You are on page 1of 11

Lab on a Chip

View Article Online


PAPER View Journal

Skin-interfaced microfluidic devices with one-


Cite this: DOI: 10.1039/d0lc00400f
opening chambers and hydrophobic valves for
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

sweat collection and analysis†


Yingxue Zhang,‡ab Yao Chen,‡a Jielong Huang,a Yangchengyi Liu,a Jinfeng Peng,c
Shangda Chen,a Kui Song,d Xiaoping Ouyang,a
Huanyu Cheng *b and Xiufeng Wang *ae

Soft, skin-interfaced microfluidic platforms are capable of capturing, storing, and assessing sweat chemistry
and total sweat loss, which provides essential insight into human physiological health. However, sweat loss
from the outlet of the microfluidic devices often leads to deviation of the measured concentration of the
biomarker or electrolyte from the actual value. Here, we introduce hydrophobic valves at the junction of
the chamber and the microfluidic channel as a new chamber design to reduce sweat evaporation. Because
the advancing front of the liquid in the hydrophilic microchannel is blocked by the hydrophobic valve, the
fluid flows into the chambers, forms the initial meniscus, and completely fills the chambers along the initial
meniscus. Fluid dynamic modeling and numerical simulations provide critical insights into the sweat
sampling mechanism into the chambers. With significantly reduced evaporation and contamination, the
Received 17th April 2020, sweat sample can be easily stored for a long time for later analysis when in situ analysis is limited.
Accepted 1st June 2020
Additionally, the design with multiple chambers can allow sequential generation of sweat collection at
DOI: 10.1039/d0lc00400f
different times for long-term analysis. The in situ real-time measurements of the sweat loss and pH value
analysis from the human subject demonstrate the practical utility of the devices in collecting, storing, and
rsc.li/loc analyzing the sweat generated from sweat glands on the skin.

1. Introduction For instance, local and full-body sweat loss rates provide
essential insights into the hydration state of an individual.6,7
Recent advances in materials, mechanics design, and device Quantitative measurements of the sweat rate can prevent
architectures have spurred the rapid development of devices heat-related illnesses, including heat stroke and heat
that are thin, soft, and mechanically compliant. Capable of exhaustion, in hot environments.8 Because the pH of sweat is
interfacing with the human epidermis, this class of emerging associated with metabolic alkalosis, noninvasive
electronics offers continuous recording of vital physiological measurement of metabolic alkalosis from perspiration is
parameters with clinical quality for health and performance possible.9,10 Additionally, sweat electrolytes and metabolites
assessment. Sweat, which can be measured noninvasively and (e.g., lactate, chloride, and glucose) can be used to inform
contains various biomarkers, is starting to gain popularity for electrolyte balance, blood glucose levels,9,11 hydration status,
physiological health monitoring and disease diagnostics.1–5 and overall health.12
Early developed sweat collection devices rely on absorbent
pads made from filter paper, cotton, gauze, or toweling.13
a
School of Materials Science and Engineering, Xiangtan University, Xiangtan, After separating the sweat from the absorbent pads by
Hunan 411105, China. E-mail: onexf@xtu.edu.cn centrifugation, sweat analysis is performed with
b
Department of Engineering Science and Mechanics, The Pennsylvania State
corresponding benchtop equipment.14 However, this process
University, University Park, PA 16802, USA. E-mail: huanyu.Cheng@psu.edu
c
School of Mechanical Engineering, Xiangtan University, Xiangtan, Hunan 411105,
requires expensive equipment and trained professionals to
China collect and analyze the data.15 As an alternative, wearable
d
College of Civil Engineering and Mechanics, Xiangtan University, Xiangtan, microfluidic devices open opportunities to collect and analyze
Hunan 411105, China sweat in situ for personalized medicine, improved sports
e
Institute of Flexible Electronics Technology of Tsinghua University, Jiaxing,
performance, and military readiness.16–21 For commonly used
Zhejiang 314006, China
† Electronic supplementary information (ESI) available. See DOI: 10.1039/
microfluidic channels that are hydrophobic, the sweat must
d0lc00400f overcome the positive capillary pressure to enter the
‡ Y. Z. and Y. C. contributed equally to this work. microchannels for capture and collection. In contrast, the

This journal is © The Royal Society of Chemistry 2020 Lab Chip


View Article Online

Paper Lab on a Chip

hydrophilic microfluidic channels that are associated with L. Riegger et al.29 were generated by enhancing the
negative capillary pressure spontaneously wick sweat into the hydrophobicity of the PDMS surface, which is associated with
microchannels more rapidly by capillary action. complex fabrication processes and a much higher bursting
Due to the inherently low concentrations of biomarkers pressure (2.5–4.5 kPa). Because this bursting pressure is
and electrolytes (e.g., sodium, chloride, and potassium), their higher than the sweat pressure, hydrophobic valves cannot
measurements are susceptible to evaporation.22 Silicone be easily integrated into microfluidic devices for sweat
elastomers such as poly(dimethylsiloxane) (PDMS) are collection. In contrast, endowing the rest of the microfluidic
commonly explored as microfluidic devices with an ultrathin device with hydrophilicity allows us to directly explore the
capping layer of 150 μm for sweat sampling; they are use of naturally hydrophobic PDMS in hydrophobic valves.
associated with ∼100% water loss within 3 hours at 37 °C.12 The fabrication process is simple and effective to generate
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

