Professional Documents
Culture Documents
CHROMSYMP. 2905
Review
ABSTRACT
The complexity of the mixtures of ecdysteroids in plants and the close similarities in their chemical structures have challenged
chemists to find suitable ways to separate and identify them. Great ingenuity has been applied to these problems and
consequently a wide range of separation and methods are available today. These methods have been reviewed with assessment of
their strengths and limitations, with the intention to guide investigators towards the methods most useful to their purpose.
CONTENTS
1. Introduction ............................................. .. 32
2. Sample preparation ........................................ .. .. .. .. .. 32
2.1. General considerations ................................. .. .. .. .. .. .. .. .. . .. 32
2.2. Partition techniques .................................... .. .. .. .. .. . .. .. .. 32
2.2.1. Solvent partitioning .............................. .. .. .. 32
2.2.2. Liquid-liquid partition techniques ................... .. .. 33
2.2.2.1. Counter-current distribution ................. .. .. 33
2.2.2.2. Droplet counter-current chromatography ....... .. .. . .. .. .. .. 34
2.3. Low-pressure column chromatography ..................... .. .. .. .. .. .. .. .. .. .. 34
2.3.1. Analytical scale: disposable cartridges for sample clean-up .. .. .. .. .. .. .. 34
2.3.1.1. Normal-phase cartridges .................... .. 34
2.3.1.2. Reversed-phase cartridges .................. .. .. 34
2.3.1.3. Immobilized phenylboronic acid (PBA) cartridges .. .. .. .. .. .. 35
2.4. Preparative scale ...................................... .. .. .. .. .. .. .. .. .. 35
2.5. Selected examples of protocols ........................... .. .. .. .. .. . .. .. .. 37
3. High-performance liquid chromatography ...................... .. .. .. .. .. .. .. 37
3.1. Correspondence between TLC and HPLC .................. .. .. 37
3.2. Chromatographic procedures ............................ .. .. .. 38
3.2.1. Normal-phase systems ............................ .. .. .. .. 38
* Corresponding author.
TABLE 1
PARTITION COEFFICIENTS OF ECDYSTEROIDS AND
VARIOUS PRECURSORS
I ~umol~Wata(P.lNNaOIi.~rAfOH)
~maJl&water(l:l:l)
1 1~ tplum>Mculfald
water (5:4:1); Ropmool-~w~....
Ecdysteroid K Ref.
Cyclohexane-n-butanol-water (5:5:10)
Ecdysone 3.54 71
Makisterone A 1.27 72
20-Hydroxyecdysone 0.52 71
3-Epi-20-hydroxyecdysone 0.52 72
26-Hydroxyecdysone 0.39 73
20,26_Dihydroxyecdysone 0.06 72
n-Butanol-water (19)
I Ecdysone cu. 10 74
FINAL PURIFICATION OR ANALYSIS 20-Hydroxyecdysone 5.3 9
~in-kyp.~&b+wWliquiddmnufography
Ethyl acetate-water (1:l)
2,22-Dideoxyecdysone 20 116
Fig. 1. General extraction and purilication chart for ecdy-
2-Deoxyecdysone 4
steroids.
Ecdysone 0.4
20-Hydroxyecdysone 0.1
Chloroform-methanol-water (2:l:l)
Other suitable partitioning systems for removing 2,22-Dideoxyecdysone 13 116
lipids include hexane-aqueous methanol (7:3, ZDeoxyecdysone 2.7
v/v) or light petroleum (b.p. 4040°C)-aqueous Ecdysone 0.4
20-Hydroxyecdysone 0.1
methanol. Mixtures of water-propanol (PrOH)
(3:1, v/v) and hexane can also be used to Chloroform-ethanol-water (1:l:l)
remove non-polar contaminants, the ecdyster- Ecdysone 4.6 116
20-Hydroxyecdysone 1.5
oids remaining in the aqueous phase. The addi-
tion of (NH,)$O, to promote the formation of Chloroform-water (1:l)
2,22,25Trideoxyecdysone (Ketodiol) 90 116
the two phases may be required with the PrOH
2,22_Dideoxyecdysone 20
mixture. The separation of polar impurities from 2-Deoxyecdysone 2.9
the ecdysteroids can be achieved by partition Ecdysone 0.06
between water and BuOH (ecdysteroids partition 20-Hydroxyecdysone 0.015
into the organic phase) and water and ethyl Hexane-acetonitrile (1:l)
acetate (EtOAc) (ecdysteroids remain in the Cholesterol 1.5 116
aqueous phase). 2,22,25-Trideoxyecdysone 0.06
The major factors governing the choice of 2,22_Dideoxyecdysone co.01
2-Deoxyecdysone co.01
solvent partition system are the type of con- Ecdysone co.01
taminants to be removed (i.e. mainly lipids or
mainly polar, etc.) and the nature of the ecdy-
steroids to be isolated. Thus the addition or
removal of one OH group, or conjugation to
polar (e.g. sulphate) or non-polar (e.g. acetate or
fatty acyl) groups can significantly affect parti- 2.2.2. Liquid-liquid partition techniques
tion ratios. Partition coefficients of representa- 2.2.2.1. Counter-current distribution (CCD).
tive ecdysteroids and precursors in a number of Counter-current distribution between BuOH and
systems are given in Table 1. water is effective for removing polar contami-
34 R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53
nants (the addition of a small amount of salt raphy) was proposed [14]. In that case, the
reduces emulsion formation) and CHCl,- column is a multi-layer coil and efficient mixing
MeOH-water or hexane-PrOH-water. is achieved through planetary rotation. Centrifu-
This technique, as is the case for all other gal force replaces normal gravity and allows
liquid-liquid partitions systems, has the advan- separation to be achieved in hours rather than
tage of a 100% sample recovery. The technique days.
is not of wide use at the present time. It has been
advantageously replaced by the more recent 2.3. Low-pressure column chromatography
DCCC and RLCC techniques (see below).
