNU/NTU to 0-2000 FNU/NTU circuit failures, etc.
■ Improved cleaning system for less maintenance ■ Multiple detector/sensor connection
 Easy cleaning cell structure  Turbidity, pH, Conductivity
■ Analog outputs, Digital communications (Modbus
RTU or Modbus TCP/IP)
■ Multilingual display
 English, German, French, Chinese, Korean,
Spanish, Portuguese, Japanese
■ Automatic analog output range switching (2 or 3
range changing)

FLEXA, FLXA, SENCOM are trademarks or registered trademarks of Yokogawa Electric Corporation.
All other company and product names mentioned in this document are trademarks or registered trademarks of their respective companies.
We do not use TM or ® mark to indicate those trademarks or registered trademarks in this document.

Yokogawa Electric Corporation GS 12E04B40-01EN


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright Jan. 2023
3rd Edition May 11, 2023
2
n System Configuration
For example, the diagram below is a system without sampling system.

Analog Output
FLXA402T
Contact Input
Liquid
Contact Output Analyzer
RS485 or Ethernet

TB830D

Power Supply

Drain
Head
Sample water
Tank

Drain

For example, the diagram below is a system with automatic cleaning and automatic zero calibration

Analog Output
Contact Input FLXA402T
Contact Output Liquid
Analyzer
RS485 or Ethernet

TB830D
Relay
box
Power supply
Drain

Head
Tank

Sample water/
Zero water
Drain
Sample water
Zero turbidity filter
Zero water

Tap water
For washing

Figure 1

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
3
n Standard Specifications Frequency: 50/60Hz
Acceptable range: 50Hz±5%, 60Hz±5%
1. TB830D Surface Scattering Light Turbidity Power consumption (Sampling system)
Detector -NN, -A5: 15 VA
-A1, -A2, -A3: 150 VA (including the converterʼs)
■ Measurement
Communication signal:
Turbidity of tap water, river water, and water used in
RS485 (Modbus RTU) 9600 bp
general processes
The maximum cable length between Analyzer and
■ Detector Detector is 20 m.
Measuring method: S
 urface scattering-light
■ Mechanical and others
measurement
Material:
Light source: White Light LED, Infrared light LED
Case; m-PPE resin
(860 nm)
Wetted parts; m-PPE, PVC (Rigid polyvinyl
Measuring range:
chloride), PE, PP, Fluororubber (FKM)
0-2 to 0-2000 NTU (Turbidity standard: Formazin,
Finish: none
White Light)
Color: Black
0-2 to 0-2000 FNU/NTU (Turbidity standard:
Protection: IP65, Type4X (electric part on upper body
Formazin, 860 nm)
of the detector)
Linearity: *: A separate rainproof cover is required for outdoor
Upper range limit of 1000 FNU/NTU or less; installation. Required a roof to avoid direct sunshine.
±2% F.S. or ±0.04 FNU/NTU Weight: approx; 3.5 kg (detector only)
whichever is greater Mounting (option):
Over 1000 FNU/NTU and 2000 FNU/NTU or less; Pipe- or rack-mounted
±4% F.S. Terminal screw size: M3 (Earth terminal: M4)
Repeatability: Cable gland:
±1% of reading or ±0.02 FNU/NTU 3 hole, M20 cable gland x 3 (for detector)
whichever is greater (For 6.5 to 12 mm outer diameter cable)
Response time: Within 2 minutes Pipe connection:
(90% response, sample flow rate 3 L/min) Detector connection: 25 mm ID hose joint
Sample conditions: Conduit adapter (option):
(with the sampling system) G1/2, 1/2NPT or M20 × 1.5 mm
Flow rate; 2 to 10 L/min Stainless steel tag plate (option)
Pressure; 20 to 500 kPa max In case you specified Option code /SCT,
Temperature; 0 to 50°C (The ambient temperature Tag plate which is printed numbers is
should not exceed 30°C.) shipped with product. You can attach this
Zero Calibration water and Cleaning (tap) water tag by wire.
conditions Warm up time: Approx. 30 min
Turbidity; 2 FNU/NTU or less Installation/Storage environment:
Temperature; 0 to 50°C (The ambient temperature Indoors (A separate rainproof cover is
should not exceed 30°C.) required for outdoor installation.)
Pressure; 100 to 500 kPa Ambient Operating Temperature;
Flow rate; (piping system-dependent) -5 to 55°C (Samples or tap water may
Zero; 2 to 10 L/min need protection against freezing.)
Cleaning water; 3 to 6 L/min Storage temperature; -30 to 70°C
Water Consumption; Humidity; 10 to 90 % RH at 40°C (No condensation)
(with automatic cleaning, with
automatic zero calibration, time ■ Sampling system, others
preset at the factory shipment) Sampling system -A1, -A2, -A3 is specified, the
Zero; Approx. 380 L/day detector, the sampling system, the solenoid valve for
(In case of 2 L/min flow rate) the sampling system, the head tank are assembled in
Cleaning water; a stanchion. In this case FLXA402T is other model.
 Approx. 90 L/day (In case of When FLXA402T -ST (Assembled in stanchion) is
3 L/min flow rate) specified, the product is shipped with the sampling
(without the sampling system) system assembled in the stanchion.
Flow rate; 1.5 to 2 L/min Piping connections:
Temperature; 0 to 50°C (The ambient temperature System with sampling system;
should not exceed 30°C.) VP16 for sample inlet, tap water inlet
■ Electrical VP40 for drain
AC Power Supply Sampling system stanchion:
Ratings (Acceptable range) (Solenoid valve for Paint; Baked polyurethane resin finish
Sampling system) Color; Munsell 0.6GY3.1/2.0 equivalent
-NN: 100 to 240 VAC (90 to 264 VAC) Material; Carbon steel or stainless steel
-10: 100 VAC (90 to 110 VAC) Weight; approx. 50 kg
-12: 110 VAC (99 to 121 VAC) (in full specifications, excluding
-20: 200 VAC (180 to 220 VAC) combination instruments)
-22: 220 VAC (198 to 242 VAC) Installation; Fixed with anchor bolts
Class D grounding (ground resistance: 100 ohm or less)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
4
Relay box: 2. FLXA402T Liquid Analyzer
Control box for solenoid valve for sampling system FLXA402T Liquid Analyzer is a common converter
Cable grand; 8 cable grands (outer diameter of for Chlorine sensor unit and Turbidity detector. In this
applicable wires; 6.5 to 12 mm) GS, only function and specification are provided for
Weight; approx. 2 kg the connection between FLXA402T and TB830D.
Installation; assembled in a stanchion
Terminal box A: A box embedded with an arrester for 2.1 Measurement
power supply and a circuit protector ■ Specification of the turbidity detector
Cable gland; 4 cable glands (outer diameter of See “1. TB830D Surface Scattering Light Turbidity
applicable wires; 6.5 to 12 mm) Detector”. TB800D or TB810D is also
Weight; approx. 2 kg connectable to FLXA402T.
Installation; assembled in a stanchion
Terminal box B: A box for power distribution Turbidity: lower limit -1.000 NTU
Cable gland; 4 cable glands (outer diameter of upper limit 9999 NTU
applicable wires; 6.5 to 12 mm) Unit: NTU, FNU, FTU, mg/L You can input any
Weight; approx. 1 kg letters by alphanumeric.
Installation; assembled in a stanchion Display resolution:
0.001 NTU (when the measured value is
■ Safety, EMC, and RoHS conformity standards less than 0.2 NTU)
Safety (*1): Maximum number of digits to display after
UL: UL 61010-1, UL 61010-2-030 the decimal point
CSA: CAN/CSA-C22.2 No.61010-1
CAN/CSA-C22.2 No.61010-2-030 ■ pH (PH)
CE, UKCA: EN 61010-1, EN 61010-2-030 When 2nd Input is -P1 (PH)
Morocco/LVD Arrêté: High accurate pH measurement by the embedded
NM EN 61010 1, NM EN 61010 2 030 calibration standard table, temperature compensation
GB: GB30439 function, stability check function.
Installation altitude: 2000 m or less • Constant diagnosis of impedance, asymmetric
Overvoltage Category based on IEC 61010: II(*) potential, slopes.
Pollution level: IEC 61010: 2 (**) • Continuous diagnosis of the condition of the
(*) Installation category, called over-voltage category, electrode (dirt, breakage, disconnection), detector
specifies impulse withstand voltage. Equipment (sample surface level)
with Category II is energy-consuming equipment • Data saving of the latest five calibration results for
to be supplied from the fixed installation. future maintenance
(**) Pollution level shows the level of existing the • Predicts the maintenance or calibration time
solid, liquid and gas which break electrical Input Specification
isolation. Category 2 is applied to normal room Dual high impedance input (≥1012 Ω)
environment. Input Range
EMC: pH: -2 to 16 pH
CE, UKCA: EN 61326-1 Class A,Table 2 (For use in Temperature:
industrial locations) Pt1000; -30 to 140ºC
EN 61326-2-3 Pt100; -30 to 140ºC
EN IEC 61000-3-2 PTC500; -30 to 140ºC
EN 61000-3-2 6k8; -30 to 140ºC
EN 61000-3-3 PTC10k; -30 to 140ºC
Morocco/EMC Arrêté: NTC 8k55; -10 to 120ºC
NM EN 61326 1 3k Balco; -30 to 140ºC
NM EN 61326 2 3 Output Range
NM EN 61000 3 2 pH: min. span 1 pH
NM EN 61000 3 3 max. span 20 pH
RCM: EN61326-1 Class A Temperature: min. span 25ºC
Korea Electromagnetic Conformity Standard max. span 170ºC
한국 전자파적합성 기준 Class A Performance (The specifications are expressed
Environmental regulation: with simulated inputs.)
RoHS Directive (*2): EN IEC 63000 pH:
REACH: Regulation EC 1907/2006 Linearity; ±0.01 pH
Waste Electrical and Electronic Equipment (WEEE) Repeatability; ±0.01 pH
Directive : Temperature: w  ith Pt1000, 6k8, PTC10k, NTC 8k55,
This product is purposely designed to be 3k Balco, PTC500
used in a large scale fixed installations Linearity; ±0.3ºC
only and, therefore, is out of scope Repeatability; ±0.1ºC
of the WEEE Directive. The WEEE with Pt100:
Directive does not apply. The WEEE Linearity; ±0.4ºC
Directive is only valid in the EU and UK. Repeatability; ±0.1ºC
(*1): Except for sampling system (-A2, -A3)
(*2:) Except for sampling system (-A1, -A2, -A3)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
5
Performance (pH detector combined in standard YOKOGAWA provides conductivity sensors of
solution) which cell constants are 0.1 to 10 cm-1.
Response time: 10 seconds (90 % response, Output item
Temperature equilibrium at 20°C select Conductivity/Temperature (linear)
using the pH detector and sufficiently Temperature compensation
agitated standard solution) Automatic or manual in the temperature
Linearity: ±0.1 pH (using PH8EFP, PH8EHP) range of the C (measurement range)
±0.15 pH (using YOKOGAWA standard Temperature configurable 0 to 100ºC
detector other than the above) (default 25ºC)
Output item: select pH/Temperature (linear) Calculation
Temperature compensation: IEC60746-3 NaCI tables (default)
Automatic or manual. Compensation to Nernst Process compensation by configurable
equation. Process compensation by configurable temperature coefficient; Manual adjustment
temperature coefficient. Manual adjustment with with automatic stability check 0 to 10.0%/ºC
automatic stability check. Calibration
Calibration Manual, Automatic. Automatic with
Semi-automatic 1or 2 point calibration using pre- preset OlML (KCl) standard solution,
configured NIST, US, DIN buffer tables 4, 7 & 9, with automatic stability check. Sample
with automatic stability check. Manual adjustment. calibration
Note: According to OIML (International Organization of
■ Conductivity (SC) Legal Metrology) No.56 (1981), Electrode conductivity.
When 2nd Input is -C1 (SC) AC output
High accurate conductivity measurement by the 2 or 4 point 4-20 mA for the display,
embedded calibration standard table, temperature recording, operation control. Parameter is
compensation function, cell check function. selected conductivity/temperature.
Compatible to both two or four-electrode detectors, 5
types of temperature detector. 2.2 Electrical
Input Specification Display
Two or four electrodes measurement with square Display: QVGA color LCD with a touch screen
wave excitation, cell constants from 0.005 to 50.0 cm-1, Message Language:
YOKOGAWA SC4AJ, SC210G, SC8SG English, German, French, Chinese,
Input Range Korean, Spanish, Portuguese, Japanese
Note: You may find some image loss, uneven brightness,
Conductivity: or afterimage on display. They are generated from
min.; 0 µS/cm the product features but not defects
max.; 200 mS x (Cell constant) AC Power supply
(max. 2000 mS/cm) Ratings: 100 to 240 V AC
Temperature: Acceptable range; 90 to 264 V AC
Pt1000; -20 to 250ºC Ratings: 50/60 Hz
Pt100; -20 to 200ºC Acceptable range; 50 Hz ± 5 %, 60 Hz ± 5%
Ni100; -20 to 200ºC Power Consumptio: 35 VA
NTC 8k55; -10 to 120ºC Output signal
Pb36(JIS NTC 6k); -20 to 120ºC Isolated outputs: 4-20 mA DC
Output Range Accuracy: ± 0.02 mA
Conductivity: Repeatability: ± 0.02 mA
min.; 0.01 µS/cm Linearity: ± 0.02 mA
max.; 2000 mS/cm (max 90% zero Maximum load: 600 Ω
suppression) Number of outputs: select by suffix code
Temperature: 2 isolated outputs; -N2
min.; span 25ºC 4 isolated outputs; -N4
max.; span 270ºC (Not isolated between
Performance (Accuracy) analog outputs 1 and 2)
(The specifications are expressed with simulated inputs.) Output range: Configure within measurement range
Conductivity Minimum range; minimum measurement range
2 µS x K cm-1 to 200 mS x K cm-1: Maximum range; maximum measurement range
Linearity; ±0.5% F.S. (*) In case automatic range switching function is
Repeatability; ±0.5% F.S. enable, low range is fixed by 0 NTU.
1 µS x K cm-1 to 2 µS x K cm-1:
Linearity; ±1.0% F.S. Automatic range switching function:
Repeatability; ±1.0% F.S. Ch1; No switching or switching in 2 or 3 points
Temperature (Pt1000, PB36NTC, Ni100): You can select the followings.
Linearity; ±0.3ºC “Manual range switching, Remote range
Repeatability; ±0.1ºC switching, Automatic 2 ranges switching,
Temperature compensation: Automatic 3 ranges switching”
NaCl table; ±1 % Ch2-Ch4; No switching
Step response: You can select the followings.
90% (< 2 decades) in 7 seconds “Manual switching, Remote 2 range
Note: “F.S.” means maximum setting value of converter switching”
output.
“K” means cell constant.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
6
Output function: Linear output or 3 points polyline 2.3 Calibration, cleaning, diagnosis and other
output (Set upper/lower value and another functions
one point. The values between them are Calibration function:
filled by linear.) Zero calibration; Calibration by zero water
Output current; 2.4 to 21.6 mA Lamp-off calibration; If the turbidity to be measured
Output signal during maintenance (HOLD): exceeds 200 NTU, zero
Enable/Disable keep output value calibration is possible by
function during setting, calibration and turning off the light source
maintenance (Automatic hold function). Slope calibration; Check tool with zero water,
You can set HOLD value the last value or standard solution.
any value between 2.2 mA to 22.0 mA Sample correction; Zero point correction and
Burn out function (NAMUR 43): sensitivity correction by any
Analyzer can output 2.2 mA (Burn down) sample (actual process)
or to alarm failure. Automatic calibration function:
Burn down; 2.2 mA Automatic zero calibration with membrane
Burn up; 22.0 mA filtered water
Contact outputs Automatic cleaning:jet cleaning (configurable cleaning
Type: No-voltage contact output time/cycle) (only for the product type with
Number: 4 contacts (S1/S2/S3/S4) automatic cleaning)
Contact S4 is for Fail-safe function. Diagnostic function:
Function: Selectable 4 functions among Alarm, Heater disconnection, light source
maintenance mode, cleaning/calibration monitoring
status, mA1 output range status. 2.4 Mechanical and others
Identification of range switching: Housing: Die-cast aluminum
Contacts (S1/S2/S3) identifies mA Case:
output range. Up to 3 contacts are used Color; Silver gray
for 3 ranges. Refer to User’s manual Finish; Urethane coating, high anti-corrosion
when using 2 contacts to identify 3 coating
ranges. Window: Polycarbonate
Switch capacity: Protection: IP66 NEMA Type 4X (USA), CSA Type 4X
Maximum values; 100 VA, 250 V AC, 5A (*) (Canada)
Maximum values; 50 W, 24 V DC, 5A (*) Please add Hood option (/H) in case of
(*): This specification is in case of “-N2”. In case of using outdoor
“-N4”, maximum switch current capacity is 4 A.
Size: 165×165×168.5 mm (W×H×D)
Contact input
Not included cable gland
Type: Non-voltage contact output
Weight: Maximum 1.5 kg
Number: 2 contacts (-N4), 1 contact (-N2)
Mounting hardware (option): stainless fittings
Function: Select among Remote 2range switching,
-Pipe and wall mounting hardware
Remote 3range switching Remote
-Panel mounting hardware
cleaning start, Remote calibration start,
Hood (option):
Disable
-Stainless steel
Remote range switching:
-Stainless steel with chemically resistant
By contact input, mA output is switched
coating
to range “2” or range “3”. Range “3” is
Terminal screw size: M3 (Earth terminal: M4)
available only to mA1.
Cable gland: 8 holes, M20 Cable gland x 8
Short-circuit resistance (on resistance):
(For 6 to 12 mm outer diameter cable)
maximum 200 Ω Note: Please screw up them by yourself
Open resistance (off resistance): Conduit adapter (option): G1/2, 1/2NPT or M20 × 1.5
minimum 100k Ω Note: Please order required numbers.
Digital communication (option) Stainless Steel Tag Plate (option): In case you
Ethernet (Modbus TCP): specified Option code /SCT, Tag plate on
10/100 Mbps which is the number is printed is shipped
Cable length; maximum 100 m with product. You can attach this tag by
RS-485 (Modbus RTU): wire.
115200/38400/9600 bps Warm up time: Approx. 30 min
Cable length; Installation/Storage environment:
115200 bps; maximum 600 m Ambient Operating Temperature; -20 to55ºC
38400 bps,9600 bps; maximum 1200 m Strage Temperature; -30 to 70ºC
Analog Input (mA, O/I code: -N4) Humidity; 10 to 90% RH at 40ºC
Temperature compensation (pH, SC) (No condensation)
Isolated Input: 4 to 20 mA DC
Accuracy: ± 0.02 mA
Number: 1

