You are on page 1of 85

715003674 REV.

C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 1 OF 85

Alliance HPLC System Troubleshooting Guide


This manual is a reference tool for field service engineers troubleshooting problems in Alliance HPLC 
Systems. 

Table of contents

Troubleshooting errors . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . . 2
All errors/warnings index . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . . 2
System level errors  . . . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . . 7
Solvent manager errors. . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 16
Sample manager errors. . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 26
Degasser errors  . . . . . . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 49
Column device errors  . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 53
Detector control errors . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 58
Dissolution errors  . . . . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 60
Alliance HPLC 2795 and 2796 errors  . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 64
Reference information  . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 73

Purpose

This document outlines recommended steps for troubleshooting system‐level symptoms in the Alliance 
HPLC System.

Scope

This document is valid for the following Alliance HPLC systems:

• Alliance 2690, 2695, and e2695 Systems
• Alliance 2790, 2795, and 2796 Systems
This document contains error troubleshooting information for each of the following modules:

• Alliance HPLC System • Degasser • Detector control


• Solvent Manager  • Column device • Dissolution
• Sample Manager

How to use

1.   Find the relevant problem in each section and click the corresponding link to troubleshoot.
2.   Perform troubleshooting steps in the numbered order listed in the troubleshooting tables.

Note:  Troubleshooting recommendations are formatted in easy‐to‐scan tables that list possible 
causes and corrective actions, ranked by number from most likely cause/least invasive action 
to least likely cause/most invasive action (typically, component replacement).

TABLE OF CONTENTS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 2 OF 85

Troubleshooting errors

This section outlines troubleshooting for the Alliance HPLC error states for the system and each system 
module. 

Additional troubleshooting information
For additional troubleshooting information for your specific Alliance HPLC System, see:
• Alliance 2690/2695/e2695 System Service Key Documents 
• Alliance 2790/2795/2796 System Service Key Documents

Major repairs and requalification
Major repairs require requalification. If troubleshooting results in a major repair or you are uncertain if a 
repair qualifies as major, consult the requisite qualification documentation for your specific system.

All errors/warnings index

Table 1 is an alphabetical list of all Alliance HPLC errors and warnings. Select the first letter of the error to 
jump to the relevant section (see below), or select the specific error/warning (within Table 1): 

A B C D E F G H I J K L M
N O P Q R S T U V W X Y Z

Table 1: All errors and warnings

Initial letter Error or warning

# • 410 detector fault, page 58
• 410/2410 detector fault, page 64
• 486 detector fault, page 59

A • A failure occurred in attempting to download an instrument method [Empower error], page 7
• Accumulator over pressure, page 16
• Auto Shutdown Invoked, page 7

B • Bath speed tolerance, page 60
• Bath temp tolerance, page 60
• Bubble found on compression, page 16
• Bubble found on intake, page 17

TROUBLESHOOTING ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 3 OF 85

Table 1: All errors and warnings

Initial letter Error or warning

C • Can't understand requested well, page 64
• Cannot reach minimum pressure, page 18
• Carousel carrier jammed, page 26
• Carousel ID sensor fault, page 26
• Carousel is busy, page 27
• Carousel malfunction, page 28
• Carousel Z axis malfunction (1), page 30
• Carousel Z axis malfunction (2), page 31
• Carousel Z axis malfunction (3), page 31
• Carousel Z axis malfunction (4), page 32
• Carousel Z axis malfunction (5), page 32
• Carrier is busy, page 65
• Carrier lockpin movement failed, page 65
• Carrier malfunction, page 65
• Column heater door open, page 53
• Column heater not ambient, page 53
• Column heater running, page 53
• Column temp h/w fault (1), page 54
• Column temp h/w fault (2), page 54
• Column temp h/w fault (3), page 54
• Column temp h/w fault (4), page 54
• Column temp h/w fault (5), page 54
• Column temp h/w fault (6), page 55
• Column temp h/w fault (7), page 55
• Column temp h/w fault (8), page 55
• Column temp over limit (X°), page 55
• Compression check failed, page 34
• Compression leak test failed, page 34
• Controller Internal Error, page 7

D • Default separation method created, page 66
• Degasser ambient pressure, page 49
• Degasser high leak rate, page 49
• Degasser h/w fault (1), page 50
• Degasser h/w fault (2), page 50
• Degasser h/w fault (3), page 50
• Degasser minimum pressure, page 51
• Degasser pressure fault (X), page 51
• Diskette removed while logging, page 7
• Dissolution, page 60
• Dissolution (bad state), page 60
• Dissolution (carousel), page 61
• Dissolution (comm error), page 61

ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 4 OF 85

Table 1: All errors and warnings

Initial letter Error or warning

D • Dissolution (couldn’t continue), page 61
• Dissolution (couldn’t dispense), page 62
(continued)
• Dissolution (dispenser move), page 62
• Dissolution (TM sensor), page 63
• Dissolution (TM syringe), page 63
• Dissolution (vial scan), page 63
• Door is Closed, page 8
• Door is Open, page 8

E • Eluent Monitor Disabled, page 66
• Eluent Monitor Error, page 66
• Error 109, page 8
• Error 110, page 8
• Error changing column position, page 56
• Error writing log to disk, page 8
• Exceeded maximum plunger stroke, page 19

F • Fan failed, page 9
• Flow too high for stroke volume, page 19

G • General electronics fault, page 9
• General instrument error, page 9
• General instrument fault, page 9
• General solv delivery fault, page 20
• GPV h/w fault, page 20

H • HWERROR1, page 9

I • Initialization fault, page 9
• Inject Valve movement failure, page 66
• Injection parameters incorrect, page 66
• Injector internal error, page 20
• Instrument Failure Error, page 10
• Internal error, page 10

L • Loop too small for sample, page 35
• Lost Prime, page 21
• Low column temp (X°), page 57
• Low sample temp (X), page 35

M • Missed a dissolution sample, page 63
• Missing carousel, page 36
• Missing vial, page 36
• Must degas solvents for SysPrep, page 66

ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 5 OF 85

Table 1: All errors and warnings

Initial letter Error or warning

N • Needle is busy, page 38
• Needle malfunction, page 38
• Needle malfunction (bottom sensor), page 39
• Needle malfunction (top sensor), page 41
• No‐Flow Shutdown Invoked, page 67

O • Options not recognized, page 10
• Overpressure: check inline filter, page 67

P • Parameter range error, page 11
• pH/Cond. Monitor Disabled, page 67
• pH/Cond. Monitor Error, page 67
• Plate height error, page 68
• Plate parameters are incorrect, page 67
• Plate X axis parameters incorrect, page 68
• Plate Y axis parameters incorrect, page 68
• Plunger homing fault, page 21
• Plunger homing over pressure, page 22
• Power down, page 11
• Power up, page 11
• Primary over pressure, page 22
• Primary transducer range (X), page 23
• Procedure interrupted, page 12

R • Restriction missing, page 42
• RS‐232 Valve Comm Error, page 68
• RS‐232 Valve Error, page 69
• RS‐232 Valve Missing, page 69

S • Sample heater/cooler running, page 42
• Sample syringe over pressure (X), page 42
• Sample temp h/w fault 1, page 43
• Sample temp h/w fault 2, page 44
• Sample temp h/w fault 3, page 44
• Sample temp h/w fault 4, page 44
• Sample temp h/w fault 5, page 45
• Sample temp h/w fault 6, page 45
• Sample temp h/w fault 7, page 45
• Sample temp not ambient, page 43
• Sample temp over limit (X), page 43
• Sample too large for configuration, page 69
• Sample transducer range (X), page 46
• Seal pack geometry fault (X), page 46
• Seal wash internal error, page 23
• Seals not adjusted, page 47

ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 6 OF 85

Table 1: All errors and warnings

Initial letter Error or warning

S • Server busy, page 12
• Solv delivery h/w fault, page 24
(continued)
• Solv delivery internal error, page 24
• Solv delivery motor lost sync (X), page 25
• Sparge internal error, page 52
• StopFlow, page 12
• Stop Flow input was received, page 13
• Stop Flow key was pressed, page 13
• Stop Flow request from remote, page 13
• Syringe movement malfunction, page 47
• SYSERROR1, page 13
• SysPrep requires a degasser, page 69
• System over pressure, page 14
• System transducer range (X), page 25
• System under pressure, page 14

T • Top and bottom needle sensors active, page 70

U • User‐programmed alert, page 15

V • Vial internal error, page 48
• Vial sensor test failed, page 48

W • Well depth error, page 70
• Wrong carousel in position, page 48

X • X axis malfunction, page 71

Y • Y axis malfunction, page 72

ALL ERRORS/WARNINGS INDEX
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 7 OF 85

System level errors

This section lists the system level errors and warnings in alphabetical order. Locate the relevant error or 
warning in the left column and follow the appropriate troubleshooting steps. 

Table 2:  System level errors or warnings

Error or warning Possible cause Corrective action

A failure occurred in attempting to  Firmware problem  Update the firmware. 


download an instrument method  (This may occur when controlling  1. Go to: e2695 Separations 
[Empower error] the 2695 with Ethernet.) Module Firmware Revision 3.02 
(667004916)
(eAlliance system error; message  Note:  This problem is corrected in  2. Read the Release Notes.
displayed at the Empower Message  eAlliance firmware version 
Center) 3. Install the firmware.
3.02.
Note:  This error only affects  4. Relaunch the Run Samples 
systems under Ethernet  application. 
control. IEEE‐controlled  Note:  If necessary, reboot the 
Alliance systems are not  node to reestablish 
affected. communication with the 
instrument.

Auto Shutdown Invoked The Auto Shutdown feature has  No action required. This is normal/


applied the initial conditions of a  expected behavior.
separation method after a set  To deactivate, click Clear.
amount of idle time Note:  You can adjust the settings 
for separation method and 
idle time from the Auto 
Shutdown window.

Controller Internal Error Firmware encountered an  1. Examine the trace log to 


unexpected condition determine instrument activity at 
This is a generic message.  the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Diskette removed while logging Floppy disk removed during a data  Reinsert the floppy disk. 


save Note:  Remove the disk only when 
The data being saved to the disk is  the instrument is in an idle 
lost. state.

Dust inside the floppy drive Clean the inside of the drive using 
compressed air. 

Disk‐drive problem Replace the drive.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 8 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

Door is Closed Door is closed when it should be  1. Open the door. 


open 2. Resume or restart the diagnostic.
The sample compartment or col‐
umn heater door is closed, but the  Door switch problem 1. Inspect the switch and the cable 
currently running diagnostic  that connects it to the PCB.
requires that it be open. 2. Replace if necessary.

Door is Open Door is open when it should be  1. Close the door. 


closed 2. Resume or restart the diagnostic.
The sample compartment or col‐
umn heater door is open, but the  Door switch problem 1. Inspect the switch and the cable 
currently running diagnostic  that connects it to the PCB.
requires that it be closed. 2. Replace if necessary.

Error 109 An internal system error has  1. Examine the trace log to 


occurred determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Error 110 An internal system error has  1. Examine the trace log to 


occurred determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Error writing log to disk Missing disk Ensure that there is a floppy disk in 


the drive. 
The system tried to write data to 
the disk, but failed. Data may be  Incorrectly formatted disk Format the disk.
lost. Write‐protected disk Ensure that the floppy disk is not 
write‐protected. 

Insufficient disk capacity Ensure that the floppy disk has 
enough free space to store the data 
being written. 
Note:  It takes a very long time to 
fill a floppy disk using the 
"Report to floppy disk" 
feature. Replace the disk if it 
has been in use for a long 
time. 

Ensure that the disk contains no 
more than 256 files.

Corrupted disk Replace the disk. 

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 9 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

Error writing log to disk Dust in disk drive Clean the inside of the drive using 


compressed air. 
(continued)
Disk drive problem Replace the disk drive.

Fan failed Cooling fan problem Replace the fan.

General electronics fault A general electronics problem 1. Examine the trace log to 


determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

General instrument error Firmware encountered an  1. Examine the trace log to 


unexpected condition  determine instrument activity at 
This is a generic message. the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

General instrument fault Firmware encountered an  1. Examine the trace log to 


unexpected condition  determine instrument activity at 
This is a generic message. the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

HWERROR1 A hardware problem has occurred 1. Examine the trace log to 


determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Initialization fault An internal system error has  1. Examine the trace log to 


occurred during initialization determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 10 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

Instrument Failure Error  Optional component problem Replace the optional component if 


necessary.
[Empower error message]

Occurs in e2695 Auto Addition with 
time delay under Ethernet control.
Notes:  —This error affects only 
systems under Ethernet 
control. 
  —Auto Addition with time 
delay under IEEE control and 
in stand‐alone mode are 
immune to this issue.

Internal error A problem in the valves, needle, or  1. Examine the trace log to 


syringe determine instrument activity at 
the time of failure.
2. Inspect, adjust, or replace 
affected valves, needle, or 
syringe.

A hardware problem has occurred 1. Examine the trace log to 
determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Options not recognized Failure of an optional component  1. Power‐off system and disconnect 


connection all optional components except 
one.
See service note: Options Not  2. Power‐on and look for the same 
Recognized (link opens in IBM  failure message.
Notes)
3. Repeat steps 1 and 2 for each 
optional component until you 
find the component causing the 
error.
4. Examine the J2 through J5 
positions on the CPU PCB and 
reconnect if necessary.
5. Replace cables if necessary.

Optional control cable problem 1. Inspect/Reconnect/Replace.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 11 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

Options not recognized CPU connector problem 1. Inspect the CPU connector and 


cable.
(continued) 2. Reconnect.
3. Try another connector (J2‐J5) to 
verify that the problem is not the 
CPU connector.
4. Replace.

Optional component problem Replace the optional component if 
necessary.

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

Parameter range error The data system sent a corrupt  Recreate the data system 


separation method to the  separation method to fix the 
This is a generic message indicating  instrument corruption. 
that the firmware encountered an 
unexpected condition. Too many gradient lines • Revise the method to require 16 
Note:  The maximum number of  gradient lines or fewer. 
gradient lines supported in  • Verify that the temperature and 
separation methods is 16. stroke volume values in the 
instrument method are within 
range.

Firmware encountered an  1. Examine the trace log to 
unexpected condition determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Power down System power is off No action required. 


This is normal/expected behavior.
This message is added to the error 
log whenever the instrument is 
powered‐off. It allows users to track 
when the instrument is powered‐
on and off.

Power up System power is on No action required. 


