Type: Software
Target processor: x86
Target compiler: VS.NET 2003 C++ (MFC)
DWG G25: -
Board: -
Description: The program for control and monitoring YLS (High Power) lasers.
Revision history:
Rev G60 Release Date Author CVS Tag Remarks
1 G60-19-030925-1 17.12.2003 MB - Version 2.00
2 G60-19-030925-2 09.01.2004 MB - Version 2.01
3 G60-19-030925-3 15.06.2004 MB - Version 2.02
4 G60-19-030925-4 30.06.2004 MB - Version 2.03. New Hardware (IG203)
5 G60-19-030925-5 29.07.2004 MB - Version 2.04
6 G60-19-030925-6 04.08.2004 MB - Version 2.05
7 G60-19-030925-7 07.08.2004 MB - Version 2.06
8 G60-19-030925-8 09.08.2004 MB - Version 2.07
9 G60-19-030925-9 10.08.2004 MB - Version 2.08
10 G60-19-030925-10 04.10.2004 MB - Version 2.09
11 G60-19-030925-11 11.11.2004 MB - Version 2.10.1
12 G60-19-030925-12 14.11.2004 MB - Version 2.11
13 G60-19-030925-13 17.11.2004 MB - Version 2.12
14 G60-19-030925-14 16.12.2004 MB - Version 2.13
Form: 19002
Software description Revision: 01
Seite/Page: 2 of 29
Version history:
Version Changes
2.00 The first release version of the program. It was tested in VW.
2.01 It is installed in TWI and Hyundai.
2.02 It is installed in KATECH.
2.03 The first version of the program for the new laser’s hardware (IG203 e-board).
It is installed in KUKA.
2.04 - is added : the ramping time controls;
- is added : the chiller page;
- is added : the power supply page;
- is changed : some signals in status and alarm pages;
- some bugs is fixed.
It is installed on 10K Laser in BIAS Rostock.
2.05 - is added : the button for change IP address of the laser;
- some bugs is fixed.
It is installed on 10K Laser in BIAS Bremen (the second installation).
2.06 - is changed : the emission button on the control page is initialized by WCE_STATUS_EMISSION_ENABLED bit of the status (before – by
Form: 19002
Software description Revision: 01
Seite/Page: 7 of 29
WCE_STATUS_EMISSION);
- is added : the switching of the program to IPG Service Mode without rebuild it;
It is installed in KUKA (the second installation).
2.07 - is added : the offline mode of modem;
- is added : the support of “access.dat” file (for permanent switching to IPG Service Mode and Modem Mode);
- some bugs is fixed.
2.08 - bug is fixed : the program was not saved a current telephone number in the offline modem mode;
- bug is fixed : the program was crash when the modem has second connection;
- bug is fixed : the state of the socket connection was not updated when the modem has disconnected.
2.09 - is added : the oscilloscope mode;
- is added : support “No power supplies monitor” configuration bit of the laser.
2.10.1 - is changed : the program is converted to UNICODE;
- is added : the support of the laser programs;
- is added : the support of the optical switch;
- some bugs is fixed.
2.11 - is added : LEDs for the chiller status.
2.12 - is added : the support of the laser test station (the button for read special file from TestStation is added).
2.16 - is added : the support of the single module mode;
- is added : the support of the hardwiring version with the input and output for the lamp test.
2.17 - is added : the support of the 7onfiguration interface (English + Germany).
2.19 - is added : graphics for the laser program;
- bug is fixed : the problem with the modem connection (this problem was presented in all UNICODE version of the program).
2.19.1 - bug is fixed : the problem of the loading of a laser program in the Germany version;
- bug is fixed : the problem with moving of the control for select of the time scale in the window of the laser program when the window size was changing.
2.19.2 - bug is fixed : the problem with the show of all events.
2.19.3 - bug is fixed : the problem with the control dialog for set a new time in the laser.
2.20 - is changed : the protocol of transfer of files through modem connection.
2.20.1 - bug is fixed : the program was crashed when it was started on a computer without modem.
2.20.2 - bug is fixed : the problem with the save old state of the hardwiring on the “Hardwiring” page.
2.20.3 - bug is fixed : the log file viewer was not showing the new version of the main laser structure (v5);
- bug is fixed : the problem with the show of any parameters when a laser was restarted without restart LaserNet.