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
7
2.5 Safety, EMC, and RoHS conformity standards
Safety:
CE: EN 61010-1, EN 61010-2-030
UL: UL 61010-1, UL 61010-2-030
CSA: CAN/CSA-C22.2 No.61010-1
CAN/CSA-C22.2 No.61010-2-030
Morocco/LVD Arrêté:
NM EN 61010 1
NM EN 61010 2 030
GB: GB30439
Installation altitude: 2000 m or less
Category based on IEC 61010: II (*)
Pollution degree based on IEC 61010: 2 (**)
(*) Installation category, called over-voltage category,
specifies impulse withstand voltage.
Equipment with Category II is energy-consuming
equipment to be supplied from the fixed installation.
(**) Pollution degree indicates the degree of existence
of solid, liquid, gas or other inclusions which may
reduce dielectric strength.
Degree 2 is the normal indoor environment.
EMC:
CE: EN 61326-1 Class A,Table 2 (For use in
industrial locations)
Influence of immunity environment
(Criteria A): Output shift is specified
within ± 10 % of F.S.
EN 61326-2-3
EN 61000-3-2
EN IEC 61000-3-2
EN 61000-3-3
RCM: EN61326-1 Class A
Korea Electromagnetic Conformity Standard
한국 전자파적합성 기준 Class A
Morocco/EMC Arrêté:
NM EN 61326 1
NM EN 61326 2 3
NM EN 61000 3 2
NM EN 61000 3 3
Environmental regulation:
RoHS Directive: EN IEC 63000
REACH: Regulation EC 1907/2006
Waste Electrical and Electronic Equipment (WEEE)
Directive:
This product is purposely designed to be
used in a large scale fixed installations
only and, therefore, is out of scope
of the WEEE Directive. The WEEE
Directive does not apply. The WEEE
Directive is only valid in the EU.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
8
n Model and Codes
Option
Model Suffix code Description
code
TB830D ········································· ·········· Surface Scattering Light Turbidity Detector
Light source -F8 ·········· White LED, Formazin, 0-2 NTU to 0-2000 NTU
and Turbidity -FS ·········· 860 nm, Formazin, 0-2 FNU/NTU to 0-2000 FNU/NTU
standard
Type -AB ·········· General purpose for CE, RCM, China standard
-AD ·········· General purpose for CSA
-AG ·········· General purpose for KC
-AJ ·········· General purpose (*1)
Sampling system -NN ·········· Without sampling system (*2)
-A1 ·········· With sampling system (without automatic cleaning, without automatic
zero calibration) (*3) (*4) (*13)
-A2 ·········· With sampling system (with automatic cleaning, without automatic zero
calibration) (*3) (*14)
-A3 ·········· With sampling system (with automatic cleaning, with automatic zero
calibration) (*3) (*14)
-A5 ·········· Without sampling system, with relay box (*2)
Material and Connection of -AN ·········· Without sampling system (Detector mounting bracket Carbon steel) (*5)
Sampling system -SN ·········· Without sampling system (Detector mounting bracket Stainless steel) (*5)
-AD ·········· Carbon steel stanchion, base mounting
-AB ·········· Carbon steel stanchion, rear mounting
-SD ·········· Stainless steel stanchion, base mounting
-SB ·········· Stainless steel stanchion, rear mounting
Solenoid valve for Sampling -NN ·········· Without Solenoid valve (*15)
system -10 ·········· 100 V AC
-12 ·········· 110 V AC
-20 ·········· 200 V AC
-22 ·········· 220 V AC
— -NN ·········· Always -NN
— -NN ·········· Always -NN
Option /L02 Connection cable for analyzer 2 m (*6)
/L03 Connection cable for analyzer 3 m (*6)
/L05 Connection cable for analyzer 5 m (*6)
/L10 Connection cable for analyzer 10 m (*6)
/L20 Connection cable for analyzer 20 m (*6)
/SCT Stainless steel tag plate
/U Pipe mounting hardware (SUS)
/R Rack or wall mounting hardware (SUS)
/CB Conduit adapter G1/2×1 pcs
/CD Conduit adapter 1/2NPT× 1 pcs
/CF Conduit adapter M20×1.5×1 pcs
/ARS With arrester (*7)
/PHN7 With pH (Using FLXA402) (without ultrasonic cleaning) (*8) (*10) (*13) (*14)
/PHU7 With pH (Using FLXA402) (with ultrasonic cleaning) (*8) (*10) (*13) (*14)
/TT3 With 500ml KCl reserve tank for pH (*10) (*11)
/FC With Free Available Chlorine Analyzer (*8) (*9) (*13) (*14)
/PHN2 With pH (Using FLXA402T 2nd input) (without ultrasonic cleaning) (*8)
(*11) (*13) (*14)
/PHU2 With pH (Using FLXA402T 2nd input) (with ultrasonic cleaning) (*8) (*11)
(*13) (*14)
/L Air denoising for Low Range (*12)
*1: “-AJ” is no standard.
*2: Install the head tank so that the flow rate of the detector is the specified flow rate (1.5 to 2 L/min). Customers have to
arrange a head tank.
*3: 1 μm zero turbidity filter is supplied with the product.
*4: Cleaning, zero calibration, span calibration are enabled manually.
*5: This is selectable only when Sampling system “-NN” or “-A5” is specified.
*6: This is not selectable when Sampling system “-A1”, “-A2”, “-A3” is specified. 1 m cable is supplied with the product. Select
this code for other cable length to change if necessary.
*7: This is not selectable when Type -AD is specified.
*8: This is selectable only when “-A1”, “-A2”, “-A3” for Sampling system is specified.
*9: For TB830D equipped with Non-reagent type Free Available Chlorine Analyzer. You need one converter just for the chlorine
analyzer. Selectable only when “-AJ” (no standard) is specified.
The TB830D cannot be equipped with both a pH sensor and a Non-reagent type Free Available Chlorine Analyzer.
Prepare the following products separately. Refer to GS 12F05B10-01EN for FC800D Non-reagent type Free Available
Chlorine Analyzer.
Sensor: FC800D-□□-AJ-NN-NN/L02
Converter: FLXA402T-A-B-AJ-CL-NN-N□-WR-N-□-J-ST-NN

