You are on page 1of 7

1740 Diabetes Care Volume 46, October 2023

Executive Summary: Guidelines David B. Sacks,1 Mark Arnold,2


George L. Bakris,3 David E. Bruns,4
and Recommendations for Andrea R. Horvath,5 Åke Lernmark,6
Boyd E. Metzger,7 David M. Nathan,8 and
Laboratory Analysis in the M. Sue Kirkman 9
EXECUTIVE SUMMARY

Diagnosis and Management of


Diabetes Mellitus

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023


Diabetes Care 2023;46:1740–1746 | https://doi.org/10.2337/dci23-0048

1
Department of Laboratory Medicine, National
Institutes of Health, Bethesda, MD
2
Department of Chemistry, University of Iowa, Iowa
City, IA
3
Department of Medicine, American Heart
Association Comprehensive Hypertension Center,
BACKGROUND Section of Endocrinology, Diabetes and Metabolism,
Numerous laboratory tests are used in the diagnosis and management of patients University of Chicago Medicine, Chicago, IL
4
Department of Pathology, University of Virginia
with diabetes mellitus. The quality of the scientific evidence supporting the use
Medical School, Charlottesville, VA
of these assays varies substantially. An expert committee compiled evidence- 5
New South Wales Health Pathology Department
based recommendations for laboratory analysis in patients with diabetes. The of Chemical Pathology, Prince of Wales Hospital,
overall quality of the evidence and the strength of the recommendations were Sydney, Australia
6
Department of Clinical Sciences, Lund University/
evaluated. The draft consensus recommendations were evaluated by invited re- CRC, Skane University Hospital Malm€ o, Malm€o,
viewers and presented for public comment. Suggestions were incorporated as Sweden
7
deemed appropriate by the authors (see Acknowledgments in the full version of Division of Endocrinology, Northwestern University,
the guideline). The guidelines were reviewed by the Evidence Based Laboratory The Feinberg School of Medicine, Chicago, IL
8
Massachusetts General Hospital Diabetes Center
Medicine Committee and the Board of Directors of the American Association for and Harvard Medical School, Boston, MA
Clinical Chemistry and by the Professional Practice Committee of the American 9
Department of Medicine, University of North
Diabetes Association. Carolina, Chapel Hill, NC
Corresponding author: David B. Sacks, sacksdb@
CONTENT mail.nih.gov
Diabetes can be diagnosed by demonstrating increased concentrations of glucose Received 19 April 2023 and accepted 12 May
in venous plasma or increased hemoglobin A1c (HbA1c) in the blood. Glycemic 2023
control is monitored by the patients measuring their own blood glucose with me- This document was approved by American
ters and/or with continuous interstitial glucose monitoring devices and also by Association for Clinical Chemistry’s (AACC’s)
Evidence Based Laboratory Medicine Subcommittee
laboratory analysis of HbA1c. The potential roles of noninvasive glucose monitor- in January 2023, the AACC Academy Council in
ing; genetic testing; and measurement of ketones, autoantibodies, urine albu- February 2023, the AACC Board of Directors
min, insulin, proinsulin, and C-peptide are addressed. in March 2023, and the American Diabetes
Association in March 2023.
SUMMARY The full “Guidelines and Recommendations
The guidelines provide specific recommendations based on published data or de- for Laboratory Analysis in the Diagnosis and
Management of Diabetes Mellitus” can be
rived from expert consensus. Several analytes are found to have minimal clinical accessed online at https://doi.org/10.2337/
value at the present time, and measurement of them is not recommended. dci23-0036.
This article has been co-published with permission
in Diabetes Care by the American Diabetes
Diabetes mellitus is a group of metabolic disorders of carbohydrate metabolism in Association and in Clinical Chemistry by Oxford
University Press on behalf of the American
which glucose is both underutilized and overproduced, resulting in hyperglycemia. The
Association for Clinical Chemistry. © This work is
disease is classified conventionally into several clinical categories. Type 1 diabetes mel- written by (a) U.S. Government employee(s) and
litus is usually caused by autoimmune destruction of the pancreatic islet b-cells, ren- is in the public domain in the U.S. (2023).
dering the pancreas unable to synthesize and secrete insulin (1). Type 2 diabetes The articles are identical except for minor stylistic
mellitus results from a combination of insulin resistance and inadequate insulin differences in keeping with each journal’s style.
diabetesjournals.org/care Sacks and Associates 1741

