You are on page 1of 8

Influence of Calibration on Densitometric

Studies of Emphysema Progression


Using Computed Tomography
David G. Parr, Berend C. Stoel, Jan Stolk, Peter G. Nightingale, and Robert A. Stockley

Lung Investigation Unit and Wellcome Trust Clinical Research Facility, Queen Elizabeth Hospital, Birmingham, United Kingdom; and Division
of Image Processing, Department of Radiology, and Department of Pulmonology, Leiden University Medical Center, Leiden, The Netherlands

The fundamental importance of calibration for any measuring de- voxels. The “voxel index” (VI) method calculates the proportion
vice is indisputable, but computed tomography (CT) calibration of low-attenuation lung voxels within a desired range, and al-
in longitudinal lung densitometry studies is largely unexplored. though various voxel index thresholds have been employed and
Although the validity of CT as a measure of emphysema has been validated against pathology (4–7), it is likely that no single thresh-
confirmed in cross-sectional studies, there are limited data on long- old is appropriate universally.
term reproducibility, and this is critically important for validating The fundamental importance of calibration in any measuring
its use as an outcome measure in therapeutic trials. A general under- device is indisputable, yet few studies have addressed the issue
standing of the strengths and pitfalls of CT densitometry is critical
of CT calibration in lung densitometry. Short-term reproducibil-
for physicians reviewing the published literature using this method-
ity has been demonstrated (8–10), but there are limited published
ology. In our study of 57 patients with alpha-1 antitrypsin deficiency
data on long-term reproducibility (11). It is recognized that X-ray
(phenotype PiZ), progression of voxel index determined from three
successive annual scans acquired with a fully calibrated scanner was
tube ageing and replacement may introduce data acquisition
intimately associated with changes in CT air densitometry, sampled errors (11), but although accurate scanner calibration is man-
from patient images. Images were therefore reanalyzed, using a datory, procedures of standardization have yet to be defined.
correction technique validated in phantom studies that adjusted Previous studies have used water phantoms alone for quality
for changes in measured air density, and the reliability of the voxel assurance (4, 8), but in a cross-sectional study to evaluate confor-
index as a measure of emphysema progression was improved. mity between scanners, Kemerink and coworkers (12) demon-
Comparison of adjusted voxel index thresholds indicated the opti- strated that although water densitometry was consistent, the
mum threshold was –950 Hounsfield units. Internal air calibration is Hounsfield number for air varied between scanners. A method
therefore critical in longitudinal and multicenter lung densitometry was proposed for correcting the CT numbers for a scanner with
studies of emphysema and incorporation of a correction factor is poor air calibration and its potential use as a means of standardiz-
essential for quantitative image analysis. ing densitometry was described (12). However, although stan-
dardization has been recognized to be crucially important (13),
Keywords: alpha-1 antitrypsin deficiency; emphysema; lung densitometry the influence of air calibration errors on lung densitometry stud-
ies in a clinical setting has yet to be addressed systematically.
Projected estimates of the global burden of chronic obstructive
Such information is crucial to determine the validity and hence
pulmonary disease have led to recognition of the need for the
interpretation of observational and interventional studies.
development of novel therapies to modify disease progression. In
The provisional assessment of CT progression in a sample of
routine clinical practice, physiologic measures such as diffusion
patients attending our program had unexpectedly suggested an
capacity and airflow obstruction are traditionally employed as
increase in lung density (consistent with resolution of emphy-
surrogate measures of emphysema severity. However, practical
sema) in a limited number of serial annual scans acquired in the
difficulties arising from the limitations of these conventional
early months of 2000 (L. Dowson, personal communication).
methods (1) necessitate the development of alternative outcome
The current study was therefore undertaken to review the consis-
measures for use in clinical trials (2).
In contrast to chronic obstructive pulmonary disease, which tency of CT densitometry for the measurement of emphysema
is defined in physiologic terms, emphysema is defined pathologi- progression in a cohort of patients with alpha-1 antitrypsin defi-
cally (3). Nevertheless, it may now be diagnosed accurately and ciency (AATD) and to standardize densitometry measurements
quantified noninvasively by computed tomography (CT). The by adjusting for any identifiable scanner errors. Validation of the
loss of lung tissue associated with emphysema can be measured methodology was performed in studies with an anthropomorphic
as a reduction in lung density by CT, using various parame- phantom. In addition, the opportunity was taken to review data
ters derived from the frequency distribution histogram of lung from the Dutch–Danish controlled trial (14) relating to scanner
calibration.
Some of the results of these studies have been reported pre-
viously in the form of an abstract (15).

