Professional Documents
Culture Documents
Vaim5975 Iss 24
Vaim5975 Iss 24
(QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 1 of 21
ELECTRICAL CONDUCTIVITY TESTING
1.0 General
1.1 This document covers the requirements for the temper inspection of aluminum alloys by
eddy current electrical conductivity measurement.
1.2 Modifications to this procedure affecting Lockheed-Martin parts shall be approved in writing
by LM Aero – Marietta Materials and Processes prior to implementation.
1.3 Definitions
2.0 Equipment
2.1 Instrument
2.1.1 Verimet M4900C (conductivity meter with a variable frequency of 50/60 kHz)
2.1.1.1 SN 1641 with probe 4862C
2.1.1.2 SN 2086 with probe 4623C
2.2.1 Common Conductivity Standards in the 30%, 40% & 50%IACS approximately.
2.2.3 A minimum of three (3) instrument conductivity standards shall be required for
testing production parts. One standard shall be within ±7% IACS of the
applicable acceptance range of the value to be measured and the other two (2)
shall bracket and be within ±25% IACS of the applicable acceptance range of
the value to be measured. A minimum of 3% IACS shall separate the
conductivity standards.
2.2.5 At least, one standard shall cover the 25 to 32% IACS range and one standard
shall cover the 38 to 62% IACS range.
2.2.7 Sanding, grinding or any method that may alter the metallurgical characteristics
of the standard is prohibited.
2.2.8 The instrument conductivity standards shall be certified at the same time the
time instrument is calibrated.
2.4 Instrument shall be capable of measuring the accuracy of any reference conductivity
standard within ±0.5%IACS of its certified value in the instrument’s certified range up to
62% IACS and within ±1.0% IACS of any reference standard value greater than 62%
IACS (applicable to bare material only and through non-conductive coatings up to 0.003”
thick).
2.6 Instrument shall be capable of determining the conductivity of aluminum alloys with an
accuracy of ±1.0%IACS or better through electrically nonconductive films and coatings
up to 0.003” thick.
2.7 The instrument shall be capable of operating linearly within the range of the instrument
standards
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 3 of 21
ELECTRICAL CONDUCTIVITY TESTING
3.0 References
Note: With reference to BAC 5946 and the three above common standard specifications, Vac
Aero has been classified as having a Type II test system with operating ranges of 18%
to 51% IACS and 18% to 60% IACS.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 4 of 21
ELECTRICAL CONDUCTIVITY TESTING
Calibration
Frequency
Frequency Written approval
OEM Specification reduction
Required required
(max)
(max)
Bombardier /
PPS 20.07 6 months Not allowed No
DeHavilland
No (objective
Mil-Std Mil-Std-1537 120 days 12 months evidence required
only)
4.2 Based on para 4.1, the conductivity meter, the probe and the standards shall be calibrated
in accordance with MIL-STD-1537 and BAC5651 as a system at least every 4 months
(120 days) by an approved and certified laboratory.
4.2.1 The calibration frequency may be reduced to a maximum of six (6) months only
if three (3) consecutive calibration results do not show any out-of-tolerance
condition.
4.3 Reference to Boeing type test system and operating range shall be on the calibration
sticker and certificate.
4.4.1 the conductivity meter lacks the capability of measuring with an accuracy of
1% IACS
4.4.2 the probe or probe cable is replaced
4.4.3 the instrument needs repair
4.4.4 instrument standards are damaged
4.4.5 the probe cable length is changed
4.5 Laboratory conductivity reference standards used by the certified calibration source shall
comply with MIL-STD-1537 and BAC5651 requirements.
4.5.1.1 The reference standards selected shall provide widely spaced data
points to verify the test instrument accuracy within the certified range.
4.5.1.2 One standard shall be within ±2.5% IACS of the low end of the range of
the certification and one standard shall be within ± 2.5% IACS of the
high end of the range of the certification.
