Professional Documents
Culture Documents
Troubleshooting The Alliance HPLC 1692029818
Troubleshooting The Alliance HPLC 1692029818
Table of contents
Troubleshooting errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . 2
All errors/warnings index . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . 2
System level errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . 7
Solvent manager errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 16
Sample manager errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 26
Degasser errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 49
Column device errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 53
Detector control errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 58
Dissolution errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 60
Alliance HPLC 2795 and 2796 errors . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 64
Reference information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 73
Purpose
This document outlines recommended steps for troubleshooting system‐level symptoms in the Alliance
HPLC System.
Scope
This document is valid for the following Alliance HPLC systems:
• Alliance 2690, 2695, and e2695 Systems
• Alliance 2790, 2795, and 2796 Systems
This document contains error troubleshooting information for each of the following modules:
How to use
1. Find the relevant problem in each section and click the corresponding link to troubleshoot.
2. Perform troubleshooting steps in the numbered order listed in the troubleshooting tables.
Note: Troubleshooting recommendations are formatted in easy‐to‐scan tables that list possible
causes and corrective actions, ranked by number from most likely cause/least invasive action
to least likely cause/most invasive action (typically, component replacement).
TABLE OF CONTENTS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 2 OF 85
Troubleshooting errors
This section outlines troubleshooting for the Alliance HPLC error states for the system and each system
module.
Additional troubleshooting information
For additional troubleshooting information for your specific Alliance HPLC System, see:
• Alliance 2690/2695/e2695 System Service Key Documents
• Alliance 2790/2795/2796 System Service Key Documents
Major repairs and requalification
Major repairs require requalification. If troubleshooting results in a major repair or you are uncertain if a
repair qualifies as major, consult the requisite qualification documentation for your specific system.
Table 1 is an alphabetical list of all Alliance HPLC errors and warnings. Select the first letter of the error to
jump to the relevant section (see below), or select the specific error/warning (within Table 1):
A B C D E F G H I J K L M
N O P Q R S T U V W X Y Z
Table 1: All errors and warnings
Initial letter Error or warning
# • 410 detector fault, page 58
• 410/2410 detector fault, page 64
• 486 detector fault, page 59
A • A failure occurred in attempting to download an instrument method [Empower error], page 7
• Accumulator over pressure, page 16
• Auto Shutdown Invoked, page 7
B • Bath speed tolerance, page 60
• Bath temp tolerance, page 60
• Bubble found on compression, page 16
• Bubble found on intake, page 17
TROUBLESHOOTING ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 3 OF 85
Table 1: All errors and warnings
Initial letter Error or warning
C • Can't understand requested well, page 64
• Cannot reach minimum pressure, page 18
• Carousel carrier jammed, page 26
• Carousel ID sensor fault, page 26
• Carousel is busy, page 27
• Carousel malfunction, page 28
• Carousel Z axis malfunction (1), page 30
• Carousel Z axis malfunction (2), page 31
• Carousel Z axis malfunction (3), page 31
• Carousel Z axis malfunction (4), page 32
• Carousel Z axis malfunction (5), page 32
• Carrier is busy, page 65
• Carrier lockpin movement failed, page 65
• Carrier malfunction, page 65
• Column heater door open, page 53
• Column heater not ambient, page 53
• Column heater running, page 53
• Column temp h/w fault (1), page 54
• Column temp h/w fault (2), page 54
• Column temp h/w fault (3), page 54
• Column temp h/w fault (4), page 54
• Column temp h/w fault (5), page 54
• Column temp h/w fault (6), page 55
• Column temp h/w fault (7), page 55
• Column temp h/w fault (8), page 55
• Column temp over limit (X°), page 55
• Compression check failed, page 34
• Compression leak test failed, page 34
• Controller Internal Error, page 7
D • Default separation method created, page 66
• Degasser ambient pressure, page 49
• Degasser high leak rate, page 49
• Degasser h/w fault (1), page 50
• Degasser h/w fault (2), page 50
• Degasser h/w fault (3), page 50
• Degasser minimum pressure, page 51
• Degasser pressure fault (X), page 51
• Diskette removed while logging, page 7
• Dissolution, page 60
• Dissolution (bad state), page 60
• Dissolution (carousel), page 61
• Dissolution (comm error), page 61
ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 4 OF 85
Table 1: All errors and warnings
Initial letter Error or warning
D • Dissolution (couldn’t continue), page 61
• Dissolution (couldn’t dispense), page 62
(continued)
• Dissolution (dispenser move), page 62
• Dissolution (TM sensor), page 63
• Dissolution (TM syringe), page 63
• Dissolution (vial scan), page 63
• Door is Closed, page 8
• Door is Open, page 8
E • Eluent Monitor Disabled, page 66
• Eluent Monitor Error, page 66
• Error 109, page 8
• Error 110, page 8
• Error changing column position, page 56
• Error writing log to disk, page 8
• Exceeded maximum plunger stroke, page 19
F • Fan failed, page 9
• Flow too high for stroke volume, page 19
G • General electronics fault, page 9
• General instrument error, page 9
• General instrument fault, page 9
• General solv delivery fault, page 20
• GPV h/w fault, page 20
H • HWERROR1, page 9
I • Initialization fault, page 9
• Inject Valve movement failure, page 66
• Injection parameters incorrect, page 66
• Injector internal error, page 20
• Instrument Failure Error, page 10
• Internal error, page 10
L • Loop too small for sample, page 35
• Lost Prime, page 21
• Low column temp (X°), page 57
• Low sample temp (X), page 35
M • Missed a dissolution sample, page 63
• Missing carousel, page 36
• Missing vial, page 36
• Must degas solvents for SysPrep, page 66
ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 5 OF 85
Table 1: All errors and warnings
Initial letter Error or warning
N • Needle is busy, page 38
• Needle malfunction, page 38
• Needle malfunction (bottom sensor), page 39
• Needle malfunction (top sensor), page 41
• No‐Flow Shutdown Invoked, page 67
O • Options not recognized, page 10
• Overpressure: check inline filter, page 67
P • Parameter range error, page 11
• pH/Cond. Monitor Disabled, page 67
• pH/Cond. Monitor Error, page 67
• Plate height error, page 68
• Plate parameters are incorrect, page 67
• Plate X axis parameters incorrect, page 68
• Plate Y axis parameters incorrect, page 68
• Plunger homing fault, page 21
• Plunger homing over pressure, page 22
• Power down, page 11
• Power up, page 11
• Primary over pressure, page 22
• Primary transducer range (X), page 23
• Procedure interrupted, page 12
R • Restriction missing, page 42
• RS‐232 Valve Comm Error, page 68
• RS‐232 Valve Error, page 69
• RS‐232 Valve Missing, page 69
S • Sample heater/cooler running, page 42
• Sample syringe over pressure (X), page 42
• Sample temp h/w fault 1, page 43
• Sample temp h/w fault 2, page 44
• Sample temp h/w fault 3, page 44
• Sample temp h/w fault 4, page 44
• Sample temp h/w fault 5, page 45
• Sample temp h/w fault 6, page 45
• Sample temp h/w fault 7, page 45
• Sample temp not ambient, page 43
• Sample temp over limit (X), page 43
• Sample too large for configuration, page 69
• Sample transducer range (X), page 46
• Seal pack geometry fault (X), page 46
• Seal wash internal error, page 23
• Seals not adjusted, page 47
ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 6 OF 85
Table 1: All errors and warnings
Initial letter Error or warning
S • Server busy, page 12
• Solv delivery h/w fault, page 24
(continued)
• Solv delivery internal error, page 24
• Solv delivery motor lost sync (X), page 25
• Sparge internal error, page 52
• StopFlow, page 12
• Stop Flow input was received, page 13
• Stop Flow key was pressed, page 13
• Stop Flow request from remote, page 13
• Syringe movement malfunction, page 47
• SYSERROR1, page 13
• SysPrep requires a degasser, page 69
• System over pressure, page 14
• System transducer range (X), page 25
• System under pressure, page 14
T • Top and bottom needle sensors active, page 70
U • User‐programmed alert, page 15
V • Vial internal error, page 48
• Vial sensor test failed, page 48
W • Well depth error, page 70
• Wrong carousel in position, page 48
X • X axis malfunction, page 71
Y • Y axis malfunction, page 72
ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 7 OF 85
This section lists the system level errors and warnings in alphabetical order. Locate the relevant error or
warning in the left column and follow the appropriate troubleshooting steps.
