Professional Documents
Culture Documents
AVE772 User Manul - 1
AVE772 User Manul - 1
User’s Manual
Building B6, Luguyuyuan, No.27 Wenxuan Road, Changsha Hi-tech Industrial Development Zone, Changsha,
1
AVE-772 User's Manual
If this instrument is used in a manner differently than specified in this manual, the protection
provided by the equipment may be impaired.
Modification history
Version Date Modification
1.0 2016.12.01 First edition
1.1 2017.7.01 Updated for hardware modification
2
AVE-772 User's Manual
1 INTRODUCTION ............................................................................................................................5
1.1 Intended Use ........................................................................................................................5
1.2 How to use the User Manual................................................................................................5
1.3 Precautions and Warnings Symbols – Definitions ...............................................................5
1.3.1 Cautions .....................................................................................................................5
1.3.2 Warnings ....................................................................................................................5
1.3.3 Biological Warnings ...................................................................................................6
1.4 Symbols................................................................................................................................6
1.5 Contact Information ..............................................................................................................7
1.6 Limitation of Liability .............................................................................................................7
1.7 Use of Third Party Computer Products ................................................................................7
2 SYSTEM DESCRIPTION ...............................................................................................................9
2.1 Operation Theory .................................................................................................................9
2.2 Performance and Technical Index .......................................................................................9
2.3 System Components ..........................................................................................................10
2.3.1 System Appearance ................................................................................................10
2.3.2 AVE-772’s structure .................................................................................................12
2.4 Software Functions ............................................................................................................12
2.4.1 Main Programme Screen .........................................................................................13
2.4.1.1 Daily Operation Main Menu ...........................................................................14
2.4.1.2 Management Main Menu...............................................................................14
2.4.1.3 Quality Control Main Menu ............................................................................15
2.4.1.4 Data Main Menu ............................................................................................15
2.4.1.5 Help Main Menu ............................................................................................16
2.4.2 Microscopy and Verification Screen ........................................................................16
2.4.2.1 Urine Real-time Monitoring Screen ...............................................................17
2.4.2.2 Urine Microscope and Verification Screen ....................................................19
3 INSTALLATION ............................................................................................................................21
3.1 Summary ............................................................................................................................21
3.2 Installation Requirements ..................................................................................................21
3.2.1 Space .......................................................................................................................21
3.2.2 Power .......................................................................................................................21
3.2.3 Environmental requirements....................................................................................21
3.2.4 Installation ................................................................................................................22
3.3 System Connection ............................................................................................................22
3.4 Pipeline connection of the analyzer ...................................................................................22
3.5 Power-cable connection.....................................................................................................23
4 SETUP ..........................................................................................................................................24
4.1 User Set .............................................................................................................................24
4.1.1 User login .................................................................................................................24
4.1.2 User Set Operations ................................................................................................25
4.1.3 Password Modification .............................................................................................25
4.2 Microscopy Setting .............................................................................................................26
4.2.1 Item Setting ..............................................................................................................27
1
AVE-772 User's Manual
2
AVE-772 User's Manual
3
AVE-772 User's Manual
4
AVE-772 Introduction
1 INTRODUCTION
1.3.1 Cautions
CAUTION: Identifies laser light. A laser is used to read the barcodes only. Protect
eyes from the laser light.
1.3.2 Warnings
5
AVE-772 Introduction
1.4 Symbols
The following is a list of symbols used on the nameplate, product packaging or within the
user manual and their corresponding meaning.
Nameplate, product
10 Date of manufacture
packaging
In vitro diagnostic medical
11 Nameplate
device
Nameplate, product
12 Manufacturer
packaging
Authorised representative in Nameplate, product
13
the European Community packaging
Nameplate, product
14 CE Mark
packaging
6
AVE-772 Introduction
7
AVE-772 Introduction
without advance approval by AVE may affect the terms of or void any warranty otherwise in
effect, covering AVE-772 supplied software and hardware on the computers and the overall
performance and reliability of the entire system.