4.5.2 The laboratory conductivity reference standards shall have an uncertainty equal
to or less than the requirements shown below (at 68°F ± 5°F)
4.6 The primary conductivity reference standards shall have an uncertainty equal to or less
than ±0.2%IACS or 0.7% of stated value, whichever is less. They shall be calibrated at
68°F ± 5°F.
4.7 Instrument conductivity standards (ref section 2.0) shall be calibrated by comparison
with laboratory conductivity reference standards and comply with MIL-STD-1537 and
BAC5651 for accuracy.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 6 of 21
ELECTRICAL CONDUCTIVITY TESTING
4.8 Conductivity reference standards (primary, laboratory and instrument) dimensions shall
be as follows:
4.8.1 A size large enough to preclude an edge effect response, unless corrected
values are provided for each test instrument, instrument frequency and
reference standard size.
4.9 Calibration shall include testing for accuracy, drift, linearity and lift-off capability. The
test instrument operates on the principle of maintaining a constant depth of penetration
by adapting the oscillator frequency to the conductivity of the part. The instrument shall
be calibrated to maintain a constant depth of penetration of 0.050”.
The maximum surface roughness of parts to be tested shall be RMS 165. The areas to
be tested shall be free of dirt, grease, oil, or other visible contaminants. If required, the
surface shall be wiped with a clean cloth soaked in a solvent, followed by wiping with a
dry clean cloth.
5.2 Thickness
5.2.1 When thin or clad material is tested for electrical conductivity, a correction factor
shall be used. Thin material limitation is different for each customer:
Bombardier (BAPS) & Boeing (BAC & DPS) Less than 0.064”
Bombardier - De Havilland (PPS) Less than 0.060”
Bell Helicopter Textron (BPS) Refer to appendix IV
SAE (AMS) & Lockheed-Martin (C) Less than 0.044”
5.2.2.1 For thin Boeing parts, correction tables shall be applied in accordance
with BSS7351 requirements table V through VIII, depending on material
condition. Corrected acceptance values shall be determined from the
tables as follows:
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 7 of 21
ELECTRICAL CONDUCTIVITY TESTING
5.2.3 Bell Helicopter (Textron) parts shall be tested with a back-up plate without any
other correction factor in accordance with BPS 4453 depending on the
thickness of the part and on the expected conductivity. The back-up plate shall
be a flat sheet of 5052 aluminum alloy in the O condition, 0.050" thick or
greater. Refer to Appendix IV to determine the minimum thickness of the part
for which conductivity may be measured without a back-up plate.
For the Verimet 4900C used at 60kHz, the minimum sample thickness without
use of the back-up plate are given below for specific conductivity.
Conductivity
10 15 20 25 30 35 40 45 50
(%IACS)
Minimum
thickness 0.087 0.072 0.062 0.055 0.050 0.047 0.044 0.042 0.039
(in)
5.2.4 Lockheed-Martin parts tested in accordance with AMS 2658 requirements and
parts heat treated in accordance with AMS 2770 do not include a correction
factor. Testing in accordance with those specifications or a specification that
doesn’t include correction factor shall be inspected and actual values shall be
corrected, as per Appendix III.
The minimum concave curvature of a surface to be tested shall be on a 10” radius (or
20” diameter). Otherwise refer to customer specification.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 8 of 21
ELECTRICAL CONDUCTIVITY TESTING
5.4.2 Refer to BPS 4453 for inspection criteria of 2024 tubing for Bell Helicopter parts.
5.4.3 Refer to BSS7351 tables IX for acceptance value of round tubing for Boeing
parts.
5.4.3.2 For part/material wall thickness equal to or greater than 0.063”, refer to
table IX for conductivity values corrected for surface curvature.
Corrected acceptance values shall be determined as follows:
d. For material curvatures that are between the diameters listed in the
tables, interpolate the corrected values from the next larger radius
and the next smaller radius corrected values in the table.