This is normal/expected behavior.
This message is added to the error 
log whenever the instrument is 
powered‐on. It allows users to track 
when the instrument is powered‐
on and off.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 12 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

Procedure interrupted A diagnostic with a countdown  No action required. 


timer (such as, sample heater/ This is normal/expected behavior. 
cooler) or other time‐critical test 
was interrupted by the user,  The procedure must be halted or 
invalidating the test results aborted before continuing.

Server busy Interactions between Empower  No action required.


and the Quickset window cause a  • This issue was resolved in 
freeze Empower 2 Feature Release 5 and 
in all subsequent releases.
• If using any version of Empower 2 
prior to FR5, avoid using the 
abort button during acquisition.
For more information, see Server 
Busy Error: Troubleshooting, page 
83

StopFlow  Firmware encountered an  1. Examine the trace log to 


unexpected condition determine instrument activity at 
The solvent management system  the time of failure. 
has been set to zero flow rate, and  2. If possible, resolve the root 
any current function has been inter‐ cause identified in the trace log.
rupted.
3. If you cannot resolve the root 
cause of the error, contact GSS 
See related:
and be prepared to describe the 
• Stop Flow input was received conditions leading to this error.
• Stop Flow key was pressed
• Stop Flow request from remote Electrical short on rear terminal  1. Disconnect the wire between the 
connector of the solvent manager  Stop Flow input terminals and 
or on the CPU test to see if the error still 
Note:  When using a MS, there is a  occurs.
wire from the rear terminal  • If the error reoccurs, the ter‐
of the pump to the MS stop  minal may be shorted.
flow terminal. During a  a. Inspect connectors for
hardware fault, the system       short and resolve.
sends a signal via this wire to  b. If connector is shorted 
stop flow to prevent      internally or cannot 
flooding the MS. A short in      otherwise be resolved, 
this wire will also cause      replace the PCB.
StopFlow. • If the error no longer occurs 
with the wire removed, 
replace the wire and retest. 
2. If you cannot resolve the root 
cause of the error, contact GSS 
and be prepared to describe the 
conditions leading to this error.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 13 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

Stop Flow input was received Accidental activation If the activation is unintentional, 


reconfigure the instrument to no 
The Stop Flow rear‐panel input was  Note:  If connected, mass  longer signal a stop flow. 
activated by an electrical signal.  spectrometers are known to  (Use the "Events In" window to 
frequently stop flow for valid  disable the Stop Flow rear‐panel 
See related: StopFlow reasons.  input.)

Wiring issue—the wrong wire may  1. Inspect the rear‐panel input 
be connected to the Stop Flow rear‐ wiring. 
panel input  2. Verify proper operation using the 
Inputs and Outputs diagnostic. 

Rear‐panel IO connector block  1. Inspect/replace. 
problem 2. Verify proper operation using the 
Inputs and Outputs diagnostic. 

Failure of electrical connections  1. Inspect/Reconnect/Replace..
between the rear‐panel IO  2. Verify proper operation using the 
connector block and the CPU PCB  Inputs and Outputs diagnostic. 
board

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814).

Stop Flow key was pressed A user pressed the front‐panel Stop  No action required.


Flow key This is normal/expected behavior.
See related: StopFlow

Stop Flow request from remote By pressing the data system’s  No action required. 


faucet icon, a user issues a stop‐ This is normal/expected behavior.
The data system (Empower, Mass‐ flow command
Lynx, etc.) sent a stop flow request 
to the instrument.

See related: StopFlow

SYSERROR1  An internal system error has  1. Examine the trace log to 


occurred determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 14 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

System over pressure  Excessive column or system  Reduce the amount of system 


pressure pressure by reducing the amount of 
System pressure exceeds the limit  restriction in the system. 
set in the current separation 
method and remains high for at  Flow rate is too high for the column  Reduce the flow rate or change the 
least five seconds. or solvents in use column or solvents. 

The separation method contains a  Change the system pressure limit in 
limit that is too low for the column,  the separation method. 
flow rate, and solvent or solvents.

Column problem 1. Remove column to reduce 
pressure and replace if 
necessary. 
2. Verify and adjust flow rate if 
necessary.
3. Clean or replace the in‐line filter.
4. Change the system pressure limit 
in the separation method if 
necessary.

Clogged in‐line filter Clean or replace the filter.
See:
• 269x: Replacing the In‐Line Filter
• 2790/2795: Replacing the In‐Line 
Filter 
• 2796: 2796 In‐Line Filter 

System under pressure Fluidics leak Inspect for fluid leakage and repair.

System pressure falls below the  Lack of solvent Ensure that there is sufficient 


limit set in the current separation  solvent. 
method and remains low for at  The minimum system pressure in  Change the system pressure limit in 
least 30 seconds. the separation method is too high  the separation method. 
for the column, flow rate, or 
solvents

Failed check valve Replace the primary inlet check 
valve. 
See:
• Check Valve Cartridge Summary
• 296x: Replacing the Check Valves 
• 2790/2795:
Replacing the Check Valves 
• 2796: Replacing the 2796 Check 
Valves

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 15 OF 85

Table 2:  System level errors or warnings (continued)

Error or warning Possible cause Corrective action

User‐programmed alert The user specifies conditions under  No action required. 


which the system issues an alert  This is normal/expected behavior.
Users can program alerts into the 
timed event tables for their separa‐
tion methods. At the given time, 
this message is displayed on the 
front‐panel.

SYSTEM LEVEL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 16 OF 85

Solvent manager errors

This section lists the system errors and warnings specific to the solvent manager in alphabetical order. 
Locate the relevant error or warning in the left column and follow the appropriate troubleshooting steps. 

Table 3:  Solvent manager errors and warnings

Error or warning Possible cause Corrective action

Accumulator over pressure  Excessive column or system  Inspect the following components 


restriction for sources of restriction and 
The pressure in the pump's accumu‐ eliminate them. 
lator head exceeded its safety limit. • in‐line filter
• vent valve
• column
• other components if necessary

Zero the accumulator pressure 
transducer by restarting the 
instrument.

The flow rate is too high for current  Reduce the flow rate or change 
column or solvent or solvents solvents. 

Inspect the cable that runs from 
the accumulator pressure 
transducer to the CPU board and 
replace.

Pressure transducer problem Replace the transducer. 
See Transducer Replacement 
For additional information, see: 
Carousel and dogpin parts diagram  Adjust the accumulator pressure 
(2695 only), page 78 transducer voltage to 0.5V ± 0.2V. 

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

Bubble found on compression  A gas bubble momentarily  Ensure that the solvent lines are 


interrupted solvent flow properly connected and are not 
The solvent management system  allowing air into the system.
detected a bubble during its com‐ • The solvent delivery system is 
pression cycle. Ensure that the solvents are 
operating at near its maximum 
properly degassed. 
• Criteria for bubble detection: pressure vs. flow envelope
•  The failure of the pre‐com‐ • High gas loading of solvents Condition the fluidics by running 
pression stroke to achieve  (inadequate solvent conditioning  the Wet Prime function with 
system pressure within 20%  within degas or sparge system) organic solvent, such as 100% 
of the working stroke. methanol. 
•  The C/D ratio is 1.8 or greater.  
(compression/decompres‐
sion)

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 17 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Bubble found on compression  Degasser vacuum chamber  1. Verify that the degasser pressure 


problem is .9 psi or less (functioning 
(continued) properly). 
2. Inspect the degasser vent lines 
for liquid. If liquid is present, 
there is a problem with the 
chamber. 
3. Replace the chamber.
See Degasser Assembly Component 
Replacement

Degasser vacuum pump problem Replace the degasser component. 
See Degasser Assembly Component 
Replacement

Plunger seal problem 1. Replace plunger seals.
2. Pressurize the seals to 4,000–
4,500 psi for 10 minutes. 

Clogged filter 1. Clean the inlet reservoir filter. 
2. Replace the filter if necessary.
See Replacing the In‐line Filter

Cable connection or problem 1. Verify that the cable connecting 
the primary pressure transducer 
to the CPU board is connected 
and operational.
2. Replace the cable if necessary. 

Pump primary pressure transducer  Replace the transducer. 
problem See Transducer Replacement 
For additional information, see: 
Carousel and dogpin parts diagram 
(2695 only), page 78

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

Bubble found on intake  A bubble in the solvent Ensure that the solvent lines are 


(degasser or sparge system) properly connected and are not 
The solvent delivery system has  allowing air into the system.
failed to detect the completion of 
the solvent decompression phase  Poor solvent conditioning, resulting  Ensure that the solvents are 
during an intake stroke. in high gas loading properly degassed. 

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 18 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Bubble found on intake  Failed degasser vacuum chamber 1. Inspect the degasser vent lines 


for liquid. If liquid is present, 
(continued) there is a problem with the 
chamber. 
2. Replace the chamber.

Degasser problem 1. Verify that the degasser pressure 
is .9 psi or less (functioning 
properly). 
2. Replace the degasser 
component. 
See Degasser Assembly Component 
Replacement

Poor fluidics conditioning Condition the fluidics by running 
the Wet Prime function with 
organic solvent, such as 100% 
methanol. 

Plunger seal problem 1. Replace plunger seals.
2. Pressurize the seals to 4,000–
4,500 psi for 10 minutes. 

Clogged filter Clean or replace the inlet reservoir 
filter. 

Cable problem Inspect/Reconnect/Replace. the 
cables that run from the pump 
pressure transducers to the CPU 
board. 

Primary pressure transducer  Replace the transducer. 
problem See Transducer Replacement
See Carousel and dogpin parts 
diagram (2695 only), page 78

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

Cannot reach minimum pressure Solvent line connection issue Ensure that the solvent lines are 


properly connected and are not 
The solvent‐management system  allowing air into the system.
generates less than the minimum 
pressure required for the static leak 
test.

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 19 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Cannot reach minimum pressure Solvent gas 1. Ensure that the solvents are 


properly degassed.
(continued) 2. Verify that the degasser pressure 
is .9 psi or less (functioning 
properly).   

Poorly conditioned fluidics Condition the fluidics by running 
the Wet Prime function with 
organic solvent, such as 100% 
methanol.

Leaky components 1. Inspect for leaks in the pin plug, 
cap, check valves, plunger seals, 
and fluid lines.
2. Repair or replace any 
components with problems.

Exceeded maximum plunger stroke  Solvent line connection issue Ensure that the solvent lines are 


properly connected and are not 
The solvent‐management system  allowing air into the system.
displaces too much volume while 
generating the required pressure.  Leaky components 1. Inspect for leaks in the pin plug, 
cap, check valves, plunger seals, 
and fluid lines.
2. Repair or replace any 
components with problems.

Flow too high for stroke volume User attempts to change either  Confirm the compatibility of flow 


flow rate or stroke volume to a  rate and stroke volume.
The requested flow rate and stroke  value incompatible with the other (See table 4) 
volume are incompatible.
For example: setting a flow rate of 
5 mL/min under a 50‐µL stroke 
volume

Table 4: Flow rates and compatible stroke volumes

Flow rate (mL/min) Stroke volume (µL)

<10 130
<3.03 100
<1.23 50
<0.53 25

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 20 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

General solv delivery fault Firmware encountered an  1. Examine the trace log to 


unexpected condition determine instrument activity at 
The solvent management system  the time of failure. 
has detected an operational error. 2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Blown 7a fuse on pump driver  Replace the 7a fuse. 
board See Pump Driver Board Fuses 

Problem with the lower  Replace the fan or heater/cooler. 
compartment cooling fan or sample 
heater/cooler fans

Power supply problem 1. Inspect the power supply.
2. Replace if necessary. 

Plunger driver board problem Replace the board. 
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

Alliance CPU PCB problem 1. Verify and reconnect the CPU to 
plunger PCB cable.
1. Replace the CPU to plunger PCB 
cable.
2. Replace plunger driver PCB.
3. Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

GPV h/w fault A short in one or more GPV  Replace the GPV. 


solenoid windings causes the  See GPV Replacement 
The Gradient Proportioning Valve  failure or malfunction of the GPV or 
(GPV) driver electronics detected a  pump driver board Replace the pump driver board. 
solenoid over‐current condition. See:  See Pump Driver Board 
Replacement
• GPV Assembly 
• Pump Driver PCB ‐ Reference 
Information 

Injector internal error A problem in the valves, needle, or  1. Examine the trace log to 


syringe has been detected determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 21 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Lost Prime Lack of solvent 1. Ensure that there is sufficient 


solvent. 
The solvent management system  2. If problem persists, inspect and 
has lost prime (run out of solvent  reconnect solvent tubing.
during operation). 
Solvents improperly prepared 1. Ensure that the solvents are 
The criteria for loss of prime is sol‐ properly degassed.
vent volume of less than 25 psi for  2. Ensure that the solvent lines 
50 cycles, or the detection of suc‐ have been wet primed. 
cessive bubbles. The system shuts 
down to prevent excessive wear of  Lack of flow rate or pressure Increase the flow rate or pressure. 
the seals and pistons. 
Failure of primary inlet check valve Replace the valve.
See:
• Check Valve Cartridge Summary
• 296x: Replacing the Check Valves 
• 2790/2795:
Replacing the Check Valves 
• 2796: Replacing the 2796 Check 
Valves

A gas bubble momentarily  Ensure that the solvent lines are 
interrupted solvent flow properly connected and are not 
Note:  The criterion for detection of  allowing air into the system. 
a bubble is the failure of the 
Ensure that the solvents are 
pre‐compression stroke to 
properly degassed. 
achieve system pressure 
within 20% of the working  Condition the fluidics by running 
stroke.  the Wet Prime function with 
• The solvent delivery system is  organic solvent, such as 100% 
operating at near its maximum  methanol. 
pressure vs. flow envelope
• High gas loading of solvents 
(inadequate solvent conditioning 
within degas or sparge system)

Plunger homing fault Loose wires Reconnect or tighten the wires to 


the pump motor driver position 
The solvent‐delivery system fails to  encoder. 
detect home position during 
plunger registration. Problem with position encoder in  Replace the encoder. 
the pump motor driver

Pump motor problem Replace the motor. 

Plunger driver board problem Replace the board. 
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 22 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Plunger homing over pressure The pin plug restriction has not  1. Verify that the pin plug 


been removed restriction has been removed.
The solvent delivery system detects  2. Remove the pin plug restriction if 
excessive pressure while the pump  necessary.
plungers move toward their home 
registration positions.  Clogged in‐line filter Clean or replace the filter.
See:
• 269x: Replacing the In‐Line Filter
• 2790/2795: Replacing the In‐Line 
Filter 
• 2796: 2796 In‐Line Filter 

Waste‐line blockage Clear all blockages. 