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
9
*10: For TB830D equipped with pH sensor. You need one converter just for the pH sensor. Selectable only when “-AJ” (no
standard) is specified.
Prepare the following products separately. Refer to GS 12B07B02-E, GS 12J05C02-00E, GS 12A01F01-01EN, GS
19C01B05-01EN.
pH sensor: PH8EFP-03-TN-TT1-N-G*A (/PHN7 or /PHU7 and without /TT3)
PH8EFP-03-TN-TT3-N-G*A (/PHN7 or PHU7 and with /TT3)
pH holder: PH8HF-PP-JPT-T-NN-NN*A (/PHN7 without ultrasonic cleaning )
PH8HF-PP-JPT-T-S3-C1*A (/PHU7 with ultrasonic cleaning)
pH converter: FLXA402-A-B-AJ-P1-NN-A□-□□-N-□-J-NN (/PHN7 or /PHU7 )
Pulse generator for clean unit: PG400-A-B-AJ-PU-00-J-ST (/PHU7 with ultrasonic cleaning)
*11: For TB830D equipped with pH sensor. One converter can be shared with the scattered light turbidity detector.
Analog pH sensors can be used.
For the 2nd Input of the converter, specify “-P1” (pH) to connect an analog pH sensor
Prepare the following products separately. Refer to GS 12B07B02-00E, GS 12J05C02-00E, GS 19C01B05-01EN.
pH sensor: PH8EFP-03-TN-TT1-N-G*A (/PHN2 or /PHU2 and without /TT3)
PH8EFP-03-TN-TT3-N-G*A (/PHN2 or /PHU2 and with /TT3 )
pH holder: PH8HF-PP-JPT-T-NN-NN*A (/PHN2 without ultrasonic cleaning)
PH8HF-PP-JPT-T-S3-C1*A (/PHU2 with ultrasonic cleaning)
Pulse generator for clean unit: PG400-A-B-□-PU-00-□-ST
*12: For low turbidity (200 NTU or less) measurement range and the application which may generate bubbles, select this code
for head tank.
Do not select this code for high turbidity (above 200 NTU) measurement range because the influence of air bubbles is
small, and the turbidity in the sample may block the air bubble removal piping or reduce the flow rate.
*13: When the Sampling system “-A1” is specified, the sampling system does not support CE or Chinese standards. The
product cannot be sold to the area to which these regulations apply.
*14: When the Sampling system “-A2” or “-A3” is specified, the sampling system does not support CE, RCM, CSA, Chinese
standards. The product cannot be sold to the area to which these regulations apply.
*15: Specify “-NN” when the Sampling system “-NN” or “-A1” is selected.

 Accessories
Name Q´ty Remarks
Silicone cloth 1 Part number: K9210KS
Check tool 1 ―
Desiccant 1 For spare. 4 pcs/Q'ty (Part number: K9657RJ)
1 For operation. 4 pcs/Q'ty (Part number: K9657RJ)
Desiccant
(When the code for “Without sampling” is specified)
Pipe mounting hardware 1 For “/P” (Option)
(Option)
Rack mounting hardware 1 For “/R” (Option)
(Option)
Soft vinyl chloride tube 1 set (1 m x 2) For piping detector
(Ø33 × Ø25 Black) (when the code for “Without sampling” is specified)
1 For piping detector
Clamp
(when the code for “Without sampling” is specified

 Spare Parts
Name Part No. Description Frequency of
Replacement
LED ASSY (white) K8004BD When “-F8” is selected. 3 years
LED ASSY (IR) K8004BE When “-FS” is selected. 3 years
Desiccant (4 pcs/Q'ty)* K9657RJ ― Yearly
Drain tube K9411JM ― Yearly
Fuse A1633EF ― ―
Fuse A1624EF for Sampling system “-A2” “-A3”. The relay ―
box uses two fuses.
1 µm filter K9008ZD ― Yearly
*: Use within a year after purchasing.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
10
 Mounting Hardware (Option)
Select each corresponding hardware based on the table below.
Mounting type
Model
Wall Pipe Panel
TB830D /R /U (Pipe mounting) NA
FLXA402T /UM* or /U /UM* or /U /UM* or /PM
*: Universal mounting kit (/UM) contains pipe, wall mounting hardware (/U) and panel mounting hardware (/PM).

 Parts for Sampling system


Name Part No.
Pinch valve for drain (100V, 110V) K9411JG
Pinch valve for drain (200V, 220V) K9411JH
Solenoid valve for cleaning water or zero water (100V) A1014MZ
Solenoid valve for cleaning water or zero water (110V) A1016MZ
Solenoid valve for cleaning water or zero water (200V) A1015MZ
Solenoid valve for cleaning water or zero water (220V) A1017MZ
Solenoid valve for sample (100V, 110V) K9411DT
Solenoid valve for sample (200V, 220V) K9411DU
Head tank (with manual valve) K9411GC
Head tank (with pinch valve 100V, 110V) K8004FB
Head tank (with pinch valve 200V, 220V) K8004FC
Head tank mounting hardware K8004LD
Zero turbidity filter (1 μm) K9411UA

 Required Number of Conduit Adapters


If you need to provide conduit work with cables, specify the conduit adapter by Option code.
Conduit adapters and dedicated cable glands, which are used in place of the standard cable gland for cable entry
holes, are supplied with the product.
You can specify the conduit adapter by Option code for both TB830D and FLXA402T, however, be aware of the
following.
• When FLXA402T Digital communication “-E” (Modbus TCP/IP) is selected
If you attach a conduit adapter on the Modbus TCP/IP cable entry, you need an Ethernet dedicated conduit
adapter. Be sure to specify FLXA402T “/C□6”.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
11
Next diagram shows the position of each cable entry () where you can attach conduit adapters for the conduit.

• TB830D-....-NN-□N-NN-NN/ARS/C□

Power Terminal TB830D *2 FLXA402T


supply box *1 Power mA output/input
A 1m *2
1st input mA output/input*3
2nd input*2 Contact outputs*3
1,2,3,5,10,20 m
Contact outputs*3
Digital Communication*3

• TB830D-....-NN-□N-NN-NN/C□

Power Terminal TB830D FLXA402T


supply box *1 Power mA output/input
B 1m 1st input mA output/input*3
2nd input*2 Contact outputs*3
1,2,3,5,10,20 m
Contact outputs*3
Digital Communication*3

• TB830D-....-A1-NN... /ARS/C□ *4

Power Terminal TB830D FLXA402T


supply box *1 Power mA output/input
A 1m 1st input mA output/input*3
1m 2nd input*2 Contact outputs*3
Contact outputs*3
Digital Communication*3

• TB830D-....-A1-NN... /C□ *4
Power Terminal TB830D FLXA402T
supply box *1 Power mA output/input
B 1m 1st input mA output/input*3
1m 2nd input*2 Contact outputs*3
Contact outputs*3
Digital Communication*3

• TB830D-....-A2--NN... /C□ *4 


TB830D-....-A3--NN... /C□ *4
  TB830D FLXA402T
*1 1m
Power mA output/input
1st input mA output/input*3
2nd input*2 Contact outputs*3
Contact outputs*3
1m
Digital Communication*3

Relay
box
Power supply

• TB830D-....-A2--NN... /ARS/C□ *4 


TB830D-....-A3--NN... /ARS/C□ *4
 
 
TB830D FLXA402T
*1 1m
Power mA output/input
1st input mA output/input*3
2nd input*2 Contact outputs*3
Contact outputs*3
1m
Digital Communication*3

Terminal Relay
box 1m box
Power
supply A

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
12

• TB830D-....-A5--NN... /C□ *4 

TB830D *2 FLXA402T
*1 Power mA output/input
*2
1st input mA output/input*3
2nd input*3 Contact outputs*3
Contact outputs*3
1m
Digital Communication*3

Relay 1,2,3,5,10,20 m
Power box
supply

• TB830D-....-A5--NN... /ARS/C□ 

TB830D *2 FLXA402T
*1 Power mA output/input
*2
1st input mA output/input*3
2nd input*3 Contact outputs*3
Contact outputs*3
1m
Digital Communication*3

Terminal Relay 1,2,3,5,10,20 m


box 1m box
Power
supply A

*1: Conduit piping is not allowed to TB830D because the wiring inlet is on the moving part.
*2: The standard cable length between FLXA402T and TB830D is 1 m. You can change the cable length by specifying a code
from Option code.
*3: The number of cable entry holes to be used are defined by the specification as below.

