1
Data on file with Diopsys.
5. Apply a pearl size amount of NuPrep® Gel to a clean
gauze pad.
 Patients should be tested with uncorrected vision. 6. Rub the gel lightly onto the skin with 4-5 small
While patient refraction has minimal influence on circular motions on the center of the patient’s
Diopsys® ffERG vision tests, testing the same forehead, just below the hairline. Gently remove all
patient the same way each time is important to remaining gel thoroughly with a dry part of the
track results. If the patient is tested with pupil gauze.
dilation, this can influence the results and it should
be marked in the “Comments” section. 7. Place a pearl-sized amount of Ten20® Conductive
Paste onto the EEG Electrode and position on the
 Lights should be left on in the testing room. recently cleansed area in the center of the
 Limit distractions to the patient. forehead.

 Do not test patient if fever or eye infection is 8. Connect the RED lead wire to the Lid Electrode
present. under the right eye and the BLACK lead wire the Lid
Electrode under the left eye with the OD and OS
 If a patient has a past eye injury or contact lens labels facing outwards.
issue such as dry-eye, or excessive blinking, these
may affect the patient’s ffERG response. Artificial 9. Adjust the lead wires so that they rest over the
tears may be helpful in this type of situation. patient’s ear using the slide to adjust tension and
prevent a downward pull.
 The patient should keep both eyes open and focus
the fellow eye (the eye not being tested) on a 10. Snap the GREEN lead wire to a Disposable EEG
stationary target, such as a door handle, image on Electrode. Adjust the lead wire so that it rests
behind the patient’s ear. Plug each lead wire into
the wall, etc.
the corresponding color jack of the Filter/Amplifier
 Make the patient aware that quality test results will Module.
only be achieved if he/she is sitting still and is 11. Give the mini-Ganzfeld to the patient, and instruct
paying full attention as instructed. them to cover his/her right eye. Take care to inform
the patient to rest the mini-Ganzfeld dome lightly
on the cheek and above the eyebrow so that the
eye is centered in the
1. Using an OCuSOFT Original Lid Scrub Pad, roll to
face of the dome.
create a rounded edge. Ask the patient to tilt their
head back and look up at the ceiling. Cleanse the 12. Begin the Diopsys®
under eye area of the right eye, close to the lash ffERG test.
line with 6 sweeps toward the temple. Then
thoroughly dry the area with gauze.
2. Turn the same OCuSOFT Original Lid Scrub Pad
inside out, and roll to create a rounded edge. Ask
the patient to tilt their head back and look up at the
ceiling. Cleanse the under eye area of the left eye,
close to the lash line with 6 sweeps toward the The Use of the Diopsys® NOVA™ (ARGOS™) Vision
temple. Then thoroughly dry the area with gauze. Testing Systems is Contraindicated for Patients
3. Repeat Step 1 and Step 2 cleansing and drying with Seizure Disorders.
process for both eyes' lower lid skin. Diopsys, Inc. has created this guide as a courtesy to help better
understand the test results from the Diopsys® ffERG/Multi-
4. Gently place Diopsys® ERG Lid Electrodes on the Luminance Flicker report. The comments and examples provided are
lower lid skin of the right and left eye, centered based on clinical experience and published literature. It is the
responsibility of the Doctor to make any diagnostic and/or
with the pupil, and as close as possible to the treatment decisions specific to each patient. Please consult the
eyelash line without touching the eye lashes. operator’s manual for indications, contraindications, and
instructions for use.