(Received in original form March 11, 2004; accepted in final form July 21, 2004)
METHODS
ADAPT program supported by a noncommercial grant from Bayer (UK); use of
Pulmo-CMS software supported by European Union funding (grant RNDV.07773). Data records of 57 patients with AATD (PiZ) were selected from
Correspondence and requests for reprints should be addressed to R. A. Stockley, among those attending our program (see the online supplement) to
M.D., F.R.C.P., D.Sc., Lung Investigation Unit, First Floor, Nuffield House, Queen include all subjects who had completed three consecutive annual assess-
Elizabeth Hospital, Birmingham B15 2TH, UK. E-mail: r.a.stockley@bham.ac.uk ments over 2 years spanning the months of concern in early 2000 (see
This article has an online supplement, which is accessible from this issue’s table Introduction). Alpha-1 antitrypsin level and phenotype were verified
of contents online at www.atsjournals.org by immunoassay and isoelectric focusing, respectively, using a dried
Am J Respir Crit Care Med Vol 170. pp 883–890, 2004
finger-prick blood spot (Heredilab, Salt Lake City, UT). The program
Originally Published in Press as DOI: 10.1164/rccm.200403-326OC on July 21, 2004 was approved by the local research ethics committee, and all subjects
Internet address: www.atsjournals.org gave written informed consent.
884 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL 170 2004