2.2.2.2. Droplet counter-current chromatog- 2.3.1. Analytical scale: disposable cartridges for
raphy (DCCC). This technique provides an effi- sample clean-up
cient means for purifying samples up to the gram 2.3.1.1. Normal-phase cartridges. The use of
range. It belongs to the family of liquid-liquid low pressure chromatography on a small column
partition chromatographic methods. has been used very early in phytoecdysteroid
DCCC equipment is compact, it can be used research for the fractionation of crude extracts.
with rather crude samples, and in favourable It was at that time either silica or alumina which
cases it can allow the preparation of “pure” was used in such columns (normal phase sys-
compounds [lo-131. In our hands, it seemed tems), generally eluted with binary mixtures -a
however to require a subsequent HPLC step in step-gradient of alcohol in chloroform or ben-
order to get fully pure ecdysteroids. Several zene. Use of these disposable cartridges is now
systems in the ascending or descending mode generally called solid-phase extraction (SPE).
have been described (Table 2). 2.3.1.2. Reversed-phase cartridges. The availa-
Rotation locular counter-current chromatog- bility of hydrophobic phases (resins like Amber-
raphy (RLCC) was designed as an alternative to lite or hydrocarbon-bonded silica) had led to a
DCCC [14]. It was applied to the purification of complete renewal of the procedures [16]. More-
Vitex strickeri methanolic extracts [15], in combi- over, the design of small cartridges or syringes
nation with recycling HPLC. containing 0.2-l g of HPLC phase has led to an
Although efficient, DCCC and RLCCC are ideally suited material for a rapid clean-up of
rather time-consuming, and processing a single small samples.
sample usually requires l-5 days [14]. This is Among them, Sep-Pak cartridges from Waters
because of the need to allow good exchange to are the most widely used (Fig. 2 [17-211. In fact
proceed between the mobile droplets and the their use has been extended not only for bio-
stationary phase. To overcome this drawback, logical extracts but also for desalting purposes,
HSCCC (high-speed counter-current chromatog- e.g. direct adsorption of ecdysteroids from cul-
TABLE 2
1 Q
ecdysteroids [22] and it could have several
I specific developments.
s ml mthanol-water 31
TABLE 3
USE OF LOW-PRESSURE CHROMATOGRAPHY FOR MEDIUM- OR LARGE-SCALE PURIFICATION
Normal phases
Silica (silica gel, silicic acid CHCI,-MeOH (step gradient) 76
or celite) CHCl,-MeOH (step gradient) 77
CHCl,-MeOH (step gradient) 78
CHCl,-MeOH (step gradient) 79
(+ 1% CH,COOH)
CHCl,-MeOH (8:2) 80
CHCl,-MeOH (955) 81
CHCl,-MeOH (100:3) 82
CHCl,-EtOH (19:l) 83
Me&O-CH,Cl,-water (2:8:1), 84
then EtOH
CHCl,-MeOH-water (82~1 or 85
7.5:2:1, lower phase)
CH,Cl,-EtOH (step gradient) 49
EtOAc 86
EtOAc-MeOH (step gradient) 87
Alumina CHCl,-MeOH (step gradient) 77
(containing usually 10% water) CHCl,-MeOH (step gradient) 78
CHCl,-MeOH (step gradient) 88
CHCl,-MeOH (step gradient) 89
CHCl,-MeOH (2:l) 90
CHCl,-EtOH (step gradient) 91
CH,Cl,-EtOH (step gradient) 92, 93
EtOAc-EtOH (step gradient) 94
Me,CO-CH,Cl,-water 84
(62.5:15:10)
CH,CI,-EtOH (9:l) 95
EtOAc-MeOH (1:l) 76
EtOAc-EtOH (1:l) 83
EtOAc-EtOH (2:l) 96
Sephadex LH-20 CH,Cl,-Me,CO 83, 97
CH,Cl,-MeOH (step gradient) 83, 97
CHCl,-EtOH (88:12) 96
Reversed-phases”
Amberlite XAD-2 Water-MeOH (step gradient) 98,99
(medium-pressure LC)
Sephadex LH-20 EtOH-water (7:3) 81
MeOH 85, 92, 93
Celite impregnated with Water (saturated with BuOH 76
BuOH + cyclohexane + cyclohexane)
Polyamide Water 80, 93, 100
(for removal of flavonoids)
1
33% Me-OH 70% MeOH (3 I) 100% MeOH (4 I)
Polarimpvitis Ponaaaonc A (60 mg)
Traces of ecdystcroids
(Ponasterme c) I
chmmatography on a silica column
I
Ponasterone A (6.1 g)
Ponasterone B (0.05 g)
Ponasuronc c (O.Sg)
2.5. Selected examples of protocols provide unambiguous evidence for the identity of
a given compound.
Some representative protocols are given in When facing a separation problem it is pos-
Figs. 4-7, that each use a part of the above sible to make a logical choice, according to the
techniques. Other complex processing schemes results of a preliminary TLC step. Current
can be found elsewhere [23-251. HPLC techniques have been designed about ten
years ago and the HPLC of ecdysteroids may
now be considered to be a mature technique (for
3. HIGH PERFORMANCE LIQUID reviews see refs. 26-31).
CHROMATOGRAPHY
I I
Methanol-WY 1:l *L, Chl~fOIm solids
Combined
Triturat~Iwithwater(-> insotubleresiduel)
Chlomfotm _I_ Methanol-water B* I
I
canccn~
Ttiturafed with water (--> insoluble tesiduc2)
Chmmatography on silicic acid
I
I
chromatography on c&e MSS
cIystalliscdinanicebatb
I I
Rccrysrallise
fromEtoAc-water
1O:l
I
2044YDROXYECDYSON!Z(170 mg)
molecules of the same family may evoke the Fig. 6. Isolation of ecdysteroids from Kalaaizna seeds [lOBI.
same AR,. Accordingly, the AR, values are
cumulative, and two independent changes 1 and AR, = log k; - log ki = log(kalki), or
2 evoke a final AR, = AR,, + AR,,.