Table 2: System level errors or warnings
Dust inside the floppy drive Clean the inside of the drive using
compressed air.
Disk‐drive problem Replace the drive.
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 8 OF 85
Table 2: System level errors or warnings (continued)
Insufficient disk capacity Ensure that the floppy disk has
enough free space to store the data
being written.
Note: It takes a very long time to
fill a floppy disk using the
"Report to floppy disk"
feature. Replace the disk if it
has been in use for a long
time.
Ensure that the disk contains no
more than 256 files.
Corrupted disk Replace the disk.
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 9 OF 85
Table 2: System level errors or warnings (continued)
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 10 OF 85
Table 2: System level errors or warnings (continued)
Occurs in e2695 Auto Addition with
time delay under Ethernet control.
Notes: —This error affects only
systems under Ethernet
control.
—Auto Addition with time
delay under IEEE control and
in stand‐alone mode are
immune to this issue.
A hardware problem has occurred 1. Examine the trace log to
determine instrument activity at
the time of failure.
2. Contact GSS and be prepared to
describe the conditions leading
to this error.
Optional control cable problem 1. Inspect/Reconnect/Replace.
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 11 OF 85
Table 2: System level errors or warnings (continued)
Optional component problem Replace the optional component if
necessary.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
Firmware encountered an 1. Examine the trace log to
unexpected condition determine instrument activity at
the time of failure.
2. Contact GSS and be prepared to
describe the conditions leading
to this error.
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 12 OF 85
Table 2: System level errors or warnings (continued)
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 13 OF 85
Table 2: System level errors or warnings (continued)
Wiring issue—the wrong wire may 1. Inspect the rear‐panel input
be connected to the Stop Flow rear‐ wiring.
panel input 2. Verify proper operation using the
Inputs and Outputs diagnostic.
Rear‐panel IO connector block 1. Inspect/replace.
problem 2. Verify proper operation using the
Inputs and Outputs diagnostic.
Failure of electrical connections 1. Inspect/Reconnect/Replace..
between the rear‐panel IO 2. Verify proper operation using the
connector block and the CPU PCB Inputs and Outputs diagnostic.
board
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814).
See related: StopFlow
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 14 OF 85
Table 2: System level errors or warnings (continued)
The separation method contains a Change the system pressure limit in
limit that is too low for the column, the separation method.
flow rate, and solvent or solvents.
Column problem 1. Remove column to reduce
pressure and replace if
necessary.
2. Verify and adjust flow rate if
necessary.
3. Clean or replace the in‐line filter.
4. Change the system pressure limit
in the separation method if
necessary.
Clogged in‐line filter Clean or replace the filter.
See:
• 269x: Replacing the In‐Line Filter
• 2790/2795: Replacing the In‐Line
Filter
• 2796: 2796 In‐Line Filter
Failed check valve Replace the primary inlet check
valve.
See:
• Check Valve Cartridge Summary
• 296x: Replacing the Check Valves
• 2790/2795:
Replacing the Check Valves
• 2796: Replacing the 2796 Check
Valves
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 15 OF 85
Table 2: System level errors or warnings (continued)
SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 16 OF 85
This section lists the system errors and warnings specific to the solvent manager in alphabetical order.
Locate the relevant error or warning in the left column and follow the appropriate troubleshooting steps.
Table 3: Solvent manager errors and warnings
Zero the accumulator pressure
transducer by restarting the
instrument.
The flow rate is too high for current Reduce the flow rate or change
column or solvent or solvents solvents.
Inspect the cable that runs from
the accumulator pressure
transducer to the CPU board and
replace.
Pressure transducer problem Replace the transducer.
See Transducer Replacement
For additional information, see:
Carousel and dogpin parts diagram Adjust the accumulator pressure
(2695 only), page 78 transducer voltage to 0.5V ± 0.2V.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 17 OF 85
Table 3: Solvent manager errors and warnings (continued)
Degasser vacuum pump problem Replace the degasser component.
See Degasser Assembly Component
Replacement
Plunger seal problem 1. Replace plunger seals.
2. Pressurize the seals to 4,000–
4,500 psi for 10 minutes.
Clogged filter 1. Clean the inlet reservoir filter.
2. Replace the filter if necessary.
See Replacing the In‐line Filter
Cable connection or problem 1. Verify that the cable connecting
the primary pressure transducer
to the CPU board is connected
and operational.
2. Replace the cable if necessary.
Pump primary pressure transducer Replace the transducer.
problem See Transducer Replacement
For additional information, see:
Carousel and dogpin parts diagram
(2695 only), page 78
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 18 OF 85
Table 3: Solvent manager errors and warnings (continued)
Degasser problem 1. Verify that the degasser pressure
is .9 psi or less (functioning
properly).
2. Replace the degasser
component.
See Degasser Assembly Component
Replacement
Poor fluidics conditioning Condition the fluidics by running
the Wet Prime function with
organic solvent, such as 100%
methanol.
Plunger seal problem 1. Replace plunger seals.
2. Pressurize the seals to 4,000–
4,500 psi for 10 minutes.
Clogged filter Clean or replace the inlet reservoir
filter.
Cable problem Inspect/Reconnect/Replace. the
cables that run from the pump
pressure transducers to the CPU
board.
Primary pressure transducer Replace the transducer.
problem See Transducer Replacement
See Carousel and dogpin parts
diagram (2695 only), page 78
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 19 OF 85
Table 3: Solvent manager errors and warnings (continued)
Poorly conditioned fluidics Condition the fluidics by running
the Wet Prime function with
organic solvent, such as 100%
methanol.
Leaky components 1. Inspect for leaks in the pin plug,
cap, check valves, plunger seals,
and fluid lines.
2. Repair or replace any
components with problems.
Table 4: Flow rates and compatible stroke volumes
Flow rate (mL/min) Stroke volume (µL)
<10 130
<3.03 100
<1.23 50
<0.53 25
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 20 OF 85
Table 3: Solvent manager errors and warnings (continued)
Blown 7a fuse on pump driver Replace the 7a fuse.
board See Pump Driver Board Fuses
Problem with the lower Replace the fan or heater/cooler.
compartment cooling fan or sample
heater/cooler fans
Power supply problem 1. Inspect the power supply.
2. Replace if necessary.
Plunger driver board problem Replace the board.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
Alliance CPU PCB problem 1. Verify and reconnect the CPU to
plunger PCB cable.