8
AVE-772 System description
2 SYSTEM DESCRIPTION
9
AVE-772 System description
value of the difference with the reference solution to not more than one order of magnitude.
Positive results do not appear negative, negative results do not appear positive;
Repeatability: See the table below
Concentration ( cell/µl) 20~100 500~1000 5000
CV(%) ≤20 ≤12 ≤8
Cross Contamination(Microscopy):≤1 cell/µl at concentration of 4600~5400 cell /µl; ≤2
cell/µl at concentration of 9200~10800 cell/µl;
Cross contamination(Chemistry):Using negative specimen to do cross test with the 3+
positive quality control, the negative sample will not show positive result;
Test Sample Volum:3ml;
Sampling and Cleaning: Automatical sampling,homogenizing,filling and cleaning;
Counting Chamber: Each channel work by time-share, and automatically cleaning,
constraint cleaning and maintenance;
Network Function:Information can be shareable through network;
Expansion Tests:Expansion tests can be carred out, such as Hydrothorax, Cerebrospinal
Fluid etc;
Report Form:Comprehensive report (including chemical, physical, particles counting and
morphological images, etc.);
Memory Capacity:≥20,000;
Microscopy Control:Five-axis automatic control, control precision is less than 0.5µm,with
automatic focus.
Quality Control Function : Record quality control parameter, do quality control result
statistics, display and print quality control statistical graph;
10
AVE-772 System description
3 4
Fig.2-1 Front view of the system
1.Microscope observation window;2.AVE-772; 3.Automatic sampler; 4.Barcode reader;
11
AVE-772 System description
1
2
Fig.2-3 Left side view of the system
1.Main power switch; 2.Automatic sampler power switch;
12
AVE-772 System description
13
AVE-772 System description
Microscopy and
Quit
Verification
Urine History
Verification Urine Inquiry User Management Data Backup
Fig. 2-6 Shortcut keys area
14
AVE-772 System description
cross-reference table.
Chemistry Setting Set Chemistry Items and Chemistry Protocol
Set the patient’s information of import, printing and
Patient Info. Setting
displaying
Mark Color Setting Set the target display color
Print Format Set report template
Print Setting Set the print and preview parameters
Adjustment Adjust parameters of the analyzer
User Set Create user, modify user password
Backup Backup and restore database and settings
15
AVE-772 System description
16
AVE-772 System description
Specimen ID Area
Fig. 2-14 Microscopy Screen
17
AVE-772 System description
the two states. Functions: The instrument will fulfill the current sample analyzing by clicking
“Start” . This operation is for normal microscopy. The instrument will stop working temporarily
on clicking “Pause” and turn to maintenance state. The user can conduct a relevant quick
maintenance and resume its work by clicking on “Continue”. It will be converted to the specific
channel by clicking the corresponding one and begin sampling, focusing and washing.
“Stop”: Stop testing normally. The system will stop working after finishing the sample,
which is currently tested in the cell by clicking on “Stop”. The next sample on the current rack
will not be tested.
“Forced”: Forced to stop testing. The system will stop testing both the sample in the cell
and the next one on the current rack by clicking on “Forced”.
18
AVE-772 System description
A. Functional Buttons
“Modify”:Clear out the counting data and cell type of the current picture and change to
another.
“Previou”: Select the previous sample.
“Next”: Select the next sample.
“Zoom”: Enlarge or diminish the selected pictures in the image showing field.
19
AVE-772 System description
C. Chemistry Data
Click the “▲”or “▼”button to preview the chemistry data received.
E. Frame
The frame information of the current sample appears here.
20
AVE-772 Installation
3 INSTALLATION
3.1 Summary
Anyone who is not trained by AVE or authorized distributors of AVE may cause damage
to the analyzer when opening the box or installing the instrument. Therefore do not open the
box or install the instrument without an authorized technician being present.
On receiving the analyzer, please check the package carefully to see whether there is a
physical damage. If there is any, please notify your local distributor immediately.
