C Setting to work
ULS 0132 SM 01/03052 /C/1.1.1
1. Changing firmware
• Switch the system off.
• Check the software version: MIB V26 1.4, APB V57 2.7.
2. Setting configuration
• Switch in the user panel VH 103 S2:4 to closed (US enabled)
- Press (”Enter”) .
C-2
C Setting to work
ULS 0132 SM 01/03052 /C/1.1.0
4. Phantom calibration
• Insert the US-phantom adjustment tool into the
transceiver holder and plug the cable to the
connector VGX9 (red).
Fig.: Bild 3
Note
The green LED at the reference transmitter
Fig.: Bild 1
C-3
C Setting to work
ULS 0132 SM 01/03052 /C/1.1.1
If not o.k., change temperature up or down for fine adjustment so that the values are in range.
• Select menu ”F2 balance”.
The green LED and both yellow LED´s at the reference transmitter must be lit.
• if green LED is blinkimg connection between SD 360 and ULS 132 is not O.K.
• Press the switch a second time (the data will be stored and displayed in the screen ”F2 balance”).
Note
The phantom calibration must be finished in the menu "F2 balance" with the "ESC" pushbutton. The calibration
accepted if the values (µ) for X,Y,Z are o.k.
C-4
C Setting to work
ULS 0132 SM 01/03052 /C/1.1.0
Fig.: Bild 2
phantom
Fig.: US-phantom2
C-5
C Setting to work
ULS 0132 SM 01/03052 /C/1.1.1
controller.
• Fix the US–positioning transmitter at the
accessory rail by using the transmitter holder.
• Press the switch at the US–transceiver holder.
• Move the phantom with the use of the control X-ray image
panel in the F2 postion (X,Y,Z=0), as shown at
the control desk.
area of tolerance
• Check with x-ray in PA and CC position that the
top of the phantom is in the F2 position ultrasound phantom
(tolerance ± 3 mm)
Cx
Note
B
D
The rather large tolerance value is caused by
propagation time deviations, depending on the water
temperature and the coupling angle of the transducer
with respect to the phantom pot.
The failure of the location system itself is by far less
than ± 3 mm. The standart deviation will be checked B
for acceptance by the software during the phantom
calibration.
Fig.: X-ray image
Example:
The demanded tolerance is ±3mm (circle with 6mm diameter (D).
The real distance between the inner tips of the cross wires is 8mm (A).
The measured distance on the monitor (inner tips of the cross) is 18mm (B).
The scale (C) = B/A = 18mm/8mm = 2.25
The dimeter of the measured circle (toleance) on the monitor
= CxD = 2.25x6mm = 13.5mm
C-6
C Setting to work
ULS 0132 SM 01/03052 /C/1.1.0
C-7