1 Overview
1.1 Major Differences in Specifications between REGIUS110 and REGIUS190 ................................ 1-1
1.2 Changes in CS-2/CS-3 Program to Support REGIUS110 ............................................................. 1-1
1.3 Specification Comparison .............................................................................................................. 1-2
1.4 Restrictions & Default Values ........................................................................................................ 1-5
1.4.1 CS-1/2/3 Program Version supporting REGIUS 110 ........................................................... 1-5
1.4.2 Default Value of Host Name & IP Address .......................................................................... 1-5
1.4.3 Setting the Host Name ........................................................................................................ 1-5
1.4.4 Setting the Exam Keys ........................................................................................................ 1-6
1.4.5 Setting the IP Address......................................................................................................... 1-6
1.5 REGIUS 110 Error Code ............................................................................................................... 1-7
1.6 Job Table of Job Manager ........................................................................................................... 1-10

Chap.2 Connecting CS-1/2/3 with REGIUS 110


2.1 Flow of Setup Procedure ............................................................................................................... 2-1
2.2 Setting Up the 1-to-1 Connection .................................................................................................. 2-4
2.2.1 Setting Up the REGIUS 110 ................................................................................................ 2-5
2.2.2 Setting Up the CS-1/2/3..................................................................................................... 2-10
2.3 Setting up the n-to-m Connection ................................................................................................ 2-17
2.3.1 Setting Up the REGIUS 110 .............................................................................................. 2-18
2.3.2 Setting Up the CS-1/2/3 of Std. JM.................................................................................... 2-23
2.3.3 Setting the CS-1/2/3 of Backup JM ................................................................................... 2-32

Chap.3 Verification & Backup of Setup Data


3.1 Checking the Image....................................................................................................................... 3-1
3.2 Backing Up the Setup Data ........................................................................................................... 3-6
3.2.1 Backing Up the Reader Setup Data .................................................................................... 3-6
3.2.2 Back Up of CS-1/2/3 Setup Data......................................................................................... 3-7

Chap.4 Adjusting the REGIUS 110


4.1 Checking and Adjusting the Image Size ........................................................................................ 4-1
4.1.1 Checking Image Size, Read Start Position.......................................................................... 4-1
4.1.2 Adjusting the Image Size and Read Start Position .............................................................. 4-3
4.2 Unevenness Calibration of Regular Cassette................................................................................ 4-6
4.3 Sensitivity Calibration of Regular Cassette.................................................................................... 4-9

Chap.5 Maintenance & Adjustment


5.1 Upgrading the Version of REGIUS 110 ......................................................................................... 5-1

i
Chap.6 Service Tool Screens
6.1 Overview of Service Tool............................................................................................................... 6-1
6.2 [Console] > [System Info.] Screen ................................................................................................. 6-2
6.2.1 "SYSTEM INFO" Screen ..................................................................................................... 6-2
6.2.2 "SYSTEM INFO • System Config.(Reader)" Screen ........................................................... 6-2
6.2.3 "SYSTEM INFO • Remote Maintenance" Screen............................................................... 6-3
6.3 [Reader] > [Reader] > "Dedicated Reader".................................................................................. 6-4
6.3.1 Common Displays of "CCU INFO" Screen .......................................................................... 6-4
6.3.2 "CCU INFO • TCP/IP" Screen ............................................................................................ 6-4
6.4 [Reader] > [Device ] ....................................................................................................................... 6-5
6.4.1 "Device" Screen (for REGIUS 110)...................................................................................... 6-5
6.4.2 "Table Management (for REGIUS 110)" Screen .................................................................. 6-6
6.5 [Reader] > [PCB]............................................................................................................................ 6-8
6.5.1 "REGIUS 170 PCB" Screen ................................................................................................ 6-8
6.6 [Reader] > [Std. Current]................................................................................................................ 6-9
6.6.1 "Standard Current" Screen .................................................................................................. 6-9
6.7 [Reader] > [Uneven. Cal.] ............................................................................................................ 6-10
6.7.1 "Unevenness Calibration" Screen ..................................................................................... 6-10
6.8 [Reader] > [Sensitivity]................................................................................................................. 6-12
6.8.1 "Sensitivity Calibration" Screen ......................................................................................... 6-12
6.9 [Reader] > [Reader Log] .............................................................................................................. 6-14
6.9.1 "Reader Log" Screen......................................................................................................... 6-14
6.10 [Reader] > [Mech. Counter]......................................................................................................... 6-15
6.10.1"Mech. Counter" Screen .................................................................................................... 6-15
6.11 [Reader] > [File Import], [Upgrade] .............................................................................................. 6-16
6.11.1 "Install File" Screen............................................................................................................ 6-16
6.12 [Reader] > [Network].................................................................................................................... 6-17
6.12.1"Network" screen ............................................................................................................... 6-17

ii
Chap.1
O v e r v i e w
How to connect and network CS-1/2/3 with REGIUS
110 is outlined in this chapter.
The procedure described in this manual is an example of
CS-2 and REGIUS 110. The same procedure can be
applied to the case of CS-1/2/3.
Blank Page
1.1 Major Differences in Specifications between REGIUS110 and REGIUS190

1.1 Major Differences in 1) Further downsizing (Compared to REGIUS 190: W130% x D60% x
Specifications H60%) is achieved, requiring an installation space smaller than that
required for REGIUS 190.
between REGIUS110
2) Suitable to the operation where the production capacity as that of REGIUS
and REGIUS190 190 is not needed.
Mono-Slot System, 45sec for Cassette Feed/Load time, 80 plates/hr for
reading throughput.
3) Mammogram examination is not available on REGIUS 110.
4) Origin of the REGIUS Plate where the reading starts is different between
REGIUS 110 and REGIUS190.
REGIUS 110:Upper right of the plate
REGIUS 190:Lower right of the plate
5) REGIUS 110 requires Service Tool partially different from REGIUS 190.

1.2 Changes in CS-2/ 1) The read start point of REGIUS 110 is changed to the upper right. Accord-
CS-3 Program to ingly the program has been changed to support the new read start point.
The image starts appearing from the top to down in the image field when
Support REGIUS110 the image data is received from REGIUS 110.

2) A new Service Tool special to REGIUS 110 is added. Some changes are
also made to the existing Service Tool due to the addition of support for
REGIUS 110.
"Sub-Log" tool is not available for REGIUS 110.
「6.1 Overview of Service Tool」
3) Icons indicated in each screen have been changed from the illustration of
REGIUS 170/190 to cassette.

4) "System Info" - Registration procedure for REGIUS 110/170/190 on the


"System Configuration" screen has been changed. Selection should be
made from "Pull-down" menu instead of ticking the check box in the past.
「6.2 [Console] > [System Info.] Screen」
5) Error Code and Error Message for REGIUS 110 have been added.

6) REGIUS 110 is newly added to “Remote Maintenance” service.

<Important>Refer to "CS-1/2/3 Program Version supporting


REGIUS 110" when the program is updated as above.

1-1
1.3 Specification Comparison

1.3 Specification Com- Specifications of REGIUS 110 and REGIUS 190 are shown below.
parison
■ Specification of Main Unit

REGIUS 110 REGIUS 190


Dimensions 740 (W)× 365 (D)× 747 (H)mm 580(W) × 580(D) × 1230(H) mm

Weight approx.100kg approx.170kg

Adjustable Level Difference in floor level at the Foot Adjuster : ± Turn the adjuster foot to lower the level. Fix the foot
above Floor 12mm or less so that the rubber part of the adjuster foot is slightly
depressed, caster on the front is in contact with floor
and the device is not easily moved to the right and
left.

Noise In operation:55dB or less 60dB or less


In standby:42dB or less
(except for warning buzzer, user's operational noise,
etc.)

Power Consumption 185Wh or less (when 14"x17"plate of exposure dose maximum 1200W (1032kcal/h) Rated
of 40mR or less is in process) Consumption:1100W(946kcal/h)

Maximum Power 700VA or less (100V, 7A or less) approx.1.1kW


Consumption

Power Supply Japan:AC100V ± 10% (50/60Hz) Japan: AC100V ± 10%(50/60Hz)


North America: AC110 / 115 / 120 /220 / 230 / 240V North America: AC115 / 120 /220 / 230 / 240V ±
± 10%(60Hz) 10%(60Hz)
Europe: AC200 / 220 / 230 / 240V ± 10% (50Hz) Europe: AC200 / 220 / 230 / 240V ± 10%(50Hz)

Operating In operation: 15 ~ 30°C / 35 ~ 80%RH Operating Environment:15 ~ 30 ℃/ 35 ~ 80%


Environment (no condensation) RH
(Temp./Humidity) Not in operation: - 10 ~ 40°C / 10 ~ 95%RH (no (no condensation)
condensation) Transport/Storage Condition:- 10 ~ 40 ℃/ 10
~ 95%RH (no condensation)

Consumable Parts Erasure Lamp Unit (Erasure Frequency: approx. Erasure Lamp Unit (Reading/Erasure Frequency:
30,000shots) approx. 10,000shots)

Accessaries 100V Power Cable (3m), Operation Manual Power Cable (3m), Operation Manual

Options REGIUS Cassette, REGIUS Plate, JM, REGIUS REGIUS Cassette, REGIUS Plate, REGIUS Console,
Console (CS-1/CS-2/CS-3/REGIUS Unitea) Monitor Arm, Operation Table/PC Rack, Monitor
Fixing Bracket

■ Usable Cassettes

REGIUS 110 REGIUS 190


Cassette types Cassette having the similar open/close mechanism as Cassette having the similar open/close mechanism as
capable to be read RC-110, RC-110R, RC-110M. RC-110, RC-110R, RC-110M.

Cassette sizes 14" × 17", 14" × 14", 11" × 14", 14" × 17", 14" × 14", 11" × 14",
capable to be read 10" × 12", 8" × 10", 24 × 30cm, 10" × 12", 8" × 10", 24 × 30cm,
18 × 24cm, 15 × 30cm 18 × 24cm, 15 × 30cm

Cassette types RC-110T(for stitched image),RC-110L(for deciding RC-110T(for stitched image),RC-110L(for deciding
exclusive to exposure the treatment position ) the treatment position )

Cassette sizes RC-110T :4 sizes (10" × 36", 11" × 28", 14" × RC-110T :4 sizes (10" × 36", 11" × 28", 14" ×
exclusive to exposure 42", 14" × 51") 42", 14" × 51")
RC-110L :3 sizes (14" × 17", 14" × 14", 10" × RC-110L :3 sizes (14" × 17", 14" × 14", 10" ×
12" 12"

1-2
1.3 Specification Comparison

■ Image Reading/Erasing Performance

REGIUS 110 REGIUS 190


Reading Resolution Normal Reading Mode:175μm Normal Reading Mode:175μm
High-Res. Reading Mode:87.5μm High-Res. Reading Mode:87.5μm
Mammo High-Res. Reading:43.75μm

Cassette Feed/Load 45sec or less (cassette size:14" × 14"/normal approx.40sec or less (image size 14" × 14", at
Time reading mode) reading mode of 175μm)
(Time required from
cassette insertion to
cassette ejection)

Processing Capacity 80 plates or more (cassette size:14" × 14"/normal approx.90 plates or more (image size 14" × 14", at
(Cassettes processed reading mode) reading mode of 175μm)
per hour when they
are inserted
continuously)

QR Value 3 values selectable between 125 ~ 500. 3 values selectable between 125 ~ 500.
・Low Sensitivity (QR Value: 125 ~ 249) ・Low Sensitivity (QR Value: 125 ~ 249)
・Normal Sensitivity (QR Value: 250) ・Normal Sensitivity (QR Value: 250)
・High Sensitivity (QR Value: 251 ~ 500) ・High Sensitivity (QR Value: 251 ~ 500)

Definition of QR:QR=200 is defined when 1535- Definition of QR:QR=200 is defined when 1535-
step is resulted from 12bit output when exposure is step is resulted from 12bit output when exposure is
made under the condition of exposure dose at 2.58 × made under the condition of exposure dose at 2.58 ×
10-7 C/kg (= 1mR)and the tube voltage at 80kV. 10-7 C/kg (= 1mR).

Range (reading) Normal D Range (4.0 digits: PMT output current Normal D Range (4.0 digits: PMT output current
range: 100nA to 1mA) range: 100nA to 1mA)

Digital Gradation 12bit (4096 steps) 12bit (4096 steps)

1-3
1.3 Specification Comparison

■ Cropped Image Size

REGIUS 110 REGIUS 190


Cassette Type Displayed Normal H-Res. Mammo Displayed Normal High- Mammo
Size (pixel) Mode H- Size (pix.) Res.Mode H-
(mm) (pix.) Res.Mode (mm) (pix.) Res.Mode
(pix.) (pix.)