5.4.3.3 For part/material wall thickness less than 0.063”, refer to para 5.2.2.1 to
obtain conductivity values corrected for thin gauge materials (table V
through VIII). Then apply the surface curvature correction values of table
IX to the thin gauge corrected values. Corrected acceptance values
shall be determined as follows:
a. Find the minimum thin gauge (corrected) conductivity value in the left
hand column of table IX. Read across the column until you are at the
curvature diameter. Note this value as the corrected minimum
acceptance limit.
d. For material curvatures that are between the diameters listed in the
tables, interpolate the corrected values from the next larger radius
and the next smaller radius corrected values in the table.
5.5.1 No reading shall be taken with the edge of the probe closer than ¼” (0.25”) to
any edge of the surface under test.
The Verimet M4900C probe housing is 1¼” (1.25”) diameter and the probe
diameter is 0.438”.
a. a minimum distance of ¾” from the center of the test probe to the edge of the
part or from an adjacent metal part shall be maintained, which gives a local
test area with a minimum width of 1½” to maintain system accuracy.
b. a minimum distance of 1” from the center of the test probe to the sides of a
recess, cutout or slot in a part.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 10 of 21
ELECTRICAL CONDUCTIVITY TESTING
5.6.1 Conductivity measurements may be taken through thin surface finishes such as
anodize (except hard anodize), zinc chromate or other non-conductive coatings,
paints and primers provided these coatings do not exceed 0.003” in thickness.
The instrument lift-off limitation is 0.003” maximum for automatic compensation.
Beyond this limit a correction factor shall be applied.
5.6.3 For Boeing parts, the lift-off shall not exceed 0.005” to maintain system
accuracy.
For long narrow parts and parts with small surface areas, the areas to which the eddy
current test apply shall be as defined in customer specification (as in BSS7351) or if
not, in DPS 1.05-8.
5.8.2 For Boeing parts, the clad correction tables shall be used (BSS7351 tables VI,
VII and VIII).
5.8.3 For Bell Helicopter parts, acceptance limits shall be as shown below. For
annealed (O temper), non-clad values for conductivity apply.
The surface of the tested part shall be flat such that all of the probe face touches the
part.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 11 of 21
ELECTRICAL CONDUCTIVITY TESTING
5.10 In cases where coatings shall be removed prior testing for compliance, coating removal
shall be limited to the minimum area necessary for testing. The area shall be buffed
lightly with a scouring pad using a pneumatic tool until a constant conductivity value is
obtained. Thickness of part at the testing location shall be measured before and after
coating removal and recorded on the process sheet.
6.1 Immediately prior to performing inspections of parts, the lift-off effect shall be determined
using a 0.003” thick nonconductive flat shim (or calibration foil) Immediately prior to
performing inspections of parts, the lift-off effect shall be determined using a 0.003” thick
nonconductive flat shim (or calibration foil) and the conductivity meter calibration shall be
verified against at least three conductivity standard blocks having values close to the
tested parts. Two of the three test blocks chosen shall bracket the expected conductivity
of the part. Adjustment shall be within ±0.3% IACS (Vac Aero practice).
6.1.1 The results of the lift-off effect verification and all standardization readings shall
be recorded on form F5953-3. The instrument standardization procedure is
detailed in Appendix I.
6.2 During continuous inspection, the instrument shall be verified at intervals no greater
than 15 minutes. A final verification shall be performed upon completion of inspection
of parts (BPS 4453 para 4.1.1.2 requirement). Discrepancies of more than ±0.3%
IACS shall be reported to the quality assurance department.
6.3 If surface of conductivity standard blocks is damaged, the block must be retested prior
to be used again to ensure accurate calibration value.
7.1 The conductivity measuring instrument, the standards, and the test material shall be
maintained within 2°F of each other (ref. BAC7351). Parts shall be tested at room
temperature (i.e. 70°F ± 15°F). Avoid performing conductivity test in air draft.