However, PDMS is still commonly used as a microfluidic hydrophobic valves with long-term stability. Additionally, the
material due to its biocompatibility, optical transparency, hydrophilic microfluidic devices in other locations can also
ease of patterning, and favorable mechanics. Depending on spontaneously wick sweat into the microfluidic channels.
the design of the microfluidic device, the measured Here, we introduce hydrophobic valves at the junction of
concentration of the biomarker or electrolyte can either the chamber and the microfluidic channel as a new chamber
slightly or greatly deviate from the actual value because of design to reduce sweat evaporation. The hydrophobic valves
water loss. Most of the previous designs of microfluidic sweat prepared by air plasma treatment guide the sweat flow to fill
sampling devices exploit chambers with valves2,3,22 as well as the chambers sequentially. Because the advancing front of
one inlet for sweat collection and one outlet connected to the the liquid in the hydrophilic microchannel is blocked by the
surroundings for releasing the air in the chambers. While the hydrophobic valve, the fluid flows into the chambers, forms
outlet avoids backpressure, it also leads to sweat evaporation the initial meniscus, and completely fills the chambers along
because most evaporation in the chambers occurs through the initial meniscus. Fluid dynamic modeling and numerical
the outlet.23 Thus, a novel design of a microfluidic sweat simulations provide critical insights into the mechanism of
sampling device to reduce unnecessary evaporation and sweat sampling into the chambers. With significantly
contamination of sweat in the collection chambers is vital for reduced evaporation and contamination, the sweat sample
accurate sweat analysis. can be easily stored for a long time for later analysis when in
As one of the most critical components in microfluidic situ analysis is limited. Additionally, the design with multiple
devices, valves are commonly used to control the direction of chambers can allow sequential sweat collection at different
the flow.24 Traditional valves with external device times for long-term analysis.
components to drive liquid flow are bulky and thus
challenging for direct integration on the skin.25 In contrast, 2. Results and discussion
passive valves without active components or power sources
allow for fluid control, presenting an attractive option for 2.1. Soft microfluidic device with hydrophobic valves for
wearable devices. Representative passive valves include sweat collection
capillary bursting valves,26 super-absorbent polymer valves,22 Fig. 1a presents an exploded view of the novel microfluidic
and hydrophobic valves22 to guide the flow in a well-defined device designed for accurate sweat collection and analysis
sequence. Capillary bursting valves are among the most along with a schematic of its working principle. The
common passive valves in wearable devices to control the representative device is 0.84 mm thick and is designed in a
sweat flow by reducing the width of the microfluidic semicircular geometry with a diameter of 2.3 cm to minimize
channels, which are associated with higher bursting pressure. the overall footprint. The proof-of-concept device includes
The preparation of the capillary bursting valves often involves three chambers for the analysis of the chloride concentration,
lithographic processes, which complicates the fabrication pH, and glucose concentration because these are common
and substantially increases the cost in rapid prototyping.27 biomarkers in sweat. Depending on need, the number of
Based on super-absorbent polymer (SAP) materials prepared chambers can be easily modulated for the target application.
by a cross-linking reaction, the SAP valve swells upon Following the fabrication process (Fig. S1†) yielded
hydration to close the inlet and prevent further flow. microfluidic devices that consisted of three layers: 1) a
However, the preparation of the SAP material is complex, and capping layer, 2) a microfluidic layer, and 3) a medical-grade
additional steps are also needed for integration of the device. acrylic adhesive film for enhanced adhesion to the skin.
By contrast, hydrophobic valves can be readily fabricated in a Because PDMS with a low effective modulus of ∼145 kPa is
high-density array. The hydrophobic valves explored in biocompatible and optically transparent, it was explored for
previous literature reports22,28 were only used to guide the both the capping and microfluidic layers. The capping layer
flow in microfluidic channels. In contrast, we report the with a thickness of 240 μm avoided self-collapse of the
design concepts, fluid mechanics principles, and applications microfluidic channel (Fig. S2†). The microfluidic layer with a
of hydrophobic valves in microfluidic devices with one- thickness of 600 μm included a microfluidic channel (300
opening chambers which can simply achieve the chrono- μm in width and height, consistent with previous designs2)
sampling of liquid/sweat. The hydrophobic valves reported by and chambers with a diameter of 3 mm. The surface of the

Lab Chip This journal is © The Royal Society of Chemistry 2020


View Article Online

Lab on a Chip Paper

capillary pressure in the hydrophilic microfluidic channels.


The device with a small thickness, soft mechanics, and a
tapered edge avoids delamination from the skin12 and
damage upon mechanical deformations such as bending and
twisting (Fig. 1b). Additionally, the device is positioned on
the skin with little deformation or strain (e.g., forehead,
forearm, and back). In addition to robust adhesion, the
adhesive layer also helps eliminate leakage during natural
skin motions from daily activities or vigorous physical
exercise.
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

2.2. Hydrophobic valves in hydrophilic microfluidic channels


While sweat excretion generates positive pressure, it must
still overcome the capillary pressure to enter the hydrophobic
microfluidic channels. In this study, we designed and
demonstrated hydrophilic microfluidic channels with
negative capillary pressure to spontaneously wick sweat into
the channel. Following the fabrication of a single
microchannel (Fig. S1†), its surface wettability was first
Fig. 1 Schematic and optical images of an epidermal microfluidic investigated. The silicone polymer PDMS was naturally
device with one-opening chambers and hydrophobic valves (HVs) for hydrophobic, with an intrinsic water contact angle (CA) of
chrono-sampling of liquid. (a) Schematic of an exploded view of the
107° (Fig. 2a, red line). After exposing the surface to air
device, highlighting various functional layers and working principles of
the HVs. (b) Optical images of the microfluidic device on the skin and plasma treatment (APT), the PDMS microfluidic channel
under mechanical deformations such as bending and twisting.

microfluidic channel was measured to have low roughness,


and the arithmetic mean height (Sa) was calculated to be
0.43 μm (Fig. S3†), comparable to that from literature reports
of 0.2 μm.30 Compared to the height of the microfluidic
channel (i.e., 300 μm), the much smaller surface roughness
exhibited a negligible effect on the sweat collection.
Additionally, the surface roughness of the PMMA mold did
not affect the PDMS bonding (Fig. S1†) because of the
demonstrated robust device performance during use even
after intensive exercise. By blocking the sweat flow, the
hydrophobic valves help route sweat into the chambers and
then open to allow sweat flow after the chambers are
completely filled. Because the chambers are filled
sequentially (labeled T1, T2, and T3 based on their filling time
points), the microfluidic device provides effective chrono-
sampling capability for in situ analysis of various biomarkers
and electrolytes. The inlet hole with a diameter of 3 mm
(corresponding to ∼9 sweat glands26) in the watertight
adhesive layer defines the localized sweat collection regions
from the skin sweat glands. The geometry of the microfluidic
channel and chambers in the device determines the sweat
Fig. 2 Design of hydrophobic valves in hydrophilic microfluidic
collection volume. After considering the target application channels. (a) The water contact angle of the PDMS surface as a
with a prescribed monitoring time, the number of sweat function of aging time before and after various surface treatments. (b)
glands within the device inlet can then be determined. The (i) Schematic of the liquid blocked by a functional hydrophobic valve in
proof-of-concept demonstration in the current study selects 9 a hydrophilic microfluidic channel. Photographs showing the liquid in
hydrophilic microfluidic channels prepared by (ii) plasma-treated PDMS
sweat glands for the chosen device geometry and a target
or (v) plasma and PVP-treated PDMS and blocked by the HV inside the
monitoring of 1 h. The sweat is driven to flow through the channel, along with (iii) the droplet residing on the inlet hole of an
inlet hole into the microfluidic system by the positive untreated hydrophilic channel and (iv) the liquid passing over the
pressure from natural sweat excretion and the negative hydrophilic channel without a hydrophobic valve.

