Professional Documents
Culture Documents
9 , 1044-1050 (1995)
SPONSOR REFEREE: Prof. Kenneth Standing, University of Manitoba, Winnipeg, Manitoba, Canada
A new delayed extraction matrix-assisted laser desorptiodionization (MALDI) time-of-flight (TOF) mass
spectrometer system is described which provides dramatically improved performance over that obtained with
identical TOF analyzers operated with constant electrical fields. In this system the ions formed by MALDI are
produced in a weak electrical field, and subsequently extracted by application of a high voltage pulse after a
predetermined time delay. Three parameters can be independently adjusted to optimize the performance. These
include the field magnitude and direction during and immediately following ion production, the magnitude of the
extraction field, and the time delay between the laser pulse and application of the extraction field. Theoretical
equations are presented which guide the selection of these parameters to optimize the performance.
Results are presented from three TOF analyzers of similar design, but ranging in size from a 1.3 m linear
analyzer to a 6.4 rn reflector. In all cases delayed extraction provides substantially improved resolution over that
obtained from static operation. In the smallest linear analyzer resolution of M/AM (full width at half maximum)
of 4000 is demonstrated for angiotensin I, and with the largest reflecting analyzer, the isotope peaks of bovine
insulin are resolved nearly to baseline. With larger proteins, the isotopic peaks are not resolved, but resolution
approaching the width of the isotopic envelope is obtained. Resolution improvement is demonstrated for proteins
up to 30 kDa. In addition to improved resolution, delayed extraction is shown to improve the quality of MALDI
mass spectra by suppressing matrix background, reducing chemical noise, and minimizing the effect of laser
intensity on performance.
Time-of-flight (TOF) mass spectrometry has been These authors also described a technique which they
experiencing a revival since the introduction of soft named 'time-lag energy focusing' for correcting for
ionization techniques compatible with this type of mass initial velocity distributions. In this instrument, the ions
analysis such as plasma desorption,' and matrix- are produced in a field-free region, and the accelerating
assisted laser desorption/ionization (MALDI).24 The field is turned on by application of a fast pulse at a
appeal of TOF mass analysis is also indicated by its predetermined delay time after initial ion formation.
recent, highly promising combination with continuous Recently, Brown, Lennon and Christie,", l2 Colby,
ionization methods such as e l e c t r o ~ p r a y .The
~ , ~ wide- King and Reilly13 and Whittal and Li14 have applied
spread use of MALDI in biological applications has similar mass analyzers to MALDI, and have shown
particularly spurred innovation in the design of TOF significantly improved resolution for selected masses as
analyzers leading to improved mass resolution, sensiti- a function of delay time and pulsed field intensity in the
vity, and mass accuracy. ion source.
A major limitation of TOF mass analyzers has often According to the theory of 'time-lag energy focusing'
been relatively poor mass resolution. The use of reflect- as originally developed by Wiley and McLaren,'" the
ing analyzers, such as that described originally by dependence of ion flight time on initial velocity can be
Mamyrin7has overcome this limitation for many appli- corrected, to first order, by delaying the extraction of
cations, but with MALDI in particular, resolution is ions from the source by an appropriate amount. If
often inadequate. Earlier studies by Beavis and Chait,' higher order terms are insignificant, then the mass
and by Zhou, Ens, Standing, and Verentchikov' indi- resolution should be determined by the ratio of the
cate that the major problem is that the ions produced total flight time to the uncertainty in the time measure-
by MALDI exhibit a rather broad energy distribution. ment. In this case, the observed mass resolution should
Initial velocity of desorbed analyte ions is nearly inde- increase in proportion to the effective length of the ion
pendent of mass; thus the initial kinetic energy is flight path. In the work of Wiley and McLaren, both
proportional to mass of the analyte. In addition, when the initial space and velocity distribution of the ions
desorption occurs in a strong electrical field, energy is were variables and simultaneous correction for both
lost presumably by collisions with the neutral plume, terms could not be accomplished. In surface ionization
and further mass dependent energy dispersion results. techniques such as MALDI, space focusing is not a
In very early work, Wiley and McLaren'" described a factor; therefore, corrections for velocity focusing will
two-field, pulsed ion source which provided first order greatly improve performance. The present work was
correction for the initial space distribution of ions. undertaken to determine the extent to which perfor-
mance of MALDI for analysis of peptides and proteins
is improved by delayed extraction, and to explore how
* Author for correspondence. this performance depends on the size of the instrument.
Sample Linear
plate Extraction
Timed ion detector
grids
\ A
I
selector
\ Reflector i
? 1"111111"
I
I
-.= i
Pum;ing
Camera
Beam guide
I11111111111
Reflector
detector
Pumping
Figure 2. Schematic diagram of time-of-flight mass analyzer employed in this work.