14" × 17" 352.0 × 2010 × 4020 × ― 352.0 × 2010 × 4020 × ―


428.5 2446 4892 428.5 2446 4892

14" × 14" 352.0 × 2010 × 4020 × ― 352.0 × 2010 × 4020 × ―


352.0 2010 4020 352.0 2010 4020

11" × 14" 275.0 × 1572 × 3144 × ― 275.0 × 1572 × 3144 × ―


352.0 2010 4020 352.0 2010 4020

10" × 12" 250.5 × 1430 × 2860 × ― 250.5 × 1430 × 2860 × ―


301.5 1722 3444 301.5 1722 3444

8" × 10" 199.5 × 1140 × 2280 × ― 199.5 × 1140 × 2280 × ―


250.5 1430 2860 250.5 1430 2860

24 × 30cm 236.5 × 1350 × 2700 × ― 236.5 × 1350 × 2700 × ―


296.5 1692 3384 296.5 1692 3384

18 × 24cm 176.4 × 1008 × 2016 × ― 176.4 × 1008 × 2016 × ―


236.3 1350 2700 236.3 1350 2700

15 × 30cm 296.5 × 1694 × 3388 × ― 296.5 × 1694 × 3388 × ―


146.5 836 1672 146.5 836 1672

24 × 30cm ― ― ― ― 296.5 × ― 3388 × 6776 ×


(mammo) 238.0 2720 5440

18 × 24cm ― ― ― ― 237.0 × ― 2709 × 5416 ×


(mammo) 177.0 2023 4040

14" × 17" (PCM) ― ― ― ― 352.0 × ― 3540 × 7080 ×


414.0 4740 9480

1-4
1.4 Restrictions & Default Values

1.4 Restrictions &


Default Values

1.4.1 CS-1/2/3 Program Ver- REGIUS 110 can be connected to the console having the program version
sion supporting listed below.

REGIUS 110
Console Version
CS-1 V2.32R00 or later

CS-2 V1.21R00 or later

CS-3 V2.32R00 or later

1.4.2 Default Value of Host In TCP/IP networks, all devices connected to the network need an
individual host name (PC name) and IP address. The network IP
Name & IP Address
address and host name must be unique.
In the factory default, IP address and host name of REGIUS 110 are
fixed. The CS-1/2/3 and REGIUS 110 are set to the manual registration
method by default.
Remains the factory default in 1-on-1 connection unless otherwise
instructed by your site.
The following is the factory default status:

REGIUS 110
•Host name r170-0001
•IP address 192.168.20.170
•Net mask 255.255.255.0

<Important>"r170-0001" has been registered with the host name


of the REGIUS 110 at the factory, allowing the REGIUS
110 to be connected to the CS-1/2/3 as default. The same
is applied to REGIUS 170/190.

<Important>When networking REGIUS 110, make sure that a cor-


rect selection is made among REGIUS 110 and REGIUS
170/190 through [System Info] -> [System Config
(reader)] before implementing the settings using Service
Tool. Setting made with incorrect reader selection may
lead to abnormality of reader function.

1.4.3 Setting the Host Name To change the host name (PC name) of the REGIUS 110, set it using
the following rules.
<Important>The host name is case sensitive. Pay attention when
setting the host name.Input lower 5 digits of REGIUS
110's serial No. after "r110-".
Example : When REGIUS 110's serial No. is "00456", the
Host Name of that cassette reader should be;
r110-00456

1-5
1.4 Restrictions & Default Values

<Important>Use 5 digits when inputting the serial No. of REGIUS


110.

1.4.4 Setting the Exam • Exam keys can be shared among REGIUS 170/190/110.
Keys • REGIUS 110 does not support mammogram.
• When REGIUS 110 alone is connected to CS-1/2/3, delete Exam
keys specific to mammogram from the User Master.

1.4.5 Setting the IP Address To connect devices to the site’s existing network, obtain an IP address
for each device from the network administrator beforehand.
If no IP address is specified, use the following addresses.

Device IP address Subnet Mask


REGIUS 110 #1 192.168.20.170 255.255.255.0
REGIUS 110 #2 192.168.20.171 255.255.255.0

1-6
1.5 REGIUS 110 Error Code

1.5 REGIUS 110 Following is the List of Error Codes that may be displayed when REGIUS 110
Error Code is networked.

Error Message
Code
6000 Justifier HP sensor NG (Move to HP)

6001 Justifier Motor Control NG (Move to HP)

6002 Justifier HP sensor NG (Cassette Fixed position)

6003 Justifier Motor Control NG (Cassette Fixed position)

6004 Justifier Motor Control NG (Cassette Support Position)

6005 Justifier Motor Control NG (Cassette Fixed position -2)

6006 Justifier HP sensor NG (Move to HP -2)

6007 Justifier Motor Control NG (Move to HP -2)

6008 Justifier HP sensor NG (Move to HP -3)

6009 Justifier Motor Control NG (Move to HP -3)

6010 Justifier HP sensor NG (Cassette Support Position)

6011 Justifier Motor Control NG (Cassette Support Position -2)

6050 Reciever HP sensor NG (Move to HP)

6051 Elevator Motor Control NG (Move to HP)

6052 Reciever HP sensor NG (Position 1)

6053 Elevator Motor Control NG (Position 1)

6054 Elevator Motor Control NG (Position 2)

6055 Elevator Motor Control NG (Slight descent at Position 2)

6056 Elevator Motor Control NG (Position 1 -2)

6057 Reciever HP sensor NG (Move to HP -2)

6058 Elevator Motor Control NG (Move to HP -2)

6059 Reciever HP sensor NG (Move to HP -3)

6060 Elevator Motor Control NG (Move to HP -3)

6061 Reciever HP sensor NG (Move to Position 2)

6062 Elevator Motor Control NG (Position 2 -2)

6100 Transporter Sensor NG (Reading Position)

6101 Transporter Motor Control NG (Reading Position)

6102 Transporter Sensor NG (Move to HP)

6103 Transporter HP sensor NG (Move to HP)

6104 Transporter Motor Control NG (Move to HP)

6105 Transporter HP sensor NG (Pressing Position)

6106 Transporter Motor Control NG (Pressing Position)

6107 Transporter Sensor NG (Reading Position 2)

6108 Transporter Motor Control NG (Reading Position 2)

1-7
1.5 REGIUS 110 Error Code

Error Message
Code
6109 Transporter Sensor NG (Pressing Position 2)

6110 Transporter Motor Control NG (Pressing Position 2)

6111 Transporter Motor Control NG (HP Passing Position)

6112 Transporter HP sensor NG (Move to HP -2)

6113 Transporter Motor Control NG (Move to HP -2)

6150 Encoder Initializing NG

6151 Subscan Motor Control NG (Mechanical HP)

6152 Subscan Motor Control NG (Reading Completion Position)

6153 Subscan Motor Control NG (Mechanical HP -2)

6200 Open Insertion Shutter NG

6201 Close Insertion Shutter NG

6250 Motor Lock Release NG (Initialize)

6251 Motor Lock Release NG (Lock)

6252 Motor Lock Release NG (Release)

6300 Justifier Sensor NG (Fixed Action)

6301 Justifier Sensor NG (Fixed)

6302 Casette Exists(Sensor Justifier)

6303 Casette Exists (Sensor Justifier)

6350 Reciever Sensor NG (Feeder)

6351 Reciever Sensor NG (Send)

6352 Reciever Sensor NG (Discharge)

6353 Reciever Sensor NG (Removed)

6354 Casette Exists (Reciever)

6355 Casette Exists (Reciever)

6400 Back Plate detection Sensor NG (Before moving to Reading


Position)

6401 Back Plate detection Sensor NG (Reading Position is Stopping)

6402 Back Plate detecti


on Sensor NG (After moving to Pressing Position)

6450 Peel Sensor Plate (Reading or Erasing)

6451 Peel Sensor NG (Non-operation)

6452 Peel Sensor NG (Start)

6453 Peel Sensor NG (Initialize)

6500 Cassette Insertion Sensor NG (Initialize)

6501 Cassette Insertion Sensor NG (Load)

6502 Cassette Insertion Sensor NG (Discharge)

6503 Back Plate Casette detection

6504 Casette Insertion Sensor NG (Discharge -2)

1-8
1.5 REGIUS 110 Error Code

Error Message
Code
6550 Polygon Rotation NG

6551 Not recieving Reading Completion signal

6552 2 Erase lamps turn off (Start)

6553 2 Erase lamps turn off (Erase)

6554 1 Erase lamp turns off (Start)

6600 Cooler fan's digital power supply disconnected

6650 FPGA Error (Event Driver Access failed)

6651 FPGA Error (Fail to access MDU)

6652 FPGA Error (Fail to access LMC)

6653 FPGA Error (I/O Control failed)

6654 FPGA Error (Fail to open Image Driver)

23000 Program Error (Main Control)

23200 Program Error (Data File Size)

23900 Program Error (Firmware Update)

24000 Network Error (JM Connection)

25000 Network Error (Non-Timeout)

25100 Network Error (Timeout:in General Communication)

25200 Network Error (Timeout:in Image Transmission)

26000 Operation Error (Invalid Job)

26010 Operation Error (Invalid Job Parameter)

26100 Operation Error (Modeless Job)

26200 Operation Error (Invalid Console)

26300 Operation Error (Reader Busy)

26900 Operation Error (PowerOFF under Job Execution)

14000 Signal Proc Error (Gain Overflow)

14001 Signal Proc Error (Gain Underflow)

14002 Signal Proc Error (Offset Overflow)

14003 Signal Proc Error (Offset Underflow)

14010 Signal Proc Error (QRV Overflow)

14011 Signal Proc Error (QRV Underflow)

1-9
1.6 Job Table of Job Manager

1.6 Job Table of Job


Manager

1. Following is the list of contents contained in the job tables, which may be
displayed on Job Manager.

Name Description

jobs:Job Table

jid Job ID automatically issued by the DB.


Range:1 ~ 2147483647

c_order Order No. effective in the console.(corresponds to "jid" by one-to-


one)
Range :1 ~ 2147483647
null :Displayed when the command other than "GetImage"
is registered.

reg_time Job registration date/time on the console.

entry_c Host name of the console with which the job is registered.
(tentative name =desti_c)

dest_c Host name of the console to which the result of implemented job is
sent.

req_r Host name of the reader that is requested to implement the job.

c_rpid Plate ID registered with the console.


00000000000000 ~ 99999999999999
:Plate ID scanned by barcode reader.
ZZZZZZZZZZZZZZ
:Plate ID for manual registration.
null :When the job does not require reading of the plate.

com_code Command Code


Range :0 ~ 199

resol Read Resolution


437 :43.75[μm]
875 :87.5 [μm]
1750 :175.0 [μm]
null :When this parameter is not applicable.

qr_val QR Value(read sensitivity)


50 :125
250 :250
500 :500
null :When this parameter is not applicable.

da_val DAC Input Value(PMT high Volt./ADC Input Range setting).


Range :0 ~ 4095
null :When this parameter is not applicable.

out_bits Bits for image output.


12 :12bit
14 :14bit
16 :16bit
null :When this parameter is not applicable.

1 - 10
1.6 Job Table of Job Manager

Name Description
sig_proc Signal in process
0 :Does not process the signal.
1 :Completes the process up to "gain/offset correction".
2 :Completes the process up to "shading correction".
3 :Completes the process up to "shading correction"
using shading correction coefficient and QRV table,
which were set at the factory.
null :When this parameter is not applicable.

ref_cur Type of reference current supply.


0 :0[nA]
50 :50[nA]
500 :500[nA]
50000 :50000[nA]
50550 :0, 50, 500, 50000[nA] switchable in that order.
null :When this parameter is not applicable.

target050 Theoretical value of 16-bit image at the high voltage setting for
Sen=50.
Range :0 ~ 65535
null :When this parameter is not applicable.

target250 Theoretical value of 16-bit image at the high voltage setting for
Sen=250.
Range :0 ~ 65535
null :When this parameter is not applicable.

target500 Theoretical value of 16-bit image at the high voltage setting for
Sen=500.
Range :0 ~ 65535
null :When this parameter is not applicable.

diag Diagnostic Mode <<NA>>


0 :Diagnosis identical to that of initializing at start up.
1 :Diagnosis of electric system except for mechanical
operation.
2 :Diagnosis of mechanical operation only.
null :When this parameter is not applicable.

firm Per each firmware type <<N/A>>


SCB :Sig firmware for REGIUS 170/190.
MCB :Mech firmware for REGIUS 170/190.
NCB :Net application for REGIUS 110/170/190.
ASC :RFPGA for EGIUS 110.
MAP :Main application for REGIUS 110.
null :When this parameter is not applicable.

sub_com Sub-system command


Commands for debug/maintenance of MCB,SCB,MAP.
null :When this parameter is not applicable.

exec_r Host name of the reader that implemented the job.

job_stat Job implementing status/implemented result.


null :Unsearched
0 :Successfully completed
1 :Searched but not implemented.
2 :Job in process.
3 :Retrying image transfer.
-1 :Failed.

1 - 11
1.6 Job Table of Job Manager

Name Description
r_rpid Plate ID detected by the reader.
Range :'00000000000000' ~ '99999999999999'
null : Displayed until the job search is implemented. Kept in
display for the job processing without plate reading.

max_val Maximum image pixels that was scanned by the reader.(8-bit repre-
sentation)
Range :0 ~ 255
null :When the command is executed while the cassette is
absent.

err_code Errors attributed to abnormal end of job.