This journal is © The Royal Society of Chemistry 2020 Lab Chip


View Article Online

Paper Lab on a Chip

becomed hydrophilic and generated negative capillary term in eqn (1) is contributed by the droplet volume at the
pressure, which quickly wicked the liquid into the channel inlet, whereas the second term is related to the hydrophilic
(Fig. 2b-ii). In contrast, the hydrophobic microfluidic channel top wall (θ1 = 42°) and three other hydrophobic walls (i.e., the
could not imbibe the droplet to wick it into the channel 1st and 2nd terms in the parenthesis, respectively). The CA at
(Fig. 2b-iii and S4†). However, the intrinsic hydrophobicity of the HV is the intrinsic contact angle of PDMS (i.e., θA = 107°),
PDMS gradually recovered ca. 5 h after the APT (Fig. 2a, whereas the CA at the top inlet depends on the treatment of
purple line). To address this challenge, subsequent low-energy the PDMS microfluidic channel (i.e., θ0 = 107°, 90°, or 42° for
polyvinylpyrrolidone (PVP) surface modification of the APT- the untreated, APT-treated, and APT-treated with PVP-
treated PDMS was exploited. While the contact angle still modified PDMS, respectively). The impeditive function on the
increased from less than 20° to ca. 42° ± 2°, it saturated at this left-hand side of eqn (1) decreases as the contact angle
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

value and remained stable for more than 24 h (Fig. 2a, experienced by the HV increases (Fig. S7†). Because the
blue line). The hydrophilic surface of the PDMS after the bursting pressure of the HV is simply the second term in eqn
oxygen plasma treatment and PVP modification can remain (1), the bursting pressure of the HVs with a contact angle of
stable for 6 months.31 Therefore, the shelf life of the 107° is calculated to be 0.04 kPa (Fig. S8,† yellow star). When
microfluidic device is over 6 months. This long-term stability of the top wall is also rendered hydrophobic, the bursting
the hydrophilic surface is attributed to the presence of CO as pressure will increase from 0.04 kPa to 0.28 kPa, representing
evidenced in the FTIR spectra (Fig. S5†) from the PVP a simple way of tuning. As eqn (1) holds true for all three
treatment, whereas the APT-treated PDMS only temporarily types of microfluidic channels even for an extremely small
changes the density of the –OH groups (water-adsorbed) at its droplet volume (V = 0.5 μL), the HV works effectively as
surface. This hydrophilic microfluidic channel from the APT designed (Fig. 2b-ii and v), whereas the fluid flows through
and PVP modification easily induced imbibition of fluid into the entire hydrophilic channel without the HV (Fig. 2b-iv).
the channel (Fig. 2b-iv).
Modulation of the sweat flow in the hydrophilic
microfluidic channel was achieved by hydrophobic valves 2.3. Working principle of HVs and their use in sequential
(HVs), which were simply created using a mask to avoid liquid sampling
exposure of PDMS to air plasma during the APT (Fig. S1†). Because chambers in microfluidic devices are often exploited
The unexposed PDMS region under the mask remained for sweat collection, novel one-opening microfluidic
hydrophobic and formed the HV in the microfluidic channel. chambers with an HV in the bridging channel were designed
Because the PVP surface modification in the microfluidic (Fig. 3a). A unit of the microfluidic devices consists of the
channel was selective only for the hydrophilic region, the main channel, a bridging channel, a collection chamber, and
obtained HV was not affected by the PVP surface an HV. In one embodiment, the bridging channel includes a
modification. In contrast, rendering the rest of the neck microchannel (length of 80 μm and width of 500 μm)
microfluidic device hydrophilic allowed us to directly explore and two transition channels with radii of 300 μm (Fig. S9†).
the use of the naturally hydrophobic PDMS in hydrophobic The longer neck microchannel forms a narrower gas outlet
valves. The fabrication process was simple and effective to between the liquid meniscus and the bridging channel,
generate hydrophobic valves with long-term stability. which traps air to increase the backpressure in the chamber.
Additionally, the hydrophilic microfluidic device in other For instance, the trapped air in the chamber prevented liquid
locations could also spontaneously wick sweat into the collection in the chamber when the length of the neck
microfluidic channels. In the proof-of-concept microchannel was increased from 80 μm in the current
demonstration, an HV with a length of l = 1 mm and a width design to 200 μm (Fig. S9c†). On the other hand, a significant
and height of w = h = 300 μm was patterned at a specific diffusion between two different fluids occurred as the width
distance (i.e., 2 cm) away from the inlet hole (Fig. 2b-i). The of the neck microchannel increased from 0.5 mm in the
fluidic flow in the hydrophilic microfluidic channel was current design to 2 mm (Fig. S9d†). Because the preparation
successfully stopped by the HV, as the air–liquid interface in of a shorter and narrower neck microchannel was associated
the channel was observed under a microscope (Fig. S6†). The with higher fabrication complexity and cost, the neck
impeditive function (PHV) of the HV can be explained by the channel with a width of 0.5 mm and a length of 80 μm was
capillary theory (Fig. S7†):28,32 chosen in this study. After the geometries of the neck
channel and chamber were chosen, the transition channels
 1
πð1 − cos θ0 Þ2 ð4 − cos θ0 − 3 cos2 θ0 Þ 3 between the two were then determined to avoid drastic
P HV ¼ 2σ
 6V  geometry changes and reduce the fluid resistance. The
w 2h þ w design of the HV is also important to allow the release of air
þσ cos θ1 þ cos θA ≤ 0 (1)
wh wh from the chambers to avoid backpressure. Although the
geometry of the HV is determined by the microfluidic
where V is the droplet volume, σ is the surface tension of the channel (Fig. S7†), its position in the main microfluidic
liquid, and θ0 and θA are the contact angles (CAs) experienced channel also affects liquid collection in the chambers. The
by the fluid at the top inlet and the HV, respectively. The first HV (see dimensions in Fig. S9†) located in the microchannel

