181057202AMETER SETTING
3 SEPARATION SYSTEM PARAL
1 Parameter Setting
The parameters are set in the operator panel
for the different separator systems, and for
different conditions within the systems.
‘There are three parameter types: Installation,
Process, and Factory Set parameters.
The parameters are initially set at installation.
The process parameters can also be set as
required during operation,
Parameters not in use are not displayed.
‘The parameter list in this booklet is valid for
Separation Units (SU) and SA systems.
General principle
The ‘ENTER’ button is used to:
enter into @ parameter list
~ enter into a parameter
— accepiistore a new parameter value.
‘The‘¥’ or the ~’ buttons are used to change the
value flashing in the display window.
2 1810572-02To change parameters, proceed as follows:
4 Push the ‘ENTER’ button,
‘Time to discharge P1 60" is shown. 'P1 is
the parameter, and '60' is the sei value.
The ‘1’ of the ‘Pt’ is flashing.
2 Push the “+ bution until the number of the
parameter you wish to program appears in
the display window,
The chosen parameter number is now
flashing
3 Push the ‘ENTER’ button.
The chosen parameter number is now
shown on the let side in the display. The
parameter value is shown flashing on the
right side of the display.
‘Schematis
diagram with
postion LEDs
Display window -push> 4 push- Enter
bution button push.
Button
181057202
1 PARAMETE
Push the ‘+’ or ~’ button to change the
parameter value,
Push the ‘ENTER’ button. The new value is
stored,
The parameter number is now flashing
Push the ‘+’ or ~ button to go to the next
parameter to be changed.
To leave the list, push the ‘+’ and the *
buttons at the same time. The display shows
‘Standst.’ of, when in operation, the trigger
value to the left, and time remaining to
discharge on the right.
LeDterheater
(green)
7 Heater onfoff
{ push-button
LED for contro!
[program running
‘green)
Control program
onlotf push-outton
LED for stop
sequence running
(yellow)
Sludge
discharge push-
ution
Alarm reset LED for commen LED showing
‘main alarm (red) this pane!
active1 PARAMETER SETTING S-TYPE SEPARATION SYSTEM PARAMETER Li
The following lists are available for viewing
from a menu in the EPC-50 control unit:
* Install (parameters)
+ Factory (parameters)
+ Alarms
+ Test
For detailed information see each list.
To open the menu and select a list, proceed as
follows:
1 Push the ‘ENTER’ button. The process
parameter list is now open.
2 Push the +’ button until ‘End is displayed,
3 Push the ‘ENTER’ button and the ‘+’ bution at
the same time. ‘Install’ will show flashing on
the display.
4 Use the +’ button to select a list.
& Use the ‘Enter’ button to enter into the
selected list
6 Toleave the list, push the +’ and the
buttons at the same time,
7 Toleave the menu, press the ‘+! button
repeatedly until ‘Exit’ is displayed. Push the
“ENTER' button.
4 1810572-028
2 Installation Parameters
The installation parameters must be set before
the initial start-up and before process
parameters are set. These parameters rarely
need to be adjusted again, If a new parameter
value is to be stored, it is necessary to go through
the whole list, and confirm at the end
nore
Lue
‘Aa Laval caso be held response for injuries
So eappaecned oy mages peeaceer anes
‘Se ciel rade ar iee ear acer sot
by Alfa Laval.
Parameter | Factory set Range Description
it n ‘en = English Language, Allinformation displayed
Se ‘on the operators panel is shown in
the selected language,
t= falian
pt-= Portuguese
fi= Finnish
0 = Swedish
Ina Aleap ‘Aloap ‘Separator type
Ing 296 211,816 ‘Separator size
200, 820, 821, 825,826 Parameters depending on separator
1300, 830.831, 235,206 «S28 a18 Set automaticaly according
tothe value in Ind
409, 840, 841, 845, 848
500, 850, 887, 865, 856
1600, 860, 861, 865, 866
700, 870, 871, 875, 875
1800, 880, 881, 885, 886
Ins 1 1-9 ‘Separator identity
lf more than one EPC on more than one separator system is
‘same Sattbus connected to the internal Sattbus
communication, each system must
bbe given its own identity (1 or 2.0r 3
etc),
1810872-02 52 INSTALLATION PARAM SEPARATION SYSTEM PARAMETER LIST
“Parameter Factory set Plant set Range Description
:
:
.