7.2 Before testing for electrical conductivity after an aging cycle, thick aluminum parts of
¾” or more shall be left at room temperature for at least five (5) hours to provide
sufficient time for dissipating residual internal heat. Typical cases are forgings.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 12 of 21
ELECTRICAL CONDUCTIVITY TESTING
8.1 Do not rub or slide the probe tip against any surface.
8.2 Wipe the probe tip before applying it on a test block.
8.3 Do not bang the probe tip against surface being measured.
8.4 Maintain a firm consistent pressure while testing.
8.5 Keep the probe tip clean by wiping often with a clean cloth.
8.6 Do not over bend the probe wire.
8.7 Do not place parts or tray on the probe wire. Always keep it free.
9.1 Before starting any measurement it is imperative to understand and follow the above-
mentioned rules and instructions.
9.2 Place the meter switch in the COMP position and let it warm up for 15 minutes. Verify and
adjust the instrument as per Appendix I.
9.3 Place the probe flat and firmly against the test surface. Read the conductivity directly from
the digital display.
9.4 Sufficient readings shall be taken at locations evenly distributed over the surfaces of the
tested part.
9.6 Conductivity shall be measured at extremes of metal thickness (thickest and thinnest) on
parts with non-uniform thickness.
9.7 Measurements shall be performed on both major surfaces (upper and lower) of a plate at
areas near corners and center.
9.8 Readings shall be taken with the probe housing edge placed no closer than 0.375” (3/8”)
to any recess, holes and cutouts (refer to para 5.5 for specific customer requirements)
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 13 of 21
ELECTRICAL CONDUCTIVITY TESTING
9.9 Conductivity variations within the part or within a lot shall not be greater than 2% IACS. If
greater variation is measured, refer to the specific customer specification for final
disposition. Refer to para 11.4 for Messier-Dowty specific requirements.
9.9.1 For Boeing parts, when hardness inspection is not required, the maximum variation
of conductivity across all measurement areas is 2% IACS. When hardness and
conductivity are both required, the maximum variation in conductivity is 3% IACS.
9.10 Inspection readings shall always strictly respect the minimum and maximum
requirements. Readings round off to the nearest 0.5% IACS shall be used only when it
is not conflicting with the minimum or maximum requirements (e.g. a reading of 29.3%
rounded off to 29.5% would not be acceptable if the minimum requirement is 29.5%).
The same rule applies when reporting the reading summary results rounded off to the
nearest 0.5% IACS.
9.10.1 For Boeing parts, results may be rounded to the nearest 0.5% according to the
following:
d. Results in the range X.751 to Y.999 shall be rounded up to Y.0 (next whole
number)
Note: For common causes of problems and their symptoms, refer to Appendix II.
9.12 For Goodrich direct orders, conductivity test shall be conducted on a 2” grid spacing on
a minimum of one part from the center of the load and two parts from other areas of the
load.
Aluminum parts shall be 100% inspected unless otherwise specified by the customer.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 14 of 21
ELECTRICAL CONDUCTIVITY TESTING
11.1 Acceptance criteria shall be in accordance with the applicable customer specification
as defined on the process sheets.
11.2 In most cases, hardness shall also be verified to obtain a meaningful conductivity test (both
hardness and conductivity values are required to define a given heat-treat temper condition).
11.2.1 For Bell Helicopter parts, it is mandatory that a hardness or tensile test be performed
in accordance with BPS 4467, BPS 4139 or the applicable material specification when
conductivity does not fall within the limits of acceptability.
11.3 For Lockheed Martin parts, acceptance criteria for 7XXX series alloys shall be based on
AMS 2658 requirements.
11.4 For Messier-Dowty parts, the range of conductivity results between the thinnest and the
thickest sections shall be lower than 3.4% IACS within the same lot and lower than 1.7%
IACS for each part.
Qualified personnel shall be those who have succeeded written and practical proficiency tests.
Qualification of personnel shall be renewed every three year (Ref. VAIM 5978).