Lab Chip This journal is © The Royal Society of Chemistry 2020


View Article Online

Lab on a Chip Paper


Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

Fig. 3 Working principle of the hydrophobic valve to route liquid into the one-opening chamber in the microfluidic device. (a) Schematic of the
one-opening chamber with the hydrophobic valve and a bridging channel in a microfluidic device. (b) Optical images (i–iv) and (c) numerical
simulations of the hydrodynamic flow process into the one-opening chamber with a hydrophobic valve, compared to (d) optical images (i–iv) and
(e) numerical simulations of the one-opening chamber without a hydrophobic valve.

needs to be positioned close to the chamber according to the wall (Fig. 3b-ii). Due to the pinning of the air–liquid interface
analysis in the following section. at both points A and B, the liquid meniscus continued to
The capability of the novel one-opening microfluidic increase (Fig. 3b-iii) and filled the entire chamber (Fig. 3b-iv).
chambers with the HV in the bridging channel was After the liquid filled the bridging channel, the air–liquid
demonstrated through a series of optical images during a interface with sufficient pressure burst open the HV and
fluid collection process (Fig. 3b). Applying a flow rate of 1.0 continued to flow.
μL min−1 as in the previous in vitro experiments,23 red-dyed As a comparison, the one-opening microfluidic chambers
water was pumped into the hydrophilic microfluidic channel. without the HV failed to collect the sweat in the chamber, as
The relatively low flow rate enabled easy observation of the demonstrated in a series of optical images during the fluid
fluid collection. Because the HV remained closed before the collection process (Fig. 3d). The air–liquid interface first
pressure across it exceeded the threshold,33 it routed the moved through the bridging channel and formed a sloped
fluid to flow into the chamber (Fig. 3b). Observation of the liquid meniscus (Fig. 3d-ii). Continuing of the fluid flow split
air–liquid interface (i.e., the two-point contact labeled A and the air–liquid interface into two meniscuses: 1) a horizontal
B) in the microfluidic channel illustrated the fluid collection meniscus that enclosed the chamber and trapped the air
process. As the liquid initially arrived at the HV, it was inside, and 2) a vertical meniscus that continued to move in
blocked; the air–liquid interface at point B was pinned at the the main channel. Due to insufficient pressure to compress
HV, whereas the air–liquid interface at point A was forced the air in the chamber, the horizontal liquid meniscus failed
through the bridging channel into the chamber. As point A to fill the chamber (Fig. 3d-iv). The designed one-opening
moved along the chamber wall clockwise and was pinned at microfluidic chambers with and without the HV also showed
the HV, an initial liquid meniscus formed on the chamber reasonably good agreement with the 2D numerical

This journal is © The Royal Society of Chemistry 2020 Lab Chip


View Article Online

Paper Lab on a Chip


Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

Fig. 4 Functional demonstrations of the microfluidic devices (a) with black-dyed water and (b) with two colors of water dyed in red and blue (the
zoomed-in image of the 3rd chamber shown in the inset indicates minimal mixing). (c) Measurements of the water loss in the chambers with and
without an outlet, indicating reduced evaporation for improved accuracy of analyte assessment.

simulations (Fig. 3c and e), which further verified the chamber, and 2) from the 1st fluid in the chamber to the 2nd
feasibility of the proposed designs. fluid in the main microfluidic channel. As demonstrated in
Next, sequential sweat collection was demonstrated for Fig. 4b, the diffusion along these two directions occurred only
potential applications in which it was necessary to analyze at the chamber wall, with less than 5% of the total volume in
the concentration changes of various biomarkers and both the 1st and 2nd chambers (Fig. S10†). Because the
electrolytes at different time points (Fig. 4). The chrono- volume of the red-dyed water was larger than that of the 1st
sampling process was illustrated by spontaneous and chamber, it was partially collected by the 2nd chamber, with a
sequential pumping of black-dyed water into the hydrophilic slightly larger mixing of 15%. The design of the 2nd chamber
microfluidic channel (Fig. 4a and Movie S1†). The flow rate was indeed chosen to buffer the transition between the two
was found to be 60 μL min−1 as the ratio of the volume in the different fluids. Because of the micro-scale dimensions in the
microfluidic channel (0.5 μL, blue region in the following device and low sweat flow rates, molecular diffusion
figure) to the corresponding filling time of 0.5 s. Although accounted for the mixing inside the chambers.23 The
the function of the HVs may be compromised when operated diffusion (or interaction) between the two different fluids only
at an even higher flow rate, the maximum sweat rate in the occurred at the chamber wall (less than 5% of the total
inlet (5.4 μL min−1 generated by 9 sweat glands, each with volume of a collection chamber), indicating minimal
the highest rate of 0.6 μL min−1 per gland as in previous interference between fluid collections at different time points.
reports23) is well within the demonstrated operation rage. This feature also allowed accurate assessment of multiple
Because the top wall in the HV is hydrophilic in the current biomarkers in different chambers with a single device.
demonstration, a small amount of the black-dyed water The novel one-opening microfluidic chambers also
wicked into the main microfluidic channel before the reduced the evaporation rate of the fluid inside the chamber.
chamber was completely filled. However, this has a negligibly The weight loss of the liquid was determined by the
small effect on the sweat collection. Endowing the top wall of difference between the remaining weight of liquid V and the
the HV with hydrophobicity could also eliminate or minimize initial weight of liquid V0; both were measured by a high-
this effect. Minimal interference with the initially collected precision electronic balance. Although weight loss was
sweat sample from the later collection was also demonstrated observed in both chambers when they were exposed to air at
using water dyed with two colors (Fig. 4b). In this 26 °C, the one-opening chamber (i.e., without an outlet)
demonstration, the 1st chamber was filled with red-dyed exhibited remarkably smaller loss than that with an outlet
water, whereas the 2nd and 3rd chambers were subsequently (0.45 mm) over a period of 10 h (Fig. 4c). Therefore, the
filled with blue-dyed water. The diffusion between the two outflow evaporation rate was much smaller than the inflow
different fluids occurred in two directions: 1) from the 2nd sweat rate, allowing the sweat to fill the chamber during the
fluid in the main microfluidic channel to the 1st fluid in the low-intensity exercise. The storage time is already sufficient

Lab Chip This journal is © The Royal Society of Chemistry 2020


View Article Online

Lab on a Chip Paper

for many applications (e.g., in situ colorimetric analysis or


short-term storage before ex situ analysis). Additionally, the
storage time can be significantly increased by reducing the
storage temperature or covering the outlet. It should be noted
that the sweat collection would be challenging if the sweat rate
were significantly reduced. Although our new design of a one-
opening chamber could help reduce the evaporation rate,
efforts are still needed to address the challenges in this new
application. Because sweat evaporation inevitably increased the
measured concentrations of the biomarkers or electrolytes (10–
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