1. Replace the CPU to plunger PCB
cable.
2. Replace plunger driver PCB.
3. Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 21 OF 85
Table 3: Solvent manager errors and warnings (continued)
A gas bubble momentarily Ensure that the solvent lines are
interrupted solvent flow properly connected and are not
Note: The criterion for detection of allowing air into the system.
a bubble is the failure of the
Ensure that the solvents are
pre‐compression stroke to
properly degassed.
achieve system pressure
within 20% of the working Condition the fluidics by running
stroke. the Wet Prime function with
• The solvent delivery system is organic solvent, such as 100%
operating at near its maximum methanol.
pressure vs. flow envelope
• High gas loading of solvents
(inadequate solvent conditioning
within degas or sparge system)
Pump motor problem Replace the motor.
Plunger driver board problem Replace the board.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 22 OF 85
Table 3: Solvent manager errors and warnings (continued)
Waste‐line blockage Clear all blockages.
Excessive restriction between the Reduce the restriction.
pump and sample loop
Vent valve problem 1. Ensure that the sample loop can
vent the pump to atmosphere by
opening valves V1, V2, and V3.
2. Inspect valves.
3. Replace valves if necessary.
Zero the accumulator pressure
transducer by restarting the
instrument.
Accumulator check valve problem Replace the check valve.
See Replacing the Check Valves
Flow rate too high for column or Reduce the flow rate or change the
solvent solvent.
Cable problem Inspect/Replace the cable that runs
from the primary pressure
transducer to the CPU board.
Incorrect transducer voltage Adjust the primary pressure
transducer voltage to 0.5V ±0.2V.
See Carousel and dogpin parts
diagram (2695 only), page 78
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 23 OF 85
Table 3: Solvent manager errors and warnings (continued)
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
Transducer voltage out of range Adjust the system pressure
transducer voltage to 0.5V ±0.2V
See service note: Transducers,
Errors, and Adjusting (link opens in
IBM Notes)
Transducer problem 1. Verify functionality of the
primary pressure transducer.
2. Replace if necessary.
See Transducer Replacement
Plunger Seal Wash Pump problem 1. Inspect the single seal wash
pump.
2. Replace if necessary.
See Plunger Seal Wash Option
Installation
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 24 OF 85
Table 3: Solvent manager errors and warnings (continued)
Cable connection problem Reseat the cable connecting the
between the plunger drive PCB and plunger driver PCB and the CPU
the CPU PCB PCB.
Plunger driver board problem Replace the board.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See Alliance CPU PCB (715003814)
Incorrect cable connection Verify that the cables from the
motors and encoders are
connected correctly.
Piston drive problem 1. Inspect the piston drives.
Note: Piston drive = torpedoes. 2. Replace if necessary.
See:
• New‐Style Piston Drive
• Piston Drive Assembly Replace‐
ment
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 25 OF 85
Table 3: Solvent manager errors and warnings (continued)
Problem with cables or incorrect 1. Inspect the cables for damage.
cable connection 2. Inspect the cables for proper
connectivity.
3. Reconnect or replace cables if
necessary.
Plunger driver board problem Replace the board.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
System pressure transducer Replace the transducer.
problem See Transducer Replacement
SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 26 OF 85
This section lists the system errors and warnings specific to the sample manager in alphabetical order.
Locate the relevant error or warning in the left column and follow the appropriate troubleshooting steps.
Table 5: Sample manager errors and warnings
Problem with cable between 1. Inspect the cable between
carrier drive position sensor and carrier drive position sensor and
fluidics driver board fluidics driver board (J4).
2. Replace if necessary.
Misaligned carrier drive position 1. Inspect position sensor
sensor alignment.
2. Adjust if necessary.
Sample compartment debris Remove debris or obstacles from
sample compartment floor.
Carrier drive problem 1. Replace the carrier motor.
2. Replace the carousel carrier
drive assembly.
See Replacing the Carousel/Carrier
Drive (715004110)
Carousel carrier drive assembly Replace the carousel carrier drive
problem assembly.
See Replacing the Carousel/Carrier
Drive (715004110)
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
The carousel ID sensor could not
detect the carousel ID magnet, or
detected multiple carousel ID mag‐
nets.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 27 OF 85
Table 5: Sample manager errors and warnings
Carousel magnet missing, Replace the carousel.
incorrectly positioned, or of
incorrect polarity
Carousel ID sensor problem 1. Inspect the sensor and the cable
See Carousel/Carrier Normal that connects it to the PCB.
Operation 2. Replace if necessary.
1. Inspect the cable that runs from
the carrier drive position sensor
to the Fluidics Driver board.
2. Replace if necessary.
Ensure that the carrier drive
position sensor is properly aligned.
Remove any debris or obstacles
from the floor of the sample
compartment that may be affecting
the carrier.
Carrier drive problem Replace the carousel carrier drive
assembly.
See Replacing the Carousel/Carrier
Drive (715004110)
Carousel carrier drive assembly Replace the carousel carrier drive
problem assembly.
See Replacing the Carousel/Carrier
Drive (715004110)
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 28 OF 85
Table 5: Sample manager errors and warnings
Carousel malfunction Symptoms
• Attempted carousel move yields inaccurate confirmation from the
position sensor.
• Carousel drive encoder wheel unmapped during start‐up.
• Carousel moves, but the position sensor is inactive at the end of the
move.
Note: Carousel malfunction vs. carousel z axis malfunction
• Carousel malfunction: The carousel fails to find home, fails to map,
or fails to locate the encoder‐wheel slot positions.
• Carousel Z‐axis malfunction (X): The Z‐ram fails to engage the
carousel for one of five reasons.
Dusty encoder wheels or sensors 1. Dampen a lint‐free cloth with
only water and carefully wipe
the dust off the carousel and
carriage encoder wheels and
sensor. (Compressed air is an
acceptable alternative to the
dampened cloth).
2. Allow the encoder wheels and
sensor to dry to avoid the
attraction of dust.
Dirty sample‐compartment floor 1. Clean floor.
2. Verify that the carousels move
easily.
Carousel’s bottom side is rough, 1. Clean carousel bottom.
causing drag 2. Carefully polish the bottom with
800‐grit sand paper.
3. Replace carousel.
Needle pierce is outside of the Adjust the carrier position.
center of septa See Adjusting the Z‐Axis Ram
Height and the Carousel Carrier
Position (715002039)
Misaligned vial sensor impedes the Push back the vial sensor tube
carousels (located on the right‐hand side).
Note: The tip of the sensor must
The result is sporadic carousel extend no more than 0.1
malfunctions or z axis 5 errors inch from the surface of the
during, or shortly after, installation foam compartment.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 29 OF 85
Table 5: Sample manager errors and warnings
Carousel is not engaged by dogpin; 1. Inspect the dogpin slot on the
as a result, carrier improperly carousel for damage.
positions carousel under drive shaft 2. If damaged, replace the carousel.
Carousel drive problem Replace the carousel drive.
See Replacing the Carousel/Carrier
Drive (715004110)
Incorrect Z ram height 1. Verify the height of the Z‐ram
using the Z‐ram height gauge
The position of the carousel drive (700001022).
shaft may be too far up or down 2. Adjust position of sensors on the
when it engages a carousel drive shaft as necessary.
