1 Overview
1.1 Major Differences in Specifications between REGIUS110 and REGIUS190 ................................ 1-1
1.2 Changes in CS-2/CS-3 Program to Support REGIUS110 ............................................................. 1-1
1.3 Specification Comparison .............................................................................................................. 1-2
1.4 Restrictions & Default Values ........................................................................................................ 1-5
1.4.1 CS-1/2/3 Program Version supporting REGIUS 110 ........................................................... 1-5
1.4.2 Default Value of Host Name & IP Address .......................................................................... 1-5
1.4.3 Setting the Host Name ........................................................................................................ 1-5
1.4.4 Setting the Exam Keys ........................................................................................................ 1-6
1.4.5 Setting the IP Address......................................................................................................... 1-6
1.5 REGIUS 110 Error Code ............................................................................................................... 1-7
1.6 Job Table of Job Manager ........................................................................................................... 1-10
i
Chap.6 Service Tool Screens
6.1 Overview of Service Tool............................................................................................................... 6-1
6.2 [Console] > [System Info.] Screen ................................................................................................. 6-2
6.2.1 "SYSTEM INFO" Screen ..................................................................................................... 6-2
6.2.2 "SYSTEM INFO • System Config.(Reader)" Screen ........................................................... 6-2
6.2.3 "SYSTEM INFO • Remote Maintenance" Screen............................................................... 6-3
6.3 [Reader] > [Reader] > "Dedicated Reader".................................................................................. 6-4
6.3.1 Common Displays of "CCU INFO" Screen .......................................................................... 6-4
6.3.2 "CCU INFO • TCP/IP" Screen ............................................................................................ 6-4
6.4 [Reader] > [Device ] ....................................................................................................................... 6-5
6.4.1 "Device" Screen (for REGIUS 110)...................................................................................... 6-5
6.4.2 "Table Management (for REGIUS 110)" Screen .................................................................. 6-6
6.5 [Reader] > [PCB]............................................................................................................................ 6-8
6.5.1 "REGIUS 170 PCB" Screen ................................................................................................ 6-8
6.6 [Reader] > [Std. Current]................................................................................................................ 6-9
6.6.1 "Standard Current" Screen .................................................................................................. 6-9
6.7 [Reader] > [Uneven. Cal.] ............................................................................................................ 6-10
6.7.1 "Unevenness Calibration" Screen ..................................................................................... 6-10
6.8 [Reader] > [Sensitivity]................................................................................................................. 6-12
6.8.1 "Sensitivity Calibration" Screen ......................................................................................... 6-12
6.9 [Reader] > [Reader Log] .............................................................................................................. 6-14
6.9.1 "Reader Log" Screen......................................................................................................... 6-14
6.10 [Reader] > [Mech. Counter]......................................................................................................... 6-15
6.10.1"Mech. Counter" Screen .................................................................................................... 6-15
6.11 [Reader] > [File Import], [Upgrade] .............................................................................................. 6-16
6.11.1 "Install File" Screen............................................................................................................ 6-16
6.12 [Reader] > [Network].................................................................................................................... 6-17
6.12.1"Network" screen ............................................................................................................... 6-17
ii
Chap.1
O v e r v i e w
How to connect and network CS-1/2/3 with REGIUS
110 is outlined in this chapter.
The procedure described in this manual is an example of
CS-2 and REGIUS 110. The same procedure can be
applied to the case of CS-1/2/3.
Blank Page
1.1 Major Differences in Specifications between REGIUS110 and REGIUS190
1.1 Major Differences in 1) Further downsizing (Compared to REGIUS 190: W130% x D60% x
Specifications H60%) is achieved, requiring an installation space smaller than that
required for REGIUS 190.
between REGIUS110
2) Suitable to the operation where the production capacity as that of REGIUS
and REGIUS190 190 is not needed.
Mono-Slot System, 45sec for Cassette Feed/Load time, 80 plates/hr for
reading throughput.
3) Mammogram examination is not available on REGIUS 110.
4) Origin of the REGIUS Plate where the reading starts is different between
REGIUS 110 and REGIUS190.
REGIUS 110:Upper right of the plate
REGIUS 190:Lower right of the plate
5) REGIUS 110 requires Service Tool partially different from REGIUS 190.
1.2 Changes in CS-2/ 1) The read start point of REGIUS 110 is changed to the upper right. Accord-
CS-3 Program to ingly the program has been changed to support the new read start point.
The image starts appearing from the top to down in the image field when
Support REGIUS110 the image data is received from REGIUS 110.
2) A new Service Tool special to REGIUS 110 is added. Some changes are
also made to the existing Service Tool due to the addition of support for
REGIUS 110.
"Sub-Log" tool is not available for REGIUS 110.
「6.1 Overview of Service Tool」
3) Icons indicated in each screen have been changed from the illustration of
REGIUS 170/190 to cassette.
1-1
1.3 Specification Comparison
1.3 Specification Com- Specifications of REGIUS 110 and REGIUS 190 are shown below.
parison
■ Specification of Main Unit
Adjustable Level Difference in floor level at the Foot Adjuster : ± Turn the adjuster foot to lower the level. Fix the foot
above Floor 12mm or less so that the rubber part of the adjuster foot is slightly
depressed, caster on the front is in contact with floor
and the device is not easily moved to the right and
left.
Power Consumption 185Wh or less (when 14"x17"plate of exposure dose maximum 1200W (1032kcal/h) Rated
of 40mR or less is in process) Consumption:1100W(946kcal/h)
Consumable Parts Erasure Lamp Unit (Erasure Frequency: approx. Erasure Lamp Unit (Reading/Erasure Frequency:
30,000shots) approx. 10,000shots)
Accessaries 100V Power Cable (3m), Operation Manual Power Cable (3m), Operation Manual
Options REGIUS Cassette, REGIUS Plate, JM, REGIUS REGIUS Cassette, REGIUS Plate, REGIUS Console,
Console (CS-1/CS-2/CS-3/REGIUS Unitea) Monitor Arm, Operation Table/PC Rack, Monitor
Fixing Bracket
■ Usable Cassettes
Cassette sizes 14" × 17", 14" × 14", 11" × 14", 14" × 17", 14" × 14", 11" × 14",
capable to be read 10" × 12", 8" × 10", 24 × 30cm, 10" × 12", 8" × 10", 24 × 30cm,
18 × 24cm, 15 × 30cm 18 × 24cm, 15 × 30cm
Cassette types RC-110T(for stitched image),RC-110L(for deciding RC-110T(for stitched image),RC-110L(for deciding
exclusive to exposure the treatment position ) the treatment position )
Cassette sizes RC-110T :4 sizes (10" × 36", 11" × 28", 14" × RC-110T :4 sizes (10" × 36", 11" × 28", 14" ×
exclusive to exposure 42", 14" × 51") 42", 14" × 51")
RC-110L :3 sizes (14" × 17", 14" × 14", 10" × RC-110L :3 sizes (14" × 17", 14" × 14", 10" ×
12" 12"
1-2
1.3 Specification Comparison
Cassette Feed/Load 45sec or less (cassette size:14" × 14"/normal approx.40sec or less (image size 14" × 14", at
Time reading mode) reading mode of 175μm)
(Time required from
cassette insertion to
cassette ejection)
Processing Capacity 80 plates or more (cassette size:14" × 14"/normal approx.90 plates or more (image size 14" × 14", at
(Cassettes processed reading mode) reading mode of 175μm)
per hour when they
are inserted
continuously)
QR Value 3 values selectable between 125 ~ 500. 3 values selectable between 125 ~ 500.
