You are on page 1of 7

D I A G N O S T I C R E P O R T

/home/hitrax/log/dsrvReports/dsrv_report_122541_20191209211724.txt
---------------------------------------------------------------------
Number of procedure errors: 0

---------------------------------------------------------------------
GENERAL 12/09/2019 21:16:48

Node Name XRAY


Machine Type HI-SCAN 100100V-2is
Serial Number 122541
Unit Network Name XRAY
Site CUN_AT6
Subsite T3-HBS
Host Address 192.168.0.1
Net Mask 255.255.255.0
Gate Address 0.0.0.0
Matrix Server Address 10.201.32.13
Software Version HX-03-11-G
Detection Code not available
Options SEN, HI-TIP-PLUS, X-PLORE, HI-SPOT, OPTI-ZOOM, IMS, X-
ACT, HDA, OJT, XPORT, HITRAX-II
Dongle ID 55A4E1B3

---------------------------------------------------------------------
KEYBOARD STATUS 12/09/2019 21:16:48

Keyboard status OK

---------------------------------------------------------------------
HARDWARE 12/09/2019 21:16:48

Sampling rate 150 Hz


Line characteristics HI-MAT, 4 LIF, 160kV
Motor characteristics frequency converter, speed 20cm/s
Project Type PT_AA_00000

---------------------------------------------------------------------
LIGHT BARRIER 12/09/2019 21:16:48

Light barrier 1 FREE


Light barrier 2 NOT_AVAILABLE
Light barrier 2x NOT_AVAILABLE
Light barrier 3 NOT_AVAILABLE
Curtain height 0
Light curtain status NOT_AVAILABLE

---------------------------------------------------------------------
POWER SUPPLY 12/09/2019 21:16:48

PI Version 3
Voltages 1: pi voltage 12.03 V
2: lif voltage 0.00 V

---------------------------------------------------------------------
LINE SCOPE 12/09/2019 21:16:48

View A - offset scope high


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 . . . . . . . . . . . . . . . . . .
70 . . . . . . . . . . . . . . . . . .
60 . . . . . . . . . . . . . . . . . .
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 ************************************
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

View A - offset scope low


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 . . . . . . . . . . . . . . . . . .
70 . . . . . . . . . . . . . . . . . .
60 . . . . . . . . . . . . . . . . . .
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 ************************************
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

View A - full scale scope high


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 ****.****** * ******* **. ******* **
70 ************************************
60 ******.***********.*****************
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 . . . . . . . . . . . . . . . . . .
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

View A - full scale scope low


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 ****.*. . * *** .*. * .** * .***.**
70 *********** ************* **********
60 ************************************
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 . . . . . . . . . . . . . . . . . .
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

View B - offset scope high


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 . . . . . . . . . . . . . . . . . .
70 . . . . . . . . . . . . . . . . . .
60 . . . . . . . . . . . . . . . . . .
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 ************************************
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

View B - offset scope low


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 . . . . . . . . . . . . . . . . . .
70 . . . . . . . . . . . . . . . . . .
60 . . . . . . . . . . . . . . . . . .
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 ************************************
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

View B - full scale scope high


100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 * ***** * **. * ****. .*********.*.
70 ************************************
60 ******* *************** .***********
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 . . . . . . . . . . . . . . . . . .
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024
View B - full scale scope low
100 . . . . . . . . . . . . . . . . . .
90 . . . . . . . . . . . . . . . . . .
80 ****.*. *** **.*****.*. ********* **
70 ********************.************ **
60 * . ******* ************* * .*******
50 . . . . . . . . . . . . . . . . . .
40 . . . . . . . . . . . . . . . . . .
30 . . . . . . . . . . . . . . . . . .
20 . . . . . . . . . . . . . . . . . .
10 . . . . . . . . . . . . . . . . . .
0 . . . . . . . . . . . . . . . . . .
+---+---+---+---+---+---+---+---+---
0 128 256 384 512 640 768 896 1024