secretion (2,3). Gestational diabetes melli- guideline committee have disclosed any users to adapt recommendations to local
tus (GDM) develops during approximately financial, personal, or professional rela- settings.
17% of pregnancies, usually remits after tionships that may constitute conflicts of
delivery, and is a major risk factor for the interest with this guideline and received RECOMMENDATIONS
development of type 2 diabetes later in no direct funding related to the develop- Capital letters denote the grade of recom-
life. Type 2 is the most common form of di- ment of the recommendations. The per- mendations and categories in parenthe-
abetes, accounting for 85% to 95% of dia- spectives and views of various international ses refer to the quality of the underlying
betes in developed countries. and national organizations, as well as body of evidence supporting each recom-
Diabetes is a common disease. World- other potential stakeholders (e.g., health mendation. The grading system is de-
wide prevalence in 2021 was estimated to care providers, people with diabetes, scribed in Tables 1 and 2.
be approximately 537 million (10.5% of policymakers, regulatory bodies, health
the global population) and is forecast to insurance companies, researchers, and 1. Glucose
reach 783 million by 2045 (4). Based on industry), were taken into account dur-
2017–2020 National Health and Nutrition a. Fasting glucose should be measured in
ing the public consultation process. The

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023


Examination Survey data and 2018–2019 venous plasma when used to establish
system developed in 2011 to grade both the diagnosis of diabetes, with a value
National Health Interview Survey data, the the overall quality of the evidence (Table 1)
U.S. Centers for Disease Control and Pre- >7.0 mmol/L (>126 mg/dL) diagnostic
and the strength of recommendations of diabetes. A (high)
vention estimated that there were 37.3 (Table 2) was used, and the key steps
million people (11.3% of the U.S. popula- b. Screening by hemoglobin A1c (HbA1c),
and evidence summaries are detailed in fasting plasma glucose (FPG), or 2-h
tion) with diabetes (5). The number of the guideline and in the Supplementary
adults with diabetes has also increased in oral glucose tolerance test is recom-
Material that accompanies the online mended for individuals who are at
other parts of the world. For example, version of this report (13). The literature
China and India were thought to have high risk of diabetes. If HbA1c is <5.7%
was reviewed to the end of 2021. (<39 mmol/mol), FPG is <5.6 mmol/L
140.9 and 74.2 million adults with diabetes This guideline focuses on the practical
in 2021 and are expected to have 174.4 (<100 mg/dL), and/or 2-h plasma glu-
aspects of care in order to assist with deci- cose is <7.8 mmol/L (<140 mg/dL),
and 124.9 million, respectively, by 2045
sions regarding the use or interpretation testing should be repeated at 3-year
(4). Approximately 45% of people with dia-
of laboratory tests while screening, diag- intervals. B (moderate)
betes worldwide are thought to be undiag-
nosing, or monitoring patients with diabe- c. Glucose should be measured in venous
nosed (4).
tes. It covers the rationale, preanalytical, plasma when used for screening of
The cost of diabetes in the U.S. in 2012
analytical, postanalytical, and, where ap- high-risk individuals. B (moderate)
was approximately $245 billion and in-
plicable, emerging considerations, which d. Plasma glucose should be measured in
creased to $327 billion by 2017 (6). The
alert the reader to ongoing studies and an accredited laboratory when used
mean annual per capita health care costs
potential future aspects relevant to each for diagnosis of or screening for diabe-
for an individual with diabetes are approx-
analyte. The recommendations intend to tes. Good Practice Point (GPP)
imately 2.3-fold higher than those for indi-
viduals who do not have diabetes (7). supplement the American Diabetes Asso- e. Routine measurement of plasma glucose
Worldwide spending in 2021 was thought ciation guidelines and thus do not address concentrations in a laboratory is not rec-
to be $966 billion. The high costs of diabe- any issues related to the clinical manage- ommended as the primary means of
tes are attributable primarily to the chronic ment of patients. The full version of this monitoring or evaluating therapy in indi-
debilitating microvascular and macrovascu- guideline and its accompanying supple- viduals with diabetes. B (moderate)
lar complications (6), which make diabetes ments are available as a Special Report f. Blood for fasting plasma glucose analy-
the fourth most common cause of death in (13). Key recommendations are summarized. sis should be drawn in the morning af-
the developed world (8). About 6.7 million These recommendations primarily tar- ter the subject has fasted overnight (at
adults worldwide were thought to have get laboratory professionals, general prac- least 8 h). B (low)
died of diabetes-related causes in 2021 (4). titioners, physicians, nurses, and other g. To minimize glycolysis, a tube contain-
The American Association for Clinical health care practitioners involved in the ing a rapidly effective glycolytic inhibi-
Chemistry and American Diabetes Associ- care of people with diabetes. The guide- tor such as granulated citrate buffer
ation issued “Guidelines and Recommen- lines can be used by individuals with dia- should be used for collecting the sam-
dations for Laboratory Analysis in the betes (where relevant, e.g., self-monitoring ple. If this cannot be achieved, the
Diagnosis and Management of Diabetes of blood glucose), policymakers, and payers sample tube should immediately be
Mellitus” in 2002 (9,10) and 2011 (11,12). for health care, as well as by researchers placed in an ice-water slurry and sub-
Here we review and update these recom- and manufacturers. Although recommen- jected to centrifugation to remove the
mendations using an evidence-based ap- dations were developed for national and cells within 15–30 min. Tubes with
proach, especially in those areas where international use and are intended to be only enolase inhibitors such as sodium
new evidence has emerged since the 2011 generic, certain recommendations may fluoride should not be relied on to pre-
publications (13). The guideline commit- not reflect views that are universally held vent glycolysis. B (moderate)
tee, whose membership was predomi- or may have limited applicability in health h. Based on biological variation, glucose
nantly from the U.S., included clinical, care settings with differing organizational, measurement should have analytical im-
laboratory, and evidence-based guideline cultural, and economic backgrounds. The precision #2.4%, bias #2.1%, and total
methodology experts. Members of the guideline committee therefore advises error #6.1%. To avoid misclassification
1742 Executive Summary Diabetes Care Volume 46, October 2023

