BT-700N

Patient Monitor

OPERATOR’S MANUAL

BT-700N

P/N : OPM(BT-700N)EN , Rev. : 02


BT-700N Operator Manual 1

Proprietary Material
Information and descriptions contained in this manual are the property of
Bistos Corporation and may not be copied, reproduced, disseminated, or
distributed without express written permission from Bistos Corporation.

Information furnished by Bistos Corporation is believed to be accurate and


reliable. However, no responsibility is assumed by Bistos for its use, or any
infringements of patents or other rights of third parties that may result from its use.
No license is granted by implication or otherwise under any patent or
patent rights of Bistos

Revision 02
April 2013

Copyright © Bistos Corporation 2013. All rights reserved.

7th Fl., A Bldg., Woolim Lions Valley 5-cha, 144-3,


Sangdaewon-dong, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Korea

Telephone: ++82 31 750 0340


Fax: ++82 31 750 0344

Printed in Korea

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 2

Table of Contents

1. SAFETY ............................................................................................ 4
1.1 Intended Use ....................................................................................................... 4
1.2 Instruction for the safe operation and use of the BT-700N ................................. 4
1.3 General Precautions on Environment ................................................................. 4
1.4 General Precautions on Electric safety ............................................................... 5
1.5 Definition of Symbols ......................................................................................... 6

2. INTRODUCTION ............................................................................ 7
2.1 About this Manual ............................................................................................... 7
2.2 BT-700N Monitor Configuration ........................................................................ 7
2.3 List of Components ............................................................................................. 7

3. INSTALLATION .............................................................................. 9
3.1 Description of the BT-700N Front Panel ............................................................ 9
3.2 Description of the BT-700N Rear Panel ............................................................. 9
3.3 Description of the BT-700N Left and right Panel ............................................. 10
3.4 Display .............................................................................................................. 11
3.5 Status icons and Display color .......................................................................... 12
3.6 Front panel LED ............................................................................................... 13
3.7 Power Adaptor Connection ............................................................................... 13
3.8 Measuremetn Cable Connections ...................................................................... 14
3.9 Battery operation............................................................................................... 15

4. BT-700N OPERATION .................................................................. 17


4.1 Turning on the monitor ..................................................................................... 17
4.2 Key and knob operation .................................................................................... 18
4.3 Display .............................................................................................................. 19
4.4 Connection and Measurement ........................................................................... 24

5. SYSTEM SETUP / MENU WINDOW / ALARM & SOUND .... 26


5.1 General .............................................................................................................. 26
5.2 Menu Structure ................................................................................................. 26
5.3 Cursor move and Select .................................................................................... 27
5.4 Boot Menu ........................................................................................................ 27
5.5 NIBP Maintenance Menu .................................................................................. 30
5.6 System Setup Menu .......................................................................................... 31
5.7 Alarm Menu ...................................................................................................... 32
5.8 Clock Menu....................................................................................................... 33
5.9 NIBP Setup Menu ............................................................................................. 33
5.10 Trend Data Menu ............................................................................................ 34
5.11 Browse Data Menu.......................................................................................... 34
5.12 SpO2 Setup Menu ........................................................................................... 35
5.13 Menu Tree ....................................................................................................... 36

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 3

5.14 Alarm and Sound ............................................................................................ 37


5.15 Alarm Priority and Messege ............................................................................ 38
5.16 Audible Alarm Suspend / Silence .................................................................... 39
5.17 Remind Beep................................................................................................... 39

6. SPO2 MONITORING .................................................................... 40


6.1 Setup Connections ............................................................................................ 41
6.2 Description of SpO2 Display and Menu ........................................................... 42

7. NIBP MONITORING .................................................................... 43


7.1 Setup Connections ............................................................................................ 44
7.2 NIBP Measurement Modes ............................................................................... 45
7.3 Descriptions of NIBP Display and Menu .......................................................... 45

8. MAINTENANCE ........................................................................... 48
8.1 Maintenance and Cleanliness ............................................................................ 48
8.2 Accessory cleaning ........................................................................................... 48
8.3 Regular Inspection ............................................................................................ 48
8.4 Disjoint and handle the internal battery ............................................................ 48

9. TROUBLESHOOTING ................................................................. 49

10. SPECIFICATION......................................................................... 50
10.1 Mechnical and Electric Specifications ............................................................ 50
10.2 Compliance ..................................................................................................... 50
10.3 Environmental Conditions .............................................................................. 50
10.4 Performance Specifications............................................................................. 51

11. WARRANTY ............................................................................... 53

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 4

Chapter 1
Safety
1.1 Intended Use
BT-700N is a Patient Monitor. It measures SpO2, blood pressure, pulse rate and
temperature. BT-700N irradiates the light signal to patient’s installation part and converts
the light signal which is penetrated patient’s blood vessel to the electrical signal. Using
this BT-700N measures arterial blood oxygen saturation and pulse rate. To measure blood
pressure, set the cuff to patient’s brachial muscle and pressurize to the cuff and measure
blood pressure. Also using noncontact infrared thermometer, it measures patient’s
temperature in a short time. It displays SpO2, blood pressure, pulse rate and temperature
to graph and number. And each value is saved internal memory. Also according to the
installed measuring modules, it can change the display with five kinds of windows.

Note : This device is for use only by trained medical personnel located in a treatment
room which is located in hospital, an operating room, a recovery room and an
intensive care unit.
Note : Because the incorrect use of this device can do patient’s life harm, this device
must be installed, used and operated by trained medical personnel who can
measure and analyze physiological signals.

1.2 Instructions for the safe operation and use of the BT-700N

Examine the monitor and any accessories periodically to ensure that the cables, line cords,
transducers, and instruments do not have visible evidence of damage that may affect patient
safety or monitoring performance. The recommended inspection interval is once per week or
less. Do not use the monitor if there is any visible sign of damage. Do not place the monitor
to incline place to avoid falling of product.

1.3 General Precautions on Environment


Do not keep or operate the equipment under the environment listed below.

Avoid placing in an area


exposed to moisture. Do not Avoid exposure to
touch the equipment with wet direct sunlight
hand.

Avoid placing in an area


where there is a high
Avoid in the vicinity
variation of temperature.
of Electric heater
Operating temperature
ranges from 10C to 35C.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 5

Avoid placing in an area


where there is an Avoid placing in
excessive humidity rise or an area where
ventilation problem. there is an excessive
Operating humidity ranges shock or vibration.
from 30% to 85%.

Avoid placing in an area Avoid dust and


where chemicals are especially metal
stored or where there is in material into the
danger of gas leakage. equipment.

Power off when


Do not disjoint or the equipment is
disassemble the equipment. not fully installed.
BISTOS Co., Ltd. does not Otherwise, the
take responsibility of it. equipment could
be damaged.

1.4 General Precautions on Electric safety


Please check the flows before using or operating the product.
 Is power supply line to adjust adaptor appropriate? (100-240V AC).
 Do you use the authorized adaptor that we offered?
 Are the all connection parts connected to the product correctly?

Classification of BT-700N :
- BT-700N’s class is Class II, type-BF about electrical shock.
- Do not use this product around a flammable anesthetic and solvent.
- Rise of the noise is A class in the standard IEC/EN 60601-1 (Safety
Caution of Electric Medical Equipment). Exemption of the noise level is B in
the standard IEC/EN 60601-1-2 (Electromagnetic Compatibility
Requirements).
To avoid electrical noise while using this product, please install this
product away from the generator, X-ray equipment, broadcasting
equipment and portable cable. If this product is installed nearby these
equipments, this can cause an incorrect result. The use of isolated
power is important to BT-700N. If power is shared with another
electronic equipments, this can cause a incorrect result.
Supporting equipments that are connected to analogue or digital
connector as RS-232 Port and USB Port in the back of the product
Warning have to meet the IEC standard regulation. Also all compositions have
to meet EN 60601-1-1:2001.
If this product is used in high temperature, this can cause the
explosion of internal battery or leakage of an electrolyte from
battery. Please use this product below 40C surely.
If skin or cloth is stained with an electrolyte from internal battery,
wash and clean with clean water immediately..
BT-700N doesn't have protection against the burn injury caused by
RF Surgical equipment. Do NOT use this product along with RF
Danger Surgical equipment.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 6

1.5 Definitions of Symbols

Symbol Definition

Power On/Off Button

This symbol identifies a safety note. Ensure you understand


the function of this control before using it. Control function is
described in the operator’s manual.