When using HPLC, the behaviour of a given AR, = log (Y
compound is expressed by its retention time (or
The consequence of this is that a given modi-
volume) t, or better by its capacity factor k’,
fication (e.g. presence or absence of a given
defined as k’ = (rR - t,lt,), where t, is the re-
-OH group) evokes a similar effect on various
tention time of a compound eluted in the void
compound pairs; for instance the (Yvalue for the
volume. The correspondence with TLC is easy to
ecdysone:20-hydroxyecdysone pair is the same
make, as
as for the 2-deoxyecdysone:2-deoxy-20-hydroxy-
f, = t,lR,, k’ = (l/R, - 1) and thus R, = log k’ ecdysone one, provided of course that isocratic
conditions are used. Moreover, it seems possible
Therefore, provided that the chromatographic
to determine the expected retention time of a
conditions and the phases are the same, it is easy
compound where no reference is available [32].
to deduce the HPLC behaviour from the TLC
Similar calculations apply to acetates, conju-
data, and the relationship between R, and k’ is
gates, etc. and may prove useful in many cases.
exemplified in Fig. 8. It is clearly apparent from
it that HPLC solvents might be chosen from
among those giving R, between 0.1 and 0.3, in 3.2. Chromatographic procedures (Table 4)
order to avoid a too rapid elution or extended
analysis times. 3.2.1. Normal-phase systems
Moreover, the selectivity factor LXfor a pair of Normal-phase (NP) systems generally use sil-
compounds 1 and 2 analyzed by HPLC can be ica columns (sometimes aminopropyl- or diol-
defined as CY= kilki, and from this bonded columns) and a standard system of
R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53 39
Dliedmaedsample TABLE 4
24
silica) can also be used instead of silica ones pressure, a problem that can be overcome by
[27,34]. Diol-bonded columns have been used increasing solvent temperature.
with Chenopodium album and Kochia scoparia
(Chenopodiaceae) extracts [35,36] whereas 3.2.2. Reversed-phase systems
aminopropyl ones proved particularly efficient Reversed-phase HPLC with C,,-bonded col-
for the separation of mixtures of 3a-OH, 3P-OH umns is the most widely used system, and it
and 3-oxoecdysteroids [37] -indeed these sepa- provides efficient separations. Methanol-water
rations were better than those achieved on silica mixtures are usual, and they represent the
columns [3]. In addition, bonded phases allow easiest way to obtain satisfactory separations,
the use of gradients without the problems linked even if they are not the most efficient ones.
with long re-equilibration times encountered Other water-miscible organic solvents may
with silica columns. equally be used, and acetonitrile provides the
Non-polar bonded-phase columns can also be advantage of forming mixtures with water that
used with the above solvents, and they provide have a much lower viscosity. On the other hand,
somewhat efficient separations with reduced re- 50% methanol is a much better solvent than
tention times [38]. In this respect, trimethylsilane 20-25% acetonitrile for preparative purposes.
(TMS) bonded phases seem particularly interest- Selectivity linked with the nature of the or-
ing: they give very symmetrical peaks and a ganic phase is an important parameter for de-
selectivity that differs from silica columns. The termining the most adequate solvent mixture for
retention of ecdysteroids on non-polar bonded a given separation: depending upon the specific
columns results from the presence of remaining problem, better results may be obtained either
free silanol groups on bonded phases (this be- with methanol, acetonitrile or THF [32,39].
comes especially evident when columns have Recycling HPLC using methanol as eluent
previously been used over a long period) which may provide an interesting method for the res-
could in theory permit both NP- and RP-HPLC olution of closely migrating compounds and it
with a single column! There exist some HPLC was successfully used for the separation of ecdy-
columns (cyanopropyl or nitrile-bonded) that steroids from the bark of Vitex strickeri [15]. This
have been designed both for NP and RP pur- method appears to have several advantages,
poses and these might probably be used for regarding its ability to separate closely related
normal-phase chromatography of polar (but of compounds at the preparative scale and its low
course non-ionic) ecdysteroids. solvent consumption [40].
Dichloromethane-based solvents, although
very efficient for chromatographic separations, 4. PLANAR CHROMATOGRAPHY
suffer from a high UV-cutoff and quenching
properties of this compound, which preclude the Planar chromatography, particularly thin-layer
use of diode-array detectors or in-line radioac- chromatography (TLC) has been used extensive-
tivity monitors. This problem may be overcome ly for the qualitative analysis and isolation
with isooctane-based mixtures [29] which in of phytoecdysteroids. Paper chromatography,
counterpart suffer from a lower efficiency (plate whilst now generally considered to be obsolete,
number) and poor ecdysteroid solubility, which has been used for the separation of ecdysteroids
may become inconvenient for (1) polar ecdy- [41] and for completeness these results are pre-
steroids and (2) preparative purposes. The selec- sented in Table 5.
tivity of such solvent systems is also very differ- NP-TLC systems using silica as stationary
ent from that of dichloromethane-based ones. phase have most frequently been used to sepa-
More recently, [32] cyclohexane-based mixtures rate ecdysteroids. A widely used general solvent
were tested: they provide much better solubili- system has been chloroform-ethanol (4:1, v/v)
ties and allow the use of diode-array detectors. with a single ascending development, however,
On the other hand, these mixtures display a many normal phase solvent systems have been
significantly higher viscosity and hence working described for these and details of a number of
R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53 41
TABLE 5
R, OF ECDYSTEROIDS ON PAPER CHROMATOGRAPHY [41]
Solvent systems are prepared by saturating the non-polar component (A) with the mixture (B + C) of polar solvents.