Table 1—Rating scale for the quality of the evidence

High: Further research is very unlikely to change our confidence in the estimate of effect. The body of evidence comes from high high-level
individual studies which are sufficiently powered; provide precise, consistent, and directly applicable results in a relevant population.
Moderate: Further research is likely to have an important impact on our confidence in the estimate of effect and may change the estimate
and the recommendation. The body of evidence comes from high/moderate moderate-level individual studies which are sufficient to
determine effects, but the strength of the evidence is limited by the number, quality, or consistency of the included studies;
generalizability of results to routine practice; or indirect nature of the evidence.
Low: Further research is very likely to have an important impact on our confidence in the estimate of effect and is likely to change the estimate
and the recommendation. The body of evidence is of low level and comes from studies with serious design flaws, or evidence is indirect.
Very low: Any estimate of effect is very uncertain. Recommendation may change when higher quality evidence becomes available. Evidence
is insufficient to assess the effects on health outcomes because of limited number or power of studies, important flaws in their design or
conduct, gaps in the chain of evidence, or lack of information.

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023


of individuals, the goal for glucose analy- 2. Glucose Meters b. Frequent blood glucose monitoring
sis should be to minimize total analytical a. Portable glucose meters should not be (BGM) is recommended for all people
error, and methods should be without used in the diagnosis of diabetes, includ- with diabetes who use intensive insulin
measurable bias. B (moderate) ing gestational diabetes. B (moderate) regimens (with multiple daily injections