Annual CT images were acquired on a General Electric Prospeed ity, by Mann–Whitney test. Comparison of the baseline and Year 2
scanner (General Electric Medical Systems, Milwaukee, WI), using a scans at the different thresholds was done with a Wilcoxon signed rank
high-resolution protocol (120 kVp [kilovolt peak], 200 mA · s [dose test for paired data. Differences were considered significant at p ⬍
rate], 1-mm collimation, bone reconstruction algorithm) on inspiration 0.05.
and expiration in the supine position as described previously (16).
The voxel indices (VI) (⫺910 Hounsfield units [HU]) were calculated
for upper (level of the aortic arch) and lower (level of the inferior
RESULTS
pulmonary veins) zone images, using the manufacturer’s Density Mask Baseline Characteristics
program.
The methodology for the current study comprised two steps before The baseline lung function and voxel indices (uncorrected thresh-
the analysis of patient data with an additional phantom study to validate old, ⫺910 HU; expressed as percentage) at baseline are shown in
the technique employed for the patient data. Table 1.
1. Error identification. Consistency of lung densitometry was ex-
plored by plotting annual VI change against scan date. Air mea- Error Identification–Discontinuity Analysis
surements acquired retrospectively from patient images, using A plot of annual VI change against scan date showed a trend
region of interest (ROI) densitometry plus weekly quality assur- occurring in March 2000 of reversed VI progression, implying
ance records of water and bone phantom measurements, were
disease improvement (Figure 1A). The best fitting model for
related to date. Both data sets were assessed for discontinuity to
identify any coincident events. voxel index progression placed this discontinuity (see the online
See Dirksen and coworkers (14), and the online supplement, for supplement) just before March 6, 2000 and produced an esti-
an analysis of the Dutch–Danish controlled trial. mated shift of 4.5% (p ⬍ 0.001). Quality assurance records of
2. Development of a correction method. Air, water, and bone den- scanner calibration showed that measurements using the manu-
sity values were compared to characterize the behavior of the facturer’s water phantom were well within the manufacturer’s
scanner over time across a wide range of density values. A correc- tolerance of 0 ⫾ 10 HU, but a concomitant shift in air density
tion equation was developed to adjust lung densitometry for air measured from patient images was observed (Figure 2A). Re-
calibration errors by predicting the potential influence of these gression analysis of the air calibration data indicated discontinu-
three measures on the VI threshold. ity with a positive shift in air number that occurred between the
Validation Studies Using a Lung Phantom observation on March 6 and the following observation on March
11, 2000 (p ⬍ 0.001) (Figure 2A). In addition, an opposing
Correction technique for longitudinal data. An anthropomorphic lung
drift of decreasing air number also existed throughout the study
phantom was scanned, using the same protocol as in the clinical study,
using a Marconi MxView scanner (Philips Medical Systems, Andover, period, described by the gradient b in the equation y ⫽ a ⫹
MA) with the facility to adjust the CT number for air. Lung densitome- bx1 ⫹ cx2 (see the online supplement). There was no difference
try was calculated for a single slice, using Pulmo-CMS software (MEDIS between the gradient of drift before and after the discontinuity
[Medical Imaging Systems], Leiden, The Netherlands) at different cali- (p ⫽ 0.77) and, after elimination of the discontinuity, the overall
bration settings without the threshold adjustment and applying the gradient drift was significantly different from zero (p ⬍ 0.001)
correction equation to the data to determine their efficacy. (Figure 2B).
Correction technique for cross-sectional multicenter data. Calibra- Measurement of blood density in the descending thoracic
tion and conformity of six third-generation multislice scanners (two
aorta did not identify any time-related trend (data not shown).
General Electric Lightspeed scanners, one Philips MxView, one Philips
MX 8000, one Toshiba Asteion, and one Siemens Sensation 16) were
Dutch–Danish Study
explored, using the dog phantom. The scanning protocol employed
was that proposed for future clinical densitometric studies (140 kVp; A similar drift but no obvious discontinuity was seen by analyzing
40 mA; 5-mm collimation; pitch, 1.5; 2.5-mm increment; field of view, ROI air measurements in the Dutch patients included in the
500 mm; and a smooth reconstruction filter) (17). Air calibration mea- Dutch–Danish study of augmentation therapy (14). The individ-
surements were acquired as described above and surrogate measures ual patient data are shown in Figure 3.
of blood density were derived from ROI measurements of the phantom
heart.
Whole lung VI of ⫺910 HU was calculated for each series with and
without the use of an internal calibration procedure and the effect on
standardization was assessed by the coefficient of variation. TABLE 1. BASELINE CHARACTERISTICS
n Absolute Value* % Predicted*
Analysis of Clinical Images: Employing the Validated
Correction Technique Age, yr 57 50 (46–61) NA
FEV1,† L 57 1.34 (0.97–2.08) 42 (30–67)
Upper-zone inspiratory images obtained previously were reanalyzed, VC,† L 57 3.82 (3.12–4.49) 98 (90–127)
using the “adjusted” threshold of ⫺910 HU. Comparison was made of RV,† L 57 2.66 (2.06–3.44) 135 (100–170)
the annual VI progression rate, using uncorrected and adjusted thresh- DLCO, mmol/min/kPa 57 5.90 (4.62–7.64) 63 (47–84)
olds for three groups of consecutive annual scans: (1 ) scans predating DLCO/VA, L mmol/min/kPa/L 57 1.03 (0.73–1.31) 67 (46–83)
the discontinuity, (2 ) scans spanning the discontinuity, and (3 ) scans UZI voxel index, –910 HU 57 34.2 (21.6–44.6) NA
following the discontinuity. UZE voxel index, –910 HU 57 19.0 (7.7–34.2) NA
In addition, the VI was calculated for corrected images, using 20-HU LZI voxel index, –910 HU 57 55.3 (33.2–64.5) NA
threshold increments from ⫺870 to ⫺950 HU and comparison of pro- LZE voxel index, –910 HU 57 47.0 (26.4–57) NA
gression over 2 years was performed.
Definition of abbreviations: DLCO ⫽ single-breath diffusing capacity of the lung
Statistical Analysis for carbon monoxide (gas transfer); DLCO/VA ⫽ DLCO normalized per liter alveolar
volume (transfer coefficient); HU ⫽ Hounsfield unit; LZE ⫽ lower-zone expiratory
Data were analyzed with a statistical software package (Statistical Pack- high-resolution computed tomography (HRCT) image; LZI ⫽ lower-zone inspira-
age for the Social Sciences [SPSS] version 10.0.5; SPSS, Chicago, IL). tory HRCT image; NA ⫽ not applicable; RV ⫽ residual volume; UZE ⫽ upper-
Air calibration was assessed with discontinuity analysis and regression zone expiratory HRCT image; UZI ⫽ upper-zone inspiratory HRCT image.
slopes were compared with the F test. The rate of CT progression in * Median (interquartile range).
the three subgroups of scan pairs was compared by Kruskal–Wallis test †
Postdual bronchodilation with nebulized salbutamol (5 mg) and ipratropium
and in the two subgroups of scan pairs, before and after the discontinu- bromide (500 ␮g).
Parr, Stoel, Stolk, et al.: Calibration in CT Lung Densitometry 885

