Professional Documents
Culture Documents
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D4107 − 20
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
1
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D4107 − 20
especially surface water sources that contain fish and other life. 7.3 Distillation Apparatus—For aqueous distillation,
The permanganate treatment oxidizes trace organics in the 250-mL and 1000-mL round bottom borosilicate flasks, con-
sample aliquots which could distill over and cause quenching necting side arm adapter,7 condenser, graduated cylinder,
interferences. A middle fraction of the distillate is collected for boiling chips, and heating mantle.
tritium analysis because the early and late fractions are more
apt to contain interfering materials for the liquid scintillation 8. Reagents and Materials
counting process. 8.1 Purity of Reagents—Reagent grade chemicals shall be
4.2 As the sample distills, there is a gradient in the tritium used in all tests. Unless otherwise indicated, it is intended that
concentration in the accumulating distillate due to isotope all reagents shall conform to the specifications of the Commit-
effects; therefore, it is important to collect the same fraction of tee on Analytical Reagents of the American Chemical Society,
the distillate for all samples and standards for tritium analysis. where such specifications are available.8 Other grades may be
used, provided it is first ascertained that the reagent is of
4.3 The collected distillate fraction is thoroughly mixed and
sufficiently high purity to permit its use without lessening the
a portion (up to 10 mL) is mixed with liquid scintillator
accuracy of the determination. Some reagents, even those of
solution, and after dark adapting, is counted in the liquid
high purity, may contain naturally-occurring radioactivity, such
scintillation counting system for tritium beta particle activity.
as isotopes of uranium, radium, actinium, thorium, rare earths,
potassium compounds, or artificially produced radionuclides,
5. Significance and Use
or combinations thereof. Consequently, when such reagents are
5.1 This test method was developed for measuring tritium in used in the analysis of low-radioactivity samples, the activity
water to determine if the concentration exceeds the regulatory of the reagents should be determined under analytical condi-
statutes of drinking water. This test method also is applicable tions that are as close as practicable to those used for the test
for the determination of tritium concentration in water as sample. The activity contributed by the reagents should be
required by technical specifications governing the operations of accounted for and applied as a correction when calculating the
nuclear power facilities. With suitable counting technique, test sample result.
sample size, and counting time a detection limit of less than 37
8.2 Purity of Water—Unless otherwise indicated, references
Bq/L (1000 pCi/L) is attainable by liquid scintillation.
to water shall be understood to mean reagent water conforming
6. Interferences to Specification D1193, Type III.
6.1 Reduced detection efficiency may result from quenching 8.3 Reagents of Distillation Treatment:
in the sample scintillator mixture. Quenching is caused by 8.3.1 Sodium Hydroxide Pellets.
impurities in the sample, which can inhibit the transfer of 8.3.2 Potassium Permanganate.
energy, or by colored materials, which may absorb some of the 8.4 Background Water, with tritium activity below the
emitted light. Corrections for quenching can be made by the minimum detectable activity (most deep well waters are low in
use of standard additions or by the ratio method.6 The approach tritium content).
described in this test method, distillation after alkaline perman- 8.5 Scintillator Solutions:
ganate treatment, eliminates quenching substances, as well as 8.5.1 Dioxane Liquid Scintillator Solution—Dissolve 4 g of
radionuclides which might be present in a volatile chemical scintillation-grade PPO (2,5-diphenyloxazole), 0.05 g of
form such as radioiodine and radiocarbon. A boiling chip must scintillation-grade POPOP [1,4-bis (5-phenyloxazolyl-2-yl)-
be used with each distillation to avoid bumping, which can benzene], and 120 g of naphthalene in 1 L of spectroquality,
amount to a carry-over excursion. 1,4-dioxane. Store the solution in a dark (amber) bottle. This