See Adjusting the Z‐Axis Ram
Height and the Carousel Carrier
Position (715002039)
The carrier plate may be out of Adjust the position sensors on the
alignment and putting too much carrier drive.
lateral force on the carousel,
preventing it from rotating
Firmware encountered an 1. Examine all carousel‐positioning
unexpected condition components and take action as
necessary.
2. If still unresolved, examine the
trace log to determine
instrument activity at the time of
failure.
3. Contact GSS and be prepared to
describe the conditions leading
to this error.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 30 OF 85
Table 5: Sample manager errors and warnings
Z axis motor problem 1. Ensure that the z axis motor is
connected to the fluidics driver
board.
2. Ensure that the motor is moving
the z axis.
3. Replace the z axis motor if
necessary.
Dogpin misaligned Carousel carrier drive has a blunt‐
cut dogpin that has difficulty
See Carousel and dogpin parts dropping into a carousel slot.
diagram (2695 only), page 78 1. Ensure that the dogpin is seated
properly within notches.
Height of carousel carrier drive in Adjust height (gauge block
compartment 700001022).
See Adjusting the Z‐Axis Ram
Height and the Carousel Carrier
Position (715002039)
Black, plastic, optical fence hits Note: The bracket may be out of
sensor or does not position into tolerance. The small change
sensor properly in tolerance is amplified
onto the optical fence
sensor.
The sensor is located at the bottom
of the metal bracket.
1. Monitor the drive while it is in
motion to determine if the
optical fence’s downward
movement causes the fence to
rub against the inside wall of the
sensor.
2. Adjust the bracket adjustment
screw until alignment is
achieved, if necessary.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 31 OF 85
Table 5: Sample manager errors and warnings
Loose encoder wheel due to 1. Ensure that the encoder staking
improper staking is within tolerance.
2. If encoder wheel is too loose,
replace carousel carrier drive.
Carousel misaligned 1. Inspect carousel.
2. Align carousel if necessary.
Debris Verify that the floor is clean and the
carrier plate moves correctly.
Misaligned vial sensor impedes Push back the vial sensor tube
carousels (located on the right‐hand side).
Note: The tip of the sensor must
extend no more than 0.1
inch from the surface of the
foam compartment.
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
Z axis motor problem 1. Ensure that the z axis motor is
connected to the fluidics driver
board.
2. Ensure that the motor is moving
the z axis.
3. Replace the z axis motor if
necessary.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 32 OF 85
Table 5: Sample manager errors and warnings
Z axis motor problem 1. Ensure that the z axis motor is
connected to the fluidics driver
board.
2. Ensure that the motor is moving
the z axis.
3. Replace the z axis motor if
necessary.
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
Z axis motor problem 1. Ensure that the z axis motor is
connected to the fluidics driver
board.
2. Ensure that the motor is moving
the z axis.
3. Replace the z axis motor if
necessary.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 33 OF 85
Table 5: Sample manager errors and warnings
Height of carousel carrier drive in Adjust height using gauge block,
compartment 700001022.
See Adjusting the Z‐Axis Ram
Height and the Carousel Carrier
Position (715002039)
Black, plastic, optical fence hits Note: The bracket may be out of
sensor or does not position into tolerance. The small change
sensor properly in tolerance is amplified
onto the optical fence
sensor.
The sensor is located at the bottom
of the metal bracket.
1. Monitor the drive while it is in
motion to determine if the
optical fence’s downward
movement causes the fence to
rub against the inside wall of the
sensor.
2. If necessary, adjust the bracket
adjustment screw until
alignment is achieved.
Carousel carrier drive shaft doesn't 1. Verify by pushing up on dogpin
move freely area.
2. Replace carousel carrier drive if
necessary.
Loose encoder wheel due to 1. Ensure that the encoder staking
improper staking is within tolerance.
2. If encoder wheel is too loose,
replace carousel carrier drive.
Carousel misaligned 1. Inspect carousel.
2. Align carousel if necessary.
Debris Verify that the floor is clean and the
carrier plate moves correctly.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 34 OF 85
Table 5: Sample manager errors and warnings
Seal problems 1. Inspect for seal leaks and resolve.
2. Rerun Adjust seal pack.
3. Replace/rebuild seals as
necessary.
See:
• Installing an Alliance injector seal
pack
• Seal Pack Replacement/Rebuild
Kits
Syringe leak 1. Inspect the syringe for presence
of bubbles.
2. Replace the syringe if necessary.
Valve leak 1. Run automated valve leak test.
2. Inspect and if necessary, replace
valves V1, V2, or V3.
Poor solvent conditioning 1. Confirm the proper degassing of
solvents.
2. Confirm the absence of bubbles
in the syringe.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 35 OF 85
Table 5: Sample manager errors and warnings
Syringe leak 1. Inspect syringe for presence of
bubbles.
2. Replace if necessary.
Valve leak 1. Inspect all valves.
2. Run automated valve test.
3. Replace valve if necessary.
Damaged cable 1. Inspect the cable that runs from
the sample heater/cooler
thermistor (J9 on PCB) to the
column heater PCB.
2. Replace.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 36 OF 85
Table 5: Sample manager errors and warnings
Sample heater/cooler PCB problem Replace the PCB.
Sample heater/cooler or sample Replace the heater/cooler or
heater/cooler hardware problem hardware.
See Installing the Sample
Compartment Heater/Cooler
Assembly (715002495)
Figure 1 ‐ Vial sensor inside curved
pathway
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 37 OF 85
Table 5: Sample manager errors and warnings
For example: when there are 10
vials in a tray and you remove vials
8 and 9, vial 8 is recognized as
missing, but vial 9 is recognized as
present.
Perform this Random Vial Position
test.
See Performing the Random Vial
Position Test, page 75
Vial transparency Ensure that the vial caps are
Note: If a vial is actually present, present and are opaque enough for
the error indication could be the vial sensor to detect them.
due to the transparency of
1. Inspect the vial sensor for dust.
the vial cap.
2. Clean if necessary.
Ensure that the vial sensor emitter
and receiver are aligned properly.
Ensure that there is nothing
blocking the vial sensor.
Cable problem 1. Inspect the fiber optics cable
that connects the vial sensor to
the fluidics driver board.
2. Replace the cable if necessary.
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 38 OF 85
Table 5: Sample manager errors and warnings
Needle mechanism problem Replace the needle mechanism.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 39 OF 85
Table 5: Sample manager errors and warnings
Incorrectly installed needle Reinstall the needle.
Needlewash issue 1. Verify that the needlewash is
using appropriate solvents.
2. Verify that the needlewash is
primed.
Firmware encountered an 1. Examine the trace log to
unexpected condition determine instrument activity at
the time of failure.
2. Contact GSS and be prepared to
describe the conditions leading
to this error.
Needle bottom sensor is dirty Clean sensor with compressed air.
Problem with needle bottom 1. Power‐off and manually move
sensor the flag in between the sensors
to reinitialize.
2. Replace the sensor.
3. Replace the injector.
Incorrectly installed needle Reinstall the needle.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 40 OF 85
Table 5: Sample manager errors and warnings
Vial/needle misalignment 1. Ensure that the vial is aligned
with the needle.
2. If the needle is misaligned, adjust
the position of the carrier and
carousel drives. (269x only)
See Carousel Rotation Adjustment
Procedure
Needle motor problem 1. Confirm that the needle motor is
connected to the PCB and the
needle moves as intended.
2. If not, replace the motor.
Roll Pin is loose or falls out when: Replace the injector.