Excessive restriction between the  Reduce the restriction. 
pump and sample loop

Vent valve problem 1. Ensure that the sample loop can 
vent the pump to atmosphere by 
opening valves V1, V2, and V3.
2. Inspect valves.
3. Replace valves if necessary.

Primary over pressure Excessive column or system  Between the two check valves, 


restriction between the two check  inspect the following components 
The pressure in the pump's primary  valves for sources of restriction and 
head exceeds the hardware limit. eliminate them. 
• in‐line filter
• vent valve
• column
• other components if necessary

Zero the accumulator pressure 
transducer by restarting the 
instrument.

Accumulator check valve problem Replace the check valve. 
See Replacing the Check Valves

Flow rate too high for column or  Reduce the flow rate or change the 
solvent solvent. 

Cable problem Inspect/Replace the cable that runs 
from the primary pressure 
transducer to the CPU board. 

Incorrect transducer voltage Adjust the primary pressure 
transducer voltage to 0.5V ±0.2V. 
See Carousel and dogpin parts 
diagram (2695 only), page 78

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 23 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Primary over pressure Pressure transducer problem Replace the transducer. 


See Transducer Replacement
(continued) For additional information, see: 
Carousel and dogpin parts diagram 
(2695 only), page 78

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

Primary transducer range (X) A blockage increased system  1. Clear the blockage.


pressure 2. Restart the instrument. 
(where X is the faulting voltage)
Cable problem or disconnection 1. Verify that the cable that runs 
Voltage offset is too high after zero‐ from the primary pressure 
ing the primary pressure trans‐ transducer to the CPU board is 
ducer. operational and connected to 
For additional information, see:  the correct port. 
Carousel and dogpin parts diagram  2. Reconnect or replace the cable if 
(2695 only), page 78 necessary. 

Transducer voltage out of range Adjust the system pressure 
transducer voltage to 0.5V ±0.2V 
See service note: Transducers, 
Errors, and Adjusting (link opens in 
IBM Notes)

Transducer problem 1. Verify functionality of the 
primary pressure transducer.
2. Replace if necessary.
See Transducer Replacement

Seal wash internal error A problem in the plunger seal wash  1. Examine the trace log to 


system has been detected determine instrument activity at 
This is a generic message indicating  the time of failure. 
that the firmware encountered an  2. Contact GSS and be prepared to 
unexpected condition. describe the conditions leading 
to this error.

Plunger Seal Wash Pump problem 1. Inspect the single seal wash 
pump. 
2. Replace if necessary.
See Plunger Seal Wash Option 
Installation 

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 24 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Seal wash internal error Plunger driver board problem Replace the board. 


Note:  Plunger driver board = pump  See Pump Driver Board 
(continued) driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

Solv delivery h/w fault  Blown fuse 7a on pump driver  Replace the fuse. 


board See Pump Driver Board Fuses 
Failure of the pump driver board 
electronics Malfunction of a lower  Replace the fan or heater/cooler.
compartment cooling fan (without  See:
the sample heater/cooler installed)  • Sample Heater/Cooler Testing, 
or a sample heater/cooler fan page 74
• Installing the Sample Compart‐
ment Heater/Cooler Assembly 
(715002495)

Cable connection problem  Reseat the cable connecting the 
between the plunger drive PCB and  plunger driver PCB and the CPU 
the CPU PCB PCB.

Plunger driver board problem Replace the board. 
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See Alliance CPU PCB (715003814)

Solv delivery internal error  Firmware encountered an  1. Examine the trace log to 


unexpected condition determine instrument activity at 
A fault in the solvent delivery  the time of failure. 
system has been detected.  2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Incorrect cable connection Verify that the cables from the 
motors and encoders are 
connected correctly. 

Piston drive problem 1. Inspect the piston drives.
Note:  Piston drive = torpedoes. 2. Replace if necessary. 
See:
• New‐Style Piston Drive 
• Piston Drive Assembly Replace‐
ment

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 25 OF 85

Table 3:  Solvent manager errors and warnings (continued)

Error or warning Possible cause Corrective action

Solv delivery internal error  Plunger driver board problem Replace the board. 


Note:  Plunger driver board = pump  See Pump Driver Board 
(continued) driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

Solv delivery motor lost sync (X) Attempted pump operation outside  1. Inspect the pressure and flow 


the recommended pressure vs.  rate.
(where X displays as 0 or 1 flow‐rate envelope 2. If the value is outside the 
• 0 = primary  recommended value, specify a 
• 1 = accumulator) value within the envelope.
See PSI vs. Flow Envelope, page 73
The pump motor is out of sync with 
its position encoder. Plunger drive problem Replace the plunger drive if 
necessary.
See Plunger Cleaning/Replacement

Problem with cables or incorrect  1. Inspect the cables for damage.
cable connection 2. Inspect the cables for proper 
connectivity.
3. Reconnect or replace cables if 
necessary.  

Plunger driver board problem Replace the board. 
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

System transducer range (X) System blockage, yielding high  1. Clear the blockage.


pressure 2. Restart the instrument. 
(where X is the faulting voltage)
Cable problem or disconnect 1. Verify that the cable that runs 
The solvent‐management system’s  from the primary pressure 
pressure transducer was zeroed,  transducer to the CPU board is 
but the resulting voltage offset is  operational and connected to 
too high. the correct port. 
For additional information, see:  2. Reconnect/Replace. 
Carousel and dogpin parts diagram 
(2695 only), page 78 Transducer voltage out of range Adjust the system pressure 
transducer voltage to 0.5V ±0.2V 
See service note: Transducers, 
Errors and Adjusting (link opens in 
IBM Notes)

System pressure transducer  Replace the transducer.
problem See Transducer Replacement

SOLVENT MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 26 OF 85

Sample manager errors

This section lists the system errors and warnings specific to the sample manager in alphabetical order. 
Locate the relevant error or warning in the left column and follow the appropriate troubleshooting steps. 

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel carrier jammed Dusty carrier drive position sensor  1. Dampen a lint‐free cloth with 


and encoder wheels prevent carrier  only water and carefully wipe 
An attempted carousel carrier  from moving to requested carousel  the dust off the encoder wheels 
move results in inaccurate confir‐ position and sensor.
mation from the position sensor. (Compressed air is an acceptable 
alternative to the dampened 
cloth.)
2. Allow the encoder wheels and 
sensor to dry to avoid the 
attraction of dust.

Problem with cable between  1. Inspect the cable between 
carrier drive position sensor and  carrier drive position sensor and 
fluidics driver board fluidics driver board (J4). 
2. Replace if necessary.

Misaligned carrier drive position  1. Inspect position sensor 
sensor  alignment.
2. Adjust if necessary.

Sample compartment debris Remove debris or obstacles from 
sample compartment floor.

Carrier drive problem 1. Replace the carrier motor.
2. Replace the carousel carrier 
drive assembly.
See Replacing the Carousel/Carrier 
Drive (715004110)

Carousel carrier drive assembly  Replace the carousel carrier drive 
problem assembly.
See Replacing the Carousel/Carrier 
Drive (715004110)

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Carousel ID sensor fault Unclean sample compartment floor Clean sample compartment floor.

The carousel ID sensor could not 
detect the carousel ID magnet, or 
detected multiple carousel ID mag‐
nets.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 27 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel ID sensor fault Carousel height 1. Disconnect carousel ID cable 


from fluidics driver PCB.
(continued) 2. Lower carousel by two turns.
3. Reconnect carousel ID cable to 
fluidics driver PCB.

Carousel magnet missing,  Replace the carousel.
incorrectly positioned, or of 
incorrect polarity

Carousel ID sensor problem 1. Inspect the sensor and the cable 
See Carousel/Carrier Normal  that connects it to the PCB.
Operation  2. Replace if necessary.

Carousel is busy This error usually occurs because a  Wait until vial‐selection operations 


user is interacting with the front‐ are complete before interacting 
The carrier/carousel positioning  panel and is inadvertently trying to  with the front‐panel. 
mechanism is currently in use for  use one of the vial‐selection  Note:  Most vial selection 
another process. components operations take place at the 
start of injections.
A vial‐selection operation was in 
progress when another process  The carrier failed to move to the  1. Ensure that the carrier drive 
tried to use one of the necessary  requested carousel position position sensor and encoder 
components (carrier drive, carousel  wheel are free of dust. 
drive, needle drive, etc.).  2. Clean them if necessary.

1. Inspect the cable that runs from 
the carrier drive position sensor 
to the Fluidics Driver board. 
2. Replace if necessary.

Ensure that the carrier drive 
position sensor is properly aligned.

Remove any debris or obstacles 
from the floor of the sample 
compartment that may be affecting 
the carrier.

Carrier drive problem Replace the carousel carrier drive 
assembly.
See Replacing the Carousel/Carrier 
Drive (715004110)

Carousel carrier drive assembly  Replace the carousel carrier drive 
problem assembly.
See Replacing the Carousel/Carrier 
Drive (715004110)

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 28 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel malfunction Symptoms
• Attempted carousel move yields inaccurate confirmation from the 
position sensor. 
• Carousel drive encoder wheel unmapped during start‐up.
• Carousel moves, but the position sensor is inactive at the end of the 
move. 
Note:  Carousel malfunction vs. carousel z axis malfunction
• Carousel malfunction: The carousel fails to find home, fails to map, 
or fails to locate the encoder‐wheel slot positions.
• Carousel Z‐axis malfunction (X): The Z‐ram fails to engage the 
carousel for one of five reasons.

Dusty encoder wheels or sensors 1. Dampen a lint‐free cloth with 
only water and carefully wipe 
the dust off the carousel and 
carriage encoder wheels and 
sensor. (Compressed air is an 
acceptable alternative to the 
dampened cloth). 
2. Allow the encoder wheels and 
sensor to dry to avoid the 
attraction of dust.

Dirty sample‐compartment floor 1. Clean floor.
2. Verify that the carousels move 
easily. 

Carousel’s bottom side is rough,  1. Clean carousel bottom.
causing drag 2. Carefully polish the bottom with 
800‐grit sand paper.
3. Replace carousel.

Needle pierce is outside of the  Adjust the carrier position. 
center of septa See Adjusting the Z‐Axis Ram 
Height and the Carousel Carrier 
Position (715002039)

Misaligned vial sensor impedes the  Push back the vial sensor tube 
carousels (located on the right‐hand side).
Note:  The tip of the sensor must 
The result is sporadic carousel  extend no more than 0.1 
malfunctions or z axis 5 errors  inch from the surface of the 
during, or shortly after, installation foam compartment.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 29 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel malfunction The carousel under the needle has  1. Recalibrate the carrier position.


been moved since the last injection 2. Clean the sample compartment 
(continued) Note:  This occurs only when a user  floor.
reaches into the sample  3. Try another carousel if necessary.
compartment and moves 
the carousel under the  4. Ensure that the carousel under 
needle, or moves the carrier  the needle is unmoved between 
enough to move the  injections.
carousel indirectly. 

Carousel is not engaged by dogpin;  1. Inspect the dogpin slot on the 
as a result, carrier improperly  carousel for damage. 
positions carousel under drive shaft 2. If damaged, replace the carousel.

Carousel drive problem Replace the carousel drive. 
See Replacing the Carousel/Carrier 
Drive (715004110)

Incorrect Z ram height 1. Verify the height of the Z‐ram 
using the Z‐ram height gauge 
The position of the carousel drive  (700001022).
shaft may be too far up or down  2. Adjust position of sensors on the 
when it engages a carousel drive shaft as necessary.
See Adjusting the Z‐Axis Ram 
Height and the Carousel Carrier 
Position (715002039)

The carrier plate may be out of  Adjust the position sensors on the 
alignment and putting too much  carrier drive.
lateral force on the carousel, 
preventing it from rotating

Firmware encountered an  1. Examine all carousel‐positioning 
unexpected condition components and take action as 
necessary.
2. If still unresolved, examine the 
trace log to determine 
instrument activity at the time of 
failure. 
3. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 30 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel Z axis malfunction (1) Sensors disconnected from fluidics  Inspect cables and reconnect if 


driver board (home or dogpin) necessary.
The z axis move was attempted but 
failed because both the dogpin and  Dusty sensors (home or dogpin) Clean the sensor.
home sensors are on. Sensor problem (home or dogpin) Replace the sensor if necessary.
Note:  This error applies only to 
Alliance 2690 and 2695. Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Z axis motor problem 1. Ensure that the z axis motor is 
connected to the fluidics driver 
board.
2. Ensure that the motor is moving 
the z axis.
3. Replace the z axis motor if 
necessary.

Dogpin misaligned Carousel carrier drive has a blunt‐
cut dogpin that has difficulty 
See Carousel and dogpin parts  dropping into a carousel slot.
diagram (2695 only), page 78 1. Ensure that the dogpin is seated 
properly within notches.

Height of carousel carrier drive in  Adjust height (gauge block 
compartment 700001022). 
See Adjusting the Z‐Axis Ram 
Height and the Carousel Carrier 
Position (715002039)

Black, plastic, optical fence hits  Note:  The bracket may be out of 
sensor or does not position into  tolerance. The small change 
sensor properly in tolerance is amplified 
onto the optical fence 
sensor.
The sensor is located at the bottom 
of the metal bracket.
1. Monitor the drive while it is in 
motion to determine if the 
optical fence’s downward 
movement causes the fence to 
rub against the inside wall of the 
sensor.
2. Adjust the bracket adjustment 
screw until alignment is 
achieved, if necessary.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 31 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel Z axis malfunction (1) Carousel carrier drive shaft doesn't  1. Verify by pushing up on dogpin 


move freely area.
(continued) 2. Replace carousel carrier drive if 
necessary. 

Loose encoder wheel due to  1. Ensure that the encoder staking 
improper staking is within tolerance.
2. If encoder wheel is too loose, 
replace carousel carrier drive.

Carousel misaligned  1. Inspect carousel.
2. Align carousel if necessary.

Debris Verify that the floor is clean and the 
carrier plate moves correctly. 

Misaligned vial sensor impedes  Push back the vial sensor tube 
carousels (located on the right‐hand side).
Note:  The tip of the sensor must 
extend no more than 0.1 
inch from the surface of the 
foam compartment.

Carousel Z axis malfunction (2) Home sensor disconnected from  Inspect cable and reconnect if 


fluidics driver board  necessary.
The z axis failed to home to top 
switch.  Dusty home sensor Clean the sensor.
Note:  This error applies only to  Home sensor problem Replace the carousel carrier drive if 
Alliance 2690 and 2695. necessary.