6.2 Scintillator stock solution or samples exposed to day- solution can be used with glass or polyethylene vials.
light must be dark-adapted. Also, toluene- or xylene-base 8.5.2 Solution G Scintillator Solution—Dissolve 18 g of
scintillators exposed to fluorescent lighting should be dark- scintillation-grade PPO (2,5-diphenyloxazole) and 3.6 g of
adapted for a minimum of 6 h and dioxane base scintillators scintillation-grade BIS-MSB [p-bis (o-methylstyryl) benzene]
exposed to fluorescent lighting for 24 h. All fluors should be in 2 L of spectroquality p-xylene. Add 1 L of Triton N-1019
checked for excitation under lighting conditions being used, detergent to the p-xylene scintillator solution. Dissolve 50 g of
and if possible, they should be exposed only to red light. SXS (sodium xylene sulfonate) in 100 mL of water and add
7. Apparatus
7
Corning Part No. 9060 has been found satisfactory for this purpose.
7.1 Liquid Scintillation Spectrometer, coincidence-type. 8
ACS Reagent Chemicals, Specifications and Procedures for Reagents and
Standard-Grade Reference Materials, American Chemical Society, Washington,
7.2 Liquid Scintillation Vials, of low-potassium glass are DC. For suggestions on the testing of reagents not listed by the American Chemical
recommended. Polyethylene vials may be used when other Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset,
than dioxane scintillator solution is used. U.K., and the United States Pharmacopeia and National Formulary, U.S. Pharma-
copeial Convention, Inc. (USPC), Rockville, MD.
9
The sole source of supply of the apparatus known to the committee at this time
is Rohm and Haas Company, Independence Mall West, Philadelphia, PA 19105. If
6
Bush, E. T., “General Applicability of the Channels Ratio Method of Measuring you are aware of alternative suppliers, please provide this information to ASTM
Liquid Scintillation Counting Efficiencies,” Analytical Chemistry, Vol 35, No. 1024, International Headquarters. Your comments will receive careful consideration at a
1963. meeting of the responsible technical committee,1 which you may attend.
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
2
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D4107 − 20
this solution to the p-xylene scintillator-Triton solution. Mix 11. Procedure
thoroughly. Store the solution in a dark (amber) bottle. This 11.1 Add 0.5 g of sodium hydroxide and 0.1 g of potassium
solution should be used with glass vials since the p-xylene permanganate to a 100-mL aliquot of the sample in a 250-mL
solvent evaporates slowly through the wall of the polyethylene distillation flask. Add a boiling chip to the flask. Connect a
vials. side-arm adapter and a condenser to the outlet of the flask.
8.5.3 Other commercially available scintillators can be Place a graduated cylinder at the outlet of the condenser. Heat
used, such as the environmentally safe di-isopropyl napthalene the sample to boiling to distill, collect the first 10 mL of
based scintillators. It is the responsibility of the user to verify distillate as a separate fraction and discard it.
the acceptability of a substitute scintillator.
11.2 Collect the next 50 mL of distillate for tritium analysis.
8.6 Tritium standard solution as tritiated water traceable to Thoroughly mix the 50-mL distillate fraction.
the SI through a national metrology institute (NMI) such as
NIST, or an ANSI N42.22 reference material provider with a NOTE 1—It is important that only the first 10-mL fraction be discarded
or the same fraction for samples and standards alike since there is a
concentration of approximately 17 kBq/mL. gradient in the tritium concentration of the distillate.
where:
distillate fraction. Repeat the distillation with two more
100-mL aliquots for triplicate analyses. This is the distilled raw ADWTS = activity of distilled water tritium standard, in
water tritium standard (DRWTS). becquerels (Bq),
Rb = background subtraction count count rate, in
10.1.3 Prepare for counting three aliquots of the DRWTS
counts per second (s−1),
distillate tritium standard solution (from 10.1.2), three aliquots RDWTS = distilled water tritium standard count rate (s−1),
of the DWTS, and three aliquots of the distilled raw water (for u(ADWTS) = standard uncertainty of the activity ADWTS (Bq),
background). Mix 4 mL of water with 16 mL of the dioxane tDWTS = count time for the distilled water tritium stan-
scintillator solution, or 10 mL of water with 12 mL of Solution dard (seconds), and
G scintillator solution in a liquid scintillator vial (glass vials tb = count time for the background sample
should be used for detergent-type scintillator solutions). Shake (seconds).