Table 2—Grading the strength of recommendations

A. STRONGLY RECOMMEND
a. adoption when:
 There is high high-quality evidence and strong or very strong agreement of experts that the intervention improves important health
outcomes and that benefits substantially outweigh harms; or
 There is moderate moderate-quality evidence and strong or very strong agreement of experts that the intervention improves important
health outcomes and that benefits substantially outweigh harms.
b. against adoption when:
 There is high high-quality evidence and strong or very strong agreement of experts that the intervention is ineffective or that benefits
are closely balanced with harms, or that harms clearly outweigh benefits; or
 There is moderate moderate-quality evidence and strong or very strong agreement of experts that the intervention is ineffective or
that benefits are closely balanced with harms, or that harms outweigh benefits.
B. RECOMMEND
a. adoption when:
 There is moderate moderate-quality evidence and level of agreement of experts that the intervention improves important health
outcomes and that benefits outweigh harms; or
 There is low low-quality evidence but strong or very strong agreement and high level of confidence of experts that the intervention
improves important health outcomes and that benefits outweigh harms; or
 There is very low-quality evidence but very strong agreement and a very high level of confidence of experts that the intervention
improves important health outcomes and that benefits outweigh harms.
b. against adoption when:
 There is moderate moderate-quality evidence and level of agreement of experts that the intervention is ineffective or that benefits are
closely balanced with harms, or that harms outweigh benefits; or
 There is low low-quality evidence but strong or very strong agreement and a high level of confidence of experts that the intervention
is ineffective or that benefits are closely balanced with harms, or that harms outweigh benefits; or
 There is very low-quality evidence but very strong agreement and very high levels of confidence of experts that the intervention is
ineffective or that benefits are closely balanced with harms, or that harms outweigh benefits.
C. THERE IS INSUFFICIENT INFORMATION TO MAKE A RECOMMENDATION
Grade C is applied in the following circumstances:
 Evidence is lacking or scarce or of very low quality, and the balance of benefits and harms cannot be determined, and there is no or
very low level of agreement of experts for or against adoption of the recommendation.
 At any level of evidence—particularly if the evidence is heterogeneous or inconsistent, indirect, or inconclusive—if there is no
agreement of experts for or against adoption of the recommendation.
GPP. GOOD PRACTICE POINT
Good Practice Points (GPPs) are recommendations mostly driven by expert consensus and professional agreement and are based on the
information listed below listed information and/or professional experience, or widely accepted standards of best practice. This category
mostly applies to technical (e.g., preanalytical, analytical, postanalytical), organizational, economic, or quality management aspects of
laboratory practice. In these cases, evidence often comes from observational studies, audit reports, case series or case studies,
nonsystematic reviews, guidance or technical documents, non-evidence–based guidelines, personal opinions, expert consensus, or
position statements. Recommendations are often based on empirical data, usual practice, quality requirements and standards set by
professional or legislative authorities or accreditation bodies, etc.
diabetesjournals.org/care Sacks and Associates 1743

or insulin pump therapy) and who are c. Consider using real-time CGM to im- (GDM) at 24 to 28 weeks of gestation.
not using continuous glucose monitor- prove HbA1c levels, time in range, and A (high)
ing (CGM). A (high) neonatal outcomes in pregnant women c. Either the one-step or two-step pro-
c. Routine use of BGM is not recom- with type 1 diabetes. B (moderate) tocol may be used, depending on re-
mended for people with type 2 diabe- d. Consider using real-time CGM and in- gional preferences. A (moderate)
tes treated with diet and/or oral agents termittently scanned CGM to lower d. Women with GDM should perform
alone. A (high) HbA1c and/or reduce hypoglycemia in fasting and postprandial BGM for opti-
d. Individuals with diabetes should be in- adults with type 2 diabetes who are mal glucose control. B (low)
structed in the correct use of glucose using insulin and not meeting glycemic e. Target glucose values are fasting plasma
meters, including technique of sample targets. B (moderate) glucose <5.3 mmol/L (<95 mg/dL) and
collection and use of quality control. e. Consider real-time-CGM or intermit- either 1-h postprandial <7.8 mmol/L
GPP tently scanned CGM in children with (<140 mg/dL) or 2-h postprandial <6.7
e. Glucose meters should report the glu- type 1 diabetes, based on regulatory mmol/L (<120 mg/dL). B (low)
cose concentrations in plasma rather approval, as an additional tool to f. Women with GDM should be tested for