---------------------------------------------------------------------
DETECTOR LINES 12/09/2019 21:16:48

View A - total high line detectors 1152


View A - defect high line detectors 0
View A - total low line detectors 1152
View A - defect low line detectors 0

View B - total high line detectors 1152


View B - defect high line detectors 0
View B - total low line detectors 1152
View B - defect low line detectors 0

---------------------------------------------------------------------
XRAY GENERATOR 12/09/2019 21:16:48

Generator 0 - (STANDBY) pos. anode voltage +0.00 kV


Generator 0 - (STANDBY) neg. anode voltage +0.00 kV
Generator 0 - (STANDBY) filament current +499.00 mA
Generator 0 - (STANDBY) anode current +0.00 uA

Generator 0 - (WORKING) pos. anode voltage +77.83 kV


Generator 0 - (WORKING) neg. anode voltage -77.83 kV
Generator 0 - (WORKING) filament current +572.25 mA
Generator 0 - (WORKING) anode current +659.23 uA

Generator 1 - (STANDBY) pos. anode voltage +0.00 kV


Generator 1 - (STANDBY) neg. anode voltage +0.00 kV
Generator 1 - (STANDBY) filament current +517.31 mA
Generator 1 - (STANDBY) anode current +0.00 uA

Generator 1 - (WORKING) pos. anode voltage +80.57 kV


Generator 1 - (WORKING) neg. anode voltage -80.11 kV
Generator 1 - (WORKING) filament current +595.14 mA
Generator 1 - (WORKING) anode current +682.12 uA

---------------------------------------------------------------------
STATISTICS 12/09/2019 21:16:48
total baggage count 147946
temporary baggage count 1
total suspicious baggage count 23961
temporary suspicious baggage count 1

---------------------------------------------------------------------
LINE INTERFACE 12/09/2019 21:16:48

LIF 0 - status OK
LIF 0 - lin slot 0
LIF 0 - lex slot 0
LIF 0 - # detectors 1152

LIF 1 - status OK
LIF 1 - lin slot 1
LIF 1 - lex slot 0
LIF 1 - # detectors 1152

LIF 2 - status OK
LIF 2 - lin slot 2
LIF 2 - lex slot 0
LIF 2 - # detectors 1152

LIF 3 - status OK
LIF 3 - lin slot 3
LIF 3 - lex slot 0
LIF 3 - # detectors 1152

---------------------------------------------------------------------
CALIBRATION DATA 12/09/2019 21:16:48

Calibration data
cal threat lut: semtex.lut,
{4095/4095, 3158/2987, 1878/1610, 1122/0882},
{2620/1948, 2021/1452, 1206/0812, 0724/0459},
{1828/0766, 1423/0585, 0860/0342, 0523/0201}
cal hard lut: hard.lut,
{4095/4095, 2620/1948, 1828/0766}

---------------------------------------------------------------------
I/O CONTROL 12/09/2019 21:16:48

IOC state OK
Config state OK
External default config ENABLED
External config DISABLED
Custom config DISABLED

Internal input configuration


1: LLI_INTERLOCK_CLOSED = I25
2: LLI_XRAY_LAMP_ERROR = I17
3: LLI_FOOT_SWITCH_CLOSED = !I8
4: LLI_XRAY1_FEEDBACK_ACTIVE = I11
5: LLI_XRAY1_VOLTAGE_ACTIVE = !I14
6: LLI_XRAY2_FEEDBACK_ACTIVE = I18
7: LLI_XRAY2_VOLTAGE_ACTIVE = !I15
8: LLI_REGENERATION1_TERMINATED = I16
9: LLI_REGENERATION2_TERMINATED = I20
10: LLI_LS1_INTERRUPTED = !I0
11: LLI_LS3_INTERRUPTED = !I1
12: LLI_LS_ERROR = !I5
13: SYMBOLIC_INPUT_REGENERATION2_TERMINATED = I20
14: opc=IOC_SYM_INPUT_READY_G_N = !I10
15: opc=IOC_SYM_INPUT_ALARM_G_N = !I13
16: opc=IOC_SYM_INPUT_PEAK_G = !I28
17: opc=IOC_SYM_INPUT_PEAK_N = !I30
18: opc=IOC_SYM_INPUT_FREQ_OK = IH
19: opc=IOC_SYM_INPUT_FU_ERROR = IL