3.2.1 Space
AVE-772 Integrated Urinalysis Analyzer (with its mainframe size of
825(L)x648(W)x550(H) mm) is a tabletop unit requiring free space on a countertop of
approximately 250x120 cm, with vertical clearance of at least 30 cm.
3.2.2 Power
Voltage Frequency Input power Fuse
AVE-772 100-240V 50/60Hz ≤650VA T3.15A 250V Φ5×20
CAUTION: UPS no less than 1500VA or voltage regulator are recommended for the
whole system operation if voltage is not stable or power outage is regular.
CAUTION: The equipment inside fixed covered areas of the instrument are not user
serviceable and may present electrical shock hazards.
WARNING: For continued protection against risk of fire and hazard, replace only with
the same type and rating fuse.
WARNING: The instrument’s main supply inlet is being used as the mains
disconnect device.
21
AVE-772 Installation
3.2.4 Installation
The AVE-772 Urinalysis System has to be installed by an authorized technician.
Warning: User has the duty to comply with the relevant regulations set by the region
and country about the treatment and discard of expired reagents and wastes.
Warning: Reagents may cause some irritant to skin and eyes. User should always
comply with the safety operation of laboratory and wear personal protective equipment (such
as protective clothing, gloves, etc.) when coming contact with reagents.
22
AVE-772 Installation
23
AVE-772 Setup
4 SETUP
The analyzer settings are already set by the factory and the screen you see after the
initial boot is set on system default. In order to meet the individual demands, some
parameters of the analyzer can be modified. Since the settings will directly affect the normal
use of the instrument, an access control has been set up for the user to ensure the safety of
the equipment.
There are two modes of use and appropriate rights: The "user" can perform clinical
urinalysis in daily use, but can not access "Management" main menu. In addition to the rights
owned by the "User", the "Administrator" can modify specific parameters on the
"Management" main menu.
CAUTION:Changing of the factory parameter settings could affect the normal use of
the instrument , if it’s necessary, please read the user’s manual first or contact manufacture.
24
AVE-772 Setup
User Set
25
AVE-772 Setup
user’s password.
2. Click the daily operation main menu, select “Password” in the drop-down submenu as
shown in Fig.4-4. The “Password” modifying window will open. The password of the login
user can be modified in this window. This is available for all users.
26
AVE-772 Setup
27
AVE-772 Setup
28
AVE-772 Setup
Add operations:
1. Choose Female or Male in the Gender box.
2. Select a blank row, enter maximum age, minimum age and the corresponding reference
values of RBC, WBC, Crystal, Yeast, Cast and Epithelium.
3. Click Save button and click “OK” in the pop-up window to validate the adding.
Delete operations:
1. Choose Female or Male in the Gender box.
2. Select the row of reference values of age range to be deleted.
3. Click Delete then Save button, and click OK in the pop-up window to confirm.
Modify operations:
1. Choose Female or Male in the Gender box.
2. Select the parameter to be altered.
3. Enter a new corresponding parameter,
4. Click Save button and click OK in the pop-up window to confirm.
5. Click Quit to exit.
29
AVE-772 Setup
30
AVE-772 Setup
31
AVE-772 Setup
2. Click the Up or Down button to move the item to where you want.
3. Click the Save button and “OK” in the pop-up window to validate the deleting.
Click the Quit button after all the operations fulfilled to exit the Chemistry Setting screen.
32
AVE-772 Setup
3.Select the corresponding chemistry analyzer in “Save”, click “Save” (lower left) to
accomplish adding.
4.3.2.3 Adding New Chemistry Protocol
1.Connect the chemistry analyzer to AVE-772 correctly.
2.Enter a new chemistry protocol in “New”, click “Add”.
3.Select the corresponding chemistry analyzer in “Protocol”, click “Test”. Choose any data of
the chemistry analyzer and transfer it to the test screen, then fill out the “Start”, “End”, “ID”,
“Y”(year), “M”(month), “D”(date). Click “Save” (upper left).
4.Select the corresponding chemistry analyzer in “Save”, click “Save” (bottom left) to
accomplish adding.