External Signal IN/OUT Port

Adapter Power Input

IPX2 IPX2 Water-proof (SpO2 sensor)

IPX1 IPX1 Water-proof(NIBP)

IPX0 IPX0 Non-protected(Thermometer)

Defibrillation-proof type BF applied part

Type BF applied part

Table 1-1. Definitions of symbols

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 7

Chapter 2
Introduction
2.1 About this Manual
This manual explains how to install and use BT-700N.
Be well-informed of the suggestions and precautions before using this device.

2.2 BT-700N Monitor Configuration


This table shows the functional combinations of BT-700N. The functional combinations
can be defined by Serial number which is stock to equipment. If the function to use is not
displayed, please check the combination by Serial Number..

Serial No. Description


BT-700N-S BT-700N Main with Battery + SpO2
BT-700N-N BT-700N Main with Battery + NIBP
BT-700N-SN BT-700N Main with Battery + SpO2 + NIBP
BT-700N-ST BT-700N Main with Battery + SpO2 + Thermometer
BT-700N-NT BT-700N Main with Battery + NIBP + Thermometer
BT-700N-SNT BT-700N Main with Battery + SpO2 + NIBP + Thermometer
Table 2-1. Configuration

2.3 List of Components


Open the package and check the list of components. Also please examine BT-700N body
and accessories do not have visible evidence of damage.

Items Qty
BT-700N body 1 ea
Re-usable SpO2 sensor probe 1 ea
Extension for Spo2 sensor 1 ea
NIBP cuff (Adult) 1 ea
Extension hose 1 ea
Power adaptor 1 ea
AC power code 1 ea
Operation manual 1 ea
Table 2-2. Standard Accessories

The follows are optional items. If you want to purchase, please contact the head office.

Items Qty
Disposable sensor probe -
Disposable neonate cuff -
Thermometer -
(Non-Contact infrared detector)
Thermometer holder -
IV-Holder -
Table 2-3. Optional Accessories

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 8

The different results can come out according to the status of patients.
If the results are doubtful, please check the patient’s status using
Warning other measuring equipment which is using in hospital.
Use only authorized accessories supplied with BT-700N. Use of any
other accessories may result in out of specification performance and
Warning possible safety hazards.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 9

Chapter 3
Installation
3.1 Description of the BT-700N Front Panel
② ⑧

③ ⑨⑩⑪ ⑦⑥⑤④
Figure 3-1. Front panel
① Power On/Off Button
② Power Indicating LED ( AC: Green / Battery: Orange)
③ TFT-Color LCD ④ Mode Change Button
⑤ Trend Save Button ⑥ Alarm On/Silent/Suspend Button
⑦ NIBP On/Off Button ⑧ Control Knob
⑨ Trend Save Status Led ⑩ Alarm Status Led
⑪ NIBP Status Led

3.2 Description of the BT-700N Rear Panel

Figure 3-2. Rear panel

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 10

① Grip ② Battery Cover


③ Power Adaptor Jack Connector ④ RS-232C Port Connector
⑤ USB Port Connector ⑥ Dip Switch
⑦ Speaker ⑧ Multi-Holder

3.3 Description of the BT-700N Left and Right Panel


Figure 3-3. Left panel Figure 3-4. Right panel

① SpO2 Sensor Connector ① Multi-Holder


② Thermometer Connector
③ NIBP Cuff Connector

USB Port and DIP switch in back panel are not for coustomer or
user. If connect this USB Port to PC USB or turn off DIP switch, this
Warning will occur failure in the system.

Do not open the battery pack cover in back panel. Do not insert or
change the battery pack that Bistos Co., Ltd. provided to other
Warning companies’ battery pack. These will cause malfunction and electric
shock. In response, Bistos Co., Ltd. dose not take responsibility of it.

To avoid electric shock, do not open the cover or power adaptor


exterior. Disassemble and repair must be performed by the qualified
Warning persons of Bistos Co., Ltd.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 11

3.4 Diplay

BT-700N can present the various windows according to the kind of connecting module.

a. Graph mode
- SpO2 + NIBP + Thermometer
- SpO2 + NIBP
- SpO2 + Thermometer
b. Numeric mode
- SpO2 + NIBP + Thermometer
- SpO2 + NIBP
- SpO2 + Thermometer
- NIBP + Thermometer
- NIBP

Figure 3-5 is default display window of BT-700N.


(S : SpO2, N : NIBP, T : Thermometer)

SpO2 Region NIBP Region

SpO2 Wave Region

Status Region Thermometer Region

Figure 3-5. Graph mode (S+N+T)

The system is set with thermometer is connected, display


Caution region in status region is changed.

When alarm is occurred, data and time in status region do not


displayed. When thermometer is connected, Beep volume icon
Caution does not displayed also.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 12

3.5 Status icons and Display color

Symbols Description Symbols Description

SpO2 index Beep volume icon


SpO2 unit Battery status icon
HR/PR detect icon Patient mode : Adult

Pulse amplitude indicator Patient mode : Neonate

HR/PR index Audible alarm Active/Silent

HR/PR unit Audible alarm Suspend

NIBP index Trend data save On/Off


2001/01/01
NIBP unit : Mercury Date and Time
12:00

NIBP unit : Pascal Thermometer index

NIBP auto mode Temp unit : Celsius

NIBP clock Temp unit : Fahrenheit

MEAN NIBP mean pressure Alarm high/low limit


NIBP stat mode
Table. 3-1. Status Icon

Function Color

SpO2 Blue / Gray


NIBP Yellow / Gray
Thermometer Orange
Beep volume icon Green / Gray
Battery status icon Green / Red
Patient mode icon Green
Alarm status icon Green / Gray
Trend status icon Green / Gray
Alarm Red
Table 3-2. Display Color

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 13

3.6 Front panel LED

Table 3-3 : The functions and components of BT-700N front panel LED.

Symbols Description

Power Status
- AC Power ON and
Battery Charging : Green LED on
-AC Power OFF and
Battery Discharging : Green LED off

Power Status
- Battery Power ON : Orange LED on
- Battery Charging : Orange LED off
Data Saving
- Trend on(Data save) : Green LED on
- Trend off : Green LED off
Alarm status
- Alarm on : Green LED on
- Alarm silent : Green LED blink
- Alarm suspend : Green LED off
NIBP on / off
- NIBP on : Green LED on
- NIBP off : Green LED off
Table 3-3. Front panel LED Symbols

3.7 Power Adaptor Connection

To ensure exact measurement and safety, do not operate the


device in damp or rainy environment. If the device is operated
Warning in that environment, we do not ensure the performance of the
device.
Do not use the device near other measuring equipments. If
other equipments are used near the device, you must check the
Warning device is operating correctly.
Do not screen a speaker. Screening speaker may result in
Warning inaudible alarms.
If the power is not appropriate or connected to outlet which is
not grounded, it may result in electric shock or fire. If the
Danger power status is doubtful, please check the power through an
electrician.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 14

1. Connect adaptor which is supplied with the device to the power. (100-240V ~50-60Hz).

2. Connect power adaptor to power connector which is located in back panel as figure 3-6.

3. If the power is supplied normally, power LED is on with green.

Figure 3-6. Power adaptor connection

3.8 Measurement Cable Connections

To get the best performance and exact measuring values,


please use accessories which is supplied with the device or is
recommended by Bistos Co.,Ltd. The standard for using
Warning accessories and facilities must meet manufacturer’s indication.
Accessories have to meet biocompatibility of ISO10993-1.