2.21 - is added : the processing of WCE_CFG_HRDW_CONTROL_BSW bit in the configuration of the laser (the control of the optical channels on the control page is
disabled in the Robot mode);
- is changed : the “Emission” button of LaserNet’s control tab is reflected the status of the emission if it is disabled;
- is added : the USA mode;
Form: 19002
Software description Revision: 01
Seite/Page: 8 of 29
2.42 - Is added the new Laser Programs for HMC and small changes in the HMC DeviceNet protocol;
- Correct some strings in the German interface;
- Is added the Korean language for GUI;
- Is added the new event filter by date;
- Fixed bug of version 2.41 in the events page.
2.42.1 - Some changes in the Korean UI.
2.42.2 - Some changes in the Korean UI;
- Support of the checkbox “Always push ‘Ctrl’ key + ‘Left’ mouse button for ‘Laser Reset’ and ‘Emission ON’ commands” in the installer.
- Fixed bug: message “An invalid argument was encountered” in the control page, when switching between laser without beam switch and with one.
2.42.3 - Fixed bug in the edit list control – crash of the program at the time of input of the new character (ex. IP Configuration dialog box)
2.43 - Is added: IPG beam coupler support;
- Is added: support of safety buttons, power monitor, MAC address, beam switch additional monitoring (water leakage, E-Stop button);
- Rewrite all pages: the new internal released: LEDs, parameters and messages.
2.43.1 Fixed bug: All LEDs on the chiller page shown alarm when only one alarm is present
2.43.2 Fixed bug: Crtl page in the robot mode was enabled after connection to the laser
2.43.3/7 Fixed some bugs during installation on DC (Interbus and Control pages)
2.43.8 Fixed bug: In the chiller folder the warning and the alarm scriptures for temperatures and conductivity are mixed;
Fixed bug: When key in Robot position and “Settings/Control/CTRL for Reset and Emission ON” is Enable – it was possible to switch on and off emission by
pushing CTRL button and Emission on button in LaserNet Control folder.
2.43.9 Fixed bug: the FFBD signal is not indicated in Beam Switch page;
Fixed bug: events is not deleted in offline mode;
Fixed bug: the Current slider is not renamed to Duty Cycle in QCW mode in Control page;
2.43.10 Fixed bug: When we set channel with hardwiring box the mirror switch to the choose channel. The green lamp of home position does not switch off. Green lamp of
choose channel switch on. After restart of LaserNet the green lamp of Home position is off. After a new channel setting the problem appears again.
Fixed bug: Internal controls enable works normal. But it is not possible to type numbers from 1 to 9 in the fields of rise and fall time. The field only accepts the 0.
2.43.11 Fixed bug: in Profibus and DeviceNet pages don’t show input/output values.
2.44 Added: the modules page with opportunity for select of some modules for work;
Fixed bug: shown signal “Lamp power test output” in the Hardwiring page (in place of Spare Output1);
2.44.1 Fixed bug: the scattered value on the BeamCoupler page was not updated at some cases.
2.44.2 Fixed bug: problem with switching between current/power on the Control page for 1 kW lasers.
2.44.3 Fixed bug: problem with current/power sliders on the Control page in the Robot mode.
2.44.4 Fixed bugs: Problem with current/power sliders on the Control page in the Robot mode.
Indicators of ramping/falling times are not updated on the Control page in the Robot mode.
The event --- Type = 21, Mask = 0 (was disabled).
The event “Beam Switch ready” present when the beam coupler installed (was renamed to “Beam Coupler ready”).
Form: 19002
Software description Revision: 01
Seite/Page: 11 of 29
2.44.5 Fixed 50% of problems with the modem connection for some new types of modems. Now the LaserNet on the customer side always connected but sometimes is not
answering (~1 answer for 5 calls). May be this problem of our telephone line or drivers of our modem. It will be fixed in the future version of the LaserNet.
2.45 Confirmation of overwrite of the program in the laser is added.
2.45.1 Japan language is added;
Visualization of empty laser programs is added (the new “Number of Laser Program” dialog) .
2.45.2 Save some laser data into the TXT-monitor file for customers is added.
2.45.3 Fixed bug: chiller warnings always shown on Hardwiring page.
2.45.4 Some strings translated into the Japanese language are added.
2.45.5 Many strings translated into the Korean language are added.
2.46.3 Opportunity of moving of the LaserNet window on screen for its any point by mouse is added.
New event type for fieldbus controller errors is added.