Complementary training may be obtained through the Bombardier qualification program.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 15 of 21
ELECTRICAL CONDUCTIVITY TESTING
13.0 Revision
Issue # Date Summary of Changes and Reason Required by Approved by
1 Feb 25 - 97 First issue M.Bergeron M.Bergeron
3 April 21 - 98 Add note at sec.3, Sec.4 refer to Boeing Type system per auditor. M.Bergeron M.Bergeron
4 Jan 4 - 99 Sec. 11 report results as reading range only, mean reading not req’d. M.Bergeron M.Bergeron
6 Apr 17 - 2000 Add at sec. 2.0 std of 20% for Westland Al-Li alloys. M.Bergeron M.Bergeron
Add BPS at sec 3.0
Add ref VAIM at 12.0
7 Apr 25 - 2001 Review 8.0 M.Bergeron M.Bergeron
Add back-up plate designation at 5.2
Add sec par at 7.0 (Bell Textron audit)
Par 6 add final checking
Para 9 j, add round off limitations
8 Apr 24 - 2002 Para 12 rewrite the text M.Bergeron M.Bergeron
Appendix I details the Verimet calibration procedure following Bell (Textron)
audit.
Para 4.0 Add calibration certify per ISO standard or equivalent per Parker
9 Sept 13 - 2003 M.Bergeron M.Bergeron
auditor.
Para 6.0 Add two test blocks shall bracket %IACS expected per Bell (Textron
10 Apr 16 - 2004 M.Bergeron M.Bergeron
Auditor)
11 Feb. 14 - 2005 Para 9.0 l) added to reflect Goodrich conductivity test procedure S. Halwani S. Halwani
12 Jan. 16 - 2006 Acceptable variation with each standard block changed to 0.3% S. Halwani S. Halwani
General review
Section 2 Mention two Verimet Conductivity meters
13 Sept 22 - 2007 Section 3 – Update list of references M. Ciarlo M. Ciarlo
Para 9 i) add max accepted variation within a lot
Quality assurance manager changed
Para 11.3 added to specify Lockheed Martin acceptance criteria on on electrical
14 Apr 4 - 2008 M. Ciarlo M. Ciarlo
conductivity for Al 7XXX series
General Review
15 Nov 15 - 2008 M. Ciarlo M. Ciarlo
Add new conductivity probe
Add Messier-Dowty requirements
16 Dec 20 - 2009 M. Ciarlo M. Ciarlo
Add instrument accuracy and sensitivity requirements
Para 1.2 added – procedure modifications shall be submitted to Lockheed-
Martin
17 Jan 25 – 2012 A. Bujeault M. Ciarlo
Modify par 5.2 to clarify use of correction factor and appendix III added
Modify para 11.3 to replace STP54-101 by AMS 2658
Add para 1.3,
Modify para 2.2.3, 2.2.4 & 2.4 to comply with Boeing requirements
Add para 2.2.6 to 2.2.8, 2.6 & 2.7
Add 4.4.1 to 4.4.5, 4.5.1.1, 4.5.1.2, 4.5.2, 4.8 & 4.9
18 Dec 1 – 2012 M. Ciarlo M. Ciarlo
Modify para 4.5.1& 4.6 to comply with Boeing requirements
Add 5.2.2.1 (a) to (e), 5.4.2, 5.4.3, 5.5.2, 5.6.2, 5.6.3 & 5.8.3
Modify para 5.2.3 to comply with Boeing requirements
Add 9.9.1 & 9.10.1 & appendix IV
19 Mar 4 -2013 Modify Para 4.0. -frequency changed to 120 days M. Ciarlo M. Ciarlo
20 Mar 14 - 2013 Modify Para 4.0. OEM calibration frequency added M. Ciarlo M. Ciarlo
21 Aug 24 - 2013 Add Verimet Instruction Manual as a reference document M. Ciarlo M. Ciarlo
APPENDIX I
STANDARDIZATION PROCEDURE FOR VERIMET M4900C CONDUCTIVITY METER
a) Ensure that probe is cleaned and that the probe connector is firmly inserted into the meter.
b) All conductivity standards shall be within 2°F (1°C) of the local room temperature where the
conductivity meter is to be used.
c) The meter may be operated from fully charged internal batteries or with the AC
adapter/charger connected to a 115 VAC outlet.
d) If batteries are weak, all displayed decimal points will lit and meter must not be used. When
this occurs, batteries shall be recharge prior to further use or the meter may be operated with
the AC adapter/charger connected.
e) Turn the NORM/OFF/COMP switch to the COMP position. The warm-up period required is
approximately 20 min. The NORM position shall not be used at any time.
f) Immediately prior to every standardization, the lift-off effect shall be determined as follows (use
form F5975-3):
Place the probe on one of the conductivity standards and record the reading obtained.
