PT. ASABA
SURVEYING DIVISION
a
! USER NAME.
i PT: BENUA BTAM JAVA MANDIR
i
! Reference No.
i Date issued
! Certificate of Calibration
NON PRISM TOTAL STATION
Instrument name IM 103,
Serial No. 28003100
Inspection date ‘September 5, 2023
Inspector name — { Indra
“This certifies that the above named instrument has been calibrated by PT. ASABA = Authorized SOKKIA
This coreg tha standard traceable to Japanese surveying manufacture Assocation: JSIMA-103-2002, —
‘according to SOKKIA Service Manual,
(Q) Theodotite
‘No. Trem Result ‘Max. tolerance
1 | Tetesone ‘ux deviation of teleoope eis Yt = OS eo a
Perpendicularity of telescope axis and
2 | Mechanism horizontal axis (Collimation) fe) seas eee
Horizontality of horizontal axis (Inclination) 4 seo. Tess than 10 sec.
3] H Encoder H Eccentricity 4 sec. ess than $ sec.
‘Vertical eccentricity 2 sec. Tess than 5 sec.
| Neca oe ‘Vertical constant difference 3 sec. Tess than_10 sec.
‘Vertical zero point calculation 1 seo. Jess than _15 sec.
> | Astomatie Compensator eror (+3 min) 2a. [ess than 6 sec.
| 5 | compensator Compensator error (3 min) sec. | less than 6 sec. 1
(2) Distance Meter
@-1) Laser power
Ne] Testitem Real Tlennse Standard
[ea Laser power 4.40mW = 0.10mW. ‘class 1, TEC60825-1 ED.2.0:2007
(2-2) Accuracy Target: APOL
[Whe ‘Nominal distance Measured distance Result Tolerance @(@2+2ppmxD)mm
1 10.692 m 10,690 m, 2m Tess ian = 2.022 mm
2 57.334 m_ 57.332 m 2mm Jess than + 2.115 mm
3 [169.103 m 169,101 m 2mm Tess than 2.338 mm,
G 754.460 m 754.458 m 2mm Tess than = 3.508 mm.
‘The inspection of EDM complies with Japanese Standard JSIMA-102 : 2006
PT ASABA isnot responsible for instrument ou of tolerance
Calibration interval: (six) months
Recommended nex calibration : March 05,2024
PT, ASABA
JL. IR. H. Juand7 Ne. 7, Jakarta, Indonesia | Telp. +6221 - 8066 4499 Fax : +6221 - 380 6174