Professional Documents
Culture Documents
Trademarks
Affymetrix®, CytoScan®, and GeneChip® are registered trademarks of Affymetrix, Inc. All other trademarks are the property of their
respective owners.
Limited License
Affymetrix hereby grants to buyer a non-exclusive, non-transferable, non-sublicensable license to Affymetrix’ Core Product IP to use the
product(s), but only in accordance with the product labels, inserts, manuals and written instructions provided by Affymetrix. “Core Product
IP” is the intellectual property owned or controlled by Affymetrix as of the shipment date of a product that covers one or more features of
the product that are applicable in all applications of the product that are in accordance with the product labels, inserts, manuals and written
instructions provided by Affymetrix. The license granted herein to buyer to the Core Product IP expressly excludes any use that: (i) is not in
accordance with the product labels, inserts, manuals and written instructions provided by Affymetrix, (ii) requires a license to intellectual
property that covers one or more features of a product that are only applicable within particular fields of use or specific applications, (iii)
involves reverse engineering, disassembly, or unauthorized analysis of the product and/or its methods of use, or (iv) involves the re-use of a
consumable product. Buyer understands and agrees that except as expressly set forth, no right or license to any patent or other intellectual
property owned or controlled by Affymetrix is granted upon purchase of any product, whether by implication, estoppel or otherwise. In
particular, no right or license is conveyed or implied to use any product provided hereunder in combination with a product or service not
provided, licensed or specifically recommended by Affymetrix for such use. Furthermore, buyer understands and agrees that buyer is solely
responsible for determining whether buyer possesses all intellectual property rights that may be necessary for buyer’s specific use of the
product, including any rights from third parties.
Patents
Array Software: Products may be covered by one or more of the following patents: U.S. Patent Nos. 5,733,729; 5,795,716; 6,066,454;
6,090,555; 6,185,561; 6,188,783; 6,223,127; 6,228,593; 6,229,911; 6,308,170; 6,361,937; 6,420,108; 6,484,183; 6,505,125; 6,510,391; 6,532,462;
6,546,340; 6,567,540; 6,584,410; 6,611,767; 6,687,692; 6,826,296; 6,882,742; 6,957,149; 6,965,704; 6,996,475; 7,068,830; 7,130,458; 7,215,804;
7,424,368; 7,634,363; 7,822,555; 7,991,564; 7,992,098; 8,190,373; 8,498,825 and other U.S. and foreign patents.
Manufacturer
Affymetrix Inc.
3420 Central Expressway
Santa Clara, CA 95051 USA
Copyright
©2014-2015 Affymetrix, Inc. All rights reserved.
IVD
Contents
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Other Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Context-Sensitive Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Complete Help System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Terminology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Text Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
European Representative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
IVD
Contents 4
Chapter 5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Types of Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Instrument Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Test Request Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Operational Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Resolving Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Instrument Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Scanner Status Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
IVD
Contents 5
IVD
Chapter 1
Introduction
This User Guide gives detailed information for use of the Affymetrix® Molecular Diagnostic Software
(AMDS) and an overview of the operation of the GeneChip® System 3000Dx v.2 (GCS 3000Dx v.2).
In this chapter:
Indications for Use
Other Documentation
Context-Sensitive Help
Terminology
Text Alerts
Technical Support
European Representative
Other Documentation
GeneChip® System 3000Dx v.2 User Guide
Affymetrix® Molecular Diagnostic Software (AMDS) Quick Reference Card
Tooltip help1
Online Help2
Context-Sensitive Help
AMDS is the operating software for the GCS 3000Dx v.2. The AMDS online help system provides
content-sensitive help.
1. Click Help.
2. Use the question mark cursor to click on parts of the window.
The help topic associated with the selected area is shown in the lower half of the center panel and
includes information about the:
Workflow Pane
Administrator Pane
User Details Pane
Device Status Pane
Alerts and Alert Details Panes
1 To use this feature, place mouse cursor over an interface component to view its description or un-truncate truncated data.