4.3.2.4 Adding Comparison Data
1.Select the corresponding item in “Items”.
2. According to the actual data of the chemistry analyzer, click the blank in “Actual data” and
input the information. Click the blank in “Report data” and input the corresponding information.
Click “Save” (upper right) and click “OK” in the pop-up window to accomplish adding.
3. Select other items and fulfill all the parameter settings referring to step 2.Click Quit to exit.
33
AVE-772 Setup
34
AVE-772 Setup
35
AVE-772 Setup
36
AVE-772 Setup
CAUTION:If change the color setting, please set different cell types in different
37
AVE-772 Setup
4.8 Tools
Select Tools in Microscopy main interface, users can perform Test Mode Setting, Users
settings, Edit ID, Unusual, Gallery, Positive List, MagnGlass On/Off etc.
38
AVE-772 Setup
39
AVE-772 Setup
40
AVE-772 Setup
41
AVE-772 Setup
42
AVE-772 Specimen processing
5 SPECIMEN PROCESSING
5.1 Power On
1. Turn on the UPS (if any).
2. Turn on the power distribution board (if any).
3. Turn on the power switches of printer, monitor, etc.
4. Turn on the main power switch(Green switch on slide, if it’s on, the yellow LED inside of the
Green switch will be lighted).
5. Turn on the auto-sampler’s power switch(Green switch on slide, if it’s on, the yellow LED
inside of the Green switch will be lighted).
6. Press the Power button of the PC Module to enter Windows OS.
5.2 Login
Double click the shortcut icon on desktop to display the login window shown in
Fig.5-1
CAUTION: Ensure that the SQL Server in task bar is started before entering the
AVE software. If not, double click the icon as shown in Fig.5-2 to pop up the SQL Server
Manager interface as shown in Fig.5-3. Click the start / continue button to launch.
43
AVE-772 Specimen processing
44
AVE-772 Specimen processing
1. Apply barcode labels to sample tubes, make sure that the barcode (not the barcode label)
is ½ inch (about 12.7 mm) below the top edge of the tube(Φ16 x 100 mm).
2. Prepare at least 4ml well-mixed native specimens into sample tube.
3. Put the tubes into the sample rack correctly and in order.
4. Launch the sample rack on the Ready Area of the sampler.
5. Mark the tubes as shown in Figure 5-5.
45
AVE-772 Specimen processing
CAUTION: Make sure the barcode labels are properly oriented in the rack. The
sample tubes must be placed straight and resting in the middle of the grommets located in the
base of the rack.
46
AVE-772 Specimen processing
CAUTION: Each time to put a new can of test strips, make sure to replace the can’s
cap into the strip tank door’s dryer hole.
47
AVE-772 Specimen processing
CAUTION: Do NOT move the sample racks in Ready Area during testing.
WARNING: Don’t place your hand below the sampling position to avoid injuries by the
sampling needle.
48
AVE-772 Specimen processing
For each testing result, the system will presents the status for this sample accordingly, the
user can define the further processing according to the status. In general, the status are:
NE (Not Examined): The microscopy examination of this sample is not finalized yet.
NV (Not Verified): The microscopy result of this sample is positive, or the strip result did
not coincide with microscopy result, further verification is recommended.
V (Verified): The microscopy result of this sample is negative, or the strip result coincide
with microscopy result, it’s not necessary to perform further verification.
R (Reported): This sample was already verified and printed out.
NV! (Not Verified!): The microscopy result of this sample is positive, or the strip result did
not coincide with microscopy result, it’s NECESSARY to perform further verification.
EV (Expert Verify): This sample was already verified by online expert. (only available for
remote connection mode)
NS (No Sample): The specified volumn of this sample is not attained, thus it was not
examined. The user need to check the volumn and test this sample again.
49
AVE-772 Specimen processing
5.8.2 Recheck
If it’s needed, users can perform Redo operation to repeat the Microcopy testing. Operation
steps are as following:
1.Prepare the emergency sample.