Spec. 2nd Input mA output/input Contact Outputs Digital Communication


Code -NN the others -N2 -N4 -WR -NR -N -E -R
Required No.of
0 1 1 2 2 0 0 1* 1
cable entry holes
*: Conduit exclusively for Ethernet
*4: When /PH□2, /PH□7, or /FC is specified, FLXA402 or FLXA402T can be connected. Conduit can be provided to these
products' mA I/O, contact output, digital communication wiring.

● Solenoid valve (SV1 to SV4)


Requirement for solenoid valves.

Port size Valve type Max. working pressure Max. Working pressure
differential
SV1 (Measure) 15A Motor valve 3 terminals 1 MPa 1 MPa
(Open, Shut, Common)
SV2 (Wash) RC3/8 Solenoid valve N.C. (*1) 1.5 MPa 0.5 MPa
SV3 (Zero) RC3/8 Solenoid valve N.C. (*1) 1.5 MPa 0.5 MPa
SV4 (Drain) Φ21×Φ18 Silicon tube Pinch valve N.C. (*1) ― ―
*1: Valve opens when energized.
Recommended cable for solenoid valves.

Cable length 1.2 m


Rated voltage 300 V AC
Specification SV1 AWG17 3-core cable, M3 round terminal
SV2, SV3, SV4 AWG20 2-core cable, M3 round terminal

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
13
FLXA402T Liquid Analyzer for Turbidity and Chlorine
Liquid Analyzer FLXA402T is a common analyzer for Turbidity detector and Chlorine sensor unit. The table below for
Model and Suffix code (MS code) applies the combination between FLXA402T and TB830D.

Model Suffix code Option code Description


FLXA402T ··········································································· ················· Liquid Analyzer for Turbidity and Chlorine
Power supply -A ················· AC version
Housing -B ················· Aluminum alloy cast + urethane coating
(*1) -D ················· Aluminum alloy cast + high anti-corrosion coating
Type -AB ················· General purpose for CE, RCM, China standard
-AD ················· General purpose for CSA
-AG ················· General purpose for KC
-AJ ················ General purpose
1st Input -TB ················· SENCOM for Turbidity
2nd Input (*2) -NN ················· Without Input
-P1 ················· pH
-C1 ················· Conductivity (SC)
-TB ················· SENCOM for Turbidity (*3)
mA Output/Input -N2 ················· 2 x 4-20 mA Output + 1 x Contact Input (without HART)
-N4 ················· 4 x 4-20 mA Output + 2 x Contact Input + 1 x 4-20
mA Input (without HART)
Contact Outputs -WR ················· Contact outputs (Wash and Fail contact outputs)
-NR ················· Without Contact outputs (without Wash and Fail
contact outputs)
― -N ················· Always -N
Digital Communication -N ················ Without Digital communication
-E ················· Modbus TCP/IP
-R ················· Modbus RTU (RS-485)
Country (*4) -N ················ Global except Japan
-J ················ Japan
Assembled in stanchion -NN ················· Not assembled in stanchion (*9)
-ST ················· Assembled in stanchion
― -NN ················· Always -NN
Option Mounting hardware (*5) /UM Universal mounting kit (*6)
/U Pipe and wall mounting hardware
/PM Panel mounting hardware
Hood /H6 Hood, stainless steel
/H7 Hood, stainless steel + urethane coating
Tag plate /SCT Stainless steel tag plate
Conduit adapter (*7) /CB4 G1/2 x 4 pcs
/CD4 1/2 NPT x 4 pcs
/CF4 M20 x 1.5 x 4 pcs
/CB6 G1/2 x 3 pcs + G1/2 for Ethernet x 1 pcs (*8)
/CD6 1/2 NPT x 3 pcs + 1/2 NPT for Ethernet x 1 pcs (*8)
/CF6 M20 x 1.5 x 3 pcs + M20 for Ethernet x 1 pcs (*8)
*1: Urethane coating is acid resistant. Heavy duty coating is acid/alkali resistant.
*2: To detectors connected to 2nd input, none of the following is available: automatic cleaning, automatic zero calibration, wiper
cleaning.
If you want to select FC800D and TB830D, select FC800D (-CL: SENCOM SA for Chlorine) on the first input. Select “-TB”
on the second input of FLXA402T in FC800D (GS 12F05B10-01EN).
*3: SENCOM Turbidity is for TB820D (GS 12E01B30-01EN) or TB830D.
*4: Select only “-J” if you use the converter in Japan. Only SI unit (International System of Units) applies.
*5: See “ Mounting Hardware (Option)”.
*6: Universal mounting kit contains pipe, wall mounting hardware (/U) and panel mounting hardware (/PM).
*7: See “ Required Number of Conduit Adapters”.
*8: Available only when “-E” (Modbus TCP/IP) to Digital communication is selected.
*9: When “-ST” is specified, neither Mounting hardware (/UM, /U, /PM) nor Hood (/H6, /H7) can be selected.
For TB830D, when a Sampling system “-A□” (with sampling system) is selected, always select “-ST”. For the second input,
always select “-NN” (Without input). However, when "/PH□2" is selected, select “-P1” (pH) for the second input.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
14
■ Optional parts
Name Parts number Quantity Remark
Mounting for pipe, wall mounting (stainless) K9703SS 1 set same as Option code /U
hardware for panel mounting (stainless) K9703ZD 1 set same as Option code /PM
Sun shade Stainless K9698WK 1 set same as Option code /H6
hood stainless + urethane K9698WL 1 set same as Option code /H7
Rubber plug attachment K9334CN 1 pcs for Cable gland
Fuse A1633EF 1 pcs 250V/2.5A (minimum 5 pcs)
SD card A1019NL 1 pcs 2 GB industrial SD card (with power failure recovery)
Customers can provide the cards with spec:
Storage capacity:
128 MB or greater
Type: SD, SDHC

n Dimensions (TB830D Surface Scattering Light Turbidity Detector)


 Without sampling system (-NN, -A5)
 TB830D Surface Scattering Light Turbidity Detector
Unit: mm

246 (22) 150


208
296

38
50

4-ø4.2 hole
14.5

142 100
4-ø6.2 hole
56

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
15
Unit: mm
 Pipe mounting hardware (/U) for -NN, -A5 Without sampling system
(22) (215)
150

Pipe 50A
(or 2B)
4-M5

Approx. 372

35
157
203
38
U-Bolt

4-ø10 56
Panel mounting
hardware: /U
70
100
208 (140)

 Rack or wall mounting hardware (/R) for -NN, -A5 Without sampling system
Rack or wall mounting
hardware: /R

3-ø6.2
Approx. 295

125
105
38

4-M5 18 ø10
56
(22) 166 25 190
220

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
16
 Terminal box A
Terminal box A is supplied when “/ARS” option code is specified.
When “/C□” option code is specified, Terminal box A can be conduit wired.
Unit: mm

275 11 92 70

17
203

75
(12)

 Installation of pipe mounting hardware (/U) on the terminal box A


275 (22) 92 (52)
70 10 4-M5
Pipe 50A
(or 2B)

35
17
203

75
277

142
U-Bolt

200
60

2-ø9
Panel mounting (24)
Panel mounting 70
hardware: /U
100
hardware: /U

 Installation of rack or wall mounting hardware (/R) on the terminal box A


275 11 92

4-M5
10
17
203

75
235

130
160

Panel mounting
70 4-ø6.5
hardware: /R
110
30
130

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
17
 Terminal box B
Terminal box B is supplied with the product in the following cases.
• When both Sampling system “-NN” and Option “/C□” (Conduit adapter) are specified, but “/ARS” (with Arrester) is
not specified.
• When Sampling system “-A1” is specified but “/ARS” (with Arrester) is not specified.
• When Sampling system “-A2” or “-A3” is specified and Option “/PH□2” or “/PH□7” or “/FC” is specified, but “/ARS”
(with Arrester) is not specified.
Terminal box B can be conduit wired.
Unit: mm

132 (31) 66

144
(31)
23.3

2-M4
114.1

 Installation of pipe mounting hardware (/U) on the terminal box B


132.5 4 66 (59)

Pipe 50A
(or 2B)
143.5

216.5

U-Bolt 2-ø9 73
60

56

Panel mounting (28) Panel mounting


hardware: /U hardware: /U 96
100
15

2-M4
27
90

114
140

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
18
 Installation of rack or wall mounting hardware (/R) on the terminal box B
Unit: mm

132.5 4 66
143.5

216.5

20
30
73
Panel mounting 2-ø10 60 2-ø5.5
hardware: /R 70
20 90

2-M4

27

90
114
140

 Relay box
A relay box is supplied with the product when “-A2”, “-A3”, or “-A5” is specified. A relay box is used to drive the
solenoid valve for Automatic cleaning/Automatic zero calibration. Select /C□ (conduit adapter) to provide conduit
wiring to the relay box.

