To cite this version : Laffont, Laure and Hezard, Teddy and Gros,
Pierre and Heimbürger, Lars-Eric and Sonke, Jeroen E. and Behra,
Philippe and Evrard, David Mercury(II) trace detection by a gold
nanoparticle-modified glassy carbon electrode using square-wave
anodic stripping voltammetry including a chloride desorption step.
(2015) Talanta, vol. 141. pp. 26-32. ISSN 0039-9140
a b s t r a c t
Gold nanoparticles (AuNPs) were deposited on a glassy carbon (GC) substrate by constant potential
electrolysis and characterized by cyclic voltammetry in H2SO4 and field emission gun scanning electron
microscopy (FEG-SEM). The modified AuNPs–GC electrode was used for low Hg(II) concentration
detection using a Square Wave Anodic Stripping Voltammetry (SWASV) procedure which included a
chloride desorption step. The comparison of the obtained results with our previous work in which no
desorption step was used showed that this latter step significantly improved the analytical performances,
Keywords: providing a three time higher sensitivity and a limit of detection of 80 pM for 300 s preconcentration, as
Hg(II) sensor well as a lower average standard deviation. The influence of chloride concentration on the AuNPs–GC
Gold nanoparticles
electrode response to Hg(II) trace amounts was also studied and its optimal value confirmed to be in the
Stripping voltammetry
10 ! 2 M range. Finally, the AuNPs–GC electrode was used for the determination of Hg(II) in a natural
Chloride desorption step
Natural water groundwater sample from south of France. By using a preconcentration time of 3000 s, a Hg(II) con-
centration of 1973 pM was found, which compared well with the result obtained by cold vapor atomic
fluorescence spectroscopy (2272 pM).
Cl ! desorption step Electrode modification procedure Linear range (nM) Normalized sensitivitya (mA nM ! 1 min ! 1) LOD (pM) Average RSD (%) Réf.
a
The value of the normalized sensitivity was calculated by dividing the slope of the linear response by the electrochemical or chemical preconcentration time. All
electrochemical measurements were performed by SWASV excepted Ref. [45] which used DPASV.
b
2-Mercaptobenzothiazole adsorbed on gold micro-/nanopore arrays.
c
Hg(II) imprinting microporous poly(2-mercaptobenzothiazole) film electropolymerized on GC electrode.
d
Hg(II) imprinting poly(2-mercaptobenzothiazole) film electropolymerized on the surface of gold nanoparticles/single-walled carbon nanotube nanohybrids modified
glassy carbon electrode.
e
Mercaptoethanesulfonate monolayer-modified AuNPs.
f
Bimetallic Au–Pt nanoparticles/organic nanofibers.
Acknowledgment
References
Fig. 4. Calibration curve obtained by SWASV on a AuNPs–GC electrode in an [1] J. Musarrat, A. Zaidi, M.S. Khan, M.A. Siddiqui, A.A. Al-Khedhairy, Environ.
acidified (0.01 M HNO3) sample of Font Estramar sample. The measurements were Pollut. 20 (2011) 323–342.
performed using 3000 s preconcentration at ! 0.4 V. White circles corresponds to [2] S. Prabhakar, A.K. Singh, D.S. Pooni, Indian J. Anim. Sci. 82 (2012) 244–255.
data which were not included in the regression. [3] E. Callender, Treatise Geochem. 9 (2004) 67–105.
[4] S.P. Roy, Ecoscan 4 (2010) 235–240.