2.Put the sample at emergency position and push it forward, wait until system pop up the
Insert STAT interface as shown in Fig.5-13.
3. Select Urine Microscopy in Micro box, Redo in Type box and Urine Redo Item in Redo
box as shown in Fig.5-13.The numbering rules is that start with 3 and last three number is as
same as Sample ID’s last three number, for example, if the Sample ID is 0013, then the Redo
ID is 3013.
50
AVE-772 Specimen processing
51
AVE-772 Specimen processing
52
AVE-772 Specimen processing
then click the "+ " symbol on the target needed to modify, the target name will change to the
selected name, and the "+ " symbol will also change. The result of current frame will increase.
Repeat this operation to modify other targets in the same frame.
4. Click Next to view all LP pictures of the sample in full-view and repeat step 2 and 3 until
finished.
HP Verification
5. Same as Basic operations step 5-6.
6. Click Modify, then click the "+ " symbol on the target needed to cancel , to cancel the
target. When the target is cancelled, the "+ " symbol will disappear, and the result of current
frame will reduce.
7. Keep Modify is selected, then click the target name in the Current Cell Counting field,
then click the "+ " symbol on the target needed to modify, the target name will change to the
selected name, and the "+ " symbol will also change. The result of current frame will increase.
Repeat this operation to modify other targets in the same frame.
8. Click Next to view all HP pictures of the sample in full-view and repeat step 6 and 7 until
finished.
Save Result
9. Click Save button under the Current Cell Counting field to confirm the verification result of
selected sample, and the sample status is changed to V.
5.9.3 Batch operations
The user can use batch operations to modify a specific targets in one frame or all frames.
Enter the Urine Verification interface and refer to the following:
LP Verification
1. Same as Basic operations step 1-2.
2. Click Modify, then choose Save Type in the Clear Mark drop-down menu, then the batch
mofiy window will appear. Choose Current/All in Picture to proceed current frame /all frames,
in Type choose the target name to be cleared, in Change to choose Cancel, then click Save.
All the "+" symbol on selected targets in current frame /all frames will disappear, and the result
in current frame /all frames will become 0. Repeat this operation to clear other targets.
3. Click Modify, then choose Save Type in the Clear Mark drop-down menu, then the batch
mofiy window will appear. Choose Current/All in Picture to proceed current frame /all frames,
in Type choose the target name to be modified, in Change to choose the target name, then
click Save. All the "+" symbol on selected targets in current frame /all frames will change, and
the result in current frame /all frames will change. Repeat this operation to modfiy other
targets.
53
AVE-772 Specimen processing
54
AVE-772 Specimen processing
55
AVE-772 Specimen processing
Quit
56
AVE-772 Specimen processing
57
AVE-772 Quality Control
6 QUALITY CONTROL
58
AVE-772 Quality Control
59
AVE-772 Quality Control
5 After QC testing is finished, the results will be displayed in Urine Special area, and
there will show QC normal/ QC warning / QC out of control messages in the status bar.
60
AVE-772 Quality Control
61
AVE-772 Quality Control
62
AVE-772 Quality Control
63
AVE-772 Quality Control
64
AVE-772 Adjustment
7 ADJUSTMENT
7.1 Summary
The AVE-772 Analyzer’s adjustment include: Microscope Focus and CCD Image
Parameters adjustment, Pump Calibration and Blank Calibration etc. Enter the main screen
and choose “Management” to open the “Adjustment” submenu (shown in Fig.7-1).
65
AVE-772 Adjustment
Steps Focus
Left-right F-Backward Up-down Lens
Channel Lamp
LP 400~800 600~1000 700~1600 50~250 700
CH1 Difference Difference
HP Same as LP 700~850 200
with LP≤120 with LP≤120
Difference
LP 400~800 700~1600 50~250 700
with CH1≈2835
CH2
Difference Difference
HP Same as LP 700~850 200
with LP≤120 with LP≤120
Difference
LP 400~800 700~1600 50~250 700
with CH2≈2835
CH3
Difference Difference
HP Same as LP 700~850 200
with LP≤120 with LP≤120
Each Counting Chamber will provide these parameters, when the Counting Chamber is
replaced, click “Management” - “Adjustment”, then click Replace chamber, and in the popup
window, enter the new parameters, click Save , then click Quit.