A B C
solvent systems used to separate the ecdysteroids This reagent gives a range of colour with differ-
are given in Table 6 [42]. The R, values obtained ent ecdysteroids as indicated in Table 8. The
for some 20 compounds in the chloroform-etha- colours produced range from pink (e.g. cyas-
nol (4:l) solvent system are provided in Table 7 terone) and red (e.g. 20-hydroxyecdysone) to
[43]. Following separation by NPTLC the detec- dark green (e.g. ecdysone), but like many such
tion of ecdysteroids on TLC plates can be colour reactions are subject to quite wide inter-
accomplished in a number of ways. Where the laboratory variations. Table 8 also lists a number
plates incorporate a suitable fluorescence in- of additional chromogenic reactions which have
dicator that can provide a general, if rather been utilised for these compounds. Scanning
non-specific method of detection. Spray reagents densitometry now provides a convenient method
have been used to provide a higher degree of for obtaining in situ UV spectra, and this can
specificity, and the vanillin-sulphuric acid re- provide increased confidence in the identification
agent has been particularly widely employed. of sample components as ecdysteroids (e.g. see
42 R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53
TABLE 6
Ecdysone 20-Hydroxyecdysone
Fig. 9). As discussed in more detail later, in situ 5P-OH compounds from their 5/3-H analogues
mass spectrometry has also been used to good (e.g. polypodine B from 20-hydroxyecdysone).
effect for the identification of phytoecdysteroids The effects of various substitution patterns on
following NP-TLC. the RP-TLC of ecdysteroids are discussed in
RP-TLC chromatography on bonded phases detail elsewhere [45]. Detection methods follow-
(C,, C,, C12, C1*, aminopropyl and cyano- ing RP-TLC are essentially the same as those
propyl) as well as on paraffin-impregnated silica employed for NP-TLC separations, although RP-
gel (e.g. refs. 39, 44 and 45) has also been TLC-MS has not proved to be so readily
employed for ecdysteroids and typical results, achieved as NP-TLC-MS .
obtained using various C,, bonded phases, for a Separations can also be modified, ‘in both NP-
range of structures are provided in Table 9 [45]. and RP-TLC systems by esterifying 20,22-diol-
In general, methanol-water (l:l, v/v) solvent containing ecdysteroids with boronic acids (e.g.
systems provide good separations but other or- phenylboronic acid) 146,471. This has the effect
ganic modifiers can be used to achieve different of reducing the polarity of such esters with a
selectivities [44]. Reversed-phase systems, like consequent effect on the chromatographic sepa-
their HPLC equivalents are poor at separating ration of an ecdysteroid mixture. This can be
R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53 43
TABLE 7
R, AND ReedysoneVALUES OBTAINED FOR ECDYSTEROIDS BY NP-TLC ON SILICA GEL TLC PLATES USING
CHLOROFORM-ETHANOL (4:l)
used to good effect to resolve ponasterone A and analysis of plant extracts expecially with the high
2-deoxyecdysone, which are otherwise difficult to concentrations of material frequently encoun-
separate [4,6]. tered in such samples. It therefore seems likely
Of the more recently developed techniques for that, with the wider availability of scanning
planar chromatography, automated multiple de- densitometers that the use of TLC and HPTLC
velopment (AMD) and over-pressure layer chro- for quantitative analysis of these compounds will
matography (OPLC) have been applied to the increase.
resolution of ecdysteroids [48,49]. AMD pro-
vided good separations but was somewhat time- 5. GAS CHROMATOGRAPHY
consuming [48]. The use of OPLC for the sepa-
ration of phytoecdysteroids of Silene nutans [49] Ecdysteroids are too polar and have too little
enabled very rapid analysis to be performed, and thermal stability to be suitable subjects for gas
it could easily be used for preparative work. chromatography (GC) but by protecting some or
However, the time required for plate prepara- all of the hydroxyl groups as silyl ethers they can
tion, etc., means that this is unlikely to supplant be thermally stabilized and their polarity reduced
conventional separation methods. so that they can be chromatographed in the
The current practice of TLC for the analysis of gaseous phase. The gas chromatography of
phytoecdysteroids is to use it for qualitative ecdysone as a trimethylsilyl ether derivative was
applications such as monitoring extractions and first described by Katz and Lensky [52], although
purification (see refs. 50 and 51 for some recent the exact nature of the product was not stated.
examples). In these applications the ease of TLC Morgan and Woodbridge [53,54] developed con-
and the generally rugged nature of the separa- ditions for the preparation of trimethylstyl ether
tion system are well suited to a rapid and high 0-methyloximes for the quantitative study of
throughput of often difficult samples. However, ecdysteroids, but later it was found that protec-
there is no practical reason why scanning den- tion of the ketone group as an 0-methyloxime
sitometry could not be used for quantitative was not necessary. Poole et al. [55] and Borst
44 R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53
TABLE 8
REACTIONS OF ECDYSTEROIDS AND THEIR DERIVATIVES WITH SPRAY REAGENTS
TABLE 8 (continued)
TABLE 9
Compound Merck C,, bonded TLC Whatman C,, bonded Macherey-Nagel C,,
plates TLC plates bonded TLC plates
tor. The flame detector precludes the use of methanol mixtures is a practical procedure, both
organic modifiers but has great sensitivity and for pure ecdysteroids and for plant extracts (see
resolution. Raynor et al. [61] have used capillary Fig. 10) [63]. Very short columns can be used
supercritical fluid chromatography (SFC) to with advantages of short retention times and
chromatograph some relatively non-polar ecdy- high resolution. The sensitivity of detection is
steroid precursors but the retention times of greater than with HPLC probably as a result of
ecdysone and hydroxyecdysone were too great the sharper peaks and greater UV transparency
for practical purposes. Perhaps these experi- of the fluid [64]. Packed-column SFC has the
ments would be worth repeating using higher retention characteristics of a NP-HPLC system.
pressures and possibly slightly higher tempera- A variety of columns from silica to ODS can be
ture. Now that the electron capture detector is used. In practice, aminopropyl silica and cyano-
becoming adaptable to SFC conditions, it be- propyl silica seem best for ecdysteroids. Some
comes more important to attempt capillary SFC conditions are given in Table 10. A further
of ecdysteroids. SFC with electron-capture de- advantage of SFC over HPLC is the avoidance of
tection could become a superior method for costly high purity solvents and the problems of
ecdysteroids in terms of resolution, sensitivity their recovery and disposal. Since many ecdy-
and selectivity. Morgan et al. [62] have shown steroids contain vicinal diols, particularly 2,3-
that packed-column SFC using carbon dioxide- and 20,22-diols, reagents specific for vicinal diols
R. Lafont et al. I J. Chromatogr. A 6.58 (1994) 31-53 47
!LL
1 dium dodecyl sulphate-methanol (5%, v/v) at
pH 9.4. Depending upon the system, separations
were performed at 20 or 8.5 kV and a run
temperature of 50°C. Increasing either the ap-
7
plied voltage or temperature reduced migration
times. The organic modifier content of the mo-
612345 012345
bile phase also affected the migration of the
Min
ecdysteroids with elution time increasing with
Fig. 10. Supercritical fluid chromatograms of plant extracts.