100 mM for sodium in sweat1), the one-opening chamber led


to a concentration measurement ca. 1.6 times more accurate
than those with an outlet after 3 h. To demonstrate the
increased sensing accuracy from the reduced evaporation rate,
a pH assay (902 07, Macherey-Nagel, Germany) was used as
evaporation from the chambers occurred at 40 °C (Fig. S11†). A
large deviation in the R index of ∼10 was first observed
between the two designs (i.e., with and without the 0.45 mm
outlet). Because the pH value was calculated from the RGB
index, the pH value at the measured time point would be
different from the time of collection due to evaporation. Fig. 5 Fluid dynamic modeling of the fluid collection process.
Therefore, a reduced evaporation rate would result in a better Schematics of the advancing front of the air–fluid interface into the
chamber (a) with and (c) without the hydrophobic valve. (b) Calculated
representation of the pH value at the time of collection.
map of the normalized capillary driving pressure versus the normalized
location variables of the advancing interface (l′ and x′). The yellow line
2.4. Fluid dynamic modeling of one-opening chambers with represents the calculated motion trajectory of the advancing front of
the HV the air–fluid interface based on the fluid dynamic model. The points
mark the experimental and simulated results, with the circle or cross
The pressure across the air–liquid interface in the channel corresponding to the success or non-success filling of the chamber.
(Pi) is obtained as (see Methods in the ESI†):
 
sinðα − θA Þ cos θA
P i ¼ 2σ þ (2)
b* h Pb ðL′ þ l′ − 0:5πÞ cos θA þ ð1 − x′Þ sin θA
r¼ 2  ; (3b)
2σ L′ þ l′2 þ 0:25π2 − L′ðπ − 2l′Þ − πl′ þ ð1 − x′Þ2
where σ is the surface tension of the liquid, θA is the constant π π 1
≤ l′ ≤ þ ;
contact angle at the channel wall, α is the apparent angle 2 2 4
between the tangent to the channel wall and the straight line
of the air–liquid interface, and b* and h are the effective where L′ is defined as L′ = L/r and x′ is related to l′ through eqn
width and height of the channel, respectively. The capillary (S5) in the ESI† (Fig. S12†). The normalized driving pressure Pb
pressure difference (ΔP) that drives the liquid advancing front r/2σ is positive for the APT-treated and PVP-modified

Pb ð4 þ L′Þ cos θA − cosð0:25 − l′ − θA Þ − x′ sin θA π 1 1


r¼ 2 2
; þ ≤ l′ ≤ π þ : (3c)
2σ ð4 þ L′Þ þ ð1 þ x′ Þ − ð8 þ 2L′Þ cosð0:25 − l′Þ þ 2x′ sinð0:25 − l′Þ 2 4 4

into the chamber is mainly contributed by Pi in eqn (2). As the hydrophilic microfluidic channel (θA = 42° ± 2° in Fig. 5b)
air–liquid interface advances in the bridging channel, the during the process of the liquid flow into the chamber.
effective channel width increases and the apparent contact Considering the additional constant driving pressure, the
angle α decreases in eqn (2). As a result, the first term in eqn resulting driving pressure spontaneously wicks the fluid into
(2) changes as the air–liquid interface advances in the bridging the one-opening chamber with the HV.
channel, whereas the second term is independent of the In the case of the one-opening chamber without the HV,
position. Therefore, the first term of the driving pressure (i.e., the resistance of compressed gas is estimated to be ca. 6 kPa
Pb = 2σ sin(α − θA)/b*) has then been systematically investigated. upon filling of the bridging channel (Fig. 5c). Because the
Denoting the advancing front of the air–liquid interface at l′ = l/ resistance is one order of magnitude higher than the
r and x′ = x/r with r as the radius of transition channels capillary drive force, the liquid does not enter the chamber.
(Fig. 5a), the normalized driving pressure is obtained as In contrast, placing the HV near the chamber routes the fluid
into the chamber. Three experimental measurements and five
Pb L′ cos θA − cosðl′ þ θA Þ − x′ sin θA π
r¼ ; 0 ≤ l′ ≤ ; (3a) calculation cases (see details in Fig. 3 and S13†) marked in
2σ 1 þ L′2 þ x′2 − 2L′ cos l′ − 2x′ sin l′ 2
Fig. 5b indicate the critical position of the HV at x′c = 3.5 (i.e.,

This journal is © The Royal Society of Chemistry 2020 Lab Chip


View Article Online

Paper Lab on a Chip

the HV needs to be placed to the left of x′c ). The HV placed 2.5. Chrono-sampling and chemical analysis of the sweat
to the left of the critical position allows the sweat to flow
along on the chamber wall, which can effectively vent air Human testing involved the evaluation of the skin-interfaced
through the HV to avoid backpressure. When the HV is microfluidic device mounted on the forehead of a young
placed to the right of the critical position, the chamber is adult volunteer during running exercises (Fig. 6a). The sweat
not filled and the liquid simply flows through the main was driven into the microfluidic device by the positive
microfluidic channel. Once the chamber is filled with the pressure from sweat excretion, assisted by the capillary action
liquid, the flow automatically bursts open the HV. During in the hydrophilic channels. The flow rate of the sweat in the
the liquid collection at a later stage, the liquid bypasses the microfluidic channel was determined by the dimension of
previous chamber without mixing with the liquid samples the microchannel, surface tension, and the positive pressure
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

collected in the previous chambers. Due to its large from sweat excretion. The sweat flow rate from the young
curvature, the curved main microfluidic channel is adult volunteer may not be constant; however, a relatively
approximated to a straight channel for easy calculation. The stable flow advancing front during the low-intensity running
simulated results further indicate no significant differences exercise was observed, indicating a relatively small effect
for sweat collection between the circular and straight main from the varying sweat flow rate. It was also noted that the
microfluidic channel (Fig. S13†). Because of the curved largely varied flow rate may lead to different timeframes to
main microfluidic channel, the effective width experienced completely fill each chamber; however, the time point for
by the liquid meniscus could be slightly different from the analysis could be adjusted accordingly. Although running
theoretical model. According to a previous literature could cause vibration, the relatively strong adhesion strength
report,34 the calculated bursting pressure would differ by between the device and skin confined the sweat within the
<5%. Because the calculated bursting pressure is much microfluidic device. Placing a red dye of water-soluble
smaller than the pressure generated by the eccrine sweat particles at the inlet of the microfluidic device allowed for
gland (i.e., 2.4–2.9 kPa), the effect from this small variation visual assessment of the cumulative sweat loss during the
is minimal. sweat collection process. The red-dyed sweat was observed to

Fig. 6 In situ chrono-sampling from various body positions during running exercise. (a) Images of microfluidic devices applied to the forehead for
sweat collection and (b) the corresponding volume measurements through a syringe pump study. Note: the syringe pump study does not include
the volume of ∼6 μL in the inlet; therefore, it must be added before filling of the 1st chamber in the human testing. (c) Changes in the RGB values
as a function of pH value. (d) Images of microfluidic devices deployed to the lower back over half an hour. (e) The sequentially measured sweat pH
values from various body positions (i.e., lower back, neck, and forehead).