168.5
158.5
165 40 80 4-M6
165

80

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
19
 Installation of pipe mounting hardware (/U) on the relay box
Unit: mm

165 168 (52)


Pipe 50A
4-M6
(or 2B)

8
165

80
235
U-Bolt

200
60

50
Panel mounting 2-ø9
(22) 70
hardware: /U
Panel mounting 80
hardware: /U 100

 Installation of rack or wall mounting hardware (/R) on the relay box


165 168 13
4-M6

8
165

80
235

200
35
15
Panel mounting 70
4-ø10
hardware: /R 80
100

 Conduit adaptor (/CB, /CD, /CF)

Case Unit: mm
(inch)
Cable gland
for adapter
Approx.55 Adapter
(2.17) 45
(1.77)

G1/2 (/CB), 1/2NPT (/CD), M20x1.5 (/CF)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
20
l Zero Turbidity Filter
OUT Unit: mm
4-M4
fixing screws

112.5
44
IN

80
ø115
ø121

Vent plug Piping inlet & outlet Rc1/2

22 22
314

Filter

Part No. Filtering size


K9411UA 1 μm
10

 With sampling system (-A1, -A2, -A3)


l Maintenance space
Maintenance space Maintenance space Unit: mm
Approx. 500 530 Approx. 500

Maintenance space
4-ø15 holes Approx. 500
(for M12 Anchor bolts) (100)

Stand 350 550

20 490 (20) 100 Maintenance space


Approx. 500

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
21
l With sampling system (with stanchion, without automatic cleaning, without automatic zero calibration)
TB830D-□-□-A1-□-NN-NN-NN
4-ø15 hole Unit: mm

(100)
For anchor bolts

*1: Applied to suffix code -AB, -SB

350
When you specified -AD and -SD, connection of
sampling system is base mounting.
*2: Applied to option code /L
20 490 (20)

100
Terminal box Turbidity converter
*2
For /ARS

Turbidity Head
detector tank

Manually Manually
*2
operated Drainage port
valve V4 operated VP40 pipe
valve V5
Manually 1453
operated Manually Sample water Tap water inlet
valve V2 operated inlet VP16 pipe VP16 pipe
valve V3
93*1
80*1

150
Zero-turbidity Manually
reference operated

253
250

filter valve V1

70 385 530
150 150
50 550

l With sampling system (with stanchion, with without automatic cleaning, without automaticzero
calibration)
TB830D-□-□-A1-□-NN-NN-NN/PH□2
4-ø15 hole Unit: mm
(100)

For anchor bolts


*1: Applied to suffix code -AB, -SB
When you specified -AD and -SD, connection of
sampling system is base mounting.
350

*2: Applied to option code /L


*3: Applied to option code /PHU2
If /PHN2 is specified, the ultrasonic oscillator will
20 490 (20) not be attached.
100

Terminal box Turbidity, pH converter KCI tank


For /ARS For /TT3
1495

Head
Ultrasonic Turbidity tank Manually
oscillator*3 detector operated
1453

valve V5
pH sensor Manually Drainage port
*2
operated VP40 pipe
Flow-through valve V6
Manually type holder
operated Sample Tap water inlet
valve V2 water inlet VP16 pipe
Manually VP16 pipe
operated
valve V4
*1

80*1

150
93

Manually
operated
253
250

valve V3
Zero-
turbidity
reference Manually
70 385 filter 530 operated
valve V1 150 150
50 550

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
22
l With sampling system (with stanchion, without automatic cleaning, without automatic zero calibration)
TB830D-□-□-A1-□-NN-NN-NN/PH□7
4-ø15 hole Unit: mm

(100)
For anchor bolts
*1: Applied to suffix code -AB, -SB
When you specified -AD and -SD, connection of
sampling system is base mounting.

350
*2: Applied to option code /L
*3: Applied to option code /PHU7
If /PHN7 is specified, the ultrasonic oscillator will
20 490 (20)

100
not be attached.
Terminal box Turbidity converter KCI tank
pH converter For /TT3
For /ARS

1480
Turbidity Head
Ultrasonic detector tank
oscillator*3 Manually
pH sensor operated Drainage port
valve V5 VP40 pipe
1453
*2
Flow-through Manually
Manually type holder operated
operated valve V6 Tap water inlet
valve V2 Sample water inlet VP16 pipe
Manually
operated VP16 pipe
valve V4

150
93 *1
80 *1
Manually
operated
valve V3

253
250

Zero-
turbidity
reference Manually
70 385 filter 530 operated
valve V1 150 150
50 *1 550

l With sampling system (with stanchion, without automatic cleaning, without automatic zero calibration)
TB830D-□-□-A1-□-NN-NN-NN/FC

Unit: mm
(100)

4-ø15 hole
For anchor bolts

*1: Applied to suffix code -AB, -SB


350

When you specified -AD and -SD, connection of


sampling system is base mounting.
*2: Applied to option code /L
20 490 (20)
100

Terminal box
Turbidity converter Free chlorine converter
For /ARS

Turbidity Head Needle


detector tank valve Manually
V7
Free chlorine operated
detector valve V5 Drainage port
*2
Manually VP40 pipe
operated
1453

Manually valve
operated V6 Sample Tap water inlet
valve V2 Manually water inlet VP16 pipe
operated VP16 pipe
valve V4
150
*1

80 *1

Manually
93

operated valve
253

V3
175

Zero-
turbidity
70 385 reference Manually
filter 530 operated
valve V1
150 150
50 *1 550

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
23
l With sampling system (with stanchion, with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN

(100)
4-ø15 hole Unit: mm
For anchor bolts

*1: Applied to suffix code -AB, -SB


When you specified -AD and -SD, connection of

350
sampling system is base mounting.
*2: Applied to option code /L
*3: Applied to option code /ARS

100
20 490 (20)

Terminal box*3 Turbidity converter


Relay box
*2

Turbidity detector Head Solenoid valve


tank SV2

Pinch valve Drainage port


*2 Manually SV4
operated
valve V4
1453 VP40 pipe
Manually Sample
operated water inlet Tap water inlet
valve V2 Needle valve VP16 pipe VP16 pipe
V3

93 *1
80 *1

150
Zero-turbidity

253
250

reference
filter

70 385 530 Manually


operated 150 150
valve V1
50 *1 550

l With sampling system (with stanchion, with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN/PH□2
(100)

4-ø15 hole Unit: mm


For anchor bolts
*1: Applied to suffix code -AB, -SB
When you specified -AD and -SD, connection of
sampling system is base mounting.
350

*2: Applied to option code /L


*3: Applied to option code /PHU2
If /PHN2 is specified, the ultrasonic oscillator will
20 490 (20)
100

not be attached.
Terminal box Turbidity converter KCI tank
For /ARS
*2
For /TT3
Relay box
1495

Turbidity Solenoid valve


detector Head SV2
tank
Ultrasonic
oscillator*3 Pinch valve Drainage port
*2
pH sensor SV4 VP40 pipe
1453

Flow- Sample water


Manually through inlet VP16 pipe
operated Tap water inlet
type holder Manually VP16 pipe
valve V2
Manually operated
operated valve V6
150

valve V4
93 *1
80 *1

Needle valve
V3
253
175

Zero-
turbidity
reference Manually
70 385 filter 530 operated
valve V1 150 150
50 *1 550

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
24
l With sampling system (with stanchion, with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN/PH□7
Unit: mm

(100)
4-ø15 hole
For anchor bolts
*1: Applied to suffix code -AB, -SB
When you specified -AD and -SD, connection of
sampling system is base mounting.

350
*2: Applied to option code /L
*3: Applied to option code /PHU7
If /PHN7 is specified, the ultrasonic oscillator
20 490 (20)

100
will not be attached.
Terminal box Turbidity converter pH sensor KCI tank
For /ARS For /TT3
*2
Relay box

1480
Head Solenoid valve
Turbidity SV2
Ultrasonic detector tank
oscillator *3
Pinch valve SV4 Drainage port
*2
pH sensor VP40 pipe
1453
Manually Flow-through Sample water
operated type holder inlet VP16 pipe Tap water inlet
valve V2 Manually VP16 pipe
Manually operated
operated valve
valve V4

150
V6
*1

80 *1
Needle valve
93
V3

253
Zero-turbidity
175

reference
filter
Manually
70 385 530 operated
valve V1 150 150
50 *1 550

l With sampling system (with stanchion, with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN/FC

Unit: mm
(100)

4-ø15 hole
For anchor bolts

*1: Applied to suffix code -AB, -SB


350

When you specified -AD and -SD, connection


of sampling system is base mounting.
*2: Applied to option code /L
20 490 (20)
100

Terminal box Turbidity


converter Free chlorine converter
For /ARS
Relay box
*2

Head Solenoid valve


SV2
1453

Turbidity tank
detector
Free chlorine Manually Pinch valve Drainage port
*2
detector operated SV4 VP40 pipe
valve V1 Sample
Manually water
Manually operated inlet
operated valve V4 Tap water inlet
Manually VP16 pipe VP16 pipe
valve V2 operated
valve V6
150
93 *1
80 *1

Needle valve
V3
253

Zero-turbidity
175

reference
filter
70 385 530 Needle valve
V7 150 150
50*1 550

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
25
l With sampling system (with stanchion, with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN

Unit: mm

(100)
4-ø15 hole
For anchor bolts

*1: Applied to suffix code -AB, -SB

350
When you specified -AD and -SD, connection
of sampling system is base mounting.
*2: Applied to option code /L
*3: Applied to option code /ARS

100
20 490 (20)

Terminal box *3 Turbidity converter *2

Relay box
Solenoid valve Solenoid valve
Turbidity Head SV3
detector SV2
tank
Pinch valve

1453
Manually SV4 Drainage port
*2 operated VP40 pipe
valve V4 Sample
Solenoid water inlet
Manually Needle valve valve Tap water inlet
SV1 VP16 pipe VP16 pipe
operated V3
valve V2

93 *1
80 *1

150
Zero-turbidity
reference filter Manually
operated
289

253
valve V1

70 385 530
150 150
50 *1
550

l With sampling system (with stanchion, with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN/PH□2
4-ø15 hole Unit: mm
(100)

For anchor bolts


*1: Applied to suffix code -AB, -SB
When you specified -AD and -SD, connection
of sampling system is base mounting.
350

*2: Applied to option code /L


*3: Applied to option code /PHU7
If /PHN7 is specified, the ultrasonic oscillator
100

20 490 (20) will not be attached.

Terminal box KCI tank


Turbidity, pH converter For /TT3
For /ARS Relay box
1495

Turbidity Head
detector tank
Solenoid Solenoid
Ultrasonic valve valve
1453

oscillator*3 SV3 SV2


*2
Pinch valve
pH Sensor SV4 Drainage port
VP40 pipe
Flow-through Manually
type holder operated Sample Tap water inlet
Manually water inlet
operated valve V1 VP16 pipe
valve V2 Manually VP16 pipe
operated
150

valve V4
93 *1
80 *1

Needle valve
V3
253
175

Zero-
turbidity
reference Manually
70 385 filter 530 operated
valve V6 150 150
50*1 550

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
26
l With sampling system (with stanchion, with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN/PH□7

(100)
4-ø15 hole Unit: mm
For anchor bolts

*1: Applied to suffix code -AB, -SB


When you specified -AD and -SD, connection

350
of sampling system is base mounting.
*2: Applied to option code /L
*3: Applied to option code /PHU7
20 490 (20) If /PHN7 is specified, the ultrasonic oscillator will not be attached.

100
Turbidity converter pH converter KCI tank
Terminal box *2 Relay box
For /ARS

Turbidity Head Manually operated

1480
detector tank valve SV2
Ultrasonic

1453
oscillator *3 pH sensor Manually
operated Drainage port
*2
Solenoid valve SV3 VP40 pipe
Flow-through valve
type holder SV1
Manually Sample Tap water inlet
operated Manually Manually water inlet VP16 pipe
valve V2 operated operated VP16 pipe
valve V4 valve V6

150
93 *1
Needle valve 80 *1
V3

253
Zero-turbidity
175

reference
filter
Manually
70 385 530 operated 150 150
50*1 550 valve V1

l With sampling system (with stanchion, with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN/FC
(100)

4-ø15 hole Unit: mm


For anchor bolts

*1: Applied to suffix code -AB, -SB


When you specified -AD and -SD, connection
350

of sampling system is base mounting.