(A) Silene otites and (B) Silene nutans [63]. Peaks: 1 = 2- organic modifier content.
deoxyecdysone; 2 = 2-deoxy-20-hydroxyecdysone; 3 = poly- The migration times of a number of common
podine B; 4 = 20-hydroxyecdysone; 5 = 26-hydroxypoly- phytoecdysteroids, in the two CZE systems ex-
podine B; 6 = integristerone A; 7 = 20,26dihydroxyecdy- amined, with the run buffer described above are
sone. Conditions: column 5 pm cyanopropyl silica (20 x 4.6
given in Table 11. Using a 50 pm I.D. capillary,
mm) CO,-MeOH (9:l) at 3 ml min-‘, 60°C and 290 bar,
detection at 235 nm. migration times ranged from 7.2 min for cyas-
terone up to 25.9 min for 2-deoxyecdysone. The
observed elution order is similar to that seen in
can be useful in selecting out such compounds RP-HPLC. This is not surprising as in MCE the
from mixtures. We have described the use of separation mechanism for ecdysteroids is based
solid-phase extraction with immobilized boronic on hydrophobic partitioning. It is interesting to
acids for the selective retention of ecdysteroids note that there is an excellent separation be-
containing 20,22-diols [22]. We have recently tween ponasterone A (16.5 min) and 2-deoxy-
extended this to the preparation and separation ecdysone (25.9 min) which have been difficult to
by SFC of methyl, butyl and phenylboronic separate by RP-HPLC, TLC, etc. The results
esters of ecdysteroids with 20,22-diol groups obtained with a 75 pm capillary were similar.
[651. MCE gave a linear calibration curve over the
range 0 to 560 pg/ml, corresponding to 0 to 2.8
7. CAPILLARY ELECTROPHORESIS ng injected on-column. The sensitivity of the
instrument was such that the detection of cu. 175
Capillary zone electrophoresis (CZE) has pg (35 pg/ml) on-column of a sample of pure
been shown to provide very efficient separations ecdysone was possible without difficulty. Ecdy-
for a wide range of substances. As the bulk of steroid-rich extracts from both plant and insect
the ecdysteroids are uncharged, simple CE, sources were subjected to analysis by MCE.
which relies on the presence of an ionised group Samples with a range of purities were examined
to attain separations, cannot be used. However, with different degrees of success. Attempted
micellar capillary electrophoresis (MCE) , where- MCE of simple methanol extracts of Silene
by a molecule such as sodium dodecyl sulphate is nutuns were unsuccessful because large and dis-
added to the buffer at a concentration above its torted peaks were obtained, apparently due to
critical micelle concentration, has been shown to column overloading and the presence of interfer-
be suitable for such compounds. In a series of ing compounds. The analysis of plant extracts
preliminary studies we have evaluated MCE after more extensive purification gave good re-
[66,67] for a range of sample types, including sults as illustrated in Fig. 11 for an extract of
plant extracts of various degrees of purity. These Silene o&es. This shows the sample to contain
studies clearly demonstrated that MCE could be 20-hydroxyecdysone , 2-deoxy-20-hydroxyecdy-
48 R. Lafont et al. I 3. Chromatogr. A 658 (1994) 31-53
TABLE 10
EXAMPLES OF CONDITIONS USED FOR SUPERCRITICAL FLUID CHROMATOGRAPHY OF ECDYSTEROIDS
1 p.s.i. = 6894.76 Pa. For many other conditions, see ref. 114.
Column type Column Mobile phase Flow-rate Pressure or Temperature Compound ra Ref.
dimensions density (“C) @in)
Cllpilltl~
Cyanopropyl- 10mxS pm CO, - 0.4-0.71 g 120 14,22,25-Tridwxyecdysone 20 61
dimethylsiloxane (1:l) cK3 2-Deoxyecdysone 31 61
Packed
Cyanopropyl 250 x 4.6 mm CO,-MeOH 3.25 I min-’ 300 bar 50 14,22,25_Trideoxyecdysone 1 61
Spherisorb 5 pm (9O:lO) gas 2-Deoxyecdysone 2.3 61
Ponasterone A 2.5 61
Ecdysone 3.2 61
Polypodine B 3.2 61
Makisterone A 3.2 61
20-Hydroxyecdysone 4.0 61
Cyasterone 7.0 61
1
a b
Od-
0 ;
0
10 20 30 Timelmin
Timelmin Fig. 12. MCE, using the Beckman P/ACE, of an ecdy-
Fig. 11. MCE of an ecdysteroid-rich extract of the plant steroid-rich extract of the plant Silene nutans using (a)
Silene otites using the ABI 270A. The extract contained, in standard conditions in which polypodine B (1) and ZO-
order of elution, (1) 20-hydroxyecdysone, (2) 2-deoxy-ZO- hydroxyecdysone (2) co-migrate, but are resolved from 2-
hydroxyecdysone and (3) 2-deoxyecdysone. Conditions: UV deoxy-20-hydroxyecdysone (3); and (b) 52.5% of methanol
absorption at 240 nm, mobile phase 5% methanol and 95% in the run buffer when 20-hydroxyecdysone elutes before
buffer containing 40 mM NaH,PO,, 20 mM Na,HBO, and polypodine B. Conditions as in Fig. 11, but varying the
20 mM sodium dodecyl sulphate, pH 9.4, 20 kV mobile phase in (b) and using 8.5 kV throughout.