1. Connect SpO2sensor or sensor’s extension cable to SpO2 connector which is located in


right panel.

2. Connect NIBP Extension tube to NIBP connector which is located in right panel.

3. If thermometer is optional item, connect thermometer to TEMP connector which is


located in right panel.

Figure 3-7 is the picture that each cable is connected.

Figure 3-7. Cable Connections

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 15

※ If thermometer is equipped for optional item, please set the multi-holder to


BT-700N body like as figure 3-8.

Figure 3-8. Thermometer Stand & Thermometer

3.9 Battery operation

The battery should be charged for a full more than one time in
Caution two months.

If the voltage is very low, the device will not operate.


Caution

Note : In the case of the BT-700N is not in use, AC power should be connected to BT-
700N for charging the battery.
Note : The battery is consumables. If the battery using period is over six months, the
operating time by the battery can be decreased. If the operating time by the battery is not
long enough, please contact service center and change the battery.

1. Operating the Monitor on Battery Power

If AC power is not supplied, BT-700N is operated by internal battery. In this case, battery
LED is on with orange and AC LED green light is off.

Symbols Description
Power status
- AC Power Connect and Battery Charging
: Green LED on
-AC Power Disconnect and Battery in use
: Green LED off
Power status
- Battery in use : Orange LED on
- Battery Charging : Orange LED off
Table 3-4. Power Status

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 16

In the case of operating BT-700N without AC power, BT-700N will not operate if the
battery is discharged completely. In this case, connect AC power over three minutes and
operate the device.
Note : If the battery is charged for a full, the device will operate during three hours.
(NIBP measuring interval : 1 minute, Alarm on).
※ Merely, the operating time can be changed by the using environment or status of the
device.

Note : In the case of the device is off with AC power is connected, LED is on.

2. Battery Status Indication

User can check the battery status by battery icon in status region during BT-700N is
operated with the battery.
Table 3-5 indicates the status of battery icon according to the residual.

Symbols Description
Full charge (Color : Green)

Using the time ≤ 10 min (Color : Orange)

Using the time ≤ 5 min (Color : Red flashing)


Table 3-5. Battery Status Icon

Note : To charge the battery over 90%, connect AC power to BT-700N and keep
this at least eight hours.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 17

Chapter 4
BT-700N Operation

When the device is on, if system check is not finished normally,


stop using the device and contact service center or agent
Warning immediately.

When the device is in use, please check Alarm limits setting is


Warning appropriate to patient’s status.

By not intended influences which are caused by internal or


external environments, the accuracy of the device can be
Warning affected. Therefore, to minimize these influences, please use the
device under the person in charge.

4.1 Turing on the Monitor


Before operating the device, please check the device is operating normally as follows.

1. Press power key( ) which is located on the top of front panel more than two
seconds. Then booting window (bistos logo) will be displayed in LCD with “di,
di, dic~” sound.

2. After displaying Booting logo, the device proceeds System check automatically.
In this procedure, the device will check module connecting status too.

3. If there is no fail in the device, Main window will be displayed in LCD.

When power is on, “di, di, ding” sound is ring. If there is no


sound or abnormal sound is ring, the device is out of function.
In this case, stop using the device immediately and contact
Caution service center or agent. If you continue to use the device with
no sound or abnormal sound, alarm sound function does not
work properly.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 18

4.2 Key and Knob operation

1. Key
In the case of Key beep is on, the beep sounds when each key is operating. User can
control or turn off the beep volume in Main window by rotating the knob.
Also user can check the bee volume by the number of green bar ( Icon) in Main
window. If the beep volume is off, the icon is changed to gray.

Symbol Description
Power ON/OFF
- Power on : Press button more than two seconds from power off status,
then the system goes to power on status.
- Power off : Press button slightly from power on status, then the system
goes to power off status. (1 Beep)
Display mode change
MODE - Whenever user press this button, display mode will be changed
stepwise. (1 Beep)
Trend ON/OFF
- Trend on : Press button in Trend off, then the system goes to Trend
TREND on(Data save). (1 Beep)
- Trend off : Press button in Trend on, then the system goes to Trend
off. (1 Beep)
Alarm ON / SILENT / SUSPEND
- Alarm silent : Press button once shortly when Alarm is on, then the
system goes to Alarm silent. (1 Beep)
ALARM - Alarm suspend : Press button once long when Alarm is ons, then the
system goes to Alarm suspend. (2 Beep)
- Alarm on : Press button once shortly in Alarm Silent or Suspend status, then
the system goes to Alarm on.(1 beep)
NIBP Stop/Start(Manual /Auto / STAT mode)
- Start Manual mode(Auto Mode Interval is set to off in Menu/NIBP item)
: Press button once shortly in NIBP stop status, then user can start NIBP
manual mode.(1 beep)
NIBP
- Start Auto mode(Auto Mode Interval time is set in Menu/NIBP item)
: Press button once shortly in NIBP stop status, then user can start NIBP auto mode.(1 beep)
- Start STAT mode : Press button once long in NIBP stop status, then user
can start NIBP stat mode. (2 beep)
Table 4-1. Key Description

Note : All keys perform the same function in Menu window, but Mode key does not
effective in Menu window.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 19

2. Knob
Knob is used for changing Beep volume (in Main window) or entering Menu. Also
knob is used for shifting and changing items of Menu item.

Display
Action Description
window
Push Enter Menu/Start up setup window

Main Window Rotate CW Up beep volume (+1)

Rotate CCW Down beep volume (-1)

Boot Menu Push Select item / Finish selecting item


& Menu Rotate(CW/CCW) Shift item and change setting value
Table 4-2. Knob Description

4.3 Displays

1. Booting logo

- When the power is on, Bistos logo is displayed with “di, di, ding” sound.

Figure 4-1. Booting window

2. System check

- While the system is booting, BT-700N will check the system and connection of module.
If the system checks errors, the system will display “Fail” in the particular item and stop
system checking.
The system will enter into “Boot menu” when user pushes the knob in system check
procedure.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 20

Figure 4-2. System check window


Note : If “Fail” sign is displayed while system check, please turn off the device and contact
service center or agent.

3. Boot menu window


- In this window, user can select items about the system maintenance.

Figure 4-6. Boot menu window

4. Main window
- BT-700N can display various windows according to the kind of connecting module (SpO2,
NIBP, Thermometer) . Each window consists of various frames. Main window is divided
into two sections. The one is Graph mode which displays signal wave and number. And the
other is Numeric mode which displays only number. Except these, there is a mode which
displays SpO2 or NIBP.

SpO2 Region NIBP Region

SpO2 Wave Region

Status Region Thermometer Region


Figure 4-7. Graph mode (S+N+T)

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 21

SpO2 Region NIBP Region

Status Region Thermometer Region


Figure 4-8. Numeric mode(S+N+T))

When Spo2, NIBP and Thermometer modules are connected, the combinations of displays
are as follows.
a. Graph mode
- SpO2 + NIBP + Thermometer
- SpO2 + NIBP
- SpO2 + Thermometer
b. Numeric mode
- SpO2 + NIBP + Thermometer
- SpO2 + NIBP
- SpO2 + Thermometer
- NIBP + Thermometer
- NIBP

Figure. 4-9. Graph mode (S+N) Figure 4-10. Numeric mode (S+N)

Figure 4-11. Graph mode (S) Figure 4-12. Numeric mode (S)

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 22

Figure 4-13. Numeric mode (N)


The system can not check the connection of temperature
module automatically. Therefore user must set and check the
connection in the “Boot menu” window.
Caution In the case of the connection of temperature module, the
display will change only in status region. The other regions
will not be changed.