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Z axis motor problem 1. Ensure that the z axis motor is 
connected to the fluidics driver 
board.
2. Ensure that the motor is moving 
the z axis.
3. Replace the z axis motor if 
necessary.

Carousel Z axis malfunction (3) Home sensor disconnected from  Inspect cable and reconnect if 


fluidics driver board  necessary.
The z axis was moved up, but failed 
because the home sensor did not  Dusty home sensor Clean the sensor.
activate.  Home sensor problem Replace the sensor.
Note:  This error applies only to 
Alliance 2690 and 2695.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 32 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel Z axis malfunction (3) Fluidics driver board problem Replace the fluidics board. 


See Fluidics Driver PCB
(continued)
Z axis motor problem 1. Ensure that the z axis motor is 
connected to the fluidics driver 
board.
2. Ensure that the motor is moving 
the z axis.
3. Replace the z axis motor if 
necessary.

Carousel Z axis malfunction (4) Sensors disconnected from fluidics  Inspect cables and reconnect if 


driver board (home or dogpin) necessary.
The z axis was moved down, but 
failed because the home sensor  Dusty sensors (home or dogpin) Clean the sensor.
was active.  Sensor problem (home or dogpin) Replace the sensor if necessary.
Note:  This error applies only to 
Alliance 2690 and 2695. Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Z axis motor problem 1. Ensure that the z axis motor is 
connected to the fluidics driver 
board.
2. Ensure that the motor is moving 
the z axis.
3. Replace the z axis motor if 
necessary.

Carousel Z axis malfunction (5)  Sensors disconnected from fluidics  Inspect cables and reconnect if 


driver board (home or dogpin) necessary.
The dogpin is off and the z axis is 
down. Dusty sensors (home or dogpin) Clean the sensor.
Note:  This error applies only to  Dogpin sensor problem  Replace carousel carrier drive if 
Alliance 2690 and 2695. necessary.

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Z axis motor problem 1. Ensure that the z axis motor is 
connected to the fluidics driver 
board.
2. Ensure that the motor is moving 
the z axis.
3. Replace the z axis motor if 
necessary.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 33 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel Z axis malfunction (5)  Dogpin misaligned Carousel carrier drive has a blunt‐


cut dogpin that has difficulty 
(continued) See Carousel and dogpin parts  dropping into a carousel slot.
diagram (2695 only), page 78 1. Ensure that the dogpin is seated 
properly within notches.

Height of carousel carrier drive in  Adjust height using gauge block, 
compartment 700001022. 
See Adjusting the Z‐Axis Ram 
Height and the Carousel Carrier 
Position (715002039)

Black, plastic, optical fence hits  Note:  The bracket may be out of 
sensor or does not position into  tolerance. The small change 
sensor properly in tolerance is amplified 
onto the optical fence 
sensor.
The sensor is located at the bottom 
of the metal bracket.
1. Monitor the drive while it is in 
motion to determine if the 
optical fence’s downward 
movement causes the fence to 
rub against the inside wall of the 
sensor.
2. If necessary, adjust the bracket 
adjustment screw until 
alignment is achieved.

Carousel carrier drive shaft doesn't  1. Verify by pushing up on dogpin 
move freely area.
2. Replace carousel carrier drive if 
necessary. 

Loose encoder wheel due to  1. Ensure that the encoder staking 
improper staking is within tolerance.
2. If encoder wheel is too loose, 
replace carousel carrier drive.

Carousel misaligned  1. Inspect carousel.
2. Align carousel if necessary.

Debris Verify that the floor is clean and the 
carrier plate moves correctly. 

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 34 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Carousel Z axis malfunction (5)  Misaligned vial sensor impedes  Push back the vial sensor tube 


carousels (located on the right‐hand side).
(continued) Note:  The tip of the sensor must 
extend no more than 0.1 
inch from the surface of the 
foam compartment.

Compression check failed Solvent‐line connection Ensure that the solvent lines are 


properly connected and are not 
The Purge function was run with  allowing air into the system.
the "Compression check” feature 
enabled, but Compression pressure  Poor solvent conditioning 1. Confirm the proper degassing of 
failed to achieve minimum within  solvents.
the configured compressible  2. Verify the absence of bubbles in 
volume. the syringe. 

Seal problems 1. Inspect for seal leaks and resolve.
2. Rerun Adjust seal pack.
3. Replace/rebuild seals as 
necessary.
See:
• Installing an Alliance injector seal 
pack
• Seal Pack Replacement/Rebuild 
Kits

Syringe leak 1. Inspect the syringe for presence 
of bubbles. 
2. Replace the syringe if necessary.

Valve leak 1. Run automated valve leak test.
2. Inspect and if necessary, replace 
valves V1, V2, or V3. 

Compression leak test failed Solvent‐line connection 1. Ensure that the solvent lines are: 


• properly connected
The system fails to maintain the 
• not allowing air into the 
pressure level required during the 
system
compression check of the surge 
• primed
function. 
2. Ensure that the degasser is 
working.

Poor solvent conditioning 1. Confirm the proper degassing of 
solvents.
2. Confirm the absence of bubbles 
in the syringe. 

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 35 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Compression leak test failed Seal problems 1. Inspect for seal leaks and resolve.


2. Rerun Adjust seal pack.
(continued)
3. Replace/Rebuild seals as 
necessary.
See:
• Installing an Alliance injector seal 
pack
• Seal Pack Replacement/Rebuild 
Kits

Syringe leak 1. Inspect syringe for presence of 
bubbles. 
2. Replace if necessary.

Valve leak 1. Inspect all valves.
2. Run automated valve test.
3. Replace valve if necessary.

Loop too small for sample Attempted injection with a sample  1. Go to Configuration.


volume larger than the installed  2. Ensure that the sample loop 
sample loop volume volume matches the volume of 
the installed sample loop, and 
that it is large enough for the 
requested sample volume. 

Low sample temp (X) Insufficient equilibration time Confirm that there is enough 


equilibration time between sample 
(Where X is the faulting tempera‐ temperature changes to allow the 
ture) temperature to reach the set 
point. 
The sample temperature falls 
below the minimum specified (X) in  Frequent door openings Ensure that the sample 
the current separation method. compartment door is not opened 
frequently. 
For testing information, see Sample  Sample heater/cooler set point too  Ensure that the temperature set 
Heater/Cooler Testing, page 74 close to ambient temperature point value is set to a minimum of 
5 °C above ambient. 

Damaged cable 1. Inspect the cable that runs from 
the sample heater/cooler 
thermistor (J9 on PCB) to the 
column heater PCB. 
2. Replace. 

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 36 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Low sample temp (X) Thermistor problem Inspect/Reconnect/Replace. 


Note:  Replacing the sample h/c 
(continued) thermistor requires 
soldering or replacing the 
engine.

Sample heater/cooler PCB problem Replace the PCB.

Sample heater/cooler or sample  Replace the heater/cooler or 
heater/cooler hardware problem hardware.
See Installing the Sample 
Compartment Heater/Cooler 
Assembly (715002495)

Missing carousel Missing carousel Ensure that there is a carousel in 


the requested position.
A new carousel was requested, but 
could not be located; the dog‐pin  Carousel missing the magnet Ensure that the carousel has not 
sensor was not activated to confirm  lost its magnet.
that a carousel was detected.  Dogpin sensor problem The dogpin sensor may be 
See: malfunctioning. Replace the dogpin 
• Carousel Initialization  sensor if necessary.
• Carousel/Carrier Normal Opera‐
tion  Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Missing vial No vial in location 1. Add missing vial.


2. Press Resume.
A reported vial was undetected 
during sequencing; in some cases, a  Top or bottom sensor misaligned Adjust the sensors as needed.
vial in the carousel is skipped.
Carefully push the sensor into the 
tube and through the curved 
pathway. The sensor stops when 
correctly positioned. 

Figure 1 ‐ Vial sensor inside curved 
pathway

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 37 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Missing vial Top or bottom sensor is out of  The top vial’s sensor optics are 


alignment tilted out of position, causing the 
(continued) sensor to recognize as present a 
nonexistent consecutive vial.

For example: when there are 10 
vials in a tray and you remove vials 
8 and 9, vial 8 is recognized as 
missing, but vial 9 is recognized as 
present.

Perform this Random Vial Position 
test.
See Performing the Random Vial 
Position Test, page 75

Vial transparency Ensure that the vial caps are 
Note:  If a vial is actually present,  present and are opaque enough for 
the error indication could be  the vial sensor to detect them.
due to the transparency of 
1. Inspect the vial sensor for dust.
the vial cap.
2. Clean if necessary.

Ensure that the vial sensor emitter 
and receiver are aligned properly.

Ensure that there is nothing 
blocking the vial sensor.

Cable problem 1. Inspect the fiber optics cable 
that connects the vial sensor to 
the fluidics driver board. 
2. Replace the cable if necessary.

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 38 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Missing vial A user added vials to the carousel  Correctly add vials while the 


while the instrument was  instrument is processing samples:
(continued) processing samples • Wait at least 30 seconds after an 
injection is made (for example, 
wait until the run time displayed 
on the front panel, or within the 
controlling software, is greater 
than 30 seconds).
• Minimize the time that the sam‐
ple compartment door is open.
• Ensure that the door is closed 
during the time injection com‐
mands (selecting vial, washing 
the needle, drawing sample, and 
so forth) are in the process of 
being carried out.

Needle is busy A user interacting with the front  Wait until all needle‐movement 


panel inadvertently activates a  operations are complete before 
While a needle‐movement opera‐ needle‐movement component interacting with the front panel. 
tion is in process, another process  Note:  Most needle movement 
requests a needle‐movement com‐ operations take place at the 
ponent (carrier drive, carousel  start of injections.
drive, needle drive, etc.). 
Carrier/carousel positioning  See: 
mechanism problem • Alliance 269x: Carousel Rotation 
Adjustment Procedure 
• Alliance 279x: Adjusting the XY 
Mechanism in 2790/2795/2796 
Separations Modules

Needle mechanism problem Replace the needle mechanism.

Needle malfunction Bent, dirty, or sticky needle 1. Verify that the needle is primed 


and working.
This error message indicates resis‐ 2. Replace the needle if necessary.
tance to needle movement.
Seal pack problem Replace the seal pack.
See also: Needle malfunction  See Seal Pack Information See:
bottom sensor (below) and the  • Installing an Alliance injector seal 
cause, Roll Pin is loose or falls out. pack
• Seal Pack Replacement/Rebuild 
Kits

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 39 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Needle malfunction Vial/needle misalignment 1. Ensure that the vial is aligned 


with the needle. 
(continued) 2. If the needle is misaligned, adjust 
the position of the carrier and 
carousel drives. (269x only)
See Carousel Rotation Adjustment 
Procedure 

Incorrectly installed needle  Reinstall the needle. 

Needlewash issue 1. Verify that the needlewash is 
using appropriate solvents.
2. Verify that the needlewash is 
primed.

Firmware encountered an  1. Examine the trace log to 
unexpected condition determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Needle malfunction (bottom  (279x only) 1. Determine the source of the 


sensor) Problem with sensor, cable, or  problem using the procedure, 
fluidics driver PCB  Troubleshooting the XYZ and 
An attempted needle move yielded  needle bottom sensor errors 
no confirmation from the bottom  (279x only), page 85. 
sensor.  When performing the 
procedure, test the combination: 
Needle bottom sensor with 
corresponding LED DS21 on the 
fluidics driver board.
2. Use the information obtained 
from that procedure to 
determine subsequent 
troubleshooting steps.

Needle bottom sensor is dirty  Clean sensor with compressed air.

Problem with needle bottom  1. Power‐off and manually move 
sensor the flag in between the sensors 
to reinitialize.
2. Replace the sensor.
3. Replace the injector.

Incorrectly installed needle Reinstall the needle.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 40 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Needle malfunction (bottom  Seal pack problem Replace the seal pack. 


sensor) See Seal Pack Information See:
• Installing an Alliance injector seal 
(continued) pack
• Seal Pack Replacement/Rebuild 
Kits

Vial/needle misalignment 1. Ensure that the vial is aligned 
with the needle. 
2. If the needle is misaligned, adjust 
the position of the carrier and 
carousel drives. (269x only)
See Carousel Rotation Adjustment 
Procedure 

Needle motor problem 1. Confirm that the needle motor is 
connected to the PCB and the 
needle moves as intended. 
2. If not, replace the motor.

Roll Pin is loose or falls out when: Replace the injector.
• the needle is in the vial position
• the needle is in a position within 
the sensors

Roll pin

Motor
screw
shaft

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 41 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Needle malfunction (top sensor) (279x only) 1. Determine the source of the 


Problem with sensor, cable, or  problem using the procedure, 
fluidics driver PCB  Troubleshooting the XYZ and 
needle bottom sensor errors 
(279x only), page 85. 
When performing the 
procedure, test the combination: 
Needle top sensor with 
corresponding LED DS22 on the 
fluidics driver board.
2. Use the information obtained 
from that procedure to 
determine subsequent 
troubleshooting steps.

Dirty needle top sensor Clean sensor with compressed air.

Problem with needle top sensor 1. Power‐off and manually move 
the flag in between the sensors 
to reinitialize.
2. Replace the sensor.
3. Replace the injector.

Incorrectly installed needle Reinstall the needle.

Seal pack problem Replace the seal pack.
See Seal Pack Information See:
• Installing an Alliance injector seal 
pack
• Seal Pack Replacement/Rebuild 
Kits

Vial/needle misalignment 1. Ensure that the vial is aligned 
with the needle. 
2. If the needle is misaligned, adjust 
the position of the carrier and 
carousel drives.
See Carousel Rotation Adjustment 
Procedure 

Needle motor problem 1. Confirm that the needle motor is 
connected to the PCB.
2. Confirm that the needle motor is 
moving the needle as intended.
3. Replace the needle motor if it is 
not operating as required.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 42 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Restriction missing Missing restrictor Install a restrictor.

During seal pack calibration, the  Air in the system 1. Ensure that there is sufficient 


system detects no restriction. The  solvent and pump is working.
current operation, however,  2. Ensure that the solvent lines are 
requires a restrictor to generate  properly connected and are not 
extra pressure. allowing air into the system.
3. Ensure that the solvents are 
properly degassed.

Fluidics leak 1. Inspect for fluid leakage and 
repair.
2. Adjust the needle seal pack.