・Low Sensitivity (QR Value: 125 ~ 249) ・Low Sensitivity (QR Value: 125 ~ 249)
・Normal Sensitivity (QR Value: 250) ・Normal Sensitivity (QR Value: 250)
・High Sensitivity (QR Value: 251 ~ 500) ・High Sensitivity (QR Value: 251 ~ 500)
Definition of QR:QR=200 is defined when 1535- Definition of QR:QR=200 is defined when 1535-
step is resulted from 12bit output when exposure is step is resulted from 12bit output when exposure is
made under the condition of exposure dose at 2.58 × made under the condition of exposure dose at 2.58 ×
10-7 C/kg (= 1mR)and the tube voltage at 80kV. 10-7 C/kg (= 1mR).
Range (reading) Normal D Range (4.0 digits: PMT output current Normal D Range (4.0 digits: PMT output current
range: 100nA to 1mA) range: 100nA to 1mA)
1-3
1.3 Specification Comparison
1-4
1.4 Restrictions & Default Values
1.4.1 CS-1/2/3 Program Ver- REGIUS 110 can be connected to the console having the program version
sion supporting listed below.
REGIUS 110
Console Version
CS-1 V2.32R00 or later
1.4.2 Default Value of Host In TCP/IP networks, all devices connected to the network need an
individual host name (PC name) and IP address. The network IP
Name & IP Address
address and host name must be unique.
In the factory default, IP address and host name of REGIUS 110 are
fixed. The CS-1/2/3 and REGIUS 110 are set to the manual registration
method by default.
Remains the factory default in 1-on-1 connection unless otherwise
instructed by your site.
The following is the factory default status:
REGIUS 110
•Host name r170-0001
•IP address 192.168.20.170
•Net mask 255.255.255.0
1.4.3 Setting the Host Name To change the host name (PC name) of the REGIUS 110, set it using
the following rules.
<Important>The host name is case sensitive. Pay attention when
setting the host name.Input lower 5 digits of REGIUS
110's serial No. after "r110-".
Example : When REGIUS 110's serial No. is "00456", the
Host Name of that cassette reader should be;
r110-00456
1-5
1.4 Restrictions & Default Values
1.4.4 Setting the Exam • Exam keys can be shared among REGIUS 170/190/110.
Keys • REGIUS 110 does not support mammogram.
• When REGIUS 110 alone is connected to CS-1/2/3, delete Exam
keys specific to mammogram from the User Master.
1.4.5 Setting the IP Address To connect devices to the site’s existing network, obtain an IP address
for each device from the network administrator beforehand.
If no IP address is specified, use the following addresses.
1-6
1.5 REGIUS 110 Error Code
1.5 REGIUS 110 Following is the List of Error Codes that may be displayed when REGIUS 110
Error Code is networked.
Error Message
Code
6000 Justifier HP sensor NG (Move to HP)
1-7
1.5 REGIUS 110 Error Code
Error Message
Code
6109 Transporter Sensor NG (Pressing Position 2)
1-8
1.5 REGIUS 110 Error Code
Error Message
Code
6550 Polygon Rotation NG
1-9
1.6 Job Table of Job Manager
1. Following is the list of contents contained in the job tables, which may be
displayed on Job Manager.
Name Description
jobs:Job Table
entry_c Host name of the console with which the job is registered.
(tentative name =desti_c)
dest_c Host name of the console to which the result of implemented job is
sent.
req_r Host name of the reader that is requested to implement the job.
1 - 10
1.6 Job Table of Job Manager
Name Description
sig_proc Signal in process
0 :Does not process the signal.
1 :Completes the process up to "gain/offset correction".
2 :Completes the process up to "shading correction".
3 :Completes the process up to "shading correction"
using shading correction coefficient and QRV table,
which were set at the factory.
null :When this parameter is not applicable.
target050 Theoretical value of 16-bit image at the high voltage setting for
Sen=50.
Range :0 ~ 65535
null :When this parameter is not applicable.
target250 Theoretical value of 16-bit image at the high voltage setting for
Sen=250.
Range :0 ~ 65535
null :When this parameter is not applicable.
target500 Theoretical value of 16-bit image at the high voltage setting for
Sen=500.
Range :0 ~ 65535
null :When this parameter is not applicable.
1 - 11
1.6 Job Table of Job Manager
Name Description
r_rpid Plate ID detected by the reader.
Range :'00000000000000' ~ '99999999999999'
null : Displayed until the job search is implemented. Kept in
display for the job processing without plate reading.
max_val Maximum image pixels that was scanned by the reader.(8-bit repre-
sentation)
Range :0 ~ 255
null :When the command is executed while the cassette is
absent.
door Notification specific to front door error among the errors attributed
to abnormal end of job.
1 :Front door open detected.
nul :Front door in normal state.
commands:Command Table
relations:Relation Table
1 - 12
1.6 Job Table of Job Manager
Name Description
1 - 13
1.6 Job Table of Job Manager
Name Description
casin Notification specific to cassette insertion error on the reader.
2 :unable to read the barcode.
3 :read barcode is abnormal.
4 :two cassettes inserted at a time.
5 :Job not registered in the manual registration mode.
6 :two cassette inserted back to back.
7 :failed to read the cassette in the device.
null :cassette insertion successful.
mainte_c Host name of the console on which the maintenance mode is set.
i_timeout N/A
1 - 14
1.6 Job Table of Job Manager
Name Description
db_ver Version of database.
1 - 15
1.6 Job Table of Job Manager
1 - 16
Connecting CS-1/2/3
Chap.2
with REGIUS 110
2.1 Flow of Setup Pro- Setup procedure for REGIUS110 networking to CS-1/2/3 as well as that for
cedure combination of REGIUS110 + REGIUS170/190 networking to CS-1/2/3 are
described. Clauses to be referred to for each step is also listed below.
■ CS-1
Start
1. Unpacking
Unpack the CS-1 and check the accessaries, etc.
listed in the packing list.
8. Calibration
Calibrate the system reader and adjust the brightness
of the CS-1 Operation Unit.
"4.1 Checking and Adjusting the Image Size"
9. Back Up "4.2 Unevenness Calibration of Regular Cassette"
"4.3 Sensitivity Calibration of Regular Cassette"
Back up the setup data of CS-1 and parameters of
REGIUS 110/170/190, etc.
(refer to each "Installation/Service Manual" for those
Indicates the step necessary for connecting
not listed in this manual)
End REGIUS 110 additionally to the existing CS-1.
