Professional Documents
Culture Documents
QS 3.5 tcm220-24023
QS 3.5 tcm220-24023
AGFA HEALTHCARE
DICOM Conformance Statement
CR QS 3.5
Status: Released
Document Information
Issued by: Agfa shall not be liable for errors contained herein or for incidental or
Agfa HealthCare consequential damages in connection with the furnishing, performance
SIV Connectivity or use of this publication. Agfa reserves the right to revise this
publication and to make changes to its content at any time, without
Septestraat 27
obligation to notify any person or entity of such revisions and changes.
B-2640 Mortsel This publication may only be used in connection with the promotion,
Belgium sales, installation and use of Agfa equipment.
Copyright ã March, 07
tel: 32.3.444.7338 Agfa HealthCare
email: connectivity@agfa.com All rights reserved
HE/001076 Page 3 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
The CR QS, implements the necessary DICOM services to facilitate receiving of unprocessed CR images
from an AGFA Digitizer, performing image processing and sending the processed images to a remote
storage or print device over a Medical Imaging Systems network. The QS may also acquire patient
information from a Radiology Information System (RIS) for use in identifying processed images.
Media Storage Application Profile Write Files (FSC or FSU) Read Files (FSR)
Compact Disk – Recordable
General Purpose CD-R Yes, if license is enabled No
HE/001076 Page 4 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Table of Contents
1 Introduction ............................................................................................7
1.1 Revision Record ................................................................................................... 7
1.2 Purpose and Intended Audience of this Document ................................................ 7
1.3 General Remarks ................................................................................................. 7
1.3.1 Integration and Validation Activities.................................................................. 7
1.3.2 Future Evolution .............................................................................................. 7
1.4 Acronyms and Abbreviations ................................................................................ 8
1.5 Related Documents .............................................................................................. 8
2 Networking .............................................................................................9
2.1 Implementation Model .......................................................................................... 9
2.1.1 Application Data Flow Diagram ........................................................................ 9
2.1.2 Functional Definitions of AE’s ........................................................................ 10
2.1.2.1 Functional Capability of DICOM Store AE ................................................. 10
2.1.2.2 Functional Capability of DICOM Print AE .................................................. 10
2.1.2.3 Functional Capability of RISLINK AE......................................................... 10
2.1.3 Sequencing of Real World Activities............................................................... 11
2.2 AE Specifications................................................................................................ 11
2.2.1 DICOM Store AE Specification....................................................................... 11
2.2.1.1 SOP Classes Supported ........................................................................... 11
2.2.1.2 Association Establishment Policies ........................................................... 12
2.2.1.2.1 General ............................................................................................... 12
2.2.1.2.2 Number of Associations ....................................................................... 12
2.2.1.2.3 Asynchronous Nature .......................................................................... 12
2.2.1.2.4 Implementation Identifying Information................................................. 12
2.2.1.3 Association Initiation Policies .................................................................... 13
2.2.1.3.1 Transmit Study .................................................................................... 13
2.2.1.3.1.1 Description and Sequencing of Activity 13
2.2.1.3.1.2 Proposed Presentation Contexts 13
2.2.1.3.1.3 SOP Specific Conformance – CR Image Storage SOP Class 13
2.2.2 DICOM Print AE Specification........................................................................ 14
2.2.2.1 SOP Classes Supported ........................................................................... 14
2.2.2.2 Association Establishment Policies ........................................................... 15
2.2.2.2.1 General ............................................................................................... 15
2.2.2.2.2 Number of Associations ....................................................................... 15
2.2.2.2.3 Asynchronous Nature .......................................................................... 15
2.2.2.2.4 Implementation Identifying Information................................................. 15
2.2.2.3 Association Initiation Policies .................................................................... 15
2.2.2.3.1 Activity – Print sheets to remote Printer................................................ 15
2.2.2.3.1.1 Description and Sequencing of Activity 16
2.2.2.3.1.2 Proposed Presentation Contexts 17
2.2.2.3.1.3 Common SOP Specific Conformance for all Print SOP Classes 17
2.2.2.3.1.4 SOP Specific Conformance – Printer SOP Class 17
2.2.2.3.1.5 SOP Specific Conformance – Film Session SOP Class 18
2.2.2.3.1.6 SOP Specific Conformance – Film Box SOP Class 19
2.2.2.3.1.6.1 Film Box SOP Class Operations (N-CREATE)............................ 19
2.2.2.3.1.6.2 Film Box SOP Class Operations (N-ACTION)............................. 20
2.2.2.3.1.7 SOP Specific Conformance – Image Box SOP Class 22
HE/001076 Page 5 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
5 Security ................................................................................................35
5.1 Security Profiles.................................................................................................. 35
5.2 Association Level Security .................................................................................. 35
5.3 Application Level Security................................................................................... 35
6 Annexes ...............................................................................................36
6.1 IOD Contents...................................................................................................... 36
6.1.1 Created SOP Instance(s)............................................................................... 36
6.1.2 Attribute Mapping .......................................................................................... 42
HE/001076 Page 6 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
1 INTRODUCTION
The user of this document is involved with system integration and/or software design. We
assume that the reader is familiar with the terminology and concepts that are used in the
DICOM 3.0 standard and the IHE Technical Framework.