2 To access, click:
IVD
Chapter 1 | Introduction 7
4. Use the navigation buttons to search for and select help topics.
Terminology
Test Request—Unique record of an assay performed on a specimen. Repeated runs of the same
specimen are represented by multiple test requests. Each test request is associated with a specific array
cartridge.
Assay Software Module (ASM)—The software package that runs an assay. ASMs are supplied by
Affymetrix or authorized partners in the form of a CD.
Assay—Procedure that processes the specimen to produce a result, includes bench work and running
through the GCS3000Dx v2 workflow. In this manual, assay refers to the ASM.
Array—A microarray is a glass substrate with oligonucleotide probes. A plastic cartridge surrounds and
protects the array and forms the array cartridge (see Figure 1.1). Array also typically refers to the array
cartridge.
Fluidics station or FS450Dx–GeneChip® Fluidics Station 450Dx v.2
Scanner–GeneChip® Scanner 3000Dx v.2 with AutoLoaderDx v.2
Text Alerts
Text alerts draw your attention to a particular piece of information. There are five types of text alerts:
Note, Important, Caution, Warning, and Danger.
NOTE: Information that may be of interest or of help to a user but is not critical to the primary
purpose of the text.
CAUTION: An alert to the user of hazards that, if not avoided, can cause minor or moderate
personal bodily injury and/or damage to an instrument or loss of data.
WARNING: Alerts the user to hazards that, if not avoided, can cause serious bodily injury or
death, or produce potentially incorrect data that could cause harm to a patient.
DANGER: Dangers alert the user to an imminent hazard that, if not avoided, will cause
serious bodily injury or death, or will produce a result that could cause serious harm to a
patient.
IVD
Chapter 1 | Introduction 8
Technical Support
Consult this user manual and online help and review the troubleshooting sections before you contact
Affymetrix technical support for unresolved issues. For complete contact information and specific
regional support contact information, go to www.affymetrix.com/browse/contactUs.jsp.
Affymetrix, Inc.
3420 Central Expressway
Santa Clara, CA 95051
USA
www.affymetrix.com
Email: support@affymetrix.com
Tel: 1-888-362-2447 (1-888-DNA-CHIP)
Fax: 1-408-731-5441
Affymetrix UK Ltd.
Voyager, Mercury Park,
Wycombe Lane, Wooburn Green,
High Wycombe HP10 0HH
United Kingdom Tel: +44 (0) 1628 552550
Fax: +44 (0) 1628 552585
France Tel. +33 08-00-91-95-05
Germany Tel. +49 01803-001334
Italy Tel. +39 800-91-59-74
Email: supporteurope@affymetrix.com
European Representative
Emergo Europe
Molenstraat 15
2513 BH, The Hague
The Netherlands
IVD
Chapter 1 | Introduction 9
Table 1.1
IVD
Chapter 2
System Overview
The GeneChip® System 3000Dx v.2 (GCS 3000Dx v.2) consists of hardware and software components
for processing microarrays based on assay parameters.The Affymetrix® Molecular Diagnostic Software
(AMDS) manages assay workflows.
In this chapter:
Hardware
Software
Hardware
The GCS 3000Dx v.2 comes with these standard hardware components:
System Workstation
Barcode Reader
Hardware Firewall
AMDS is loaded on the workstation (Figure 2.1). It provides the hardware interface to other system
components. Because the workstation is a dedicated system, it is locked down to prevent other uses. A
monitor and soundbar (not pictured) are part of the workstation system.
IVD
Chapter 2 | System Overview 11
The Fluidics Station (Figure 2.2) consists of four modules installed in a single housing. Each module
holds a single array and performs the washing and staining functions for that array. Up to 8 stations can
communicate with one system workstation.
Each module controls the addition of target and staining fluids to the array cartridge and array washing.
Each module contains a pump, valve, thermo-electric system, and LCD that are controlled by scripts
provided in the Assay Software Module (ASM). These scripts are automatically downloaded to each
module and stored in its electronic memory.