2-1
2.1 Flow of Setup Procedure
■ CS-2
Start
1. Unpacking/Connection
8. Operation Check
"3.1 Checking the Image"
Check the image output from REGIUS 110. "3.2 Backing Up the Setup Data"
"4.1 Checking and Adjusting the Image Size"
9. Back Up "4.2 Unevenness Calibration of Regular Cassette"
"4.3 Sensitivity Calibration of Regular Cassette"
Back up the setup data of CS-2, parameters of
REGIUS 110, etc.
(refer to each "Installation/Service Manual" for those
Indicates the step necessary for connecting not listed in this manual)
End REGIUS 110 additionally to the existing CS-1.
2-2
2.1 Flow of Setup Procedure
■ CS-3
Start
1. Unpacking
Unpack the CS-3 and check the accessaries, etc.
listed in the packing list.
8. Calibration
Calibrate the system reader and adjust the luminance
of the CS-3 Operation Unit.
9. Back Up
2-3
2.2 Setting Up the 1-to-1 Connection
2.2 Setting Up the 1-to-1 This section describes how to connect a single CS-1/2/3 to a single
REGIUS 110 using a 1-to-1 connection.
Connection
Small-sized networks such as connecting a single CS-1/2/3, REGIUS
110 and KonicaMinolta printer can be created without changing the
factory default settings. When a factory default CS-1/2/3 is connected
to a REGIUS 110, the manual registration method is used.
However, for the case described below, change the setting of CS-1/2/3
and REGIUS 110 according to the procedure described in this section.
• When it is necessary to change the address because the system is
connected to the site's existing network.
• To connect several 1-to-1 connection units to the same network.
Printer
Viewer, Server
Ethernet
HUB
CS-1/2/3
Serial No. 0025 REGIUS 110
Host Name CS2-0025 Serial No. 0100
IP Address 128.100.50.1 Host Name r170-0100
Subnet Mask 255.255.0.0 IP Address 128.100.50.11
Cassette Regist. Auto-Regist. Subnet Mask 255.255.0.0
2-4
2.2 Setting Up the 1-to-1 Connection
Temporary Connection of CS-1/2/3 To connect a CS-1/2/3 to a REGIUS 110 via Ethernet, the devices have
and REGIUS 110 been configured so that they can communicate using the factory
default. The set up of the REGIUS 110 below shall be made by
temporary connecting the CS-1/2/3 in a 1-to-1 configuration.
2. Using a cross cable or hub, connect the CS-1/2/3 and REGIUS 110 in
a 1-to-1 configuration.
or
Cross Cable
2. Press the power switch on the CS-1/2/3 Control Unit to start up the CS-
1/2/3. Wait until the initial screen shows after the CS-1/2/3 starts up.
Wait for the REGIUS 110 to complete initialization and to show
"READY" on the LCD.
2-5
2.2 Setting Up the 1-to-1 Connection
Setting Up the REGIUS 110 Following procedures should be initiated by CS-1/2/3 Control Unit.
2-6
2.2 Setting Up the 1-to-1 Connection
2. Start the "Service Tool" from the "REGIUS Service" screen, and open the
"Network Setup" screen. (click [Reader] -> [Network])
Network information received from the REGIUS 110 will be displayed on the
"Network Setup" screen.
"CS-1" has been registered as the Host on the factory default on REGIUS 110.
2-7
2.2 Setting Up the 1-to-1 Connection
3. Set the network condition, host name for the REGIUS 110, CS-1/2/3 and JM
referring to the following table.
Host Name Input the Host Name of the REGIUS 110 itself.
r110-*****:***** indicates a serial number.
DBHOST
Host Name Input the Host Name of the CS-1/2/3 that is connected
to the REGIUS 110.
4. Click [Send].
Settings displayed in the screen will be sent to the REGIUS 110 and the net-
work information is updated.
• The updated settings will be written in a CF (Compact Flash Memory) of
the REGIUS 110.
5. Click [End] to close the "Network Setup" screen. Return to the "Ser-
vice Tool" screen (Reader).
2-8
2.2 Setting Up the 1-to-1 Connection
6. Press and hold the [operation] switch for 3 seconds to turn OFF the REGIUS
110.
2-9
2.2 Setting Up the 1-to-1 Connection
2.2.2 Setting Up the CS-1/2/ This section describes how to set up JM on the CS-1/2/3, the settings
3 required to connect the CS-1/2/3 to the REGIUS 110 and the network
settings for the CS-1/2/3 itself.
Communication Setting
2 - 10
2.2 Setting Up the 1-to-1 Connection
■ Setting up the JM database If the host name of the CS-1/2/3 or REGIUS 110 has been changed
from the factory default, you must use the following procedure to
update the information registered in the database.
2 - 11
2.2 Setting Up the 1-to-1 Connection
1. When the Host Name of CS-1/2/3 is changed, change the registration as fol-
lows.
• < "c_status" screen >
Rewrite "CS1-0001" with the new Host Name (uppercase).
c_name c_stat
Host Name of CS-1/2/3 2
2. When the Host Name of REGIUS 110 is changed, rewrite the name as fol-
lows.
• < "r_status" screen >
Rewrite "r170-0001" with the new Host Name (lowercase) of REGIUS
110.
r_name r_stat
Host Name of REGIUS 110 2
2 - 12
2.2 Setting Up the 1-to-1 Connection
• < "r_version" screen >
Rewrite "r170-0001" with the new Host Name (lowercase) of REGIUS
110.
r_name
Host Name of REGIUS 110
3. When the Host Name of CS-1/2/3 or REGIUS 110 is changed, change the reg-
istration as follows.
< "relations" screen >
Rewrite "CS1-0001" with the new Host Name (uppercase). When the Host
Name of REGIUS 110 is changed, click [r170-0001], and rewrite the Host
Name with the new one (lowercase).
c_name r_name
Host Name of CS-1/2/3 Host Name of REGIUS 110
2 - 13
2.2 Setting Up the 1-to-1 Connection
reg_type
0 or 1
2 - 14
2.2 Setting Up the 1-to-1 Connection
Host Name (on the Input the Host Name assigned to the REGIUS 110.
Job Manager)
<Important>Always use lowercase to
input.
1. Terminate the Service Tool and click [Windows Desktop] on the REGIUS
Service screen.
Windows desktop will be displayed.
2 - 15
2.2 Setting Up the 1-to-1 Connection
1. Click [Shutdown Option (U)] -> [Shut down (U)] of the [Start] menu to turn
OFF the CS-1/2/3.
2. Remove the Ethernet cables temporarily connected from the CS-1/2/3 Control
Unit and REGIUS 110.
2 - 16
2.3 Setting up the n-to-m Connection
2.3 Setting up the n-to- This section describes the n-to-m configuration that includes two CS-1/2/3s
m Connection and two REGIUS 110s.
In this example, the settings are based on the conditions requested by the site.