Internal output configuration


1: LLO_MOTOR_ON = O3
2: LLO_MOTOR_REVERSE = !O4
3: LLO_FORWARD = O11
4: LLO_REVERSE = O12
5: LLO_WATCHDOG = O31
6: LLO_BUZZER_ON = O16
7: LLO_BAG_COUNTER = O13
8: LLO_ALARM_COUNTER = O17
9: LLO_GENERATOR1_ON1 = !O0
10: LLO_GENERATOR1_ON2 = O1
11: LLO_GENERATOR1_SOP_ON = O2
12: LLO_GENERATOR2_ON1 = !O5
13: LLO_GENERATOR2_ON2 = O6
14: LLO_GENERATOR2_SOP_ON = O7
15: LLO_LS_TEST_ON = !O10
16: SYMBOLIC_OUTPUT_REGENERATION1_ON = OX
17: SYMBOLIC_OUTPUT_REGENERATION2_ON = OX
18: IOC_SYM_OUTPUT_GAMMA_ON = O9

Internal input states


1: LLI_INTERLOCK_CLOSED = ON
2: LLI_XRAY_LAMP_ERROR = OFF
3: LLI_FOOT_SWITCH_CLOSED = ON
4: LLI_XRAY1_FEEDBACK_ACTIVE = ON
5: LLI_XRAY1_VOLTAGE_ACTIVE = ON
6: LLI_XRAY2_FEEDBACK_ACTIVE = ON
7: LLI_XRAY2_VOLTAGE_ACTIVE = ON
8: LLI_REGENERATION1_TERMINATED = ON
9: LLI_REGENERATION2_TERMINATED = ON
10: LLI_LS1_INTERRUPTED = OFF
11: LLI_LS3_INTERRUPTED = OFF
12: LLI_LS_ERROR = OFF
13: SYMBOLIC_INPUT_REGENERATION2_TERMINATED = ON
14: opc=IOC_SYM_INPUT_READY_G_N = OFF
15: opc=IOC_SYM_INPUT_ALARM_G_N = OFF
16: opc=IOC_SYM_INPUT_PEAK_G = OFF
17: opc=IOC_SYM_INPUT_PEAK_N = OFF
18: opc=IOC_SYM_INPUT_FREQ_OK = ON
19: opc=IOC_SYM_INPUT_FU_ERROR = OFF

Internal output states


1: LLO_MOTOR_ON = OFF
2: LLO_MOTOR_REVERSE = OFF
3: LLO_FORWARD = OFF
4: LLO_REVERSE = OFF
5: LLO_WATCHDOG = OFF
6: LLO_BUZZER_ON = OFF
7: LLO_BAG_COUNTER = OFF
8: LLO_ALARM_COUNTER = OFF
9: LLO_GENERATOR1_ON1 = OFF
10: LLO_GENERATOR1_ON2 = OFF
11: LLO_GENERATOR1_SOP_ON = OFF
12: LLO_GENERATOR2_ON1 = OFF
13: LLO_GENERATOR2_ON2 = OFF
14: LLO_GENERATOR2_SOP_ON = OFF
15: LLO_LS_TEST_ON = OFF
16: SYMBOLIC_OUTPUT_REGENERATION1_ON = UNKNOWN
17: SYMBOLIC_OUTPUT_REGENERATION2_ON = UNKNOWN
18: IOC_SYM_OUTPUT_GAMMA_ON = OFF

---------------------------------------------------------------------
TEMPERATURES 12/09/2019 21:16:48

Internal 0 C
External 0 C

---------------------------------------------------------------------
UPS 12/09/2019 21:16:48

UPS Installed NO

---------------------------------------------------------------------
RADIOACTIVITY DETECTOR SUBSYSTEM 12/09/2019 21:16:48

Installed NO

D I A G N O S T I C R E P O R T E N D
---------------------------------------------------------------------

You might also like