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023


than in whole blood to facilitate com- help improve glucose control and prediabetes or diabetes 4–12 weeks
parison with plasma results of assays reduce the risk of hypoglycemia. B postpartum using nonpregnant oral
performed in accredited laboratories. (low) glucose tolerance test criteria.
GPP f. Consider using professional CGM data A (moderate)
f. Glucose meters should meet rele- coupled with diabetes self-management g. Lifelong screening for diabetes should
vant accuracy standards of the U.S. education and medication dose ad- be performed in women with a his-
Food and Drug Administration in the justment to identify and address pat- tory of GDM using standard nonpreg-
U.S. or comparable analytical perfor- terns of hyper- and hypoglycemia in nant criteria at least every 3 years. A
mance specifications in other locations. people with type 1 or type 2 diabetes. (high)
GPP GPP h. There is ongoing research, but insuffi-
g. In hospitals and acute-care facilities, g. For individuals using CGM devices that cient evidence at this time, to recom-
point-of-care testing personnel, includ- require calibration by users, a blood mend testing for GDM before 20
ing nurses, should use glucose meters glucose meter should be used to cali- weeks of gestation. C (low)
that are intended for professional use. brate the CGM. Calibration should be
GPP done at a time when glucose is not ris- 6. Urine Glucose
h. When testing newborns, personnel ing or falling rapidly. For all individuals
a. Urine glucose testing is not recom-
should use only meters that are in- using CGM, BGM should be done dur-
mended for routine care of patients
tended for use in newborns. GPP ing periods when CGM results are not
with diabetes mellitus. B (low)
i. Unless CGM is used, people using mul- available or are incomplete, or when
tiple daily injections of insulin should the CGM results are inconsistent with
be encouraged to perform BGM at a the clinical state or suspected to be in- 7. Ketone Testing
frequency appropriate for their insulin accurate. GPP a. Individuals who are prone to ketosis
dosage regimen, typically at least 4 h. CGM data reports should be available (those with type 1 diabetes, history of
times per day. B (moderate) in consistent formats that include stan- diabetic ketoacidosis [DKA], or treated
j. Manufacturers should continue to im- dard metrics such as time in range, with sodium-glucose transport protein
prove the analytical performance of time in hyperglycemia, time in hypo- 2 inhibitors) should measure ketones
meters. GPP glycemia, mean glucose, and coeffi- in urine or blood if they have unex-
cient of variation. GPP plained hyperglycemia or symptoms of
3. Continuous Glucose Monitoring ketosis (abdominal pain, nausea) and
4. Noninvasive Glucose Sensing implement sick-day rules and/or seek
a. Real-time CGM should be used in
a. Overall, noninvasive glucose measure- medical advice if urine or blood ke-
conjunction with insulin as a tool to
ment systems cannot be recommended tones are increased. B (moderate)
lower HbA1c concentrations and/or re-
as replacements for either BGM or b. Specific measurement of b-hydroxybu-
duce hypoglycemia in teens and adults
CGM technologies at this time. C (very tyrate in blood should be used for di-
with type 1 diabetes who are not meet-
low) agnosis of DKA and may be used for
ing glycemic targets or have hypoglyce-
monitoring during treatment of DKA.
mia unawareness and/or episodes of
B (moderate)
hypoglycemia. A (high) 5. Gestational Diabetes Mellitus
c. Blood ketone determinations that rely
b. Consider using intermittently scanned a. All pregnant women with risk factors on the nitroprusside reaction should
CGM in conjunction with insulin as a for diabetes should be tested for un- not be used to monitor treatment of
tool to lower HbA1c concentrations diagnosed prediabetes and diabetes at DKA. B (low)
and/or reduce hypoglycemia in adults the first prenatal visit using standard
with type 1 diabetes who are not diagnostic criteria. A (moderate) 8. Hemoglobin A1c
meeting glycemic targets or have hypo- b. All pregnant women not previously a. Laboratory-based HbA1c testing can be
glycemia unawareness and/or episodes known to have diabetes should be eval- used to diagnose (a) diabetes, with a val-
of hypoglycemia. B (moderate) uated for gestational diabetes mellitus ue $6.5% ($48 mmol/mol) diagnostic
1744 Executive Summary Diabetes Care Volume 46, October 2023

of diabetes, and (b) prediabetes (or high h. HbA1c measurements in individuals 9. Genetic Markers
risk for diabetes) with an HbA1c level of with disorders that affect red blood cell a. Routine determination of genetic
5.7% to 6.4% (39–46 mmol/mol). An turnover may provide spurious (gener- markers such as HLA genes or single nu-
NGSP-certified method should be ally falsely low) results regardless of cleotide polymorphisms is of no value
performed in an accredited labora- the method used, and glucose testing at this time for the diagnosis or man-
tory. A (moderate) will be necessary for screening, diagno- agement of patients with type 1 diabe-
b. Point-of-care HbA1c testing for diabe- sis, and management. GPP tes. Typing for genetic markers and the
tes screening and diagnosis should be i. Assays of other glycated proteins, use of genetic risk scores are recom-
restricted to FDA-approved devices at such as fructosamine or glycated mended for individuals who cannot be
CLIA-certified laboratories that per- albumin, may be used in clinical
clearly classified as having type 1 or
form testing of moderate complexity settings where abnormalities in red
type 2 diabetes. A (moderate)
or higher. B (low) blood cell turnover, hemoglobin var-
b. For selected diabetes syndromes, in-
c. HbA1c should be measured routinely iants, or other interfering factors com-
cluding neonatal diabetes and maturity-
(usually every 3 months until accept- promise interpretation of HbA1c test