*2: Applied to option code /L

20 490 (20)
100

Terminal box Turbidity Free chlorine converter


For /ARS converter *2
Relay box
Turbidity
detector Head Solenoid Solenoid valve
tank
1453

valve SV3 SV2


Free chlorine
detector Needle Pinch valve
valve V7 SV4 Drain outlet
*2
VP40 pipe
Manually Solenoid
Manually operated valve
operated SV1 Sample inlet
valve V4 Tap water inlet
valve V2 VP16 pipe VP16 pipe
Needle Manually
operated
valve V3 valve V6
150
93 *1
80 *1

Zero-turbidity
reference filter
253
175

Manually
70 385 530 operated valve
V1
150 150
50*1 550

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
27
n Dimensions (FLXA402T Liquid Analyzer for Turbidity and Chlorine)
FLXA402T Liquid Analyzer for Turbidity and Chlorine

168.5 Unit: mm
158.5
165 40 80
165

137
80
For Sensor modules 4-M6 depth 8
For IO module

For Communication module

For power supply

For Relay module


137

l Pipe mounting hardware (/U, /UM)


Unit: mm
165
(222.5)
168.5 Pipe 50A
(ø60.5)
4-M6 *
184.5

200
50

M8 U-bolt
Pipe mounting (Horizontal) Pipe mounting (Vertical)
100
*: Tighten the four screws to a torque of 2 N•m.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
28
Wall mounting hardware (/U, /UM)

Unit: mm
165 181.5 4-M6 *
13
165
234.5

200
50
For wall mounting
3-ø10 holes

15
70
100
*: Tighten the four screws to a torque of 2 N•m.
Note: For wall mounting, the wall should be strong enough to bear the weight of 8 kg or more.
l Panel mounting hardware (/PM, /UM)

Unit: mm
Panel thickness
1 to 12

+1
135

2-M5 length 35 138 0


+1
138 0

4-M6 * 195
100

185

Spacing panel cutout

*: Tighten the four screws to a torque of 2 N•m.


178

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
29
Conduit adaptor (/CB4, /CD4, /CF4/, /CB6, /CD6, /CF6)

Case Case Unit: mm


(inch)
Cable gland Cable gland
for adapter for Ethernet
Approx.55 Adapter Approx.55 Adapter
(2.17) 45 (2.17) 45 for Ethernet
(1.77) (1.77)

G1/2 (/CB4, /CB6) G1/2 (/CB6)


1/2NPT (/CD4, /CD6) 1/2NPT (/CD6)
M20x1.5 (/CF4, /CF6) M20x1.5 (/CF6)
Note: Universal Mounting kit (/UM) contains pipe, wall mounting hardware (/U) and panel mounting hardware (/PM).

l Hood Stainless steel (/H6, /H7)


184 Unit: mm
165 220 9

57
72
199

165

(50)
(70)
(100)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
30
n Piping
 -A1 With sampling system (without automatic zero cleaning, without automatic zero calibration)
TB830D-□-□-A1-□-□-NN-NN
<Tubing materials>
Head tank
Ø8 / Ø6 Polyethylene tube Detector *1
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
VP16 Rigid PVC pipe
VP25 Rigid PVC pipe
VP40 Rigid PVC pipe
Manually
operated
Manually operated valve V1 valve V5
Sample water (VP16)

Tap water (VP16) *2


Manually Manually
operated operated
valve V2 valve V3
Zero-turbidity
filter
PVC tray
Manually
operated valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.
 -A1 With sampling system (without automatic zero cleaning, without automatic zero calibration)
TB830D-□-□-A1-□-□-NN-NN/PH□

<Tubing materials>
Ø8 / Ø6 Polyethylene tube
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
VP16 Rigid PVC pipe Head tank
Detector *1
VP25 Rigid PVC pipe
VP40 Rigid PVC pipe

Sample water (VP16)


Manually
operated
valve V1 Manually
operated
valve V5
Manually
operated pH sensor
valve V6
Tap water (VP16) *2
Manually Manually
operated operated
valve V2 valve V3
Zero-turbidity
filter
PVC tray
Manually
operated valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
31
 -A1 With sampling system (without automatic cleaning, without automatic zero calibration)
TB830D-□-□-A1-□-□-NN-NN/FC
<Tubing materials>
Ø8 / Ø6 Polyethylene tube
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
VP16 Rigid PVC pipe Free available chlorine
VP25 Rigid PVC pipe analyzer detector Head tank Turbidity
detector *1
VP40 Rigid PVC pipe

Needle valve
V7
Sample water (VP16)
Manually
operated Manually
valve V1 operated
valve V5
Manually
operated
valve V6
Tap water (VP16) *2
Manually Manually
operated operated
valve V2 valve V3
Zero-turbidity
filter
PVC tray
Manually
operated valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

 -A2 With sampling system (with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN

<Tubing materials>
Head tank
Ø8 / Ø6 Polyethylene tube Detector *1
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
VP16 Rigid PVC pipe
Silicone tube
VP40 Rigid PVC pipe

Manually operated valve V1 Solenoid valve


Sample water (VP16) Pinch SV2
valve
SV4
Tap water (VP16) *2
Manually Needle
operated valve V3
valve V2
Zero-turbidity
filter
PVC tray
Manually operated
valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
32
 -A2 With sampling system (with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN/PH□
<Tubing materials>
Ø8 / Ø6 Polyethylene tube
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black) Head tank
VP16 Rigid PVC pipe Detector *1
Silicone tube
VP40 Rigid PVC pipe

Sample water (VP16)


Manually
operated
valve V1
Pinch
valve
Manually SV4
operated Solenoid valve
pH sensor SV2
valve V6
Tap water (VP16) *2
Manually Needle
operated valve V3
valve V2

Zero-turbidity reference filter


PVC tray
Manually operated
valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

 -A2 With sampling system (with automatic cleaning, without automatic zero calibration)
TB830D-□-□-A2-□-□-NN-NN/FC

<Tubing materials>
Ø8 / Ø6 Polyethylene tube
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
Free available chlorine
VP16 Rigid PVC pipe Turbidity
analyzer detector Head tank
Silicone tube detector
*1
VP40 Rigid PVC pipe

Needle valve
V7
Sample water (VP16)
Manually
operated
valve V1 Pinch
valve
Manually SV4
operated Solenoid
valve V6 valve
SV2
Tap water (VP16) *2
Manually Needle
operated valve V3
valve V2
Zero-turbidity
filter
PVC tray
Manually operated
valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
33
 A3 With sampling system (with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN
<Tubing materials>
Ø8 / Ø6 Polyethylene tube Head tank
Detector *1
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
VP16 Rigid PVC pipe
Silicone tube
VP40 Rigid PVC pipe

Motor-operated Solenoid
valve SV1 valve
Pinch SV2
Sample water (VP16) valve
Manually operated valve V1 SV4

Tap water (VP16) *2


Manually Needle Solenoid valve
operated valve V3 SV3
valve V2

Zero-turbidity filter

PVC tray
Manually operated
valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

 A3 With sampling system (with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN/PH□
<Tubing materials>
Ø8 / Ø6 Polyethylene tube
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net
Ø33 / Ø25 Soft PVC tube (black)
VP16 Rigid PVC pipe Head tank
Detector *1
Silicone tube
VP40 Rigid PVC pipe

Motor-operated valve SV1


Sample water (VP16)
Manually
operated Pinch
valve V1 valve
SV4
Manually
operated pH sensor Solenoid Solenoid valve
valve V6 valve SV2
SV3
Tap water (VP16) *2
Manually
operated Needle
valve V2 valve V3

Zero-turbidity
filter
PVC tray
Manually operated
valve V4
Drain (VP40)

*1: Option Code /L (Air denoising for Low Range)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
34
 A3 With sampling system (with automatic cleaning, with automatic zero calibration)
TB830D-□-□-A3-□-□-NN-NN/FC
<Tubing materials>
Ø8 / Ø6 Polyethylene tube
Ø12 / Ø9 Polyethylene tube
Ø22 / Ø15 PVC tube with soft net Free available chlorine
Ø33 / Ø25 Soft PVC tube (black) analyzer detector Turbidity
Head tank
VP16 Rigid PVC pipe detector *1
Silicone tube
VP40 Rigid PVC pipe Motor-
operated
Needle valve valve SV1
V7
Sample water (VP16)
Manually
operated

M
valve V1

Manually Pinch valve Solenoid valve


operated SV4 SV2
valve V6

Tap water (VP16) *2


Manually Needle Solenoid
operated valve V3 valve
valve V2 SV3

Zero-turbidity filter

PVC tray
Manually operated
valve V4
Drain (VP40)

*1: Option Code /L (For bubble retardant)


*2: You should install a check valve just before the tap water pipe to prevent backflow.

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
35
n Wiring
● TB830D-□-□-NN-NN-NN-NN-NN
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Liquid Analyzer for Turbidity and Chlorine
L (101) L
L (101) Power (201) L
Power N (102)
N (102) Converter (202) N N Power (61) + Analog output (-N2, -N4)
Supply G (001) mA1 (62) –
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


(21)
(22) Contact input (-N2, -N4)
(32) NC (24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO
B− (92) RS485 (-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet (-E)

● TB830D-□-□-NN-NN-NN-NN-NN /ARS
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box A for power supply
Liquid Analyzer for Turbidity and Chlorine
L (101) L (201) L
L (101) Power (201) L
N (102) CP N (202)
Power N (102) Converter (202) N N Power (61) + Analog output (-N2, -N4)
G (001) G (003) mA1 (62) –
Supply (002) G G
G (001)
(65) +
mA2 (66) – Analog output (-N2, -N4)
AR
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


(21)
(22) Contact input (-N2, -N4)
(32) NC (24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO
B− (92) RS485 (-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet (-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
36
● TB830D-□-□-NN-NN-NN-NN-NN /C□
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box B for power supply
Liquid Analyzer for Turbidity and Chlorine
L (101) L (201) L
L (101) Power (201) L
N (102) N (202)
Power N (102) Converter (202) N N Power (61) + Analog output (-N2, -N4)
G (001) G (003) mA1 (62) –
Supply (002) G G
G (001)
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


(21)
(22) Contact input (-N2, -N4)
(32) NC (24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO
B− (92) RS485 (-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet (-E)

● TB830D-□-□-A1-□-NN-NN-NN
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box B for power supply Liquid Analyzer for Turbidity and Chlorine
L (101) L (201) L
Power L (101) Power (201) L
Supply N (102) N (202)
N (102) Converter (202) N N Power (61) +
G (001) G (003) mA1 (62) – Analog output (-N2, -N4)
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32)NC (24)
(-N2, -N4)
(31)C S1
(33)NO (87) + Analog input (-N4)
(42)NC (88) –
(41)C S2
(43)NO SHLD (89)
(52)NC Termination (95)
(51)C S3 A+ (91)
(53)NO RS485(-R)
B− (92)
Contact output (72)NO GND (93)
(-WR) (71)C S4
(73)NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
37
● TB830D-□-□-A1-□-NN-NN-NN /ARS
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box A for power supply Liquid Analyzer for Turbidity and Chlorine
L (101) L (201) L
Power L (101) Power (201) L
N (102) CP N (202)
Supply N (102) Converter (202) N N Power (61) +
G (001) G (003) mA1 (62) – Analog output (-N2, -N4)
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
AR
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32)NC (24)
(-N2, -N4)
(31)C S1
(33)NO (87) + Analog input (-N4)
(42)NC (88) –
(41)C S2
(43)NO SHLD (89)
(52)NC Termination (95)
(51)C S3 A+ (91)
(53)NO RS485(-R)
B− (92)
Contact output (72)NO GND (93)
(-WR) (71)C S4
(73)NC SHLD (94)