Figure 2. (A ) CT numbers for air (open circles) and water (solid circles)
over time. CT numbers for water (taken from region of interest [ROI]
Figure 1. Annual progression rate in upper-zone inspiratory (UZI) scans measurements of quality assurance water phantom) show consistency
at a threshold of ⫺910 HU before (A ) and after (B ) correction for air over time. In comparison, the trend in ROI measurements of ambient
calibration. Solid triangles ⫽ voxel index (VI) progression between base- air (taken retrospectively from apical images) shows a gradual long-
line and the first year; open circles ⫽ voxel index progression between term negative drift, interrupted by a positive shift occurring in March
the first and second years. Emphysema progression is indicated by points 2000. Discontinuity analysis of air number data, using the equation
lying above zero. The arrows indicate the period when the reversal of y ⫽ a ⫹ bx1 ⫹ cx2, gives a value for x1 of 6/3/00 to 11/3/00 (p ⬍
the trend (emphysema improvement) was suspected. 0.001), where y is air density value, x1 is scan date (measured as days
since 31/12/1899), x2 is 0 before shift and 1 after the air calibration
shift, a is the y intercept, b is gradient (or “drift”), and c is the shift in
air calibration. Comparison of the gradients of negative drift before and
Development of a Correction Method after the discontinuity by residual sum of squares showed no significant
difference (p ⫽ 0.77). (B ) Plot of ROI measurements for air, adjusted
Changes in air, water, and bone density values indicated that
to eliminate the effect of the shift (c). The gradient (b) is significantly
the effect of an increase in air number would be to increase
different from zero (p ⬍ 0.001).
proportionally the Hounsfield number of those voxels with a
density intermediate between water and air (see Figure 4A).
Therefore, correction for air calibration errors on VI thresh-
old could be achieved with the following formula: corrected
air calibration, using the method described here and used in the
threshold ⫽ (air calibration value ⫺ water calibration value /
main study, but the use of blood calibration did not reduce
⫺1,000) ⫻ index threshold (see Figure 4B).
variability (see Table 3).
Validation Studies with Lung Phantom
CT Progression: Application of the Correction Method
Correction technique for longitudinal data. The distribution of
voxels with the dog lung phantom was similar to that seen in The observed reversal in annual voxel index progression that
patients with AATD (data not shown), suggesting that it was a occurred in the patient cohort was eliminated on use of the
good model for emphysematous lung. Adjusting the scanner air correction technique (see Figure 1B). For the uncorrected data,
calibration (see Methods) resulted in a proportional shift in the the progression rate for scan pairs spanning the discontinuity
lung voxel distribution histogram and marked changes to the was significantly different (p ⬍ 0.001) from that of the other
voxel index (⫺910 HU) as predicted from the studies described two groups, in which the scan pair either predated or postdated
above. Application of the correction technique reproduced the the discontinuity (see Figure 5A) and suggests an overall im-
same results at each calibration setting (see Table 2). provement over this time period. This difference was abolished
Correction technique for cross-sectional multicenter data. In- when the corrected data for the three groups were compared
terscanner variability for air densitometry was greater than for (p ⫽ 0.13) (see Figure 5B). However, there was no significant
blood densitometry and there was wide discrepancy in the mea- difference in VI progression rate for scan pairs obtained either
surement of dog lung voxel indices (VI, ⫺910 HU). There was before or after the discontinuity in both the uncorrected (p ⫽
a significant reduction in this discrepancy by standardizing for 0.92) and corrected data (p ⫽ 0.72) (see Figures 5A and 5B).
886 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL 170 2004

TABLE 2. SINGLE IMAGE VOXEL INDEX AT A THRESHOLD


OF ⫺910 HU FOR DOG LUNG PHANTOM
Adjusted Air Calibration

⫺1,000 HU ⫺975 HU ⫺950 HU ⫺925 HU ⫺900 HU

Uncorrected VI 30.27 7.91 1.05 0.08 0.00


Corrected VI NA 30.29 30.43 30.74 30.90

Definition of abbreviations: NA ⫽ not applicable; VI ⫽ voxel index.


Table shows the effects of changing air calibration and the corrected value
achieved through application of the formula presented in text.

cally significant only for the upper-zone inspiratory scans (see


Figure 6A) as shown previously (8). However, after adjustment
with correction for air calibration, significant progression at the
Figure 3. Air calibration data for the Philips Tomoscan SR7000 scanner threshold of ⫺910 HU was confirmed in both upper- and lower-
used in the Dutch–Danish trial of alpha-1 antitrypsin augmentation zone scans (see Figure 6B). The annual VI progression rates at
therapy (14). ROI measurements were acquired from patient images a corrected threshold of ⫺910 HU are shown in Table 4, ex-
as described for the current study (see METHODS and online supplement). pressed as median and interquartile range, with data from the
Air density values (in Hounsfield units) are expressed as the difference previous study (8) for comparison.
from baseline values and results from individual patients are shown by The upper-zone voxel index showed significant progression
the joined lines. at all thresholds used, with marginal statistical superiority of
the inspiratory images over the expiratory images (Figure 6A).
Furthermore, in the lower-zone scans the corrected expiratory
images now also showed significant progression whatever the
Adjusting the analysis threshold for air calibration also im- threshold and the corrected inspiratory images demonstrated
proved the detection of progression over 2 years as shown by a significant progression at thresholds of ⫺950, –930, and ⫺910 HU
comparison of Z statistic (see Figure 6). Progression of the voxel (Figure 6B). Although all thresholds showed highly signifi-
index at the uncorrected threshold of ⫺910 HU was statisti- cant progression of VI, the Z statistic was best for ⫺950 HU,