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023


onset diabetes of the young, valuable in-
able, individualized targets are achieved results, although they reflect a shorter
period of average glycemia than HbA1c. formation including treatment options can
and then no less than every 6 months) be obtained with definition of diabetes-
in most individuals with diabetes melli- GPP
j. HbA1c cannot be measured and associated mutations. A (moderate)
tus to document their degree of glyce- c. There is no role for routine genetic test-
mic control. A (moderate) should not be reported in individuals
who do not have hemoglobin A, e.g., ing in people with type 2 diabetes.
d. Treatment goals should be based on
those with homozygous hemoglobin These studies should be confined to the
American Diabetes Association rec-
variants, such as hemoglobin SS or he- research setting and evaluation of spe-
ommendations which include main-
moglobin EE; glycated proteins, such cific syndromes. A (moderate)
taining HbA1c concentrations <7%
as fructosamine or glycated albumin,
(<53 mmol/mol) for many nonpreg-
may be used. GPP
nant people with diabetes and more 10. Autoimmune Markers
k. Laboratories should use only HbA1c
stringent goals in selected individuals a. Standardized islet autoantibody tests
assay methods that are certified by
if this can be achieved without signifi-
the NGSP as traceable to the Diabetes are recommended for classification of
cant hypoglycemia or other adverse
Control and Complications Trial refer- diabetes in adults in whom there is
effects of treatment. (Note that these
ence. The manufacturers of HbA1c as- phenotypic overlap between type 1
values are applicable only if the assay
says should also show traceability to and type 2 diabetes and uncertainty as
method is certified by the NGSP as
the International Federation of Clini- to the type of diabetes. GPP
traceable to the Diabetes Control and cal Chemistry and Laboratory Medi- b. Islet autoantibodies are not recom-
Complications Trial reference.) A (high) cine reference method. GPP mended for routine diagnosis of diabe-
e. Higher target ranges are recommended l. Laboratories that measure HbA1c should tes. B (low)
for children and adolescents and are participate in an accuracy-based pro- c. Longitudinal follow-up of subjects with
appropriate for individuals with limited ficiency-testing program that uses two or more islet autoantibodies is rec-
life expectancy, extensive comorbid fresh whole blood samples with tar- ommended to stage diabetes into stage
illnesses, a history of severe hypogly- gets set by the NGSP Laboratory Net-
cemia, and advanced complications. 1: two or more islet autoantibodies,
work. GPP normoglycemia, no symptoms; stage 2:
A (high) m. The goals for imprecision for HbA1c
f. During pregnancy and in preparation two or more islet autoantibodies, dys-
measurement are intralaboratory CV
for pregnancy, women with diabetes glycemia, no symptoms; and stage 3:
<1.5% and interlaboratory CV <2.5%
should try to achieve HbA1c goals that two or more islet autoantibodies, dia-
(using at least 2 control samples with
are more stringent than in the nonpreg- betes and symptoms. GPP
different HbA1c concentrations) and
nant state, aiming ideally for <6.0% d. Standardized islet autoantibody tests
ideally no measurable bias. B (low)
(<42 mmol/mol) during pregnancy to are recommended in prospective stud-
n. HbA1c should be reported as a per-
protect the fetus from congenital mal- centage of total hemoglobin or as ies of children at increased genetic risk
formations and the baby and mother mmol/mol of total hemoglobin. GPP of type 1 diabetes following HLA typing
from perinatal trauma and morbidity o. HbA1c may also be reported as esti- at birth or in first-degree relatives of
owing to large-for-date babies. mated average glucose to facilitate individuals with type 1 diabetes. B
A (moderate) comparison with the self-monitoring re- (low)
g. Laboratories should be aware of po- sults obtained by patients and make e. Screening for islet autoantibodies in
tential interferences, including hemo- the interpretation of the HbA1c more relatives of individuals with type 1 dia-
globin variants that may affect HbA1c accessible to people with diabetes. GPP betes or in people in the general popu-
test results depending on the method p. Laboratories should verify by repeat test- lation is recommended in the setting
used. In selecting assay methods, lab- ing specimens with HbA1c results below of a research study or can be offered
oratories should consider the poten- the lower limit of the reference interval as an option for first-degree relatives
tial for interferences in their particular or greater than 15% (140 mmol/mol) of a proband with type 1 diabetes. B
patient population. GPP HbA1c. GPP (low)
diabetesjournals.org/care Sacks and Associates 1745