Ethernet(-E)

● TB830D-□-□-A2-□-□-NN-NN
TB830D-□-□-A3-□-□-NN-NN
RELAY BOX
O
(11)
Measure (12) S
L (101) SV1 (-A3)
Power (13) C
Supply N (102) Power
G (001) (21)
Wash SV2 (-A2, -A3)
(22)
L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (002)
(41)
Drain SV4 (-A2, -A3)
(42)
L (301) Converter
N (302)
CN1
G (003)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine
L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)
SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC
(24)
(-N2, -N4)
(31) C S1
(33) NO (87) +
(42) NC (88) – Analog input (-N4)
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
38
● TB830D-□-□-A2-□-□-NN-NN /ARS
TB830D-□-□-A3-□-□-NN-NN /ARS
RELAY BOX
Terminal box B for power supply (11)
O
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) CP N (202) (13) C
N (102) Power
Supply G (001) G (003)
G (001) (21)
Wash SV2 (-A2, -A3)
(22)
AR L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (002)
(41)
Drain SV4 (-A2, -A3)
(42)
L (301) Converter
N (302)
CN1
G (003)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine
L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)
SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC
(24)
(-N2, -N4)
(31) C S1
(33) NO (87) +
(42) NC (88) – Analog input (-N4)
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
39
● TB830D-□-□-A5-NN-NN-NN-NN
RELAY BOX
O
(11)
Measure (12) S
Power L (101) SV1 *1
C
Supply N (102) Power (13)
G (001) (21)
Wash SV2 *1
(22)
L (201) Detector
(31)
N (202) Zero SV3 *1
(32)
G (002)
(41)
Drain SV4 *1
(42)
L (301) Converter
N (302)
CN1
G (003)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine
L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) + Analog output (-N2, -N4)
mA2 (66) –
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) + Analog output (-N4)
mA4 (86) –

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32) NC (24)
(-N2, -N4)
(31) C S1
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485 (-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet (-E)

*1: In case of selecting -A5, purchase the solenoid valve corresponding input power supply and frequency and cables.
See ● Solenoid valve (SV1 to SV4).

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
40
● TB830D-□-□-A5-NN-NN-NN-NN /ARS
RELAY BOX
Terminal box A for power supply O
(11)
L (101) L (201) Measure (12) S
L (101) SV1 *1
Power CP C
N (102) N (202) (13)
Supply N (102) Power
G (001) G (003)
G (001) (21)
Wash SV2 *1
(22)
AR
L (201) Detector
(31)
N (202) Zero SV3 *1
(32)
G (002)
(41)
Drain SV4 *1
(42)
L (301) Converter
N (302)
CN1
G (003)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine
L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) + Analog output (-N2, -N4)
mA2 (66) –
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) + Analog output (-N4)
mA4 (86) –

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32) NC (24)
(-N2, -N4)
(31) C S1
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485 (-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet (-E)

*1: In case of selecting -A5, purchase the solenoid valve corresponding input power supply and frequency and cables.
See ● Solenoid valve (SV1 to SV4).

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
41
● TB830D-□-□-A1-□-NN-NN-NN/PH□7
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box B for power supply Liquid Analyzer for Turbidity and Chlorine
L (101) L (201)
Power L (101) Power (201) L L
N (102) N (202)
Supply N (102) (202) N N Power (61) +
G (001) G (002) Converter mA1 (62) – Analog output (-N2, -N4)
G (001) for chlorine (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
(81) +
SENCOM Sensor1 mA3 (82) – Analog output (-N4)
(85) +
Analog output (-N4)
G (004)

G (003)
N (302)
mA4 (86) –
L (301)
N (12)
L (11)

SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC (-N2, -N4)
(24)
(31) C S1
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2 SHLD (89)
(43) NO
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

FLXA402
pH Converter

L
N Power (61) +
mA1 (62) – Analog output (-A2, -A4)
G
(65) +
mA2 (66) – Analog output (-A2, -A4)
(81) + Analog output (-A4)
Sensor1 mA3 (82) –
(85) +
mA4 (86) – Analog output (-A4)
pH Sensor
SHLD (63)
Sensor2
(21)
Contact input
(22)
Flow-Through type holder (32) NC
(24)
(-A2, -A4)
(31) C S1
(33) NO (87) +
(88) – Analog input (-A4)
(42) NC
Contact output (41) C S2
(43) NO
SHLD (89)
(-WR)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485 (-R)
B− (92)
(72) NO GND (93)
(71) C S4
(73) NC SHLD (94)
PG400 Pulse Generator for Clean Unit
Ethernet (-E)

U1
L1 US
U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
42
● TB830D-□-□-A1-□-NN-NN-NN/PH□7/ARS
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box A for power supply Liquid Analyzer for Turbidity and Chlorine
L (101) L (201)
Power L (101) Power (201) L L
N (102) CP N (202)
Supply N (102) (202) N N Power (61) +
G (001) G (003) Converter mA1 (62) – Analog output (-N2, -N4)
G (001) for chlorine (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
AR
(81) +
SENCOM Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC (-N2, -N4)
(24)
(31) C S1
(87) +
G(004)
N(302)
L(301)
L(401)

G(005)
N(402)

(33) NO Analog input (-N4)


(42) NC (88) –
(41) C S2 SHLD (89)
(43) NO
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

FLXA402
pH Converter

L
N Power (61) +
mA1 (62) – Analog output (-A2, -A4)
G
(65) +
mA2 (66) – Analog output (-A2, -A4)
(81) + Analog output (-A4)
Sensor1 mA3 (82) –
(85) +
mA4 (86) – Analog output (-A4)
pH Sensor
SHLD (63)
Sensor2
(21)
Contact input
(22)
Flow-Through type holder (32) NC
(24)
(-A2, -A4)
(31) C S1
(33) NO (87) +
(88) – Analog input (-A4)
(42) NC
Contact output (41) C S2
(43) NO
SHLD (89)
(-WR)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485 (-R)
B− (92)
(72) NO GND (93)
(71) C S4
(73) NC SHLD (94)
/PHU7
PG400 Pulse Generator for Clean Unit
Ethernet (-E)

U1
L1 US
U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
43
● TB830D-□-□-A2-□-□-NN-NN/PH□7
TB830D-□-□-A3-□-□-NN-NN/PH□7
RELAY BOX
Terminal box B for power supply (11)
O
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) N (202) (13) C
N (102) Power
Supply G (001) G (002)
G (001) (21)
Wash SV2 (-A2, -A3)
(22)
L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (002)
(41)
G (004)

G (003)
N (302)
L (301)
N (12)

Drain
L (11)

SV4 (-A2, -A3)


(42)
L (301) Converter
N (302)
CN1
G (003)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine

L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) + Analog output (-N4)
Sensor1 mA3 (82) –
(85) + Analog output (-N4)
mA4 (86) –

SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC
(24)
(-N2, -N4)
(31) C S1
(33) NO (87) + Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
Contact output (53) NO
B− (92) RS485(-R)
(-WR) (72) NO GND (93)
(71) C S4
(73) NC SHLD (94)

Ethernet(-E)

FLXA402
pH Converter

L
N Power (61) + Analog output (-A2, -A4)
mA1 (62) –
G
(65) +
mA2 (66) – Analog output (-A2, -A4)
(81) + Analog output (-A4)
Sensor1 mA3 (82) –
(85) + Analog output (-A4)
pH Sensor mA4 (86) –

SHLD (63)
Sensor2
(21)
Contact input
(22)
Flow-Through type holder (32) NC
(24)
(-A2, -A4)
(31) C S1
(33) NO (87) +
Analog input (-A4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
Contact output
(-WR) (52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
(72) NO GND (93)
(71) C S4
/PHU7 (73) NC SHLD (94)
PG400 Pulse Generator for Clean Unit
Ethernet(-E)

U1
L1
US U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
44
● TB830D-□-□-A2-□-□-NN-NN/PH□7/ARS
TB830D-□-□-A3-□-□-NN-NN/PH□7/ARS
Terminal box A RELAY BOX
for power supply (11)
O
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) CP N (202) (13) C
N (102) Power
Supply G (001) G (003)
G (001) (21)
Wash SV2 (-A2, -A3)
AR (22)
L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (002)
(41)
Drain SV4 (-A2, -A3)
(42)
L (301) Converter
N (302)
CN1
G (003)
G (005)

G(004)
N(302)
L(301)
N (402)
L (401)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine

L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) + Analog output (-N4)
Sensor1 mA3 (82) –
(85) + Analog output (-N4)
mA4 (86) –

SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC
(24)
(-N2, -N4)
(31) C S1
(33) NO (87) + Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
Contact output (53) NO
B− (92) RS485(-R)
(-WR) (72) NO GND (93)
(71) C S4
(73) NC SHLD (94)

Ethernet(-E)

FLXA402
pH Converter

L
N Power (61) + Analog output (-A2, -A4)
mA1 (62) –
G
(65) +
mA2 (66) – Analog output (-A2, -A4)
(81) + Analog output (-A4)
Sensor1 mA3 (82) –
(85) + Analog output (-A4)
pH Sensor mA4 (86) –

SHLD (63)
Sensor2
(21)
Contact input
(22)
Flow-Through type holder (32) NC
(24)
(-A2, -A4)
(31) C S1
(33) NO (87) +
Analog input (-A4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
Contact output
(-WR) (52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
(72) NO GND (93)
(71) C S4
(73) NC SHLD (94)
PG400 Pulse Generator for Clean Unit
Ethernet(-E)

U1
L1
US U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
45
● TB830D-□-□-A1-□-NN-NN-NN/PH□2
TB830D
Surface Scattered Light Turbidity Detector
Terminal box B FLXA402T
for power supply Liquid Analyzer for Turbidity and Chlorine
L (101) L (201)
L (101) Power (201) L L
Power N (102) N (202)
Supply N (102) (202) N N Power (61) +
G (001) G (002) Converter mA1 (62) – Analog output (-N2, -N4)
G (001) for chlorine (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
Sensor1 (81) +
SENCOM mA3 (82) – Analog output (-N4)
(85) +
G (004)

mA4 (86) – Analog output (-N4)


N (12)
L (11)

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32) NC (-N2, -N4)
(31) C S1
(24)
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
pH Sensor (53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(71) C S4
(-WR) (73) NC SHLD (94)

Ethernet(-E)