Readers not familiar with DICOM 3.0 terminology should first read the appropriate parts of
the DICOM standard itself, prior to reading this conformance statement.
Although the use of this conformance statement in conjunction with the DICOM 3.0 standard
is intended to facilitate communication with Agfa CR imaging equipment, it is not sufficient to
guarantee, by itself, the inter-operation of the connection. The following issues need to be
considered:
2 NETWORKING
Transmit Study
Remote AE (Pacs)
DICOM Store
receives images from
Application Entity the study
Study with
autorouting is
completed &
released
Print sheets
Remote AE (RIS)
RISLINK
Reload Worklist provides the worklist
Application Entity items
Dicom Standard
Interface
Ø The DICOM Store AE sends images to a remote AE (e.g a PACS system). It can be
activated manually (Transmit Study) or automatically (Auto routing on completion of a
study)
Ø The DICOM Print AE prints sheets on a remote printer. It can be activated manually
(Print Study or via Print Composer) or automatically (Auto printing on completion of a
study)
Ø The RISLINK AE receives Worklist information from a remote AE (e.g. a RIS system).
It is activated manually with a Query RIS request in ID Viewer
3. Identify the exposed cassette(s) for the study and digitize them
2.2 AE Specifications
2.2.1.2.1 General
The DICOM standard Application context is always proposed:
Table 2.2-6: DICOM implementation Class and Version for DICOM Store AE
Warning Coercion of Data B000 The routingstatus of the study will be set to done.
Elements A warning will be logged
Data Set does not B007 The routingstatus of the study will be set to done.
match SOP Class A warning will be logged
Elements Discarded B006 The routinstatus of the study will be set to done
A warning will be logged
Failure Refused: Out of A7xx The routingstatus of the study will bet set to failed.
Resources An error will be logged.
Error: Data Set does A9xx The routingstatus of the study will be set to failed
not match SOP Class An error will be logged
Error: Cannot Cxxx The routingstatus of the study will be set to failed
understand An error will be logged
Exception Behavior
Timeout The routingstatus of the study will be set to failed. An error will be logged..
Association aborted by The routingstatus of the study will be set to failed. An error will be logged.
SCP or network layers
2.2.2.2.1 General
Table 2.2-11: DICOM Application Context
The Print Jobs are created either automatically with auto printing or when the user
manually asks prints (e.g. via Print Composer). The film sheets are internally
processed and rendered to STANDARD/1,1 pages. Then these pages will be sent
from QS to Printer.
HE/001076 Page 16 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
DICOM
Printer
Print
Open Association
N-GET-RQ (Printer)
N-GET-RSP
N-ACTION-RQ
N-ACTION-RSP
N-DELETE-RQ
N-DELETE-RSP
Close Association
A typical sequence of DIMSE messages sent over an association between the DICOM
Print AE and a Printer is illustrated in Figure 2.2-2:
Note :
The Presentation LUT SOP Class is only proposed when the printer is configured with Rescale Type
“Perception”
Exception Behavior
Timeout An error is logged in the event viewer and shown to the user. The job gets an error status
in the windows spooler.
Association aborted by An error is logged in the event viewer and shown to the user. The job gets an error status
SCP or network layers in the windows spooler.
FAILURE An error (Printer Status (2110,0010)=FAILURE!) is logged in the event viewer and shown to
the user. The job gets an error status in the windows spooler.
The attributes supplied in an N-CREATE request are listed in the table below:
Table 2.2-20: Film Session SOP CLASS N-CREATE Response Status Handling Behavior
Table 2.2-22: Film Box SOP CLASS N-CREATE Response Status Handling Behavior
Table 2.2-24: Film Box SOP CLASS N-CREATE Response Status Handling Behavior
Failure Unable to create C602 The association is aborted. An error is logged in the event
Print Job SOP viewer and shown to the users. The job gets an error
Instance; print status in the windows spooler
queue is full
Table 2.2-26: Presentation LUT SOP CLASS N-CREATE Response Status Handling Behavior
Failure Image size is The association is aborted. An error is logged in the event
larger than image viewer and shown to the users. The job gets an error
box size status in the windows spooler
Failure Insufficient
memory in printer
to store the image
Table 2.2-28: Presentation LUT SOP CLASS N-CREATE Response Status Handling Behavior
* * Any other
status code
2.2.3.2.1 General
The DICOM standard Application context shall be specified.
Exception Behavior
Timeout A message box showing the error message is displayed.
Acquired images will always use the Study Instance UID specified for the
Scheduled Procedure Step (if available). If an acquisition is unscheduled, a Study
Instance UID will be generated locally.
The table below provides a description of the Worklist Request identifiers and
specifies the attributes that are copied into the image.
2.4 Configuration
The configuration is done in the CR QS ConfigView.
HE/001076 Page 30 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Multiple remote Store SCPs can be configured in the ConfigView under Devices ->
Destination.
Multiple remote Print SCPs can be configured in the ConfigView under Devices ->
Medical Printers.
3 MEDIA INTERCHANGE
Ø The DICOM Export Application Entity exports images to a CD-R Storage medium. It is
associated with the “DICOM Export” that can be activated in IPD Selector for the selected
studies
3.2 AE Specifications
Table 3.2-1: Application Profiles, Real World Activities and Roles for DICOM Export AE
Table 3.2-2: IODs, SOP Classes and Transfer Syntaxes for DICOM Export AE
Information Object Definition SOP Class UID Transfer Syntax Transfer Syntax UID
5 SECURITY
The CR QS can be configured to use TLS (for authentication and data integrity) on its
connections to a PACS system.
6 ANNEXES
The following tables use a number of abbreviations. In the “Presence of…” column:
Ø VNAP Value Not Always Present (attribute sent zero length if no value is present)
Ø ANAP Attribute Not Always Present
Ø ALWAYS Always Present with a value
Presence of
Attribute Name Tag VR Value Source
Value
Patient IE
Patient Module
Patient’s Name (0010,0010) PN VNAP USER or
MWL
Patient ID (0010,0020) LO ALWAYS USER, MWL
or AUTO
Patient’s Birth (0010,0030) DA ALWAYS USER or
Date MWL
Patient’s Sex (0010,0040) CS ALWAYS USER or
MWL
Other Patient (0010,1000) LO VNAP MWL
IDs
Ethnic Group (0010,2160) SH VNAP MWL
Patient (0010,4000) LT (Limited to 64 characters) VNAP USER or
Comments MWL
Study IE
General Study
Study Instance (0020,000D) UI ALWAYS MWL or
UID AUTO
Study Date (0008,0020) DA ALWAYS AUTO
Study Time (0008,0030) TM ALWAYS AUTO
Accession (0008,0050) SH ALWAYS USER, MWL
Number or AUTO
Referring (0008,0090) PN VNAP MWL
Physician’s
Name
Name of (0008,1060) PN VNAP AUTO
Physician(s)
Reading Study
Study ID (0020,0010) SH VNAP AUTO
Study (0008,1030) LO Study type as selected in ID viewer ALWAYS AUTO
Description
Patient Study
Admitting (0008,1080) LO VNAP MWL
Diagnoses
Description
Patient’s Size (0010,1020) DS VNAP MWL
Patient’s Weight (0010,1030) DS VNAP MWL
Occupation (0010,2180) SH VNAP MWL
Additional (0010,21B0) LT VNAP MWL
Patient History
Series IE
General Series
Modality (0008, 0060) CS CR ALWAYS AUTO
Series Instance (0020,000E) UI ALWAYS AUTO
UID
Series Number (0020,0011) IS ALWAYS AUTO
HE/001076 Page 38 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Presence of
Attribute Name Tag VR Value Source
Value
Patient Position (0018,5100) CS VNAP ?