f. Routine screening for islet autoantibod- albuminuria to be useful for patient Photonics, a for-profit company working to
ies in people with type 2 diabetes is not screening. B (moderate) develop wearable technology for fitness and
wellness applications, including noninvasive glu-
recommended at present. B (low) j. Practitioners should strictly adhere to cose measurements, consulting for Foundation
g. There is currently no role for measure- manufacturer’s instructions when us- for Innovative New Diagnostics concerning the
ment of islet autoantibodies in the ing a semiquantitative uACR dipstick development of glucose sensing technologies for
monitoring of individuals with estab- test and repeat it for confirmation to low- and middle-income countries, consulting
achieve adequate sensitivity for de- for LifePlus concerning their interest in the de-
lished type 1 diabetes. B (low)
velopment of a noninvasive glucose sensor for
h. It is important that islet autoantibodies tecting moderately increased albumin- management of diabetes, and patent application
be measured only in an accredited lab- uria. B (moderate) filed (Olesberg, JT, Arnold, MA, Urea monitoring
oratory with an established quality k. Positive urine albumin screening re- during dialysis for improved quality control and
control program and participation in a sults by semiquantitative tests should treatment guidance, U.S. Provisional Application
63/087,600 filed on 5 October 2020, Interna-
proficiency testing program. GPP be confirmed by quantitative analysis
tional Patent Application No. PCT/US2021/
in an accredited laboratory. GPP 053598 filed on 5 October 2021, U.S. Patent Ap-
11. Urine Albumin l. Currently available proteinuria dipstick plication 18/030,339 filed on 5 April 2023).

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023


tests should not be used to assess al- G.L.B., consultant or member of clinical trial
a. Annual testing for albuminuria should steering committee for Bayer, KBP Biosciences,
begin in pubertal or post-pubertal indi- buminuria. B (moderate)
Ionis, Alnylam, AstraZeneca, Novo Nordisk, Jans-
viduals 5 years after diagnosis of type 1 sen InREGEN and consulted for DiaMedica and
diabetes and at the time of diagnosis 12. Miscellaneous Potentially Quantum Genomics. A.R.H., Elsevier, royalty for
Important Analytes Tietz Textbook of Clinical Chemistry and Molecular
of type 2 diabetes, regardless of treat-
a. In most people with diabetes or at risk Diagnostics, 6th edition, and Tietz Fundamentals of
ment. A (high) Clinical Chemistry and Molecular Diagnostics, Chair
b. Urine albumin should be measured annu- for diabetes or cardiovascular disease, of the Preanalytical glucose working party of the
ally in adults with diabetes using morning routine testing for insulin or proinsulin Australasian Association of Clinical Biochemistry
spot urine albumin-to-creatinine ratio is not recommended. These assays are and Laboratory Medicine, Chair of the Analytical
useful primarily for research purposes. Performance Specifications based on Outcomes
(uACR). A (high) Task Force Group of the European Federation of
c. If estimated glomerular filtration rate B (moderate)
Clinical Chemistry and Laboratory Medicine. Å.L.,
is <60 mL/min/1.73 m2 and/or albu- b. Although differentiation between type 1 DiaMyd AB, Stockholm, Sweden, consulting fee. No
minuria is >30 mg/g creatinine in a and type 2 diabetes can usually be other potential conflicts of interest relevant to this
spot urine sample, the uACR should be made based on the clinical presenta- article were reported.
tion and subsequent course, C-peptide Author Contributions. The corresponding au-
repeated every 6 months to assess thor takes full responsibility that all authors on
change among people with diabetes measurements may help distinguish this publication have met the following required
and hypertension. A (moderate) type 1 from type 2 diabetes in ambigu- criteria of eligibility for authorship: (a) significant
d. First morning void urine sample should ous cases, such as individuals who contributions to the conception and design, ac-
have a type 2 phenotype but present quisition of data, or analysis and interpretation
be used for measurement of albumin- of data, (b) drafting or revising the article for
to-creatinine ratio. A (moderate) in ketoacidosis. B (moderate)
intellectual content, (c) final approval of the
e. If first morning void sample is difficult c. If required by the payer for coverage published article, and (d) agreement to be ac-
to obtain, to minimize variability in of insulin pump therapy, measure fast- countable for all aspects of the article thus en-
test results, all urine collections should ing C-peptide level when simultaneous suring that questions related to the accuracy or
fasting plasma glucose is 12.5 mmol/L integrity of any part of the article are appropri-
be at the same time of day. The indi- ately investigated and resolved. Nobody who
vidual should be well hydrated and (<220 mg/dL). GPP
qualifies for authorship has been omitted from
should not have ingested food within d. Insulin and C-peptide assays should be the list. A detailed list of author contributions to
the preceding 2 h or have exercised. standardized to facilitate measures of “Guidelines and Recommendations for Labora-
insulin secretion and sensitivity that tory Analysis in the Diagnosis and Management
GPP of Diabetes Mellitus” can be found within the
f. Timed collection for urine albumin will be comparable across research
full guidelines online at https://doi.org/10.2337/
should be done only in research set- studies. GPP dci23-0036.
tings and should not be used to guide e. There is no published evidence to sup-
clinical practice. GPP port the use of insulin antibody testing References
g. The analytical performance goals for for routine care of people with diabe- 1. Casta~ no L, Eisenbarth GS. Type-I diabetes: a
tes. C (very low) chronic autoimmune disease of human, mouse,
urine albumin measurement should and rat. Annu Rev Immunol 1990;8:647–679
be between-day precision #6%, bias 2. Reaven GM. Banting lecture 1988. Role of
#7% to 13%, and total allowable error insulin resistance in human disease. Diabetes 1988;
#24% to 30%. GPP Funding. D.B.S., National Institutes of Health In- 37:1595–1607
tramural Research Program. Å.L., Swedish Re- 3. Sacks DB, McDonald JM. The pathogenesis of
h. Semiquantitative uACR dipsticks can search Council. type II diabetes mellitus. A polygenic disease. Am
be used to detect early kidney disease Duality of Interest. D.B.S., Associate Editor for J Clin Pathol 1996;105:149–156
and assess cardiovascular risk when Clinical Chemistry, American Association for Clini- 4. International Diabetes Federation. IDF Diabetes
quantitative tests are not available. B cal Chemistry, and Chair, Scientific Review Com- Atlas. 10th ed. 2021. Accessed 23 May 2022.
(moderate) mittee on Laboratory Analysis in the Diagnosis Available from https://diabetesatlas.org/idfawp/
and Management of Diabetes Mellitus. M.A.A., resource-files/2021/07/IDF_Atlas_10th_Edition
i. Semiquantitative or qualitative screen- travel expenses paid by American Diabetes Asso- _2021.pdf
ing tests should be positive in >85% of ciation for attendance at the in-person manu- 5. Centers for Disease Control and Prevention.
individuals with moderately increased script review meeting, consulting for Rockley National Diabetes Statistics Report, 2022.
1746 Executive Summary Diabetes Care Volume 46, October 2023

