Professional Documents
Culture Documents
21 Evaluation For Different Models of Analyzer
21 Evaluation For Different Models of Analyzer
Author(s) Approver
1. Purpose
The purpose of this study is to various performance characteristics of the 6 models of Automated
Mass Spectrometry Microbial Identification System.
The 6 models of Automated Mass Spectrometry Microbial Identification System used in this
study were listed in table 1.
Table 2 Evaluated instruments
3. Protocol
1 / 12
20211201:01
3.2.1 Sensitivity
Study Description: Inoculate BSA to the target plate and dry at room temperature, overlay the
sample spot with 1µL HCCA, dry at room temperature. Then identify the sample with MALDI-
TOF MS. The information of BSA is listed in the table.
Data analysis: Read the data in the displayed graph peak directly. Acceptance Criteria: The
signal-to-noise ratio (S/N) should be ≥ 10.
Study Description: Inoculate 1µL angiotensinogen, 1µL cytochrome C and 1µL mix protein to
the target plate separately and dry at room temperature, overlay each sample spot with 1ul
HCCA, dry at room temperature. Then identify the sample with MALDI-TOF MS. The
information of angiotensinogen, cytochrome C and mix protein is listed in the table.
Data analysis: Select the method of Peptide-3k and method of Microbe as the calibration method
and test method. The laser is used for 40 times, and each sample is measured for more than 3
effective tests. The resolution of each test and the corresponding instrument parameters were
recorded. Read the data in the displayed graph peak directly.
Acceptance Criteria: The mass resolution of single protein and mixed protein should meet the
following criteria.
Study Description: Calibrate the device with the calibrator, acquisit the map of Insulin (m/z=
5808.57(M+H)+) and Cytochrome C (m/z= 12362.96(M+H)+) separately, choose the mass-to-
charge ratio of corresponding mass spectrum peak, then carry out 3 effective tests and calculate
the average value. The information of calibration protein is listed in the table.
2 / 12
20211201:01
Data analysis: Test the Insulin and Cytochrome C after calibration of the device, take 3 maps of
each sample spot. The laser is used for 40 times. Calculate the mass accuracy of each peak in the
Study Description: Calibrate the device with bacterial protein calibrator, acquisit the map of
Insulin and Cytochrome C separately, choose the mass-to-charge ratio of corresponding mass
spectrum peak, then carry out 6 effective tests. The information of calibration protein is listed
in the table.
Data analysis: Test the Insulin and Cytochrome C after calibration of the device, take 6 maps of
each sample spot. The laser was used for 40 times. Calculate the mass repeatability of each peak
√
n
1
in the map with the formula: CV = ∑ ( m −m)2 ×10%
n−1 i=1 i
m
CV —— Mass variation coefficient of the corresponding mass spectrum peak of the map;
mi —— Measured value of the corresponding mass spectrum peak i of the map;
m——average value of 6 measurements of the corresponding mass spectrum peak of the map.
Acceptance Criteria: The CV of peak mass in the chromatograms of insulin and cytochrome C
should be less than 0.02%.
3 / 12
20211201:01
Study Description: Calibrate the device with bacterial protein calibrator, acquisit the map of
Insulin and Cytochrome C separately, choose the mass-to-charge ratio of corresponding mass
spectrum peak, then carry out 3 effective tests and calculate the average value. Repeat the above
testing process after 24 hours and compare the mass deviation with initial test results. The
information of calibration protein is listed in the table.
Data analysis: Calculate the mass stability of each peak in the map with the formula:
|m2−m1|
D= × 106
m1
m2 ——Average value of the 3 measurements of the initial mass spectrum peaks after 24 hours.
Acceptance Criteria: The relative mass deviation of the test results of peak masses obtained in
the chromatograms of insulin and cytochrome C 24 h after the instrument is turned on and stable
from the initial test result when the instrument is stable should be ≤ 300 ppm.
4 / 12
20211201:01
1) Parameter requirements: set grounding resistance tester test parameters: current 10A (AC),
test time 20S, grounding resistance upper limit 0.1 ω;
4. Results
4.1.1. Sensitivity
5 / 12
20211201:01
Mixed protein
21 879 937 948 948 >700
12362.96
6 / 12
20211201:01
Conclusion: According to the above results, the mass resolution of the 6 models met the criteria.
7 / 12
20211201:01
8 / 12
20211201:01
Autof ms Insulin 5808.46 5808.677 5808.543 5808.671 5808.551 5808.595 8
2600 Cytochrom 12362.452 12363.14 12362.945 12362.886 12362.576 12362.917 4
eC
Conclusion: According to the above results, the mass stability of the 6 models met the criteria.
9 / 12
20211201:01
coli(ATCC25922)
Enterococcus
9.694 9.628 9.721 9.759 9.737 9.736 9.628
faecium(ATCC19434)
Staphylococcus
9.759 9.792 9.782 9.768 9.794 9.789 9.759
aureus(ATCC29213)
Bacteroides
9.702 9.724 9.736 9.723 9.759 9.775 9.702
fragilis(ATCC 25285)
Pseudomonas
aeruginosa(ATCC27853 9.656 9.644 9.557 9.623 9.648 9.669 9.623
)
Candida
parapsilosis(ATCC 9.644 9.663 9.653 9.667 9.666 9.678 9.644
22019)
10 / 12
20211201:01
1 2 3 4 e5
Enterococcus
9.648 9.712 9.683 9.689 9.673 9.672 9.648
faecalis(ATCC51299)
Escherichia coli(ATCC25922) 9.659 9.691 9.705 9.696 9.678 9.684 9.659
Enterococcus
9.739 9.733 9.734 9.758 9.781 9.757 9.733
faecium(ATCC19434)
Staphylococcus
9.715 9.720 9.711 9.727 9.724 9.695 9.695
aureus(ATCC29213)
Bacteroides fragilis(ATCC
9.619 9.710 9.727 9.695 9.673 9.707 9.619
25285)
Pseudomonas
9.612 9.583 9.670 9.672 9.659 9.643 9.583
aeruginosa(ATCC27853)
Candida parapsilosis(ATCC
9.646 9.666 9.653 9.420 9.656 9.662 9.420
22019)
Conclusion: According to the above results, the identification accuracy of the 6 models met the
criteria.
Contact current
Current value
value under Result Result
Model type under single
normal (mA) (mA)
failure condition
conditions
Autof ms600 <0.5mA 0.03 <3.5mA 2.237
Autof ms800 0.03 2.282
11 / 12
20211201:01
Conclusion: Conclusion: According to the above results, the contract current result of the 6
models met the criteria.
5. Conclusion
According to the results above, the performance parameter and electric safety of 6 models of
Automated Mass Spectrometry Microbial Identification System can meet the criteria, there
is no difference between the 6 models during the usage.
12 / 12