Ontario Ambulance Services Provincial Equipment Standards 2023 Version
Ontario Ambulance Services Provincial Equipment Standards 2023 Version
Provincial Equipment
Standards for Ontario
Ambulance Services
Version 3.7
Comes into force
February 1, 2023
Emergency Health
Regulatory and
Accountability
Branch
Ministry of Health
To all users of this publication:
The information contained in this standard has been carefully compiled and is believed to
be accurate at date of publication.
For further information on the Provincial Equipment Standards for Ontario Ambulance
Services, please contact:
ParamedicStandards@ontario.ca, or
Document Control
Preamble
Preface
Part A of this document contains the following equipment lists:
1. List 1 – Land Ambulance/ERV Accessory Equipment
List 1(a) – Land Ambulance Accessory Equipment
List 1(b) – ERV Accessory Equipment
2. List 2 – Primary Care Paramedic Land Ambulance Equipment and Medication
3. List 3 – Primary Care Paramedic ERV Equipment
List 3(a) – ERV Responder Equipment and Medication
List 3(b) – ERV Support/Command Equipment
4. List 4 – Advanced Care Paramedic Land Ambulance/ERV Equipment and
Medication
5. List 5 – Land Ambulance Auxiliary Equipment and Medication
Lists 1 - 4 specify the minimum quantities of each piece of equipment that are
required to be carried on a land ambulance or emergency response vehicle to
provide care for a minimum of two (2) patients, and to transport a minimum of one
(1) patient. The lists also identify standards that apply to specific equipment items.
Where a standard has been developed by Emergency Health Regulatory and
Accountability Branch (EHRAB), the applicable EHRAB Equipment Standards
Reference Number has been provided. The lists, in some cases, reference standards
developed by an external organization. In such cases, the equipment identified shall
meet or exceed these referenced standards.
Part B and C of this document contain the actual equipment standards for those
items in Part A identified as having to meet minimum standards developed by
EHRAB.
4. Ambulance service operators shall ensure that all electrical medical and
accessory equipment being operated, either on 110/120 volt (AC) or low
voltage, have been approved or inspected for use by CSA or C-UL.
5. Ambulance service operators shall ensure that controlled substances
(opiates and benzodiazepines) are stored in different carrying cases than
other medications.
Modifications
1. Modifications shall not be made that would adversely affect the safety or
other performance characteristics of any piece of equipment.
Documentation from the original equipment manufacturer shall be kept on
file confirming the acceptability of any modifications.
2. Any accessory components added to a piece of equipment (e.g. brackets,
shelves, etc.) shall be able to support a minimum of ten (10) times the
weight of the component plus any item it is intended to hold or carry.
Equipment Manuals
Manufacturer’s equipment manuals shall be readily available as local reference
documents. These references will assist quality assurance personnel, management
staff, paramedics and review teams in determining the appropriate use,
maintenance and compliance of applicable medical and accessory equipment.
Materials
Latex
Equipment used in an ambulance or emergency response vehicle shall be latex-
free, where available.
Hypoallergenic Materials
Equipment shall be hypoallergenic, where available.
Gauge G
Gram g
International Standards Organization ISO
Intravenous IV
Kilogram kg
Metre m
Metered Dose Inhaler MDI
Microgram mcg
Milliequivalent mEq
Millilitre ml
Millimetre mm
Multi-casualty incident MCI
Ministry of Health MOH
Pound lbs
Pounds per square inch psi
Primary Care Paramedic PCP
Underwriters Laboratories of Canada C-UL
United States National Institute for Occupational NIOSH
Safety and Health
United States Pharmacopoeia USP
Water H20
Definitions
Approximately
means a range within 10% (plus or minus) of the specified figure.
Disposable
means a product that is designed for a single use or application.
Reusable
means a product that is designed for multiple uses or application.
Sterile
condition of being aseptic or free from live bacteria or other microorganisms
especially pathogenic microorganisms and their spores.