5. Status region
- Date/Time, Beep volume, Battery level, Patient mode, Alarm icon and Trend icon are
presented in this region. When temperature module is connected, the location of each item is
changed and the system will present the value of measured temperature instead of Alarm and
Trend icon.
- When alarm is activated, the system will blink Alarm index with red color in Date/Time
region. (When temperature module is connected, Beep volume region will be changed to
Alarm index region.)

Figure 4-14. ~ Figure 4-17 present displayed items according to Thermometer on/off and
Alarm on/off in status region.

Figure 4-14. Status Region(Thermometer OFF & Alarm OFF)

Figure 4-15. Status Region(Thermometer ON & Alarm OFF)

Figure 4-16. Status Region(Thermometer OFF & Alarm ON)

Figure 4-17. Status Region(Thermometer ON & Alarm ON)

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 23

Table 4-3 : The functions of status region icon

Icon Name Description


Beep volume icon Indicates Beep(HR/Key) volume status.
Battery status icon Indicates battery charge status.
Patient mode icon The object of measurement is adult.

Patient mode icon The object of measurement is neonate.


Indicates alarm on status. (Green)
Alarm status icon
Indicates alarm silent status. (Green blink)
Alarm status icon Indicates alarm suspend status. (Red)
Indicates trend status.
Trend status icon - Trend on : Green(Icon is roataing.)
- Trend off : Gray(Icon is stop.)
Table 4-3. The functions of status region icon.

6. Thermometer region
- In the case of temperature item is on in Boot up setup when the system is booting,
thermometer region will be created in status region.
Thermometer icon, unit and the measured value are presented in this region.
Figure 4-18 showes displayed items in thermometer region.

Thermometer unit

Thermometer icon Thermometer value

Figure 4-18 Thermometer region

7. Setting basic setup Parameters


- Push the knob in Main window and enter Menu. In this window user can change the system
setting according to the purpose by rotating the knob. To select the item, push the knob.

Figure 4-19. Initial Menu Window

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 24

Menu Description
System Setup User can set items related to the system and the alarm sound.

NIBP Setup User can set items related to NIBP.

SpO2 Setup User can set items related to SpO2.

Trend Data User can set items related to the patient’s data saving.
Table 4-4. Initial Menu

Note : If user pushes the knob during system check, the system enter to Boot Menu.

4.4 Connection and Measurement

1. SpO2
- Probe connection
Connect SpO2 probe to the “SpO2” connector in the left panel.

- Measurement of SpO2 and Pulse rate


Clean the measuring part(ex, finger and so on) with alcohol. Insert finger into a probe.
Avoid the use of other medical devices which can disturb the blood flow.
Avoid attaching the probe to the position which can bring up the interference of
measurement. For the stable measurement, please fix the probe cable properly using finger
and a sticking plaster with attention to prevent the disturbing of blood flow.
Check the status of skin and connection of probe at regular interval.

2. NIBP
- Cuff connection and disconnection
Push NIBP cuff in the “NIBP” connector in the left panel. NIBP cuff is connected with a
clack sound. Check the locking status of connector. To disconnect a cuff from “NIBP”
connector, slide a lock slide to the back and unlock the lock slide.

- Measurement of NIBP
Sit on the chair calmly and comfortably with back about 5 minutes and stabilize the body.
Measure NIBP in a sitting or lying position (Provided, the height of arm is same as heart.).
The object of measurement must not smoke and drink coffee before the measurement of
blood pressure. After taking off the object’s top or rolling up object’s shirt’s sleeve with no
arm fastening, measure blood pressure. The size of cuff’s horizontal axis must be large as
possible as can wrap 80% of arm. Measure blood pressure over two times at the intervals of
two minutes and average these measured values.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 25

3. Thermometer
- Thermometer connection
Connect thermometer module to the “TEMP” connector in the left panel.
Because the system can not recognize thermometer connection automatically, user have to
set with thermometer is connected in Boot Menu when the system is booting.

- Measurement of Temperature
Set a thermometer facing the forehead. Take a thermometer’s position about 3cm far from
the forehead. Push “START” button at the end of a thermometer and start to measure
temperature.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 26

Chapter 5
System Setup / Menu window /
Alarm & Sound
5.1 General

BT-700N has two windows to set the system. The first window is Start up setup window that
can set the system and items related maintenance when the system is booting. The other
window is Menu window that can set the various system setting while the system is working.
Pushing knob cause that BT700N will enter Boot Menu window or Menu window.

5.2 Menu Structure

Figure 5-1. shows Menu Structure.

Figure 5-1. Menu Structure

- Path : Indicate the name and location of Menu window at present.


- Menu list : Indicate the item list of the window at present.
- Select item : Indicate the selected item which cursor is pointed.
- Step up : Move to upper stage from the current Menu.
- Exit : Exit Menu window.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 27

5.3 Cursor move and Select

Knob rotating makes cursor movement between items in the window. User can move
between items through the rotating knob to clockwise or counterclockwise. The item that
cursor is pointed will be reversal (blue color).
Figure 5-2. shows cursor is pointed the first item in Boot menu window.

Figure 5-2. Cursor Move

After move cursor to item to select by rotating knob and push knob, then the system select
the item. If item is selected, the color of selected item is changed to orange.
Figure 5-3 shows second item is selected in Boot menu window.

Figure 5-3. Item Select

Select an item and then change setting what you want by rotating the knob. And then push
the knob to exit from the selected item.

5.4 Boot Menu

Incorrect setting for Boot Menu can cause a critical influence to


operating the device. Setting for Boot Menu must be performed
Warning by authorized person.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 28

User can set items related System Maintenance in Boot menu and check the system status.
The system enters to Boot menu by pushing the knob during system check.

Figure 5-4. System Check

Figure 5-5. shows the initial window of Boot Menu window.

Figure 5-5. Boot Menu

-Factory Setting
This item initializes the system status. If “Yes” is selected, the system is initialized to factory
shipping status. In this case, trend data is deleted completely.
Table 5-1. showes the factory default values..

Setting Parameter Factory Default

Temperature Off
Demo mode Off
ID 00000
Target Adult
Language English
Unit Temp ℃
Bright 3
Volume(beep/alarm) 2
Silence Interval 30 sec

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 29

Suspend Interval 10 Min


Remind Beep Interval 3 Min
Set Date 2000.01.01
Set Time 12:00
Data Format yy/mm/dd
Auto Mode Interval Off
STAT Mode Time 5 Min
Initial Inflate Pressure 180
Unit NIBP mmHg
Systolic Limit H:270, L:30
Mean Limit H:260, L:20
Diastolic Limit H:250, L:10
NIBP PR Limit H:250, L:20
Save Interval 1Min
Delete All Data No
SpO2 Limits H:100, L:20
PR Limits H:250, L:20
Table 5-1. Factory Default Value

- Thermometer
In this item, user can select the status of thermometer connection. If “ON” is selected, the
system recognizes that thermometer is connected.
※ In the case of SpO2 and NIBP, the system recognizes connection status automatically, but
in the case of thermometer, the user must select item manually.

- NIIBP Maintenance
This is the item related NIBP. When “Enter” is selected, the system moves to NIB
Maintenance window.

- Firmware Upgrade
This is the item related to upgrade of SpO2/NIBP module’s firmware through USB memory
stick. When “SpO2” or “NIBP” is selected, the system moves to F/W upgrade window of
each module.

- Demo mode
In this item, user can select Demo mode status (On/Off). When “On” is selected, SpO2 is
working with Demo mode in Main window.