Barcode Reader
REGIUS 110 #1
Serial No. 0100
Host Name r170-0100
Ethernet IP Address 128.100.50.11
CS-1/2/3 #1(Std.JM
Serial No. 0025
Host Name CS2-0025
IP Address 128.100.50.1 HUB
Subnet Mask 255.255.0.0 Printer,
Cassette Regist. Barcode Regist. Host
REGIUS 110 #2
Serial No. 0101
Host Name r170-0101
CS-1/2/3 #2 IP Address 128.100.50.12
(Backup JM
Serial No. 0026
Host Name CS2-0026
IP Address 128.100.50.2
Subnet Mask 255.255.0.0
Cassette Regist. Barcode Regist
2 - 17
2.3 Setting up the n-to-m Connection
Temporarily connecting the CS-1/2/ To enable the CS-1/2/3 to implement the set up of REGIUS 110 via
3 and REGIUS 110 Ethernet, both have been set up at the factory so that they can
communicate using the factory setting. The set up of REGIUS 110 in
the below shall be made by temporarily connecting the CS-1/2/3 in the
"1-to-1" configuration.
or
Cross Cable
2. Press the power switch on the CS-1/2/3 Control Unit to start up the CS-
1/2/3. Wait until the initial screen shows after the CS-1/2/3 starts up.
Wait for the REGIUS 110 to complete initialization and to show
"READY" on the LCD.
2 - 18
2.3 Setting up the n-to-m Connection
Setting Up the REGIUS 110 Following procedures should be initiated by CS-1/2/3 Control Unit.
2 - 19
2.3 Setting up the n-to-m Connection
2. Start the "Service Tool" from the "REGIUS Service" screen, and open the
"Network Setup" screen. (click [Reader] -> [Network])
"Reader" screen will be displayed.
The reader device will be exhibited in the window in order of "Display Name"
and "Reader Type".
"CS-1" has been registered as the Host on the factory default on REGIUS 110.
2 - 20
2.3 Setting up the n-to-m Connection
4. Set the network condition, host name for the REGIUS 110, CS-1/2/3 and JM
referring to the following table.
Host Name Input the Host Name of the REGIUS 110 itself.
r110-*****:***** indicates a serial number.
DBHOST
Host Name Input in each column the Host Names of each CS-1/2/3
that is connected to the REGIUS 110.
5. Click [Send].
Settings displayed in the screen will be sent to the REGIUS 110 and the net-
work information is updated.
• The updated settings will be written in a CF (Compact Flash Memory) of
the REGIUS 110.
6. Click [Exit] to close the "Network Setup" screen. Return to the "Ser-
vice Tool" screen (Reader).
2 - 21
2.3 Setting up the n-to-m Connection
7. Press and hold the [operation] switch for 3 seconds to turn OFF the REGIUS
110.
2 - 22
2.3 Setting up the n-to-m Connection
2.3.2 Setting Up the CS-1/2/ This section describes how to set up JM on the CS-1/2/3 of Std. JM,
3 of Std. JM the settings required to connect the CS-1/2/3 to the REGIUS 110 and
the network settings for the CS-1/2/3 itself. CS-1/2/3 of Backup JM will
be described in " Implement the setting of CS-1/2/3 of Backup JM, the
setting for connection of CS-1/2/3 with REGIUS 110 and the network
setting of CS-1/2/3." .
2 - 23
2.3 Setting up the n-to-m Connection
Communication Setting
2 - 24
2.3 Setting up the n-to-m Connection
■ Setting up the JM database If the host name of the CS-1/2/3 or REGIUS 110 has been changed
from the factory default, you must use the following procedure to
update the information registered in the database.
2 - 25
2.3 Setting up the n-to-m Connection
c_status c_stat
Host Name of CS-1/2/3 #1 2
r_name r_stat
Host Name of REGIUS 110 #1 2
2 - 26
2.3 Setting up the n-to-m Connection
• < "r_version" screen >
<Important>Use lowercase to input the Host Name of REGIUS 110
when using PostgreSQL tool.
r_name
Host Name of REGIUS 110 #1
Setting the Relations Set the relations between the CS-1/2/3(s) and REGIUS 110(s). In the
"n-to-m connection", setting the relations between the CS-1/2/3 and
REGIUS 110 will enable a linked-operation as to the power control,
error information between the CS-1/2/3 and corresponding REGIUS
110.
This section describes settings used to associate all of CS-1/2/3 and
REGIUS 110s.
1. Rewrite as the following, and set the relation between the CS-1/2/3 and
REGIUS 110.
• < "relations" screen >
<Important>Use uppercase to input the Host Name of CS-1/2/3
and lowercase to input the Host Name of REGIUS 110.
2 - 27
2.3 Setting up the n-to-m Connection
Setting the cassette registration This section describes how to set the cassette registration method.
method By default, the method is set to "1" (Manual Registration or Auto-
registration).
"reg_type" set here and the cassette registration method set in " ■ Setting the
JM information" have the relation as the following. Always make settings so
that the following relation is established.
Auto-Regist. 1
Barcode Regist. 0
■ Backing up JM information Back up the settings to a floppy disk after completing the settings of the
JM database.
1. Insert a floppy disk for backup in the floppy disk drive of CS-1/2/3 Control
Unit.
2 - 28
2.3 Setting up the n-to-m Connection
3. Remove the backup floppy disk from the floppy disk drive.
2 - 29
2.3 Setting up the n-to-m Connection
■ Setting the CS-1/2/3 to connect This section describes how to set up the REGIUS 110s connected to
with the REGIUS 110 the CS-1/2/3 and the device name for each REGIUS 110.
2 - 30
2.3 Setting up the n-to-m Connection
Host Name Input the Host Name set for the REGIUS 110.
Host Name Input the Host Name set for REGIUS 110 using lower-
(on Job Manager) case.
Switch between the device names exhibited at the upper left of the screen, and
set each of connecting REGIUS 110 as above.
2 - 31
2.3 Setting up the n-to-m Connection
1. Terminate the Service Tool and click [Windows Desktop] on the REGIUS
Service screen to display the Windows desktop.
1. Click [Shutdown Option (U)] -> [Shut down (U)] of the [Start] menu to turn
OFF the CS-1/2/3.
2.3.3 Setting the CS-1/2/3 of Implement the setting of CS-1/2/3 of Backup JM, the setting for connection of
Backup JM CS-1/2/3 with REGIUS 110 and the network setting of CS-1/2/3.
2 - 32
2.3 Setting up the n-to-m Connection
2. Set as follows.
Setting Item Setting Details
Standard Setting
Communication Setting
Device Registered ON
2 - 33
2.3 Setting up the n-to-m Connection
■ Setting the database of the Register with the backup JM the same database that has been set in
backup JM " ■ Setting up the JM database".
2 - 34
2.3 Setting up the n-to-m Connection
2 - 35
2.3 Setting up the n-to-m Connection
■ Setting to connect CS-1/2/3 with Implement the setting of the number of REGIUS 110 that shall be connected
REGIUS110 with CS-1/2/3 and the name of each REGIUS 110.