Part A – Land
Ambulance/ERV
Equipment
List 1 – Land Ambulance/ERV
Accessory Equipment
List 1 (a) – Land Ambulance Accessory Equipment
Equipment Description EHRAB Minimum
Standard Quantity
No.
1. radio equipment: type and power, approved by the N/A 1
Director
Immobilization Equipment
8. cervical collars, adult 070 2 each size
or
equivalent
multi-size
9. cervical collars, pediatric 070 2 each size
or
equivalent
multi-size
10. spinal board, quick connect 325 1
Blankets/Linens
78. blanket N/A 2
79. blankets, disposable N/A 2
80. pillow N/A 2
81. pillow case N/A 2
82. sheets N/A 2
83. towel N/A 4
Miscellaneous
84. bed pan N/A 1
85. emesis bag N/A 4
86. tissues, facial N/A 2 boxes
87. tissue, toilet N/A 1 roll
88. urinal N/A 1
Immobilization Equipment
7. spinal immobilization extrication device 330 1
8. splint, multi-purpose/malleable 335 2
9. splint, traction 340 1
Intravenous Supplies
21. alcohol preps N/A 10
22. dressing, clear sterile N/A 6
23. intraosseous (IO) needles 170 2 each size
24. intravenous catheters 175 2 each size
25. intravenous drip tubing 180 2
26. intravenous drip tubing, pediatric 181 2
27. intravenous pressure infuser 185 1
28. intravenous solution (0.9% Sodium Chloride) 190 2 litres, any
combination
29. needles 240 2 each size
30. saline flush solutions N/A 2
31. syringe, medical 405 2 each size
32. tourniquet N/A 2
Medication
33. adenosine (6 mg and/or 12 mg preparations) N/A 36 mg total
34. atropine sulphate injection N/A 2 mg
35. calcium gluconate (1 g/10 mL = 10%) N/A 4g
36. 10% dextrose in water injection (10 g/100 ml) N/A 4
OR
50% dextrose in water injection (25 g/50 ml)
37. dopamine HCL solution (200 mg/250 ml) OR N/A 2 bags or 2
dopamine HCL injection (200 mg/5 ml)
38. epinephrine 0.1 mg/ml = 1:10,000 N/A 10 mg
39. ketamine N/A 1000 mg
40. lidocaine injection OR amiodarone injection N/A 600 mg or
600 mg
41. lidocaine spray (10 mg/spray) with 2 spray N/A 2
nozzles/canister
42. midazolam injection N/A 20 mg
43. morphine sulphate injection N/A 40 mg
44. sodium bicarbonate injection (50 mEq/50 ml) N/A 2
Equipment
1. Continuous Positive Airway Pressure Unit A-910 N/A
2. Endotracheal Tube – Nasal A-912 1 each size
3. Meconium Aspirator Adapter A-915 N/A
Medication
6. fentanyl N/A 400 mcg
7. furosemide N/A 200 mg
8. topical antibiotic N/A sufficient
quantities to
treat 2
patients
9. xylometazoline 0.1% N/A 2 bottles
Part B – Equipment
Standards
Reference
N/A
Reference
ANSI Z79.3 Anaesthetic Equipment-Oropharyngeal and Nasopharyngeal Airways
Reference
ANSI Z79.3 Anaesthetic Equipment-Oropharyngeal and Nasopharyngeal Airways
Reference
N/A
Minimum requirements
Intranasal atomization device shall:
1. be constructed with a soft, conical plug;
2. be constructed to form a seal within the nostril and medication being
delivered;
3. be capable of delivering atomized drugs into a fine mist of particles (<100
microns in size);
4. be individually packaged;
5. be for single patient use;
6. be in a sealed package, capable of maintaining the clean integrity of the
atomizer under normal conditions of shipping and storage; and;
7. attach to or is part of a device (e.g. syringe) that allows appropriate unit
dose measurements.