Range :'00000' ~ '99999'
null :Except for "job_sat:-1(abnormal end)", or when
abnormal end occurs due to signal processing error/
polygon error/laser error.

door Notification specific to front door error among the errors attributed
to abnormal end of job.
1 :Front door open detected.
nul :Front door in normal state.

sig Notification specific to signal processing error among the errors


attributed to abnormal end of job.
1 :Uses the shading correction coefficient set at the
factory.
2 :Uses the QRV table set at the factory.
3 :Uses the shading correction coefficient/ARV table set
at the factory.
null :No error in signal processing

poli Notification specific to polygon error among the errors attributed to


abnormal end of job.
1 :Polygon PLL error detected.
null :Polygon in normal state.

laser Notification specific to laser error among the errors attributed to


abnormal end of job.
1 :Laser error detected.
null :Laser in normal state.

getj_time Job search date/time of the reader.

commands:Command Table

com_code Command code.


0 ~ 99 :Command for the case with use of cassette (with plate
reading)
100 ~ 199:Command for the case without use of cassette (with-
out plate reading)

com_name Command name.

com_pri Command priority.


0(priority=highest) ~ 7(priority=lowest)

relations:Relation Table

c_name Host name of the console.

r_name Host name of the reader.

alert Error notification from the reader.


0 :Does not notify the error to the associated console.
1 :Notifies the error to the associated console.

1 - 12
1.6 Job Table of Job Manager

Name Description

c_status:Console Status Table

c_name Host name of the console.

img_ack Transmission of "Ack" for received image from the console.


null :Does not send "Ack".
0 :Does not send "Ack".
1 :Sends "Ack".

c_stat Console status.


0 :ready = Low processing load
(capable of receiving the result of job implementation
from the reader)
1 :busy = High processing load
(not capable of receiving the result of job implementa-
tion from the reader)
2 :off = Halting state
(not capable of receiving the result of job implementa-
tion from the reader)
-1 :error = Abnormality exits.
(not capable of receiving the result of job implementa-
tion from the reader)

r_status:Reader Status Table

r_name Host name of the reader.

r_type Device code of the reader.


M :REGIUS 170
V :REGIUS 190
J :REGIUS 110

r_stat Reader status.


0 :ready = ready to start reading.
(capable of receiving/implementing the job sent from
the console)
1 :busy = job in process
(capable of partially receiving the job but not capable
of implementing the job)
2 :off = Halting state
(not capable of receiving/implementing the job sent
from the console)
3 :standby = in standby state.
(capable of receiving/implementing "Wakeup", "Pow-
erOFF" sent from the console.
4 :wait = ready to update the firmware.
(capable of receiving/implementing
"EnterMaintenance", "UpdateFirm" sent from the
console.
-1 :error = Abnormality exists(Fatal Error)
(not capable of receiving/implementing the job sent
from the console)

exec_jid Job ID currently processed on the reader.


Range :1 ~ 2147483647
null :No job under processing.

err_code Error attributed to abnormality of the reader.


Range :00000 ~ 99999
null :except for "r_sat:-1(in error state)"

door Notification specific to front door of the reader.


1 :Front door open
null :Front door in normal state (closed)

1 - 13
1.6 Job Table of Job Manager

Name Description
casin Notification specific to cassette insertion error on the reader.
2 :unable to read the barcode.
3 :read barcode is abnormal.
4 :two cassettes inserted at a time.
5 :Job not registered in the manual registration mode.
6 :two cassette inserted back to back.
7 :failed to read the cassette in the device.
null :cassette insertion successful.

casout Notification specific to cassette ejection error on the reader.


8 :ejection stacker is full.
9 :ejection slot is open.
null :cassette ejection successful.

lamp Notification specific to erasure lamp error on the reader.


A :one erasure lamp is dead.
C :one fluorescent tube is dead.
null :erasure lamp in normal state.

mainte_c Host name of the console on which the maintenance mode is set.

sys_config:System Configuration Table

reg_type Job registration type.


0 :Manual registration mode
1 :Barcode registration mode

n_timeout Timeout setting for network.


0 ~ 10 :0 ~ 10[sec]
null :10[sec]

i_timeout N/A

r_version:Reader Version Table

r_name Host name of the reader.

version Version of the reader update kit.


V :Major Minor1 Minor0
R :Major Minor1
T :Major Minor1

ncb_version Version of Network Control Firmware on the reader.


Net Application @R170/190/110

scb_version Version of SignalControl or System Control Firmware on the


reader.
Sig Firmaware @R170/190 or Main Application @R110

mcb_version Version of Mech Control Firmware or Mech Control FPGA on the


reader.
Mech Firmware @R170/190 or FPGA Config. Data @R110

cf_version Version of the contents in CF card on the reader.

up_time Date/time of updating.

r_log:Reader Log Table<<N/A>>

r_name Host name of the reader.

log_name Reader log file name.

log_data Reader log data.

up_time Date/time of reader log updating.

jm_jnfo:JM Information Table

1 - 14
1.6 Job Table of Job Manager

Name Description
db_ver Version of database.

jm_ver Version of Job Manager.

jbt_ver Version of Job Table.

cmt_ver Version of Command Table.

rlt_ver Version of Relation Table.

cst_ver Version of Console Status Table.

rst_ver Version of Reader Status Table.

sct_ver Version of System Configuration Table.

rvt_ver Version of Reader Version Table.

rgt_ver Version of Reader Log Table.

jit_ver Version of JM Information Table.

getjp_ver Version of Job Search Function (for cassette kick).

getjn_ver Version of Job Search Function (for console kick).

1 - 15
1.6 Job Table of Job Manager

1 - 16
Connecting CS-1/2/3
Chap.2
with REGIUS 110

Make setting of CS-1/2/3,REGIUS110 and JM so that


they can be networked.Because different setting is
required depending on the system configuration,
description is separately made for each of Manual Reg-
istration, Barcode Registration (JM: built-in) and Bar-
code Registration (JM:enabled). Only principles of
setting are described in this chapter. Please refer to
"Ch.6 Service Tool Screens" for details of setting items
on each setup screen.
Blank Page
2.1 Flow of Setup Procedure

2.1 Flow of Setup Pro- Setup procedure for REGIUS110 networking to CS-1/2/3 as well as that for
cedure combination of REGIUS110 + REGIUS170/190 networking to CS-1/2/3 are
described. Clauses to be referred to for each step is also listed below.
■ CS-1
Start

1. Unpacking
Unpack the CS-1 and check the accessaries, etc.
listed in the packing list.

2. Checking the operation of CS-1

Connect the mouse and check the operation.

3. Installing Lead Battery/Option Board

Install the UPS(lead battery) and option board on the


CS-1 Control Unit.

4. Setting the Network "2.2 Setting Up the 1-to-1 Connection"


"2.3 Setting up the n-to-m Connection"
Set the network condition for REGIUS 110/170/190
JM CS-1 in this order.

5. Connecting to the System


Connect to the actual network, and connect various
devices necessary to complete the system.
Printer/RIS/HOST/IDS, etc.
System reader

6. Setting the CS-1


Set the information of other than network such as
site’s information,connecting device information, etc.
using CS-1.

7. Operation Check "3.1 Checking the Image"


"3.2 Backing Up the Setup Data"
Check the image output from REGIUS 110/170/190.

7.5. Installing the Mammo System

8. Calibration
Calibrate the system reader and adjust the brightness
of the CS-1 Operation Unit.
"4.1 Checking and Adjusting the Image Size"
9. Back Up "4.2 Unevenness Calibration of Regular Cassette"
"4.3 Sensitivity Calibration of Regular Cassette"
Back up the setup data of CS-1 and parameters of
REGIUS 110/170/190, etc.
(refer to each "Installation/Service Manual" for those
Indicates the step necessary for connecting
not listed in this manual)
End REGIUS 110 additionally to the existing CS-1.

2-1
2.1 Flow of Setup Procedure

■ CS-2

Start

1. Unpacking/Connection

2. Setting the CS-2 Operation Unit

3. Updating the CS-2 PROGRAM

"2.2 Setting Up the 1-to-1 Connection"


4. Connecting CS-2 and REGIUS 110
"2.3 Setting up the n-to-m Connection"
Implement settings to network CS-2, REGIUS 110
and JM.

5. Connecting to the System


Move the CS-2 and REGIUS 110 to final installation
site or location and connect to the network. Then
connect optional devices. At the same time, connect a
densitometer to adjust the luminance of CS-2
Operation Unit.

6. Setting the CS-2

Set the information of the site, peripheral devices, etc.

7. Adjusting Luminance of CS-2 Operation Unit

8. Operation Check
"3.1 Checking the Image"
Check the image output from REGIUS 110. "3.2 Backing Up the Setup Data"
"4.1 Checking and Adjusting the Image Size"
9. Back Up "4.2 Unevenness Calibration of Regular Cassette"
"4.3 Sensitivity Calibration of Regular Cassette"
Back up the setup data of CS-2, parameters of
REGIUS 110, etc.
(refer to each "Installation/Service Manual" for those
Indicates the step necessary for connecting not listed in this manual)
End REGIUS 110 additionally to the existing CS-1.

2-2
2.1 Flow of Setup Procedure

■ CS-3

Start

1. Unpacking
Unpack the CS-3 and check the accessaries, etc.
listed in the packing list.

2. Checking the Operation of CS-3

Connect the mouse and check the operation.

3. Installing Option Board

Install an option board on the CS-3 Control Unit.

4. Setting the Network "2.2 Setting Up the 1-to-1 Connection"


"2.3 Setting up the n-to-m Connection"
Set the network condition for REGIUS 190/170 JM
CS-3 in this order.

5. Connecting to the System


Connect to the actual network, and connect various
devices necessary to complete the system.
Printer/RIS/HOST/IDS, etc.
System reader
Magnetic card reader, Barcode reader

6. CSetting the CS-3


Set the information of other than network such as
institution’s information,connecting device information,

7. Operation Check "3.1 Checking the Image"


"3.2 Backing Up the Setup Data"
Check the image output from REGIUS 190/170.
"4.1 Checking and Adjusting the Image Size"
"4.2 Unevenness Calibration of Regular Cassette"
7.5. Installing the Mammo System
"4.3 Sensitivity Calibration of Regular Cassette"

8. Calibration
Calibrate the system reader and adjust the luminance
of the CS-3 Operation Unit.

9. Back Up

Back up the setup data of CS-1 and parameters of


REGIUS 190/170, etc.
(refer to each "Installation/Service Manual" for those
not listed in this manual)
Indicates the step necessary for connecting
End REGIUS 110 additionally to the existing CS-1.

2-3
2.2 Setting Up the 1-to-1 Connection

2.2 Setting Up the 1-to-1 This section describes how to connect a single CS-1/2/3 to a single
REGIUS 110 using a 1-to-1 connection.
Connection
Small-sized networks such as connecting a single CS-1/2/3, REGIUS
110 and KonicaMinolta printer can be created without changing the
factory default settings. When a factory default CS-1/2/3 is connected
to a REGIUS 110, the manual registration method is used.
However, for the case described below, change the setting of CS-1/2/3
and REGIUS 110 according to the procedure described in this section.
• When it is necessary to change the address because the system is
connected to the site's existing network.
• To connect several 1-to-1 connection units to the same network.

The explanation in this section is based on the settings instructed by


the institute as following.

Printer
Viewer, Server

Ethernet

HUB

CS-1/2/3
Serial No. 0025 REGIUS 110
Host Name CS2-0025 Serial No. 0100
IP Address 128.100.50.1 Host Name r170-0100
Subnet Mask 255.255.0.0 IP Address 128.100.50.11
Cassette Regist. Auto-Regist. Subnet Mask 255.255.0.0

<Important>It is possible to configure a 1-to-1 connection system


to use barcode registration; however, as a rule, you
should use the manual registration or auto-registration
method. It is necessary to arrange a barcode reader
when barcode registration is selected.

2-4
2.2 Setting Up the 1-to-1 Connection

2.2.1 Setting Up the


REGIUS 110

Temporary Connection of CS-1/2/3 To connect a CS-1/2/3 to a REGIUS 110 via Ethernet, the devices have
and REGIUS 110 been configured so that they can communicate using the factory
default. The set up of the REGIUS 110 below shall be made by
temporary connecting the CS-1/2/3 in a 1-to-1 configuration.

1. Plug the REGIUS 110 into a wall outlet.

2. Using a cross cable or hub, connect the CS-1/2/3 and REGIUS 110 in
a 1-to-1 configuration.

<Important>This configuration is a temporary case purely for the


purposes of configuring the REGIUS 110. Do not connect
several REGIUS 110s to the hub at the same time. Doing
so prevents correct recognition of each device due to
the fact that all REGIUS 110s are preset with the same IP
address, and in consequence it becomes impossible to
set up the device.

Temporary Connected Hub REGIUS 110


CS-1/2/3 to be set

or

Cross Cable

<Important>If the CS-1/2/3 or JM settings are changed before set-


ting up the REGIUS 110, the REGIUS 110 cannot be set
up because communication with the REGIUS 110 is dis-
abled. If this happens, reset the CS-1/2/3 to the factory
defaults and try again.

Starting up a CS-1/2/3 and REGIUS


110

1. Turn ON the circuit breaker for the REGIUS 110.

2. Press the power switch on the CS-1/2/3 Control Unit to start up the CS-
1/2/3. Wait until the initial screen shows after the CS-1/2/3 starts up.
Wait for the REGIUS 110 to complete initialization and to show
"READY" on the LCD.

2-5
2.2 Setting Up the 1-to-1 Connection

Setting Up the REGIUS 110 Following procedures should be initiated by CS-1/2/3 Control Unit.