- F/W Ver. : In this item user can check Firmware Version of each module.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 30

- Power check : In this item, user can check the power status of the system.

5.5 NIBP Maintenance Menu

In NIBP Maintenance menu, user can set items related the maintenance and management of
NIBP.
Figure 5-6.shows NIBP Maintenance menu.

Figure 5-6. NIBP Maintenance Menu

- Calibrate Pressure
This is the item to calibrate NIBP pressure sensor. Before using this item, connect a
manometer, cuff and a manual/automatic pressure device to NIBP connector. Using “Y” or “T”
connector, connect a manometer, cuff and a pressure device to the system.
.

Figure 5-7. Calibrate Pressure sensor

After setting the outer measurement device, push “Enter” button. “200mmHg” item will be
activated. Calibrate a manometer’s pressure value as “200mmHg”. Select “Set” item by
rotating knob and push knob. Now setting for standard value (200mmHg) of NIBP Pressure
sensor is completed. After setting of standard value (200mmHg), setting item for standard
value of 0mmHg will be activated automatically. To set standard value of 0mmHg, it needs to
apply pressure to 0mmHg. After stopping decompression, when manometer indicates 0mmHg,
operate knob in the same way for setting 200mmHg standard value. When NIBP Pressure

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 31

sensor’s 0mmHg standard value setting is completed, all pressure is eliminated and the mode
goes to previous step of Calibrate Pressure selection. With this NIBP Pressure sensor’s
calibration is completed.

NIBP Pressure sensor’s calibration must be performed by and


expert. Do not turn off the device while calibrating. (Please
Caution use AC power.)

- Check Pressure
This item is used for the check whether NIBP pressure sensor sends the exact values or not.
And using this item, the user can test the Over pressure measurement function. If “Yes” is
selected, the number and unit (mmHg) are activated. Now pressurize to NIBP connector, and
then the pressure value is displayed. If more than 300mmHg is pressurized, warning phase is
displayed and the system escapes this item after pressure exhaust.

5.6 System Setup Menu

User can enter System setup window by selecting Menu / System setup item. In System setup
window, user can set items relevant to the overall system operating and management.
Figure 5-8 shows System setup window.

Figure 5-8. System Setup Menu

- ID : User can set to 5 unit numbers for patient’s ID.

- Patient Mode : Select the object to measure. User can select “Adult” or “Neonate”.
According to the object, the measurement range is changed.

- Language : Select the language for user.

- Alarm : Set system alarm. If “Enter” is selected, the system goes to window which can
select item relevant to alarm.
- Clock : Set system clock. If “Enter” is selected, the system goes to window which can select

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 32

item relevant to clock.

- Temp. Unit : Set temperature unit. User can select the unit as ‘˚C’ or ‘˚F’.

5.7 Alarm Menu

User can enter Alarm window by selecting alarm item in System setup window. In Alarm
window user can set items relevant to alarm sound.
Figure 5-9. shows Alarm window.

Figure 5-9. Alarm Menu

- Alarm Volume
This item is used for control alarm sound. The control steps are composed with 6 steps and
the sound becomes louder as the number is increased. But the system does not support the
mute.

- Silent Interval
This item is used for setting Alarm sound silence interval. User can select the one among 30,
60, 90 and 120 seconds. During setting time the alarm is silent. After setting time is over, the
alarm sounds again.

- Suspend Interval
This item is used for setting Alarm sound suspend interval. User can select the one among
10, 20, 30 and 60 minutes. During setting time the alarm sound is suspended. After setting
time is over, the alarm sounds again.

- Remind Beep Interval


During Alarm sound suspend mode, user can set Remind beep interval. User can select the
one among off, 3, 5 and 10minnutes.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 33

5.8 Clock Menu

User can enter Clock menu by selecting clock item in System setup window. User can set
items relevant clock in Clock window.
Figure 5-10. shows Clock menu.

Figure 5-10. Clock Menu

- Set Data : User can set date. (yy(year)/ mm(month) /dd(day))

- Set Time : User can set time.(h(hour)/m(min)/s(sec))

- Date Format : User can change the date display format.

5.9 NIBP Setup Menu

User can enter NIBP setup window by selecting NIBP item in Menu. User can set items
relevant to NIBP in NIBP setup window.
Figure 5-11.shows NIBP setup window.

Figure 5-11. NIBP Setup Menu

Refer to the Menu clause of NIBP Monitoring about indications for each Menu.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 34

5.10 Trend Data Menu

User can enter Trend Data menu by selecting Trend Data item in Menu. User can set items
relevant to Trend Data in Trend Data menu.
Figure 5-12.shows Trend Data menu.

Figure 5-12. Trend Data Menu

- Save Interval : User can set Save interval of Trend Data. It is possible to set the interval
from 1 to 60 minutes. The system saves data at the intervals of setting time. The system can
save 4088 data.

- Browse Data : User can view the saved Trend Data with the passage of time.

- Delete All Data : User can delete all saved Trend Data.

5.11 Browse Data Menu

User can enter Browse data menu by selecting Browse Data item in Trend data menu. User
can check the saved results (measured values) in Trend data window.
Figure 5-13.shows Browse datamenu.

Figure 5-13. Browse Data Menu

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 35

- Path : Indicate window’s location and name.

- Save Date/Time : Indicate the date and time when Trend data is saved.

- Save Data : Indicate the saved results(measured values).

- Data move : Shift the page of the saved data.


‘↓’ : Shift this page to the below part.
‘↑’ : Shift this page to the upper part.

- Window move : Move to the upper step window or escape Menu.

5.12 SpO2 Setup Menu

User can enter SpO2 Setup menu by selecting SpO2 Setup item in Menu. User can set items
relevant to SpO2 setup in SpO2 Setup menu.
Figure 5-14. shows SpO2 Setup menu.

Figure 5-14. SpO2 Setup Menu

Refer to the Menu clause of SpO2 Monitoring about indications for each Menu.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 36

5.13 Menu Tree


Menu Tree

Factor Setting

Thermometer

NIBP Maintenance
Boot Menu
Firmware Upgrade

Demo Mode

F/W Ver

Power Check

ID

Patient Mode

Language
System Setup
Alarm Sound

Clock

Thermometer Unit

Brightness

Auto Mode Interval

STAT Mode Time

Menu
Initial Inflate Pressure

NIBP Unit
NIBP Setup
Systolic Limit

Mean Limit

Diastolic Limit

NIBP PR Limit

Save Interval

Trend Data Browse Data

Delete All Data

SpO2 Limit
SpO2 Setup
PR Limit

Figure 5-15. Menu Tree

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 37

5.14 Alarms and Sound

When BT-700N is under using, please check Alarm limits


Warning setting is suitable for the status of patients.

Using the same or similar devices with different setting for


Warning alarm in the same area may result in the potential risk factors.

1. General

In the case of BT-700N measured a particular matter that the user have to know from the
patient, BT-700N goes to Alarm status. BT-700N indicates the alarm status as follows.

- Visual alarm indication


- Audible alarm indication
- Nurse call indication

2. Changing Alarm Volume

The user can adjust the alarm volume from 1 to 6 levels in Menu/System Setup/Alarm Sound.

3. Visual Alarm Indication

- In the case of Visual alarm occurs, the alarm index is displayed in Status region. The alarm
is reversed periodically. If several alarms occur in the same time, all alarms are displayed in
serial order

Figure 5-16. Visual Alarm Indication (Thermometer disconnected)

Figure 5-17. Visual Alarm Indication (Thermometer connected)

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 38

5.15 Alarm Priority and Message

Alarm sounds consist of High, Middle and Low sound. The highest level alarm sounds
among occurring alarms at present.