well, dark-adapt the vials overnight, and count in a liquid
scintillation counter. Count each vial long enough to meet the 12.2 Recovery Correction Factor, F:
required detection (0.037 Bq/mL) or longer (see Appendix X1 R DWTS 2 R b
F5 (3)
for calculating required counting time). ε 3 A RWTS
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
3
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D4107 − 20
where:
RDRWTS = count rate of distilled raw water standard (s−1),
LC 5
Œ
1.65 S D
R b 3 t a 3 11
ta
tb
(7)
and ε 3 t a 3 F 3 V 3 e 2λt
ARWTS = activity of (undistilled) raw water tritium standard NOTE 2—The k factor of 1.645 assumes a Type I error rate of 5 %.
Bq. Factors corresponding to other error rates may be applied as appropriate
depending of the data/measurement quality objectives.
12.3 Sample Tritium Activity, AC, for Each Aliquot:
12.7 When the detection criterion of 12.6 is used, calculate
Ra 2 Rb the a priori minimum detectable activity concentration (MDC)
AC 5 (4)
ε 3 F 3 V 3 e 2λt as follows:
where:
Ra = STS gross count rate (s−1),
MDC 5
2.7113.29 Œ S D
R b 3 t a 3 11
ta
tb
(8)
ε = detection efficiency, as determined in Eq 1, ε 3 t a 3 F 3 V 3 e 2λt
V = volume of the distillate added to STS (mL), NOTE 3—Eq 8 is only valid for Type I and Type II error rates of 5 %.
F = recovery factor, as determined in Eq 3, The formulation will need to be modified if data/measurement quality
λ = decay constant for tritium, (ln 2) / t1/2, objectives specify different tolerable decision error rates.
t1/2 = half-life of tritium, 4497 days,10,11 and
t = elapsed time between sampling and counting, in days. 13. Quality Control
13.1 In order to provide reasonable assurance that the
12.4 The result of the measurement has an uncertainty due analytical results obtained using this test method are valid and
to counting statistics (counting uncertainty). The component of accurate within the confidence limits of this test method,
the combined standard uncertainty of the tritium concentration quality control (QC) samples are analyzed with each batch of
in the sample due to counting statistics, ucC(AC), is given by: samples undergoing analysis. Each batch should include not
more than 20 samples, including those used for QC purposes.
u cC~ AC! 5
Œ
Ra Rb
1
ta tb
(5)
Laboratory or project quality assurance plans may contain
more restrictive process QC requirements. The following
ε 3 F 3 V 3 e 2λt minimum QC procedures must be followed when running the
where: test method:
tb = count time of the background sample, in seconds. 13.2 Initial Demonstration of Laboratory/Instrument Qual-
12.5 The combined standard uncertainty, uc(AC), of the ity:
measured concentration can be calculated as follows: 13.2.1 If a laboratory or analyst has not performed this test
before or there has been a major change in the measurement
u c ~ AC! 5 Œ 2
u cC~ AC! 1AC2 FS ~ !D S ~ !D S ~ !D G
u F
F
2
1
u V
V
2
1
u ε
ε
2
system, a precision and bias study must be performed to
demonstrate laboratory/instrument capability. A significant
(6) change is defined as any change, repair, or alteration of any
component in the system which maybe expected to affect the
where: response of the measurement system. See Practices D3648 for
AC = measured tritium concentration (Bq/mL), from recommended practices.