66
AVE-772 Adjustment
67
AVE-772 Adjustment
Note: The F-Backward step of CH2 & CH3 is 3385 & 6220.
68
AVE-772 Adjustment
go to Step4.
69
AVE-772 Adjustment
4. Switch to CH2 and CH3 respectively, and repeat step 1-3 to finalize the settings.
5. Reture to CH1 and tap LP Reset button to switch to LP lens, and click Auto Focus and
then the microscope will perform focus calibrating automatically.
70
AVE-772 Adjustment
71
AVE-772 Adjustment
72
AVE-772 Adjustment
11. Switch to CH3 and repeat step 8-10 to adjust the focus of CH3.
73
AVE-772 Adjustment
7. Check “LP” in the “Precise Adjust” field to enter the LP filed; the operations are the
same as HP. Click “Save All” to save all the adjusted parameters.
7.3.1 Operations
The pump calibration can be performed at the lower right of the adjustment screen.
7.3.2 Caution
If the wash/sample calibration value is abnormal in the process of Pump Calibration,
please check whether the pipeline is clogged.
74
AVE-772 Adjustment
CAUTION:If the dirty spot exceeds 5, please perform special wash, then perform
Blank Calibration again; If the dirty spot still exceeds 5, it is recommended to replace the
counting chamber.
75
AVE-772 Results
8 RESULTS
8.1.1 QC Inquiry
Click “Data”-“Urine Data”-“QC Inquiry” to enter the QC Inquiry Screen.
The user can inquire and backup previous QC records.
76
AVE-772 Results
The user can inquiry, backup, preview and print history records (Shown in Fig.8-3).
77
AVE-772 Results
statistics screen. Workload, urine result and chemistry result can be counted here. The type of
the statistics varies with the static option. Both test results of the day and those without
patient information cannot be counted.
78
AVE-772 Results
79
AVE-772 Results
80
AVE-772 Results
81
AVE-772 Maintenance and Service
9.1 Maintenance
9.1.1 Precautions
82
AVE-772 Maintenance and Service
CAUTION: Do not wipe back and forth, wipe in one direction to avoid scratches on the
counting chamber surface!
9.2.4 Cleaning Waste Tank
CAUTION: Immediately power off the sampler and clean it while a whole rack samples
are pouring onto the sampler.
Tools required: Disinfectant/detergent, gauze, and protective gloves.
Procedure:
1 Power off the instrument. Ensure that the maintenance is performed under power cut-off
condition.
83
AVE-772 Maintenance and Service
2 Moisten gauze with disinfectant/cleanser and wipe the sampler to remove any deposits.
3 Check the flexibility of pawls. Moisten cotton swabs with disinfectant/cleanser to clean
pawls, remove any deposits.
4 Wipe again using distilled water.
5 Power on the instrument until the sampler is totally dry.
84
AVE-772 Maintenance and Service
Procedure2: Move the pinch pipelines upper or down 2mm per month to prevent adhesion or
broken after long time clamped.
Contact the service for replacement if you need a new pinch pipeline.
9.4.2 Cleaning the Waste Pool
In order to prevent pipeline block and invalidation of liquid detection, the waste pool is suggested
to perform cleaning every month. This includes cleaning the inner wall, cleanser nozzle upper and
waste nozzle at bottom.
Tools required: Deionized water, cotton swabs and protective gloves.
Procedure:
1 Make sure the analyzer is in shutdown state.
2 Open the front cover to access the waste pool. Manually lift the sampling needle out.
3 Using a tube brush moistened with deionized water, remove salt deposits present on the
waste pool. Clean the inner wall.
4 Using a cotton swab moistened with deionized water to swab cleanser nozzle upper and
waste nozzle at bottom.
