$:10Feot: potentiometer manually given voltage input +, Potentiometer manually given voltage input - Foot 11: Automatic Display Switch Input Foot 12:24, Foot 13: 24V GND 14.15 Foot: Empty calibration order, Empty calibration indication 16 foot: section selection 17-18 Foot: Full calibration order, Full calibration indication 20.21 Foot :4-20 mA Set level height signal 22.23 Foot :0-10 V Set level height signal Area 19. s: + 25 core standard double-row pins: 3.4,5 feet: Upper limit alarm :NO/ middle contact point / NC 7.8,9 feet: Lower limit alarm :NO/ middle contact point / NC 11.12,13 feet: MEU normal : NO/middle contact point /NC Area20. 220 V Power outlet Area 21, Instrument grounding or receiver); 1. Check MEU display information, data should be like follow: Ay The N value is big enough 8. VO value 4-8 Cy The V6 value is about v0.5-1 Note: Mean while, the sen: and radioactive source should be fixed tightly in the mold_and source contait should be_in_epen condition: Mould cooling water should be full of mould jacket, Not allow other metal stu inside of mould; 2. If the V6 value is not about 0.5-1 V and the N value is small, it is necessary to Sxecute the peak stabilization operation.:By Pressing "MOD" input "9999" and Press "1", the peak voltage can be adjusted automatically. (Prohibit in casting) ; 3. ifthe automatic V6 can not be adjusted, it i necessary to execute the manual peak stabilization operation,