07:59 18/08/2023 Reset TnT Data


Topic ID: id_26824230
Version: 2.0
Date: Dec 23, 2020 9:10:38 PM

Reset TnT Data

Prerequisites
Table 1. Personnel requirements
Required persons Preliminary Procedure Finalization
requirements
1 N/A 15 minutes N/A

Table 2. Required conditions


Condition Reference Effectivity
- -
Recommend procedure completion
by GE Service trained personnel.

- -
Confirm [120 VAC], [AXIAL DRIVE
ENABLE], and [HVDC ENABLE]
switches on Service Switch Panel are
turned on.

- -
Confirm Gantry Controls Table
Collision Sensor Indicator has been
reset.

- -
Confirm Scan Hardware has been
successfully reset.

- -
Confirm no flashing indicators on
Gantry Controls or Display.

Overview

This document defines the procedures to reset the Generator Tracking and Trending data and filament aging
calibration to factory values.

Last Revised: 29 December, 2014

Procedure

file:///G:/My Drive/Lab-N.Đăng/1.Việc_Chính/3.Library/3.CT-GE/GE-Revolution ACTs/index_frames.html 1/2


07:59 18/08/2023 Reset TnT Data

1. On the upper right corner of the Desktop, click to select SERVICE MODE.
2. Select DIAGNOSTICS from Service Desktop.
3. Select GENERATOR TOOL, or X-RAY GENERATION, GENERATOR TOOL.
4. Click RESET TNT DATA.
5. Select data areas to be re-initialized.

a. By default both TUBE TNT DATA and GENERATOR TNT DATA are highlighted.
b. If GENERATOR TNT DATA is highlighted then TUBE TNT DATA will always be highlighted.
c. If GENERATOR TNT DATA is not selected then TUBE TNT DATA can either be selected or un-selected.
d. For tube replacement, un-select GENERATOR TNT DATA, select (highlight) TUBE TNT DATA.
e. For all other procedures, select (highlight) GENERATOR TNT DATA.

6. Click RESET.
7. Click OK in Attention window.
8. Click CANCEL in Reset TnT Data window.

NOTE:
If process fails, take appropriate corrective action and repeat procedure. If the process fails persistently, perform
Initialize Generator Database.

9. Click DISMISS in Generator Tool window.

Finalization
1. Filament Aging Calibration
2. Save System State.