Bag-Valve-Mask Resuscitator,
Adult - No. 015
Minimum requirements
Bag-Valve-Mask resuscitators shall include all of the following components:
1. one (1) adult ventilation bag with a minimum capacity sufficient to
adequately ventilate a large adult;
2. one (1) non-rebreathing valve with single diaphragm;
3. one (1) intake valve;
4. one (1) way exhalation port that vents expired gases to the atmosphere;
5. one (1) reservoir assembly with a capacity equal to or greater than the
ventilation bag capacity.
Reference
AARC Guidelines, September 2004, Vol 49/9
AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular
Care, 2015
Bag-Valve-Mask Resuscitator,
Pediatric - No. 020
Minimum requirements
Bag-valve-mask resuscitators shall include of the following components:
1. one (1) child ventilation bag with a minimum capacity sufficient to
adequately ventilate a pediatric patient;
2. pop-off pressure relief valve;
3. one (1) patient non-rebreathing valve with single diaphragm;
4. one (1) intake valve;
5. one (1) reservoir assembly with a capacity equal to or greater than the
ventilation bag capacity.
9. have an integral oxygen stem attached to the inlet valve which shall allow
gas to flow into the intake valve when used with oxygen and vent excess
oxygen to the atmosphere;
10. have a non-rebreathing valve which includes a single diaphragm to
prevent rebreathing of the patient’s exhaled gas;
11. allow the patient to breathe freely from the atmosphere should the device
become occluded;
12. not allow the non-rebreathing valve to jam despite temperature extremes,
moisture, or supplemented oxygen inflow as great as 15 litres per minute;
13. have a valve assembly capable of releasing excess pressure within the
bag;
14. resist jamming due to the presence of debris;
15. be easy to disassemble and clean, with components designed to prevent
incorrect assembly when reusable equipment is utilized;
16. not exceed 10 ml valve dead space;
17. have a dead space for the valve and mask not exceeding 175 ml;
18. supply oxygen concentrations in excess of 40% with supplemental oxygen
when used without an oxygen reservoir;
19. maintain an oxygen concentration not less than 100% with the reservoir
assembly attached and supplied with an oxygen inflow of 12 litres per
minute, a tidal volume of 300 ml and a cycle rate of 12 cycles per minute.
Reference
AARC Guidelines, September 2004, Vol 49/9
AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular
Care, 2015
Reference
ISO 81060 – 1:2007
Reference
ISO 81060 – 1:2007
Reference
N/A
Cardiac Monitor/Defibrillator
and Patient Monitoring Devices
(ACP) - No. 040
Minimum requirements
Cardiac monitor/defibrillators and patient monitoring devices shall:
1. be capable of patient monitoring, defibrillation, and synchronized
cardioversion;
2. have monitoring technologies with diagnostic measurements for S-T
segment analysis for 12-lead ECG acquisition;
3. be able to monitor oxygen saturation of arterial hemoglobin (SpO 2);
4. be able to monitor end-tidal CO2, including numeric value and waveform;
5. be programmable to energy settings for defibrillation, transcutaneous
pacing and synchronized cardioversion as described in the current MOH
Advanced Life Support Patient Care Standards;
6. be able to automatically analyze electrocardiographic rhythms to
determine if defibrillation is required;
7. be easy to clean and disinfect;
8. be water-resistant;
9. be capable of ‘3’, ‘4’ or ‘5’ lead ECG monitoring through electrodes and/or
ECG monitoring through defibrillation pads;
10. have data collection capability and provide specific event summary
reports;
11. have a strip chart printer and ECG paper designed for the device;
12. have internal operational checks;
13. have individual, adjustable volume of QRS beeper, voice prompts, and
other standard alerts;
14. have both AC and DC power modules;
15. have one spare battery;
16. have a software configuration mode;
17. have a diagnostic mode;
Reference
ANSI/AAMI IEC 60601-2-4:2010
ANSI/AAMI EC13:2002 (R2007)
Cardiac Monitor/Defibrillator
and Patient Monitoring Devices
(PCP) - No. 045
Minimum requirements
Cardiac monitor/defibrillators and patient monitoring devices shall:
1. be capable of patient monitoring and defibrillation;
2. have monitoring technologies with diagnostic measurements for S-T
segment analysis for 12-lead ECG acquisition;
3. be able to monitor oxygen saturation of arterial hemoglobin (SpO 2);
4. be programmable to energy settings for defibrillation as described in the
current MOH Advanced Life Support Patient Care Standards;
5. be able to automatically analyze electrocardiographic rhythms to
determine if defibrillation is required;
6. be easy to clean and disinfect;
7. be water-resistant;
8. be capable of ‘3’, ‘4’ or ‘5’ lead ECG monitoring through electrodes and/or
ECG monitoring through defibrillation pads;
9. have data collection capability and provide specific event summary
reports;
10. have a strip chart printer and ECG paper designed for the device;
11. have internal operational checks;
12. have individual, adjustable volume of QRS beeper, voice prompts, and
other standard alerts;
13. have both AC and DC power modules;
14. have one spare battery;
15. have a software configuration mode;
16. have a diagnostic mode;
17. be equipped with a carrying case.