1. Open "SYSTEM INFO (reader)" screen and set up as follows.


([SYSTEM INFO] → [System Config. (reader)])

Setting Item Setting Details


REGIUS 110/170/190

Reader-#1 Select "REGIUS 110"

Displayed Name Input the display name of REGIUS 110.


This name appears in field of "System Info" screen or
"Service Tool" screen of the CS-1/2/3 application for
selection of target REGIUS 110.

2-6
2.2 Setting Up the 1-to-1 Connection

2. Start the "Service Tool" from the "REGIUS Service" screen, and open the
"Network Setup" screen. (click [Reader] -> [Network])
Network information received from the REGIUS 110 will be displayed on the
"Network Setup" screen.

<Important>If failed to receive the network information from


REGIUS 110, a message "Failed to output file" will be
displayed.

"CS-1" has been registered as the Host on the factory default on REGIUS 110.

2-7
2.2 Setting Up the 1-to-1 Connection

3. Set the network condition, host name for the REGIUS 110, CS-1/2/3 and JM
referring to the following table.

Setting Item Setting Details


Reader

Host Name Input the Host Name of the REGIUS 110 itself.
r110-*****:***** indicates a serial number.

<Important>Always use lowercase to


input.

IP Address Input the IP address of the REGIUS 110.

Subnet Mask Input the Subnet Mask of the network.

DBHOST

DBHOST Input the Host Name of the JM.


Fixed at "jm1-0001".

<Important>Always use lowercase to


input.
Hosts

Host Name Input the Host Name of the CS-1/2/3 that is connected
to the REGIUS 110.

<Important>Always use uppercase to


input.

IP Address Input the IP Address of the CS-1/2/3 that is connected


to the REGIUS 110.

Host Name Input the Host Name of the JM.


Fixed at "jm1-0001".

<Important>Always use lowercase to


input.

IP Address Input the IP Address of the JM.

<Important>Assign the same IP address to the Host name CS2-


0025 and jm1-0001. Because an access to the Console
(CS-1/2/3) and JM is made for different purpose, strictly
follow the setting as shown above.

4. Click [Send].
Settings displayed in the screen will be sent to the REGIUS 110 and the net-
work information is updated.
• The updated settings will be written in a CF (Compact Flash Memory) of
the REGIUS 110.

<Important>When the connection is made in 1:1, only one unit of


CS-1/2/3 that is connected to the REGIUS 110 shall be
registered in the "Hosts" list.

5. Click [End] to close the "Network Setup" screen. Return to the "Ser-
vice Tool" screen (Reader).

2-8
2.2 Setting Up the 1-to-1 Connection

6. Press and hold the [operation] switch for 3 seconds to turn OFF the REGIUS
110.

7. Turn OFF the circuit breaker of the REGIUS 110.


Now the setup of REGIUS 110 is completed. Next, proceed to the setup of
CS-1/2/3.

<Important>The REGIUS 110 whose network condition has been


changed is no more capable of communicating with the
CS1/2/3 until the network condition of the CS-1/2/3 (and
JM) is newly set up. Upon the start up of the REGIUS 110,
a network error appears on the LCD after completing the
initialization, and IP address will be displayed.

2-9
2.2 Setting Up the 1-to-1 Connection

2.2.2 Setting Up the CS-1/2/ This section describes how to set up JM on the CS-1/2/3, the settings
3 required to connect the CS-1/2/3 to the REGIUS 110 and the network
settings for the CS-1/2/3 itself.

Setting up the JM on the CS-1/2/3

■ Setting the JM information

1. Click the [Job Manager] button of the "Input/Output".


"JOBM INFO" screen appears.

By default, "CS1-0001" is registered as the Host name in JM on the


CS-1/2/3. This is so to make it possible to connect the CS-1/2/3 to a
REGIUS 110 using the factory defaults.

2. Set each item as follows.


Setting Item Setting Details
Basic Setting

Cassette Regist. "Manual", "Auto" or "Barcode" registration

Host Name regis- Input the Host Name of CS-1/2/3 itself.


tered with JM
<Important>Always use uppercase to
input.

Communication Setting

Built-in Job Manager ON

3. Click [OK] button, then click [Yes] in the confirmation dialog.


"Service Tool(Console)"screen appears.

2 - 10
2.2 Setting Up the 1-to-1 Connection

■ Setting up the JM database If the host name of the CS-1/2/3 or REGIUS 110 has been changed
from the factory default, you must use the following procedure to
update the information registered in the database.

<Important>Always use uppercase characters to input the host


name for the CS-1/2/3 and lowercase to input the host
name for the REGIUS110.

1. Click [Adjustment] of "Service Tool" screen, then click [Configure] of "Job


Manager".
"PostgreSQL access" tool starts up.

<Important>On each of "PostgreSQL" screen, click [Reload] fol-


lowed by [Close] to close the screen after completing
input.

2 - 11
2.2 Setting Up the 1-to-1 Connection

Changing the Registered Informa-


tion of CS-1/2/3 and REGIUS 110

1. When the Host Name of CS-1/2/3 is changed, change the registration as fol-
lows.

• < "c_status" screen >
Rewrite "CS1-0001" with the new Host Name (uppercase).

c_name c_stat
Host Name of CS-1/2/3 2

<Important>By default, "CS1-0001" is registered as the Host name


in JM on the CS-1/2/3. This is so to make it possible to
connect the CS-1/2/3 to a REGIUS 110/170/190 using the
factory defaults.

2. When the Host Name of REGIUS 110 is changed, rewrite the name as fol-
lows.
• < "r_status" screen >
Rewrite "r170-0001" with the new Host Name (lowercase) of REGIUS
110.

<Important>Always use lowercase to input the Host Name of the


REGIUS 110 when using "PostgreSQL".

r_name r_stat
Host Name of REGIUS 110 2

2 - 12
2.2 Setting Up the 1-to-1 Connection

• < "r_version" screen >
Rewrite "r170-0001" with the new Host Name (lowercase) of REGIUS
110.

r_name
Host Name of REGIUS 110

3. When the Host Name of CS-1/2/3 or REGIUS 110 is changed, change the reg-
istration as follows.
< "relations" screen >
Rewrite "CS1-0001" with the new Host Name (uppercase). When the Host
Name of REGIUS 110 is changed, click [r170-0001], and rewrite the Host
Name with the new one (lowercase).

<Important>Always use uppercase to input the Host Name of CS-


1/2/3, lowercase to input the Host Name of REGIUS 110
when using "PostgreSQL".

c_name r_name
Host Name of CS-1/2/3 Host Name of REGIUS 110

2 - 13
2.2 Setting Up the 1-to-1 Connection

Setting the cassette registration Set the cassette registration method.


method "1" (manual registration or auto-registration) is set by default.
< "sys_config" screen >

reg_type
0 or 1

1. Set "reg_type" to "1" (manual registration or auto-registration) or to


"0"(barcode registration).

2. Click [Database] menu of "PostgreSQL access" screen, and select "Exit".

Returns to "Service Tool(Adjustment)" screen exiting "PostgreSQL access"


tool.

2 - 14
2.2 Setting Up the 1-to-1 Connection

■ Establishing a network connec-


tion between the CS-1/2/3 and
REGIUS 110

1. Open "CCU INFO" screen, and set up as follows.


([Reader] →[Reader])

Setting Item Setting Details


TCP/IP

Host Name (on the Input the Host Name assigned to the REGIUS 110.
Job Manager)
<Important>Always use lowercase to
input.

IP Address Input the IP Address assigned to the REGIUS 110.

■ Network setting of CS-1/2/3 itself

1. Terminate the Service Tool and click [Windows Desktop] on the REGIUS
Service screen.
Windows desktop will be displayed.

2. Open "Properties" screen of "Internet Protocol (TCP/IP)".


([Control Panel] → [Network & Internet Connection] → [Network Connec-
tion] → [Local Area Connections] → [Properties(P)] → [Internet Proto-
col(TCP/IP)])

3. Set the IP Address and Subnet Mask.


4. Click [OK] to close all dialogues.

2 - 15
2.2 Setting Up the 1-to-1 Connection

■ Shutting down the system

1. Click [Shutdown Option (U)] -> [Shut down (U)] of the [Start] menu to turn
OFF the CS-1/2/3.

2. Remove the Ethernet cables temporarily connected from the CS-1/2/3 Control
Unit and REGIUS 110.

Implementing the above procedures completes the set up of one each


of REGIUS 110 and CS-1/2/3.
To set up several CS-1/2/3s and REGIUS 110s, repeat the procedures
in and after "2.2.1 Setting Up the REGIUS 110".

2 - 16
2.3 Setting up the n-to-m Connection

2.3 Setting up the n-to- This section describes the n-to-m configuration that includes two CS-1/2/3s
m Connection and two REGIUS 110s.
In this example, the settings are based on the conditions requested by the site.

Barcode Reader
REGIUS 110 #1
Serial No. 0100
Host Name r170-0100
Ethernet IP Address 128.100.50.11

CS-1/2/3 #1(Std.JM
Serial No. 0025
Host Name CS2-0025
IP Address 128.100.50.1 HUB
Subnet Mask 255.255.0.0 Printer,
Cassette Regist. Barcode Regist. Host

REGIUS 110 #2
Serial No. 0101
Host Name r170-0101
CS-1/2/3 #2 IP Address 128.100.50.12
(Backup JM
Serial No. 0026
Host Name CS2-0026
IP Address 128.100.50.2
Subnet Mask 255.255.0.0
Cassette Regist. Barcode Regist

<Important>In the n-to-m configuration, the registration mode


must be set to "Barcode Registration".

<Important>To start operating the system, always turn on the


power of the CS-1/2/3 which incorporates the standard
JM, and turn off this CS-1/2/3 last when terminating the
operation. Instruct the users to observe this rule in daily
operation.

2 - 17
2.3 Setting up the n-to-m Connection

2.3.1 Setting Up the


REGIUS 110

Temporarily connecting the CS-1/2/ To enable the CS-1/2/3 to implement the set up of REGIUS 110 via
3 and REGIUS 110 Ethernet, both have been set up at the factory so that they can
communicate using the factory setting. The set up of REGIUS 110 in
the below shall be made by temporarily connecting the CS-1/2/3 in the
"1-to-1" configuration.

1. Plug the REGIUS 110 into a wall outlet.


2. Connect the CS-1/2/3 and REGIUS 110 in 1-to-1 using cross cable or hub.
<Important>This connection is a temporary case purely for the
purpose of setting the REGIUS 110. Do not connect sev-
eral units of REGIUS 110 to hub at the same time. Doing
so hinders recognition of each device due to the fact that
all of REGIUS 110 are preset at the same IP address and
in consequence it becomes impossible to set up.

Temporary Connected Hub REGIUS 110


CS-1/2/3 to be set

or

Cross Cable

<Important>If the CS-1/2/3 or JM settings are changed before set-


ting up the REGIUS 110, the REGIUS 110 cannot be set
up because communication with the REGIUS 110 is dis-
abled. If this happens, reset the CS-1/2/3 to the factory
defaults and try again.

Starting up a CS-1/2/3 and REGIUS


110

1. Turn ON the circuit breaker for the REGIUS 110.

2. Press the power switch on the CS-1/2/3 Control Unit to start up the CS-
1/2/3. Wait until the initial screen shows after the CS-1/2/3 starts up.
Wait for the REGIUS 110 to complete initialization and to show
"READY" on the LCD.

2 - 18
2.3 Setting up the n-to-m Connection

Setting Up the REGIUS 110 Following procedures should be initiated by CS-1/2/3 Control Unit.

1. Open "SYSTEM INFO(reader)" screen, and set as the following table.


([SYSTEM INFO] → [System Configuration (reader)])

Setting Item Setting Details


REGIUS 110/170/190

Reader-1 Select "REGIUS 110"

Displayed Name Input the display name of REGIUS 110.


This name is the name which is displayed in the field
for selecting an individual REGIUS 110 for CS-1/2/3
application on the “System Info” screen or “Service
Tool”.

Reader-2 Select "REGIUS 110"

Displayed Name Input the display name of REGIUS 110.


This name appears in field of "System Info" screen or
"Service Tool" screen of the CS-1/2/3 application for
selection of target REGIUS 110.
"Displayed Name" is the name which is displayed in the field for selecting an individ-
ual REGIUS 110 for CS-1/2/3 application on the “System Info” screen or “Service
Tool”.
Input the name that can be easily recognized by the user. (for example, the
name of X-ray room where the REGIUS 110 is installed)
Ex.:X-ray Room No.1, X-ray Room No.2, etc.

2 - 19
2.3 Setting up the n-to-m Connection

2. Start the "Service Tool" from the "REGIUS Service" screen, and open the
"Network Setup" screen. (click [Reader] -> [Network])
"Reader" screen will be displayed.
The reader device will be exhibited in the window in order of "Display Name"
and "Reader Type".

3. Select the REGIUS 110 to be set up, and click [OK].


Network information received from the REGIUS 110 will be displayed on the
"Network Setup" screen.

<Important>If failed to receive the network information from


REGIUS 110, a message "Failed to output file" will be
displayed.

"CS-1" has been registered as the Host on the factory default on REGIUS 110.

2 - 20
2.3 Setting up the n-to-m Connection

4. Set the network condition, host name for the REGIUS 110, CS-1/2/3 and JM
referring to the following table.

Setting Item Setting Details


Reader

Host Name Input the Host Name of the REGIUS 110 itself.
r110-*****:***** indicates a serial number.