50 R. Lafont et al. / J. Chromatogr. A 6.58 (1994) 31-53
8. HYPHENATED TECHNIQUES OF
CHROMATOGRAPHY AND MASS SPECTROMETRY
10
The large number of phytoecdysteroids that
have been encountered in plant extracts clearly
poses problems in their identification based
merely on chromatographic properties alone. SCAN No.
Whilst selectivity can be attained by the use of a Fig. 13. HPLC-MS of ecdysteroids present in an extract of
number of different chromatographic systems Silene otites separated by RP-HPLC on a C,, bonded column
(see also Lafont et al. [32]), linked chroma- (ODS Spherisorb 20 x 0.46 cm) using acetonitrile-0.1 M
ammonium acetate (4060) at 1 ml min-’ with a thermospray
tography-mass spectroscopy provides an alter-
interface (VG Quatro, capillary temperature 270°C). (a) UV
native system for unequivocally establishing trace at 240 nm; (b) ion current for m/z 481 for 20-hydroxy-
identity. The use of GC-MS for ecdysteroids is ecdysone; (c) ion current m/z 465 for 2-deoxy-2-hydroxy-
well established for insect and other arthropod ecdysone and (d) ion current m/t 449 for 2-deoxyecdysone.
samples but has not been widely applied to plant
samples. The reason for this is that in the case of at m/z 481, 465 and 449 are shown using the
arthropod samples it is the sensitivity as well of thermospray interface. The resulting spectrum
the specificity of the mass spectrometer as a for 20-hydroxyecdysone (m/z 481) is given in
detector which is the driving force for its use. In Fig. 14A. It is typical of the results obtained with
the case of plant samples the quantities of this interface following reversed-phase HPLC
material are usually such that it is considered to (methanol-ammonium acetate buffer, see cap-
be easier to isolate material for mass spec- tion for details). A number of ions are associated
trometry rather than go through the process of with the 20-hydroxyecdysone peak corre-
derivatisation which is necessary before GC can sponding to the protonated molecular (MH) ion
be used. The introduction of SFC has also (481), MH - 18 (463) and MH -2 x 18 (445)
enabled the ecdysteroids to be introduced into (loss of one or two molecules of water) and an
the mass spectrometer without the need for ion corresponding to M + MeCN + NH: (539),
derivatisation, and SFC-MS has been used for highlighting the difficulties of interpretation en-
the identification of a number of ecdysteroids in countered with this mode of ionisation. In order
extracts of Silene nutans and Silene otites [63]. to characterise more fully such compounds with
The results obtained were either chemical ioniza- this interface we have recently employed MS-
tion- or electron-impact-like, depending upon MS techniques. An example of the results ob-
the temperature of the jet block, with higher tained is shown in Fig. 14B where the daughter
temperatures associated with greater fragmenta- ion spectrum obtained for the ion at m/z 481 is
tion. Conditions could thus be varied to give shown. The weak ion at m/z 463 (not marked)
molecular mass or diagnostic fragmentation data. and ions at m/z 445 and 427 correspond to loss
In the case of HPLC-MS for phytoecdyster- of 1, 2 and 3H,O respectively, but no simple
oids progress has been limited by the available explanation can be given for the other ions.
interfaces. In Fig. 13 the chromatograms ob- Similar spectra were obtained for 2-deoxy-20-
tained for HPLC-MS of an extract of Silene hydroxyecdysone and 2-deoxyecdysone.
otites with UV and MS data, monitoring the ions In addition to the hyphenation of SFC and
R. Lafont et al. I J. Chromatogr. A 658 (1994) 31-53 51
REFERENCES
18 R. Lafont, J.-L. Pennetier, M. Andrianjalintrimo, J. 42 E.D. Morgan and C.F. Poole, Adv. Insect Physiol., 12
Claret, J.-F. Modde and C. Blais, J. Chromutogr., 236 (1976) 17; and references cited therein.
(1982) 137. 43 I.D. Wilson, J. Chromatogr., 318 (1985) 373.
19 R.D. Watson and E. Spaziani, J. Liq. Chromatogr., 5 44 I.D. Wilson, S. Scalia and E.D. Morgan, J. Chroma-
(1982) 525. togr., 212 (1981) 211.
20 G.D. Primprikar, M.J. Coign, H. Sakurai and J.R. 45 I.D. Wilson, J. Planar Chromatogr., 1 (1988) 116.
Heitz, J. Chromutogr., 317 (1984) 413. 46 J. Pis and J. Harmatha, 1. Chromatogr., 596 (1992) 271.
21 H.H. Rees and R.E. Issac, Methods Enzymol., 111 47 I.D. Wilson, J. Planar Chromatogr., 5 (1992) 316.
(1985) 377. 48 I.D. Wilson and S. Lewis, J. Chromatogr., 408 (1987)
22 S.J. Murphy, E.D. Morgan and I.D. Wilson, in A.R.
445.
McCaffery and I.D. Wilson (Editors), Chromatography
49 H. Read, I.D. Wilson and R. Lafont, in A.R. McCaffery
and Isolation of Insect Hormones and Pheromones,
and I.D. Wilson (Editors), Chromatography and Zsola-
Plenum Press, London, 1990, p. 131.
tion of Insect Hormones and Pheromones, Plenum Press,
23 J.-P. Girault, M. Bathori, E. Varga, K. Szendrei and R.
London, New York, 1990, p. 129.
Lafont, J. Nat. Produca, 53 (1990) 279.
50 M. Bathori, K. Szendrei, H. Kalasz, R. Lafont and J.-P.
24 D. Rudel, M. Bathori, J. Gharbi, J.-P. Girault, I. Racz,
Girault, Chromatographia, 25 (1988) 627.
K. Melis, K. Szendrei and R. Lafont, Planta Med., 58
51 H. Kalasz, M. Bathori, L. Kerecsen and L. Toth, J.
(1992) 358.