1046 DELAYED EXTRACTION MALDI TOF MASS SPECTROMETRY
I I
(RPI- 1 lRPl
Res.:4000 - Res.:4600
1 %
I I I I I I I I I I I t I I
I
El I m- 11 E l
I
Res.: 5800 Res. :6600 - Res.23600
1
Iv J
I I 1 I I I 4 I I I I I I I I I I I
(4 I
#
Res.: 5100
(M+H)+
d r
(M+2H)Z+
I -25Vbias I
(M+H)+
2455 2465
d Z
2475
I (M+2H)2+
[M+H~+ 1 (M+207)'
shown in Fig. 5. With the shortest analyzer (RP) the neutrals may interfere with the detection of larger
isotopes are not resolved, but with the intermediate proteins at low levels. Ion gating systems have been
system (EL) the isotopes are clearly resolved, corres- developed to overcome this problem, and the analyzers
ponding to a resolution (M/AM) (FWHM) of slightly used in the present work are equipped with a deflection
more than 6000, while in the largest system (XL) the gate for removing unwanted low-mass ions, and vari-
isotopic peaks are resolved nearly to baseline. In this able voltage detector gates for preventing detector
latter case the measured resolution, M/AM (FWHM) = saturation by low-mass neutrals. Nevertheless, chemi-
12 500. cal noise may still limit the sensitivity for higher-mass
For larger proteins the isotopes are not resolved, but species. We have found with delayed extraction that
nearly isotopically limited resolution can be obtained the matrix background can be very efficiently sup-
with delayed extraction in both linear and reflecting pressed by applying a retarding field (reverse bias) to
analyzers, and the results are almost independent of the the sample prior to extraction. An example is presented
drift distance. This is due to the fact that all the in Fig. 6, where mass spectra of bovine insulin were
analyzers in the delayed extraction mode have an recorded in CHCA matrix at three different initial
inherent resolution specification of greater than M/AM sample potentials using delayed extraction with the
(FWHM) = 3000, while the isotopically limited resolu- extraction field applied about 700ns after the laser
tion of larger proteins is approximately M/AM pulse. The same spot was irradiated in all the three
(FWHM) = 2000. examples. Figure 6(a) corresponds to approximately
One of the problems that is frequently encountered zero field prior to ion extraction. Equally strong signals
in application of MALDI to larger proteins is that the from the peptide and matrix are observed. In Fig. 6(b)
intense matrix peaks, and associated fragments and the sample was biased approximately 25 V below the
Res.: 1100 -
(RNAse A ) '
grid potential. The matrix ions are slightly less abun- (GlcNAc),Man,, and (GlcNAc),Man, , with calculated
dant, and, in particular, monomer ions of the matrix masses of 14899.1, 15 061.3, 15 223.4, 15 385.5, and
are suppressed. Figure 6(c) illustrates the effect of 15 547.7, respectively. The last component is often
applying a 50 V reverse bias on the sample; the matrix undetected under continuous extraction MALDI
ions are completely absent from the mass spectrum. ~0nditions.l~Measured mass resolution for all five gly-
Mass resolution in all cases was in the range M/AM coforms is in excess of MIAM (FWHM) = 1100. The
(FWHM) = 950-1000, corresponding approximately to unglycosylated protein is slightly out of the focus and
the width of the isotopic distribution. the resolution is 'only' M/AM (FWHM) = 900.
This method for suppression of matrix background is
also effective for larger analytes. As an illustration, a
DE-MALDI mass spectrum of cytochrome c in a sina- CONCLUSIONS
pinic acid matrix is shown in Fig. 7. Sample voltage was As predicted by theory, delayed extraction
about 70 V below the grid potential prior to application MALDI-TOF provides substantial improvement in
of the extraction field. The normal low-mass gating resolution over that obtained under static conditions
system was not used in these measurements. The back- for both linear and reflecting analyzers. For the three
ground is very clean and matrix ions are of low abun- analyzers tested, the resolution in reflector mode was
dance in the mass spectrum, even using sinapinic acid, very nearly proportional to total effective ion flight
which normally gives a very strong matrix background. path, but the resolution of the linear analyzers
Mass resolution in this example is M/AM (FWHM) = increased rather more slowly with increasing flight
1150. As the peak labels indicate, it was possible path. Using delayed extraction, the mass resolution for
to resolve two matrix adducts to the molecular ion peptides was apparently limited in the shortest analyzer
+
differing by only 18 mass units (M 207 and M 225, + by resolution of the time measurement rather than by
respectively). the initial velocity distribution, but with the larger
The largest protein to date for which data have been analyzers higher order terms which are independent of
obtained, that clearly show the advantages of DE, is flight distance become more important.