<Important>Always use lowercase to


input.

IP Address Input the IP address of the REGIUS 110.

Subnet Mask Input the Subnet Mask of the network.

DBHOST

DBHOST Input the Host Name of the JM.


Fixed at "jm1-0001".

<Important>Always use lowercase to


input.
Hosts

Host Name Input in each column the Host Names of each CS-1/2/3
that is connected to the REGIUS 110.

<Important>Always use uppercase to


input.

IP Address Input in each column the IP Address of each CS-1/2/3


that is connected to the REGIUS 110.

Host Name Input the Host Name of the JM.


Fixed at "jm1-0001".

<Important>Always use lowercase to


input.

IP Address Input in each column the IP Address of the std. JM and


backup JM.

<Important>Assign the same IP address to the Host name CS2-


0025 and jm1-0001. Because an access to the Console
(CS-1/2/3) and JM is made for different purpose, strictly
follow the setting as shown above. Same should be
applied to "CS2-0026" and "jm1-0002".

5. Click [Send].
Settings displayed in the screen will be sent to the REGIUS 110 and the net-
work information is updated.
• The updated settings will be written in a CF (Compact Flash Memory) of
the REGIUS 110.

6. Click [Exit] to close the "Network Setup" screen. Return to the "Ser-
vice Tool" screen (Reader).

2 - 21
2.3 Setting up the n-to-m Connection

7. Press and hold the [operation] switch for 3 seconds to turn OFF the REGIUS
110.

8. Turn OFF the circuit breaker of the REGIUS 110.


<Important>When CS-1/2/3 and REGIUS 110 are connected in n-
to-m connection, set "Console/JobManage" and "Hosts"
for all REGIUS 110s connected to the same while set
"Host Name" and "IP Address" of the "Reader" to identi-
cal per each REGIUS 110.

<Important>The REGIUS 110 whose network condition has been


changed is no more capable of communicating with the
CS1/2/3 until the network condition of the CS-1/2/3 (and
JM) is newly set up. Upon the start up of the REGIUS 110,
a network error appears on the LCD after completing the
initialization, and IP address will be displayed.

9. Turn OFF the CS-1/2/3.


10. Remove the temporarily connected Ethernet cable from REGIUS 110.
Now the setup of REGIUS 110 is completed. Repeat the procedure in and
from " Temporarily connecting the CS-1/2/3 and REGIUS 110" to set the rest
of REGIUS 110s that are connected to the network.

2 - 22
2.3 Setting up the n-to-m Connection

2.3.2 Setting Up the CS-1/2/ This section describes how to set up JM on the CS-1/2/3 of Std. JM,
3 of Std. JM the settings required to connect the CS-1/2/3 to the REGIUS 110 and
the network settings for the CS-1/2/3 itself. CS-1/2/3 of Backup JM will
be described in " Implement the setting of CS-1/2/3 of Backup JM, the
setting for connection of CS-1/2/3 with REGIUS 110 and the network
setting of CS-1/2/3." .

Setting up the JM on the CS-1/2/3


of Std. JM

■ Setting the JM information

1. Click the [Job Manager] button of the "Input/Output".


"JOBM INFO" screen appears.

By default, "CS1-0001" is registered as the Host name in JM on the


CS-1/2/3. This is so to make it possible to connect the CS-1/2/3 to a
REGIUS 110 using the factory defaults.

2 - 23
2.3 Setting up the n-to-m Connection

2. Set each item as follows.


Setting Item Setting Details
Basic Setting

Cassette Regist. "Barcode Regist."

<Important>Always set to "Barcode Reg-


ist." when the connection is n-to-
m.

Host Name regis- Input the Host Name of CS-1/2/3 itself.


tered with JM
<Important>Always use uppercase to
input.

Communication Setting

Built-in Job Manager ON

Communication Setting (Backup)

Device Registered Tick the checkbox when the backup JM is used.

IP Address Input the IP Address of backup JM.

3. Click [OK] button, then click [Yes] in the confirmation dialog.


"Service Tool(Console)"screen appears.

2 - 24
2.3 Setting up the n-to-m Connection

■ Setting up the JM database If the host name of the CS-1/2/3 or REGIUS 110 has been changed
from the factory default, you must use the following procedure to
update the information registered in the database.

<Important>Always use uppercase characters to input the host


name for the CS-1/2/3 and lowercase to input the host
name for the REGIUS110.

1. Click [Adjustment] of "Service Tool" screen, then click [Configure] of "Job


Manager".
"PostgreSQL access" tool starts up.

<Important>On each of "PostgreSQL" screen, click [Reload] fol-


lowed by [Close] to close the screen after completing
input.

2 - 25
2.3 Setting up the n-to-m Connection

Registering CS-1/2/3 and REGIUS


110

1. Rewrite as follows, and register each of CS-1/2/3.


• < "c_status" screen >
<Important>Use uppercase to input the Host Name of CS1/2/3
when using PostgreSQL tool.

c_status c_stat
Host Name of CS-1/2/3 #1 2

Host Name of CS-1/2/3 #2 2

<Important>By default, "CS1-0001" is registered as the Host name


in JM on the CS-1/2/3. This is so to make it possible to
connect the CS-1/2/3 to a REGIUS 110 using the factory
defaults.

2. Rewrite as follows, and register each of REGIUS 110.


• < "r_status" screen >
<Important>Use lowercase to input the Host Name of REGIUS 110
when using PostgreSQL tool.

r_name r_stat
Host Name of REGIUS 110 #1 2

Host Name of REGIUS 110 #2 2

2 - 26
2.3 Setting up the n-to-m Connection

• < "r_version" screen >
<Important>Use lowercase to input the Host Name of REGIUS 110
when using PostgreSQL tool.

r_name
Host Name of REGIUS 110 #1

Host Name of REGIUS 110 #2

Setting the Relations Set the relations between the CS-1/2/3(s) and REGIUS 110(s). In the
"n-to-m connection", setting the relations between the CS-1/2/3 and
REGIUS 110 will enable a linked-operation as to the power control,
error information between the CS-1/2/3 and corresponding REGIUS
110.
This section describes settings used to associate all of CS-1/2/3 and
REGIUS 110s.

1. Rewrite as the following, and set the relation between the CS-1/2/3 and
REGIUS 110.
• < "relations" screen >
<Important>Use uppercase to input the Host Name of CS-1/2/3
and lowercase to input the Host Name of REGIUS 110.

Example:All of CS-1/2/3 is associated with all of REGIUS 110.

2 - 27
2.3 Setting up the n-to-m Connection

Setting the cassette registration This section describes how to set the cassette registration method.
method By default, the method is set to "1" (Manual Registration or Auto-
registration).

1. Rewrite as follows, and set the cassette registration method.


• < "sys_config" screen >

"reg_type" set here and the cassette registration method set in " ■ Setting the
JM information" have the relation as the following. Always make settings so
that the following relation is established.

Cassette Regist. reg_type


Manual Regist. 1

Auto-Regist. 1

Barcode Regist. 0

2. Click [Database] menu of "PostgreSQL access" screen, and select "Exit".

Returns to "Service Tool(Adjustment)" screen exiting "PostgreSQL access"


tool.

■ Backing up JM information Back up the settings to a floppy disk after completing the settings of the
JM database.

1. Insert a floppy disk for backup in the floppy disk drive of CS-1/2/3 Control
Unit.

2 - 28
2.3 Setting up the n-to-m Connection

2. Click [Backup] of the "Job Manager".

A command prompt appears on the screen and backup starts.


Backup data is saved in a file "jm_backup.bat" of the floppy disk with
file names as “c_status”, “r_status”, “relations”, “kddm_jm”, “r_version”,
“sys_config”. When a floppy disk is not available, it will be saved in
“C:\KonicaMinolta\JM\backup” with file names as “c_status”,
“r_status”, “relations”, “kddm_jm”, “r_version”, “sys_config”.
Upon completion of backup, the command prompt closes.

3. Remove the backup floppy disk from the floppy disk drive.

2 - 29
2.3 Setting up the n-to-m Connection

■ Setting the CS-1/2/3 to connect This section describes how to set up the REGIUS 110s connected to
with the REGIUS 110 the CS-1/2/3 and the device name for each REGIUS 110.

1. Open "SYSTEM INFO(reader)" screen, and set as the following table.


([SYSTEM INFO] → [System Configuration (reader)])

Setting Item Setting Details


REGIUS 110/170/190

Reader-1 Select "REGIUS 110"

Displayed Name Input the display name of REGIUS 110.


This name is the name which is displayed in the field
for selecting an individual REGIUS 110 for CS-1/2/3
application on the “System Info” screen or “Service
Tool” screen.

Reader-2 Select "REGIUS 110"

Displayed Name Input the display name of REGIUS 110.


This name appears in field of "System Info" screen or
"Service Tool" screen of the CS-1/2/3 application for
selection of target REGIUS 110.
"Displayed Name" is the name which is displayed in the field for selecting an individ-
ual REGIUS 110 for CS-1/2/3 application on the “System Info” screen or “Service
Tool” screen.
Input the name that can be easily recognized by the user. (for example, the
name of X-ray room where the REGIUS 110 is installed)
Ex.:X-ray Room No.1, X-ray Room No.2, etc.

2 - 30
2.3 Setting up the n-to-m Connection

2. Open "CCU INFO" screen and set as follows.


([Reader] →[Reader])

Setting Item Setting Details


TCP/IP

Host Name Input the Host Name set for the REGIUS 110.

Host Name Input the Host Name set for REGIUS 110 using lower-
(on Job Manager) case.

<Important>Always use lowercase to


input.

IP Address Input the IP address set for the REGIUS 110.

Switch between the device names exhibited at the upper left of the screen, and
set each of connecting REGIUS 110 as above.

3. Click [OK] button, then click [Yes] in the confirmation dialog.


"Service Tool(Reader)"screen appears.

2 - 31
2.3 Setting up the n-to-m Connection

■ Setting the network of CS-1/2/3

1. Terminate the Service Tool and click [Windows Desktop] on the REGIUS
Service screen to display the Windows desktop.

2. Open "Properties" screen of "Internet Protocol (TCP/IP)".


([Control Panel] → [Network & Internet Connection] → [Network Connec-
tion] → [Local Area Connections] → [Properties(P)] → [Internet Proto-
col(TCP/IP)])

3. Set the IP Address and Subnet Mask.


4. Click [OK] to close all dialogues.
■ Shutting down the system

1. Click [Shutdown Option (U)] -> [Shut down (U)] of the [Start] menu to turn
OFF the CS-1/2/3.

Implementing the above procedure complete the set up of CS-1/2/3.


Next, proceed to set up of CS-1/2/3 of Backup JM.

2.3.3 Setting the CS-1/2/3 of Implement the setting of CS-1/2/3 of Backup JM, the setting for connection of
Backup JM CS-1/2/3 with REGIUS 110 and the network setting of CS-1/2/3.

■ Setting the JM information

1. Click [Job Manager] of "I/O".


"JOBM INFO" screen will be displayed.

2 - 32
2.3 Setting up the n-to-m Connection

2. Set as follows.
Setting Item Setting Details
Standard Setting

Cassette Regist. "Barcode Registration"


Method
<Important>In the case of "m-to-n connec-
tion", always set to "Barcode
Registration".

Host Name regis- Input the host name of CS-1/2/3.


tered on the JM
<Important>Always use uppercase to
input.

Communication Setting

Built-in Job Manager OFF

IP Address Input the IP address of CS-1/2/3 for standard JM.

Communication Setting (Backup)

Device Registered ON

Built-in Job Manager ON

IP Address Input the IP address of backup JM.

3. Click [OK], and click [Yes] of the confirmation dialogue.


Returns to "Service Tool (Console)" screen.

2 - 33
2.3 Setting up the n-to-m Connection

■ Setting the database of the Register with the backup JM the same database that has been set in
backup JM " ■ Setting up the JM database".

1. Using [Adjustment] → [Restore] of "Service Tool" screen and "Return to


the”Service Tool(Reader)” screen , restore the Job Manager settings.

Command Prompt will be displayed on the screen, and restoration starts.


Command prompt will be closed when restoration is completed.
Completing the procedure will set the database of the built-in JM in CS-1/2/3
same as that set in CS-1/2/3 #1.

2. Check that the setting is made as follows on "c_status" window of


PostgreSQL. ([Adjustment] -> [Setting] ->[c_status])
• If the value of "c_stat" is not "2", click on the value and change it to "2".

3. Check that the setting is made as follows on "r_stat" window of PostgreSQL.


([Adjustment] -> [Setting] ->[r_stat])
• If the value of "r_stat" is not "2", click on the value and change it to "2".

2 - 34
2.3 Setting up the n-to-m Connection

4. Click [Database] menu of "PostgreSQL access" screen, and select [Exit].

Return to "Service Tool" screen after closing "PostgreSQL access" screen.

2 - 35
2.3 Setting up the n-to-m Connection

■ Setting to connect CS-1/2/3 with Implement the setting of the number of REGIUS 110 that shall be connected
REGIUS110 with CS-1/2/3 and the name of each REGIUS 110.
In this procedure, set the settings the same configuration that has been set in
the built-in standard JM.