In this procedure, set the settings the same configuration that has been set in
the built-in standard JM.
1. Using the same procedure practiced for " ■ Setting the CS-1/2/3 to connect
with the REGIUS 110", implement the setting of device name of REGIUS 110
and CCU INFO on "SYSTEM INFO" screen.
1. Exit "Service Tool" screen. Then set the network for CS-1/2/3 in the same
manner as practiced for " ■ Setting the network of CS-1/2/3"
1. Click [Shutdown Option (U)] -> [Shut down (U)] of the [Start] menu to turn
OFF the CS-1/2/3.
2 - 36
Chap.3 Verification & Backup of
Setup Data
Scan the solid density image, and check that the cap-
tured image is normal.
Then back up the setting data and complete the installa-
tion procedure.
Blank Page
3.1 Checking the Image
3.1 Checking the Image Using REGIUS 110, scan the cassette on which the solid density image was
exposed, and check the image using "Service Tool".
2. Press [operation] switch and start up REGIUS 110 that should be checked.
3. Input the tentative Patient Information on CS-1/2/3 Control Unit, and switch
to "Image Verification" screen.
4. Touch the rectangular(others) at the lower left of the anatomy, then touch
[TEST] to select.
2m
6. If the system is set to "Barcode Registration", read the barcode of the cassette
that was exposed.
"Barcode Registration" : the bar code of the cassette is displayed in the order.
"Manual Registration" : "1st" is indicated.
"Auto-registration" : Nothing on the screen.
8. [Touch [OK].
3-1
3.1 Checking the Image
3-2
3.1 Checking the Image
• Select the latest file(with the largest number) if several image file is
listed in the "Open File" dialogue.
3-3
3.1 Checking the Image
19. Adjust the slider bar so that the difference between "Black" and "White" on
the "Contrast" screen becomes approximately 400 steps.
20. Drag the slider of the "Center" to right and left so that the unevenness of the
image becomes most evident.
21. Using a yellow cross cursor, select the entire image tracing its border.
22. Horizontal profile of the area that is enclosed by the cross cursor will be
displayed on the "Horizontal Profile" screen.
3-4
3.1 Checking the Image
25. In a same manner, check the image read by the rest of all REGIUS 110s.
If a steep slope or winding is found in the horizontal profile, it may
mean there is an abnormality on the REGIUS 110. Take necessary
action by referring to the REGIUS 110 Service Manual.
3-5
3.2 Backing Up the Setup Data
3.2 Backing Up the After completing the image check of the REGIUS 110, back up the setup
data following the procedures below, and complete the installations.
Setup Data
3.2.1 Backing Up the Upload and save to the CS-1/2/3 the settings and correction data
saved in the connected REGIUS 110.
Reader Setup Data
3. Click [Reader].
"Reader" screen of the Service Tool will be shown.
4. Click [PCB].
"Reader Device" screen will be shown.
5. Select the REGIUS 110 whose setup data shall be backed up.
<Important>[Restore] -> [CS-1 -> REGIUS 170] button is disabled
if the setup data is not backed up on the CS-1/2/3.
3-6
3.2 Backing Up the Setup Data
7. Touch [Exit].
The PCB screen closes, and switches to the "Reader" screen.
Repeat from Step 3 in 「3.2.1 Backing Up the Reader Setup Data」, and
back up the setup data on all REGIUS 110s on the CS-1/2/3.
3.2.2 Back Up of CS-1/2/3 <Important>Using the floppy disk on which a specific CS-1/2/3
Setup Data has been backed up, you can restore the same set ups
on the other CS-1/2/3. However, note that an error will
occur if the restored CS-1/2/3 is interfaced to the same
network as the original because the setup data unique to
the original CS-1/2/3 are also restored on the other CS-1/
2/3. If such happens, disconnect the restored CS-1/2/3
from the network, and set up the CS-1/2/3 uniquely from
others.
2. Click [System].
"Save System Setup File" screen will be shown.
3-7
3.2 Backing Up the Setup Data
3. Click [>>].
4. Click [Save].
A message prompting an insertion of the disk into drive "A" will be
shown.
7. Click [Yes].
Saving of the CS-1/2/3 setup data will start.
Progress will be indicated on "Saving ...." message while saving
continues.
Upon completion of saving, a message to inform completion will be
shown.
8. Click [Yes].
Returns to "Save System Set Up File" screen.
9. Click [Exit].
Confirmation dialogue will be shown.
3-8
3.2 Backing Up the Setup Data
3-9
3.2 Backing Up the Setup Data
3 - 10
Chap.4
Adjusting the REGIUS 110
4.1 Checking and Check the pixels in the main scanning direction of the REGIUS 110, and
adjust PLL setting if the value is out of the standard value.
Adjusting the Image
Size
4.1.1 Checking Image Size, Reading the cassette on the REGIUS 110, which is exposed to produce
solid density image, and measure the pixels in the main scanning
Read Start Position
direction.
2. Touch the square (Other) in the lower left corner of the body diagram and
touch to select [TEST].
2m
4. If the system is set to "Barcode Registration", read the barcode of the cassette
that was exposed.
"Barcode Registration": the bar code of the cassette is displayed in the order.
"Manual Registration" : "1st" is indicated.
"Auto-registration" : Nothing on the screen.
4-1
4.1 Checking and Adjusting the Image Size
6. Touch [OK].
7. Touch [Exam completed].
8. Touch [KONICAMINOLTA] and start Service Tool from the system menu or
open the REGIUS Service screen and start the Service Tool.
The Service Tool (Console) screen will be shown.
9. Touch [Analyze].
"Analyze" screen will be shown.
4-2
4.1 Checking and Adjusting the Image Size
14. Check that the measured pixels fall within the allowance (Hor. and Ver. direc-
tion(sub-scan): 2022pixels ± 10pixels/Ver. direction: 2460 ± 12pixels, 2022
± 12pixels for 14x17in ).
15. Select [View Size (V)] in the [Display (V)] menu, and select [No Pixel Skip
(N)] in the submenu.
16. Using mouse, check that the margin of H_sync (the left side of image) on the
plate falls within the allowance (33 ± 10 pixels) and the margin on the top
falls within the allowance (33 ± 10 pixels).
4-3
4.1 Checking and Adjusting the Image Size
3. Select the REGIUS 110 on which the image was read, and click [OK].
"Device Setup" screen will be shown.
To Change Pixels
4-4
4.1 Checking and Adjusting the Image Size
1. Input the amount of the distance which the image would like to be shifted to
“Hor.Shift”. (input a positive/negative value)
Inputting a positive value moves the image to the right, while a negative value
to the left.
2. Input the amount of the distance which the image would like to be shifted to
“Ver.Shift”. (input a positive/negative value)
If inputting a positive value, the image goes upward and if inputting a nega-
tive value, the image goes downward.
3. Click [Send].
Revised value will be sent to the REGIUS 110.
4. Click [Exit].
Returns to "Service Tool (Reader)" screen.