Eq 4, 13.2.2 Analyze seven replicates of a standard solution
ucC(AC) = standard counting uncertainty of AC, from Eq 5, prepared from an independent reference material (IRM) con-
u(V) = standard uncertainty of the volume, V, taining at 3H concentrations sufficient to reduce the relative
u(ε) = standard uncertainty of the detection efficiency, ε, standard counting uncertainty to 1 % or less. The matrix used
from Eq 2, and for the demonstration should represent a water sample typical
u(F) = standard uncertainty of the recovery factor. The
for which the procedure will be used, for example, drinking
standard uncertainty of F could be determined by
water.
repeated measurements; however, since F will
13.2.3 Calculate the mean and standard deviation of the
generally be close to 1, the uncertainty may be
seven values and compare to the acceptable ranges of precision
assumed to be 0.
and mean bias of 10 % and 610 %, respectively, based on a
12.6 For each sample measurement, calculate the critical review of the collaborative study data. Practice D2777 should
activity concentration, LC, in becquerels per milliliter (Bq/mL), be consulted on the manner by which precision and mean bias
as follows: are determined from the initial demonstration study. The study
should be repeated until the precision and bias are within the
given limits.
10
“Table of Radionuclides,” BIPM Monographie 5, Vol 3, pp. 1–4, available 13.2.4 Analyze three replicates of a blank solution matrix.
from www.bipm.org/utils/common/pdf/monographieRI/Monographie_BIPM-5_ The matrix used for the demonstration should represent a water
Tables_Vol3.pdf, accessed July 22, 2019.
11
sample typical for which this test method will be used, for
Two online references for expertly evaluated nuclear data, including radionu-
example, drinking water. The total dissolved solids (TDS) of
clide half-lives and their uncertainties, are recommended: BIPM-5 and NUDAT2.
Other sources of nuclear data may be used at the user’s discretion. In all cases, the the matrix should approximate that which may be encountered
source should be clearly documented. in normal use.
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
4
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D4107 − 20
13.2.5 Calculate the 3H activity for each of the three blank ? AC original 2 ACdup ?
DER 5 (10)
solutions. This test method shall not be used for official =u 2c ~ ACoriginal! 1u 2c ~ ACdup!
samples until the results for each of the blank solutions is
below half of the associated MDC. where:
13.2.6 This test method shall not be used for official samples ACoriginal = original sample activity concentration,
until precision, bias, and blank requirements are met. ACdup = duplicate sample activity concentration,
uc(ACoriginal) = combined standard uncertainty of the origi-
13.3 Laboratory Control Sample (LCS): nal sample, and
13.3.1 To ensure that the test method is in control, analyze uc(ACdup) = combined standard uncertainty of the dupli-
an LCS with each batch of no more than 20 samples. The cate sample.
activity added to reagent water should be appropriate for the
type of samples analyzed and should allow sufficient precision 13.6.2 In those cases where there is insufficient sample
to ensure meaningful assessment of accuracy. The LCS must be volume to allow performance of a duplicate sample analysis, a
taken through all the steps of the analytical method including duplicate LCS should be performed and analyzed using the
sample preservation and pretreatment. The result for the LCS same DER criteria.
shall fall with the limit of 625 % of the expected value. 13.6.3 The value of DER should be less than or equal to 3.0.
13.3.2 If the result is not within the limit, analyses should be If the sample duplicate or LCS duplicate result is greater than
stopped and the reason for failure identified and resolved. An 3.0, reanalyze all samples in the batch or provide an explana-
indication of the occurrence should accompany the reported tion in a case narrative.
results. 13.7 Independent Reference Material (IRM):
13.4 Method Blank (Blank): 13.7.1 In order to verify the quantitative value produced by
13.4.1 Analyze a reagent water test blank or distilled RWS the test method, analyze an IRM sample, which was submitted
(if reagent water is known to contain 3H) with each batch of no on at least a single-blind basis (if practical) to the laboratory at
more than 20 samples. The concentration of 3H found in the least once per quarter. The concentration of analyte in the
blank should be less than the LC. traceable reference material should be appropriate to the
13.4.2 If the concentration of 3H is found above this level, typical purpose for which the method is used. The value
provide an explanation in a case narrative. obtained shall demonstrate acceptable performance as defined
by the program or the outside source.
13.5 Matrix Spike (MS): 13.7.2 In the absence of other acceptance criteria for the
13.5.1 Analyze at least one matrix spike sample with each IRM sample, compare the IRM sample result to the IRM
batch of no more than 20 samples by spiking an aliquot of a known value as follows:
sample within the batch with a known 3H concentration.