1. Using the same procedure practiced for " ■ Setting the CS-1/2/3 to connect
with the REGIUS 110", implement the setting of device name of REGIUS 110
and CCU INFO on "SYSTEM INFO" screen.

■ Setting the network of CS-1/2/3 used


as the backup JM

1. Exit "Service Tool" screen. Then set the network for CS-1/2/3 in the same
manner as practiced for " ■ Setting the network of CS-1/2/3"

■ Shutting down the system

1. Click [Shutdown Option (U)] -> [Shut down (U)] of the [Start] menu to turn
OFF the CS-1/2/3.

2 - 36
Chap.3 Verification & Backup of
Setup Data

Scan the solid density image, and check that the cap-
tured image is normal.
Then back up the setting data and complete the installa-
tion procedure.
Blank Page
3.1 Checking the Image

3.1 Checking the Image Using REGIUS 110, scan the cassette on which the solid density image was
exposed, and check the image using "Service Tool".

<Important>Refer to the Operation Manual of CS-1/2/3 for the


operation of "Exam. Registration" screen or "Image Veri-
fication" screen.

1. Turn on the power switch to start CS-1/2/3.


Application software of CS-1/2/3 starts, and "Initial" screen will be displayed.

2. Press [operation] switch and start up REGIUS 110 that should be checked.
3. Input the tentative Patient Information on CS-1/2/3 Control Unit, and switch
to "Image Verification" screen.

4. Touch the rectangular(others) at the lower left of the anatomy, then touch
[TEST] to select.

5. Expose the solid density image on the test cassette.


• Make an exposure at the condition of Tube Voltage: 80kV, 10mAs, Dis-
tance at approx. 2m.
• Align the long side of the cassette with the vertical direction of the X-ray
tube.

2m

Vertical direction Position the test cassette Long side of


of X-ray tube to allow the blue line to the cassette
lie on the bottom

6. If the system is set to "Barcode Registration", read the barcode of the cassette
that was exposed.
"Barcode Registration" : the bar code of the cassette is displayed in the order.
"Manual Registration" : "1st" is indicated.
"Auto-registration" : Nothing on the screen.

7. Insert the cassette into the REGIUS 110.


As the image in the cassette is read, the image is displayed on the
screen of operation unit.

8. [Touch [OK].

3-1
3.1 Checking the Image

9. Touch [Exam. Completed].


10. Returns to the initial screen.
11. Touch [KonicaMinolta].
System menu will be shown.

12. Touch [Utilities] to start the "Service Tool".


13. Touch [Analyze].
"Analyze" screen will be shown.

14. Click [Image].


Image-check software will start.

3-2
3.1 Checking the Image

15. Select [Open (O)] in the [File (F)] menu.


[Open] dialogue will be shown.

16. Select "1.img", and touch [Open].


Solid density image that is read by the REGIUS 110 will be displayed
on the screen of image-check software.

• Select the latest file(with the largest number) if several image file is
listed in the "Open File" dialogue.

17. Select [Density • Contrast (C)] in the [Display (V)] menu.


"Contrast" screen will be shown.

18. Select [Profile(Horizontal) (S)] in the [Analysis (A)] menu."Horizontal


Profile" screen will be shown.

3-3
3.1 Checking the Image

19. Adjust the slider bar so that the difference between "Black" and "White" on
the "Contrast" screen becomes approximately 400 steps.

20. Drag the slider of the "Center" to right and left so that the unevenness of the
image becomes most evident.

21. Using a yellow cross cursor, select the entire image tracing its border.

22. Horizontal profile of the area that is enclosed by the cross cursor will be
displayed on the "Horizontal Profile" screen.

3-4
3.1 Checking the Image

23. Check that the horizontal profile is almost flat.


If the image correction is not appropriate, this horizontal profile curve
will show a mountain shape.
ex) Profile with no correction.

24. Select [Exit (X)] in the [File (F)] menu.


Closing all windows, and returns to the "Service Tool" screen.

25. In a same manner, check the image read by the rest of all REGIUS 110s.
If a steep slope or winding is found in the horizontal profile, it may
mean there is an abnormality on the REGIUS 110. Take necessary
action by referring to the REGIUS 110 Service Manual.

3-5
3.2 Backing Up the Setup Data

3.2 Backing Up the After completing the image check of the REGIUS 110, back up the setup
data following the procedures below, and complete the installations.
Setup Data

3.2.1 Backing Up the Upload and save to the CS-1/2/3 the settings and correction data
saved in the connected REGIUS 110.
Reader Setup Data

1. Using a mouse or touch panel, open the "REGIUS Service Screen".


2. Click [Service Tool].
"Service Tool" screen will be shown.

3. Click [Reader].
"Reader" screen of the Service Tool will be shown.

4. Click [PCB].
"Reader Device" screen will be shown.

5. Select the REGIUS 110 whose setup data shall be backed up.
<Important>[Restore] -> [CS-1 -> REGIUS 170] button is disabled
if the setup data is not backed up on the CS-1/2/3.

3-6
3.2 Backing Up the Setup Data

6. Click [REGIUS 110 -> CS-1/2/3] in the [PCB] screen.


The setup data will be uploaded to the CS-1/2/3 from the REGIUS 110.

7. Touch [Exit].
The PCB screen closes, and switches to the "Reader" screen.

Repeat from Step 3 in 「3.2.1 Backing Up the Reader Setup Data」, and
back up the setup data on all REGIUS 110s on the CS-1/2/3.

3.2.2 Back Up of CS-1/2/3 <Important>Using the floppy disk on which a specific CS-1/2/3
Setup Data has been backed up, you can restore the same set ups
on the other CS-1/2/3. However, note that an error will
occur if the restored CS-1/2/3 is interfaced to the same
network as the original because the setup data unique to
the original CS-1/2/3 are also restored on the other CS-1/
2/3. If such happens, disconnect the restored CS-1/2/3
from the network, and set up the CS-1/2/3 uniquely from
others.

Note that if the license file is restored on another CS-1/2/


3, the first CS-1/2/3 will not start up.

1. Click [Backup] of the "Service Tool" screen.


"Backup" screen of the Service Tool will be shown.

2. Click [System].
"Save System Setup File" screen will be shown.

3-7
3.2 Backing Up the Setup Data

3. Click [>>].
4. Click [Save].
A message prompting an insertion of the disk into drive "A" will be
shown.

5. Insert a floppy disk into the floppy disk drive.


"Browse for Folder" dialogue will be shown.

6. Check that "3.5 Inch FD (A:)” is selected, click [OK].


Confirmation screen will be shown.

7. Click [Yes].
Saving of the CS-1/2/3 setup data will start.
Progress will be indicated on "Saving ...." message while saving
continues.
Upon completion of saving, a message to inform completion will be
shown.

8. Click [Yes].
Returns to "Save System Set Up File" screen.

9. Click [Exit].
Confirmation dialogue will be shown.

3-8
3.2 Backing Up the Setup Data

10. Click [Yes].


Returns to the "Service Tool" screen.
Implementing the above procedures will complete the back up for one
CS-1/2/3.
When several CS-1/2/3s are set up, use a different floppy disk to
implement the back up of setup data for each CS-1/2/3.

3-9
3.2 Backing Up the Setup Data

3 - 10
Chap.4
Adjusting the REGIUS 110

This section describes how to adjust the REGIUS 110.


Blank Page
4.1 Checking and Adjusting the Image Size

4.1 Checking and Check the pixels in the main scanning direction of the REGIUS 110, and
adjust PLL setting if the value is out of the standard value.
Adjusting the Image
Size

4.1.1 Checking Image Size, Reading the cassette on the REGIUS 110, which is exposed to produce
solid density image, and measure the pixels in the main scanning
Read Start Position
direction.

<Important>For details of how to use the Examination Registra-


tion screen and Image Confirmation screen, refer to the
CS-1/2/3 Instruction Manual.

1. On the CS-1/2/3 Operation Unit, input a temporary patient information for


checking the image, then change the screen to the “Image Confirmation”
screen.

2. Touch the square (Other) in the lower left corner of the body diagram and
touch to select [TEST].

3. Expose the solid density image on the test cassette.


• Bulb Voltage - 80kV, 10mAs / Distance - approximately 2m
• Align the long side of the cassette with the vertical direction of the
X-ray tube.

2m

Vertical direction Position the test cassette Long side of


of X-ray tube to allow the blue line to the cassette
lie on the bottom

4. If the system is set to "Barcode Registration", read the barcode of the cassette
that was exposed.
"Barcode Registration": the bar code of the cassette is displayed in the order.
"Manual Registration" : "1st" is indicated.
"Auto-registration" : Nothing on the screen.

4-1
4.1 Checking and Adjusting the Image Size

5. Insert the cassette into the REGIUS 110.


As the image in the cassette is read, the image is displayed on the
screen of operation unit.

6. Touch [OK].
7. Touch [Exam completed].
8. Touch [KONICAMINOLTA] and start Service Tool from the system menu or
open the REGIUS Service screen and start the Service Tool.
The Service Tool (Console) screen will be shown.

9. Touch [Analyze].
"Analyze" screen will be shown.

10. Click [Image].


Image-check software will start.

4-2
4.1 Checking and Adjusting the Image Size

11. Select [Open (O)] in the [File (F)] menu.


[Open] dialog will be shown.

12. Open the image data read in step 5.


13. Select [Display (V)] -> [View Size (V)] -> [No Pixel Skip (N)]. Then using
mouse, read the X-coordinate of the right and left edge and the Y-coordinate
of the top and bottom edge of the plate to measure the number of fluorescent
pixels(image pixels) in the horizontal direction (main-scan) of the image.

14. Check that the measured pixels fall within the allowance (Hor. and Ver. direc-
tion(sub-scan): 2022pixels ± 10pixels/Ver. direction: 2460 ± 12pixels, 2022
± 12pixels for 14x17in ).

15. Select [View Size (V)] in the [Display (V)] menu, and select [No Pixel Skip
(N)] in the submenu.

16. Using mouse, check that the margin of H_sync (the left side of image) on the
plate falls within the allowance (33 ± 10 pixels) and the margin on the top
falls within the allowance (33 ± 10 pixels).

17. Exit "Image Check" tool.


Returns to the Service Tool (Analyze) screen.
Fluorescent Pixel Size or pixels in the margin is out of the allowance, adjust
the PLL, reader start position by the following procedure.
4.1.2 Adjusting the Image
Size and Read Start
Position

1. Click [Reader] of the Service Tool screen.


The Service Tool (Reader) screen will be shown.

4-3
4.1 Checking and Adjusting the Image Size

2. Click [Device] of [System].


"Reader Device" screen will be shown.

3. Select the REGIUS 110 on which the image was read, and click [OK].
"Device Setup" screen will be shown.

To Change Pixels

1. Change the setting of "Image Size (PLL)".


Measured Pixels Setting of Horizontal Magnification

Larger than 2022 ± 10 pixels Decrease to bring within the allowance.

Smaller than 2022 ± 10 pixels Increase to bring within the allowance.

Measured Pixels Setting of Vertical Magnification

Larger than 2460 ± 12 ( 2022 Decrease to bring within the allowance.


± 10 for 11x14in) pixels

Smaller than 2460 ± 12 ( 2022 Increase to bring within the allowance.


± 10 for 11x14in) pixels
Target: Fluolescent pixels will be increased by “1” when horizontal size is
changed by “+20” or vertical size is changed by “+2”.

4-4
4.1 Checking and Adjusting the Image Size

To Change the Image Position

1. Input the amount of the distance which the image would like to be shifted to
“Hor.Shift”. (input a positive/negative value)
Inputting a positive value moves the image to the right, while a negative value
to the left.

2. Input the amount of the distance which the image would like to be shifted to
“Ver.Shift”. (input a positive/negative value)
If inputting a positive value, the image goes upward and if inputting a nega-
tive value, the image goes downward.

3. Click [Send].
Revised value will be sent to the REGIUS 110.

4. Click [Exit].
Returns to "Service Tool (Reader)" screen.

4-5
4.2 Unevenness Calibration of Regular Cassette

4.2 Unevenness Cali- If the optical unit of REGIUS 110 is replaced or used the REGIUS Plate
with new version, implement the unevenness calibration by the following
bration of Regular procedure.
Cassette
<Important>Use the largest cassette that is used in the site for the
unevenness calibration. The correction value will be
invalid for the sizes larger than the one used for the cali-
bration.

1. Expose the solid density image on the cassette for the unevenness calibration
to the following conditions.

• Exposure Condition
Volt : 80kV (fixed)

mAs value : Set the X-ray dose so that the maximum data value falls
within 1500 ~ 3000 Step. (target : 10 to 20mAs at 2m)

Plate - Bulb distance : 2m or more (recommended)

• Align the long side of the cassette with the vertical direction of the
X-ray tube.

2m

Vertical direction Position the test cassette Long side of


of X-ray tube to allow the blue line to the cassette
lie on the bottom

2. Click [Reader] of the "Service Tool".


The Service Tool (Reader) screen will be shown.

4-6
4.2 Unevenness Calibration of Regular Cassette

3. Click [Uneven. Cal.].


"Reader Device" screen will be shown.

4. Select the cassette reader on which the unevenness should be calibrated, and
click [OK].
• "Unevenness Calibration" screen will be shown.

5. Select the resolution to be calibrated in "Resolution".


Select "High (Auto)" (87.5µm) or "Standard" (175µm).