IVD
Chapter 2 | System Overview 12
The Scanner 3000Dx v.2 with AutoLoaderDx v.2 (Figure 2.3) is a wide-field, epifluorescent, confocal,
scanning laser microscope.It scans the chip after staining on the Fluidics Station. Array cartridges are
loaded into the scanner by the AutoLoaderDx v.2 (an automatic handler) before scanning and returned
to the AutoLoaderDx v.2 when scanning finishes.
Barcode Reader
The barcode reader (Figure 2.4) is used to enter barcode-encoded information from:
Array cartridge
Fluidics station
Specimen identification
IVD
Chapter 2 | System Overview 13
Hardware Firewall
The hardware firewall (Figure 2.5) protects against unauthorized access to the workstation. It is a
required component of the system.
Table 2.1
IVD
Chapter 2 | System Overview 14
Requirements for a hybridization oven that does not connect to AMDS are:
Table 2.2
Communications NA
Software
There are two types of software:
Affymetrix ® Molecular Diagnostic Software (AMDS)
AMDS
AMDS is the user interface for the system. It provides control of the instruments, assay workflow, and
the data collection and management processes. When array scanning finishes, data is passed through
AMDS to the ASM. Alternatively, AMDS transfers the data to a network location for downstream
analysis. AMDS also assists you in installing and deactivating an ASM.
AMDS tools:
Control assay workflows and show test results
ASM
ASMs provide the necessary parameters for processing an array through its entire workflow. ASMs
contain the algorithms and reporting functions to produce the result specific to that ASM. ASM user
manuals give detailed instructions that are unique to specific assays.
NOTE: For ASMs that allow auto-transfer of data, AMDS auto-transfers data (including DAT
and CEL files) to your network where it is available for downstream analysis. Your network
location must have ample storage and the correct access permissions for this auto-transfer
feature to work.
Center Panel–Gives information based on your selection in the left pane; also shows online help when
selected
Right Panel–User Details, Device Status, Alert, and Alert Details panes
IVD
Chapter 2 | System Overview 15
Center Panel
Workflow Pane
Buttons are:
Active Worklist
Registration
Hybridization Oven
Fluidics
Scanner
Non-Active Worklist
When you select an item, the associated worklist opens in the center panel. A blue arrow next to the
selected item shows that it has been selected.
Active Worklist
A test request is active until processing finishes or it is canceled. While a test request is active, you can
view its status by clicking Active Worklist. Data associated with a test request appear only on the Active
Worklist and the current workflow stage of the test request. Active Worklist enables you to:
View the current status of all active test requests
You cannot edit information in the Active Worklist. This worklist is maintained and updated by AMDS.
IVD
Chapter 2 | System Overview 16
Registration Worklist
The Registration Worklist is the next step in the workflow. This step allows you to associate a test request
with an array cartridge and adds the target preparation reagent kit lot numbers (if these are supported by
the ASM).
Fluidics Worklist
The Fluidics step allows you to associate a test request with a fluidics station and module and runs the
associated fluidics script. The selected station and module wash and stain the array.
Scanner Worklist
The Scanner selection opens the Scanner Worklist. AMDS automates control of the scanner and uses
the assay information provided by the ASM to manage the scanner settings for each type of array. After
scanning, AMDS automatically grids the image, computes intensities, and either runs an ASM-specific
analysis algorithm to generate a result or transfers the files to data storage.
Non-Active Worklist
The Non-Active Worklist selection shows non-active test requests. The system moves test requests to
the Non-Active Worklist when processing in AMDS finishes.
Administrator Pane
The Administrator Pane (Figure 2.7) is below the Workflow Pane. It is visible to everyone, but only
Laboratory Supervisors and System Maintainers use it.
IVD
Chapter 2 | System Overview 17
Change passwords
This pane is an instrument status dashboard, providing color-coded status for each system device.
Green–Device is operating normally
AMDS provides status for the fluidics station, scanner, and printer, and for hybridization oven, only if
connected.
IVD
Chapter 2 | System Overview 18
When you click on an alert in this pane, information about the alert is shown in the Alert Details pane
(Figure 2.11).