Lab Chip This journal is © The Royal Society of Chemistry 2020


View Article Online

Lab on a Chip Paper

be confined within the microfluidic channel and chambers analysis. By reducing sweat evaporation, the new chamber
even during low-intensity physical exercise for ca. one hour, design with only one opening provides a more accurate
consistent with literature reports.2,9,23 The robust adhesion assessment of the biomarker or electrolyte in the sweat. Fluid
strengthened from the chosen adhesive layer avoided the dynamic modeling and numerical simulations have been
detachment of the device from the skin. However, long-term explored to validate the effectiveness of the design. Human
use of the device for sweat monitoring may result in the testing on young healthy volunteers demonstrates the
detachment, which will be systematically studied in the chrono-sampling capabilities of sweat collection, storage, and
future. With a volume capacity of 2.12 μL in each chamber analysis. Taken together, the microfluidic device with one-
and a capacity of 3.64 μL in the microfluidic channel, the opening chambers and hydrophobic valves with significantly
total volume of ∼10 μL (verified by a syringe pump study in reduced fabrication complexity provides a simple alternative
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

Fig. 6b) in the microfluidic device corresponded to the route to study sweat physiology for healthcare.
quantitative operational requirements. By taking the ratio of
the volumes (Fig. 6b) to times (Fig. 6a) after the sweat filled 4. Materials and methods
each chamber, the average sweat rate from each sweat gland 4.1. Materials
(9 in total) was calculated to be 0.065, 0.04, and 0.018 μL
min−1 from the three chambers during the chrono-sampling Materials used in this study include polymethylmethacrylate
over the running exercise. Notably, the peak sweat rate from (PMMA; Microchem, MA, China), polydimethylsiloxane
the measurement of the 1st chamber corresponded to the (PDMS; Sylgard 184, Dow Corning, US), surface treating agent
initial stage of the exercise with the largest sweat rate. The (JD-206, Dongguan Jiudian Co. Ltd, China), adhesive film
average sweat rate per sweat gland during the entire period (IV300, Smith & Nephew Medical Ltd, UK), polyvinyl
of the exercise was estimated to be 0.04 μL min−1. The pyrrolidone (PVP, Sinopharm Chemical Reagent Co., Ltd,
measured effective sweat rate is larger than the value China), methyl red (Tianjin Guangfu Fine Chemical Research
measured by conventional absorbent pads (0.006 μL min−1),35 Institute, China), hydrogen phosphate (Sinopharm Chemical
mainly due to compensatory effects (i.e., higher sweat rates Reagent Co., Ltd, China), and acid monohydrate (Aladdin
in the open sweat glands because the adjacent regions are Industrial Corporation).
blocked by the adhesive).23 Nevertheless, the increased
effective sweat rate helps increase the speed of the sweat 4.2. Design and fabrication of the microfluidic device
collection in the microfluidic device. The microfluidic devices consisted of three layers: a PDMS
While the collected sweat samples in these chambers can cover, a microfluidic layer, and an ultrathin adhesive layer.
be retrieved and analyzed later when in situ analysis is The microfluidic microchannels and chambers on the surface
limited, instantaneous and quantitative chemical assessment of polymethylmethacrylate (PMMA; Microchem, MA, China)
of the sweat is also possible through colorimetric schemes. were created by micromachining (CNN, JK-DK40, Wenhao
As a simple yet representative demonstration, the pH of Co. Ltd, China). Polydimethylsiloxane (PDMS; Sylgard 184,
sweat in the different chambers was analyzed using methyl Dow Corning, US) was mixed at a 20 : 1 ratio by weight to
red as a pH indicator. Calibration of paper-based colorimetric prepare the PDMS cover and microfluidic layer. Pouring 0.5 g
pH assays with buffer solutions over a medically relevant of the mixture into the PMMA molds, followed by degassing
range (pH from 4.5 to 6.0) indicated a significant change in in a vacuum desiccator for removing bubbles and curing at
the G value among the three RGB values (Fig. 6c). Mounting 80 °C for 50 min, formed the microfluidic layer with a
the microfluidic devices on various body positions (i.e., lower thickness of ∼600 μm. The PDMS cover with a thickness of
back, neck, and forehead) quantified the sweat pH over half ∼300 μm was created by spinning (300 rpm for 30 s on a
an hour (lower-back, Fig. 6d). Using the calibration curve smooth PMMA, curing at 80 °C for 50 min). The pH assay
(Fig. 6c), the G value obtained by the software from the papers were loaded into the chambers. The PDMS cover and
digital color images was converted into concentrations of the microfluidic layer were bonded by exposure to air plasma
target analyte (e.g., pH). The sequentially measured sweat pH (Harrick Plasma Cleaner, PDC-002-HP, Harrick Plasma, US)
was also different among the lower back, neck, and forehead with ‘high’ power for 3 min. Next, the surface of the device
(Fig. 6e). While the sweat pH from the lower back was covered with a surface treating agent (JD-206, Dongguan
sequentially decreased, no substantial differences were Jiudian Co. Ltd, China). A double-sided adhesive was applied
observed in the sweat pH from the forehead and neck, to connect the adhesive layer to the device. A thin adhesive
providing critical insights into the underlying physiological film (IV300, Smith & Nephew Medical Ltd, UK) was placed at
variations. the bottom of the device to adhere to the skin. After the mask
for the HVs was designed by AutoCAD, a laser cutting system
3. Conclusions (NARVEL 6000) was used to cut the mask from stainless steel.
Next, a microscope on the high-precision three-axis micron
This paper presents the design and demonstration of a novel positioning stage was used to place the mask at the
skin-interfaced microfluidic device with one-opening prescribed location in the main microfluidic channel close to
chambers and hydrophobic valves for sweat collection and the chambers. Exposing the microfluidic device to air plasma

