You are on page 1of 2

Model N100

UV Fluorescence SO₂ Analyzer

► Single or dual range capability


► Hydrocarbon “kicker” scrubber
► Internal DC-powered vacuum pump
► Customizable alerts and continuous
self-checking
► Wide operating temperature range
► Internal zero and span valves
(optional)
► Optional 47mm membrane or long-
life sample particulate filter

N Series Platform Features The Model N100 instrument uses the proven UV-fluorescence principle,
combined with a state-of-the-art modular architecture, and intuitive operating
Color Touch-Screen software to provide accurate and dependable measurements of low-level Sulfur
Graphics Display Dioxide (SO₂) gas.

Two Front Panel USB Ports The instrument’s UV light source is continuously flashed to provide exceptional
stability and long lifetime. Any potential lamp intensity fluctuations or detector
Modular Internal Hardware drift is automatically managed in real-time using a sophisticated combination
Design of reference readings during the lamp light/dark cycles. Hydrocarbon gases
that fluoresce at the same wavelength as SO₂ gas are removed from the sample
All DC-powered Internal stream through a scrubber (i.e. kicker) and prior to detection to eliminate any
potential interferences. The long-life sample filter option further improves
Components
efficiency with a ~6 month exchange interval in ambient air quality monitoring
Large Internal Data Storage applications.

Instrument functions and controls are managed through a series of integrated


Serial and TCP/IP Ethernet
microprocessor-controlled modules utilizing a simple and reliable CAN Bus
Included communications architecture. Each module is independently assembled and
calibrated allowing easy and fast field replacement to maximize instrument
Digital and Analog
uptime.
Expansion Options
Intuitive operation and calibration of all N Series products is achieved through
Indicator Illuminated Soft the NumaView™ Software interface. The graphical user interface (GUI) is
Power Switch customizable, giving the user fast and efficient access to instrument status,
as well as measurement data and diagnostic parameters in either numeric or
Split Fold-Down Rear Panel graphical form. NumaView™ Remote Software (included at no charge) provides
the same virtual interface and complete instrument control, as well as access to
the instrument’s large internal data storage buffer from a remote PC or tablet.

WWW.TELEDYNE-API.COM
N100 Specifications

Measurement Units ppb, ppm, μg/m³, mg/m³ (selectable)


Response Time < 120 seconds to 95%
Ranges Min: 0 - 50 ppb full scale
Max: 0 - 20,000 ppb full scale (selectable, dual-range supported)
Sample Flow Rate 650 cc/min ±10%
Zero Noise < 0.2 ppb (RMS)
Span Noise < 0.5% of reading (RMS) above 50 ppb
Lower Detectable Limit < 0.4 ppb
Precision 0.5% of reading above 50 ppb
Linearity 1% of full scale
Zero Drift < 0.5 ppb/24 hours
Span Drift < 0.5% of full scale/24 hours
Included I/O 1 x Ethernet (TCP/IP)
1 x RS232
2 x front panel USB device ports
Optional I/O Universal Analog Output Board includes (all user-definable):
4 x Isolated Voltage Outputs (5V, 10V; user-selectable)
3 x Individually Isolated Current Outputs (4-20mA)
Digital I/O Expansion Board includes:
3 x Isolated Digital Input Controls
5 x Isolated Digital Output Controls (user-definable)
3 x Form C Relay Alarm Outputs (user-definable)
Weight 31 lbs (14.1 kg)
Dimensions (HxWxD) 7” x 17” x 24.3” (178 x 432 x 617 mm)
Operating Temperature 0 - 45°C (with US EPA Approval)
Power 100V-240V, 50/60 Hz, Typical consumption 55W
Certifications US EPA: EQSA-0495-100

Specifications subject to change without notice.


All specifications are based on constant conditions.

All N Series instruments include a 2-year manufacturer’s warranty as well as


email and phone support for the lifetime of the instrument.

For more information about Teledyne API


instruments, visit our website at:

www.teledyne-api.com
9970 Carroll Canyon Road, San Diego, CA 92131
© 2021 Teledyne API
Phone 858-657-9800 
Fax 858-657-9816
Printed documents are uncontrolled. SAL000114B
Email api-sales@teledyne.com (DCN 8409) 12.17.21