[5] G.N. Mhatre, Encycl. Environ. Control Technol. 8 (1995) 427–454.
amounts of Hg(II) were spiked into the Font Estramar sample. The [6] W.F. Fitzgerald, C.H. Lamborg, Geochemistry of Mercury in the Environment,
Oxford, Pergamon, 2007.
corresponding analytical responses are depicted in Fig. 4. Only the
[7] F.M.M. Morel, A.M.L. Kraepiel, M. Amyot, Annu. Rev. Ecol. Syst. 29 (1998)
first two spiked concentrations allowed a linear regression curve to 543–566.
be calculated, the two higher ones inducing a saturation phenom- [8] D. Mergler, H.A. Anderson, L.H.M. Chan, K.R. Mahaffey, M. Murray,
enon in a similar trend to that reported above, mainly because of the M. Sakamoto, A.H. Stern, Ambio 36 (2007) 3–11.
[9] M. Sakamoto, K. Murata, A. Kakita, M. Sasaki, in: Y. Cai, G. Liu, N. O'Driscoll
significantly longer preconcentration time. On the basis of the (Eds.), Advances in Environmental Chemistry and Toxicology of Mercury,
regression equation (Δip ¼0.688172.9963 [Hg(II)]; r2 ¼0.9940), the Wiley-VCH, Weinheim, 2012, pp. 501–516, Chapter 15.
Hg(II) concentration in the Font Estramar sample was estimated to [10] D. Cossa, C. Garnier, R. Buscail, F. Elbaz-Poulichet, N. Mikac, N. Patel-Sorrentino,
E. Tessier, S. Rigaud, V. Lenoble, C. Gobeil, Biogeochemistry 119 (2014) 35–43.
be 1973 pM. This value was compared to reference technique [11] R. Eisler, Environ. Geochem. Health 25 (2003) 325–345.
analysis, namely CV-AFS, which yielded [Hg(II)]¼ 2272 pM. Thus, [12] Q. Wang, D. Kim, D.D. Dionysiou, G.A. Sorial, D. Timberlake, Environ. Pollut.
both the reference method and the AuNPs–GC electrode provided 131 (2004) 323–336.
[13] Guidelines for Drinking-water Quality, Vol. 1 third ed., World Health Organi-
comparable values for the Hg(II) content at the 95% confidence level.
zation, Geneva, 2008. 〈http://www.who.int/water_sanitation_health/dwq/
This last result makes us confident concerning the selectivity of the gdwq3rev/en/〉 (accessed 16.06.11).
sensor and rules out major interferences and especially Cu(II). The [14] S.L. Suib, Front. Chem. 1 (2013) 1–2.
[15] K. Leopold, M. Foulkes, P. Worsfold, Anal. Chim. Acta 663 (2010) 127–138.
concentration of this latter metal cation in submarine groundwater
[16] N. Bloom, Can. J. Fish. Aquat. Sci. 46 (1989) 1131–1140.
discharge such as Font Estramar has been reported to be around [17] N.S. Bloom, M. Horvat, C.J. Watras, Water Air Soil Pollut. 80 (1995) 1257–1268.
20 nM [63]. Also, Cu(II) average concentration in the Rhône River [18] D. Cossa, L.-E. Heimbürger, D. Lannuzel, S.R. Rintoul, E.C.V. Butler, A.R. Bowie,
(closed to Font Estramar) was 32.7712.6 nM [64], which confirmed B. Averty, R.J. Watson, T. Remenyi, Geochim. Cosmochim. Acta 75 (2011)
4037–4052.
the value reported by Beck et al. Thus, the favorable comparison of [19] K.L. Bowman, C.R. Hammerschmidt, Limnol. Oceanogr.—Methods (2011)
the results obtained by AFS and the AuNPs–GC sensor suggests that 121–128.
Hg(II) detection using this latter is not affected by Cu(II) con- [20] C.-M. Tseng, C.R. Hammerschmidt, W.F. Fitzgerald, Anal. Chem. 76 (2004)
7131–7136.
centration in a 1/1000 Hg to Cu ratio. In a similar way, it may be [21] M. Monperrus, E. Tessier, S. Veschambre, D. Amouroux, O. Donard, Anal.
assumed that organic matter did not undergo interferences when Bioanal. Chem. 381 (2005) 854–862.
present at concentrations up to 1.2 mg L ! 1, which corresponded to [22] D. Point, W.C. Davis, J.I. Garcia Alonso, M. Monperrus, S.J. Christopher,
O.F.X. Donard, P.R. Becker, S.A. Wise, Anal. Bioanal. Chem. 389 (2007)
the dissolved organic carbon concentration found in Font Estramar
787–798.
water. These results proved that our system was reliable and [23] S.K. Pandey, K.-H. Kim, R.J.C. Brown, Trends Anal. Chem. 30 (2011) 899–917.
potentially suitable for in situ analysis of natural water samples. [24] C.H. Lamborg, C.R. Hammerschmidt, G.A. Gill, R.P. Mason, S. Gichuki, Limnol.
Work is currently in progress in our laboratory in this sense. Oceanogr.—Methods 10 (2012) 90–100.
[25] C.R. Hammerschmidt, K.L. Bowman, M.D. Tabatchnick, C.H. Lamborg, Limnol.
Oceanogr.—Methods 9 (2011) 426–431.
[26] J. Buffle, M.-L. Tercier-Waeber, Trends Anal. Chem. 24 (2005) 172–191.
4. Conclusions [27] C. Gao, X.-J. Huang, Trends Anal. Chem. 51 (2013) 1–12.
[28] J.M. Pinilla, L. Hernandez, A.J. Conesa, Anal. Chim. Acta 319 (1996) 25–30.
[29] S. Meyer, F. Scholz, R. Trittler, Fresenius' J. Anal. Chem. 356 (1996) 247–252.
In this work, we showed that the addition of a chloride deso- [30] G.G. Muntyanu, J.Anal. Chem. 56 (2001) 546–551.
rption step in the analytical procedure aiming at Hg(II) trace [31] H. Yi, Anal. Bioanal. Chem. 377 (2003) 770–774.
[32] M. Shamsipur, J. Tashkhourian, B. Hemmateenejad, H. Sharghi, Talanta 64
determination allows to significantly improve the analytical per- (2004) 590–596.
formances of a AuNPs–GC electrode. In particular, the sensitivity [33] J. Svarc-Gajic, Z. Stojanovic, Z. Suturovic, N. Marjanovic, S. Kravic, Desalination
was three times better and the LOD was as low as 80 pM for 300 s 249 (2009) 253–259.
[34] Y. Bonfil, M. Brand, E. Kirowa-Eisner, Anal. Chim. Acta 424 (2000) 65–76.
preconcentration. The influence of chloride amount on the ana-
[35] E.A. Viltchinskaia, L.L. Zeigman, D.M. Garcia, P.F. Santos, Electroanalysis 9
lytical performances was also studied and it was proved that (1997) 633–640.
AuNPs behave in a similar trend to a Au microwire from this point [36] L.A. Khustenko, L.N. Larina, B.F. Nazarov, J. Anal. Chem. 58 (2003) 262–267.
of view. Finally, the AuNPs–GC electrode succeeded in the deter- [37] S. Daniele, C. Bragato, M.A. Baldo, J. Wang, J. Lu, Analyst 125 (2000) 731–735.
[38] P. Salaün, C.M.G. van den Berg, Anal. Chem. 78 (2006) 5052–5060.
mination of Hg(II) trace in a natural groundwater sample, allowing [39] C. Garnier, L. Lesven, G. Billon, A. Magnier, O. Mikkelsen, I. Pizeta, Anal.
19 pM Hg(II) to be detected for 3000 s preconcentration. Further Bioanal. Chem. 386 (2006) 313–323.
[40] O. Ordeig, C.E. Banks, Jd Campo, F.X. Munoz, R.G. Compton, Electroanalysis 18 [54] USEPA, Method 1631, Revision E: Mercury in Water by Oxidation, Purge and
(2006) 573–578. Trap, and Cold Vapor Atomic Fluorescence Spectrometry, 2002.
[41] L. Pujol, D. Evrard, K. Groenen-Serrano, M. Freyssinier, A. Ruffien-Cizsak, [55] D. Cossa, B. Averty, J. Bretaudeau, A.S. Senard, Speciation du Mercure Dissous
P. Gros, Front. Chem. 2 (2014) 19. Dans les Eaux Marines: Analytical Methods for the Marine Environments,
[42] X.H. Gao, W.Z. Wei, L. Yang, T.J. Yin, Y. Wang, Anal. Lett. 38 (2005) 2327–2343. Ifremer and French Ministry for Ecology and Sustainable Development Pub-
[43] O. Abollino, A. Giacomino, M. Malandrino, G. Piscionieri, E. Mentasti, Elec- lication MA0303, Paris, France, 2003.
troanalysis 20 (2008) 75–83. [56] L. Sipos, H.W. Nürnberg, P. Valenta, M. Branica, Anal. Chim. Acta 115 (1980)
[44] A. Safavi, E. Farjami, Anal. Chim. Acta 688 (2011) 43–48. 25–42.
[45] X.-C. Fu, J. Wu, L. Nie, C.-G. Xie, J.-H. Liu, X.-J. Huang, Anal. Chim. Acta 720 [57] Analytical Methods Committee, Analyst 112 (1987) 199–204.
(2012) 29–37. [58] J. Gong, T. Zhou, D. Song, L. Zhang, X. Hu, Anal. Chem. 82 (2010) 567–573.
[46] Y. Liu, G. Su, B. Zhang, G. Jiang, B. Yan, Analyst 136 (2011) 872–877. [59] X.-C. Fu, X. Chen, Z. Guo, C.-G. Xie, L.-T. Kong, J.-H. Liu, X.-J. Huang, Anal. Chim.
[47] C.M. Welch, R.G. Compton, Anal. Bioanal. Chem. 384 (2006) 601–619. Acta 685 (2011) 21–28.
[48] T. Hezard, K. Fajerwerg, D. Evrard, V. Colliere, P. Behra, P. Gros, J. Electroanal. [60] C. Faller, N.Y. Stojko, G. Henze, K.Z. Brainina, Anal. Chim. Acta 396 (1999)
Chem. 664 (2012) 46–52. 195–202.
[49] T. Hezard, K. Fajerwerg, D. Evrard, V. Colliere, P. Behra, P. Gros, Electrochim. [61] G. Sandmann, H. Dietz, W. Plieth, J. Electroanal. Chem. 491 (2000) 78–86.
Acta 73 (2012) 15–22. [62] M. Ueda, H. Dietz, A. Anders, H. Kneppe, A. Meixner, W. Plieth, Electrochim.
[50] T. Hezard, L. Laffont, P. Gros, P. Behra, D. Evrard, J. Electroanal. Chem. 697 Acta 48 (2002) 377–386.
(2013) 28–31. [63] A.J. Beck, Y. Tsukamoto, A. Tovar-Sanchez, M. Huerta-Diaz, H.J. Bokuniewicz,
[51] A. Giacomino, O. Abollino, M. Malandrino, E. Mentasti, Talanta 75 (2008) S.A. Sañudo-Wilhelmy, Appl. Geochem. 22 (2007) 477–490.
266–273. [64] P. Ollivier, O. Radakovitch, B. Hamelin, Chem. Geol. 285 (2011) 15–31.
[52] E. Pinilla Gil, P. Ostapczuk, Anal. Chim. Acta 293 (1994) 55–65. [65] X.-C. Fu, X. Chen, Z. Guo, L.-T. Kong, J. Wang, J.-H. Liu, X.-J. Huang, Electrochim.
[53] O. Abollino, A. Giacomino, M. Ginepro, M. Malandrino, I. Zelano, Electro- Acta 56 (2010) 463–469.
analysis 24 (2012) 727–734.