Figure 4. Plot of actual CT numbers for air, water phantom,


and bone phantom on a single occasion, demonstrating
correct water densitometry but incorrect values for air and
bone. Perfect calibration values are shown for comparison.
(A ) All data points; (B ) the influence on low-density values,
assuming linearity between air and water. The effect of a
shift in air calibration of ⫹30 to ⫺970 HU on an intended VI
threshold of ⫺910 HU is shown by the arrow, demonstrating
how this adjusted threshold can be calculated, assuming a
water value of zero, using the following formula: corrected
threshold ⫽ (air calibration value ⫺ water calibration value/
⫺1,000) ⫻ index threshold.
Parr, Stoel, Stolk, et al.: Calibration in CT Lung Densitometry 887

TABLE 3. MEAN, STANDARD DEVIATION, AND COEFFICIENT OF VARIATION FOR BLOOD


AND AIR DENSITY VALUES MEASURED ON SIX DIFFERENT SCANNERS
VI, ⫺910 HU

Blood Density (HU) Air Density (HU) No Calibration Blood Calibration Air Calibration

Mean 41.09 ⫺1,005.1 36.23 35.56 34.02


Standard deviation 5.36 13.73 8.95 9.27 3.33
Coefficient of variation 13.04 1.37 24.70 26.06 9.79

Voxel index values (VI, ⫺910 HU) for whole dog lung are shown, demonstrating the influence of standardization on scanner
variability by calibrating on the basis of either blood or air densitometry. The latter resulted in an almost threefold reduction in
variability.

suggesting it is the most appropriate for assessing progression that voxel index progression in a cohort of patients with AATD
among the patients studied here. was associated with changes in air density measurements ob-
tained retrospectively from patient images. A clear shift occurred
DISCUSSION in air calibration in March 2000 that produced an obvious rever-
sal in voxel index progression that not only aroused our initial
The current study demonstrates the importance of precise scan- suspicion of an error, but also allowed us to causally relate the
ner and image calibration in longitudinal lung densitometry stud- two events. An isolated drift in air calibration would probably
ies for the measurement of emphysema progression. We found have remained undetected without specific observation using an
ROI calibration and the effect on VI progression would have
been one of overestimation. No significant difference was found
between the rate of air calibration drift before and after March
2000 and this is reflected in the comparable rates of apparent
VI progression during these periods.
The underlying cause of the changes to our scanner calibra-
tion remains uncertain despite careful inspection of the service
records. Although the most likely cause for our observations is
the replacement of the X-ray tube and detectors, which occurred
in January 2000, no definite explanation can be offered for the
calibration errors. Importantly, the changes in measured air den-
sity occurred despite the apparent stability of CT calibration
when defined by the CT numbers for air and water obtained
during regular routine scanner maintenance and from the weekly
quality assurance measurements using the manufacturer’s water
phantom. A similar finding has previously been reported in a
longitudinal study documenting changes in measurements of
aortic blood density, where full routine scanner calibration was
documented monthly (11). Measurements obtained from patient
images therefore provide essential calibration information in
addition to routine procedures in which densitometry is applied
to simple water phantoms or air measurements obtained with
an empty gantry.
The scanner used in our study was a third-generation single-
slice helical CT. Although technological improvements and the
incorporation of features such as daily autocalibration into mod-
ern scanners may improve the accuracy and precision of lung
densitometry studies, these procedures ensure only a value of
⫺1,000 HU for air with an empty gantry, which may engender
its own problems. For instance, using the technique described
Figure 5. Annual progression in upper-zone inspiratory voxel index in the current study, we have identified a similar drift in air
(threshold, ⫺910 HU) (UZI VI) before (A ) and after (B ) adjustment for calibration in other scanners, including the Philips Tomoscan
air calibration. Data are shown as the median and interquartile range. SR7000 scanner (see Figure 3) that was used in the Dutch–Danish
p Values relate to the Kruskal–Wallis test comparing VI progression rate
trial of alpha-1 antitrypsin augmentation therapy (14). These
in the scan pairs spanning the discontinuity with the other two groups
errors occurred in spite of the regular and frequent procedure
and to the Mann–Whitney U test comparing scan pairs before and after
of CT number adjustment for air calibration that was performed
the discontinuity. For the unadjusted data (A ) there is a significant
difference in the rate of VI progression between the scan pairs spanning
with an empty gantry. The drift is likely to have influenced the
the time of calibration discontinuity and the scan pairs before and after consistency of lung densitometry because we have demonstrated
the discontinuity (p ⬍ 0.001), but no significant difference between a close association between these measures and VI changes. This
scan pairs before and scan pairs after the discontinuity (p ⫽ 0.92). (B ) provides further evidence of the importance of including an
After adjustment for air calibration there is no significant difference internal calibration procedure to adjust for potential errors
between the scan pairs spanning the discontinuity and the other two in air densitometry in longitudinal studies of emphysema. The
groups (p ⫽ 0.13) or between the groups before and after the disconti- error correction described in our study overcame the problem
nuity (p ⫽ 0.72). of drift in air calibration, providing a more accurate measure of
888 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL 170 2004