Priority Condition Messages


SpO2 probe is not connected. Probe Off
SpO2 signal is not detected. No Signal,
SpO2 or NIBP module is disconnected. Module Disconnect
High
Faulty SpO2 sensor is connected Sensor error
Battery voltage falls to below the fixed
Cri-Low Battery
voltage.
SpO2, NIBP High Alarm limit over. High PR/SpO2/Sys p/.. Limit
SpO2, NIBP Low Alarm limit over. Low PR/SpO2/Sys p/.. Limit
SpO2 wave is detecting. Check Signal
NIBP measuring is failed. NIBP Measure Error
Medium NIBP cuff pressure is over 300mmHg. NIBP Overpressure

NIBP measuring time is over. NIBP Time Over

NIBP cuff is disconnected. NIBP Cuff

Battery voltage falls to below. Low Battery


Low
Alarm is suspend. Alarm Suspend
Table 5-2. Alarm Priority

The sound according to Alarm priority.

Priority Sound

High Priority “BeepBeepBeep--BeepBeep”

Medium Priority “BeepBeepBeep”

Low Priority “Beep-Beep”


Table 5-3. Alarm Sound

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 39

5.16 Audible Alarm Suspend/Silence

Do not suspend or make silence alarm if patient’s safety is not


guaranteed. If the alarm is suspended or silent, the alarm does
Warning not sound during setting time.

To stop the alarm sound, user can set the alarm to Alarm Suspend or Alarm Silence. If the
alarm is suspended or silent, the alarm does not sound during setting time.

Note : If new alarm occurs, Alarm Suspend or Silence mode is released automatically.

- Alarm silent : Press button once shortly in Alarm on status, then the system goes to Alarm
silent. (1 Beep)
- Alarm suspend : Press button once long in Alarm on status, then the system goes to Alarm
suspend. (2 Beep)
- Alarm on : Press button once shortly in Alarm Silent or Suspend status, then the system goes to
Alarm on.(1 beep)

1. Alarm Silence
- Press Alarm button shortly in front panel. The system enters to Alarm silence mode. (1 key
beep)
- To release Alarm silence mode, press Alarm button shortly in front panel. (1 key beep)
- When Alarm silence, the front LED is blinking. (When Alarm icon is displayed, Icon and
LED are blinking at the same time.)
- User can select Alarm Silence interval among 30, 60, 90 and 120 seconds.

2. Alarm Suspend
- Press Alarm button about 2 seconds in front panel. The system enters to Alarm suspend
mode. (2 key beep)
- To release Alarm suspend mode, press Alarm button shortly in front panel. (1 key beep)
- When Alarm suspend, the front LED is off. At the same time, the visual alarm indication –
Alarm Suspend - is displayed. (When Alarm icon is displayed, if the icon is displayed with
“X” signal and red, LED is off.)
- User can select Alarm Suspend interval among 10, 20, 30 and 60 minutes.

5.17 Remind Beep

If the device is set to Alarm Suspend, this sound informs user that the alarm is suspended. In
Menu/System Setup/Alarm Sound, the user can set the interval by 3, 5 and 10 minutes or off
the remind beep.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 40

Chapter 6
SpO2 Monitoring
To get the best performance and exact measuring values, please
use accessories which is supplied with the device or is
Warning recommended by Bistos Co.,Ltd. Use of any other accessories
may result in out of specification performance.
If SpO2 sensor does not equipped to the patient correctly, it may
Warning result in inaccuracy of the device.
Do not use a damaged SpO2 sensor. Do not expose a sensor to
Warning water or dampness. It may result in patient’s harzards.
The following environment can cause measurement errors.

 In the case of using SpO2 sensor which is not supported by


a manufacturer or is out of the provided standard.
 A serious Dysfunctional hemoglobin
(Carboxyhemoglobin or Methemoglobin)
 In the case of using intravascular dye with indocyanine
green or methylene blue.
 In the case of the exposure to strong light sources such as
Surgical lamps, Bilirubin lamps, Fluorescent lights,
Infrared heating lamps and Direct sunlight.
 In the case of the happening of patient’s excessive
Warning movement.
 In the case of the interference is happened by an
electrosurgical equipment or a defibrillator.
 In the case of using the device to a vein.
 In the case of a SpO2 sensor is located to the end of the
NIBP cuff, arterial catheter, or intravascular line.
 In the case of the patient has symptoms such as
hypotension, severe vasoconstriction, server anemia,
hypothermia, cardiac arrest, or shock.
 In the case of the device using environment is unstable.
 In the case of using a sensor extension cable which
exceeds the regulated length.

When the alarm occurs which is related to sensors such as


SpO2 sensor Probe off or Check signal, please check the status
Caution of SpO2 sensor or installation to patient. In case of need,
please change a SpO2 sensor.
If a reusable sensor is installed in a long time to the patient’s
measuring point, please check the skin status of the patient and
Caution the installed position of the sensor.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 41

6.1 Setup Connections

1. Determine a suitable SpO2 sensor to the patient.

2. Connect extension cable to SpO2 connector in the left panel.

3. After connect SpO2 and extension cable, lock the connection by taking down a transparent
cover.

4. Install SpO2 sensor to the patient’s finger such as figure 6-1. Be cautious about the position
of SpO2 sensor. The inaccurate position of the sensor may result in inaccurate measuring
result.

Figure 6-1. SpO2 Sensor Position

6.2 Description of SpO2 Display and Menu

1. Display
The values of SpO2 and PR(Pulse rate) that is extracted from SpO2 wave are displayed
with the relevant unit(%, bpm). If there are no signals, the device will display the sign “---”
instead of value. In the alarm situation, the values will be displayed with red color.

Figure 6-2.shows the values in SpO2 region.

Figure 6-2. SpO2 Region

- Heart icon : This icon will be displayet during the heart rate is measuring.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 42

- SpO2 bar : This bar indicates the intensity of the SpO2 signal.

The follow shows the SpO2 wave.

SpO2 wave region

Figure 6-3. SpO2 Wave


2. Menu
The user can set items related to SpO2 alarm limit in Menu/SpO2 Setup.

Figure 6-4. SpO2 Setup Window

- SpO2 Limit : In the case of the measured SpO2 value is out of definite range, the user can
set the range to make alarm.
- PR Limit : In the case of the measured SpO2 PR value is out of definite range, the user can
set the range to make alarm..

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 43

Chapter 7
NIBP Monitoring
To get the best performance and exact measuring values, please
use accessories which is supplied with the device or is
Warning recommended by Bistos Co.,Ltd. Use of any other accessories
may result in out of specification performance.
The followings may result in inaccurate measuring result.
 In the case of the installing cuff to the patient with too tight
or loose.
 In the case of the using not appropriate cuff size to the
patient.
Warning  In the case of the cuff location is different from the heart’s.
 In the case of the air leakage from cuff and hose.
 In the case of the patient’s movement exceed the measuring
standard.
In some cases, rapid, prolonged cycling of an oscillometric,
noninvasive blood pressure monitor cuff has been associated
with any or all of the following : ischemia, purpura, or
neuropathy. Periodically observe the patient’s limb to make sure
Warning that the circulation is not impaired for a prolonged period of
time. Also make sure the cuff is placed according to directions
in this manual and the cuff directions for use.
Do not place the cuff, the catheter or SpO2 sensor on an
extremity being used for intravenous infusion or any area where
Warning circulation is compromised or has the potential to be
compromised
AS with all automatically inflatable blood pressure devices,
continual cuff measurements can cause injury to the patient
Warning being monitored. Weigh the advantages of frequent
measurement and/or use of STAT mode against the risk of injury
Minimize the patient’s movement during measuring NIPB. The
Warning patient’s movement may result in inaccurate measurement.
Do not use the cuff if there is trace of the vein injection or there
Warning is a damage or damage possibility of blood circulation.
Never measure neonate with the setting for adult. Although the
cuff is suitable for neonate, the pressure standard for adult
Warning exceed the pressure standard for neonate.