6. If the system is set to "Bar Code Registration", read the bar code on the
cassette using a bar code reader. The bar code data read by the bar code reader
will be displayed on the "Bar Code Registration".
• If the system is set to “Manual Registration”, it is not necessary to
read the bar code.

7. Click [OK].
A dialog indicating the progress of the Unevenness Calibration will be
shown.

4-7
4.2 Unevenness Calibration of Regular Cassette

8. Insert the cassette into the REGIUS 110.


<Important>Image will be read from the cassette, and Uneven-
ness Calibration will be implemented when the dis-
charged cassette is removed. Note that the calibration
will not start until the discharged cassette is removed.

Upon completion of the Unevenness Calibration, calibrated result will


be shown in the "Result". Check that the “Result” is “OK”.

9. Click [Exit].
Returns to the Service Tool (Reader) screen.
The result of the unevenness calibration will be saved in the
unevenness calibration table of the CS-1/2/3.

• For the Unevenness Calibration in normal case, implement the both


resolution of "General High" (87.5µm) and "General Standard"
(175µm).

4-8
4.3 Sensitivity Calibration of Regular Cassette

4.3 Sensitivity Calibra- Implement the sensitivity calibration if the X-ray generator device was
changed or calibrating the sensitivity difference on each REGIUS 110.
tion of Regular Cas-
sette <Important>Use the largest cassette that is used in the site for the
sensitivity calibration.

1. Using the condition shown below, expose a solid density image on the cassette
for the unevenness correction.

• Exposure Condition
Voltage: 80kV (fixed)

Standard X-ray Dose : 10m +/-2

Grid : None

• Align the long side of the cassette with the vertical direction of the
non-grid X-ray tube.

2. Click [Reader] of the "Service Tool".


Service Tool (Reader) screen will be shown.

4-9
4.3 Sensitivity Calibration of Regular Cassette

3. Click [Sensitivity].
"Reader Device" screen will be shown.

4. Select the cassette reader on which the unevenness should be calibrated, and
click [OK].
• "Sensitivity Calibration" screen will be shown.

5. Tick the "Automatic Calculation", and select "87.5μm(General High)" from


resolution.
• Ticking and selecting the "Automatic Calculation" will automatically
create calibration data of "General Standard" using the calibration
data of "General High" (this means calibrations will be completed by
single exposure).

6. Input the X-ray dose that was used for the exposure in "Exposure Dose".
7. If the system is set to "Bar Code Registration", read the bar code on the
cassette using a bar code reader. The bar code data read by the bar code reader
will be displayed on the "Bar Code Registration".
• If the system is set to “Manual Registration”, it is not necessary to
read the bar code

4 - 10
4.3 Sensitivity Calibration of Regular Cassette

8. Click [OK].
A dialog indicating the progress of the sensitivity calibration will be
shown.

9. Insert the cassette into the REGIUS 110.


As the image is read from the cassette, sensitivity calibration is
implemented.
Upon completion of the sensitivity calibration, the corrected result will
be shown in the "Result".

10. Click [Exit].


Returns to the Service Tool (Reader) screen.
The result of the sensitivity calibration will be saved in the sensitivity
calibration table of the CS-1/2/3.

<Important>Calibration will not be started before the discharged


cassette is removed.

4 - 11
4.3 Sensitivity Calibration of Regular Cassette

4 - 12
Chap.5 Maintenance &
Adjustment
This chapter describes procedures for maintenance and
adjustment of the CS-1/2/3.
Blank Page
5.1 Upgrading the Version of REGIUS 110

5.1 Upgrading the Ver- Update the firmware (program and data stored in CF) of the cassette
reader.
sion of REGIUS 110
To update the software version of the REGIUS 110, use the "Service
Tool" of the CS-1/2/3 that is connected to the REGIUS 110.

<Important> REGIUS 110 might not be started up if making a


wrong operation. Therefore, strictly follow the procedure
below.
Once the procedure for upgrade has started on the CS-1/
2/3, never implement other operation. Also, during the
upgrade procedure, never turn off the power of the cas-
sette reader.
In case that the REGIUS 110 is unable to be started up, it
is required to replace the CF card of REGIUS 110.

<CAUTION>When updating the software version of the REGIUS


110 in "n to m connection", never operate the cassette
reader(s) other than the targeted one as well as the CS-1/
2/3 for the updating operation.

1. Restart the power (OFF/ON) of all devices connected to the same network.
Shut down the backup JM CS-1/2/3
(not required for a 1-to-1 configuration)

Shut down the standard JM CS-1/2/3

Turn ON the standard JM CS-1/2/3

Turn ON the backup JM CS-1/2/3
(not required for a 1-to-1 configuration)

Turn ON all REGIUS 110s.

2. Start up the "Service Tool" at the REGIUS Service screen.


The “Service Tool (Reader)” screen will be shown.

5-1
5.1 Upgrading the Version of REGIUS 110

3. Click [Reader].
The “Service Tool (Reader)” screen will be shown.

4. Click [File Import] of [Program]


5. Select the REGIUS 110 that should be upgraded.
"Install File" screen will be shown.

6. Insert the floppy disk that contains a new software version for REGIUS 110
into the floppy disk drive of the CS-1/2/3.

7. Click [ • • • ].
"Open" dialog will be shown.。

5-2
5.1 Upgrading the Version of REGIUS 110

8. Select the new program version of REGIUS 110 Install File, and click
[Open (O)].

Returns to the “Install File” screen, and the selected file name will be
shown on the “File Name”.

• The file name of REGIUS 110 Install software is ;


"110ProgRxTx.lzh" (x varies depending on the version)

9. Click [OK] of the [Install File].


A dialog indicating the installation process will be shown.
Upon completion of the installation, the dialog closes, and returns to
the Service Tool (Reader) screen.

10. Click [Upgrade] of [Program].


"Reader Device" screen will be shown.

11. Select REGIUS 110 that should be upgraded, and click [OK].
"Controller Program" screen will be shown.

5-3
5.1 Upgrading the Version of REGIUS 110

12. Select the latest version, and click [OK].


A dialog indicating the updating process of the program will be shown.

• It takes 3 to 4 minutes to complete the program change.


<Important>Never touch the "Cancel" on this dialog. Do not
attempt to implement any operation on the CS-1/2/3.

<Important>Never turn off the power of the REGIUS 110.

Upon completion of updating, the dialog indicating the progress closes,


and a dialog to indicate the completion of updating will be shown.

13. Click [Yes].


Returns to the “Service Tool (Reader)” screen.

14. Restart the REGIUS 110 which has been updated.


When "READY" appears on the LCD of the REGIUS 110 after
completing the restart, upgrading is successful.
The program version currently running on the REGIUS 110 can be
verified on the "Version" of the "Controller Program" screen. (refer to
step 9 through 10 to open the "Controller Program" screen)

15. Start up the "PostgreSQL access" of the JM, and confirm on the "r_version"
screen that the version and updated data of the REGIUS 110 have been
updated.

When several REGIUS 110s are connected to the system, upgrade


each REGIUS 110 in a same manner.

5-4
Chap.6
Service Tool Screens
This chapter describes the "Service Tool" screens that
are used to interface with REGIUS 110.
Refer to Installation/Service manual of each console for
the rest of screens.
Blank Page
6.1 Overview of Service Tool

6.1 Overview of Service Tool


The below shows the changes from the Service Tool for REGIUS 170/190.
(Classification: Describes the classification of [New/Changed] from REGIUS
170/190)

Tool Class. Remarks Ref.

[Console] > [System Config.] > [System Changed Registration method of cassette reader was changed from p.6-2
Config.(Reader)] selection of availability of the device by checkbox to pull-
down menu("170/190" or "110" or "No Registration").

[Console] > [System Config.] > [Remote Changed REGIUS 110 was added to the reader selection. p.6-3
Maint.]

[Reader] > [Reader] > Changed Reader type denotation was added to the reader selection p.6-4
[Cassette Reader] field.

[Reader] > [Device Set] New Tools for REGIUS 110 was separately created. p.6-5

[Reader] > [PCB] Changed Resolution selection was changed due to nonsupport of p.6-8
mammogram.

[Reader] > [Std.Curr.] Changed Resolution selection was changed due to nonsupport of p.6-9
mammogram.

[Reader] > [Unevenness] Changed Resolution selection was changed due to nonsupport of p.6-10
mammogram.

[Reader] > [Sensitivity] Changed Resolution selection was changed due to nonsupport of p.6-12
mammogram.

[Reader] > [Sublog] Unchanged Tools for REGIUS 170/190 only. On REGIUS 110, have all -
tools in [Reader] -> [Reader Log].

[Reader] > [Reader Log] Changed Supports REGIUS 110. p.6-14

[Reader] > [Mech. Counter] New Supports REGIUS 110/170/190. p.6-15

[Reader] > [Install] Changed Supports REGIUS 110. p.6-16

[Reader] > [Program] Changed Supports REGIUS 110. p.6-16

[Reader] > [Network] Changed Supports REGIUS 110. p.6-17

[Reader] > [Unit Test] New Tools for REGIUS 110 only. -


Refer to REGIUS 110 Service Manual for operation.
With REGIUS 170/190, test shall be made by connecting to
maintenance PC. No test can be performed from CS-1/2/3.

6-1
6.2 [Console] > [System Info.] Screen

6.2 [Console] > [System Info.] Screen

6.2.1 "SYSTEM INFO" Screen

Key/Item Function
[Save & Exit] Click to exit from "System Info." screen after saving the settings.
[Cancel] Click to exit from "System info." screen without saving the settings.

6.2.2 "SYSTEM INFO • System Config.(Reader)" Screen


Using this screen, register the REGIUS 110 in the system or set the names used in the system.
Related Setting ;Section 6.3, " [Reader] > [Reader] > "Dedicated Reader""

Start up from the system menu Start up from the REGIUS Service Screen

Display Procedure • • • • [ [Console] --> [System Info.] --> (Select [System (Reader)] in the left menu.)

Key/Item Function
In the case of CS-1/3, items for upright/table readers will be displayed above the REGIUS 110.
REGIUS 110
Model Name Input a name which is for specifying the several REGIUS to identify in collective within 1
byte x 10 characters.
Reader 1 ~ 2 Check "Registered" for the reader device to be used in the system.
Disp. Name Input the name used in the system with 1 byte x 10 characters.
<Important>Default name of the reader devices displayed in the user screens is set to "REGIUS 170".
When connect to the REGIUS, the displayed name can be changed by inputting the data in the
“Disp.Name” as above screen.

6-2
6.2 [Console] > [System Info.] Screen

6.2.3 "SYSTEM INFO • Remote Maintenance" Screen


Implement the settings used in remote maintenance.

Start up from the system menu Start up from the REGIUS Service Screen

Display Procedure • • • • [Console] --> [System Info.] --> Select "Remote Maintnance" in the left menu.

Key/Item Function
Remote Maintenance
IA Auto-Start Tick to enable "Remote Maintenance" function.
Ccu Maint. Info. Acquisition
Setting
Ccu1 Tick to acquire the maintenance information of REGIUS 110/170/190 that has been registered as
"Reader #1".
Select the reader type from the pull-down menu when ticked.
Select the reader type from REGIUS 110/REGIUS 170/REGIUS 190.
Ccu2 Tick to acquire the maintenance information of REGIUS 110/170/190 that has been registered as
"Reader #2".
Select the reader type from the pull-down menu when ticked.
Select the reader type from REGIUS 110/REGIUS 170/REGIUS 190.

6-3
6.3 [Reader] > [Reader] > "Dedicated Reader"

6.3 [Reader] > [Reader] > "Dedicated Reader"

6.3.1 Common Displays of "CCU INFO" Screen

Key/Item Function
Reader Device Selection (upper left Select the reader device name to be set if there are several reader devices con-
menu) nected.
[Save & Exit] Click this button to exit from "CCU Info" screen after saving the setting.
[Cancel] Click this button to exit from "CCU Info" screen without saving the setting.

6.3.2 "CCU INFO • TCP/IP" Screen

Start up from the system Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Reader] --> select "TCP/IP" in the left menu

Key/Item Function
TCP/IP
Host Name (Hosts) Input the host name of REGIUS 110 using 1 byte x less than 32 characters. (default:
R170-0001)
Host Name (JM) Input the host name of REGIUS 110, which the JM uses, using 1 byte x less than 32
characters.
(default: r170-0001)
IP Address Input the REGIUS 110’s IP address. (default: 192.168.20.170)
Port Number (Reader) Input the REGIUS 110’s port No. (default: 8018)
Port Number (Console) Input the CS-1/2/3’s port No. (default: 8018)
Timeout 1 Do not change unless otherwise instructed. (default: 180000)

6-4
6.4 [Reader] > [Device ]

6.4 [Reader] > [Device ]

6.4.1 "Device" Screen (for REGIUS 110)

Start up from the system menu Start up from the REGIUS Service Screen

Display Procedure • • • [Reader] --> [Device ] --> select the reader name from the "Reader Device" screen --> [OK]

Key/Item Function
PLL(image size)
H-Magnification Set the horizontal magnification of the image size.
Default: Adjustment is made on each device (each serial No.) before shipment.
Pixel number of fluorescence substance should be within -5 to +10 pixels(4041 to 4056
pixels) of 4046 pixels.
V-Magnification Set the vertical magnification of the image size.
Default: Adjustment is made on each device (each serial No.) before shipment.
Pixel number of fluorescence substance should be within 2046 ± 10 pixels(2036 to
2056) (2022 ± 10pixels for 11/14in).
Reader Start Point
H-Shift Set the margin on the left side of the image (H-Sync side of the plate).
Input + value to increase the margin and - value to reduce the margin that the margin is
to be within 33 ± 10 pixels.
Default: 0
V-Shift Set the margin on the bottom side of the image.
Input + value to increase the margin and - value to reduce the margin that the margin is
to be within 33 ± 10 pixels.
Default: 0
[Send] Sends the set up value to REGIUS 110.
[Table Refer...] Reads and displays the data stored in SIU (CF) of REGIUS 110.