4-5
4.2 Unevenness Calibration of Regular Cassette
4.2 Unevenness Cali- If the optical unit of REGIUS 110 is replaced or used the REGIUS Plate
with new version, implement the unevenness calibration by the following
bration of Regular procedure.
Cassette
<Important>Use the largest cassette that is used in the site for the
unevenness calibration. The correction value will be
invalid for the sizes larger than the one used for the cali-
bration.
1. Expose the solid density image on the cassette for the unevenness calibration
to the following conditions.
• Exposure Condition
Volt : 80kV (fixed)
mAs value : Set the X-ray dose so that the maximum data value falls
within 1500 ~ 3000 Step. (target : 10 to 20mAs at 2m)
• Align the long side of the cassette with the vertical direction of the
X-ray tube.
2m
4-6
4.2 Unevenness Calibration of Regular Cassette
4. Select the cassette reader on which the unevenness should be calibrated, and
click [OK].
• "Unevenness Calibration" screen will be shown.
6. If the system is set to "Bar Code Registration", read the bar code on the
cassette using a bar code reader. The bar code data read by the bar code reader
will be displayed on the "Bar Code Registration".
• If the system is set to “Manual Registration”, it is not necessary to
read the bar code.
7. Click [OK].
A dialog indicating the progress of the Unevenness Calibration will be
shown.
4-7
4.2 Unevenness Calibration of Regular Cassette
9. Click [Exit].
Returns to the Service Tool (Reader) screen.
The result of the unevenness calibration will be saved in the
unevenness calibration table of the CS-1/2/3.
4-8
4.3 Sensitivity Calibration of Regular Cassette
4.3 Sensitivity Calibra- Implement the sensitivity calibration if the X-ray generator device was
changed or calibrating the sensitivity difference on each REGIUS 110.
tion of Regular Cas-
sette <Important>Use the largest cassette that is used in the site for the
sensitivity calibration.
1. Using the condition shown below, expose a solid density image on the cassette
for the unevenness correction.
• Exposure Condition
Voltage: 80kV (fixed)
Grid : None
• Align the long side of the cassette with the vertical direction of the
non-grid X-ray tube.
4-9
4.3 Sensitivity Calibration of Regular Cassette
3. Click [Sensitivity].
"Reader Device" screen will be shown.
4. Select the cassette reader on which the unevenness should be calibrated, and
click [OK].
• "Sensitivity Calibration" screen will be shown.
6. Input the X-ray dose that was used for the exposure in "Exposure Dose".
7. If the system is set to "Bar Code Registration", read the bar code on the
cassette using a bar code reader. The bar code data read by the bar code reader
will be displayed on the "Bar Code Registration".
• If the system is set to “Manual Registration”, it is not necessary to
read the bar code
4 - 10
4.3 Sensitivity Calibration of Regular Cassette
8. Click [OK].
A dialog indicating the progress of the sensitivity calibration will be
shown.
4 - 11
4.3 Sensitivity Calibration of Regular Cassette
4 - 12
Chap.5 Maintenance &
Adjustment
This chapter describes procedures for maintenance and
adjustment of the CS-1/2/3.
Blank Page
5.1 Upgrading the Version of REGIUS 110
5.1 Upgrading the Ver- Update the firmware (program and data stored in CF) of the cassette
reader.
sion of REGIUS 110
To update the software version of the REGIUS 110, use the "Service
Tool" of the CS-1/2/3 that is connected to the REGIUS 110.
1. Restart the power (OFF/ON) of all devices connected to the same network.
Shut down the backup JM CS-1/2/3
(not required for a 1-to-1 configuration)
↓
Shut down the standard JM CS-1/2/3
↓
Turn ON the standard JM CS-1/2/3
↓
Turn ON the backup JM CS-1/2/3
(not required for a 1-to-1 configuration)
↓
Turn ON all REGIUS 110s.
5-1
5.1 Upgrading the Version of REGIUS 110
3. Click [Reader].
The “Service Tool (Reader)” screen will be shown.
6. Insert the floppy disk that contains a new software version for REGIUS 110
into the floppy disk drive of the CS-1/2/3.
7. Click [ • • • ].
"Open" dialog will be shown.。
5-2
5.1 Upgrading the Version of REGIUS 110
8. Select the new program version of REGIUS 110 Install File, and click
[Open (O)].
Returns to the “Install File” screen, and the selected file name will be
shown on the “File Name”.
11. Select REGIUS 110 that should be upgraded, and click [OK].
"Controller Program" screen will be shown.
5-3
5.1 Upgrading the Version of REGIUS 110
15. Start up the "PostgreSQL access" of the JM, and confirm on the "r_version"
screen that the version and updated data of the REGIUS 110 have been
updated.
5-4
Chap.6
Service Tool Screens
This chapter describes the "Service Tool" screens that
are used to interface with REGIUS 110.
Refer to Installation/Service manual of each console for
the rest of screens.
Blank Page
6.1 Overview of Service Tool
[Console] > [System Config.] > [System Changed Registration method of cassette reader was changed from p.6-2
Config.(Reader)] selection of availability of the device by checkbox to pull-
down menu("170/190" or "110" or "No Registration").
[Console] > [System Config.] > [Remote Changed REGIUS 110 was added to the reader selection. p.6-3
Maint.]
[Reader] > [Reader] > Changed Reader type denotation was added to the reader selection p.6-4
[Cassette Reader] field.
[Reader] > [Device Set] New Tools for REGIUS 110 was separately created. p.6-5
[Reader] > [PCB] Changed Resolution selection was changed due to nonsupport of p.6-8
mammogram.
[Reader] > [Std.Curr.] Changed Resolution selection was changed due to nonsupport of p.6-9
mammogram.
[Reader] > [Unevenness] Changed Resolution selection was changed due to nonsupport of p.6-10
mammogram.
[Reader] > [Sensitivity] Changed Resolution selection was changed due to nonsupport of p.6-12
mammogram.
[Reader] > [Sublog] Unchanged Tools for REGIUS 170/190 only. On REGIUS 110, have all -
tools in [Reader] -> [Reader Log].
6-1
6.2 [Console] > [System Info.] Screen
Key/Item Function
[Save & Exit] Click to exit from "System Info." screen after saving the settings.
[Cancel] Click to exit from "System info." screen without saving the settings.
Start up from the system menu Start up from the REGIUS Service Screen
Display Procedure • • • • [ [Console] --> [System Info.] --> (Select [System (Reader)] in the left menu.)
Key/Item Function
In the case of CS-1/3, items for upright/table readers will be displayed above the REGIUS 110.
REGIUS 110
Model Name Input a name which is for specifying the several REGIUS to identify in collective within 1
byte x 10 characters.
Reader 1 ~ 2 Check "Registered" for the reader device to be used in the system.
Disp. Name Input the name used in the system with 1 byte x 10 characters.
<Important>Default name of the reader devices displayed in the user screens is set to "REGIUS 170".
When connect to the REGIUS, the displayed name can be changed by inputting the data in the
“Disp.Name” as above screen.