In auto mode, the last measured value is displayed. This value


Caution does not reflect the patient’s status during measuring interval.
If NIBP cuff measurement and SpO2 measurement are
performed in the same patient and measuring point, this may
Caution result in inaccurate SpO2 measurement by the cuff’s
pressurization.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 44

Be cautious about hose’s twist or imperforation.


Caution

7.1 Setup Connections

1. Determine a suitable NIBP cuff to the patient.

2. Connect extension cable to NIBP connector in the left panel. The cuff part and connecting
part of extension cable to the device have same shape.

3. Wrap the cuff to the patient’s arm. The cuff must be wrapped to skin or thin cloth. If the
cuff is wrapped to thick cloth, it may result in inaccurate result.

Please wrap the cuff tight to the point of inserting two fingers.
Caution

4. Adjust patient’s arm with the palm is head to upside. Adjust the cuff with the “ARTERIA”
sign is on the patient’s artery. Check the cuff is suitable for the patient. Be cautious about the
hose’s bending. Hose must be located with parallel to the patient’s arm.

5. Adjust and maintain the patient’s arm to the same height of heart.

Figure 7-1. NIBP Cuff Position

Note : Measuring blood pressure can be influenced by external factors such as the patient’s
position, measuring part and the status of patient.
Note : If the patient has arrhythmia, the heart rate is unstable. In this case, it is difficult to
measure NIBP or the result of measuring can be inaccurate.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 45

7.2 NIBP Measurement Modes

1. BT-700Nsupports 3 modes to measure blood pressure.

- Manual mode : Measure blood pressure once and stop to measuring.

- Auto mode : BT-700N measures blood pressure repeatedly at the interval of setting time.
Setting interval time is off, 1, 2, 3, 5, 10, 15, 30, 60 and 90 minutes.

- STAT mode : During setting time, BT-700N measures blood pressure repeatedly as many as
possible. Setting interval time is 1, 2, 3, 4, 5 and 10 minutes.

2. The follows are the starting method of each measuring mode.

- Manual mode : In the status of Auto Mode Interval is set to off in Menu/NIBP Setup item,
press NIBP Start/Stop key shortly once. (1 beep sounds.)

- Auto mode : In the status of Auto Mode Interval time is set in Menu/NIBP Setup item

( is displayed in Main window), press NIBP Start/Stop key shortly once. (1 beep

sounds and Icon is rotating.)

- STAT mode : Independent to Manual or Auto mode, press NIBP Start/Stop key about 2
seconds (2 beep sound.). Icon is displayed in Main window.

3. The follow is stopping method of each measuring mode.

- After starting Manual / Auto / STAT mode, press NIBP Start/Stop key once more and then
this function will stop.

7.3 Description of NIBP Display and Menu

1. Display

The values of Systolic, Diastolic and MAP are presented with the relevant unit (mmHg
or kPa). In the case of the value of NIBP PR, the value of PR that is measured from
NIBP signal is presented. If the value is not measured, “---” is displayed in the region
which is displayed measured value instead of measured value. In the alarm status, the
corresponding value is displayed with red. In addition the value of Initial inflate pressure,
NIBP mode icon and the measuring time of NIBP are displayed.

Figure 7-2. shows relevant values and items in NIBP region.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 46

Figure 7-2. NIBP Region

- Initial Inflate Pressure : The system pressurize to the setting pressure from the initial
pressure.(provided, if the significant wave(systolic part) is detected the system will stop the
pressurization and start to measure the pressure.)

- NIBP mode
a. Manual mode : There in no icon displayed.
b. Auto mode : icon is displayed.
c. STAT mode : icon is displayed.
(provided, in the case of the device is set to Auto mode, icon is displayed on
icon with overlapped.)

- NIBP timer : In Auto mode, the device counts up the total measuring time(This is similar
with stop watch.). In STAT mode, the device counts down the measuring time from the
start time(This is similar with a timer.).

2. Menu

The device enters to NIBP setup window by selecting Menu / NIBP setup item. The user can
set items relevant to NIBP in NIBP setup window.
Figure 7-3. shows NIBP setup window.

Figure 7-3. NIBP Setup Window


- Auto Mode Interval : During NIBP operating mode, user can set interval relevant to Auto

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 47

mode. If the time is set to voluntary value except “Off”, the device mode is changed to Auto
Mode from Manual Mode. Setting range is from “Off” to 90 minutes.

- STAT Mode Time : User can set the operating time of STAT mode from 1 to 10 minutes.

- Initial Inflate Pressure : User can set the initial pressurization value to measure NIBP.
During pressurization, if the wave of measured NIBP is detected before the initial
pressurization value, the device stops pressurization and starts to measure NIBP. If the device
determines that the pressurization to measure must be higher than the initial inflating pressure,
the device performs re-pressurization.

- NIBP Unit : User can select NIBP unit between “mmHg” and “kPa”.

- Systolic Limit : In the case of Systolic value(one of the measured NIBP values) is out of
definite range, user can set the range to make alarm.

- Mean Limit : In the case of Mean value(one of the measured NIBP values) is out of definite
range, user can set the range to make alarm.

- Diastolic Limit : In the case of Diastolic value(one of the measured NIBP values) is out of
definite range, user can set the range to make alarm.

- NIBP PR Limit : In the case of NIBP PR value(one of the measured NIBP values) is out of
definite range, user can set the range to make alarm.

Note : NIBP operating mode


- Manual mode : The device measures NIBP once and stops to measure NIBP. If user push
NIBP operating key shortly (1 Beep) when Auto Mode Interval item is off, the device will
operate with Manual mode.

- Auto mode : When the time is set in Auto mode Interval, the system measures blood
pressure repeatedly at the intervals of setting time. If total measurement time is longer than
setting time, 30 seconds after end of the prior measuring step the device measures blood
pressure again. Press NIBP operating key shortly (1 Beep) then the device starts to measuring.

- STAT mode : During the setting time, the device measures blood pressure repeatedly as
many as possible. The measuring interval is two seconds. The device will operate with STAT
mode by pressing NIBP operating key long(2 Beep) in Manual or Auto mode.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 48

Chapter 8
Maintenance
8.1 Maintenance and Cleanliness
1. The user can clean and maintain BT-700N by various methods. But please follow methods
that Bistos Co.,Ltd. recommends to avoid a damage or pollution.
2. If the materials (not authorized materials) which can cause damage to the device are used,
it is impossible to get free repair service although the device is in warranty period.

After cleaning the device, please check the body and sensor carefully. If
Caution the device is decrepit or damaged, do not use that device.

3. Please clean the body and measuring sensor by alcoholic soft fabric once in a month. Do
not use lacquer, thinner, ethylene and oxidizer.
4. Please maintain cables with no dust and soil. And clean cables with wet fabric (water
temperature is about 40˚C / 104℉.). Please clean cables using clinical alcohol once in a week.
5. Do not immerse the device or sensor in liquid or a cleanser. Also all liquid must not be
permitted to enter in the device or sensor.

8.2 Accessory cleaning


Please use clinical sterilizer to sterile or clean the body, sensor and thermometer.

8.3 Regular Inspection


Similar to most medical equipment, BT-700N has to be inspected once in a year. Refer to
the service manual which is supported by Bistos Co.,Ltd. about inspectional items.

Regular inspection must be performed by the company’s technician or the


technician who is authorized by manufacturer. The user and operator must
Warning not disjoint or remodel the device.

8.4 Disjoint and handle the internal battery


The internal battery type is Ni-MH battery. Therefore if disjoint the battery, please be
cautious about disjointing.
- The internal battery must be handled by the company’s technician only.
- Please keep the internal battery below 60 ℃ and if an electrolyte from battery is contacted
to the clothes or skin, wash the polluted parts by clean water immediately.
- If you want to change the internal battery, please contact the head office.