6-5
6.4 [Reader] > [Device ]

6.4.2 "Table Management (for REGIUS 110)" Screen


Screen to allow management of the table.

Start up from the system menu Start up from the REGIUS Service Screen

Display Procedure • • • [Reader] --> [Device ] --> select the name of REGIUS 100 from the "Reader Device" screen --> [OK]

Key/Item Function
[Back] Closes the "Table Management" screen, and returns to "Device Setup" screen.
Table Displays the value of PLL, H-Starting Point (87.5μm), H-Starting Point (175μm) and V-Start-
ing Point.
[Output] Outputs the table values in text format (.txt) file.
Default file name: Device name_y/m/d_CcuPllRet.txt」

How to calculate the image size 1. Expose the solid density image on a 14"x17" or 14"x14" cassette by High
Res. mode, and measure the size of fluorescence substance by using an
inspection tool.
2. Change the image size to be within the allowable range of fluorescence sub-
stance(Hor. Magnification: 4046 pixels -5 to +10 pixels/Ver. Magnifica-
tion: 4920 pixels -5 to +10 pixels).

6-6
6.4 [Reader] > [Device ]

Image Position Adjustment 1. Expose the 14"x17" cassette using High Res. mode, and measure the
margin on H-sync side of the plate using an inspection tool.
2. Set the H-Shift and V-Shift values to be within the
allowable range of the margin(30 to 45 pixels).

<Important>Always implement "Unevenness Calibration" when


the setting of image size and/or image position has been
changed.

6-7
6.5 [Reader] > [PCB]

6.5 [Reader] > [PCB]

6.5.1 "REGIUS 170 PCB" Screen

Start up from the system Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [PCB] --> (select the reader from the "Reader Device" screen)

Key/Item Function
Backup The stored data in CIU(CF) of the REGIUS 110 will be uploaded to CS-1/2/3
after clicking “Console -> REGIUS 170/190” and saved.
Restore The stored data in CS-1/2/3 will be uploaded to CIU(CF) of the REGIUS 110
after clicking “REGIUS 170/190 -> Console” and saved.
• "Program replacement for REGIUS170/190" is not available on REGIUS 110.
<Important>"Program replacement for REGIUS170/190" button is
invalid if the backup data is not stored in CS-1/2/3.
[Exit] Exit the "Change Board" screen.
• Data to be stored
• Shading data
• Sensitivity calibration data(settings made on "Sensitivity Calibration" screen)
• PLL adjusting value(settings made on "Device Setup" screen)
• Image position(settings made on "Device" screen)
• PMT Gain Table (settings made on "Unevenness Calibration" screen)
• Various setting data for network
• Mechanical counter info.

6-8
6.6 [Reader] > [Std. Current]

6.6 [Reader] > [Std. Current]

6.6.1 "Standard Current" Screen

Start up from the system Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Std. Current] --> (select the reader from the "Reader Device" screen --> [OK])

Key/Item Function
Reading Condition
Image Size Select the image size from the following when reading Standard Current.
• "175µm(General Std)", "87.5µm(Gen. H/Man. Std)".
• REGIUS 110 does not support Mammogram exposure.
Standard Current Select the current value for dummy reading from the following.
• "0nA", "50nA", "500nA", "50000nA", "All".
Unevenness Select the unevenness calibration for dummy reading from the following.
• "ON", "OFF".
[Delete All Test] 1) Clicking on this button will display a dialogue confirming the deletion of the dummy
image.
2) Clicking "Yes" will delete only the dummy images at one time.
[Execution] Execute the dummy reading with the standard current.
Implementing [Execution] will create “(Device Name)_CcuTest.img” file is created in the
following folder.
For CS-1 : E:\Image
For CS-2 : C:\KonicaMinolta\CS-2\Image
For CS-3 : C:\KonicaMinolta\CS-3\Image
After checking the resulted image, delete the image by implementing [Delete All Test
Images].
[Exit] Switches to "Reader" screen without implementing the dummy reading.

6-9
6.7 [Reader] > [Uneven. Cal.]

6.7 [Reader] > [Uneven. Cal.]

6.7.1 "Unevenness Calibration" Screen

Start up from the system Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Uneven. Cal.] --> (select the reader from the "Reader Device" screen --> [OK])

Key/Item Function
Calibration
Resolution Select the pixel size for calibration from "175µm(General Std)", "87.5µm(Gen. H)".
• REGIUS 110 does not support Mammogram examination.
PMT Use the default (2200)
Barcode Registration Use this function when the bar code registration is employed.
[OK] Starts the unevenness calibration.
Create Shipping Table Not to be used at the user. (do not tick the check).
(Factory use only)
Result Displays the result of the unevenness calibration.
[Table Control] Displays the information of the unevenness calibration.
• Shows the list of Plate Versions and Cassette Sizes that have been used for calibration in the
past.
[Text File] Exports the result of the unevenness calibration as a text file(.txt).
[Exit] Exits the unevenness calibration.
• Correction data
10 tables of correction data for each of "87.5µm", "175µm" per Pv (plate version) can be registered. (total 10 tables).

• Plate size
For calibration purpose, use the largest size that is used at the site. (except mammography use). For the sizes larger than the
one used for calibration, the calibrated data is no more valid.

6 - 10
6.7 [Reader] > [Uneven. Cal.]

As an example, the case where the calibration is made using Pv00 + 14" x 17", Pv01 + 10" x 12" is described in the follow-
ing table.

Read Plate Size Plate Version Calibration Data to be applied


8" x 10" Pv01 Pv01+ 10" x 12"
14" x 17" Pv01 Pv00 + 14" x 17"
10" x 12" Pv00 Pv00 + 14" x 17"

• Notes when implementing the unevenness calibration.


Use the X-ray dose with bulb voltage 80kV, and mAs value to generates the signal value within the range of
1500 ~ 3000STEP. Distance between the plate and bulb should be 2m or more.

6 - 11
6.8 [Reader] > [Sensitivity]

6.8 [Reader] > [Sensitivity]

6.8.1 "Sensitivity Calibration" Screen

Start up from the sytem Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Sensitivity] --> (select the reader from the "Reader Device" screen --> [OK])

Key/Item Function
Calibration
Automatic calculation from Selecting this option enables to automatically create the calibration data for low
high resolution data resolution using calibration data for high resolution.
Resolution When "Automatic calculation from high resolution data" is not selected, select the pixel
size for sensitivity calibration.
• For regular cassette, select from the "175µm (standard)", "87.5µm (high resolution mode
A)" .
When "Automatic calculation from high resolution data" is selected, select the pixel size
for sensitivity calibration.
• For regular cassette, select "87.5µm (high resolution mode A)"
Exposure Dose Input the irradiation dose.
• 8 ~ 20mR (10mR or higher is recommended)
Sensitive Input "0.94" for regular cassette.
• In "Auto Create" mode, sensitivity data for low resolution will be calculated using these set-
tings.
Barcode Registration Use only for "Bar Code Registration" is selected.
[OK] Executes the sensitivity calibration.
Create Shipping Table Not to be used at the user. (do not tick the check).
(Manufacturer Use Only)
Low sensitivity QR value of the low sensitivity region that shall be calibrated. (default : 125)
Standard sensitivity QR value of the normal sensitivity region that shall be calibrated. (default : 250)
High sensitivity QR value of the high sensitivity region that shall be calibrated. (default : 500)

6 - 12
6.8 [Reader] > [Sensitivity]

Key/Item Function
Result Displays the result of sensitivity calibration.
[Table Control] Displays the correction data for the reader device.
• Lists all history of calibrations with Plate version (Pv), size, cassette in the past.
[Text File] Exports the result of calibration result as a file.
[Exit] Exits the sensitivity calibration.

• Sensitivity calibration
• Regular (87.5µm, 175µm) cassette will be automatically recognized by reading the barcode on the cassette.
• Selecting "Automatic calculation from high resolution data" will implement the calibration by reading at high resolution,
and automatically create the sensitivity data for low resolution. (calibration completes with one exposure)
• Use 14" x 17" film for regular (87.5µm, 175µm) . (use the largest size that is actually used at the institution)
• Standard Exposure Dose : 10mR or higher is recommended (use the exposure dose at which the x-ray irradiation
becomes stable)

6 - 13
6.9 [Reader] > [Reader Log]

6.9 [Reader] > [Reader Log]

6.9.1 "Reader Log" Screen

<Important>Implementing this function will download the CIU Logs of the REGIUS 110.

Start up from the sytem Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Reader Log] --> (select the reader from the "Reader Device" screen)

Key/Item Function
Reader Select Select the REGIUS for acquiring a log when there are several reader devices which are
connected.
For REGIUS170/190, implement [Reader] -> [Create Sub-log] before implementing
[Reader Log].
[Create Sub-log] is not necessary for REGIUS 110.
[Reader Log] Receives the logs from the REGIUS 110.
[Exit] Exits the “Reader Log” screen.

6 - 14
6.10 [Reader] > [Mech. Counter]

6.10 [Reader] > [Mech. Counter]

6.10.1 "Mech Count" Screen

Start up from the sytem Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Mech. Counter]

Key/Item Function
Number of Process Shows the number of process of each cassette size fed by the Reader.
TTL :Total of "Read", "Erase (Quick)" and "Erase (Slow)".
Read :Number of process for cassette reading.
Erase(Quick):Number of process for cassette erasure(Erase Mode-1).
Erase(Slow):Number of process for cassette erasure(Erase Mode-1).
Halogen Lamp <Case of REGIUS 110>
Warning Settings :
Displays the maximum number lighting of Halogen lamp in terms of count of number of
lighting. The warning that is for replacing the halogen lamp will be displayed if the number
reaches this limits.

Number of Lighting :
Displays how many times the lamp is actually turned on in the past.

<Case of REGIUS 170/190 >(reference)


Warning Settings :
Displays the rest of usable life of Halogen lamp in terms of count of number of lighting for
the one still alive after one of two lamps has gone.

Number of Lighting :
Displays how many times the lamp is actually turned on in the past.
[File Output] Exports the Mech. Count Information as a text file(.txt).
[Back] Exit the "Mech. Counter" screen.

6 - 15
6.11 [Reader] > [File Import], [Upgrade]

6.11 [Reader] > [File Import], [Upgrade]

6.11.1 "Install File" Screen

Start up from the sytem Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [File Import]

Key/item Function
[...] (reference key) Select the holder where the install program resides. Normally select from "A" drive (floppy
disk drive).
[OK] The selected program will be decompressed.
[Cancel] Exit the "Install File" screen after cancelling the selection of the file.

6.11.2 "Reader Program" Screen


Start up from the sytem Start up from the REGIUS Service
menu Screen

Display Procedure • • • [Reader] --> [Upgrade] --> (select the reader from the "Reader Device" screen)

Key/Item Function
Version Indicates the program version which is currently running on the REGIUS 110.
The Current Installed Version list of the REGIUS 110’s programs stored in the CS-1/2/3. The program selected
here will be written in the reader device.
[OK] Execute the switch over to the program version listed in the reader program version list.
[Cancel] Exit the "Reader Program (change)" screen after cancelling the installation of program.

6 - 16
6.12 [Reader] > [Network]

6.12 [Reader] > [Network]

6.12.1 "Network" screen


Write the network information on the REGIUS 110 using this screen. Communicating with the REGIUS 110 in FTP mode
when the screen starts, displays the obtained REGIUS 110 information.

Start up from the sytem Start up from the REGIUS Service


menu Screen

Display Procedure • • • [Reader] --> [Network] --> (select the reader from the "Reader Device" screen --> [OK])

Name of currently
connected
REGIUS 110

Key/Item Function
Reader
Host Name Input the host name of the REGIUS 110. (default: r170-0001)
IP Address Input the IP address of the REGIUS 110. (default: 192.168.20.170)
Netmask Input the subnet mask of the REGIUS 110. (default: 255.255.255.0)
JM
Host Name Input the Host Name(jml-0001 fixed) of JM which controls this REGIUS 110.
Hosts
Host Name Displays the host name and IP address of CS-1/2/3s which are connected to this
REGIUS 110.
[New] Click this button when newly inputting the host name and IP address of the CS-1/2/3
which uses this REGIUS 110.

[Edit] Click this button to change the host name or IP address of the CS-1/2/3 or JM listed on
the "Hosts" window
[Delete] Deletes the CS-1/2/3 or JM from the "Hosts" window.
[Send] Send the displayed settings to the REGIUS 110 in FTP mode.
[Exit] Exit the "Network" screen. No change will be caused on the REGIUS 110.

6 - 17
6.12 [Reader] > [Network]

6 - 18