6-2
6.2 [Console] > [System Info.] Screen
Start up from the system menu Start up from the REGIUS Service Screen
Display Procedure • • • • [Console] --> [System Info.] --> Select "Remote Maintnance" in the left menu.
Key/Item Function
Remote Maintenance
IA Auto-Start Tick to enable "Remote Maintenance" function.
Ccu Maint. Info. Acquisition
Setting
Ccu1 Tick to acquire the maintenance information of REGIUS 110/170/190 that has been registered as
"Reader #1".
Select the reader type from the pull-down menu when ticked.
Select the reader type from REGIUS 110/REGIUS 170/REGIUS 190.
Ccu2 Tick to acquire the maintenance information of REGIUS 110/170/190 that has been registered as
"Reader #2".
Select the reader type from the pull-down menu when ticked.
Select the reader type from REGIUS 110/REGIUS 170/REGIUS 190.
6-3
6.3 [Reader] > [Reader] > "Dedicated Reader"
Key/Item Function
Reader Device Selection (upper left Select the reader device name to be set if there are several reader devices con-
menu) nected.
[Save & Exit] Click this button to exit from "CCU Info" screen after saving the setting.
[Cancel] Click this button to exit from "CCU Info" screen without saving the setting.
Display Procedure • • • [Reader] --> [Reader] --> select "TCP/IP" in the left menu
Key/Item Function
TCP/IP
Host Name (Hosts) Input the host name of REGIUS 110 using 1 byte x less than 32 characters. (default:
R170-0001)
Host Name (JM) Input the host name of REGIUS 110, which the JM uses, using 1 byte x less than 32
characters.
(default: r170-0001)
IP Address Input the REGIUS 110’s IP address. (default: 192.168.20.170)
Port Number (Reader) Input the REGIUS 110’s port No. (default: 8018)
Port Number (Console) Input the CS-1/2/3’s port No. (default: 8018)
Timeout 1 Do not change unless otherwise instructed. (default: 180000)
6-4
6.4 [Reader] > [Device ]
Start up from the system menu Start up from the REGIUS Service Screen
Display Procedure • • • [Reader] --> [Device ] --> select the reader name from the "Reader Device" screen --> [OK]
Key/Item Function
PLL(image size)
H-Magnification Set the horizontal magnification of the image size.
Default: Adjustment is made on each device (each serial No.) before shipment.
Pixel number of fluorescence substance should be within -5 to +10 pixels(4041 to 4056
pixels) of 4046 pixels.
V-Magnification Set the vertical magnification of the image size.
Default: Adjustment is made on each device (each serial No.) before shipment.
Pixel number of fluorescence substance should be within 2046 ± 10 pixels(2036 to
2056) (2022 ± 10pixels for 11/14in).
Reader Start Point
H-Shift Set the margin on the left side of the image (H-Sync side of the plate).
Input + value to increase the margin and - value to reduce the margin that the margin is
to be within 33 ± 10 pixels.
Default: 0
V-Shift Set the margin on the bottom side of the image.
Input + value to increase the margin and - value to reduce the margin that the margin is
to be within 33 ± 10 pixels.
Default: 0
[Send] Sends the set up value to REGIUS 110.
[Table Refer...] Reads and displays the data stored in SIU (CF) of REGIUS 110.
6-5
6.4 [Reader] > [Device ]
Start up from the system menu Start up from the REGIUS Service Screen
Display Procedure • • • [Reader] --> [Device ] --> select the name of REGIUS 100 from the "Reader Device" screen --> [OK]
Key/Item Function
[Back] Closes the "Table Management" screen, and returns to "Device Setup" screen.
Table Displays the value of PLL, H-Starting Point (87.5μm), H-Starting Point (175μm) and V-Start-
ing Point.
[Output] Outputs the table values in text format (.txt) file.
Default file name: Device name_y/m/d_CcuPllRet.txt」
How to calculate the image size 1. Expose the solid density image on a 14"x17" or 14"x14" cassette by High
Res. mode, and measure the size of fluorescence substance by using an
inspection tool.
2. Change the image size to be within the allowable range of fluorescence sub-
stance(Hor. Magnification: 4046 pixels -5 to +10 pixels/Ver. Magnifica-
tion: 4920 pixels -5 to +10 pixels).
6-6
6.4 [Reader] > [Device ]
Image Position Adjustment 1. Expose the 14"x17" cassette using High Res. mode, and measure the
margin on H-sync side of the plate using an inspection tool.
2. Set the H-Shift and V-Shift values to be within the
allowable range of the margin(30 to 45 pixels).
6-7
6.5 [Reader] > [PCB]
Display Procedure • • • [Reader] --> [PCB] --> (select the reader from the "Reader Device" screen)
Key/Item Function
Backup The stored data in CIU(CF) of the REGIUS 110 will be uploaded to CS-1/2/3
after clicking “Console -> REGIUS 170/190” and saved.
Restore The stored data in CS-1/2/3 will be uploaded to CIU(CF) of the REGIUS 110
after clicking “REGIUS 170/190 -> Console” and saved.
• "Program replacement for REGIUS170/190" is not available on REGIUS 110.
<Important>"Program replacement for REGIUS170/190" button is
invalid if the backup data is not stored in CS-1/2/3.
[Exit] Exit the "Change Board" screen.
• Data to be stored
• Shading data
• Sensitivity calibration data(settings made on "Sensitivity Calibration" screen)
• PLL adjusting value(settings made on "Device Setup" screen)
• Image position(settings made on "Device" screen)
• PMT Gain Table (settings made on "Unevenness Calibration" screen)
• Various setting data for network
• Mechanical counter info.
6-8
6.6 [Reader] > [Std. Current]
Display Procedure • • • [Reader] --> [Std. Current] --> (select the reader from the "Reader Device" screen --> [OK])
Key/Item Function
Reading Condition
Image Size Select the image size from the following when reading Standard Current.
• "175µm(General Std)", "87.5µm(Gen. H/Man. Std)".
• REGIUS 110 does not support Mammogram exposure.
Standard Current Select the current value for dummy reading from the following.
• "0nA", "50nA", "500nA", "50000nA", "All".
Unevenness Select the unevenness calibration for dummy reading from the following.
• "ON", "OFF".
[Delete All Test] 1) Clicking on this button will display a dialogue confirming the deletion of the dummy
image.
2) Clicking "Yes" will delete only the dummy images at one time.
[Execution] Execute the dummy reading with the standard current.
Implementing [Execution] will create “(Device Name)_CcuTest.img” file is created in the
following folder.
For CS-1 : E:\Image
For CS-2 : C:\KonicaMinolta\CS-2\Image
For CS-3 : C:\KonicaMinolta\CS-3\Image
After checking the resulted image, delete the image by implementing [Delete All Test
Images].
[Exit] Switches to "Reader" screen without implementing the dummy reading.
6-9
6.7 [Reader] > [Uneven. Cal.]
Display Procedure • • • [Reader] --> [Uneven. Cal.] --> (select the reader from the "Reader Device" screen --> [OK])
Key/Item Function
Calibration
Resolution Select the pixel size for calibration from "175µm(General Std)", "87.5µm(Gen. H)".
