1 Cause analysis: The module is the latest upgraded hardware, and the

blockage is detected through the pressure sensor inside.


The previous version used a gauge pressure sensor, and now an absolute
pressure sensor is used.
For the application of absolute pressure sensors: it is necessary to detect the
zero point when powering on, as the baseline for pressure monitoring.The
previous version of the gauge pressure sensor is not required

Place the module in an environment with stable pressure. Then power on the
module again. The module can work.

The customer can reproduce this phenomenon: simulate a low pressure


environment in a closed container, and when the container pressure is stable,
restart the module and the module works normally.
If the pressure in the container is unstable and has been decreasing, the
module may still be not working properly.

Upgrade method,
Upgrade the software ANES_1.bin to the three modules whose serial
numbers are AG10023064690 AG10023064691 AG10023064692
Upgrade the software ANES_2.bin to the modules whose serial
numbers are AG10023074707