25 M. Wessner, B. Champion, J.-P. Girault, N. Kaouadji, Planar Chromatogr., 6 (1993) 38.
B. Saidi and R. Lafont, Phytochemistry, 31 (1992) 3785. 52 M. Katz and Y. Lensky, Experientia, 26 (1970) 1043.
26 R. Lafont, G. Sommt-Martin, B. Mauchamp, B.F. 53 E.D. Morgan and A.P. Woodbridge, Chem. Commun.,
Maume and J.-P. Delbecque, in J.A. Hoffmann (1971) 475.
(Editor), Progress in Ecdysone Research, Elsevier/ 54 E.D. Morgan and A.P. Woodbridge, Org. Mass Spec-
North-Holland, Amsterdam, 1980, p. 45. tram., 9 (1974) 102.
27 R. Lafont, P. Beydon, B. Mauchamp, G. Somme-Mar- 55 C.F. Poole, E.D. Morgan and P.M. Bebbington, J.
tin, M. Andrianjafintrimo and P. Krien, in F. Sehnal, A. Chromatogr., 104 (1975) 172.
Zabaz, J.J. Menn and B. Cymborowski (Editors), Regu- 56 D.W. Borst and J.D. O’Connor, Steroids, 24 (1974) 637.
lation of Insect Development and Behaviour, Wroclaw 57 C.R. Bielby, A. Gande, E.D. Morgan and I.D. Wilson,
Technical University Press, Wroclaw, 1981, p. 125. J. Chromatogr., 194 (1980) 43.
28 J.C. Touchstone, in G. Zweig and J. Sherma (Editors), 58 R.P. Evershed, J.G. Mercer and H.H. Rees, J. Chroma-
CRC Handbook of Chromatography, Steroids, CRC togr., 390 (1987) 357.
Press, Boca Raton, FL, 1986, p. 119. 59 R.P. Evershed, M. Kabbouh, M.C. Prescott, J.L. Maggs
29 R. Lafont, in H. Kalasz and S. Ettre (Editors), Chroma- and H.H. Rees, in A.R. McCaffery and I.D. Wilson
tography ‘87, Akademia Kiado, Budapest, 1988, p. 1. (Editors), Chromatography and Isolation of Insect Hor-
30 M.J. Thompson, J.A. Svoboda and M.F. Feldlaufer, in mones and Pheromones, Plenum Press, London, New
W.D. Nes and E.J. Parish (Editors), Analysis of Sterols York, 1990, p. 103.
and Other Biologically Significant Steroids, Academic 60 J.E. Lovelock, in C.M. White (Editor), Modern Super-
Press, San Diego, CA, 1989, p. 81. critical Fluid Chromatography, VCH, Weinlieim, 1988.
31 R. Lafont and I.D. Wilson, in A.R. McCaffery and I.D. 61 M.W. Raynor, J.P. Kithinji, I.K. Barker, K.D. Bartle
Wilson (Editors), Chromatography and Isolation of and I.D. Wilson, J. Chromatogr., 436 (1988) 497.
Insect Hormones and Pheromones, Plenum Press, Lon- 62 E.D. Morgan, S.J. Murphy, D.E. Games and I.C.
don, New Ybrk, 1990, p. 79. Mylchreest, J. Chromatogr., 441 (1988) 165.
32 R. Lafont, N. Kaouadji, E.D. Morgan and I.D. Wilson, 63 M.W. Raynor, J.P. Kithinji, K.D. Bartle, D.E. Games,
J. Chromatogr. A, 658 (1994) 55. I.C. Mylchreest, E.D. Morgan, R. Lafont and I.D.
33 R. Lafont, G. Sommt-Martin and J.-C. Chambet, J. Wilson, J. Chromatogr., 467 (1989) 292.
Chromatogr., 170 (1979) 185. 64 E.D. Morgan, H.-P. Huang and I.D. Wilson, in A.R.
34 L.N. Dinan, P.L. Donnahey, H.H. Rees and T.W. McCaffery and I.D. Wilson (Editors), Chromatography
Goodwin, J. Chromatogr., 205 (1981) 139. and Isolation of Insect Hormones and Pheromones,
35 L.N. Dinan, Phyrochem. J., 3 (1992) 132. Plenum, London, New York, 1990, p. 95.
36 L.N. Dinan, J. Chromatogr. A, 658 (1994) 69. 65 J.-H. Shim, E.D. Morgan and I.D. Wilson, J. Chroma-
37 C. Blais and R. Lafont, Hoppe-Seylers’ Z. Physiol. togr., 639 (1993) 281.
Chem., 365 (1984) 809. 66 T. Large, R. Lafont, E.D. Morgan and I.D. Wilson,
38 N. Kaouadji and R. Lafont, J. Chromatogr., 505 (1990) Anal. Proc., 29 (1992) 386.
408-412. 67 P. Davis, R. Lafont, T. Large, E.D. Morgan and I.D.
39 I.D. Wilson, C.R. Bielby and E.D. Morgan, J. Chroma- Wilson, Chromatographia, 37 (1993) 37.
togr., 238 (1982) 97. 68 I.D. Wilson, R. Lafont and P. Wall, 1. Planar Chroma-
40 I. Kubo and T. Nakatsu, LC . GC, 8 (1992) 933. fogr., l(l988) 357.
41 M.W. Gilgan and T.E. Farquharson, Steroids, 22 (1973) 69 I.D. Wilson, R. Lafont, R.G. Kingston and C.J. Porter,
366. J. Planar Chromatogr., 3 (1990) 359.
R. Lafont et al. I J. Chromarogr. A 658 (1994) 31-53 53
70 R. Lafont, C.J. Porter, E. Williams, H. Read, E.D. 93 M. Bathori, I. Toth, K. Szendrei and J. Reisch, Phyro-
Morgan and I.D. Wilson, J. Planar Chrornalogr., in chemistry, 10 (1982) 2225.
press. 94 T. Takemoto, S. Ogawa, N. Nishimoto, K.-Y. Yen, K.
71 J.N. Kaplanis, M.J. Thompson, R.T. Yamamoto, W.E. Abe, T. Sato, K. Osawa and M. Takahashi, Yakugaku
Robbins and S.J. Louloudes, Sferoids, 8 (1966) 605. Zasshi, 87 (1967) 1521.