2.46.4 Some changes in the IPG Beam Coupler page
2.46.5 Some changes in the IPG Beam Coupler page
2.46.6 Hide all flow rates in IPG Beam Coupler page without the bit 0x100 in Config2.
Shown all IPG Beam Coupler status bits in the events page.
2.46.7 Fixed the problem with processor usage (decrease from >55% to 10%)
2.46.8/10 Fixed bug: “It was found that sometimes the information on the laser status is missing after the switching of the control from one laser to another. For example, the
“Laser Emission”, “Laser On”, “External Control”, “Analog Control” buttons, the pattern of the modules selected for emission, etc.”
2.46.11 Fixed bug: redraw power values (under the power slider) after switch from current to power.
Rename the event for the signal “beam switch warning lamp failed” (old string: “Reserved”)
2.47.0/1 The option “Warning at switching lasers” in the Control Settings was added (MHI request).
Now the selection of the current laser in the lasers list (on the left side of the program) is always marked by blue and uses the bold and underlined font (MHI request).
2.47.2 Some strings translated to Germany and Japanese.
Added the message box which shown during the loading of new events from the laser after established connection.
2.47.3 The speed of the clearing events list on the events page is increased.
2.48 The work with 1.000.000 events has been optimized (updating from the laser, loading from the disk during the start).
Fixed some problem with events during intensive work of the laser via fieldbus/hardwiring (more 7 full-events per second). Currently version has been tested during
60 hours (50 full-events per second, together with WCELaser v.1.37).
2.48.1 Fixed bug: When activate “Ctrl for Reset and Emission On” and change the “Control” page and go back, is not possible to switch off emission without pressing Ctrl.
2.48.2 Fixed bug: When the maximum of output power of the laser is not rounded by 100W, the set power command return the answer ‘error’.
The filtered events list has been optimized.
2.49 Was added:
- Scalability of the program window (resizing);
- Support of QCW+ mode;
- Display of version numbers of programs (only for IPG Service Mode);
Form: 19002
Software description Revision: 01
Seite/Page: 12 of 29
2.67.3/4 The view for dynamic structures in the laser.dat was added to the log-file viewer.
Small changes in the German translation.
2.67.5 Small changes on IPG-Chiller page (only labels).
2.67.6 Small changes in Italian language.
2.67.7/8 Small changes for IPG chiller type “g”.
2.67.9 Fix small problem in the log file viewer.
2.67.10/12 Small changes for IPG chiller type “g”.
2.67.13 Small changes for IPG Beam Coupler.
Fix the small problem in the log file viewer.
2.67.14 Changes in German and French localizations.
2.67.15 “Fiber Interlock” was renamed on “Process Fiber Interlock” for a beam coupler.
The lamp “Water inside beam coupler” was removed from alarm page for IPG Beam Coupler Type 1.
2.68 The support of 8 words fieldbus protocol and extension of laser programs for use ext 3 words were added.
2.69 The support of IG320 with 6-th COM-CCA modules (PROFIBUS & DeviceNet) was added.
2.69.1 The one string on German was corrected.
2.70 The fieldbus page for Daimler protocol was added.
2.70.1 Fix the problem with menu item “Settings\Laser\Update Configuration”
2.71 The full support of remote update configuration of a laser.
The support of High Power Single Mode laser
2.71.1 Two forgotten strings for events decoding have been added: “Updating of PC104 firmware” and “Updating of the laser configuration”
2.71.2 The bug was fixed: the Profibus/DeviceNet pages did not show.
2.71.4 The bug in the LogFileViewer was fixed: for laser.dat files some last records did not show.
New features in the LogFileViewer: “frozen” names of columns and “zebra”.
Added SI2 and SO2, SO3 in the LaserProgramsEditor for AUDI fieldbus protocol.
2.71.5 Some bugs in the LaserProgramsEditor was fixed.
2.71.6/7 All translations for all languages (except Spanish and Italian) were brought to actual state.
Support of Audi-Fieldbus protocol was added.
The warning: “Insufficient amount of active laser modules” “Reduced active modules” was added.
The alarm event: “Power supplies overheat” was added.
2.71.8 Fixed problem: the configuration bit “Power Monitor is absent” has worked only for Single Mode and Multi Outputs lasers.
2.71.9 The bug with access levels menu was fixed.
Italian translation was brought to actual state.
2.72.0 “Laser temp. is too low” and “Laser power out of spec” – warnings and alarms was added.
2.72.1 The problem of save logfile data in text format was fixed.
2.72.2 The problem with beam switch guide lasers was fixed.
Form: 19002
Software description Revision: 01
Seite/Page: 15 of 29
2.84.8 The bug with enabled “External” button in Robot mode was fixed.
2.84.9 The bug with enabled “External” button in “Ext Sync 16onfiguration” in some cases was fixed.
2.85 Fieldbus page was upgraded to v4 of Kuka/Kuka-8W/Standard (Default) protocols
2.87.2 QCW Programs Editor was added
2.87.6 Some problems for HP SC laser was fixed.
2.88.0 The new filedbus protocol VW was supported.
2.89.0 The support of new high contrast lasers was added.
Some bugs were fixed.
Two output signals for hardwiring “Reduced active modules” and “Reserved module is active” was added.
2.89.1 The new filedbus protocol Ford was supported.
Some bugs in the fieldbus page was fixed.
2.89.2/3 The support of multi switch board (IO User Interface type) was added.
2.89.4 The support of Bystronic was added (hide the logo on the main window by VersionsSwitcher utility).
2.89.5 A lot of bugs was fixed (at the case if 1-ch beam switch is installed, in fieldbus page for VW & MPI protocols, warnings about pulse library for non pulsed lasers)
2.89.7 The “External Modulation” option for SuperCompact laser was added
2.89.8 The support of LSS Fieldbus protocol was added;
The bug in the Debug Fieldbus windows was fixed: The message box with “Invalid command…” was present, if the first FB controller was installed not in the first
slot…
2.90.0/7 A lot of bugs were fixed:
- in help files paths;
- in the QCW program editor;
- in logfile-viewer;
- ‘the answer ‘error’ at command 29’ on Interbus page was fixed;
- Analog/External buttons interplay on the control page;
- etc.
The folders politics was changed.
Dew point sensor page was added.
The indicator of the laser mode (test/robot) was added.
2.90.8 The bug in QCW program editor was fixed (GOTO SI)
2.90.9 The bug in QCW program editor was fixed
2.91.3 The feature for switching off some master modules for high brightness laser was added.
2.91.4/6 The paths dialog, denied areas for pulse sliders were added.
The restriction of an emission time on the hot reserve mode has been added (but not tested!!! The warning messages were not approved by EA!!!).
2.92 The support of VW Fast protocol was added.
2.92.1 The bug with firmware for IG320 v2.14 was fixed.
Form: 19002
Software description Revision: 01
Seite/Page: 17 of 29
2.92.2 The bug with select channel command in the program editor for 1-channel beam switch was fixed.
2.93 The Adige protocol was changed for R3 support.
2.93.1 Add ConfigView, add max power (%) in BeamSwitch page, add change network settings Ethernet IP in FieldBus settings dialog.
2.93.3 Append support MPI Toyota, Air Conditioner
2.93.4 Append support MBDA chiller (type=12)
2.93.6 Fixed bugs
2.93.7 Refresh build
2.93.9 Reverse bytes in IP Properties dialog for setup ip,mask, gateway in COMX modules (Motorola byte order)
2.94.3 Add support for Pump Combining Laser (Free space aux user interface)
2.94.4 Fixed bugs
2.94.5 Remove limitation of lines in GOTO command in Pulsed Program. Fixed UI interface Adige protocol (Work Time Emission Time …). Fixed Bugs.
2.94.6 Append red led for HBR to Pump Combining Page.
2.94.7 - the command line parameter \port= was added
- information about beam coupler FF/PF FFBD was added into the logfile viewer
2.94.8 - the correlation between the signal “cabinet door is open” the configuration bit “no monitoring of power supply” is broken
2.95.0 Fix log file viewer (ipg chiller parameters)
2.95.1 Fix log (dew point)
2.95.2 SI2 was added in the laser programs editor for the Ford protocol
2.95.3 Append to config view output Profile ID
2.95.4 Disable output Emission Time and Emission Time Today (except ipgaccess)
2.95.5 Append support for 128 laser modules
2.95.6 Fixed bugs in Control page.
2.95.7 The Modulation button is always enabled now
The Update Chiller Firmware button was enabled (if laser/chiller firmware are supported this update)
2.95.8 HPL_RemoteUpdateCreator was removed from the solution (singled out into separate project)
Help files was removed from the solution (singled out into separate project)
2.95.9 Append support of fieldbus protocol VASS 2 (VW Wolfsburg)
2.96.1 Append support of IPG Beam Switch with CAN interface (the first version of IPG Bsw Page was removed)
2.96.2 Append support of BMW 16W filedbus protocol
Bugs with error messages in Program Editor are fixed. (Misoperation of “Out of range” error message for “Wait” command took a place. Also error messages for
2.96.4
“Pulse” command were not complete.)
2.96.5 Bug is fixed in Program Editor, that appeared on pressing “Write” button, but currently editing cell was on focus. (Changed value couldn’t take effect)
2.96.6 Fixed bug with crash program after close main window when once opened Configuration View dialog.
The support of chiller type 13 was added.
2.97.1
Few names of I/O signals were corrected for BMW16 FB protocol.
Form: 19002
Software description Revision: 01
Seite/Page: 18 of 29
- append tooltip for showing thresholds parameters (Flow Meters: Status Page, Beamswitch Page, Coupler Page)
- add support flow meters in lasers with coupler
- chiller page was is divided into three pages( Status, Alarms, Warning)
- add water supplying led on chiller status page
2.107 The support of Remote FFC was added (the page + events)
Source code redesign of Pages for AuxUserInterafece
Fixed:
- BeamCoupler Page for hide values of Flow Meters if in config not setted bit.
- Save, Write config of LaserNet
- Events List View in Events page. End Time in list showed as Begin Time.
New:
2.107.1/.9
- Support Brazing Laser
- Add “Show Average Power in Pulsed Mode” in Control Settings dialog
- Add support “External laser enable” status for lasers with configuration: SICK + (FFC or FF).
- Add support “External permission malfunction” alarm for lasers with configuration: SICK + (FFC or FF).
- Add support hysteresis parameters of Riedel’s chiller temperatures (IG412)
- Added the ability to view the CS.dat log file
Fixed:
- flicker sliders on the Pulse Mode page
New:
- Include all files to export log (modules, events, alarms, times, e.t.c).
2.108.1/6 - Add button “Make New Folder” in Select Folder dialog.
- Add event “High Pulse Energy”
- Add RDPL fieldbus protocol support
- Add QCW Ext fieldbus protocol support
- Add VASS 3 fieldbus protocol support
2.109.0 Add VW Pamplona protocol support
2.109.1/6 Fixed:
- viewing problem with Adige protocol
- viewing problem with Brose protocol
New:
- Add event “Module Fiber Interlock” for Brazing Laser
- MPI VASS4
- FBUS power value for SP command in Laser Program (Brazing)
- MPI BROSE V2
Form: 19002
Software description Revision: 01
Seite/Page: 21 of 29
New:
2.109.7/8
- Add in monitor system power from brazing modules
2.110.0 Same changes for the YLP-HP-1kW-CABINET laser
Fixed bug found by A.Guenther (letter to N. Semko from 23.07.15)
2.110.1
Added possibility to archive log-files for a time interval
Changed MPI VW Pamplona (swap input signals Laser Program, Optical Channel).
2.110.2/6 Fixed Laser Program Editor (FBUS not used in the command SP when selected MPI VW Pamplona)
Fixed problems with HP RDPL lasers.
The alarm led “Workcell interlock malfunction” was added on Alarm Page. Condition of visibility: from v2.71 (IG277) & v3.45 (IG320) + IPG CAN Beam Switch
2.111.0
(ExtOpticsType == 3)
The alarm led “Ext. interlock fault” was added on Beam Switch page for each channel. Condition of visibility: from v2.72 (IG277) & v3.46 (IG320) + WCELaser
2.111.1/3 v3.209 + IPG CAN Beam Switch (ExtOpticsType == 3)
Some modification for HP RDPL lasers.
Add support MPI FordBrazing.
2.111.4/6 Fixed problem with view of MPI VW Pamplona (optical channel <-> program number).
Append bit 11 “High Back Reflection” in Status2 for KUKA, KUKA8W, KUKA (QCW Extension).
2.111.7 The new status warning bit about “module(s) with no power supply voltage” was added into Standard protocol FB-page
2.111.8/9 Few changes in the new FB page in VW family protocols, for the support of a beam coupler without additional IB-gateway.
2.112 Three parameters: cabinet temperature, humidity and dew point has been added into monitor file
Splice Box alarm has been enabled for IG203 board.
2.112.1/2
Changed access from IPG Service Mode to Engeneer for Pulsed Laser Settings
2.112.3 The wrong string of the event “Alarm FFBD” for Remote FFC was fixed.
2.113 The test release version with support VASS_W8 protocol (test version of special protocol was made by request VW (Mishke), for test only)
2.113.1/2 The support of ESD Generic file for EthernetIP was added into Fieldbus Advanced Settings dialog
2.114.5 Added: Maintenance page and Page for Beam coupler for YLP-HP laser
2.115.0/2 Added: Pulse Generator Page
2.115.3 Added two new parameters for recording to the monitor file.
Added words “Alarms”, “Warnings”, etc. in Monitor File Window.
2.115.4 Pulse Generator Page is reworked (Waveform generator functionality)
Added possibility to work with 200 laser programs
2.115.5 Fixed bug in Monitor File Window
2.115.8 Added new parameter FBUS_W8 for Brazing VASS protocol. Added field to change byte order for Brazing VASS protocol.
2.115.9 Changes are made for Profinet page (for RD-PL protocol) according to D. Deyneka specification.
2.115.10 Changes for Control page (for RD-PL protocol) ) according to D. Deyneka specification (Email from 20.05.2016)
Form: 19002
Software description Revision: 01
Seite/Page: 22 of 29
Added:
- Maintenance settings (for correct work need WCELaser min 3.220)
- Alarm: Module(-s) have no supply voltage
2.117.0/5
Fixed:
- Maintenance (Replace DI Cartridge)
- Profibus address settings
Added:
2.117.6
- Show Help ID by mouse click in Service Mode
Added:
- To show the label “Dew Point Temp. is OK” on Hardwiring page instead of “Spare Output” for firmware since versions: 2.78 & 3.55
2.117.7/9 - Output power tune
- Custom Power via FieldBus
Fixed some bugs
Added:
- some new Help IDs for parameters on Dew Point and Status pages
- blocked dependence of SICK alarm and warning “External laser enable”, from the installation of a beam switch
- possibility viewing values of FFBD thresholds in logs
- Digital potentiometer and user digital potentiometer values in configuration view
2.117.10/18 - Help IDs for stripping modules page
Fixed:
- problem with check boxes in Laser System Settings
- format User_DigPot in logs
- some bugs
- Help IDs on chiller alarms, status and warnings pages
Form: 19002
Software description Revision: 01
Seite/Page: 23 of 29
Fixed:
- Problem with running on Windows XP
Added:
- Monitoring power supply more than four (IG412 HPL)
- FPGA firmware update (IG412 HPL)
2.117.19/25
- Fieldbus protocol Kuka Welding
- Fieldbus protocol QCW V2
- Fieldbus protocol QCWExt V2
- Warning bit for laser water flow
- Short description for Status and Alarms in log viewer for MO module (IG412 RDPL)
Added:
- new description for laser program interruption (incorrect QCW operational mode)
- in config and log view serial number of modules (min WCELaser 3.231)
2.117.26/28
- Reactivation modules (min WCELaser 3.231)
Fixed:
- QCW Laser Program Editor for QCW Ext V2 protocol (WAIT->EXT)
Fixed:
- configuration of swap bytes for PC-104 FieldBus Interfaces.
Added
- support monitoring temperature of combiner
2.117.29/32 - add alarm and warning for combiner temperature
- view combiners temperature in log
- support 200 CW laser programs
- swap bytes for FieldBus COMX interfaces (IG412)
- added alarm and warning “Laser Module(s) has no emission”
Added
2.117.33
- FBUS_W4 in EXTPWR (KukaWelding)
Fixed:
2.117.34
- writing settings in advanced settings dialog
Fixed:
- some bugs
2.118.1/6 Added:
- some events for HP RDPL lasers
- support environment sensors
2.119.0/7 Fixed:
- view for Air Cooled Laser with beam coupler
Form: 19002
Software description Revision: 01
Seite/Page: 24 of 29
Added:
- IPG chiller serial number for IG277 and IG320 (IG277 v2.91, IG320 v3.74, Chiller Firmware 2.41)
- showing description of Error Code for CAN Power Supply.
- Parameters (Scattering, Temp, DewPoint, Humidity) in log for Fiber Coupler (RDPL, IG412)
2.222.0/ - Dual feeding fiber laser configuration (IG412)
2.223.0 - Firmware update for Pulse Generator (IG412 v.6.26, WCELaserArm 5.64)
- Limitation for commands WAIT, GOTO in QCW Laser Program
- setting for changing led color (Green<->Yellow) and event type (Info<->Warning) for Water supplying (IPG Chiller)
Fixed:
- FieldBus debug window
- Low refrigerant pressure value for IPG Chiller (Type – 6)
- Displaying thresholds on the chiller page
2.223.1/15 Added:
2.224.0 - PowCheckCounter and PowCheckDeviation in log view for output power check option
- FF Flow Meters for Dual Feeding Fiber configuration for IG277 (MCU 2.93)
- In log view Read/Write DPM memory counters per second (WCELaseArm 5.69.2, WCELaserX86 7.10.1)
- In log view OSImage version (WCELaseArm 5.69.2, WCELaserX86 7.10.1)
- Fieldbus protocol Bystronic
- New leds (Channel overload, Standby mode) for IPG CAN power supply
Fixed:
- Digital potentiometer, % in configuration view (IG412)
- Changed firmware update timeout for Pulse Generator (IG412)
- View of Environment Sensors (IG488) log
- Firmware update for laser modules (IG243, IG512)
Form: 19002
Software description Revision: 01
Seite/Page: 26 of 29
Changed:
- alarms levels on HPP Page can be viewed only in IPG Service Mode
Fixed:
- set channel numbers in FieldBus settings dialog
Added:
- for peak hold power and energy levels (IG412 HPP)
- visual limitation of the frequency and duration of the pulse (IG412 HPP)
2.229.0/2
- alarm duty cycle to high (IG412 HPP)
Fixed:
- incorrect display of the chiller page
Fixed:
- Adjusted time column in the viewing window of log file for laser module.
- program freezing in case of recording a monitor file.
2.230.0/3
- writing DewPoint values in Monitor File
Changed:
- Disabled menu Program Editor for lasers with IPGP scanner configuration.
Added:
- disable binding optical channel number to field bus slot (VASS protocols, except VASS1, MCU IG412 v.6.62, MCU IG320 v.3.86)
- possibility to reset Supervisor password
2.231.0/4
Fixed:
- selection of laser modules for a 120 kilowatt laser.
- multi chiller interface (IG412)
Fixed:
- view of KukaTechPack fieldbus protocol adjusted to specification
2.232.0/1 - incorrect showing Beam Switch page if laser with Beam Coupler.
Added:
- support DLS laser (IG412 MCU 9.*, WCELaserArm 5.84)
- support AMB2 laser (IG412 MCU 10.*, WCELaserArm 5.84)
2.233.0/12
Fixed:
- time in log viewer for lasers without dynamic structures
- time in command WAIT of laser program
Form: 19002
Software description Revision: 01
Seite/Page: 28 of 29
Fixed:
- program crashes when switching to fieldbus page
2.233.13/14
Fixed:
- incorrect display of water flowmeter 2 for laser configuration with FF flow switch.
2.233.15
Fixed:
- no display of the ramp time on the control page
2.233.16/19 - data on chiller page for AMB2 laser
Changed:
- removed the ability to update the web interface
Added:
- support AMB2 laser based on IG612 (MCU IG612 12.X, WCE 5.85)
- support laser modules based on IG606
Fixed:
2.234.0/4
- reset Supervisor password
- random show command SPTP in Pulse Program Editor.
Changed:
- alarms to warnings for EnvironmentSensors (IG488)
Added:
- support house (customer) water pressure difference for IPG chiller (WCE 5.85.6, IG412 MCU 6.75, CHILLER 2.46 or 3.10)
- support HPP mode for Hardwiring
- support SPTA command in laser programs (valid only for AMB2 lasers based on IG612)
2.235.0/4 - additional checks when loading a log file
- disallowed control on pulse generator page for monitor access level
- control for enable Ext. Start activates low power mode in Laser->Advanced settings dialog (IG412 MCU 6.75)
Fixed:
- save events to file
Form: 19002
Software description Revision: 01
Seite/Page: 29 of 29
Fixed:
- program crash on close Laser->Advanced settings dialog with OK button.
2.236.0/1 - changing IP address of the laser.
Fixed:
- output power display for center and ring (AMB2 laser)
- laser program SPTA command for FBUS parameters
2.237.0/4
- support peak power for HPP lasers (IG412 MCU 6.76)
- the state of the HPP mode button on the HPP mode page