5 Move the sampling needle back into the waste pool, close the front cover.
6 Power on the instrument and login the software. Click Wash All 1-2 times in Adjustment
interface to wash all the channels.
CAUTION: Do not insert the cotton swab inside the waster pool; cotton particles may
clog the tubing connectors.
9.4.3 Pump Calibration
Refer to 7.3 Pump Calibration for details.
9.4.4 Cleaning Sampling needle
Tools required: 95% absolute alcohol, cotton swabs and protective gloves
Procedure:
1 Make sure the analyzer is in shutdown state.
2 Open the front cover to access the waste pool. Manually lift the sampling needle out.
3 Using a cotton swab moistened with absolute alcohol to swab the interface of needle,
4 Move the sampling needle back into the waste pool, close the front cover.
9.4.5 Cleaning Photoelectric Sensors on auto-sampler
Cleaning photoelectric sensors on auto-sampler is recommended to perform every month.
Tools required: 95% absolute alcohol, cotton swabs and protective gloves
Procedure:
1 Power off the instrument, make sure the operation is in shutdown state.
2 Using a cotton swab moistened with absolute alcohol to swab the surface of photoelectric
sensors
3 Swabs dry them by using a dry cotton swab.
85
AVE-772 Maintenance and Service
and glove to protect you from the waste remaining in the pump tube!
86
AVE-772 Maintenance and Service
87
AVE-772 Maintenance and Service
CAUTION: Make sure stretching the new pump tube before replacement.
CAUTION: After installing the new pump tube into the pump, make sure the rotor can
spin smoothly, , the pump tube should be no distortion, not protrude from the pump.
88
AVE-772 Maintenance and Service
CAUTION: In order to prevent affecting the pump accuracy, refer to 7.3 Pump
Calibration to perform the pump calibration after replacing the pump tube!
89
AVE-772 Maintenance and Service
CAUTION: Make sure the replacement is done in power off state! To prevent burning
by the lampshade, make sure it is already cooled before replacement!
Replacement procedure:
1 Power off the instrument; make sure the lampshade is already cooled. Open the front cover to
find the fix screw located on the base.
90
AVE-772 Maintenance and Service
Lamp shelter
Lamp
CAUTION: Socket oxidation may cause contact problem. If the lamp still does not
light on, please repeat plug several times.
91
AVE-772 Maintenance and Service
92
AVE-772 Maintenance and Service
CAUTION: The installation trends should follow this: the face with eminence is upper
surface and the other one is on the bottom; the side with serial number printed on should
be toward inside. The chamber pipeline should be plugged to 1/2 of iron tube (as shown
below).
93
AVE-772 Maintenance and Service
94
AVE-772 Maintenance and Service
95
AVE-772 Maintenance and Service
CAUTION: To prevent losing data, do not perform any other operation during the
process of backup.
9.7.2 Data Restore
CAUTION: Only properly trained and authorized service personnel can operate
restoring. Make sure that your operation are under the instruction of authorized service!
Operation procedure:
1 Click Browse button to choose the specified restore path.
2 Click File button to select files (bak. format) need to be restored in specified path.
3 Click Open button to start restoring your selected data.
4 Select tick box in Urine Data or Parameter area to restore data need to import.
5 Click Import button to import the selected item from the specified path.
96
AVE-772 Maintenance and Service
CAUTION: To prevent losing data, do not perform any other operation during the
process of restore.
9.8 Troubleshooting
9.8.1 Shutting Down the Instrument
When it is necessary to restart the instrument, shut down the instrument and the
computers in the following order:
1 Click Quit button on the Microscopy screen.
2 Select Quit and auto turn off, unselect Quit and maintain, then click Quit button.
3 After the device shutting down, turn off the power of sampler.
4 Turn off the switch at the left of instrument.
9.8.2 Restarting the Instrument
1 Turn on the switch of the instrument.
2 Turn on the power of sampler.
3 Press on the power switch on the front panel to start the system.
4 Double-click the software icon on the desktop. Login the system.
After starting the examination, the 1 Check whether the switch of sampler is turn
system notice as ‘Sampler on. If not, turn on the switch;
communication timeout’. 2 Confirm the error message, click Forced
button to forced stop the examination;
2
3 After the examination is stopped, click Start
button for recheck.