13.5.2 The spike should produce a concentration of 3H that R5
? IRM found 2 IRMknown ? (11)
is 2 to 5 times the anticipated sample concentration or as =u ~ IRMfound! 1u 2c ~ IRMknown!
2
c
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
5
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D4107 − 20
14.2 These collaborative test data were obtained using 14.4 Bias—The bias of this test method, based upon the
reagent grade water. For other matrices, these data may not collaborative test data, was found to vary with level as
apply. presented in Table 1.
14.3 Precision—The overall and single operator precision
have been found to vary with level as presented in Table 1. 15. Keywords
15.1 drinking water; 3H; liquid scintillation; radioactivity;
TABLE 1 Observed Tritium Precision and Bias Data radioisotope; tritium
Added, Measured Absolute Precision
Percent
Bq/mL Bq/mL Bias, Bq/mL So St
0.26 0.25 0.01 −3.8 0.015 0.035
1.01 0.95 0.06 −5.9 0.020 0.120
11.4 10.5 0.9 −7.9 0.170 1.370
APPENDIX
(Nonmandatory Information)
X1. DETERMINATION OF THE NECESSARY COUNTING TIME FOR A REQUIRED DETECTION LIMIT (SENSITIVITY)
X1.1 Section 141.25 (c) of the National Primary Drinking X1.1.3 Since the gross count rate, Ra, is equal to the net
Water Regulations (NPDWR) define detection limit as follows: count rate, N, plus the background count rate, Rb, Ra = N + Rb.
For the purpose of monitoring radioactivity concentrations so:
in drinking water, the required sensitivity of the radioanalysis
N12R b
is defined in terms of a detection limit. The detection limit shall σ N2 5 (X1.4)
t
be that concentration which can be counted with a precision of
6100 % at the 95 % confidence level (1.96 σN where σN is the X1.1.3.1 Now, combine Eq X1.1 and Eq X1.4:
standard deviation of the net counting rate of the sample).
X1.1.1 In this definition, the following equation is implied: σ N2 5
N12R b
t
5
N
1.96S D 2
(X1.5)
N
σN 5
1.96
(X1.1) Solve for t:
3.84 3 N17.68 3 R b
where: t5 (X1.6)
N2
N = net count rate, Ra – Rb.
X1.1.2 Also, the standard deviation of the net count rate can X1.2 Example—The required detection limit for tritium in
be calculated from the equation: drinking water is 0.037 Bq/mL. Determine the counting time
that is required to meet that detection limit when a 10-mL
σN 5 Œ Ra Rb
1
ta tb
(X1.2) sample aliquot is counted at a detection efficiency of 15 % and
a background count rate of 0.27 s–1.
where: X1.2.1 Calculation—Calculate the absolute tritium activity,
Ra = gross count rate (s–1), in becquerels, present in a 10 mL sample aliquot when the
Rb = background count rate (s–1), sample is at the required detection limit:
ta = counting time for the sample aliquant (seconds), and
tb = counting time for the background (seconds). 0.037 Bq/mL 3 10 mL 5 0.37 Bq 5 0.37 s 21
σN 5 Œ R a 1R b
t
(X1.3)
0.37 s 21 3 0.15 5 0.0555 s 21 5 N
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
6
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D4107 − 20
RELATED MATERIAL
ASTM D3856 Guide for Management Systems in Laboratories Engaged ASTM D6001 Guide for Direct-Push Groundwater Sampling for Envi-
in Analysis of Water ronmental Site Characterization
ASTM D4448 Guide for Sampling Ground-Water Monitoring Wells ASTM D7282 Practice for Set-up, Calibration, and Quality Control of
ASTM D5847 Practice for Writing Quality Control Specifications for Instruments Used for Radioactivity Measurements
Standard Test Methods for Water Analysis
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
&RS\ULJKWE\$670,QW
ODOOULJKWVUHVHUYHG:HG6HS*07
7
'RZQORDGHGSULQWHGE\
8&58QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG