Professional Documents
Culture Documents
SSE 7-INF.4 - Dcacacas...
SSE 7-INF.4 - Dcacacas...
Information and interim report evaluating dry chemical powder agents according to
standard ISO 7202 criteria
SUMMARY
Output: OW 39
Action to be taken: 8
Related documents: MSC 98/23; SSE 5/9, SEE 5/17; SSE 6/7, SSE 6/18; SSE 7/7 and
SSE 7/7/1
General
1 At its ninety-eighth session, the Maritime Safety Committee (MSC) tasked the
Sub-Committee to review and amend the Guidelines for the approval of fixed dry chemical
powder fire-extinguishing systems for the protection of ships carrying liquefied gases in bulk
(MSC.1/Circ.1315).
I:\SSE\07\SSE 7-INF.4.docx
SSE 7/INF.4
Page 2
3 The test methods and criteria in standard ISO 7202 include fire test procedures, as
well as material testing of dry powder agents to determine their properties, including fluidity
and susceptibility to caking and clogging.
4 The United States sponsored research on the standard ISO 7202 test methods and
acceptability criteria to assess the suitability of standard ISO 7202 for incorporation into
MSC.1/Circ.1315. This research addresses concerns related to clogging, caking, clumping and
fire performance of different fire suppression agents. The agents tested in this research include
sodium bicarbonate, potassium bicarbonate and agents based on neither sodium nor
potassium.
5 The United States developed an interim report based on the results gathered to date,
as set out in the annex. The research project is not complete, however, a final report will be
published prior to this session.
7 Once the full testing program is complete, the final report may be obtained in portable
data format (PDF) from the U.S. Coast Guard Office of Design and Engineering Standards,
Lifesaving and Fire Protection Division (CG-ENG-4) web page using the following URL:
https://www.dco.uscg.mil/CG-ENG-4/Research/
8 The Sub-Committee is invited to note the results of the testing when considering
document SSE 7/7/1, as set out in the annex, and to consider the findings in the development
of updates to the MSC.1/Circ.1315.
***
I:\SSE\07\SSE 7-INF.4.docx
ANNEX SSE 7/INF.4
0
EVALUATION OF SODIUM AND POTASSIUM
CHEMICAL FIRE EXTINGUISHING STANDARDS
Prepared for:
Jason P. Huczek
Principal Engineer Director
Engineering and Research Section Fire Technology Department
This report is for the information of the client. It may be used in its entirety for the purpose of securing product acceptance from duly
constituted approval authorities. This report shall not be reproduced except in full, without the written approval of SwRI. Neither this
report nor the name of the lnntitute shall be used in publicity or advertiaing.
Benefiting government, industry and the public through innovalive science and technology
TABLE OF CONTENTS
PAGE
1.0 BACKGROUND ................................................................................................................... 5
2.0 TEST SPECIMENS................................................................................................................ 6
3.0 TEST PROCEDURES ............................................................................................................ 8
3.1 ISO 7202........................................................................................................................ 8
3.1.2 Section 13.5 – Test for Resistance to Caking and Clumping ................................ 9
4.2 ISO 7202 – Section 13.5 – Test for Resistance to Caking and Clumping ................... 15
4.4 ISO 7202 – Section 13.7 – Test Method for Moisture Content ................................... 18
It is also of interest to investigate the performance effects of dry chemical agents subjected to
thermal cycling and environmental conditioning. After evaluating specimens to ISO 7202 and UL
299, an assessment of the test methods will be conducted in order to identify if the standards adequately
predict the tendency of an agent to clog or cake in the marine environment.
After the preliminary testing of dry chemical agent properties, pool fire testing will be
conducted per ISO 7165, Fire-Fighting — Portable Fire Extinguishers — Performance and
Construction (2017 Edition) and UL 711, Rating and Fire Testing of Fire Extinguishers (2018
Edition). Testing will be conducted using both weathered and not weathered, commercially available
sodium and potassium-based fire extinguishing dry chemical systems as well as the Contractor made
dry chemical mixtures to evaluate their firefighting performance, as well as their caking, clogging, and
discharge characteristics.
Phase 1 testing per ISO 7202 and UL 299 was initiated on September 26, 2019, and completed
on November 26, 2019, by Southwest Research Institute’s (SwRI) Fire Technology Department
(FTD), located in San Antonio, Texas.
United States Coast Guard 5 SwRI Project No.: 01.24804.01.001
The results presented in this report apply only to the materials tested, in the manner tested,
and not to any similar materials or material combinations. This testing was conducted in accordance
with the relevant sections of the applicable standards and, to the best of our knowledge, it contains
no errors, omissions, or false statements.
Na1a 100 0 0 0 No No
Sodium Bicarbonate (NaHCO3)
SwRI procured the base dry chemical from Earthborn Elements. Both were Food and USP
pharmaceutical grade. For the additives, silica dioxide (CAS: 68611-44-9) and Attapulgite clay (CAS:
12174-11-7) were chosen based on a review of SDS information from several commercial products.
These were both procured from EastChem. Figure 1 shows a photograph of each of these materials.
The samples were mixed in one of three ways depending on Table 1. For samples that did
not require milling or sifting per the matrix, they were simply mixed with a stand mixer. For samples
that were milled and sifted, a flour/grain style mill was used and subsequent to the milling, the sample
was sifted with an electric sieve/shaker appliance. Figure 2 shows a photograph of each of these
devices.
Figure 2. Left: Stand Mixer, Center: Flour/Grain Mill, Right: Sieve Shaker.
The purpose of the weathering of a subset of the samples was to evaluate if a given product
was more or less susceptible to clogging or caking when subjected to marine environmental conditions.
The weathered samples were conditioned in metal tin cans of a nominal 1-gal capacity. The
exposures were carried out with the lids off these sample containers. The environmental exposure
testing was performed by the Structural Dynamics group in the Mechanical Engineering Division at
SwRI. Figure 3 shows a photograph of the thermal chamber and salt fog apparatus used for this part
of testing.
In order to approximate the marine environment, two types of extended exposures were
imposed on the weathered samples. First, they were subjected to thermal cycling in accordance with
MIL-STD-810H, Environmental Engineering Considerations and Laboratory Tests. This test
consisted of 10 cycles, with each cycle having an 8 h dwell at -30°C and an 8 h dwell at +65°C. The
transition between temperatures was maintained to be no more than 3°C per minute to avoid thermal
shock conditions. Testing was initiated by heating the samples to 65 ºC at a rate of 2-3 ºC/min. The
temperature was held at 65 ºC for 8 h and then ramped down to -30 ºC at 2-3 ºC. This temperature
The day after thermal cycling was complete, the specimens were subjected to salt fog exposure
using ASTM B117-18, Standard Practice for Operating Salt Spray (Fog) Apparatus, as guidance.
This test involves exposure to a fog generated from 5% salt water solution (by mass) with a fallout
rate and pH that comply with the specification. The temperature in the exposure zone of the salt/fog
chamber is required to be 35 ± 2°C (95 ± 3°F). While ASTM B117 does not specify an exposure
duration, 240 h (10 days) is a common duration and was used for this testing. After the exposure was
completed, the samples were allowed to come to equilibrium under standard temperature conditions
(20-25 ºC). This exposure was completed on October 14, 2019. The sample container lids were
installed and the samples were transported back to the Fire Technology for subsequent ISO 7202 and
UL 299 testing, which was conducted over the following six weeks.
Figure 4. Left: Schematic of Fluidity Test Device, Right: Photograph of SwRI’s Fluidity Device.
The Petri dish is placed in a desiccator for 24 h, maintained at 20 ± 5 ºC, that contains a
saturated sodium chloride solution, which provides approximately 75 % relative humidity. After
the desiccator exposure, the samples are moved into an oven maintained at 48 ± 3 ºC for an
additional 24-hr period. After the oven exposure, the samples are covered and allowed to cool for
1 h, after which time, the Petri dish is overturned on a clean sheet of paper for further evaluation.
The powder is allowed to slide into a 425 µm sieve such that any formed lumps are not
crushed. The sieve is shaken gently with horizontal revolving movements to separate the formed
lumps from free flowing powder without crushing the lumps.
A spatula is used to lift the lumps, drop them from a height of 200 ± 10 mm onto a clean
sheet of paper, which is placed on a hard surface. The process is repeated and any remaining powder
is sieved and any remaining lumps are dropped a second time.
Figure 5. Left: Desiccator with Salt Solution, Right: Oven used for Temperature Exposure.
3.1.3 Section 13.6 – Water Repellence Test
Samples are prepared in the same way in the same type of Petri dish as described in the
previous section. On three different areas of the powder surface, a drop of distilled water is applied
gently. The Petri dish is placed in the same desiccator as previously described for 2 h at 20 ± 5 ºC.
The Petri dish is removed and the drops are examined. Per Section 9 of ISO 7202, there shall be no
complete absorption of the water droplets into the powder. Figure 6 shows two photographs
illustrating complete absorption and no absorption.
Figure 6. Left: Samples that completely absorbed Water Droplets, Right: Samples that did not
completely absorb the Water Droplets.
3.2 UL 299
Section 48.1.1 of UL 299 stipulates that when subjected to the tests specified in 48.2.1 –
48.3.3, the dry chemical specimen shall be free flowing, except that should there be any tendency
to lump, harden, or cake, such portions shall be friable when dropped from a height of 100 mm (4 in)
onto a smooth, hard surface. Prior to the tests, all dry chemical samples are conditioned in a
desiccator containing anhydrous calcium chloride for 48 hours at room temperature. Each of the
following subsections provides additional details for the specific test methods utilized in this project.
3.2.1 Section 48.2 – Elevated Temperature Test
The procedure requires that a 150 g specimen of powder be placed inside seamless tinned
steel cups measuring approximately 75 mm (3 in) in diameter and 50 mm (2 in) deep with closely
fitting flanged covers. The specimens are then placed in an oven maintained at 60 ºC (140ºF). After
one week, the samples are removed from the oven, allowed to cool for three days and then examined
for evidence of caking. Any lumps present are to be dropped from a height of 100 mm (4 in) onto
a smooth hard surface to determine whether they are friable. Per Section 48.1.1 of UL 299, any
lumps present shall be friable. Figure 7 shows the tin cans for sample preparation and the oven used
for the testing.
Figure 7. Left: Samples in Tin Cans, Right: Oven for Elevated Temperature Test.
Most modern test equipment of this type can also be used to make energy measurements,
which are derived based on temperature rise of the instrument in specific locations (typically, at or
just under the sample cup). ISO 11358-1 makes reference to ISO 11357-1, Plastics — Differential
Scanning Calorimetry (DSC) — Part 1: General Principles, for this application.
In general, the reactions which cause the mass of a test specimen to change are
decomposition or oxidation reactions or the volatilization of a component. The change in mass is
recorded as a thermogravimetric (TG) curve.
The change in mass of a material as a function of temperature and the extent of this change
are indicators of the thermal stability of the material. TG data can therefore be used to evaluate the
relative thermal stability of polymers of the same generic family and polymer-polymer or polymer-
additive interactions, using measurements made under the same test conditions.
It is also of note that TG data may be used for process control, process development and
material evaluation. Long-term thermal stability is a complex function of service and environmental
condition, therefore, TG data alone may not be able to describe the long-term thermal stability of a
polymer.
Figure 9 shows SwRI’s TGA/DSC apparatus, which is a Netzsch Model STA 449 F3
Jupiter.
12
10
8
6
4
2
0
Na1a Na1b Na2a Na2b Na3a Na3b Na4a Na4b K1a K1b K2a K2b K3a K3b K4a K4b
Sample ID
Figure 10. Left: Schematic of Fluidity Test Device, Right: Photograph of SwRI’s Fluidity Device.
Table 3. Phase 1 Test Results – ISO 7202 Section 13.5 (Sodium-Based Samples).
still remaining
same as
some caking, same as same as replicate lumps after same as FAIL
K1a small lumps replicate 1 1 and 2 second replicate 1
replicate 1 and
2
sieve/drop
still remaining
very heavy same as
same as same as replicate lumps after same as FAIL
K2a caking, large
replicate 1 1 and 2 second replicate 1
replicate 1 and
lumps 2
sieve/drop
seemed like
silica dioxide still same
same as
formed very same as same as replicate small lumps same as FAIL
K3a small lumps replicate 1 1 and 2 after second replicate 1
replicate 1 and
2
that would not sieve
break apart
yes, still
fine powder,
remaining same as
lots of very same as same as replicate same as FAIL
K3b small masses replicate 1 1 and 2
small masses
replicate 1
replicate 1 and
after second 2
after sieve/drop
sieve/drop
no caking, no
lumping, no
same as same as replicate PASS
K4a second
replicate 1 1 and 2
N/A N/A N/A
sieve/drop
required
yes, still
fine powder,
remaining same as
lots of very same as same as replicate same as FAIL
K4b small masses replicate 1 1 and 2
small masses
replicate 1
replicate 1 and
after second 2
after sieve/drop
sieve/drop
Four out of the sixteen samples met the test requirements. The repeatability between
replicates was good and in only one case were replicate results slightly different (Na1b), although
this didn’t affect the pass/fail result. The sodium-based samples (3 out of 8 passed) performed
slightly better than the potassium-based samples (1 out of 8 passed). Overall, the weathered samples
performed poorly compared to the non-weathered samples.
Figure 11 shows a photograph array of the results of this testing. These photographs, along
with the observations in Tables 3 and 4 are used to assign ranks to the sample results for future
analysis of data trends.
Na3a, Fail, Rank 5 Na3b, Fail, Rank 5 Na4a, Pass, Rank 1 Na4b, Fail, Rank 5
K1a, Fail, Rank 11 K1b, Fail, Rank 13 K2a, Fail, Rank 12 K2b, Fail, Rank 13
K3a, Fail, Rank 5 K3b, Fail, Rank 5 K4a, Pass, Rank 1 K4b, Fail, Rank 5
Figure 11. Photograph Array of Final Results for Testing per ISO 7202, Section 13.5.
There were three basic types of performance in this test. The first kind of result was that
the drops immediately soaked into the sample and were completely absorbed. The second kind of
result was that the drops soaked into the sample slightly, but not completely. The third kind of result
was that the drops did not soak into the sample at all. These three types of result aided in assigning
rank to the results, described in the next section of this report.
For this testing, the presence of additives seemed to be the main driver of good performance.
Overall, the weathered samples performed slightly worse compared to the non-weathered samples,
but not to the point where it changed the pass/fail result for a given mixture.
Table 5. Phase 1 Test Results – ISO 7202 Section 13.6 (Water Repellence).
Are Drops Completely Absorbed by
the Powder?
Sample PASS /
Additional Comments
ID FAIL?
Replicate 1 Replicate 2 Replicate 3
Na1a Yes Yes Yes FAIL drops immediately soaked into sample
Na1b Yes Yes Yes FAIL drops immediately soaked into sample
Na2a Yes Yes Yes FAIL drops immediately soaked into sample
Na2b Yes Yes Yes FAIL drops immediately soaked into sample
Na3a No No No PASS drops intact sitting on top of samples
drops seemed to soak into the sample a
Na3b No No No PASS
little bit, but not completely
Na4a No No No PASS drops intact sitting on top of samples
drops seemed to soak into the sample a
Na4b No No No PASS
little bit, but not completely
K1a Yes Yes Yes FAIL drops immediately soaked into sample
K1b Yes Yes Yes FAIL drops immediately soaked into sample
K2a Yes Yes Yes FAIL drops immediately soaked into sample
K2b Yes Yes Yes FAIL drops immediately soaked into sample
K3a No No No PASS drops intact sitting on top of samples
drops seemed to soak into the sample a
K3b No No No PASS
little bit, but not completely
drops intact, but seemed to soak into
K4a No No No PASS
sample slightly
drops seemed to soak into the sample a
K4b No No No PASS
little bit, but not completely
4.4 ISO 7202 – Section 13.7 – Test Method for Moisture Content
The procedure requires that a 20 g specimen of powder be placed inside a Petri dish, which
is placed is a desiccator, with concentrated sulfuric acid, for 48 h at a temperature of 20 ± 3 ºC.
After removal of the Petri dish from the desiccator, the sample is reweighed and the weight loss is
Table 6 provides the results of this testing. Figure 12 shows this data graphically and Figure
13 shows the same data, zoomed in on the failure threshold of a 0.25% moisture content mass
fraction, which is depicted by a black dotted line.
All of the weathered samples had a much higher moisture content than the non-weathered
samples, which is understanding considering they were exposed to salt/fog conditions with the lids
of the batch containers removed. For the non-weathered specimens, the potassium-based samples
exhibited a lower moisture content than the sodium-based samples. Also, the milling/sifting process
for the sodium-based samples seemed to aerate the specimens slightly, which seemed to impact the
ability for the water to be absorbed into the powder (e.g., compare Na1a to Na2a or Na3a to Na4a).
This effect was not seen for the potassium-based samples.
Table 6. Phase 1 Test Results – ISO 7202 Section 13.7 (Moisture Content).
Initial Mass (g) Final Mass (g) Mass Fraction (%)
Average
Sample Mass PASS /
#1 #2 #3 #1 #2 #3 #1 #2 #3
ID Fraction FAIL?
(%)
Na1a 20.000 20.000 20.000 20.000 20.000 20.000 0.000 0.000 0.000 0.000 PASS
Na1b 20.001 20.001 20.001 19.242 19.219 19.257 3.795 3.910 3.720 3.808 FAIL
Na2a 20.000 20.000 20.000 19.991 20.000 19.994 0.045 0.000 0.030 0.025 PASS
Na2b 20.001 20.001 20.001 19.091 19.380 19.305 4.550 3.105 3.480 3.711 FAIL
Na3a 20.000 20.000 20.000 19.985 19.994 19.989 0.075 0.030 0.055 0.054 PASS
Na3b 20.001 20.001 20.001 19.650 19.762 19.720 1.755 1.195 1.405 1.452 FAIL
Na4a 20.000 20.000 20.000 19.964 19.981 19.978 0.180 0.095 0.110 0.128 PASS
Na4b 20.001 20.001 20.001 19.792 19.682 19.681 1.045 1.595 1.600 1.413 FAIL
K1a 20.000 20.000 20.000 20.000 20.000 20.000 0.000 0.000 0.000 0.000 PASS
K1b 20.000 20.001 20.001 18.027 18.122 18.313 9.865 9.395 8.440 9.233 FAIL
K2a 20.000 20.000 20.000 20.000 20.000 20.000 0.000 0.000 0.000 0.000 PASS
K2b 20.001 20.001 20.000 18.142 18.368 18.347 9.295 8.165 8.265 8.575 FAIL
K3a 20.000 20.000 20.000 20.000 20.000 20.000 0.000 0.000 0.000 0.000 PASS
K3b 20.001 20.001 20.001 19.753 19.755 19.738 1.240 1.230 1.315 1.262 FAIL
K4a 20.000 20.000 20.000 20.000 20.000 20.000 0.000 0.000 0.000 0.000 PASS
K4b 20.001 20.001 20.001 19.204 19.150 19.221 3.985 4.255 3.900 4.046 FAIL
0.45
Average Water Mass Fraction (%)
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05
0.00
Na1a Na1b Na2a Na2b Na3a Na3b Na4a Na4b K1a K1b K2a K2b K3a K3b K4a K4b
Sample ID
Figure 13. Graphical Results of ISO 7202, Section 13.7 Testing (Zoomed in on Y-Axis).
On average, the sodium-based samples performed better than the potassium-based samples
and that was especially the case for the samples with additives.
Table 7. Phase 1 Test Results – ISO 7202 Section 13.9 (Moisture Absorption).
Moisture Absorption (% w/w)
Average
Moisture PASS /
Sample ID RANK
Replicate 1 Replicate 2 Replicate 3 Absorption FAIL?
(%)
0.80
0.70
Moisture Absorption (%)
0.60
0.50
0.40
0.30
0.20
0.10
0.00
Na1a Na1b Na2a Na2b Na3a Na3b Na4a Na4b K1a K1b K2a K2b K3a K3b K4a K4b
Sample ID
This test method is similar in principle to Section 13.5 of ISO 7202. However, the exposure
is longer and at a higher temperature in the standard UL test, as compared to the ISO test. Also, the
drop test in this UL method is from 100 mm (4 in.), whereas the drop is from 200 mm (8 in.) in the
ISO test method. Finally, the UL method doesn’t specify a sieve for friability quantification and
only says that the lumps should be friable, which is interpreted that they break apart when dropped,
not that they break to a specific particle size when dropped.
Table 8 shows the results for the sodium-based samples when tested per Section 48.2 of
UL 299. Table 9 shows the results for the potassium-based samples when tested per Section 48.2
of UL 299.
Four out of the sixteen samples met the test requirements. The repeatability between
replicates was good and in only one case were replicate results slightly different (Na4b and K1a),
although this didn’t affect the pass/fail result. The sodium-based samples (3 out of 8 passed)
Figure 15 shows a photograph array of the results of this testing. These photographs, along
with the observations in Tables 8 and 9 are used to assign ranks to the sample results for future
analysis of data trends.
some caking
Yes, several
throughout, same as
same as same as replicate non-friable same as
Na1a lots of lumps,
replicate 1 1 and 2 lumps left after replicate 1
replicate 1 and FAIL
small to big in 2
drop test
size
lots of masses Yes, several
same as
throughout same as same as replicate non-friable same as
Na1b sample, lots of replicate 1 1 and 2 lumps left after replicate 1
replicate 1 and FAIL
2
caking drop test
some caking
Yes, several
throughout, same as
same as same as replicate non-friable same as
Na2a lots of lumps,
replicate 1 1 and 2 lumps left after replicate 1
replicate 1 and FAIL
small to big in 2
drop test
size
lots of masses Yes, several
same as
throughout same as same as replicate non-friable same as
Na2b sample, lots of replicate 1 1 and 2 lumps left after replicate 1
replicate 1 and FAIL
2
caking drop test
no caking, but
seemed to have same as
same as same as replicate same as
Na3a small masses of
replicate 1 1 and 2
No lumps
replicate 1
replicate 1 and PASS
the silica 2
dioxide on top
very small
masses of silica
no lumps, just same as
dioxide on top, same as same as replicate same as
Na3b no caking other replicate 1 1 and 2
these very tiny
replicate 1
replicate 1 and PASS
SiO2 masses 2
than those
masses
No, remaining
same as
very little same as same as replicate lumps, they same as
Na4a caking replicate 1 1 and 2 completely fell replicate 1
replicate 1 and PASS
2
apart
very little similar to
Yes, several
caking, a few replicate 1 and
same as same as replicate non-friable same as
Na4b masses
replicate 1 1 and 2 lumps left after replicate 1
2, but fewer FAIL
throughout lumps
drop test
sample remaining
some caking,
some caking,
some caking, mostly small Yes, small same as
mostly small same as
K1a mostly small lumps, but
lumps, but some
lumps
replicate 1
replicate 1 and FAIL
lumps some larger remaining 2
larger lumps too
lumps too
very wet
sample with No defined
same as
consistency of same as same as replicate lumps to drop, same as
K1b sludge/mud, replicate 1 1 and 2 sample is too replicate 1
replicate 1 and FAIL
2
orange-pink wet
color
heavy caking,
Yes, little to no same as
entire sample same as same as replicate same as
K2a consists of replicate 1 1 and 2
breaking apart
replicate 1
replicate 1 and FAIL
after drop test 2
lumps
very wet
sample with No defined
same as
consistency of same as same as replicate lumps to drop, same as
K2b sludge/mud, replicate 1 1 and 2 sample is too replicate 1
replicate 1 and FAIL
2
orange-pink wet
color
almost no Yes, no
same as
caking, very same as same as replicate breaking apart same as
K3a few small replicate 1 1 and 2 of any of the replicate 1
replicate 1 and FAIL
2
lumps small lumps
small masses yes, small
same as
throughout, same as same as replicate SiO2 lumps same as
K3b look like SiO2 replicate 1 1 and 2 did not break replicate 1
replicate 1 and FAIL
2
lumps apart
same as
same as
K4a no caking no caking no caking no lumps
replicate 1
replicate 1 and PASS
2
Na3a, Pass, Rank 1 Na3b, Fail, Rank 3 Na4a, Pass, Rank 3 Na4b, Fail, Rank 8
K1a, Fail, Rank 8 K1b, Fail, Rank 15 K2a, Fail, Rank 10 K2b, Fail, Rank 15
K3a, Fail, Rank 5 K3b, Fail, Rank 5 K4a, Pass, Rank 1 K4b, Fail, Rank 5
Figure 15. Photograph Array of Final Results for Testing per UL 299, Section 48.2.
This test method is similar in principle to Section 13.9 of ISO 7202. However, the exposure
is slightly different. In section 13.9 of ISO 7202, there is only one cycle of 80% and 60% RH for a
period of 2 days each. In the UL test, the cycle is repeated for 3 weeks. Also, while the method
doesn’t specify the environment that anhydrous calcium chloride provides in equilibrium, this
information is readily available in the literature and ranges from ~15-30 % RH at the test temperature
range. So, the UL test specifies a more extreme RH range (~20 – 80 % RH). However, since it is
not required to measure mass changes during the exposure, it is difficult to compare the results
between methods. This may be something to consider in Phase 2 of the project.
Also, the drop test in this UL method is from 100 mm (4 in.), whereas the drop is from
200 mm (8 in.) in the ISO test method. Finally, the UL method doesn’t specify a sieve for friability
quantification and only says that the lumps should be friable, which is interpreted that they break
apart when dropped, not that they break to a specific particle size when dropped.
Table 10 shows the results for the sodium-based samples when tested per Section 48.2 of
UL 299. Table 11 shows the results for the potassium-based samples when tested per Section 48.2
of UL 299.
some caking
throughout, same as same as replicate
Na1a small masses replicate 1 1 and 2
No No No PASS
throughout
some caking
throughout, same as same as replicate
Na2a small masses replicate 1 1 and 2
No No No PASS
throughout
lots of silica
dioxide masses same as same as replicate No lumps to No lumps to No lumps to
Na3a on top, no replicate 1 1 and 2 test test test PASS
visible caking
some caking
throughout, same as same as replicate
Na3b small masses replicate 1 1 and 2
Yes Yes Yes FAIL
throughout
lots of masses,
same as same as replicate
Na4b caking
replicate 1 1 and 2
Yes Yes Yes FAIL
throughout
caking
throughout, a same as same as replicate
K1a couple of replicate 1 1 and 2
Yes Yes Yes FAIL
visible lumps
wet spots,
small masses same as same as replicate No lumps to No lumps to No lumps to
K3a on top of replicate 1 1 and 2 test test test
PASS
sample
small masses
same as same as replicate
K3b throughout,
replicate 1 1 and 2
Yes Yes Yes FAIL
little caking
small masses
same as same as replicate
K4b throughout,
replicate 1 1 and 2
Yes Yes Yes FAIL
little caking
Six out of the sixteen samples met the test requirements. The repeatability between
replicates was good and all the replicate results were qualitatively the same. The sodium-based
samples (4 out of 8 passed) performed slightly better than the potassium-based samples (2 out of 8
passed). Overall, the weathered samples performed poorly compared to the non-weathered samples.
Figure 16 shows a photograph array of the results of this testing. These photographs, along
with the observations in Tables 10 and 11 are used to assign ranks to the sample results for future
analysis of data trends.
Na3a, Pass, Rank 1 Na3b, Fail, Rank 9 Na4a, Pass, Rank 1 Na4b, Fail, Rank 10
K1a, Fail, Rank 13 K1b, Fail, Rank 13 K2a, Fail, Rank 13 K2b, Fail, Rank 13
K3a, Fail, Rank 1 K3b, Fail, Rank 7 K4a, Pass, Rank 1 K4b, Fail, Rank 7
Figure 16. Photograph Array of Final Results for Testing per UL 299, Section 48.3.
Figure 17 and Figure 18 show the preliminary results for the K1a and Na1a samples,
respectively. Figure 19 shows preliminary results for Na3a samples.
For the limited replicate tests that have been performed to date, the repeatability is good.
The initial major transition temperature is approximately the same for potassium-based samples
(~120 – 140 ºC) and sodium-based samples (~100 – 120 ºC). For samples with additives, the onset
of this second transition appears to start at 750 – 800 ºC, but this will be confirmed after testing is
complete.
Figure 17. TGA/DSC Test Results for K1a at 5 K/min Ramp Rate.
Figure 19. TGA/DSC Test Results for Na3a at 5 K/min Ramp Rate.
The initial approach to that problem is to assign ranking values to each sample ID for each
test method and to sum these up and plot them in different ways to try to determine which materials
perform best and which test variables seem to drive the results. For test methods that have
quantitative results, the ranking is straightforward (i.e. lower ranking = better result) and these tests
were given a value between 1 and 16 depending on severity of pass/fail. This was the case for
Sections 13.4, 13.7, and 13.9 of ISO 7202. For the remaining test methods (Section 13.5 and 13.6
of ISO 7202 and the UL 299), some judgement was required based on a comparative analysis of the
test observations and the photographs. These rankings helped separate performance between
samples that otherwise were only initially placed into one of two categories (i.e., pass or fail).
Tables 12 and 13 show the pass/fail results for all the samples when tested per ISO 7202
and UL 299, respectively. It can be seen that not a single sample met the requirements of all seven
test methods. For ISO 7202 only, none of the samples passed all the tests, but four of them (Na3a,
Na4a, K3a and K4a) met the requirements of four of the five of the procedures. For UL 299, there
were three samples (Na3a, Na4a and K4a) that met the requirements of both procedures.
Tables 14 and 15 show a summary of all the rankings for ISO 7202 and UL 299,
respectively. The Pass Rank is based on the results in Tables 12 and 13, the Section number rankings
reflect how each sample did in each procedure and the total rank score is the sum of all the ranks
for a given sample. The total rank score column is shaded in a gradient that reflects the ranking.
The lightest shading is the best (lowest) ranking and the shading gets darker as the ranking increases
in value. Figure 20 shows this data graphically in a stacked histogram. Figures 21 and 22 show the
same data for only the non-weathered and weathered samples, respectively.
Table 13. Phase 1 Test Results – Summary of Pass/Fail Results for UL 299.
Sample
48.2 - Elevated Temp 48.3 - Hygroscopicity
ID
Na1a FAIL PASS
Na1b FAIL FAIL
Na2a FAIL PASS
Na2b FAIL FAIL
Na3a PASS PASS
Na3b PASS FAIL
Na4a PASS PASS
Na4b FAIL FAIL
K1a FAIL FAIL
K1b FAIL FAIL
K2a FAIL FAIL
K2b FAIL FAIL
K3a FAIL PASS
K3b FAIL FAIL
K4a PASS PASS
K4b FAIL FAIL
Table 15. Phase 1 Test Results – Summary of Test Result Rankings – UL 299.
TOTAL
Sample PASS 48.2 48.3
RANK
ID RANK RANK RANK
SCORE
Na1a 4 10 5 19
Na1b 8 10 11 29
Na2a 4 10 5 19
Na2b 8 10 11 29
Na3a 1 1 1 3
Na3b 4 3 9 16
Na4a 1 3 1 5
Na4b 8 8 10 26
K1a 8 8 13 29
K1b 8 15 13 36
K2a 8 10 13 31
K2b 8 15 13 36
K3a 4 5 1 10
K3b 8 5 7 20
K4a 1 1 1 3
K4b 8 5 7 20
Figure 21. Summary of Stacked Total Rank Scores for Non-Weathered Specimens.
Table 16 organizes the total rank values by test variable. Figure 23 shows this data
graphically. The two most important test variables seem to be the presence of additives and if the
sample was weathered.
Table 16. Phase 1 Test Results – Summary of Ranking Results by Test Variable.
TOTAL
TOTAL TOTAL
Agent Milling / TOTAL by by
Weathering? Additives? by Base by
Base Sifting? Weathering Milling /
(Na or K) Additives
Sifting
Milling /
Na Weathered Additives 417 557 283 459
Sifting
No
Non- No
K Milling / 480 340 614 438
Weathered Additives
Sifting
___________