Laterality (0020,0060) CS VNAP ?
Series (0008,103E) LO Concatenation of Substudy and exposure ALWAYS AUTO
Description
Performing (0008,1050) PN VNAP AUTO
Physician’s
Name
Operator’s (0008,1070) PN VNAP AUTO
Name
CR Series Module
Body Part (0018,0015) CS Automatically filled in with the Body Part VNAP AUTO
Examined that was configured for the taken
exposure in the exam tree. The list of
body parts can be extended in the config
view
View Position (0018,5101) CS Automatically filled in with the View ALWAYS AUTO
Position that was configured for the taken
exposure in the exam tree. (AP, PA, LL,
RL)
Fitler Type (0018,1160) SH VNAP AUTO
Collimator/Grid (0018,1180) SH VNAP AUTO
Name
Focal Spot (0018,1190) DS VNAP AUTO
Plate Type (0018,1260) SH ALWAYS AUTO
Equipment
General Equipment Module
Manufacturer (0008,0070) LO AGFA ALWAYS FIXED
Institution Name (0008,0080) LO ALWAYS CONFIG
Station Name (0008,1010) SH ALWAYS AUTO
Institutional (0008,1040) LO ALWAYS CONFIG
Department
Manufacturer’s (0008,1090) LO ALWAYS AUTO
Model Name
Device Serial (0018,1000) LO ALWAYS AUTO
Number
Software (0018,1020) LO ALWAYS AUTO
Version
Image
General Image Module
Instance (0020,0013) IS ALWAYS AUTO
Number
Patient (0020,0020) CS ALWAYS AUTO
Orientation
Image (0020,4000) LT ALWAYS USER
Comments
Image Type (0008,0008) CS ALWAYS AUTO
Acquisition Date (0008,0022) DA ALWAYS AUTO
Acquisition (0008,0032) TM ALWAYS AUTO
Time
Presentation (2050,0020) CS ANAP AUTO
LUT Shape
HE/001076 Page 39 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Presence of
Attribute Name Tag VR Value Source
Value
Image Pixel Module
Samples per (0028,0002) US 1 ALWAYS AUTO
pixel
Rows (0028,0010) US ALWAYS AUTO
Columns (0028,0011) US ALWAYS AUTO
Pixel Aspect (0028,0034) IS 11 ALWAYS AUTO
Ratio
Bits allocated (0028,0100) US 8 or 16 ALWAYS CONFIG
Bits Stored (0028,0101) US 8, 12 or 15 ALWAYS CONFIG
High bit (0028,0102) US ALWAYS AUTO
Pixel (0028,0103) US ALWAYS AUTO
Representation
Pixel Data (7FE0,0010) OW ALWAYS AUTO
or
OB
Contrast / Bolus Modules
Contrast/Bolus (0018,0010) LO VNAP AUTO
Agent
Contrast/Bolus (0018,1041) DS VNAP AUTO
Volume
Contrast/Bolus (0018,1044) DS VNAP AUTO
Total Dose
CR Image Module
KVP (0018,0060) DS VNAP AUTO
Plate ID (0018,1004) LO ALWAYS AUTO
Exposure Time (0018,1150) IS VNAP AUTO
X-Ray Tube (0018,1151) IS VNAP AUTO
current
Exposure in uAs (0018,1153) IS VNAP AUTO
Imager Pixel (0018,1164) DS VNAP AUTO
Spacing
Acquisition (0018,1401) LO VNAP AUTO
Device
Procession
Code
Cassette (0018,1402) CS ALWAYS AUTO
Orientation
Cassette Size (0018,1403) CS ALWAYS AUTO
Exposures on (0018,1404) US ALWAYS AUTO
Plate
Relative X-ray (0018,1405) IS Is equal to Lgm value in (0019,1015) * ALWAYS AUTO
Exposure 1000
Sensitivity (0018,6000) DS ALWAYS AUTO
Photometric (0028,0004) CS ALWAYS AUTO
Interpretation
Modality LUT Module
Rescale (0028,1052) DS VNAP AUTO
Intercept
Rescale Slope (0028,1053) DS VNAP AUTO
Rescale Type (0028,1054) LO OD, OD REL, LOG_E REL, P, P REL VNAP AUTO
VOI LUT Module (See note)
HE/001076 Page 40 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Presence of
Attribute Name Tag VR Value Source
Value
Window Width (0028,1051) DS ANAP AUTO
Window Center (0028,1050) DS ANAP AUTO
VOI LUT (0028,3010) DS ANAP AUTO
Sequence
>Lut Descriptor (0028,3002) US ANAP AUTO
>Lut Data (0028,3006) US ANAP AUTO
>Lut (0028,3003) LO ANAP AUTO
Explanation
SOP Common Module
Specific (0008,0005) DS ANAP CONFIG
Character Set
SOP Class UID (0008,0016) UI ALWAYS AUTO
SOP Instance (0008,0018) UI ALWAYS AUTO
UID
Instance (0008,0012) DA ALWAYS AUTO
Creation Date
Instance (0008,0013) TM ALWAYS AUTO
Creation Time
Additional Attributes
Medical Alerts (0010,2000) LO VNAP MWL
Contrast (0010,2110) LO VNAP MWL
Allergies
Estimated (0018,1114) DS VNAP AUTO
Radiographic
Magnification
Factor
Grid (0018,1166) CS VNAP AUTO
Anode Target (0018,1191) CS VNAP AUTO
Material
Body Part (0018,11A0) DS VNAP AUTO
Thickness
Compression (0018,11A2) DS VNAP AUTO
Force
Positioner (0018,1510) DS VNAP AUTO
Primary Angle
Filter Material (0018,7050) CS VNAP AUTO
Exposure (0018,7060) CS VNAP AUTO
Control Mode
Exposure (0018,7062) LT VANP AUTO
Control Mode
Description
Phototimer (0018,7065) DS VNAP AUTO
Setting
Pixel Spacing (0028,0030) DS VNAP AUTO
Special Needs (0038,0050) LO VNAP AUTO
Current Patient (0038,0300) LO VNAP AUTO
Location
Confidentiality (0040,3001) LO VNAP AUTO
Constraint on
Patient Data
Description
HE/001076 Page 41 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Presence of
Attribute Name Tag VR Value Source
Value
Entance Dose (0040,8302) DS VNAP AUTO
in mGy
Polarity (2020,0020) CS VNAP AUTO
Source Number (300A,0212) IS VNAP AUTO
Source Type (300A,0214) CS VNAP AUTO
Approval status (300E,0002) CS VNAP AUTO
Private Attributes
Private Creator (0019,00FF) SH AGFA ALWAYS AUTO
Image (0019,FF10) ST ALWAYS AUTO
processing
parameters
Identification (0019,FF11) LO ALWAYS AUTO
Data
Sensometry (0019,FF13) LO VNAP AUTO
Name
Dose monitoring (0019,FF15) LO VNAP AUTO
(Lgm)
Clipped (0019,FF1A) LO VNAP AUTO
Exposure
Deviation
Logarithmic PLT (0019,FF1B) LO VNAP AUTO
Full Scale
Nr of series (0019,FF60) US ALWAYS AUTO
ID station name (0019,FF62) SH VNAP AUTO
Number of (0019,FF65) US VNAP AUTO
Images in Study
to be
transmitted.
Only send if
autoverify is on
Nr of images (0019,FF70) US VNAP AUTO
Geometrical (0019,FF80) SH VNAP AUTO
Transformations
Roam Origin (0019,FF81) ST VNAP AUTO
Zoom factor (0019,FF82) US VNAP AUTO
Status (0019,FF93) CS VNAP AUTO
XRay (0019,FF98) LO VNAP AUTO
Equipment Type
HE/001076 Page 42 of 44
Document number 001076 Revision 1.2
NodeID Livelink : 9970858
Agfa HealthCare 8 March, 2007
Notes
Has the form “MENU=menu CC=cc EC=ec LR=lr NR=nr” where:
menu: an integer specifying the image processing number
cc: an integer specifying the collimation configuration
mc: a real number specifying the Musica contrast
ec: a real number specifying the edge contrast
lr: a real number specifying the latitude reduction
nr: a real number specifying the noise reduction
It is however possible in ID Viewer to map data from an arbitrary DICOM field to the exam
tree, in order to enable automatic study selection based on data coming from the RIS.
file://D:\sData\Adlib\Express\temp\9970858.5info.html 3/8/2007