72 J.N. Kaplanis, S.R. Dutky, W.E. Robbins, M.J. Thomp- 95 R. Hardman and T.V. Benjamin, J. Chromatogr., 131
son, E.L. Lindquist, D.H.S. Horn and M.N. Galbraith, (1977) 468.
Science, 190 (1975) 681. 96 A. Suksamram, J.S. Wilkie and D.H.S. Horn, Phyro-
73 J.N. Kaplanis, W.E. Robbins, M.J. Thompson and S.R. chemistry, 25 (1986) 1301.
Dutky, Science, 180 (1973) 307. 97 G.B. Russel and D.R. Greenwood, in J. Koolman
74 P. Karlson, Angew. Chem., Znt. Ed. Engl., 2 (1963) 175. (Editor), Ecdysone, From Chemistry io Mode of Action,
75 I. Kubo and F.J. Hanke, in J.R. Miller and T.A. Miller Georg Thieme Verlag, Stuttgart, 1989, p. 97.
(Editors), Insec&Planr Interactions (Springer Series in 98 M. Hori, Seroids, 14 (1969) 33.
Experimental Entomology), Springer, New York, 1986, 99 D.A. Schooley, G. Weiss and K. Nakanishi, Steroids, 19
p. 225. (1973) 377.
76 M.N. Galbraith and D.H.S. Horn, Aust. J. Chem., 22 100 R. Hardman and T.V. Benjamin, Planta Med., 39 (1980)
(1969) 1045. 148.
77 S. Imai, M. Hori, S. Fujioka, E. Murata, M. Goto and 101 H.N. Nigg, M.J. Thompson, J.N. Kaplanis, J.A.
K. Nakanishi, Tetrahedron Letr., 36 (1968) 3883. Svoboda and W.E. Robbins, Steroids, 23 (1974) 507.
78 A. Banerji, G.J. Chintalwar, N.K. Joshi and M.S. 102 E.D. Morgan and C.F. Poole, J. Chromatogr., 116
Chadha, Phytochemisrry, 10 (1971) 2225. (1976) 333.
79 M.L. Colombo, F. Tome, 0. Servettaz and C. Bugatti, 103 B. Stay, L.S. Ostedgaard, S.S. Tobe, A. Strambi and E.
Int. J. Crude Drug Res., 28 (1990) 219. Spaziani, J. Insect Physiol., 30 (1984) 643.
80 J. Jizba, L. Dolejs and V. Herout, Phyrochemisrry, 13 104 A. Moribayashi, H. Kurahashi and T. Ohtaki, Arch.
(1974) 1915. Insect Biochem. Physiol., 2 (1985) 237.
81 T. Yokota, M. Arima and N. Takahashi, Tetrahedron 105 P. Beydon and R. Lafont, in A.P. Gupta (Editor),
Leu., 23 (1982) 1275. Recent Advances in Comparative Arthropod Morpholo-
82 A. Fukuzawa, M. Miyamoto, Y. Kumagai and T. gy, Physiology and Developmenr, Vol. I, Rutgers Uni-
Masamune, Phytochemistry, 25 (1986) 1305. versity Press, New Brunswick, NJ, 1989, p. 455.
83 G.B. Russel, P.G. Fenemore, D.H.S. Horn and E.J. 106 P.V. Ripa, E.A. Martin, SM. Cocciolone and J.H.
Middleton, Aust. J. Chem., 25 (1972) 1935. Adler, Phytochemishy, 29 (1990) 425.
84 G. Heinrich and H. Hoffmeister, Experientia, 23 (1967) 107 G.M. Holman and R.W. Meola, Znsecl Biochem., 8
995. (1978) 275.
85 N. Nishimoto, Y. Shiobara, M. Kujino, S.S. Inoue, T. 108 L. Canonica, B. Danieli, G. Ferrari, J. Krepinsky and I.
Takemoto, F. De Oliveira, G. Akisue, M.K. Akisue, G. Weisz-Venze, Phytochemistry, 14 (1975) 525-527.
Hashimoto, 0. Tanaka, R. Kasai and H. Matsuura, 109 G.B. Russel and J.G. Fraser, Ausr. J. Chem., 1805,
Phytochemislry, 26 (1987) 2505. (1973).
86 A.C. Santos, M.T. Chua, N. Eufemio and C. Abela, 110 J. Koolman and K.-D. Spindler, Hoppe-Seyler’s 2.
Experientia, 26 (1970) 1053. Physiol. Chem., 358 (1977) 1339.
87 H. Hikino, K. Mohri, T. Okuyama and K.-Y. Yen, 111 M.J. Thompson, J.N. Kaplanis, W.E. Robbins and R.T.
Steroids, 28 (1976) 649. Yamamato, Chem. Commun., (1967) 650.
88 F. Camps, J. Co11 and 0. Dargallo, Ann. Quim., 81C 112 W.-S. Chou and M.-S. Lu, in J.A. Hoffmann (Editor),
(1975) 74. Progress in Ecdysone Research, Elsevier, Amsterdam,
89 R. Hardman and S.M.A. Mahar, Planra Med., 33 (1978) 1980, p. 281.
278. I13 A. Faux, M.N. Galbraith, D.H.S. Horn, E.J. Middleton
90 A.U. Mamatkhanov, M.I. Shamsutdinov and T.T. and J.A. Thomson, Chem. Commun., (1970) 243.
Shakirov, Khim. Prir. Soedin., 5 (1967) 667. 114 H.P. Huang, Ph.D. Thesis, Keele University, Keele,
91 Z. Saatov, M.B. Gorovits, N.D. Abdullaev, B.Z. Us- 1992.
manov and N.K. Abubakirov, Khim. Prir. Soedin., 5 115 G.B. Russell, D.G. Fenemore, D.H.S. Horn and E.J.
(1985) 611. Middleton, Chem. Commun., (1968) 971.
92 I. T6th, M. BBthory, K. Szendrei, E. Minker and G. 116 R. Lafont, unpublished results.
Blazs6, Fitoterapia, 52 (1981) 77.