en —s
%
‘Steam (steam heater)
into 6 7,8, 14,16, 22, 24, 96,40, El heater size
50, $6, 65, 72, 86,96, 98, activated when El. is selected in
100, 110, 112, 126,128, ing
130, 144 kw
n20 120 (0-999 seconds ‘Steam-valve run time
‘Activated when Steam is selected in
inte
6 1810572-02s PARATI 2 INSTALLATION PARAMETERS
Parameter | Factory set Plantset | Range Description
value value
ine 6 (= 80 seconds Alarm delay imme
in23 6 0-240 seconds CCharige-over valve V1 temp. alarm
delay
Prevents false alarm due to any
fluctuations in temperature caused
by valve actions.
nae No No (Change-over valve V1 action at extra
Aca alarm
| When the extra alarm inputs used,
action at alarm can be none (No) or
“The feed flow measurement, based
‘on inletpressure, can be displayed in
cone of these ways.
feed off (es)
| n25 Yes No Power fil alarm used
Yes I this alarm isnot required, set the
parameter to'No*
nas No. No “Type of computer communication
ExiPC = Fiokibus a
naz ° 0-256 Intemet IP adress, part 1
nae ° 0-256 Intemet IP adress, part 2
in29 ° 0-256 Intemet IP adress, part
n30 ° 0-288 Intemet IP address, part 4
Inst No No Feed pump controlled
Yes Feed pump has to be connected by
EPC when running in auto mode.
nse mh mth Flow rate in mh per hour or
sam US gallons per hour
1810572-02 73 Process Parameters
‘The process parameters can be adjusted easily
and as often as required, even during operation,
to meet changes in the operating conditions, e.g.
Lime between sludge discharges, oil temperature,
alarm points
(vee
Regularly check the condition of the bowl. If there is
no noticeable amount of hard sludge in the sludge
‘space, the discharge intorval may be increased by 30,
minutes at a time. The total discharge interval must
not exceed three times the factory set value in order
to avoid an uneven build-up of sludge
PARAMETER LIST
rameter | Factory set Range
value
Pri 120 71= 300 minutes: Time to discharge
Pr2 HF 360 MDO = Marine Diesel Ol ——Ciltype
GO =Gas oi)
Fax = Intermediate Fuel Oil (ox represents viscosity in Cst at
Hic = Heavy Fuel Oil 50°C)
LO TP = Lubricating Oil for
Trunk Piston Engines
LOCH = Lubricating Oil for
Gross Head Engines,
Pra OP loo: (OPloe = Local OP in control Control locaton
(OPrem = Remote OP (Pleo is standard. OPrem, ExtPC
Rifts pane ig ‘and FlamSW are shown only when
Sa installed,
ExtPC = Extern Computer
Board
FlemSw = Remots switches
Pra on On ‘Alcap On/Standoy
ste MT50 function disabled when in
standby (Sth). Discharge every 18
‘minutes (overrides Prt). No
leplacement water added
prs 00 0.0 =Notused ‘Countdown service timer
0.1 t0 10.0x1000 hours
‘Acton is to be taken when the timer
reaches 0.
1810572-021 SEPARATION
STEM PARAMETER UST
Parameter Factory set Plantset Range Description
value value
Pre ° O= service tine lapsed Type of eervce
1-92 Servce text 1..9 When service timer PrS reaches 0,
analarmie given and a meseage
shown
Pr? ° Information Keys to special functions used by
1 = timers shown cee
es When P17 is set to 1 during normal
operation, the actual timer is
splayed if the sutton i pressed
once.
Pr 10 3.0 0.0~80bar High cil pressure limit
Gos olpressure sensor (cleaned oll outa)
disabled
Prt 10 0.0-60bar Low oil pressure iit
Priz 55 0.0-60bar High water prossure lit
20 water (water outer)
disabled
Pris 02 0.0-60bar Low water pressure limit
Prag 30 0.0-60ber High feed pressure limit
0.0 = feed pressure sensor
disabled
Pets 020 000-600 bar Low feed pressure it
Low pressure indicates insufcient
flow. The low pressure mit must be
higher than any static pressure and
lower than normal pressure.
Pre 105°C 0-115 Cor HFOILO high temp. iit
0-255°F
0 = Temp sensor disabled
Pra as 0-118 Cor HFOILO low temp. lint
0-255°F
1810872-02PARAMETERS S-TYPE SEPARATION SYS!
Parameter Factory set Plant set Range Description
| jue value
Pris 95°C 0-110 °C or HFOLO setpoint”?
