Automated Hematology and

ESR Analyzer MEK-1305


This guide shows

Operation Guide
simply workflow.
For more detail and
information, see the
operation manual.

Start up
Power ON
Status Indicator

1 Press the power switch on the front.

2 Select operator and Logon.

3 Preform Self-check. (About 6 minutes)


Check the status indicator becomes green.

(*)The background check can be included


Check back ground in the self-check by settings.

From the Home screen, touch [Maintenance]


1 [Maintenance Operation]key.

2 Touch [Background Check].


Back ground
WBC 0.2 × 103/μL or less
RBC 0.02 × 106/μL or less
HGB 0.10 g/dL or less
PLT 10.0 × 103/μL or less
Press [Yes]. It
takes about 1
minute.

When the information key [ ] is turn on red or flashes, touch it to


display. Take the displayed action, and touch [ ] key .
Quality Control For quality control, lot registration is
required every 3 months, when a new
blood control lot is available, make a
Register QC lot of control new registration.

From home screen, touch [QC] and [Lot Management].


1 And press [New] key to register a new lot.

Read the barcode printed on the assay sheet in the


2 order of 1 2 3 with a handy barcode reader. The
assay value, range, expiration date and lot number
are automatically entered.

3 Touch the [Enable] check box and then touch [OK].


(*) Whether to use for quality control can be switched
later by checking the target lot on the lot management
screen and then touching [Status].

Take out the blood control from the


refrigerator and left for 10 minutes in
Run QC room temperature.
Mix with inverting 20 times or more.

1 Touch [ ] key.
Input QC lot number into the sample ID column. If the sample ID
2 is same as registered QC lot, the parameter set is automatically
set to "QC".
3 Touch the parameter set key on the measurement screen and
select the measurement method.
Measure the control. If the all results is within the
4 reference value for all registered concentrations, the quality
control status will be green. (PASS state)

5 Out-of-range results are displayed as


red dots on the QC graph screen.

6 Press [ ] key and enter the period to print the QC report.


Measurement
Prepare for measurement
1 Touch the [ ] bottom.

2 Select the parameter set.

3 Input the sample ID number.


(Read barcode by handy barcode reader. )

4 Mix with inversion the sample tube 20 times or more.

*Until the sample nozzle rises automatically,


Open mode keep sample tube at the position.

Remove the lid of the sample tube


1
and insert the sample nozzle at the
position of right figure.

2 Press the measurement switch.

Pre dilution mode (*) CBC mode only

1 Press [ ] key. The sample


nozzle will raise and ready to dispense
the diluent.
When "Dispense diluent" is
2 displayed, press “Yes” to dispense
the diluent into the assist tube.

Dispense the blood into the assist tube with a


3 dropper. Cover the lid of assist tube and mix.

4 Remove the lid of the assist tube and insert the


sample nozzle as shown in the right figure.

5 Press the measurement switch.


Check measurement results

CBC result is shown within about 1 minute.


CBC+ESR result is shown within 2 minutes.
The result will printout from the printer automatically.
Input next sample ID
number.

Check the data.

Display the histograms

Display the normal


range bar graph.

Check the flag message.

Printout. Transfer the result.

Smart ColoRerun Assist

“Smart Color Rerun Assist” is a function to


check the cause of re-measurement at a
glance with color messages.

Yellow message Orange Message Red Message


Panic value. The result may not be correct During measurement, the
The result should be due to sample origin or technical device message appears.
reported to the Dr. factors. Result may not be correct.

When the color message is shown, please check the flag screen.
Please re-measure the sample again to confirm at least.
Reagent Management

Replace Reagents (*)Example of diluent

1 Replace the empty reagent bottle with a new one.

2 From the [Home] screen, Touch [Manage Reagents] key.

3 Read the barcode of the


reagent to be replaced.

(*) Multiple reagents can be replaced.

Lot number and expiration date are


automatically entered.
4 Touch [Replace] key. Press [Yes]. It
takes about 2
minutes.
(*)Refer to the operation manual for reagent management.

Empty Waste container

1 Empty the waste container.

2 From the [Home] screen, touch [Manage Reagents] key.

3 Check the box of “WASTE”, and touch [Replace].

Press [Yes]
Maintenance

Recommended
Self-Check : once a day
Background check : once a day Before the inspection begins.
QC : once a day
Cleaning : once a day After the inspection.
(it’s included power off function)

Clean Protein

1 From the Home screen, touch [Maintenance]


[Maintenance Operation] key.
Perform when the message
2 Touch [Clean Protein].
of “Remove Protein” is
displayed or once a month.

Press [Yes]. It
takes about 20
minutes.

Maintenance parts
Every 6,000 measurements Every 24,000 measurements

x2

Power OFF

Press the power switch on the front.


The instrument perform Cleaning (about 10 minutes)
automatically, after that the power will be OFF.
(When the protein cleaning period comes, it will be done automatically.)

If you have any problem, Call :


The person in charge: ___________________
30th May 2020