Sample heater/cooler running Insufficient power to the column  1. Power‐off the sample heater.


heater/cooler 2. Start the column heater/cooler 
The sample heater is running, con‐ Note:  The column heater (or CHC)  or CHC diagnostic.
suming power required by the col‐ diagnostic requires that the 
umn heater/cooler.  column heater (or CHC) be 
Note:  This error occurs during  running at full power. The 
column heater/cooler  sample heater/cooler shares 
operations. a power supply with the 
See related: Column heater  column heater (or CHC). 
running, page 53 Neither can run at full power 
when both are in use. 
Note:  Power sharing may 
adversely affect test results.

Sample syringe over pressure (X) Waste‐line blockage Clear the waste line of any 


blockages.
(where X is the faulting pressure)
Syringe or waste valve problem 1. Inspect the syringe and valves.
Pressure over 350 psi was detected  2. Replace if necessary.
in the sample loop with the syringe  See Performing the Valve Leak Test
valve open.
Syringe problem 1. Inspect the syringe.
2. Replace if necessary.

Incorrect seal position Adjust the seal position.
See Adjusting the Seals

Valve V3 problem Replace the valve. 

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 43 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Sample temp not ambient Chamber is too warm or too cool 1. Cool or warm the compartment 


by opening the sample heater/
The sample chamber is outside the  cooler door until the chamber 
18 °C to 28 °C temperature range,  equilibrates to room 
which is required to begin the  temperature. 
heater/cooler diagnostic. 2. Close the door before repeating 
the diagnostic.
See Sample Heater/Cooler Testing, 
page 74

Adjust the sample chamber set 
point to 23 °C.

Thermistor connection problem  1. Verify connection to PCB.
with PCB 2. Replace if necessary. 

Thermistor problem Inspect/Reconnect/Replace.

Sample temp over limit (X) Insufficient equilibration time  Allow enough time between 


sample temperature changes for 
The sample temperature exceeds  the temperature to adjust to set 
the limit specified in the current  point. 
separation method (where X is the 
faulting temperature). Ambient operating temperature  Ensure that the ambient 
limits exceeded temperature is within the 
supported range.

Thermistor problem Inspect/Reconnect/Replace. 

Sample heater/cooler PCB problem  1. Test the PCB.
See Sample Heater/Cooler Testing, 
page 74
2. Replace if necessary.

Sample temp h/w fault 1 Cable connection problem Confirm proper cable connection 


between the sample heater/cooler 
Communications failure between  and the CPU. 
the instrument and the sample 
heater/cooler. PCB problem Replace the sample heater/cooler 
PCB. 

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See technical reference sheet, 
Alliance CPU PCB (715003814)

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 44 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Sample temp h/w fault 2 Cable connection problem Confirm proper cable connection 


between the sample heater/cooler 
The sample heater/cooler electron‐ and the CPU. 
ics signal a hardware watchdog 
time‐out. PCB problem Replace the sample heater/cooler 
PCB. 

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.
See Alliance CPU PCB (715003814)

Sample temp h/w fault 3 External fan problem If an Alliance instrument or Sample 


Heater/Cooler sub‐assembly 
Sample heater/cooler electronics  For additional information, see  (WAT270804 or WAT270802) fail by 
signal an over‐temperature condi‐ service note: exhibiting a “Sample Temp H/W 
tion. Alliance: Sample Temp H/W Fault  Fault #3 or #5 error”, both external 
#3, and/or #5 error caused by  fans should be replaced.
faulty external fans (link opens in  Note:  Replacement fans from 
IBM Notes) Sanyo‐Denki (S‐D) are not 
subject to this failure mode.

Sample heater/cooler thermistor  3. Inspect/Reconnect/Replace. 
problem

Fan problem 1. Inspect the circulating fan and 
the external fan. 
2. Replace if necessary.

PCB problem Replace the sample heater/cooler 
PCB. 

Sample temp h/w fault 4 Internal fan problem 1. Inspect the fan. 


2. Replace if necessary.
Sample heater/cooler electronics 
have signaled that the internal fan  PCB problem Replace the sample heater/cooler 
has failed. PCB. 

Sample heater/cooler engine  Replace the heater/cooler.
problem See Installing the Sample 
Compartment Heater/Cooler 
Assembly (715002495)

Plunger driver board problem Replace the board. 
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 45 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Sample temp h/w fault 5 External fan problem If an Alliance instrument or Sample 


Heater/Cooler sub‐assembly 
Sample heater/cooler electronics  For additional information, see  (WAT270804 or WAT270802) fail by 
signal external fan failure. service note: exhibiting a “Sample Temp h/w 
Alliance: Sample Temp H/W Fault  fault #3 or #5 error”, replace both 
#3, and/or #5 error caused by  external fans.
faulty external fans (link opens in  Note:  Replacement fans from 
IBM Notes) Sanyo‐Denki (S‐D) are not 
subject to this failure mode.

Sample heater/cooler PCB problem Replace the PCB.

Sample heater/cooler engine  Replace the heater/cooler.
problem See Installing the Sample 
Compartment Heater/Cooler 
Assembly (715002495)

Plunger driver board problem Replace the board. 
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information

Sample temp h/w fault 6 Sample heater/cooler thermistor  Inspect/Reconnect/Replace. 


problem
Sample heater/cooler electronics 
signal a thermistor‐high condition.  Sample heater/cooler PCB problem Replace the PCB.

Sample heater/cooler engine  Replace the heater/cooler.
The value returned by the thermis‐ problem See Installing the Sample 
tor is not recognized as valid. Compartment Heater/Cooler 
Assembly (715002495)

Sample temp h/w fault 7 Sample heater/cooler thermistor  Inspect/Reconnect/Replace. 


problem
Sample heater/cooler electronics 
signal a thermistor‐low condition.  Sample heater/cooler PCB problem Replace the PCB.

Sample heater/cooler engine  Replace the heater/cooler.
The value returned by the thermis‐ problem See Installing the Sample 
tor is not recognized as valid. Compartment Heater/Cooler 
Assembly (715002495)

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 46 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Sample transducer range (X) System blockage 1. Clear the blockage.


2. Restart the instrument. 
(where X is the faulting voltage)
Cable connection issue 1. Inspect and reconnect the cable 
Indicates an excessive offset in the  that runs from the sample 
sample loop pressure transducer. pressure transducer to the CPU 
board. 
(When the sample pressure trans‐ 2. Replace if necessary.
ducer is zeroed, the resulting volt‐
age offset is too high.) Sample pressure transducer out of  Adjust the sample pressure 
specification transducer voltage to 0.5 ± 0.2 V.
For additional information, see: 
Transducer problem Replace the transducer.
Carousel and dogpin parts diagram 
See Transducer Replacement
(2695 only), page 78

Seal pack geometry fault (X) Seal pack problem 1. Adjust the sensor bracket to 


See Seal Pack Information 5/16 inch.
(where X is a value from 1 to 4 2. Raise or lower the bracket as 
necessary.
Seal pack geometry is based on the 
3. Move flag between sensors and 
needle moving up/down the bot‐
reboot 2695.
tom seal and sensing pressure to 
find Top, Bottom, and Width of  Note:  This resolves the problem in 
seal. most cases.
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Air in the system 1. Ensure that the solvent lines are 
Seal pack geometry fault (1) properly connected and are not 
allowing air into the system.
The seal pack top edge, as calcu‐
lated by the Adjust Seals diagnostic,  2. Remove any air in the system.
is too high. Fluidics leak Inspect for fluid leakage and repair.
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Seal pack geometry fault (2)

The seal pack bottom edge, as cal‐
culated by the Adjust Seals diagnos‐
tic, is too low.
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Seal pack geometry fault (3)

The seal pack, as calculated by the 
Adjust Seals diagnostic, is too wide.
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Seal pack geometry fault (4)

The seal pack, as calculated by the 
Adjust Seals diagnostic, is too nar‐
row.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 47 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Seal pack geometry fault (X) Leaky seal pack 1. Inspect seal packs.


2. Repair or replace.
(continued)
See:
• Rebuilding the Seal Pack
• Seal Pack Replacement/Rebuild 
Kits
• Installing an Alliance Injector Seal 
Pack

Damaged needle or incorrectly  1. Replace or rebuild the seal pack.
installed needle See:
• Rebuilding the Seal Pack
(This is a known issue with the flag/
stop on the injector and with the  • Seal Pack Replacement/Rebuild 
injector tester used in  Kits
manufacturing, the needle may  • Seal Pack Rebuild Kit w/Needle 
have been stressed enough during  • Installing an Alliance Injector Seal 
manufacturing testing to cause it to  Pack
bow or bend.) 2. Replace the needle. 
(The internal seals are the same 
as the original seal pack.)

Injector valve problem 1. Run the automated valve test.
2. Run compression test.
3. Repair or replace valves if 
necessary.
See Rebuilding the Injector Valves 

Seals not adjusted The Adjust Seals diagnostic must  Run the Adjust Seals diagnostic, 


be run before making injections Main > Diag > Adjust Seals.
The sample‐management system  See Adjusting the Seals
requires optimization values for the 
needle positioner. Problem with battery not retaining  Replace battery.
memory

Problem with sample transducer 1. Inspect sample transducer.
2. Correct any issues discovered or 
replace the transducer.

Syringe movement malfunction Syringe home sensor problem 1. Inspect and reconnect.


2. Ensure that it is free of dust.
During an attempted syringe move, 
3. Replace.
the home sensor fails to activate 
and provides no confirmation. Syringe drive problem Replace the syringe drive.
See Syringe Drive Assembly
Fluidics driver board problem Replace the fluidics board. 
See:
Alliance 269x: Fluidics Driver PCB
Alliance 279x: Fluidics Driver PCB

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 48 OF 85

Table 5:  Sample manager errors and warnings

Error or warning Possible cause Corrective action

Vial internal error Problem with connections to the  1. Inspect the fiber optics 


vial sensor connections. 
The vial/carousel positioning  2. Reconnect.
system detects a fault. 
3. Replace.

Fluidics driver board problem Replace the fluidics board. 
See:
Alliance 269x: Fluidics Driver PCB
Alliance 279x: Fluidics Driver PCB

Firmware encountered an  1. Examine the trace log to 
unexpected condition determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Vial sensor test failed Dusty vial sensor Use compressed air to clean the 


fiber optics cables (vial sensor).
During instrument start‐up, the vial 
sensor is tested by rotating the car‐ Misalignment of sensor emitter  1. Inspect the alignment of the vial 
rier plate 180° and monitoring the  and receiver sensor emitter and receiver.
state of the sensor. This message  See Vial Present Sensor 
displays when the sensor state dis‐ Maintenance 
plays no change while the carrier  2. Adjust as necessary.
plate is rotating.
See: 
• Vial Present Sensor
• Vial Present Sensor Adjustment 

Blocked sensor Remove obstruction. 
See Vial Present Sensor 
Maintenance 

Wrong carousel in position A carousel was placed in the wrong  Ensure that all carousels are in the 


location on the carrier plate correct locations on the carrier 
The system detects a misplaced car‐ plate.
ousel. 
Carousel ID magnet is missing,  Replace the carousel. 
Note:  The five carousels (A, B, C, D, 
incorrectly positioned, or of the  See Replacing the Carousel/Carrier 
E) are keyed so the system 
wrong polarity Drive (715004110)
can ensure that they are 
placed in the correct  Carousel ID sensor problem 1. Inspect the sensor and the cable 
positions in the carrier.  that connects it to the PCB. 
To eliminate restrictions in 
2. Replace if necessary.
carousel placement, go to 
the Configuration window 
and click Main > Config, and 
clear Verify carousel 
placement.

SAMPLE MANAGER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 49 OF 85

Degasser errors

This section lists the system errors and warnings specific to the degasser in alphabetical order. Locate the 
relevant error or warning in the left column and follow the appropriate troubleshooting steps.

Note:  See also PerformancePLUS Degasser Troubleshooting

Table 6:  Degasser errors and warnings

Error or warning Possible cause Corrective action

Degasser ambient pressure Vacuum leak 1. Tighten fittings.


2. Replace loose or damaged 
The degasser pressure fails to reach  manifold tubing.
0.9 psi during the first 10 minutes 
3. Ensure that all solvent lines are 
of power‐up.
in solvent and primed.
4. Look for liquid in the degasser 
vent line. (This indicates a 
chamber problem.)
See Degasser Assembly Component 
Replacement

Pressure sensor problem 1. Inspect the pressure sensor on 
the degasser PCB.
2. Replace if necessary.

Degasser PCB problem Replace the degasser PCB.
See Degasser PCB (700002511)  See Degasser Assembly Component 
(715003571) Replacement

Vent line blockage Primary symptom is a loud, high‐
pitched sound emitting from the 
degasser vacuum pump.
1. Inspect the vent tube for any 
kinks or blockages.
2. Repair or replace.

Degasser high leak rate Vacuum line or chamber leak Restart the degasser from the 


Status page.
Degasser pressure is changing too 
rapidly. 1. Inspect the degasser vacuum 
lines and chambers.
2. Replace if necessary. 
See Degasser Assembly Component 
Replacement

DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 50 OF 85

Table 6:  Degasser errors and warnings

Error or warning Possible cause Corrective action

Degasser h/w fault (1) Disconnected control cable  Ensure that the control cable 


(between the degasser and Alliance  between the degasser and Alliance 
Communications failure between  system) system is connected properly at 
instrument and degasser. each end.
(Optional control cable connects to 
CPU at J2 through J5.)

Problem with control cable  Replace the control cable 
(between the degasser and Alliance  (WAT270908).
CPU)

Degasser PCB problem 1. Verify that the dip switches on 
See Degasser PCB (715003571) the PCB are set correctly.
2. Replace the PCB if necessary.
See Degasser Assembly Component 
Replacement

Degasser h/w fault (2) Disconnected control cable  Ensure that the control cable 


(between the degasser and Alliance  between the degasser and Alliance 
Electronics have signaled a hard‐ system) system is connected properly at 
ware watchdog time‐out. each end.

Problem with control cable  Replace the control cable 
(between the degasser and Alliance  (WAT270908).
system)

Degasser vacuum pump problem 1. Inspect the degasser vacuum 
pump.
2. Replace.
See Degasser Assembly Component 
Replacement

Degasser PCB problem 1. Verify that the dip switches on 
See Degasser PCB (715003571) the PCB are set correctly.
2. Replace the PCB if necessary.
See Degasser Assembly Component 
Replacement

Degasser h/w fault (3) A vacuum leak or leaking chamber  Troubleshoot the vacuum pump.


allows liquid into the vacuum pump See Troubleshooting the Vacuum 
The electronics system signals an  Pump, page 84
over‐current condition. This error 
appears when the system fails to 
pressurize to 0.9 psi within 10 min‐
utes of power‐up. 

DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 51 OF 85

Table 6:  Degasser errors and warnings

Error or warning Possible cause Corrective action

Degasser h/w fault (3) Vent line blockage Primary symptom is a loud, high‐


pitched sound emitting from the 
(continued) degasser vacuum pump.
1. Inspect the vent tube for any 
kinks or blockages.
2. Repair or replace.

Degasser vacuum pump problem 1. Inspect the vent tube for any 
kinks or blockages.
2. Inspect the vacuum pump.
3. Replace if necessary.
See Degasser Assembly Component 
Replacement

Degasser PCB problem Replace the PCB.
See Degasser PCB (715003571) See Degasser Assembly Component 
Replacement

Degasser minimum pressure  Vacuum leaks 1. Inspect the degasser vacuum 


lines and chambers for leaks. 
The degasser vacuum pump is  2. Replace lines and chambers 
unable to generate the required  (700001218) if necessary. 
vacuum. 
Vent line blockage Primary symptom is a loud, high‐
pitched sound emitting from the 
degasser vacuum pump.
1. Inspect the vent tube for any 
kinks or blockages.
2. Repair or replace.

Vacuum pump problem Replace the vacuum pump.
See Degasser Assembly Component 
Replacement

Degasser pressure fault (X) Vacuum leaks 1. Inspect the degasser vacuum 


lines and chambers for leaks. 
(where X is the faulting pressure or  2. Replace lines and chambers 
leak rate) (700001218) if necessary. 
See Degasser Assembly Component 
The degasser is unable to maintain 
Replacement
the pressure required for reliable 
performance. Vent line blockage Primary symptom is a loud, high‐
pitched sound emitting from the 
degasser vacuum pump.
1. Inspect the vent tube for any 
kinks or blockages.
2. Repair or replace.

DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 52 OF 85

Table 6:  Degasser errors and warnings

Error or warning Possible cause Corrective action

Degasser pressure fault (X) Degasser vacuum pump problem 1. Inspect the vacuum pump.


2. Replace the vacuum pump 
(continued) (700001352) if necessary.
See Degasser Assembly Component 
Replacement

Vacuum chamber problem 1. Inspect the vacuum chamber.
2. Replace the vacuum chamber 
(700001218) if necessary.
See Degasser Assembly Component 
Replacement

Sparge internal error Firmware encountered an  1. Examine the trace log to 


unexpected condition determine instrument activity at 
This is a generic message.  the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Sparge assembly problem Replace the sparge assembly.
See Sparge Assembly 

Plunger driver board problem Replace the Plunger PCB.
Note:  Plunger driver board = pump  See Pump Driver Board 
driver board. Replacement
See Pump Driver PCB ‐ Reference 
Information 

DEGASSER ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 53 OF 85

Column device errors

This section lists the system errors and warnings specific to the column device in alphabetical order. 

Table 7:  Column device errors and warnings

Error or warning Possible cause Corrective action

Column heater door open The door is open for more than 60  Close the door. 


seconds
Proper heating requires that the 
column heater door remain closed. Door switch problem 1. Inspect and reseat the 
connectors on the PCB. 
• J1 = door sensor
• J10 = thermistor
2. Replace door switch or cable if 
necessary.

Column heater not ambient Room temperature is outside of the  1. Power‐off the column heater.


specified operating range 2. Open the column heater door 
The current diagnostic requires the  until the unit equilibrates to 
column heater to be within the  room temperature.
range of 18 °C to 28 °C (approxi‐
3. Power‐on the column heater.
mately 23 °C). 
Inadequate clearance around the  Adjust the system setup to provide 
instrument, hampering air flow adequate airflow. 
Required clearances:
• Sides: 3 inches
• Rear: 6 inches

Cable problem 1. Inspect the cable that connects 
the column heater thermistor to 
the PCB. 
2. Replace if necessary.

Thermistor problem Inspect/Reconnect/Replace. 

Column heater PCB problem Replace the PCB.

Column heater running Insufficient power to the sample  1. Power‐off the column heater.


heater/cooler 2. Start the sample heater/cooler 
The column heater (or CHC) is run‐ Note:  The sample heater/cooler  diagnostic.
ning, consuming power required by  diagnostic requires that the  3. Verify that the power supply is 
the sample heater/cooler.  sample heater/cooler be  operating normally (24V).
Note:  This error occurs during  running at full power. The 
sample heater/cooler  sample heater/cooler shares  Note:  Power sharing may 
operations. a power supply with the  adversely affect test results.
See related: Sample heater/cooler  column heater (or CHC). 
running, page 42 Neither can run at full power 
when both are in use. 

COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 54 OF 85

Table 7:  Column device errors and warnings (continued)

Error or warning Possible cause Corrective action

Column temp h/w fault (1)  Problem with interface cable  1. Ensure that the interface cable 


connecting the column heater to  connecting the column heater 
Communications failure between  the Alliance system  and the Alliance system is 
the Alliance system and the column  connected and functioning 
heater. properly.
2. Replace if necessary. 

Blown fuse on column heater PCB. Replace the fuse or the PCB. 

Blown fuse on column heater  Replace the fuse or the PCB. 
connection board (attached 
internally to column heater 
connector)

Column heater PCB problem Replace the PCB.

Column temp h/w fault (2) Column heater PCB problem Replace the PCB.

Hardware watchdog time‐out for 
the column heater.

Column temp h/w fault (3) Thermistor problem Inspect/Reconnect/Replace. 

Over‐temperature. Thermal snap switch problem 1. Inspect the snap switch.


2. Replace the PCB if necessary. 

Column heater engine problem Replace the column heater engine.

Column heater PCB problem Replace the PCB.

Column temp h/w fault (4) Column heater fan problem Replace the fan (700005104).

Fan failure. Blown fuse on column heater PCB Replace the fuse. 

Blown fuse on column heater  Replace the fuse or the PCB. 
connection board (attached 
internally to column heater 
connector) 

Column heater PCB problem Replace the PCB.

Column temp h/w fault (5) Blown fuse on column heater PCB Replace the fuse. 

Over‐current. Blown fuse on column heater  Replace the fuse or the PCB. 


connection board (attached 
internally to column heater 
connector) 

Column heater PCB problem Replace the PCB.

Column heater engine problem Replace the column heater engine.

COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 55 OF 85

Table 7:  Column device errors and warnings (continued)

Error or warning Possible cause Corrective action

Column temp h/w fault (6) Thermistor problem Inspect/Reconnect/Replace. 

Thermistor low–the value returned  Column heater PCB problem Replace the PCB.


by the thermistor is unreasonably 
low.

Column temp h/w fault (7) CHC unplugged or powered‐off Ensure that the unit is plugged in 


and powered‐on. 
Internal power failure.
CHC power supply problem 1. Inspect fuse on PCB.
Replace fuse or PCB if necessary.
2. Replace the CHC power supply.

Column temp h/w fault (8) Column heater temperature  Inspect/Reconnect/Replace. 


thermistor problem
Thermistor high–the value returned 
by the thermistor is unreasonably  Column heater PCB problem Replace the PCB.
high.

Column temp over limit (X°) Insufficient equilibration time Allow enough equilibration time 


between column temperature 
(where X is the faulting tempera‐ changes for the temperature to 
ture) reach the new set point. 

Ambient operating temperature  Determine the room temperature 
limit exceeded and adjust if necessary.
See related: Column heater not 
ambient, page 53

Cable problem or disconnection 1. Inspect the cable that runs from 
the column heater thermistor to 
the column heater PCB.
2. Replace if necessary.

Thermistor problem 3. Inspect/Reconnect/Replace.  

PCB problem 1. Inspect the PCB. 
2. Replace if necessary. 

Column heater hardware problem 1. Inspect the column heater 
hardware. 
2. Replace engine.

COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 56 OF 85

Table 7:  Column device errors and warnings (continued)

Error or warning Possible cause Corrective action

Error changing column position No column‐switching valve is  1. Confirm that the column 


installed, or the instrument cannot  switching valve is installed.
The column switcher indicates an  detect it 2. Install valve if necessary.
error in moving to the requested 
position. Damaged cable 1. Inspect the cable that connects 
the column switching valve to 
the column heater PCB.
2. If the valve is installed, verify 
correct setting in back of column 
heater. 
See Column device heater dial/
potentiometer settings, page 75

Incorrect setting Ensure that the dial/potentiometer 
setting on the column heater is 
correct for the installed switching 
valve. 
See Column device heater dial/
potentiometer settings, page 75

Column heater configuration  Power cycle the column heater to 
misread force the system to reread the 
column heater configuration. 

Separation method issue Modify the separation method to 
eliminate the column position 
change. 

Valve initialized incorrectly 1. Power‐down Alliance.
2. Wait one minute and power‐on.
3. After Alliance initialization, use 
the front panel to reposition the 
switching valve. 
If the valve position remains 
unchanged, reboot again using 
the one‐minute wait time.
4. If you are unsuccessful after 
three attempts, additional 
troubleshooting is required. 
Contact GSS.

Valve PCB problem Replace the PCB. 

Column heater PCB problem 1. Inspect the PCB.
2. Replace if necessary.

COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 57 OF 85

Table 7:  Column device errors and warnings (continued)

Error or warning Possible cause Corrective action

Low column temp (X°) Insufficient equilibration time Allow enough equilibration time 


between column temperature 
where X is the faulting temperature changes for the temperature to 
reach the new set point. 
The column temperature is below 
the minimum specified in the cur‐ Frequent door openings Ensure that the column heater door 
rent separation method. is closed long enough to prevent a 
drop in temperature. 

Problem with door gasket Repair or replace.

Column heater set point too close  Ensure that the temperature set 
to ambient point is a minimum of 5 °C above 
ambient. 

Ambient operating temperature  1. Determine the room 
limit exceeded temperature. 
2. Adjust if necessary.
See related: Column heater not 
ambient, page 53

Damaged cable 1. Inspect the cable that runs from 
the column heater thermistor to 
the column heater PCB.
2. Replace if necessary. 

Column heater temperature  Inspect/Reconnect/Replace. 
thermistor problem

Column heater PCB problem Replace the PCB.

Low column temp (X°) Column heater hardware or column  Replace the hardware or column 


heater problem heater.

COLUMN DEVICE ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 58 OF 85

Detector control errors

This section lists the system errors and warnings specific to the control of the detector in alphabetical 
order. Locate the relevant error or warning in the left column and follow the appropriate troubleshooting 
steps.
For troubleshooting information pertinent to each detector, see:

• ACQUITY QDa Detector: troubleshooting information
• 2414 RI Detector: troubleshooting information
• 2424 ELS Detector: troubleshooting information
• 2998 PDA Detector: troubleshooting information
• 2489 UV/Vis Detector: troubleshooting information
• 2475 FLR Detector: troubleshooting information

Table 8:  Detector control errors and warnings

Error or warning Possible cause Corrective action

410 detector fault Communications error  1. Ensure that the detector is 


powered‐on.
Occurs while the instrument is con‐ 2. Rescan the IEEE bus from the 
trolling a 410 detector over IEEE.  detector window. 

Incorrect detector address 1. Ensure that the IEEE address of 
the detector is correct. 
2. Rescan the IEEE bus from the 
detector window. 

Problem with IEEE cable between  1. Inspect the IEEE cable between 
Alliance system and detector the Alliance system and the 
detector.
Note:  Power‐off before removing 
or attaching the IEEE cable.
2. Reconnect.
3. Replace if necessary.
4. Rescan the IEEE bus from the 
detector window. 

410 detector CPU board problem 1. Replace the 410 detector CPU 
board.
2. Rescan the IEEE bus from the 
detector’s window.

Alliance CPU PCB problem 1. Replace the Alliance CPU PCB 
board.
2. Rescan the IEEE bus from the 
detector’s window.
See technical reference sheet, 
Alliance CPU PCB (715003814)

DETECTOR CONTROL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 59 OF 85

Table 8:  Detector control errors and warnings

Error or warning Possible cause Corrective action

486 detector fault Communications error  1. Ensure that the detector is 


powered‐on.
(Also known as: "Absorbance detec‐ 2. Rescan the IEEE bus from the 
tor fault) detector window. 
The detector, under IEEE bus 
control, is taken off‐line. Incorrect detector address 1. Ensure that the IEEE address of 
the detector is correct. 
2. Rescan the IEEE bus from the 
detector window. 

Problem with IEEE cable between  1. Inspect the IEEE cable between 
Alliance system and detector the Alliance system and the 
detector.
2. Reconnect the cable.
3. Replace if necessary.
4. Rescan the IEEE bus from the 
detector window. 

486 detector CPU problem Replace the 486 detector CPU 
board. 

Alliance CPU PCB problem 1. Replace the Alliance CPU PCB 
board.
2. Rescan the IEEE bus from the 
detector’s window.
See technical reference sheet, 
Alliance CPU PCB (715003814)

DETECTOR CONTROL ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 60 OF 85

Dissolution errors

This section lists the system errors and warnings specific to Dissolution in alphabetical order. Locate the 
relevant error or warning in the left column and follow the appropriate troubleshooting steps.

Table 9:  Dissolution errors and warnings

Error or warning Possible cause Corrective action

Bath speed tolerance The connected bath reported a  Bath mechanical system requires 


paddle speed out of tolerance error service. Considerations:
• RPM test
• Loose or worn belt
• Bent paddle
• Excessive vibration
Note:  Only Hanson Research is 
authorized to perform major 
service on the bath 
mechanism.

Bath temp tolerance The connected bath reported a  1. Verify proper bath circulation.


temperature out of tolerance error 2. Verify that the heater plate is 
positioned over the heater.
3. Inspect water level.
4. Verify that the heater is heating. 

Allow for a greater tolerance in the 
bath temperature by changing the 
limits.

Dissolution An error occurred during the  1. Verify the cable connection 


synchronized operation of the  between 2695 comm port B and 
This is a generic message. separations module and transfer  WTM comm port 1.
module 2. Replace cable.
3. Verify that the WTM rear panel 
LED is green.
4. Replace WTM CPU PCB.

Dissolution (bad state) Firmware encountered an  1. Verify the cable connection 


unexpected condition between 2695 comm port B and 
This is a generic message. WTM comm port 1.
2. Replace cable.
3. Inspect power supply.
4. Verify that the WTM rear panel 
LED is green.
1. Replace WTM CPU PCB.

DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 61 OF 85

Table 9:  Dissolution errors and warnings

Error or warning Possible cause Corrective action

Dissolution (carousel) Carousel carrier drive problem 1. Verify that the compartment 


floor is clean and free of debris, 
Indicates that the required vials  and that there are no physical 
could not be located under the dis‐ obstructions of the carousel.
solution dispenser when needed. 2. Verify that carousel is in correct 
location.

Vial sensor problem 1. Clean sensor.
2. Verify sensor alignment.
3. Replace vial sensor.
4. Confirm carousel carrier 
movement.
5. Verify that there are no 
obstructions.

Wrong carousel Verify proper carousel and sensor 
identification of that carousel.

Z axis problem Verify z axis operation/depth.
See Adjusting the Z‐Axis Ram 
Height and the Carousel Carrier 
Position (715002039)

Dissolution (comm error) Power supply problem  1. Verify that the power supply is 


(not supplying 24 volts) providing 24 volts.
Indicates that communications was  2. Replace power supply if 
temporarily lost with the Waters  necessary.
Transfer Module or bath. 
Problem with cable from 2695 port  1. Verify cable connectivity and 
B to WTM Comm1 functionality.
2. Replace if necessary.

Problem with cable from WTM  1. Verify cable connectivity and 
Comm2 to Bath Comm1 functionality.
2. Replace if necessary.

WTM/Bath CPU problem 1. Verify that the rear panel LED is 
green. 
• If green, CPU is good.
• If red, replace CPU.

Dissolution (couldn’t continue) Dissolution method problem Create a new method.

Indicates that the Waters Transfer  WTM initialization problem/lost  1. Reboot WTM.


Module failed to perform the post‐ communication Note:  Backflush is default setting.
sample backflush or media replace‐ Media Replace is a SW 
ment.  selection.

DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 62 OF 85

Table 9:  Dissolution errors and warnings

Error or warning Possible cause Corrective action

Dissolution (couldn’t dispense) Incorrect needle height Adjust needle height.

Indicates the Waters Transfer Mod‐ Tubing disconnected or kinked 1. Inspect the tubing connections 


ule failed to dispense the sample  and verify that there are no 
when requested to do so.  kinks.
2. Replace tubes if necessary.

Dispenser assembly problem 1. Verify dispenser operation.
2. Replace if necessary.

Fluidic driver PCB problem Firmware should be >3.0.
(See service note) (link opens in 
IBM Notes)

Firmware problem Verify that the fluidic driver PCB is 
the correct revision.
See service note (rev 0 or 6) (link 
opens in IBM Notes)

Dissolution (dispenser move) Power supply problem 1. Verify that the power supply is 


(not supplying 24 volts) providing 24 volts.
Indicates that the dissolution dis‐ 2. Replace power supply if 
penser needles failed to move up or  necessary.
down. 
Needle problem 1. Verify that the needles are 
functioning properly.
2. Replace if necessary.

Dispenser assemble jam 1. Verify dispenser operation.

Fluidics driver board problem Firmware should be >3.0.
(See service note) (link opens in 
To help determine the source of  IBM Notes)
the problem, use the procedure: 
Troubleshooting the XYZ and  Verify that the fluidic driver PCB is 
needle bottom sensor errors (279x  the correct revision.
only), page 85. See service note (rev 0 or 6) (link 
When performing the procedure,  opens in IBM Notes)
test the combination: Replace the fluidics board. 
• Dissolution top sensor with corre‐ See:
sponding LED DS19 on the fluidics  Alliance 269x: Fluidics Driver PCB
driver board. Alliance 279x: Fluidics Driver PCB
• Dissolution bottom sensor with 
corresponding LED DS20 on the 
fluidics driver board.
Use the information obtained from 
that procedure to determine the 
cause and corrective action.

DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 63 OF 85

Table 9:  Dissolution errors and warnings

Error or warning Possible cause Corrective action

Dissolution (TM sensor) Transfer Module movement  1. Verify the up/down movement 


restricted from diagnostics.
Indicates that the Waters Transfer  2. Verify that the sensor flag is 
Module has not reached sensor  functioning properly.
position. 
3. Lubricate guide rails.
4. Replace rear panel PCB.

Dissolution (TM syringe) Loose syringe Tighten syringe.

Indicates that the Waters Transfer  Worn plunger Replace syringe.


Module syringe transport is  Syringe misaligned 1. Lubricate WTM guide rails.
jammed. 
2. Perform a prime or rinse 
function.

WTM guide post alignment  1. Remove all syringes.
problem 2. Verify up/down movement.
3. Lubricate guide post.
4. If problem remains unresolved, 
contact GSS.

Dissolution (vial scan) Missing vial Ensure that the vials are inserted.

Indicates that an automatic pre‐ Inconsistent vial height 1. Verify that all vials are the same 


transfer scan has determined that  height.
one or more required vials is miss‐ 2. Verify that all vial caps are black 
ing.  or blue.

Scan sensor optics problem 1. Verify that fiber optic cable is in 
correct position.
2. Verify that fiber optic cable is 
working properly.

Fluidics driver board problem Replace the fluidics board. See:
Alliance 269x: Fluidics Driver PCB
Alliance 279x: Fluidics Driver PCB

Missed a dissolution sample An unexpected timing conflict  1. Verify dissolution method 


prevents the transfer module from  parameters.
collecting a scheduled sample 2. Test functions:
• Purge
• Prime syringes
3. Test the dispense function from 
diagnostics.

DISSOLUTION ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 64 OF 85

Alliance HPLC 2795 and 2796 errors

This section lists the system errors and warnings specific to Alliance HPLC 2795 and 2796 in alphabetical 
order. Locate the relevant error or warning in the left column and follow the appropriate troubleshooting 
steps.

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings

Error or warning Possible cause Corrective action

410/2410 detector fault Communications error  1. Ensure that the detector is 


powered‐on.
The instrument is controlling a 410/ 2. Rescan the IEEE bus from the 
2410 detector over IEEE, but a com‐ detector window. 
munications error has occurred and 
the detector has been taken offline.  Incorrect detector address 1. Ensure that the IEEE address of 
the detector is correct. 
2. Rescan the IEEE bus from the 
detector window. 

Problem with IEEE cable between  1. Inspect the IEEE cable between 
Alliance system and detector the Alliance system and the 
detector.
2. Reconnect or replace the cable if 
necessary.
3. Rescan the IEEE bus from the 
detector window. 

410/2410 detector CPU board  Replace the detector CPU board 
problem and then rescan the IEEE bus from 
the detector’s window.

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.

See technical reference sheet, 
Alliance CPU PCB (715003814)

Can't understand requested well Incorrect configuration Do one of the following:


• Select a different well for the 
The well specified for the current  injection.
injection does not exist.  • Ensure that the plate type in use 
contains the required well loca‐
tion.
• Ensure that the plate type in use 
matches the plate configuration. 

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 65 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

Carrier is busy A user is interacting with the front‐ Wait until vial‐selection operations 


panel and is inadvertently trying to  are complete before interacting 
A vial selection operation was in  use one of the vial selection  with the front‐panel. 
progress when another process  components Note:  Most vial‐selection 
tried to use one of the necessary  operations take place at the 
components (carrier drive, carousel  start of injections.
drive, needle drive, etc.). 

Carrier lockpin movement failed Problem with cable connecting the  1. Verify that the cable is connected 


lockpin sensor to the CPU board properly.
The carrier lockpin moved up, but  2. Replace cable if necessary.
the home sensor did not activate.
Misalignment or problem with flag  1. Realign flag.
mounted on lockpin lead screw 2. Replace if necessary.

Lead screw not moving Use diagnostics to test movement.

Problem with sensor Replace the sensor.
See: 
• Lockpin Motor and Sensor 
• Lockpin Drive Sensor

Problem with lockpin motor Replace the motor.
See: 
• Lockpin Motor and Sensor 
• Lockpin Drive Motor

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Carrier malfunction Dusty carrier drive position sensor Clean sensor.

The carrier failed to move to the  Dusty encoder wheel Clean wheel.


requested carousel position.  Misaligned carrier drive position  Align sensor.
sensor

Debris/obstacles on compartment  Remove any obstacles and clean 
floor affecting carrier the compartment floor.

Carrier drive assembly problem Replace the carrier motor or drive 
See also: assembly.
• Carrier Assembly
• Carrier Drive 

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 66 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

Default separation method created The system finds no default  No action required. 


separation method The system creates a default 
method.
Note:  This message is added to the 
error log whenever the 
default separation method is 
recreated. This occurs on the 
first start‐up after the flash 
memory is cleared. 

Eluent Monitor Disabled Firmware encountered an  1. Examine the trace log to 


unexpected condition when trying  determine instrument activity at 
This is a generic message. to control the eluent monitor the time of failure. 
Note:  This error applies only to  2. Contact GSS and be prepared to 
Alliance 2796 systems. describe the conditions leading 
to this error.

Eluent Monitor Error Firmware encountered an  1. Examine the trace log to 


unexpected condition when trying  determine instrument activity at 
This is a generic message. to control the eluent monitor the time of failure. 
Note:  iThis error applies only to  2. Contact GSS and be prepared to 
Alliance 2796 systems. describe the conditions leading 
to this error.

Injection parameters incorrect Error in either plate location or  Select a well compatible with the 


well‐identification format plate configuration or change the 
The system is unable to find the  plate configuration. 
specified well. 

Inject Valve movement failure Inject valve problem Replace the inject valve.


See: 
The inject valve was moved, but the  • 2796 Inject Valve
reading from the position sensor  • 2790/2795 Inject Valve Rebuild 
did not match the requested posi‐ Kit 
tion.
Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Must degas solvents for SysPrep Degasser powered‐off Power‐on degasser.

A degasser must be installed and  Degasser lost connectivity to  1. Inspect connectivity.


powered‐on to run the System Prep  system 2. Replace cables if necessary.
function. 
No degasser installed 1. Install degasser.
2. Rerun System Prep.

Degasser PCB problem 1. Replace degasser PCB.

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 67 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

No‐Flow Shutdown Invoked The No‐Flow Shutdown routine  No action required.


runs because the pump flow rate  This is normal/expected behavior. 
The No‐Flow Shutdown feature has  remains at 0‐ml/min for the time 
turned off the pump or degasser.  specified on the Auto Shutdown 
page

Overpressure: check inline filter Pin plug restrictor was not removed Ensure that the pin plug restrictor is 


removed.
A large amount of pressure was 
detected while moving the pump  Clogged in‐line filter Clean or replace the filter.
plungers to their home positions.  See:
• 2790/2795: Replacing the In‐Line 
Filter 
• 2796: 2796 In‐Line Filter 

Blocked waste line Clear any blockage of the waste 
line.

pH/Cond. Monitor Disabled A serious error disabled the pH/ 1. Examine the trace log to 


conductivity monitor determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

pH/Cond. Monitor Error The pH/conductivity monitor  1. Examine the trace log to 


experienced an unexpected error determine instrument activity at 
the time of failure. 
2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Plate parameters are incorrect Incorrect parameters Do one of the following:


• Change the plate dimensions to 
The user‐configured plate dimen‐ fit within the field of movement 
sions extend beyond the range of  of the X/Y mechanism. 
the X/Y mechanism movement  • Select a different plate type.
field. 

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 68 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

Plate height error Mismatch of measured plate height  Review the plate‐height parameter 


and selected plate type in the specified plate configuration 
The plate height detected does not  Note:  Occurs when the separation  and do one of the following:
match the current plate configura‐ method enables the Check  • Ensure that the correct plate type 
tion.  plate height option. is in use. 
• Change the plate configuration to 
match the physical plate in use. 

Needle bottom sensor problem 1. If dusty, clean sensor.
2. Verify that the cable is connected 
properly between sensor and 
PCB.
3. If malfunctioning, replace.

Needle deck height mis‐calibrated Run the Replace Needle diagnostic 
to recalculate the deck height.

Plate X axis parameters incorrect Specified plate dimensions exceed  • Change the plate dimensions to 


the range of X movement fit within X/Y mechanism's field 
The user‐configured plate dimen‐ of movement. 
sions extend beyond the range of  • Select a different plate type. 
the X/Y mechanism movement 
field. 

Plate Y axis parameters incorrect Specified plate dimensions exceed  • Change the plate dimensions to 


the range of Y movement fit within X/Y mechanism's field 
The user‐configured plate dimen‐ of movement. 
sions extend beyond the range of  • Select a different plate type. 
the X/Y mechanism movement 
field. 

RS‐232 Valve Comm Error Incorrect RS‐232 valve address Correct the address.

An error occurred while trying to  Problem with cable 1. Inspect the cable and reconnect.


communicate with an external  (441000459) 2. Replace if necessary.
RS‐232 valve. 
RS‐232 valve electronics problem Replace the valve.

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.

See technical reference sheet, 
Alliance CPU PCB (715003814)

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 69 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

RS‐232 Valve Error Firmware encountered an  1. Examine the trace log to 


unexpected condition determine instrument activity at 
An unknown error occurred while  the time of failure. 
trying to control the RS‐232 valve. 2. Contact GSS and be prepared to 
describe the conditions leading 
to this error.

Problem with cable 1. Inspect the cable and reconnect.
(441000459) 2. Replace if necessary.

RS‐232 valve electronics problem Replace the valve.

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.

See technical reference sheet, 
Alliance CPU PCB (715003814)

RS‐232 Valve Missing Incorrect RS‐232 valve address Correct the address.

The RS‐232 valve could not be  Problem with cable 1. Inspect the cable and reconnect.


detected. (441000459) 2. Replace if necessary.

RS‐232 valve electronics problem Replace the valve.

Alliance CPU PCB problem Replace the Alliance CPU PCB 
board.

See technical reference sheet, 
Alliance CPU PCB (715003814)

Sample too large for configuration Incorrect configuration Change the injection volume or 


syringe and sample loop size to be 
The total injection volume exceeds  compatible. 
the limit allowed for the currently  Note:  For full loop injections, the 
configured syringe size and sample  recommended holding loop 
loop size. size is also a consideration.

SysPrep requires a degasser No degasser installed 1. Install degasser.


2. Rerun SysPrep.
The SysPrep function requires an 
operating vacuum degasser. Degasser lost connectivity to  1. Inspect connectivity.
system 2. Replace cables if necessary.

Degasser PCB problem Replace degasser PCB.