• REGIUS 110 does not support Mammogram examination.
PMT Use the default (2200)
Barcode Registration Use this function when the bar code registration is employed.
[OK] Starts the unevenness calibration.
Create Shipping Table Not to be used at the user. (do not tick the check).
(Factory use only)
Result Displays the result of the unevenness calibration.
[Table Control] Displays the information of the unevenness calibration.
• Shows the list of Plate Versions and Cassette Sizes that have been used for calibration in the
past.
[Text File] Exports the result of the unevenness calibration as a text file(.txt).
[Exit] Exits the unevenness calibration.
• Correction data
10 tables of correction data for each of "87.5µm", "175µm" per Pv (plate version) can be registered. (total 10 tables).
• Plate size
For calibration purpose, use the largest size that is used at the site. (except mammography use). For the sizes larger than the
one used for calibration, the calibrated data is no more valid.
6 - 10
6.7 [Reader] > [Uneven. Cal.]
As an example, the case where the calibration is made using Pv00 + 14" x 17", Pv01 + 10" x 12" is described in the follow-
ing table.
6 - 11
6.8 [Reader] > [Sensitivity]
Display Procedure • • • [Reader] --> [Sensitivity] --> (select the reader from the "Reader Device" screen --> [OK])
Key/Item Function
Calibration
Automatic calculation from Selecting this option enables to automatically create the calibration data for low
high resolution data resolution using calibration data for high resolution.
Resolution When "Automatic calculation from high resolution data" is not selected, select the pixel
size for sensitivity calibration.
• For regular cassette, select from the "175µm (standard)", "87.5µm (high resolution mode
A)" .
When "Automatic calculation from high resolution data" is selected, select the pixel size
for sensitivity calibration.
• For regular cassette, select "87.5µm (high resolution mode A)"
Exposure Dose Input the irradiation dose.
• 8 ~ 20mR (10mR or higher is recommended)
Sensitive Input "0.94" for regular cassette.
• In "Auto Create" mode, sensitivity data for low resolution will be calculated using these set-
tings.
Barcode Registration Use only for "Bar Code Registration" is selected.
[OK] Executes the sensitivity calibration.
Create Shipping Table Not to be used at the user. (do not tick the check).
(Manufacturer Use Only)
Low sensitivity QR value of the low sensitivity region that shall be calibrated. (default : 125)
Standard sensitivity QR value of the normal sensitivity region that shall be calibrated. (default : 250)
High sensitivity QR value of the high sensitivity region that shall be calibrated. (default : 500)
6 - 12
6.8 [Reader] > [Sensitivity]
Key/Item Function
Result Displays the result of sensitivity calibration.
[Table Control] Displays the correction data for the reader device.
• Lists all history of calibrations with Plate version (Pv), size, cassette in the past.
[Text File] Exports the result of calibration result as a file.
[Exit] Exits the sensitivity calibration.
• Sensitivity calibration
• Regular (87.5µm, 175µm) cassette will be automatically recognized by reading the barcode on the cassette.
• Selecting "Automatic calculation from high resolution data" will implement the calibration by reading at high resolution,
and automatically create the sensitivity data for low resolution. (calibration completes with one exposure)
• Use 14" x 17" film for regular (87.5µm, 175µm) . (use the largest size that is actually used at the institution)
• Standard Exposure Dose : 10mR or higher is recommended (use the exposure dose at which the x-ray irradiation
becomes stable)
6 - 13
6.9 [Reader] > [Reader Log]
<Important>Implementing this function will download the CIU Logs of the REGIUS 110.
Display Procedure • • • [Reader] --> [Reader Log] --> (select the reader from the "Reader Device" screen)
Key/Item Function
Reader Select Select the REGIUS for acquiring a log when there are several reader devices which are
connected.
For REGIUS170/190, implement [Reader] -> [Create Sub-log] before implementing
[Reader Log].
[Create Sub-log] is not necessary for REGIUS 110.
[Reader Log] Receives the logs from the REGIUS 110.
[Exit] Exits the “Reader Log” screen.
6 - 14
6.10 [Reader] > [Mech. Counter]
Key/Item Function
Number of Process Shows the number of process of each cassette size fed by the Reader.
TTL :Total of "Read", "Erase (Quick)" and "Erase (Slow)".
Read :Number of process for cassette reading.
Erase(Quick):Number of process for cassette erasure(Erase Mode-1).
Erase(Slow):Number of process for cassette erasure(Erase Mode-1).
Halogen Lamp <Case of REGIUS 110>
Warning Settings :
Displays the maximum number lighting of Halogen lamp in terms of count of number of
lighting. The warning that is for replacing the halogen lamp will be displayed if the number
reaches this limits.
Number of Lighting :
Displays how many times the lamp is actually turned on in the past.
Number of Lighting :
Displays how many times the lamp is actually turned on in the past.
[File Output] Exports the Mech. Count Information as a text file(.txt).
[Back] Exit the "Mech. Counter" screen.
6 - 15
6.11 [Reader] > [File Import], [Upgrade]
Key/item Function
[...] (reference key) Select the holder where the install program resides. Normally select from "A" drive (floppy
disk drive).
[OK] The selected program will be decompressed.
[Cancel] Exit the "Install File" screen after cancelling the selection of the file.
Display Procedure • • • [Reader] --> [Upgrade] --> (select the reader from the "Reader Device" screen)
Key/Item Function
Version Indicates the program version which is currently running on the REGIUS 110.
The Current Installed Version list of the REGIUS 110’s programs stored in the CS-1/2/3. The program selected
here will be written in the reader device.
[OK] Execute the switch over to the program version listed in the reader program version list.
[Cancel] Exit the "Reader Program (change)" screen after cancelling the installation of program.
6 - 16
6.12 [Reader] > [Network]
Display Procedure • • • [Reader] --> [Network] --> (select the reader from the "Reader Device" screen --> [OK])
Name of currently
connected
REGIUS 110
Key/Item Function
Reader
Host Name Input the host name of the REGIUS 110. (default: r170-0001)
IP Address Input the IP address of the REGIUS 110. (default: 192.168.20.170)
Netmask Input the subnet mask of the REGIUS 110. (default: 255.255.255.0)
JM
Host Name Input the Host Name(jml-0001 fixed) of JM which controls this REGIUS 110.
Hosts
Host Name Displays the host name and IP address of CS-1/2/3s which are connected to this
REGIUS 110.
[New] Click this button when newly inputting the host name and IP address of the CS-1/2/3
which uses this REGIUS 110.
[Edit] Click this button to change the host name or IP address of the CS-1/2/3 or JM listed on
the "Hosts" window
[Delete] Deletes the CS-1/2/3 or JM from the "Hosts" window.
[Send] Send the displayed settings to the REGIUS 110 in FTP mode.
[Exit] Exit the "Network" screen. No change will be caused on the REGIUS 110.
6 - 17
6.12 [Reader] > [Network]
6 - 18