Professional Documents
Culture Documents
1. Find the dial switch on the CPU board (Chest stand and table). Put No.1 to
ON together (ON – Calibration mode, shield potentiometer value)
2. Chest stand Sw1-2 (ON- Table is on the left side. OFF- Table is on right
side). Table SW1-2 no need to setting.
Potentiometer adjustment of Tube column and chest stand: ( MCU board from chest stand)
1. Tube horizontal movement potentiometer measurement (Pull-line potentiometer of rail).
TP11-TP10 (Value is around from 100mv – 2V).
2. Tube up and down potentiometer adjustment. Adjust tube center to ground is
1m ,measure TP14-TP10 should be 0.963±0.01V.
3. Chest stand up and down potentiometer adjustment. Adjust chest detector center to
ground is 1 m, measure TP13-TP10 is 1.304±0.01V.
Tube rotation ③
Column brake ④
⑤
⑥
⑦ & ⑧
9 10 11 12
Calibration instruction:
1. Enter calibration mode.
Put SW1-1 to ON for both MCU board. Press button 1, 2, 6 together to enter calibration
mode. (The program LED blink)
2. Press button 4 to move the tube column to left side 30cm to the end of rail. (see picture ).
Press button 1 and 4 together (Press button more than 3-5 seconds, it is better to press 3
times).
3. Press button 4 to move the tube column to right side 30cm to the end of rail. ( see picture ).
Press button 2 and 4 together (Press button more than 3-5 seconds, it is better to press 3
times).
4. Move the tube column to the detector holder center!!!! Press button 1 and 2
together more than 3 seconds (Try 3 times). Then press footswitch 10 and 11
together more than 3 seconds (try 3 times).
5. Move the tube column to right side ( if the table is on left side ), make sure the
distance between center of tube arm and edge of table is 36±1cm. (See picture) .
Press tube up button to move the tube to 1.5m position (Tube to floor is 1.5m).
Rotate the tube to chest stand (90°) and move the chest detector holder to 1.5 m
also. Press chest stand button 7 and 8 together to complete following movement
calibration and table edge memory.
Table edge
Rotate the tube to 90°, measure the FFD( Distance between tube focus and chest
detector) such as 100cm. Exit WD-Acquire software and modify Distance=2xxx
(reference ranges=2000-3000). Then, save this file and close it. Start WD-acquire
software again and check the PAD value and modify the Distance=2xxx until Pad
value to 100cm.
Rotate the tube to 0°, measure the FFD( Distance between tube focus and table
detector) such as 100cm. Exit WD-Acquire software and modify Framebox=500
(reference ranges=500). Then, save this file and close it.
EleConfig=3