• the needle is in the vial position
• the needle is in a position within
the sensors
Roll pin
Motor
screw
shaft
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 41 OF 85
Table 5: Sample manager errors and warnings
Dirty needle top sensor Clean sensor with compressed air.
Problem with needle top sensor 1. Power‐off and manually move
the flag in between the sensors
to reinitialize.
2. Replace the sensor.
3. Replace the injector.
Incorrectly installed needle Reinstall the needle.
Seal pack problem Replace the seal pack.
See Seal Pack Information See:
• Installing an Alliance injector seal
pack
• Seal Pack Replacement/Rebuild
Kits
Vial/needle misalignment 1. Ensure that the vial is aligned
with the needle.
2. If the needle is misaligned, adjust
the position of the carrier and
carousel drives.
See Carousel Rotation Adjustment
Procedure
Needle motor problem 1. Confirm that the needle motor is
connected to the PCB.
2. Confirm that the needle motor is
moving the needle as intended.
3. Replace the needle motor if it is
not operating as required.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 42 OF 85
Table 5: Sample manager errors and warnings
Fluidics leak 1. Inspect for fluid leakage and
repair.
2. Adjust the needle seal pack.
Incorrect seal position Adjust the seal position.
See Adjusting the Seals
Valve V3 problem Replace the valve.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 43 OF 85
Table 5: Sample manager errors and warnings
Adjust the sample chamber set
point to 23 °C.
Thermistor connection problem 1. Verify connection to PCB.
with PCB 2. Replace if necessary.
Thermistor problem Inspect/Reconnect/Replace.
Thermistor problem Inspect/Reconnect/Replace.
Sample heater/cooler PCB problem 1. Test the PCB.
See Sample Heater/Cooler Testing,
page 74
2. Replace if necessary.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 44 OF 85
Table 5: Sample manager errors and warnings
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See Alliance CPU PCB (715003814)
Sample heater/cooler thermistor 3. Inspect/Reconnect/Replace.
problem
Fan problem 1. Inspect the circulating fan and
the external fan.
2. Replace if necessary.
PCB problem Replace the sample heater/cooler
PCB.
Sample heater/cooler engine Replace the heater/cooler.
problem See Installing the Sample
Compartment Heater/Cooler
Assembly (715002495)
Plunger driver board problem Replace the board.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 45 OF 85
Table 5: Sample manager errors and warnings
Sample heater/cooler PCB problem Replace the PCB.
Sample heater/cooler engine Replace the heater/cooler.
problem See Installing the Sample
Compartment Heater/Cooler
Assembly (715002495)
Plunger driver board problem Replace the board.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
Sample heater/cooler engine Replace the heater/cooler.
The value returned by the thermis‐ problem See Installing the Sample
tor is not recognized as valid. Compartment Heater/Cooler
Assembly (715002495)
Sample heater/cooler engine Replace the heater/cooler.
The value returned by the thermis‐ problem See Installing the Sample
tor is not recognized as valid. Compartment Heater/Cooler
Assembly (715002495)
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 46 OF 85
Table 5: Sample manager errors and warnings
The seal pack bottom edge, as cal‐
culated by the Adjust Seals diagnos‐
tic, is too low.
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Seal pack geometry fault (3)
The seal pack, as calculated by the
Adjust Seals diagnostic, is too wide.
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Seal pack geometry fault (4)
The seal pack, as calculated by the
Adjust Seals diagnostic, is too nar‐
row.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 47 OF 85
Table 5: Sample manager errors and warnings
Damaged needle or incorrectly 1. Replace or rebuild the seal pack.
installed needle See:
• Rebuilding the Seal Pack
(This is a known issue with the flag/
stop on the injector and with the • Seal Pack Replacement/Rebuild
injector tester used in Kits
manufacturing, the needle may • Seal Pack Rebuild Kit w/Needle
have been stressed enough during • Installing an Alliance Injector Seal
manufacturing testing to cause it to Pack
bow or bend.) 2. Replace the needle.
(The internal seals are the same
as the original seal pack.)
Injector valve problem 1. Run the automated valve test.
2. Run compression test.
3. Repair or replace valves if
necessary.
See Rebuilding the Injector Valves
Problem with sample transducer 1. Inspect sample transducer.
2. Correct any issues discovered or
replace the transducer.
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 48 OF 85
Table 5: Sample manager errors and warnings
Fluidics driver board problem Replace the fluidics board.
See:
Alliance 269x: Fluidics Driver PCB
Alliance 279x: Fluidics Driver PCB
Firmware encountered an 1. Examine the trace log to
unexpected condition determine instrument activity at
the time of failure.
2. Contact GSS and be prepared to
describe the conditions leading
to this error.
Blocked sensor Remove obstruction.
See Vial Present Sensor
Maintenance
SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 49 OF 85
Degasser errors
This section lists the system errors and warnings specific to the degasser in alphabetical order. Locate the
relevant error or warning in the left column and follow the appropriate troubleshooting steps.
Note: See also PerformancePLUS Degasser Troubleshooting
Table 6: Degasser errors and warnings
Pressure sensor problem 1. Inspect the pressure sensor on
the degasser PCB.
2. Replace if necessary.
Degasser PCB problem Replace the degasser PCB.
See Degasser PCB (700002511) See Degasser Assembly Component
(715003571) Replacement
Vent line blockage Primary symptom is a loud, high‐
pitched sound emitting from the
degasser vacuum pump.
1. Inspect the vent tube for any
kinks or blockages.
2. Repair or replace.
DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 50 OF 85
Table 6: Degasser errors and warnings
Problem with control cable Replace the control cable
(between the degasser and Alliance (WAT270908).
CPU)
Degasser PCB problem 1. Verify that the dip switches on
See Degasser PCB (715003571) the PCB are set correctly.
2. Replace the PCB if necessary.
See Degasser Assembly Component
Replacement
Problem with control cable Replace the control cable
(between the degasser and Alliance (WAT270908).
system)
Degasser vacuum pump problem 1. Inspect the degasser vacuum
pump.
2. Replace.
See Degasser Assembly Component
Replacement
Degasser PCB problem 1. Verify that the dip switches on
See Degasser PCB (715003571) the PCB are set correctly.
2. Replace the PCB if necessary.
See Degasser Assembly Component
Replacement
DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 51 OF 85
Table 6: Degasser errors and warnings
Degasser vacuum pump problem 1. Inspect the vent tube for any
kinks or blockages.
2. Inspect the vacuum pump.
3. Replace if necessary.
See Degasser Assembly Component
Replacement
Degasser PCB problem Replace the PCB.
See Degasser PCB (715003571) See Degasser Assembly Component
Replacement
Vacuum pump problem Replace the vacuum pump.
See Degasser Assembly Component
Replacement
DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 52 OF 85
Table 6: Degasser errors and warnings
Vacuum chamber problem 1. Inspect the vacuum chamber.
2. Replace the vacuum chamber
(700001218) if necessary.
See Degasser Assembly Component
Replacement
Sparge assembly problem Replace the sparge assembly.
See Sparge Assembly
Plunger driver board problem Replace the Plunger PCB.
Note: Plunger driver board = pump See Pump Driver Board
driver board. Replacement
See Pump Driver PCB ‐ Reference
Information
DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 53 OF 85
This section lists the system errors and warnings specific to the column device in alphabetical order.