4 If the problem cannot be eliminated, please
contact the authorized service personnel.
The system notice as ‘Push-in motor 1 Check whether the sampler is placed at a
at ready area reaching position direct strong light area. If so, prevent the
error.’ sampler from strong light.
3 2 Confirm the error message for its own
eliminated.
3 If the problem still remains, shut down the
instrument and then restart.
97
AVE-772 Maintenance and Service
During examination, the tube rack 1、 Click Forced button to forced stop the
cannot be shifted by the pawls at examination. Take the tube rack out from
testing area. testing area.
2、 Check whether the tube rack is for specialized
use. Change the tube rack if it is not.
4 3、 Check the flexibility of pawls. Moisten cotton
swabs with disinfectant/cleanser to clean
pawls, remove any deposits.
4、 Restart the instrument for further check. If the
problem cannot be eliminated, please
contact the authorized service personnel.
Sample was missed examination by 1 Shut down the instrument and then restart.
5 instrument. 2 If the problem cannot be eliminated, please
contact the authorized service personnel.
Free position on tube rack was exam 1 Check whether the sampler is placed at a
by accident. direct strong light area. If so, prevent the
sampler from strong light.
6
2 Shut down the instrument and then restart.
3 If the problem cannot be eliminated, please
contact the authorized service personnel.
The system notice as ‘Push-out 1 Check whether the sampler is placed at a
motor at tested area reaching direct strong light area. If so, prevent the
position error.’ sampler from strong light.
7
2 Shut down the instrument and then restart.
3 If the problem cannot be eliminated, please
contact the authorized service personnel.
The system notice as ‘Tested area is 1 Remove all the finished tube racks at tested
full.’ area and confirm the error message for its
own eliminated.
2 If the problem still remains, check whether the
sampler is placed at a direct strong light area.
8 If so, prevent the sampler from strong light.
Confirm the error message for its own
eliminated.
3 Shut down the instrument and then restart.
4 If the problem cannot be eliminated, please
contact the authorized service personnel.
98
AVE-772 Maintenance and Service
99
AVE-772 Maintenance and Service
100
AVE-772 Maintenance and Service
101
AVE-772 Maintenance and Service
102
AVE-772 Maintenance and Service
103
AVE-772 Maintenance and Service
104
AVE-772 Maintenance and Service
105
AVE-772 Maintenance and Service
The system notice as ‘Failed to read 1 Restart the instrument to check whether the
or write registry!’ problem is eliminated.
49
2 If the problem cannot be eliminated, please
contact the authorized service personnel.
To calibrate the pump system, the 1 Restart the instrument to check whether the
system notice as ‘Valve Test failed!’ problem is eliminated.
50
2 If the problem cannot be eliminated, please
contact the authorized service personnel.
To calibrate the pump system in 1 Re-calibrate the pump system to eliminate
Adjustment interface, the system the error automatically.
notice as ‘Reagent Stop;’ 2 If the problem still remains, check if the
cleanser A was expired or with precipitate
inside. Replace a new one if it was expired.
51 Shake it well if there is precipitate inside.
3 If the problem still remains, refer to 9.4.2
Cleaning the Waste Pool to maintain the
waste pool.
4 If the problem cannot be eliminated, please
contact the authorized service personnel.
The system notice as ‘The pump 1 Re-calibrate the pump system to eliminate
calib time is longer than expected’. the error automatically.
52
2 If the problem cannot be eliminated, please
contact the authorized service personnel.
The system notice as ‘Abnormal 1 Restart the instrument to check whether the
Data:’ problem is eliminated.
53
2 If the problem cannot be eliminated, please
contact the authorized service personnel.
106
AVE-772 Appendix
10 APPENDIX
107
AVE-772 Appendix
108