Do not dispose the internal Ni-MH battery as a general waste. If it is


dispose as a general waste, it may results in human and environmental
Caution harms. When disposing of internal Ni-MH battery, adhere to all applicable
laws regarding recycling.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 49

Chapter 9
Troubleshooting

Prohibit to disjointing the device by not authorized person. We


do not compensate for a malfunction or damage for human life
Warning by a voluntary disjoint.
In the case of not intended result is displayed, please check bio-
Warning signals of the patient or the status of the patient.

When the following problems are happened, before contact the head office, please check
following measures.

1. When you can’t power on the device.


- Please check that an adaptor of the device is connected to the AC connector properly.
- Please check the LED in the front panel is green or yellow.
- Please change an adaptor.

2. When the measured value is not displayed in LCD.


- Please check the connection of the sensor or expanded cable to the connector in left panel
properly.
- After connect the sensor to connector of the device and check the operation of the sensor.

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 50

Chapter 10
Specification
10.1 Mechanical and Electrical Specifications
Dimension 250(W) x 114(L) x 90(H) (Unit : mm).

Weight 2.5kg Under (Include battery).

Display 320 x 240 TFT color LCD.


Screen dimension 64.8(W) x 48.6(H) (Unit : mm).

Communication PC communication with RS-232 (with Nurse call state).

Firmware upgrade Connect PC to USB port.

Power Medical adaptor.


Input AC100 – 240V 50/60H, 0.4A.
Output +9V DC / 1.5A.

Battery operating time SpO2 Only : Min 7 hours depending on use.


NIBP Only : Min 4.5 hours depending on use.
SpO2 + NIBP +Temp. : Min 3 hours depending on use.
Battery recharge time 90% 8 hours.

Memory 4088 data can be stored.

※ Simultaneously operates monitor and charges battery when AC power adapter in plugged

10.2 Compliance
Classification Class II Internally powered

Type of protection Type BF - Applied part

Mode of operation Continuous

Degree of protection : NIBP IPX1,


SpO2 sensor IPX2
Thermometer IPX0

10.3 Environmental Conditions


Operation Temperature 10°C ~ 35°C (50°F ~ 95°F)

Storage Temperature –20°C ~ 60°C (–4°F ~ 140°F)

Humidity 20% to 90%, Non-condensing

Altitude 0 -3048m (0 -10,000 ft)

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 51

10.4 Performance Specifications


1. Noninvasive Blood Pressure (NIBP)
Technique Oscillometric Measurement

Measurement modes AUTO, MANUAL and STAT

AUTO mode Automatic NIBP measurements at interval of


1,2,3,5,10,15,30,60,90 minutes

MANUAL mode Single measurement initiated by NIBP Start/Stop button

STAT mode Series of consecutive measurements for (1,2,3,4,5,10) minutes

2. NIBP measurement range


Systolic pressure range Adult /Pediatric : 30 mmHg to 250 mmHg
Neonate : 20 mmHg to 120 mmHg

Diastolic pressure range Adult /Pediatric : 20 mmHg to 210 mmHg


Neonate : 10 mmHg to 110 mmHg

Mean pressure range Adult /Pediatric : 20 mmHg to 235 mmHg


Neonate : 10 mmHg to 110 mmHg

Pressure display accuracy Meets ANSI/AAMI SP 10:2002 +A1 : 2003

Cuff pressure range 0 to 300 mmHg (0 to 40kPa)

Initial cuff inflation


Adult/Pediatric:
120, 140, 160(Default), 180, 200, 220, 240, 260, 280 mmHg
(16.0, 18.7, 21.3(Default), 24.0, 26.7, 29.3, 32.0, 34.7, 37.3 kPa)

Neonate :
80, 90, 100(default), 110, 120, 130, 140 mmHg
(10.7, 12.0(Default), 13.3, 14.7, 16.0, 17.3, 18.7 kPa)

Overpressure protector Adult /Pediatric : 300mmHg ~ 330 mmHg


Neonate : 150 mmHg ~ 165 mmHg

Pulse rate range Adult/Pediatric : 40 bpm to 220 bpm


Neonate : 40 bpm to 240 bpm

Pulse rate accuracy ±2 bpm or ±2%, whichever is greater

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 52

3. Pulse Oximetry(SpO2)
Measurement range
SpO2 : 0 to 100%
Pulse rate : 30 to 250 bpm

Saturation without motion


Adults/Pediatric : 70 to 100% ±2 digits
Neonate : 70 to 100% ±3 digits

Saturation with motion Adults/Pediatric, Neonate : 70 to 100% ±3 digits

Low perfusion 70 to 100% ±2 digits, 0 to 69% unspecified

Pulse rate
Without motion : 30 to 250 bpm ±3 digits
With motion : normal physiologic range 55 to 125 bpm±5 digits
Low perfusion : 20 to 250 bpm ±3 digits

4. Thermometer
Measurement method Infrared measurement, Non-contact measurement

Operation temperature 15 ~ 40℃ (59 ~ 104℉), Relative humidity ≤ 95%

Storage temperature 20 ~ 50℃ (-4 ~ 122℉)

Measurement distance 2~3cm from the forehead center

Measurement range
Body temperature : 34.0 ~ 42.5℃ (93.2~108.5℉)
Accuracy ±0.2℃ (36.0 ~ 39℃
±0.3℃ (34.0 ~ 35.9℃ ℃

Wide temperature : 15~60℃ ℉


Accuracy ±2℃ or ±2%

Response time 2 seconds

Resolution 0.1℃

Memory Automatically body temperature data(maximum 32) stored

Back-light If measurement complete, lighted 3 seconds

Automatic power off If not press any button in a minute, automatically off

Power supply DC3V(AAA 1.5V Battery 2ea)

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 53

Chapter 11
Warranty
Product Name Patient monitor

Model Name BT-700N

Permission No. 1) 제허 09-368 호

Serial No.

Warranty Period 2 years from purchasing date

Date of purchase

C Hospital :
O
S Address :
T
M Name :
E
R Contact :

Agent / Dealer

Manufacturer Bistos Co., Ltd.

※ Thanks for your purchasing our product.


※ This product is manufactured and passed through strict quality control system and inspection.
※ Compensation standard concerning repair, replacement, refund of the product complies with
“Consumer’s protection low” noticed by Economic Planning Dept.
※ If an abnormal condition is happened with normal use in the term of guarantee, the service
center will offer free repair service.
※ Please inform us model, the serial number, purchasing date and problems when the problem is
happened in the term of guarantee.

Note 1) ‘Permission No.’ is a identify number by Korean Food and Drug Administration.

Manufacturer Information

Head office / 7th Fl., A Bldg., Woolim Lions Valley 5-cha,144-3,


Manufacturing Sangdaewon-dong, Jungwon-gu, Seongnam-si,
site Gyeonggi-do, Korea
Manufacturer Telephone 82-31-750-0340
Tele fax 82-31-750-0344
Web site www.bistos.co.kr
e-mail bistos@bistos.co.kr

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)


BT-700N Operator Manual 54

Contact to Technical support

Telephone: +82 31 750 0340


Fax: +82 31 750 0344

Bistos Co., Ltd.


th
7 Fl., A Bldg., Woolim Lions Valley 5-cha, 144-3,
Sangdaewon-dong, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Korea

www.bistos.co.kr
bistos@bistos.co.kr

Model Name: BT-700N

EC Representative: Medical Econet Gmbh


Im Erlengrund 20 / D-46149 Oberhausen / Germany

Telephone: +49 (0)208 377 890-0


Fax: +49 (0)208 377 890 55

P/N : OPM(BT-700N)EN Bistos Co.,Ltd. Rev. 02(2013.04.12)