Table 7: Column device errors and warnings
Cable problem 1. Inspect the cable that connects
the column heater thermistor to
the PCB.
2. Replace if necessary.
Thermistor problem Inspect/Reconnect/Replace.
Column heater PCB problem Replace the PCB.
COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 54 OF 85
Table 7: Column device errors and warnings (continued)
Blown fuse on column heater PCB. Replace the fuse or the PCB.
Blown fuse on column heater Replace the fuse or the PCB.
connection board (attached
internally to column heater
connector)
Column heater PCB problem Replace the PCB.
Hardware watchdog time‐out for
the column heater.
Column heater engine problem Replace the column heater engine.
Column heater PCB problem Replace the PCB.
Blown fuse on column heater Replace the fuse or the PCB.
connection board (attached
internally to column heater
connector)
Column heater PCB problem Replace the PCB.
Column heater PCB problem Replace the PCB.
Column heater engine problem Replace the column heater engine.
COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 55 OF 85
Table 7: Column device errors and warnings (continued)
Ambient operating temperature Determine the room temperature
limit exceeded and adjust if necessary.
See related: Column heater not
ambient, page 53
Cable problem or disconnection 1. Inspect the cable that runs from
the column heater thermistor to
the column heater PCB.
2. Replace if necessary.
Thermistor problem 3. Inspect/Reconnect/Replace.
PCB problem 1. Inspect the PCB.
2. Replace if necessary.
Column heater hardware problem 1. Inspect the column heater
hardware.
2. Replace engine.
COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 56 OF 85
Table 7: Column device errors and warnings (continued)
Incorrect setting Ensure that the dial/potentiometer
setting on the column heater is
correct for the installed switching
valve.
See Column device heater dial/
potentiometer settings, page 75
Column heater configuration Power cycle the column heater to
misread force the system to reread the
column heater configuration.
Separation method issue Modify the separation method to
eliminate the column position
change.
Valve initialized incorrectly 1. Power‐down Alliance.
2. Wait one minute and power‐on.
3. After Alliance initialization, use
the front panel to reposition the
switching valve.
If the valve position remains
unchanged, reboot again using
the one‐minute wait time.
4. If you are unsuccessful after
three attempts, additional
troubleshooting is required.
Contact GSS.
Valve PCB problem Replace the PCB.
Column heater PCB problem 1. Inspect the PCB.
2. Replace if necessary.
COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 57 OF 85
Table 7: Column device errors and warnings (continued)
Problem with door gasket Repair or replace.
Column heater set point too close Ensure that the temperature set
to ambient point is a minimum of 5 °C above
ambient.
Ambient operating temperature 1. Determine the room
limit exceeded temperature.
2. Adjust if necessary.
See related: Column heater not
ambient, page 53
Damaged cable 1. Inspect the cable that runs from
the column heater thermistor to
the column heater PCB.
2. Replace if necessary.
Column heater temperature Inspect/Reconnect/Replace.
thermistor problem
Column heater PCB problem Replace the PCB.
COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 58 OF 85
This section lists the system errors and warnings specific to the control of the detector in alphabetical
order. Locate the relevant error or warning in the left column and follow the appropriate troubleshooting
steps.
For troubleshooting information pertinent to each detector, see:
• ACQUITY QDa Detector: troubleshooting information
• 2414 RI Detector: troubleshooting information
• 2424 ELS Detector: troubleshooting information
• 2998 PDA Detector: troubleshooting information
• 2489 UV/Vis Detector: troubleshooting information
• 2475 FLR Detector: troubleshooting information
Table 8: Detector control errors and warnings
Incorrect detector address 1. Ensure that the IEEE address of
the detector is correct.
2. Rescan the IEEE bus from the
detector window.
Problem with IEEE cable between 1. Inspect the IEEE cable between
Alliance system and detector the Alliance system and the
detector.
Note: Power‐off before removing
or attaching the IEEE cable.
2. Reconnect.
3. Replace if necessary.
4. Rescan the IEEE bus from the
detector window.
410 detector CPU board problem 1. Replace the 410 detector CPU
board.
2. Rescan the IEEE bus from the
detector’s window.
Alliance CPU PCB problem 1. Replace the Alliance CPU PCB
board.
2. Rescan the IEEE bus from the
detector’s window.
See technical reference sheet,
Alliance CPU PCB (715003814)
DETECTOR CONTROL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 59 OF 85
Table 8: Detector control errors and warnings
Problem with IEEE cable between 1. Inspect the IEEE cable between
Alliance system and detector the Alliance system and the
detector.
2. Reconnect the cable.
3. Replace if necessary.
4. Rescan the IEEE bus from the
detector window.
486 detector CPU problem Replace the 486 detector CPU
board.
Alliance CPU PCB problem 1. Replace the Alliance CPU PCB
board.
2. Rescan the IEEE bus from the
detector’s window.
See technical reference sheet,
Alliance CPU PCB (715003814)
DETECTOR CONTROL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 60 OF 85
Dissolution errors
This section lists the system errors and warnings specific to Dissolution in alphabetical order. Locate the
relevant error or warning in the left column and follow the appropriate troubleshooting steps.
Table 9: Dissolution errors and warnings
Allow for a greater tolerance in the
bath temperature by changing the
limits.
DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 61 OF 85
Table 9: Dissolution errors and warnings
Vial sensor problem 1. Clean sensor.
2. Verify sensor alignment.
3. Replace vial sensor.
4. Confirm carousel carrier
movement.
5. Verify that there are no
obstructions.
Wrong carousel Verify proper carousel and sensor
identification of that carousel.
Z axis problem Verify z axis operation/depth.
See Adjusting the Z‐Axis Ram
Height and the Carousel Carrier
Position (715002039)
Problem with cable from WTM 1. Verify cable connectivity and
Comm2 to Bath Comm1 functionality.
2. Replace if necessary.
WTM/Bath CPU problem 1. Verify that the rear panel LED is
green.
• If green, CPU is good.
• If red, replace CPU.
DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 62 OF 85
Table 9: Dissolution errors and warnings
Dispenser assembly problem 1. Verify dispenser operation.
2. Replace if necessary.
Fluidic driver PCB problem Firmware should be >3.0.
(See service note) (link opens in
IBM Notes)
Firmware problem Verify that the fluidic driver PCB is
the correct revision.
See service note (rev 0 or 6) (link
opens in IBM Notes)
Dispenser assemble jam 1. Verify dispenser operation.
Fluidics driver board problem Firmware should be >3.0.
(See service note) (link opens in
To help determine the source of IBM Notes)
the problem, use the procedure:
Troubleshooting the XYZ and Verify that the fluidic driver PCB is
needle bottom sensor errors (279x the correct revision.
only), page 85. See service note (rev 0 or 6) (link
When performing the procedure, opens in IBM Notes)
test the combination: Replace the fluidics board.
• Dissolution top sensor with corre‐ See:
sponding LED DS19 on the fluidics Alliance 269x: Fluidics Driver PCB
driver board. Alliance 279x: Fluidics Driver PCB
• Dissolution bottom sensor with
corresponding LED DS20 on the
fluidics driver board.
Use the information obtained from
that procedure to determine the
cause and corrective action.
DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 63 OF 85
Table 9: Dissolution errors and warnings
WTM guide post alignment 1. Remove all syringes.
problem 2. Verify up/down movement.
3. Lubricate guide post.
4. If problem remains unresolved,
contact GSS.