Assay Name
Workflow Step
Date/Time
Alert Details
Recommended resolution
IVD
Chapter 3
System Operation Basics
In this chapter:
Start and Shut Down the System
Change Password
E-Signatures
When AMDS completes the self-test, the AMDS User Logon window opens.
IVD
Chapter 3 | System Operation Basics 20
AMDS authenticates your credentials and provides functionality based on your user role.
IMPORTANT: To preserve the lifetime of the scanner laser, turn off the scanner when it is not
in use for an extended period of time, such as overnight or a weekend.
Log Off
Logging off does not terminate the active workflow processes. If there are test requests in hybridization,
fluidics or scanning, AMDS continues to process them.
1. Go to User Details pane (Figure 3.2)
IVD
Chapter 3 | System Operation Basics 21
Shut Down
1. Go to User Details pane.
2. Click Shut Down.
3. Turn off instrumentation.
IMPORTANT: If there are test requests in progress, a confirmation dialog box opens. Click
Yes to initiate an orderly shutdown of the system. All tests requests that are “In Progress”
go into an error state and must be re-run.
Change Password
1. Go to User Details pane.
2. Click Change Password.
3. In the Change Password dialog box, enter your old password (Figure 3.3).
IVD
Chapter 3 | System Operation Basics 22
or
Click in field in last row and scan barcode
or
Click in field in the last row, release the Ctrl key, and enter value
or
Click in field in last row and scan barcode
or
Click in field in last row, release Ctrl key, and enter value
IVD
Chapter 3 | System Operation Basics 23
Front
Probes on a
glass substrate
Back
Although the inner glass surface is protected, any contamination or scratches on the outer surface of the
glass can compromise the integrity of the scan.
IMPORTANT: Do not touch the surface of the glass with your fingers. Skin oils and other
substances, such as lotions or ink, can fluoresce. If the array surface is dirty, carefully clean it
with a nonabrasive laboratory tissue
E-Signatures
E-signatures help track user activity on the system. At times after completing procedures in AMDS, you
are prompted to confirm a change with your password and the reason for the change (Figure 3.5).
IVD
Chapter 4
Running Test Requests
Introduction
A test request is a unique record of an assay performed on a particular specimen. Repeated runs of the
same specimen require multiple test requests. Each test request is associated with a unique array cartridge.
In this chapter:
Working with Test Requests
Oven Hybridization
Data Processing
Non-Active Worklist
Active Worklist
The Active Worklist contains all active test requests, regardless of their step in the workflow. After all
steps of the workflow are complete, you can approve or reject test request results from the Active
Worklist. Once approved, rejected, or canceled, test requests are shown in the Non-Active Worklist.
IVD
Chapter 4 | Running Test Requests 25
NOTE: You can create more than one test request at a time.
3. In Enter Test Request window, enter the Specimen ID barcode using the barcode reader (Figure 4.3).
NOTE: If your specimen does not have a barcode, manually enter the name of the
Specimen ID in the Specimen ID field. Specimen IDs need not be unique and can have up
to 40 characters.
4. Under Assay Name, click a down-arrow and select an Assay Name from the drop-down list.
5. Click Submit.
Note that:
You can submit one or more test requests at the same time after all Specimen IDs and Assay Names
are entered.
When you click Submit your test request is created and becomes active.
The Enter Test Request window always closes after you click either Submit or Delete.
You can find your newly created test request in the Active Worklist based on the Specimen ID, Assay
Name, and current status.
IVD
Chapter 4 | Running Test Requests 26
NOTE: If you delete your test request before you click Submit, the test request is not
created.
3. Edit the Specimen ID–this is the only field you can edit.
4. Click anywhere on the tool bar or Specimen ID row.
5. Click Save Changes.
6. In the Confirm Step dialog box (Figure 4.6), enter:
Your password
IVD
Chapter 4 | Running Test Requests 27
7. Click OK.
The edited Specimen ID is now shown on the Active Worklist.
Undo Changes
1. To undo your changes, click Discard Changes (Figure 4.7).
IMPORTANT: Batch editing is possible only with test requests that use the same Assay
Software Module (ASM). Batch Edit works only when no arrays associated with the selected
test requests have started scanning.