0-240°F Setpoint for feed temperature i valid
OaTemp sensorcisabled only whenatomperature controls
instalag inthe EPC.
Prt9 40°c 0-118 °Cor 0 bigh temp mit
0-255 °F
Pr20 20°¢ 0-118 *Cor 0 tow temp. init |
0-255 °F |
Prat 30°C 0-110 °C or DO setpoint”? |
0-240 Setpoint rte temperatures vad |
only when a temperature controler is
installed inthe EPC. |
Pra 40 10-50% P-eonstant?
| Prose value based on experience
| wi Alfa Laval Heatpac heaters
| Pr2a 12 01-100 minutes L-eonstant!?
| Pieeaiscioe feed ener
| wih Afa Laval Heatpac heaters,
| Pras ° o-9 Heater contel location
0 = daabled ‘To be set whan more than one EPC
| temperature contolarcan be used
| forone heater
Pras 2 0240 seconds Sludge tank alarm delay (uring
Ti 63)
Prag 60 0-100 discharges ‘Test of sludge in boat
Prat 18 0-100% Max. decrease of displacement
water
') Parameter appears only when heater board installed.
10 181057202RY SET
4 Factory Set Parameters
Fa21
Fa22
Fa23
meter Factory set
value
See tabla on page
02
See table on page
‘See table on page
300
‘See table on page
‘See table on page
os
3.0 if Pr2= fuel oll
06 Pr2= lube cil
100
100
1Oi#Pr2 = fuel of
eS#Pr2=TPLO
eit Pr2.=CHLO
‘Piant set
Range
value
0.1 98,99 lie
00-9bar
(0 14000 rpm
(0 sensor disabled
Is set depending on
‘separator size
0 ~ 14000 rpm
Is set depending on
‘separator size
| 02000 1pm
00-25
(0.0 = sensor disabled
00-25
|
00-25
| 001-1009
0- 250%
0- 250%
0- 999.9 pF
(= water transducer
disabied
1810872-02
Description
‘Conditioning water volume to be fed
after discharge
PTS pressure increase or decrease,
(Test during 7159, Ti 64, T170, Ti71
Ti 2) Signal thatthe bow! is filed
High speed
Low speed
‘Speed decrease during discharge
Shutdown limit for separator spindle
amplitude
Vibration alarm prewarning
‘Vibration sensor adjustment zero
point. Zero point should be within
152Fa18
Alcap trigger factor
This is the amplification factor for the
transducer signal co thatthe proper
{rigger range is achieved.
“Tiager limit HO
“Tigger limit LODO
‘Alarm limit for high water transducer
value.
"14 FACTORY SET PARAMETERS
PARATION SYSTEM PARAMETER LIST
Parameter Factoryset Plant set | Range Description
value value
[raze | 70 (0-999.9 pF ‘Alarm limit for low water transducer
value
Fa25 25 (0-250 seconds Pulse time for open drain valve
Fa26 5 0-30 seconds Drain pause time. With the valve
closed, the system wats for the
result of the draining,
Fa27 70 0 100% The limit below which the trigger
value must be to interrupt draining.
Fa26 6 0-20 Limit for number of HFO drainings. It
the trigger value has not fallen below
the value of Fa 27 after Fa 28, an
| slam is given
Fao 5 0-40 | Test of water transducer tigger
| o=elsabled | signal every x
T75Paussard
discharge
feedback
7 Discharge
L__J
30 1810572-02PATION SYSTEM PARAMETER LIST 5 SEQUENCE DIAGRAM
st
‘Start Without ae ee,
Calibration
UST Faed [iraFeecor
speed, ana tap |
chock
[raevirne]
¥ | sew Hesterot
TiSS Standby. Ti6s Yes L
Push Seperation intruptedby 1
toeomne Granny? a
fushing
Ti 60 Discharge | Vv
[T8 Sep. motor
TT DSEL ot Crack
water check cecresang tena]
Tr6t Pause |
! Ee Tiss Check
a2 Bow cosa Displacement eR
y Ti 73 Water 7
FresConcnioing
water cule sting. | Tae
Tet Feod en 7a tage Y
Tie Wing fr
7 O-speed
TEMES reise
signal test, discharge Y
teodback ss
(Tera) etd stat
y
Tey NTE he
y
Ties Time to
discharge
Y
Ties
elapsed?
Tie2
Yes
[Discharge q_¥
1810572-02 31