Scan sensor optics problem 1. Verify that fiber optic cable is in
correct position.
2. Verify that fiber optic cable is
working properly.
Fluidics driver board problem Replace the fluidics board. See:
Alliance 269x: Fluidics Driver PCB
Alliance 279x: Fluidics Driver PCB
DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 64 OF 85
This section lists the system errors and warnings specific to Alliance HPLC 2795 and 2796 in alphabetical
order. Locate the relevant error or warning in the left column and follow the appropriate troubleshooting
steps.
Table 10: Alliance HPLC 2795 and 2796 errors and warnings
Problem with IEEE cable between 1. Inspect the IEEE cable between
Alliance system and detector the Alliance system and the
detector.
2. Reconnect or replace the cable if
necessary.
3. Rescan the IEEE bus from the
detector window.
410/2410 detector CPU board Replace the detector CPU board
problem and then rescan the IEEE bus from
the detector’s window.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 65 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Lead screw not moving Use diagnostics to test movement.
Problem with sensor Replace the sensor.
See:
• Lockpin Motor and Sensor
• Lockpin Drive Sensor
Problem with lockpin motor Replace the motor.
See:
• Lockpin Motor and Sensor
• Lockpin Drive Motor
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
Debris/obstacles on compartment Remove any obstacles and clean
floor affecting carrier the compartment floor.
Carrier drive assembly problem Replace the carrier motor or drive
See also: assembly.
• Carrier Assembly
• Carrier Drive
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 66 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Degasser PCB problem 1. Replace degasser PCB.
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 67 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Blocked waste line Clear any blockage of the waste
line.
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 68 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Needle bottom sensor problem 1. If dusty, clean sensor.
2. Verify that the cable is connected
properly between sensor and
PCB.
3. If malfunctioning, replace.
Needle deck height mis‐calibrated Run the Replace Needle diagnostic
to recalculate the deck height.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 69 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Problem with cable 1. Inspect the cable and reconnect.
(441000459) 2. Replace if necessary.
RS‐232 valve electronics problem Replace the valve.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
RS‐232 valve electronics problem Replace the valve.
Alliance CPU PCB problem Replace the Alliance CPU PCB
board.
See technical reference sheet,
Alliance CPU PCB (715003814)
Degasser PCB problem Replace degasser PCB.
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 70 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Incorrect needle installation Reinstall the needle.
See:
• 2790/2795 Needle
• 2796 Needle
• Rebuild/Replace 2795 Needle
Assembly
To help determine the source of the problem, use the procedure,
Troubleshooting the XYZ and needle bottom sensor errors (279x only),
page 85.
When performing the procedure, test the combination:
• Needle bottom sensor with corresponding LED DS13 on the fluidics
driver board.
• Needle top sensor with corresponding LED DS16 on the fluidics driver
board.
Use the information obtained from that procedure to determine the cause
and corrective action below.
Sensor problem Replace the respective sensor.
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
Needle bottom sensor problem 1. If dusty, clean sensor.
2. Verify that the cable is connected
properly between sensor and
PCB.
3. If malfunctioning, replace.
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 71 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
X axis malfunction To help determine the source of the problem, use the procedure,
Troubleshooting the XYZ and needle bottom sensor errors (279x only),
An x axis movement failed to reach page 85.
the requested position. When performing the procedure, test the combination:
• X axis home sensor with corresponding LED DS14 on the fluidics driver
board.
See:
• Alliance 279x: Fluidics Driver PCB
• Adjusting the XY Mechanism in 2790/2795/2796 Separations Modules
Dusty x axis rails Clean the x axis rails.
Dusty x axis home sensor Clean the x axis home sensor.
X axis home sensor problem 1. Adjust x axis sensor.
2. Replace the x axis home sensor if
necessary.
See X‐Axis Home Sensor
Cable disconnection or problem 1. Inspect connectivity of cables
and reconnect if required.
2. If cable is damaged or
malfunctioning, replace.
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
The x axis belt may be slipping or Tighten or replace the belt.
broken See X Traverse Belt
Misaligned x axis rails 1. Verify that the belt and pulley
are not to tight or loose. Adjust if
necessary.
2. Align the x axis rails.
Problem with x axis motor Replace the x axis motor.
See X‐Axis Motor
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 72 OF 85
Table 10: Alliance HPLC 2795 and 2796 errors and warnings (continued)
Y axis malfunction To help determine the source of the problem, use the procedure,
Troubleshooting the XYZ and needle bottom sensor errors (279x only),
A y axis movement failed to reach page 85.
the requested position. When performing the procedure, test the combination:
• Y axis home sensor with corresponding LED DS15 on the fluidics driver
board.
Use the information obtained from that procedure to determine the cause
and corrective action below.
See:
• Alliance 279x: Fluidics Driver PCB
• Adjusting the XY Mechanism in 2790/2795/2796 Separations Modules
Dusty y axis rails Clean the y axis rails.
Dusty y axis home sensor Clean the y axis home sensor.
Y axis home sensor problem Replace the y axis home sensor.
See:
• Y‐Axis Home Sensor
• YZ and Needle Bottom Breakaway
PCB Assembly
Cable disconnection or problem 1. Inspect connectivity of cables
and reconnect if required.
2. If cable is damaged or
malfunctioning, replace.
Fluidics driver board problem Replace the fluidics board.
See Fluidics Driver PCB
The y axis belt may be slipping or Tighten or replace the y axis belt.
broken See Y Traverse Belt
Misaligned y axis rails Verify that the pulley and belt are
not too tight or loose and adjust if
necessary.
Problem with y axis motor Replace the y axis motor.
See Y‐Axis Drive Motor
ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 73 OF 85
Reference information
This section describes common tests and procedures that are referenced within this guide.
PSI vs. Flow Envelope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 73
Sample Heater/Cooler Testing . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 74
Troubleshooting Transducer Problems . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 74
Column device heater dial/potentiometer settings . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 75
Performing the Random Vial Position Test . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 75
Needle Piercing Alignment Procedure . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 76
Carousel and dogpin parts diagram (2695 only) . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 78
Z axis ram height gauge (700001022) settings . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 79
2690 Fluidics Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 80
2690 Tubing Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 81
2690 Interconnect Diagram . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 82
Server Busy Error: Troubleshooting . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 83
Troubleshooting the Vacuum Pump. . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . 84
Troubleshooting the XYZ and needle bottom sensor errors (279x only). . . . . . .. . . . . . . . . . . . . . . . . . . . . 85
When testing the Sample Heater/Cooler with a calibrated thermocouple: place the thermocouple in the
compartment over the center of the carousel.
Notes:
—Be aware that temperature degradation occurs at ambient temperatures above 29 °C.
—The heater/cooler can only attain internal temperatures of 4 °C or 25 °C below
ambient. For example, if the ambient temperature is 30 °C the compartment
temperature can only attain a temperature of 5 °C or a 25 °C differential.
A transducer failure can occur during initialization. If the software cannot electronically zero the
transducer, an "Out of range" error is displayed.
Note: During initialization, the transducer voltage must be within a specific range. The range to pass
power‐up initialization is 0.5V +/‐0.2V (0.3V ‐ 0.7V). If the voltage is outside of this range, an
"Out of range" error occurs. You can view the voltage in Service diagnostics, Service Utilities in
a pane containing the voltage offsets. An ideal transducer has a reading of 0.5V.