Reference
ANSI/AAMI IEC 60601-2-4:2010
ANSI/AAMI EC13:2002 (R2007)
Cardiac Monitor/Defibrillator,
Defibrillation Pads - No. 050
Minimum requirements
Defibrillation pads shall:
1. be disposable;
2. be soft and flexible;
3. be able to function in a variety of temperature and environmental
conditions;
4. be constructed of a thin flexible conductor, protective polymer backing
and a hydro-gel adhesive;
5. provide delivery of current for defibrillation, transcutaneous pacing and
synchronized cardioversion;
6. be soft, flexible and oval-shaped to conform to body contours for larger
surface contact and adhesion;
7. serve various functions: ECG monitoring, defibrillation, transcutaneous
pacing and synchronized cardioversion;
8. be individually packaged;
9. be labeled with size.
Reference
ANSI/AAMI DFSO:2003
ANSI/AAMI/ISO EC13:2002 (R) 2007
Cardiac Monitor/Defibrillator,
ECG Cable - No. 055
Minimum requirements
ECG cables shall:
1. provide ‘3’, ‘4’ or ‘5’ lead ECG monitoring through electrodes;
2. provide 12-lead ECG monitoring through electrodes;
3. be easy to clean and disinfect;
4. be water-resistant;
5. be colour-coded and/or labeled to designate appropriate limb
designation.
Reference
ANSI/AAMI EC53:2002 (R) 2007
Cardiac Monitor/Defibrillator,
ECG Monitoring Electrodes - No.
060
Minimum requirements
ECG monitoring electrodes shall:
1. be capable of monitoring neonatal, pediatric and adults patients;
2. be disposable;
3. have fluid resistant adhesive;
4. be capable of interfacing with ECG cable;
5. have a conductive gel centre;
6. provide for general purpose monitoring and diagnostics;
7. be flexible and conform to body surfaces;
8. be sealed in light resistant packaging.
Reference
ANSI/AAMI EC53:2002 (R) 2007
Reference
ASTM F1559-94 (2007)
Reference
N/A
Reference
Canadian Community Health Survey
Reference
Canadian Community Health Survey
Reference
N/A
1
See p.6 for details. N.B. Reusable equipment should be cleaned, disinfected,
maintained, and cared for in accordance with the manufacturer’s instructions.
Reference
N/A
Reference
Reference
ISO 5361:1999
Reference
N/A
Reference
N/A
Reference
CSA Standard Z94.3 Eye and Face Protectors
ANSI/ISEA Z87.1 Personal Eye and Face Protection Devices
Reference
CSA Standard Z94.3 Eye and Face Protectors
ANSI/ISEA Z87.1 Personal Eye and Face Protection Devices
The case, or combination of cases, shall contain the following oxygen and
accessory equipment:
Equipment Description EHRAB Minimum
Standard Quantity
No.