Flow-Through type holder

PG400 Pulse Generator for Clean Unit

U1
L1 US
U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
46
● TB830D-□-□-A1-□-NN-NN-NN/PH□2/ARS
TB830D
Surface Scattered Light Turbidity Detector
Terminal box A FLXA402T
for power supply Liquid Analyzer for Turbidity and Chlorine
L (101) L (201)
L (101) Power (201) L L
Power N (102) CP N (202)
Supply N (102) (202) N N Power (61) +
G (001) G (003) Converter mA1 (62) – Analog output (-N2, -N4)
G (001) for chlorine (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
AR
Sensor1 (81) +
SENCOM mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


(21)
Contact input
(22)
G (005)
N (402)

(-N2, -N4)
L (401)

(32) NC
(31) C S1
(24)
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
pH Sensor (53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(71) C S4
(-WR) (73) NC SHLD (94)

Ethernet(-E)

Flow-Through type holder

PG400 Pulse Generator for Clean Unit

U1
L1 US
U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
47
● TB830D-□-□-A2-□-□-NN-NN/PH□2
TB830D-□-□-A3-□-□-NN-NN/PH□2
Terminal box B RELAY BOX
for power supply (11)
O
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) N (202) (13) C
Supply N (102) Power
G (001) G (002)
G (001) (21)
Wash SV2 (-A2, -A3)
BK (22)
L (201) Detector
WH (31)
N (202) Zero SV3 (-A3)
(32)
G (004)

RD
N (12)

G (002)
L (11)

(41)
Drain SV4 (-A2, -A3)
BK (42)
L (301) Converter
WH
N (302)
RD CN1
G (003)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine

L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) + Analog output (-N4)
Sensor1 mA3 (82) –
(85) + Analog output (-N4)
mA4 (86) –

Sensor2 SHLD (63)


(21)
pH Sensor Contact input
(22)
(32) NC (-N2, -N4)
(24)
(31) C S1
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2 SHLD (89)
Flow-Through type holder (43) NO
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

/PHU2
PG400 Pulse Generator for Clean Unit

U1
L1 US U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
48
● TB830D-□-□-A2-□-□-NN-NN/PH□2 /ARS
TB830D-□-□-A3-□-□-NN-NN/PH□2 /ARS
RELAY BOX
Terminal box A for power supply O
(11)
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) CP N (202) (13) C
Supply N (102) Power
G (001) G (003)
G (001) (21)
Wash SV2 (-A2, -A3)
(22)
AR
L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (002)
(41)
Drain SV4 (-A2, -A3)
(42)
L (301) Converter
N (302)
CN1
G (003)
G (005)
N (402)
L (401)

TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
RELAY Liquid Analyzer for Turbidity and Chlorine

L (101) Power L
N (102) N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) + Analog output (-N4)
Sensor1 mA3 (82) –
(85) + Analog output (-N4)
mA4 (86) –

Sensor2 SHLD (63)


(21)
pH Sensor Contact input
(22)
(32) NC (-N2, -N4)
(24)
(31) C S1
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2 SHLD (89)
Flow-Through type holder (43) NO
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

/PHU2
PG400 Pulse Generator for Clean Unit

U1
L1 US U2
L2 Power Oscillator
S
G

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
49
● TB830D-□-□-A1-□-NN-NN-NN/FC
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box B
Liquid Analyzer for Turbidity and Chlorine
for power supply
L (101) L(201) L
L (101) Power (201) L
Power N (102) N(202) N
N (102) Converter (202) N Power (61) +
Supply G (001) G(002) mA1 (62) – Analog output (-N2, -N4)
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
G (003)
N (302)
L (301)

(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)

(21) Contact input


(22) (-N2, -N4)
(32) NC (24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO
B− (92) RS485(-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

FC800D
Non-reagent Type Free Available
Chlorine Sensor Unit
FLXA402T
Disuse Liquid Analyzer for Turbidity and Chlorine

L (101) Power (201) L L


N (102) Converter (202) N N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM Sensor1 (81) + Analog output (-N4)
mA3 (82) –
(85) + Analog output (-N4)
mA4 (86) –

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32) NC (-N2, -N4)
(31) C S1
(24)
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO B− (92) RS485(-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
50
● TB830D-□-□-A1-□-NN-NN-NN/FC /ARS
TB830D
Surface Scattered Light Turbidity Detector
FLXA402T
Terminal box A for power supply Liquid Analyzer for Turbidity and Chlorine

L (101) L(201) L
L (101) Power (201) L
Power N (102) CP N(202)
N (102) Converter (202) N N Power (61) +
Supply G (001) G(003) mA1 (62) – Analog output (-N2, -N4)
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
AR
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

Sensor2 SHLD (63)


G (004)
N (302)
L (301)

(21) Contact input


(22) (-N2, -N4)
(32) NC (24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO
B− (92) RS485(-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

FC800D
Non-reagent Type Free Available
Chlorine Sensor Unit
FLXA402T
Disuse Liquid Analyzer for Turbidity and Chlorine

L (101) Power (201) L L


N (102) Converter (202) N N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) (002) G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM Sensor1 (81) + Analog output (-N4)
mA3 (82) –
(85) + Analog output (-N4)
mA4 (86) –

Sensor2 SHLD (63)


(21)
Contact input
(22)
(32) NC (-N2, -N4)
(31) C S1
(24)
(33) NO (87) +
Analog input (-N4)
(42) NC (88) –
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO B− (92) RS485(-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
51
● TB830D-□-□-A2-□-□-NN-NN/FC
TB830D-□-□-A3-□-□-NN-NN/FC
Terminal box B RELAY BOX
for power supply (11)
O
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) N (202) (13) C
N (102) Power
Supply G (001) G (002)
G (001) (21)
Wash SV2 (-A2, -A3)
(22)
L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (003)
N (302)
L (301)

G (002)
(41)
Drain SV4 (-A2, -A3)
(42)
L (301) Converter
N (302)
CN1
G (003)

TB830D FLXA402T
Surface Scattered Light Liquid Analyzer for Turbidity and Chlorine
Turbidity Detector
L
RELAY
N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
L (101) G
Power (65) +
N (102) mA2 (66) – Analog output (-N2, -N4)
G (001) SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)
SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC
(24) (-N2, -N4)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)
FC800D Ethernet(-E)
Non-reagent Type Free Available
Chlorine Sensor Unit
FLXA402T
Disuse Liquid Analyzer for Turbidity and Chlorine
L (101) Power (201)L L
N (102) Converter (202)N N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) (002)G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC (-N2, -N4)
(24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2 SHLD (89)
(43) NO
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO B− (92) RS485(-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
52
● TB830D-□-□-A2-□-□-NN-NN/FC/ARS
TB830D-□-□-A3-□-□-NN-NN/FC /ARS
Terminal box A RELAY BOX
for power supply (11)
O
L (101) L (201) Measure (12) S
L (101) SV1 (-A3)
Power N (102) CP N (202) (13) C
N (102) Power
Supply G (001) G (003)
G (001) (21)
Wash SV2 (-A2, -A3)
AR (22)
L (201) Detector
(31)
N (202) Zero SV3 (-A3)
(32)
G (002)
(41)
Drain SV4 (-A2, -A3)
(42)
L(301)

G(004)
N(302)

L (301) Converter
N (302)
CN1
G (003)

TB830D FLXA402T
Surface Scattered Light Liquid Analyzer for Turbidity and Chlorine
Turbidity Detector
L
RELAY
N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
L (101) G
Power (65) +
N (102) mA2 (66) – Analog output (-N2, -N4)
G (001) SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)
SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC
(24) (-N2, -N4)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2
(43) NO SHLD (89)
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO RS485(-R)
B− (92)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)
FC800D Ethernet(-E)
Non-reagent Type Free Available
Chlorine Sensor Unit
FLXA402T
Disuse Liquid Analyzer for Turbidity and Chlorine
L (101) Power (201)L L
N (102) Converter (202)N N Power (61) +
mA1 (62) – Analog output (-N2, -N4)
G (001) (002)G G
(65) +
mA2 (66) – Analog output (-N2, -N4)
SENCOM (81) +
Sensor1 mA3 (82) – Analog output (-N4)
(85) +
mA4 (86) – Analog output (-N4)

SHLD (63)
Sensor2
(21)
Contact input
(22)
(32) NC (-N2, -N4)
(24)
(31) C S1
(33) NO (87) +
(88) – Analog input (-N4)
(42) NC
(41) C S2 SHLD (89)
(43) NO
(52) NC Termination (95)
(51) C S3 A+ (91)
(53) NO B− (92) RS485(-R)
Contact output (72) NO GND (93)
(-WR) (71) C S4
(73) NC SHLD (94)

Ethernet(-E)

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
53
n System Configuration (Multiple Connections)
FLXA402T enables multiple sensor/detector connection for free chlorine , turbidity, pH, conductivity.
Select 2nd Input for pH or conductivity parameter.
Select -TB (TB830D-□□-A□-NN-□N-NN-NN-NN) for 1st input.
2nd Input -TB -P1 -C1
Connectable TB820D TB830D FU20 SC4A
sensor/detector FU24 SC42
(to be purchased SM21/SR20 PH20 SX42
separately) SC25V SC4AJ
SC21 SC8SG
SC29C PH8E□P SC210G
PH4□
PH4□□
Conditions TB820D-□□-A□-NN- TB830D-□□-A□-NN- Except for Variopin connector Except for Variopin connector
□□-CT-NN □N-NN-NN-NN (-VS) of FU20,FU24 (-VS) of SC4A
Donʼt select relay box Select always “-NN” Except for Variopin connector Except for Variopin connector
for solenoid valve (Without sampling (-B□P25) of SC25V (-□V,-□G) of SC42
(-A5) system) for the Except for Variopin connector Except for Variopin connector
Sampling system. (-V) of PH8E□P (-□V) of SX42
Except for Variopin connector
(-VS) of SC4AJ, SC8SG.
To connect FLXA402T to both a chlorine sensor and a turbidity detector, select CL (FC800D-□□-A□-NN-NN) for the 1st Input.
Invalid combination:
“-TB” (SENCOM for Turbidity) on 1st Input and “-CL” (SENCOM for Chlorine) on 2nd Input.
To connect FLXA402T to both a chlorine sensor and a turbidity detector, read the general specification GS 12F05B10-
01EN for FC800D-FLXA402T.
FLXA402T 2nd input “-TB”
Analog Output
Contact Input FLXA402T
Liquid
Contact Output
Analyzer
RS485 or Ethernet TB830D *

Power Supply
TB830D

Power Supply Sample Drain


water

Drain
Sample water

Drain

*: TB820 is also connectable.

Drain
FLXA402T 2nd input “-P1” or ”-C1”
Analog Output
FLXA402T
Contact Input
Liquid
Contact Output Analyzer
RS485 or Ethernet

TB830D
pH sensor
or
Power Supply Conductivity sensor

Sample water

Drain

Figure 2

All Rights Reserved. Copyright © 2023, Yokogawa Electric Corporation GS 12E04B40-01EN May 11, 2023-00
Subject to change without notice.