You can swap cable connections at the CPU board location to troubleshoot, isolate and verify when a
specific transducer has failed. If the problem follows the transducer, attempt to adjust or replace the
transducer. You should also inspect the cables and connections.
If back pressure is present on the transducers, you could see the same error. Verify that all pressure has
been released from the fluid components while troubleshooting this failure.
Re‐initializing
Transducers are auto‐zeroed during re‐initialization.
Transducer Adjustments
You can manually make transducer adjustments by pre‐zeroing the transducer to bring it into an
acceptable range for the electronic zero during power‐up. Remove the black rubber stopper on the
transducer to access the adjustment potentiometer. You can perform the adjustment by either trial/error,
or by carefully checking the voltage on a microchip.
Note: Ideal transducer reading is 0.5V. ±0.2V.
Trial‐and‐error method
Record the Out of range number displayed, power‐off the instrument, make a slight adjustment to the
transducer, then power‐on the instrument. Continue this adjustment procedure until the Out of range
number is decreased to eliminate the error.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 75 OF 85
Monitoring voltage method
Power‐on the instrument and connect the positive lead of the voltmeter to the following chips on the CPU
PCB to monitor the transducer offset voltage and connect the negative lead to chassis ground. If possible,
adjust the potentiometer to approximately 0.5V.
Table 11:
Transducer Chip and Pin #
System: U49 Pin 7
Primary: U49 Pin 1
Sample: U51 Pin 7
Note: Content for this topic sourced from service note.
Ensure that the dial/potentiometer setting on the column device heater is correct for the installed
switching valve. The switch is located on the back of the column module.
Switch position and valve type:
• 0: No valve
• 1: Three‐Column Select Valve Upgrade Kit (205000162)
• 7: 2‐Position Regeneration Valve Upgrade Kit (205000163)
• F: 6‐Column Select Valve Upgrade Kit (205000164)
This procedure is applicable to Alliance 269x.
1. Install vials with caps in positions 59, 60, 62, 63, 83, 84, 86, and 87. Install no vials in positions 61
and 85.
2. Run the method.
3. Ensure that the sample set stops at vial 85 and the alarm “VialMissingError” displays.
4. Install a vial in carousel position 85 and resume injections.
5. Ensure that the sample set stops at vial 61 (the alarm “VialMissingError” displays).
6. Install a vial in position 61 and resume injections.
• If injections continue as specified in the method, the test passes and no adjustment is
needed.
• If the test fails, the vial’s sensor optics are out of position and require adjustment.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 76 OF 85
The content that follows was originally publish as a service note. See service note: Needle Piercing
Alignment.
Scope Technical Discussion
If the injector is piercing off center and the injector motor loses steps due to striking the vial cap or the
side of the vial the error Injector malfunction "bottom sensor error" (bse) will be displayed. In this case,
the BSE is not related to an issue with the injector, but due to a misalignment of the carrier portion of the
carousel‐carrier drive. The Alliance methods below should be used to diagnose the carrier misalignment
problem globally with all the carousels. The carousel carrier drive may be adjusted if any of the piercings
are grossly out of alignment.
Product Affected
2695, 2695D
Tools required:
Adjustment requires a T‐10 Torx® driver.
Alignment Procedure
Determine current needle piercing locations
1. Use the piercing template on all the carousels or some other covering, such as parafilm, to see
the needle holes.
Note: If you use masking tape, be sure to thoroughly wipe the needle with a lintless wiper (for
example, KimWipe), moistened with methanol, after completion of the test.
2. Open the service note Needle Piercing Alignment, and download the Alliance methods below to a
diskette, and import them into the 2695.
• OPTOTEST.MTH
• TESTCCD.MTH
3. Verify that the needle wash system has the appropriate solvent and is primed.
4. Go into Menu > Status and set up the Optotest as your Separation Method (0 ml/min flow).
5. Run the Testccd Sample Method (runs 20 injections of 0 ul that have a runtime of .01 min).
6. Inspect all the needle holes for alignment and determine if the carousel needs to be adjusted
clockwise (CW) or counter‐clockwise (CCW).
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 77 OF 85
Align the needle piercing locations
7. Loosen the sensor lockdown screw located to the left of the carousel rotation motor.
8. Rotate the spring‐tensioned sensor adjustment screw located on the right‐hand side of the
carousel rotation motor.
• To rotate the carousel position CW, turn the sensor screw CCW.
• To rotate the carousel position CCW, turn the sensor screw CW.
9. After each adjustment, tighten the lockdown screw and run test injections in several locations on
multiple carousels to ensure that the adjustment is optimized for all carousels.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 78 OF 85
See the following service notes for more information regarding the use of the ram height gauge:
• Carousel Carrier Drive Sensor Adjustments
• 2695 Zram Height and Carrier Plate Adjustment
• Z Axis Malfunction vs Carousel Malfunction
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 80 OF 85
Possible Cause:
Interactions between Empower and the Quickset window cause a freeze.
Affected systems:
• Single systems
• Multiple systems configured on the acquisition machine.
Affected configuration:
A 2695 with both a 2489 and 2996 detector configured.
1. The 2695 pressure console freezes at the last recorded pump pressure.
2. Task Manager indicates that the instrument server is using approximately 25% of one of the CPUs.
3. Approximately two minutes after the abort/inject, Server Busy displays on the Quickset window.
4. A single error message is recorded in the Message Center => "2489XXXX instrument currently in
use" (where XXXX is the instrument serial number).
Corrective Action:
No action required.
This issue was resolved in Empower 2 Feature Release 5 and in all subsequent releases.
If using any version of Empower 2 prior to FR5, avoid using the abort button during acquisition.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 84 OF 85
1. Inspect and resolve the following:
• Inspect the vent line for liquid.
• Inspect manifold tubing for kinks, holes, and a tight fit.
• Pull off and reseat ALL vacuum connections to eliminate the flashing effect from the manifold
molding process.
2. If pump fails to power‐on, inspect:
• Power cable connection.
• Vacuum pump motor connection.
• Voltage at PCB connection.
• Vacuum pump to see if it stalled.
3. If error still occurs, troubleshoot the pathway by bypassing/removing the chambers and only use
vacuum pump and PCB.
• Pinch off tubing to chambers to troubleshoot individual chamber. Confirm that the pressure
reaches 0.9 psi.
• You might have to change the tubing configuration.
Connect the vacuum pump directly to the PCB Transducer (use the port closest to the edge
with the solder leads).
Confirm that the pressure reaches 0.9 psi.
• Check LEDs. See Degasser PCB (715003571).
4. Replace components as necessary.
See Degasser Assembly Component Replacement
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 85 OF 85
Troubleshooting the XYZ and needle bottom sensor errors (279x only)
You can use the LEDs on the fluidics driver board to determine if an error is caused by the sensor, cable, or
fluidics driver board. See also: Fluidics Driver PCB LEDs and Sample Handling Sensors.
This procedure is useful when troubleshooting the following errors:
• Top and bottom needle sensors active, page 70
• X axis malfunction, page 71
• Y axis malfunction, page 72
These instructions are written generically. Adapt for the specific sensor/LED combination in question.
Note: DS13, 14, 15, and 16 are visible through a cutout on the back side of the PCB.
To identify the malfunctioning component:
Place the flag into the sensor.
• If the corresponding LED turns ON, the sensor, cable and fluidics driver board are functioning
correctly.
If the corresponding LED remains OFF, troubleshoot the sensor, cable, PCB chain.