1. airway, oropharyngeal 010 1 each
size
2. bag-valve-mask resuscitator, adult 015 1
3. bag-valve-mask resuscitator, pediatric 020 1
4. hydrophobic submicron filter 160 1
5. blood pressure cuff/manometer (manual), adult 025 1
6. blood pressure cuff/manometer (manual), pediatric 030 1
7. oxygen cylinder, portable 247 1
8. oxygen mask, adult high concentration 255 1
9. oxygen mask, pediatric simple 265 1
10. oxygen nasal cannula 270 1
11. oxygen pressure regulator, first response kit 275 1
12. resuscitation mask, adult 310 1
13. stethoscope N/A 1 or
personal
issue
The case, or combination of cases, shall contain the following first aid supplies:
Equipment Description EHRAB Minimum
Standard Quantity
No.
1. bandage, conforming gauze roll N/A 2
2. bandage, triangular N/A 6
3. cervical collar 070 1 each
size or
multi-size
4. dressing, hemostatic 092 2
5. dressing, pressure N/A 2
6. eye pad N/A 2
7. gauze pad N/A 4
8. gloves, non-sterile 130 2 pairs
9. hand rub, antiseptic 150 1
10. irrigation fluid 195 Minimum
1,000 ml
11. tape, adhesive N/A 2 rolls
12. tourniquet, arterial 011 1
4. allow the kit to remain stable when placed on the ground or other flat
surfaces and should not tip over easily;
5. be easily cleaned and disinfected;
6. remain water resistant under normal operating conditions.
Reference
N/A
Reference
ISO 11193-2:2006
Reference
ISO 10282
Reference
ISO 15197:2003
Diabetes Care (2000) 23:1143 – 1148
Reference
Infection Prevention and Control Best Practices Manual for Land Paramedics,
Version 1.0 – March 2007
Reference
ISO 23328-1:2003
Reference
AAMS Pediatric Restraint Position Paper, October 2007
Reference
O. Reg. 474/07: Needle Safety under the Occupational Health and Safety Act
Reference
ISO 10555-5:1996
Reference
N/A
Reference
N/A
Reference
N/A
Reference
ISO 15747:2003
Reference
The United States Pharmacopeia (USP)
Reference
ISO 7376:2003
Reference
ISO 7176-24:2004
Canadian Community Health Survey
Reference
ISO 7176-24:2004
Canadian Community Health Survey
Reference
ASTM F1638-95(2008) E1
Reference
N/A
Reference
ISO 5361:1999
Reference
ISO/DIS 27427:2009
Reference
ISO/DIS 27427:2009
Reference
ISO 9626:1991
O. Reg. 474/07: Needle Safety under the Occupational Health and Safety Act
Reference
AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular
Care, 2015
Reference
N/A
Reference
N/A
Reference
N/A
Reference
ISO 15001:2003
Reference
ISO 15001:2003
Reference
ISO 15001:2003
Reference
ISO 15001:2003
NOTE: Oxygen regulators intended for aeromedical use shall only be used in
conjunction
with devices intended and tested for the aeromedical environment.
Reference
ASTM G175-03
CGA
Reference
ASTM G175-03
CGA
Reference
ISO 15001:2003
Reference
N/A
2
See p.6 for details. N.B. Reusable equipment should be cleaned, disinfected,
maintained, and cared for in accordance with the manufacturer’s instructions.
Operators shall have a redundancy plan in place should a powered or reusable
respirator fail.
Reference
N/A
Reference
Canadian Community Health Survey
Reference
Canadian Community Health Survey
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 100
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 15001:2003
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 101
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 15001:2003
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 102
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO/NP 23908-3
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 103
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ASTM F1557-94(2007)
Canadian Community Health Survey
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 104
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ASTM F1556-94(2007)
Canadian Community Health Survey
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 105
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Splint, Multi-purpose/Malleable
- No. 335
Minimum requirements
A multi-purpose splint shall:
1. be capable of being cleaned and disinfected, or be disposable;
2. be constructed of rigid material and have rounded edges;
3. be adjustable for arm or lower leg use in adults and arm or leg use in
children;
4. be x-ray translucent;
5. be individually packaged with easy to understand instructions for use;
6. under normal use conditions, support and stabilize injured extremities and
maintain alignment in as near neutral position as possible, or maintain the
position of splinting.