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 70 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

Top and bottom needle sensors  Dirty sensors Clean the sensors.


active
Cable disconnection or problem 1. Inspect connectivity of cables 
Both the top and bottom needle  and reconnect if required.
sensors are active, which should  2. If cable is damaged or 
not be physically possible.  malfunctioning, replace.

Incorrect needle installation Reinstall the needle.
See:
• 2790/2795 Needle 
• 2796 Needle 
• Rebuild/Replace 2795 Needle 
Assembly 

To help determine the source of the problem, use the procedure, 
Troubleshooting the XYZ and needle bottom sensor errors (279x only), 
page 85.
When performing the procedure, test the combination:
• Needle bottom sensor with corresponding LED DS13 on the fluidics 
driver board.
• Needle top sensor with corresponding LED DS16 on the fluidics driver 
board.
Use the information obtained from that procedure to determine the cause 
and corrective action below.

Sensor problem Replace the respective sensor.

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

Well depth error Bad or deformed plate 1. Try another plate.

The measured well depth does not  Mismatch of well depth and plate  1. Review the well‐depth 


match the plate type selected. type parameter in the specified plate 
configuration.
Note:  This error occurs when the 
separation method enables  2. Ensure that the specified well 
the Seek well bottom  depth matches the plate 
option. configuration.
Note:  Consider adding a 
well‐depth offset to the 
separation method.

Needle bottom sensor problem 1. If dusty, clean sensor.
2. Verify that the cable is connected 
properly between sensor and 
PCB.
3. If malfunctioning, replace.

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 71 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

X axis malfunction To help determine the source of the problem, use the procedure, 
Troubleshooting the XYZ and needle bottom sensor errors (279x only), 
An x axis movement failed to reach  page 85.
the requested position.  When performing the procedure, test the combination:
• X axis home sensor with corresponding LED DS14 on the fluidics driver 
board.
See:
• Alliance 279x: Fluidics Driver PCB
• Adjusting the XY Mechanism in 2790/2795/2796 Separations Modules

Dusty x axis rails  Clean the x axis rails.

Dusty x axis home sensor Clean the x axis home sensor.

X axis home sensor problem 1. Adjust x axis sensor.
2. Replace the x axis home sensor if 
necessary.
See X‐Axis Home Sensor 

Cable disconnection or problem 1. Inspect connectivity of cables 
and reconnect if required.
2. If cable is damaged or 
malfunctioning, replace.

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

The x axis belt may be slipping or  Tighten or replace the belt.
broken See X Traverse Belt

Misaligned x axis rails 1. Verify that the belt and pulley 
are not to tight or loose. Adjust if 
necessary.
2. Align the x axis rails.

Problem with x axis motor  Replace the x axis motor.
See X‐Axis Motor

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 72 OF 85

Table 10:  Alliance HPLC 2795 and 2796 errors and warnings (continued)

Error or warning Possible cause Corrective action

Y axis malfunction To help determine the source of the problem, use the procedure, 
Troubleshooting the XYZ and needle bottom sensor errors (279x only), 
A y axis movement failed to reach  page 85.
the requested position.  When performing the procedure, test the combination:
• Y axis home sensor with corresponding LED DS15 on the fluidics driver 
board.
Use the information obtained from that procedure to determine the cause 
and corrective action below.
See:
• Alliance 279x: Fluidics Driver PCB
• Adjusting the XY Mechanism in 2790/2795/2796 Separations Modules

Dusty y axis rails  Clean the y axis rails.

Dusty y axis home sensor Clean the y axis home sensor.

Y axis home sensor problem Replace the y axis home sensor.
See:
• Y‐Axis Home Sensor 
• YZ and Needle Bottom Breakaway 
PCB Assembly

Cable disconnection or problem 1. Inspect connectivity of cables 
and reconnect if required.
2. If cable is damaged or 
malfunctioning, replace.

Fluidics driver board problem Replace the fluidics board. 
See Fluidics Driver PCB

The y axis belt may be slipping or  Tighten or replace the y axis belt.
broken See Y Traverse Belt

Misaligned y axis rails Verify that the pulley and belt are 
not too tight or loose and adjust if 
necessary.

Problem with y axis motor  Replace the y axis motor.
See Y‐Axis Drive Motor

ALLIANCE HPLC 2795 AND 2796 ERRORS
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 73 OF 85

Reference information

This section describes common tests and procedures that are referenced within this guide.
PSI vs. Flow Envelope  . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 73
Sample Heater/Cooler Testing . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 74
Troubleshooting Transducer Problems  . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 74
Column device heater dial/potentiometer settings . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 75
Performing the Random Vial Position Test  . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 75
Needle Piercing Alignment Procedure . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 76
Carousel and dogpin parts diagram (2695 only) . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 78
Z axis ram height gauge (700001022) settings  . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 79
2690 Fluidics Diagram . . . . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 80
2690 Tubing Connections . . . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 81
2690 Interconnect Diagram  . . . . . . . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 82
Server Busy Error: Troubleshooting . . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 83
Troubleshooting the Vacuum Pump. . . . . . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . .. . . . . . . .  . . . . . . . . . . . . . 84
Troubleshooting the XYZ and needle bottom sensor errors (279x only). . . . . . .. . . . . . . .  . . . . . . . . . . . . . 85

PSI vs. Flow Envelope


715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 74 OF 85

Sample Heater/Cooler Testing

When testing the Sample Heater/Cooler with a calibrated thermocouple: place the thermocouple in the 
compartment over the center of the carousel. 

Notes:  
—Be aware that temperature degradation occurs at ambient temperatures above 29 °C. 
—The heater/cooler can only attain internal temperatures of 4 °C or 25 °C below 
ambient. For example, if the ambient temperature is 30 °C the compartment 
temperature can only attain a temperature of 5 °C or a 25 °C differential. 

Troubleshooting Transducer Problems

A transducer failure can occur during initialization. If the software cannot electronically zero the 
transducer, an "Out of range" error is displayed.

Note:  During initialization, the transducer voltage must be within a specific range. The range to pass 
power‐up initialization is 0.5V +/‐0.2V (0.3V ‐ 0.7V). If the voltage is outside of this range, an 
"Out of range" error occurs. You can view the voltage in Service diagnostics, Service Utilities in 
a pane containing the voltage offsets. An ideal transducer has a reading of 0.5V.

You can swap cable connections at the CPU board location to troubleshoot, isolate and verify when a 
specific transducer has failed. If the problem follows the transducer, attempt to adjust or replace the 
transducer. You should also inspect the cables and connections. 
If back pressure is present on the transducers, you could see the same error. Verify that all pressure has 
been released from the fluid components while troubleshooting this failure. 

Re‐initializing 
Transducers are auto‐zeroed during re‐initialization. 

Transducer Adjustments 
You can manually make transducer adjustments by pre‐zeroing the transducer to bring it into an 
acceptable range for the electronic zero during power‐up. Remove the black rubber stopper on the 
transducer to access the adjustment potentiometer. You can perform the adjustment by either trial/error, 
or by carefully checking the voltage on a microchip. 

Note:  Ideal transducer reading is 0.5V. ±0.2V.

Trial‐and‐error method
Record the Out of range number displayed, power‐off the instrument, make a slight adjustment to the 
transducer, then power‐on the instrument. Continue this adjustment procedure until the Out of range 
number is decreased to eliminate the error.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 75 OF 85

Monitoring voltage method
Power‐on the instrument and connect the positive lead of the voltmeter to the following chips on the CPU 
PCB to monitor the transducer offset voltage and connect the negative lead to chassis ground. If possible, 
adjust the potentiometer to approximately 0.5V.

Table 11: 
Transducer Chip and Pin #
System: U49 Pin 7
Primary: U49 Pin 1
Sample: U51 Pin 7

Note:  Content for this topic sourced from service note.

Column device heater dial/potentiometer settings

Ensure that the dial/potentiometer setting on the column device heater is correct for the installed 
switching valve. The switch is located on the back of the column module.
Switch position and valve type:
• 0: No valve
• 1: Three‐Column Select Valve Upgrade Kit (205000162)
• 7: 2‐Position Regeneration Valve Upgrade Kit (205000163)
• F: 6‐Column Select Valve Upgrade Kit (205000164)

Performing the Random Vial Position Test

This procedure is applicable to Alliance 269x.

1.   Install vials with caps in positions 59, 60, 62, 63, 83, 84, 86, and 87. Install no vials in positions 61 
and 85.
2.   Run the method.
3.   Ensure that the sample set stops at vial 85 and the alarm “VialMissingError” displays.
4.   Install a vial in carousel position 85 and resume injections.
5.   Ensure that the sample set stops at vial 61 (the alarm “VialMissingError” displays).
6.   Install a vial in position 61 and resume injections.
•   If injections continue as specified in the method, the test passes and no adjustment is 
needed. 
•   If the test fails, the vial’s sensor optics are out of position and require adjustment.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 76 OF 85

Needle Piercing Alignment Procedure

The content that follows was originally publish as a service note. See service note: Needle Piercing 
Alignment.

Scope Technical Discussion    
If the injector is piercing off center and the injector motor loses steps due to striking the vial cap or the 
side of the vial the error Injector malfunction "bottom sensor error" (bse) will be displayed. In this case, 
the BSE is not related to an issue with the injector, but due to a misalignment of the carrier portion of the 
carousel‐carrier drive. The Alliance methods below should be used to diagnose the carrier misalignment 
problem globally with all the carousels. The carousel carrier drive may be adjusted if any of the piercings 
are grossly out of alignment. 

Product Affected                  
2695, 2695D

Tools required:
Adjustment requires a T‐10 Torx® driver.

Alignment Procedure
Determine current needle piercing locations

1.   Use the piercing template on all the carousels or some other covering, such as parafilm, to see 
the needle holes.

Note:  If you use masking tape, be sure to thoroughly wipe the needle with a lintless wiper (for 
example, KimWipe), moistened with methanol, after completion of the test.

2.   Open the service note Needle Piercing Alignment, and download the Alliance methods below to a 
diskette, and import them into the 2695. 
•   OPTOTEST.MTH
•   TESTCCD.MTH
3.   Verify that the needle wash system has the appropriate solvent and is primed. 
4.   Go into Menu > Status and set up the Optotest as your Separation Method (0 ml/min flow).  
5.   Run the Testccd  Sample Method (runs 20 injections of 0 ul that have a runtime of .01 min). 
6.   Inspect all the needle holes for alignment and determine if the carousel needs to be adjusted 
clockwise (CW) or counter‐clockwise (CCW).
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 77 OF 85

Align the needle piercing locations
7.   Loosen the sensor lockdown screw located to the left of the carousel rotation motor. 

8.   Rotate the spring‐tensioned sensor adjustment screw located on the right‐hand side of the 
carousel rotation motor.
•   To rotate the carousel position CW, turn the sensor screw CCW.
•   To rotate the carousel position CCW, turn the sensor screw CW.

9.   After each adjustment, tighten the lockdown screw and run test injections in several locations on 
multiple carousels to ensure that the adjustment is optimized for all carousels. 
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 78 OF 85

Carousel and dogpin parts diagram (2695 only)


715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 79 OF 85

Z axis ram height gauge (700001022) settings

See the following service notes for more information regarding the use of the ram height gauge:

• Carousel Carrier Drive Sensor Adjustments
• 2695 Zram Height and Carrier Plate Adjustment
• Z Axis Malfunction vs Carousel Malfunction
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 80 OF 85

2690 Fluidics Diagram


715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 81 OF 85

2690 Tubing Connections


715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 82 OF 85

2690 Interconnect Diagram


715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 83 OF 85

Server Busy Error: Troubleshooting

Empower Error Message: Server busy

Possible Cause:
Interactions between Empower and the Quickset window cause a freeze. 

Affected systems:
• Single systems
• Multiple systems configured on the acquisition machine.

Affected configuration:
A 2695 with both a 2489 and 2996 detector configured.

How the system is affected:

1.   The 2695 pressure console freezes at the last recorded pump pressure.
2.   Task Manager indicates that the instrument server is using approximately 25% of one of the CPUs.
3.   Approximately two minutes after the abort/inject, Server Busy displays on the Quickset window.
4.   A single error message is recorded in the Message Center => "2489XXXX instrument currently in 
use" (where XXXX is the instrument serial number).

Corrective Action:
No action required.
This issue was resolved in Empower 2 Feature Release 5 and in all subsequent releases.
If using any version of Empower 2 prior to FR5, avoid using the abort button during acquisition.
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 84 OF 85

Troubleshooting the Vacuum Pump

1.   Inspect and resolve the following:
•   Inspect the vent line for liquid.
•   Inspect manifold tubing for kinks, holes, and a tight fit.
•   Pull off and reseat ALL vacuum connections to eliminate the flashing effect from the manifold 
molding process.
2.   If pump fails to power‐on, inspect:
•   Power cable connection.
•   Vacuum pump motor connection.
•   Voltage at PCB connection.
•   Vacuum pump to see if it stalled.
3.   If error still occurs, troubleshoot the pathway by bypassing/removing the chambers and only use 
vacuum pump and PCB.

•   Pinch off tubing to chambers to troubleshoot individual chamber. Confirm that the pressure 
reaches 0.9 psi.
•   You might have to change the tubing configuration. 
Connect the vacuum pump directly to the PCB Transducer (use the port closest to the edge 
with the solder leads). 
Confirm that the pressure reaches 0.9 psi.
•   Check LEDs. See Degasser PCB (715003571).
4.   Replace components as necessary.
See Degasser Assembly Component Replacement
715003674 REV. C COMPANY CONFIDENTIAL - PROPRIETARY INFORMATION - FOR INTERNAL USE ONLY 85 OF 85

Troubleshooting the XYZ and needle bottom sensor errors (279x only)

You can use the LEDs on the fluidics driver board to determine if an error is caused by the sensor, cable, or 
fluidics driver board. See also: Fluidics Driver PCB LEDs and Sample Handling Sensors.
This procedure is useful when troubleshooting the following errors:
• Top and bottom needle sensors active, page 70
• X axis malfunction, page 71
• Y axis malfunction, page 72
These instructions are written generically. Adapt for the specific sensor/LED combination in question.

Note:  DS13, 14, 15, and 16 are visible through a cutout on the back side of the PCB.

To identify the malfunctioning component: 
Place the flag into the sensor.
• If the corresponding LED turns ON, the sensor, cable and fluidics driver board are functioning 
correctly.
If the corresponding LED remains OFF, troubleshoot the sensor, cable, PCB chain.

You might also like