Accessed 9 August 2022. Available from https:// 9. Sacks DB, Bruns DE, Goldstein DE, Maclaren of the American Association for Clinical Chemistry.
www.cdc.gov/diabetes/data/statistics-report/index NK, McDonald JM, Parrott M. Guidelines and Guidelines and recommendations for laboratory
.html recommendations for laboratory analysis in the analysis in the diagnosis and management of
6. American Diabetes Association. Economic costs diagnosis and management of diabetes mellitus. diabetes mellitus. Diabetes Care 2011;34:
of diabetes in the U.S. in 2017. Diabetes Care Diabetes Care 2002;25:750–786 e61–e99
2018;41:917–928 10. Sacks DB, Bruns DE, Goldstein DE, Maclaren 12. Sacks DB, Arnold M, Bakris GL, et al. Guidelines
7. American Diabetes Association. Economic costs NK, McDonald JM, Parrott M. Guidelines and and recommendations for laboratory analysis in the
of diabetes in the U.S. in 2007. Diabetes Care recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.
2008;31:596–615 diagnosis and management of diabetes mellitus. Clin Chem 2011;57:e1–e47
8. Roglic G, Unwin N, Bennett PH, et al. The Clin Chem 2002;48:436–472 13. Sacks DB, Arnold M, Bakris GL, et al. Guidelines
burden of mortality attributable to diabetes: 11. Sacks DB, Arnold M, Bakris GL, et al.; and recommendations for laboratory analysis in the
realistic estimates for the year 2000. Diabetes National Academy of Clinical Biochemistry; diagnosis and management of diabetes mellitus.
Care 2005;28:2130–2135 Evidence-Based Laboratory Medicine Committee Diabetes Care 46:e151–e199

Downloaded from http://diabetesjournals.org/care/article-pdf/46/10/1740/734759/dci230048.pdf by guest on 22 October 2023

You might also like