Figure 6. Voxel index progression over 2 years in up-


per-zone images (A ) and lower-zone images (B ), ex-
pressed as a Z statistic obtained by Wilcoxon signed
rank test. Adjusted thresholds are indicated in Houns-
field units. The threshold for uncorrected data at ⫺910
HU is shown for comparison. Solid boxes represent ex-
piratory images, and open boxes represent inspiratory
images. The significance of the change over 2 years is
shown for all data. Data labels refer to p values.

emphysema progression. Our results confirm that our previous tion between scanners coexists with great variability in air cali-
finding of highest VI progression rate in the upper-zone inspira- bration. It is not surprising, as the density of blood is close to
tory images (8) remains valid after adjustment of scanner calibra- that of water, that interscanner variability in lung densitometry is
tion errors. Furthermore, adjustment for air calibration resulted not improved by an internal calibration method that incorporates
in improved variability and the detection of statistically significant blood density values alone whereas variability is improved when
progression in all images, even at the threshold of ⫺910 HU used adjusting for differences in air calibration. However, the applica-
in our original study (8). Consequently, this correction method tion of blood calibration may be of greater importance for lung
has now been incorporated in the analysis software being used densitometry studies of conditions such as pulmonary fibrosis,
for our clinical trials and includes calibration of the entire image which produce changes in tissues of greater density than those
data before lung segmentation and histogram analysis (18). affected by the emphysematous process (19).
Adjustment of lung densitometry to correct for errors in air Good correlation has been reported between quantitative CT
calibration provides a mechanism to improve scanner consistency and other measures of emphysema both for the voxel index
in longitudinal studies and scanner comparability in multicenter method (6, 7, 20–24) and the percentile method (22). Gevenois
studies as validated in the phantom studies. The voxel distribution and coworkers have shown that the optimal VI threshold (using
histograms for the dog lung phantom compared well with those a high-resolution protocol) is ⫺950 HU in cross-sectional studies
from patients with emphysema, confirming that the phantom comparing CT with both macroscopic (20) and microscopic mor-
was an appropriate model to validate the correction method. phometry (6). In the current study, several VI threshold values
Although the scanning protocols differed between clinical and were used because it was considered that sensitivity to the
phantom studies, it was necessary in the multicenter phantom changes of emphysema progression would be threshold depen-
study to use a whole lung volume scanning protocol and whole dent and, although the correction method was expected to im-
lung densitometry to reduce sampling errors. prove variability, it was recognized that the effect could also be
The multiscanner phantom study confirms the findings of to change the average VI threshold to one less sensitive than
Kemerink and coworkers (12) that consistency of water calibra- ⫺910 HU. There are limited data on the measurement of em-
Parr, Stoel, Stolk, et al.: Calibration in CT Lung Densitometry 889