carbonic anhydrase; an example is shown in Fig. 8. In In addition to improved mass resolution, delayed
this case the measured resolution is M/AM (FWHM) = extraction provides other improvements in the quality
840, and a fragment or impurity peak 44 mass units less of MALDI spectra, which are more difficult to quan-
than the molecular mass is partially resolved (-CO,). tify. These include a lower level of chemical noise due
Since the loss of small neutral fragments has frequently to fragmentation in the ion accelerating fields, reduc-
been observed in MALDI of small proteins,' the mea- tion or elimination of matrix background, and signifi-
sured resolution may be limited by unresolved frag- cant reduction in the dependence of ion flight times on
ments such as that produced by loss of ammonia, or laser intensity. With delayed extraction, peak widths
carbon dioxide. and centroids are much less dependent on the absolute
The advantages of DE are more readily apparent intensity of the ion signal; thus, it is not necessary to
with more complex samples. Figure 9 presents a operate near the threshold for ion production to obtain
DE-MALDI mass spectrum of ribonuclease B, a small both high resolution and improved mass accuracy.
glycoprotein with an inhomogeneous, high-mannose In this work, we have demonstrated advantages of
type glycan. The matrix was sinapinic acid. The mass of delayed extraction MALDI-TOF for applications to
the unglycosylated protein, ribonuclease A, is peptides and proteins. We anticipate that this new
13 682 Da, and it is present as a minor component. The technique may prove to be of even greater value for
glycoforms are easily identified in this mass spectrum as applications to more fragile biopolymers such as oligo-
(GlcNAc),Man, , (GlcNAc),Man, , (GlcNAc),Man, , nucleotides, where the performance of conventional
1050 DELAYED EXTRACTION MALDI TOF MASS SPECTROMETRY
MALDI has not been nearly as good as it is for Zagulin, Souiet Phys. JETP 37, 45-48 (1973).
polypeptides. Work on the application of DE-MALDI
to DNA sequencing is in progress and results will be
;: R. C. Beavis and B. T. Chait, Chem. Phys. Lett. 181,479 (1991).
J. Zhou, W. Ens, K. Standing and A. Verentchikov, Rapid
Commun. Mass Spectrom. 6,671-678 (1992).
presented elsewhere. 10. W. C. Wiley and I. H. McLaren, Rev. Sci. Instrum. 26, 1150-
1157 (1953), W. C. Wiley, U.S. Patent 2,685,035.
11. R. S. Brown, J. J. Lennon and D. Christie, Desorption '94-Mass
REFERENCES Spectrometry of Large Organic Ions by Particle and Photon
1. R. D. MacFarlane, R. P. Skowronski and D. F. Torgerson, Induced Desorption, March 27-31, 1994, Sunriver OR, USA,
Biochem. Biophys. Res. Commun. 60,616-624 (1974). p. 63.
2. M. Karas, D . Bachmann, U. Bahr and F. Hillenkamp, Int. J. 12. R. S. Brown and J. J. Lennon, Anal. Chem. 67, 1998-2003
Mass Spectrom. Ion Processes 78, 53-68 (1987). (1995).
3. K. Tanaka, H. Waki, Y. Ido and S. Akita, Rapid Commun. Mass 13. S. M. Colby, T. B. King and J. P. Reilly, Rapid. Commun. Mass
Spectrom. 3 , 151-153 (1988). Spectrom. 8, 865-868 (1994).
4. F. Hillenkamp and M. Karas, Methods in Enzymol. 193,280-295 14. R. M. Whittal and L. Li, Anal. Chem. 67, 1950-1954 (1995).
(1990). 15. These are commercially available analyzers from Perseptive
5. 0. A. Mirgorodskaya, A. A. Shevchenko, I. V. Chernushevich, Biosystems, Framingham MA, USA.
A. F. Dodonov and A. I. Miroshinikov, Anal. Chem. 66,99-107 16. M. Karas, U. Bahr, K. Strupat, F. Hillenkamp, A.
(1994). Tsourbopoulos and B. N. Pamanik, Ann. Chem. 67, 675-679
6. A. N. Verentchikov, W. Ens and K. G. Standing, Anal. Chem. (1995).
66, 126-133 (1994). 17. R. S. Annan, J. A. Hill, H. Koechling and K. Biemann, Rapid
7. B. A. Mamyrin, V. J. Karatajev, D. V. Smikk and V. A. Commun. Mass Spectrom. 6,298-302 (1992).