Reference
ASTM F1555-94 (2007)
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 106
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ASTM F1555-94 (2007)
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 107
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
Canadian Community Health Survey
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 108
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 109
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 8836:2007
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 110
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 8836:2007
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 111
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 8836:2007
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 112
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 8836:2007
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 113
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 114
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 115
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
With regards to batteries and electrical power supply for the suction unit:
1. If the integral battery is rechargeable, it shall be rechargeable from the
ambulance’s 12V DC electrical system and from 115V AC. The 115-V AC
charging system need not be integral to the portable suction unit.
2. The battery shall be rechargeable while installed in the aspirator case.
Batteries shall not be physically disconnected in order to initiate their
recharge.
3. External battery chargers shall be CSA approved for “Electro-medical Use”
and bear the applicable CSA file number.
4. The suction unit shall include electrical cords and plugs that will interface
within the ambulance conversion design.
Reference
ANSI / ISO10079.1
AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular
Care, 2015
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 116
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 117
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ANSI / ISO10079.1
AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular
Care, 2015
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 118
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Shall contain the following medications and equipment; along with any auxiliary
medications as applicable:
Equipment Description EHRAB Minimum
Standard Quantity
No.
Medications
1. acetaminophen (960 – 1,000 mg/dose) N/A 1920 mg
2. acetylsalicylic acid (ASA) (80 or 81 mg/tablet) N/A 6 tablets
3. dexamethasone N/A 16 mg
4. dimenhydrinate N/A 100 mg
5. diphenhydramine N/A 100 mg
6. epinephrine 1 mg/ml = 1:1,000 N/A 10 mg
7. glucagon N/A 2 mg
8. glucose, oral (paste, tablets or other formulation) N/A 2
15 g/dose
9. ibuprofen (400 mg/dose) N/A 800 mg
10. ketorolac (10-15 mg/dose) N/A 2
vials/amps
11. naloxone N/A 12 mg
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 119
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Equipment
17. alcohol preps N/A 10
18. atomization device, intranasal 012 2
19. glucometer 145 1
20. metered dose inhaler (MDI), valved holding 230 2
chamber
21. needles 240 2 each size
22. syringe, medical 405 2 each size
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 120
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 7886-1:1993
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 121
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 80601-2-56: 2009
ASTM E1965-98
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 122
Part B – Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 123
Part B – Equipment Standards
Part C – Auxiliary
Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Part C – Auxiliary
Equipment Standards
Continuous Positive Airway
Pressure Unit - No. A-910
Minimum requirements
Continuous positive airway pressure (CPAP) units shall:
1. be driven/powered by oxygen with an anti-suffocation valve;
2. have independent adjustable settings CPAP pressure from 5 – 10 cm H 2O;
3. have an airway pressure relief mechanism;
4. have the ability to provide a minimum FIO2 of 50% with optional FIO2 of
100% oxygen;
5. use a disposable circuit including mask and head harness for CPAP
delivery;
6. include oxygen hose/tubing that will allow the device to be attached to
an oxygen source at a minimum distance of two (2) metres;
7. have the ability to provide nebulized and/or metered dose inhaler
medication;
8. have a method for attaching a hydrophobic submicron filter to decrease
the risk of airborne contagion.
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 125
Part C – Auxiliary Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
ISO 5361:2012
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 126
Part C – Auxiliary Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
N/A
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 127
Part C – Auxiliary Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
Canadian Community Health Survey
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 128
Part C – Auxiliary Equipment Standards
Emergency Health Regulatory and Accountability Branch, Ontario Ministry of Health
Reference
Canadian Community Health Survey
Provincial Equipment Standards for Ontario Ambulance Services – Version 3.7 129
Part C – Auxiliary Equipment Standards
This page is intentionally left blank.