Conflict of Interest Statement : D.G.P. does not have a financial relationship with
TABLE 4. ANNUAL PROGRESSION RATES OF SINGLE IMAGE a commercial entity that has an interest in the subject of this manuscript; B.C.S.
does not have a financial relationship with a commercial entity that has an interest
VOXEL INDEX AT A THRESHOLD OF ⫺910 HU, ADJUSTED in the subject of this manuscript; J.S. does not have a financial relationship with
FOR CHANGES IN AIR CALIBRATION WITH a commercial entity that has an interest in the subject of this manuscript; P.G.N.
DATA PREVIOUSLY REPORTED AND INCLUDED does not have a financial relationship with a commercial entity that has an interest
FOR COMPARISON* in the subject of this manuscript; R.A.S. has lectured widely for non-promotional
purposes for GlaxoSmithKline, Bayer, and Eli Lilly and acts on Advisory Boards for
Median Adjusted Rate Mean Annual Rate Reported AstraZeneca (AZ) and Baxter and has two non-commercial research grants—one
Image† (IQR)† Previously (SE)‡ from AZ (£125,000 since 2002) and one from Bayer (£500,000 per annum since
1996).
UZI 1.67 (–0.23 to 3.28) 2.82 (0.55)
LZI 0.61 (–1.63 to 2.72) 1.03 (0.56) Acknowledgment : The authors thank KCARE (Kings College Hospital, London)
UZE 0.58 (–0.50 to 2.98) 1.67 (0.38) for the loan of the lung phantom and Derek Tarrant for technical support with
LZE 1.17 (–0.29 to 3.85) 1.66 (0.77) the use of the Marconi MxView scanner. R. A. Stockley and J. Stolk are members
of AIR (the Alpha-1 International Registry, www.aatregistry.org) and thank the
Definition of abbreviations: IQR ⫽ interquartile range; LZE ⫽ lower-zone expir- other council members of AIR for valuable discussions on the results of this
atory; LZI ⫽ lower-zone inspiratory; SE ⫽ standard error; UZE ⫽ upper-zone study.
expiratory; UZI ⫽ upper-zone inspiratory.
* See Reference 8. References

n ⫽ 57.
1. Burrows B. A clinical trial of efficacy of antiproteolytic therapy: can it

n ⫽ 43.
be done? Am Rev Respir Dis 1983;127:S42–S43.
2. Croxton TL, Weinmann GG, Senior RM, Wise RA, Crapo JD, Buist
AS. Clinical research in chronic obstructive pulmonary disease: needs
and opportunities. Am J Respir Crit Care Med 2003;167:1142–1149.
physema progression by serial quantitative CT, but in a study 3. Snider GL, Kleinerman J, Thurlbeck WM, Bengali ZH. The definition
of emphysema: report of a National Heart, Lung, and Blood Institute,
of 22 patients with severe AATD over 2–4 years Dirksen and Division of Lung Diseases Workshop. Am Rev Respir Dis 1985;132:
coworkers (4) showed that the optimum VI threshold for whole 182–185.
lung analysis was ⫺930 HU. The data presented here on two 4. Dirksen A, Friis M, Olesen KP, Skovgaard LT, Sorensen K. Progress of
different single lung slices confirm that ⫺930 HU was able to emphysema in severe ␣1-antitrypsin deficiency as assessed by annual
detect progression with a high degree of statistical significance, CT. Acta Radiol 1997;38:826–832.
5. Gevenois PA, Zanen J, de Maertelaer V, De Vuyst P, Dumortier P,
as were the other thresholds. The threshold that appeared statis- Yernault JC. Macroscopic assessment of pulmonary emphysema by
tically the best for progression using this high-resolution protocol image analysis. J Clin Pathol 1995;48:318–322.
(as determined by the Z statistic) was ⫺950 HU, although the 6. Gevenois PA, De Vuyst P, de Maertelaer V, Zanen J, Jacobovitz D, Cosio
practicality of validating this in a longitudinal study using pathol- MG, Yernault JC. Comparison of computed density and microscopic
ogy is preclusive. Nevertheless, the good correlation of this morphometry in pulmonary emphysema. Am J Respir Crit Care Med
1996;154:187–192.
threshold with pathology in cross-sectional studies (6, 20) sug- 7. Müller NL, Staples CA, Miller RR, Abboud RT. Density mask: an objec-
gests that the changes in our study (at least on inspiratory scans tive method to quantitate emphysema using computed tomography.
that would not be influenced by changes in air trapping, as in Chest 1988;94:782–787.
expiratory scans) also reflect progression of emphysema. 8. Dowson LJ, Guest PJ, Stockley RA. Longitudinal changes in physiologi-
cal, radiological, and health status measurements in ␣1-antitrypsin de-
Notwithstanding the errors resulting from variability in sam-
ficiency and factors associated with decline. Am J Respir Crit Care
pling and inspiratory level, the single-slice protocol employed Med 2001;164:1805–1809.
here is sensitive to emphysema progression and also limits radia- 9. Gierada DS, Yusen RD, Pilgram TK, Crouch L, Slone RM, Bae KT,
tion exposure. Volume scanning protocols may prove to be more Lefrak SS, Cooper JD. Repeatability of quantitative CT indexes of
reproducible and sensitive to progression and the development emphysema in patients evaluated for lung volume reduction surgery.
Radiology 2001;220:448–454.
of low radiation dose protocols, in combination with faster scan-
10. Kemerink GJ, Lamers RJ, Thelissen GR, van Engelshoven JM. CT densi-
ning of the chest, currently facilitates the use of volume scans tometry of the lungs: scanner performance. J Comput Assist Tomogr
in ongoing clinical trials. These newer protocols allow correction 1996;20:24–33.
for differences in inspiration level between scans, which may 11. Stoel BC, Vrooman HA, Stolk J, Reiber JH. Sources of error in lung
improve consistency further in longitudinal studies and also en- densitometry with CT. Invest Radiol 1999;34:303–309.
12. Kemerink GJ, Lamers RJ, Thelissen GR, van Engelshoven JM. Scanner
able the retrieval of single slices from whole lung image sets to
conformity in CT densitometry of the lungs. Radiology 1995;197:749–
identify regions in the lung with a better response to treatment 752.
for emphysema. 13. Morgan MD. Detection and quantification of pulmonary emphysema by
In summary, air calibration is crucial to the consistency of lung computed tomography: a window of opportunity. Thorax 1992;47:1001–
densitometry studies in emphysema, where tissue is replaced by 1004.
14. Dirksen A, Dijkman JH, Madsen F, Stoel BC, Hutchison DC, Ulrik CS,
air. Air number measurements obtained from patient images Skovgaard LT, Kok-Jensen A, Rudolphus A, Seersholm N, et al. A
are more relevant to lung densitometry than standard quality randomized clinical trial of ␣1-antitrypsin augmentation therapy. Am
assurance values and may be used in a simple correction process J Respir Crit Care Med 1999;160:1468–1472.
that adjusts for any air calibration errors. Although this process 15. Parr DG, Stoel BC, Stolk J, Stockley RA. Lung densitometry: achieving
may also be applicable as the first step toward the standardiza- standardization by overcoming variability in scanner performance [ab-
stract]. Am J Respir Crit Care Med 2004;169:A881.
tion of lung densitometry between centers, further studies are 16. Dowson LJ, Newall C, Guest PJ, Hill SL, Stockley RA. Exercise capacity
needed. Details of calibration method are essential to the inter- predicts health status in ␣1-antitrypsin deficiency. Am J Respir Crit
pretation of lung densitometry, and quality assurance procedures Care Med 2001;163:936–941.
for densitometry studies should include the technique of air 17. Stolk J, Ng WH, Bakker ME, Reiber JH, Rabe KF, Putter H, Stoel BC.
calibration described here. These data must be reported in lon- Correlation between annual change in health status and computer
tomography derived lung density in subjects with ␣1-antitrypsin defi-
gitudinal studies of disease progression and when direct com- ciency. Thorax 2003;58:1027–1030.
parison is made between densitometry performed on different 18. Stoel BC, Bakker ME, Stolk J, Dirksen A, Stockley RA, Piitulainen E,
scanners to ensure validity. Russi EW, Reiber JH. Comparison of the sensitivities of 5 different
890 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL 170 2004