This journal is © The Royal Society of Chemistry 2020 Lab Chip


View Article Online

Paper Lab on a Chip

treatment endowed the PDMS with hydrophilicity in the of 35 °C. A volume of water was imbibed during the exercise,
regions not covered by the mask. The unexposed PDMS and the sweat collection continued until the microfluidic
region under the mask remained hydrophobic and formed device was completely filled with sweat.
the HV in the microfluidic channel.
Conflicts of interest
4.3. Surface modification
The authors declare no competing financial interest.
The PDMS layers were washed with ethanol and deionized (DI)
water, followed by drying with N2. The PDMS layers were Acknowledgements
treated with air plasma with ‘high’ power for 3 min to activate
the surface. Next, the PDMS layers were immersed in 10% (w/v) This research was supported by the National Natural Science
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

polyvinyl pyrrolidone (PVP, Sinopharm Chemical Reagent Co., Foundation of China (11872326 and 11702236), Natural Science
Ltd, China) solution for 3 min. After washing with DI water Foundation of Hunan Province (2018JJ2379 and 2018JJ2396),
and drying with N2, PDMS layers with long-term hydrophilic Scientific Research Fund (18B089) of Hunan Provincial
surfaces were obtained. Spin casting of the 20 : 1 PDMS on a Education Department, and Open Fund from Institute of
smooth PMMA layer at 3000 rpm provided an uncured PDMS Flexible Electronics Technology of THU (2019KF1102). H. C.
thin layer. Dipping the capping layer against this uncured also acknowledges the support from Pennsylvania State
PDMS thin layer generated a tacky surface on the capping layer University and American Chemical Society Petroleum Research
to facilitate its bonding with the microfluidic layer. Fund (59021-DNI7). The helpful discussion with W. Xia at Xi'an
Jiaotong University is also acknowledged.
4.4. Contact angle measurements
References
The contact angles of the samples were measured using a
Model 250 (p/n 250-F1) optical contact-angle goniometer 1 Z. Sonner, E. Wilder, J. Heikenfeld, G. Kasting, F. Beyette, D.
(Rame-hart, USA) at ambient temperature. After dropping 3 Swaile, F. Sherman, J. Joyce, J. Hagen, N. Kelley-Loughnane
μL of deionized water onto the sample surface with an and R. Naik, Biomicrofluidics, 2015, 9, 031301.
automatic dispense controller, the contact angles were 2 A. Koh, D. Kang, Y. Xue, S. Lee, R. M. Pielak, J. Kim, T.
determined automatically using the Laplace–Young fitting Hwang, S. Min, A. Banks, P. Bastien, M. C. Manco, L. Wang,
algorithm. The average contact angle was based on five K. R. Ammann, K.-I. Jang, P. Won, S. Han, R. Ghaffari, U.
independent contact angle measurements. Paik, M. J. Slepian, G. Balooch, Y. Huang and J. A. Rogers,
Sci. Transl. Med., 2016, 8, 366ra165.
4.5. Chemicals for the pH assay 3 S. B. Kim, K. Lee, M. S. Raj, B. Lee, J. T. Reeder, J. Koo, A.
Hourlier-Fargette, A. J. Bandodkar, S. M. Won, Y. Sekine, J.
Methyl red (Tianjin Guangfu Fine Chemical Research
Choi, Y. Zhang, J. Yoon, B. H. Kim, Y. Yun, S. Lee, J. Shin, J.
Institute, China) served as a chromogen for pH detection. A
Kim, R. Ghaffari and J. A. Rogers, Small, 2018, 14, e1802876.
homogenous 0.5% (w/v) solution of methyl red was prepared
4 V. F. Curto, S. Coyle, R. Byrne, N. Angelov, D. Diamond and
in ethanol. A filter paper served as the assay carrier of the
F. Benito-Lopez, Sens. Actuators, B, 2012, 175, 263–270.
reaction and was hole-punched into circular shapes with
5 Z. Wang, T. Liu, L. Jiang, M. Asif, X. Qiu, Y. Yu, F. Xiao and
diameters of 2.5 mm. Next, each piece of filter paper was
H. Liu, ACS Appl. Mater. Interfaces, 2019, 11, 32310–32319.
dipped in the solution for 30 s and dried at room
6 L. B. Baker, C. T. Ungaro, B. C. Sopeña, R. P. Nuccio, A. J.
temperature for 30 min to form a stable pH assay. Mixing 0.2
Reimel, J. M. Carter, J. R. Stofan and K. A. Barnes, J. Appl.
M disodium hydrogen phosphate (Sinopharm Chemical
Physiol., 2018, 124, 1304–1318.
Reagent Co. Ltd, China) and 0.1 M citric acid monohydrate
7 L. B. Baker, P. J. D. De Chavez, C. T. Ungaro, B. C. Sopeña,
(Aladdin Industrial Co.) yielded pH buffer solutions with
R. P. Nuccio, A. J. Reimel and K. A. Barnes, Eur. J. Appl.
various pH values (i.e., 4.5, 5.0, 5.5, and 6.0) as validated by a
Physiol., 2019, 119, 361–375.
pH meter (PHS-3B, Shanghai Shiheng Co. Ltd, China). After
8 Y. Epstein and W. O. Roberts, Scand. J. Med. Sci. Spor.,
dropping 1 μL of buffer solution on the pH essay papers,
2011, 21, 742–748.
Photoshop was used to determine the RGB values of each
9 Y. Zhang, H. Guo, S. B. Kim, Y. Wu, D. Ostojich, S. H. Park, X. Wang,
chromogenic paper, which corresponded to the
Z. Weng, R. Li, A. J. Bandodkar, Y. Sekine, J. Choi, S. Xu, S. Quaggin,
concentrations of the selected biomarkers.
R. Ghaffari and J. A. Rogers, Lab Chip, 2019, 19, 1545–1555.
10 C. Zhu, A. Chortos, Y. Wang, R. Pfattner, T. Lei, A. C. Hinckley,
4.6. Chrono-sampling of sweat in microfluidic devices I. Pochorovski, X. Yan, J. W. F. To, J. Y. Oh, J. B. H. Tok, Z. Bao
Testing involved healthy young adults as volunteers during and B. Murmann, Nat. Electron., 2018, 1, 183–190.
normal physical activities, and the volunteer subjects gave 11 A. Mena-Bravo and M. D. Luque de Castro, J. Pharm. Biomed.
informed consent. The skin was wiped with alcohol to clean Anal., 2014, 90, 139–147.
grease and ash to form strong adhesion to the device. Low- 12 J. T. Reeder, J. Choi, Y. Xue, P. Gutruf, J. Hanson, M. Liu, T.
intensity running was carried out at an ambient temperature Ray, A. J. Bandodkar, R. Avila, W. Xia, S. Krishnan, S. Xu, K.

Lab Chip This journal is © The Royal Society of Chemistry 2020


View Article Online

Lab on a Chip Paper

Barnes, M. Pahnke, R. Ghaffari, Y. Huang and J. A. Rogers, 23 J. Choi, D. Kang, S. Han, S. B. Kim and J. A. Rogers, Adv.
Sci. Adv., 2019, 5, eaau6356. Healthcare Mater., 2017, 6, 1601355.
13 S. Coyle, K. Lau, N. Moyna, D. O. Gorman, D. Diamond, F. D. 24 Y. Liu, M. Pharr and G. A. Salvatore, ACS Nano, 2017, 11,
Francesco, D. Costanzo, P. Salvo, M. G. Trivella, D. E. D. Rossi, 9614–9635.
N. Taccini, R. Paradiso, J. Porchet, A. Ridolfi, J. Luprano, C. 25 J. T. Reeder, Y. Xue, D. Franklin, Y. Deng, J. Choi, O. Prado,
Chuzel, T. Lanier, F. Revol-Cavalier, S. Schoumacker, V. R. Kim, C. Liu, J. Hanson, J. Ciraldo, A. J. Bandodkar, S.
Mourier, I. Chartier, R. Convert, H. De-Moncuit and C. Bini, Krishnan, A. Johnson, E. Patnaude, R. Avila, Y. Huang and
IEEE T. Inf. Technol. B., 2010, 14, 364–370. J. A. Rogers, Nat. Commun., 2019, 10, 5513.
14 Y. Sekine, S. B. Kim, Y. Zhang, A. J. Bandodkar, S. Xu, J. 26 C. Jungil, X. Yeguang, X. Wei, R. R. Tyler, T. R. Jonathan,
Choi, M. Irie, T. R. Ray, P. Kohli, N. Kozai, T. Sugita, Y. Wu, J. B. Amay, K. Daeshik, X. Shuai, H. Yonggang and A. R.
Published on 18 June 2020. Downloaded by Uppsala University on 6/18/2020 11:06:20 AM.

K. Lee, K. T. Lee, R. Ghaffari and J. A. Rogers, Lab Chip, John, Lab Chip, 2017, 17, 2572–2580.
2018, 18, 2178–2186. 27 S. J. Keating, M. I. Gariboldi, W. G. Patrick, S. Sharma, D. S.
15 M. Barfield, N. Spooner, R. Lad, S. Parry and S. Fowles, Kong and N. Oxman, PLoS One, 2016, 11, e0160624.
J. Chromatogr., B, 2008, 870, 32–37. 28 K. Handique, D. T. Burke, C. H. Mastrangelo and M. A.
16 A. J. Bandodkar and J. Wang, Trends Biotechnol., 2014, 32, Burns, Anal. Chem., 2000, 72, 4100–4109.
363–371. 29 L. Riegger, M. M. Mielnik, D. Mark, W. Streule, M. Clad, R.
17 S. Choi, H. Lee, R. Ghaffari, T. Hyeon and D. H. Kim, Adv. Zengerle and P. Koltay, Teflon-carbon black as new material
Mater., 2016, 28, 4203–4218. for the hydrophobic patterning of polymer labs-on-a-chip, in
18 T. Someya, Z. Bao and G. G. Malliaras, Nature, 2016, 540, Proceedings of the Solid-state Sensors, Actuators & Microsystems
379–385. Conference, Transducers International, IEEE Transducers,
19 X. He, T. Xu, Z. Gu, W. Gao, L. P. Xu, T. Pan and X. Zhang, 2009.
Anal. Chem., 2019, 91, 4296–4300. 30 M. Villegas, Z. Cetinic, A. Shakeri and T. F. Didar, Anal.
20 A. J. Bandodkar, J. Choi, S. P. Lee, W. J. Jeang, P. Agyare, P. Chim. Acta, 2018, 1000, 248–255.
Gutruf, S. Wang, R. A. Sponenburg, J. T. Reeder, S. Schon, 31 S. Hemmilä, J. V. Cauich-Rodríguez, J. Kreutzer and P. Kallio,
T. R. Ray, S. Chen, S. Mehta, S. Ruiz and J. A. Rogers, Adv. Appl. Surf. Sci., 2012, 258, 9864–9875.
Mater., 2019, 31, 1902109. 32 Y. Feng, Z. Zhou, X. Ye and J. Xiong, Sens. Actuators, A,
21 L.-C. Tai, T. S. Liaw, Y. Lin, H. Y. Y. Nyein, M. Bariya, W. Ji, 2003, 108, 138–143.
M. Hettick, C. Zhao, J. Zhao, L. Hou, Z. Yuan, Z. Fan and A. 33 T. Ishida, D. McLaughlin, Y. Tanaka and T. Omata, Sens.
Javey, Nano Lett., 2019, 19, 6346–6351. Actuators, B, 2018, 254, 1005–1010.
22 S. B. Kim, Y. Zhang, S. M. Won, A. J. Bandodkar, Y. Sekine, Y. 34 J. Choi, Y. Xue, W. Xia, T. R. Ray, J. T. Reeder, A. J.
Xue, J. Koo, S. W. Harshman, J. A. Martin, J. M. Park, T. R. Ray, Bandodkar, D. Kang, S. Xu, Y. Huang and J. A. Rogers, Lab
K. E. Crawford, K. T. Lee, J. Choi, R. L. Pitsch, C. C. Grigsby, Chip, 2017, 17, 2572–2580.
A. J. Strang, Y. Y. Chen, S. Xu, J. Kim, A. Koh, J. S. Ha, Y. 35 N. A. S. Taylor and C. A. Machado-Moreira, Extreme Physiol.
Huang, S. W. Kim and J. A. Rogers, Small, 2018, 14, e1703334. Med., 2013, 2, 4.

This journal is © The Royal Society of Chemistry 2020 Lab Chip

You might also like