Place a 0.003” thick nonconducting calibration foil between the probe and the same
conductivity standard and record the reading obtained.
The difference in the two readings shall not exceed 0.5 %IACS. If the difference exceeds
0.5 %IACS, do not use the instrument and notify the quality assurance representative. The
instrument shall be quarantined and returned to the calibration facility for repair.
g) The meter shall be standardized immediately after determining the lift-off effect and at minimum
15-minute intervals during continuous operations using three certified standards as follows:
- Press the probe flat and firmly in the center of the low conductivity standard. Use the
compensation adjustment knob until the conductivity reading matches the IACS value
of the certified standard within 0.3% IACS.
- Repeat the operations with the high conductivity standard and with the middle
conductivity standard.
- Repeat the steps above until the Verimet readings are within ±0.3% IACS for each of
the three (3) conductivity standards.
Note: The Verimet M4900C accuracy is 0.38% IACS. However, practically Vac Aero has always
succeeded to set the Verimet within 0.3% IACS. A re-calibration shall be required when the
discrepancy is above 0.5% IACS.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 17 of 21
ELECTRICAL CONDUCTIVITY TESTING
APPENDIX II
COMMON CAUSES OF PROBLEMS IN CONDUCTIVITY AND THEIR SYMPTOMS
APPENDIX III
CORRECTION FACTOR FOR VERIMET M4900C
When this appendix is applicable as per para. 5.2, actual values of conductivity obtained shall be
divided by correction factor applicable. This factor is dependent of part thickness.
* Do not conductivity test parts under 0,015'' thick, as compensation (COMP) doesn't work at those
thicknesses.
** No correction factor required for thickness over 0,044''.
Those values are approximated values obtained from figure II of Verimet M4900C instruction manual.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 19 of 21
ELECTRICAL CONDUCTIVITY TESTING
APPENDIX IV
DEPTH OF PENETRATION VS CONDUCTIVITY (Ref BPS 4453)
Note: When the combination of a given material thickness and expected conductivity in %IACS lies
below the curve for the frequency used for the measurement (Verimet 4900C uses 60 kHz), a backup
plate shall be used.
VAC-AERO INTERNATIONAL INC. (QUEBEC DIVISION)
VAIM Procedure No 5975
Issue No 24
August 25, 2021
Page 20 of 21
ELECTRICAL CONDUCTIVITY TESTING
APPENDIX V
DEFINITIONS
Conductivity ()
Reciprocal of volume resistivity () or 1/. It is the relative value of conductivity of any nonferrous
metal, to the arbitrary value set for commercially pure copper and is usually expressed in percent
Annealed Copper Standard (percent IACS or %IACS), at 20°C by definition, where copper is 100
%IACS.
Measure of the ability of a material to conduct electricity
Eddy current
Electrical current induced in a conductor by means of time-varying electromagnetic magnetic field
Lift-Off Effect
Distance between the probe and the surface of the part, typically due to electrically non-conductive
films or coatings (primer, paints, etc.) above which the conductivity varies more than ±0.5% IACS,
or,
Effect observed in the signal of equipment due to a change of the magnetic coupling when the
distance between the part and the probe varies,
or,
The change in eddy current meter indication when the spacing between the surface probe and the
test object changes from an ‘in contact’ position to a position away from the surface of the test
object. The resulting gap between the probe and the object surface is filled with nonconductive air,
coatings or shims
Probe
Many turns of fine wire usually wound on a ferrite or laminated iron core in such a way as to produce
a varying electromagnetic field when energized by an alternating current, usually 60 kHz or above for
conductivity measurements
where is the standard penetration depth in inches, is the relative permeability (1.0 for
nonmagnetic materials), is the conductivity in %IACS and is the frequency in Hz