computed tomography scanners for the assessment of the progression 22. Gould GA, MacNee W, McLean A, Warren PM, Redpath A, Best JJ,
of pulmonary emphysema: a phantom study. Invest Radiol 2004;39:1–7. Lamb D, Flenley DC. CT measurements of lung density in life can
19. Best AC, Lynch AM, Bozic CM, Miller D, Grunwald GK, Lynch DA. quantitate distal airspace enlargement: an essential defining feature
Quantitative CT indexes in idiopathic pulmonary fibrosis: relationship of human emphysema. Am Rev Respir Dis 1988;137:380–392.
with physiologic impairment. Radiology 2003;228:407–414. 23. Hayhurst MD, MacNee W, Flenley DC, Wright D, McLean A, Lamb
20. Gevenois PA, de Maertelaer V, De Vuyst P, Zanen J, Yernault JC. D, Wightman AJ, Best J. Diagnosis of pulmonary emphysema by com-
Comparison of computed density and macroscopic morphometry in puterised tomography. Lancet 1984;2:320–322.
pulmonary emphysema. Am J Respir Crit Care Med 1995;152:653–657. 24. Kuwano K, Matsuba K, Ikeda T, Murakami J, Araki A, Nishitani H,
21. Gevenois PA, De Vuyst P, Sy M, Scillia P, Chaminade L, de Maertelaer Ishida T, Yasumoto K, Shigematsu N. The diagnosis of mild emphy-
V, Zanen J, Yernault JC. Pulmonary emphysema: quantitative CT sema: correlation of computed tomography and pathology scores. Am
during expiration. Radiology 1996;199:825–829. Rev Respir Dis 1990;141:169–178.

You might also like