You can use Batch Edit to associate information with multiple test request records.
Batch Edit
1. In any worklist, select one or more test requests associated with the same ASM.
2. Click Batch Edit.
3. In the Assay Batch Information dialog box, proceed according to your ASM user manual.
IVD
Chapter 4 | Running Test Requests 28
Add Information
1. In any worklist, click on a Specimen ID hyperlink.
2. In the Additional Information dialog box, proceed according to your ASM user manual.
3. Enter required information in Confirm Step dialog box and click OK.
Affymetrix® Molecular Diagnostic Software (AMDS) moves the test request to the Non-Active
Worklist.
NOTE: A test request “In Progress” on either the Fluidics or Scanner Worklist cannot be
canceled by the Cancel Test Request button.
1. Click Shut Down to restart your workstation. This action forces all In Progress test requests into an
error state, and stops all processing.
2. In Active Worklist, select a test request.
3. Click Cancel Test Request.
4. Enter required information in Confirm Step dialog box and click OK.
AMDS moves the test request to the Non-Active Worklist.
IVD
Chapter 4 | Running Test Requests 29
NOTE: Not all assays require a Target Preparation Reagent Lot ID. For some assays, you enter
reagent data using the Reagent Information tab of the Assay Home window. Refer to your
ASM user manual for details.
2. In the Registration Worklist, scan the Specimen ID barcode (Figure 4.10). AMDS selects the test
request associated with the Specimen ID.
You can also manually select the Specimen ID from the worklist.
3. Scan the array barcode from the array cartridge. AMDS populates the Array ID field.
You can also enter this information manually.
NOTE: An AMDS alert opens if this Array ID is already in use, if the array is expired, or if
the array type is not compatible with the assay type.
IVD
Chapter 4 | Running Test Requests 30
This step completes the registration and moves the test requests to the next step in the workflow,
which depends on the type of assay you are using. Refer to your ASM user manual for details.
NOTE: Some assays require additional information for test requests. The required
information must be entered before the test request moves off the Registration Worklist.
Refer to your ASM user manual for details.
Troubleshooting Registration
You must correct the data entry if you get a message that:
The Array ID already exists
Oven Hybridization
IMPORTANT: Make sure you understand how to operate your hybridization oven.
NOTE: You cannot change any parameter in the Time/Rotation/Duration field in the
Hybridization Oven Worklist. These are the optimal hybridization parameters for the assay.
IVD
Chapter 4 | Running Test Requests 31
IMPORTANT: Follow the Step 2 only if you are using an oven tray.
2. In Hybridization Oven Worklist (Figure 4.12), associate test requests with an oven tray.
3. Select all test requests for the arrays you place into the oven.
Select one test request for a tray and click Select Tray
or
Manually select test requests
or
For an unconnected oven, enter E
5. Click outside field to populate oven number for all selected Specimen IDs.
IVD
Chapter 4 | Running Test Requests 32
NOTE: The background color of the Reported Elapsed Hybridization Time cell turns yellow
when the time is less than the minimum time required by the assay parameters, green
when it falls within the acceptable range, and red when it exceeds the maximum
allowable time.
End Hybridization
1. Upon completion, remove arrays from oven.
2. Select test requests associated with the arrays.
3. Click End.
4. Click Complete Step.
The test requests move to the Fluidics Worklist.
Hybridization Errors
AMDS checks your data as you enter the tray and oven information.
You must correct the issue if you get an error message that arrays are not compatible for:
Inclusion on same tray
IVD
Chapter 4 | Running Test Requests 33
Preparation
Set Up Fluidics Station
1. Click Fluidics (Figure 4.13).
IVD
Chapter 4 | Running Test Requests 34
3. Under Assay Name, select an assay for each fluidics stations you will use.
4. Enter Wash Buffer A and Wash buffer B lot numbers (optional).
5. Under Station #, select fluidics station(s).
6. Click Prime Fluidics Station and follow fluidics station LCD instructions.
IMPORTANT: Prime fluidics station when your first start it, change wash buffers, after a
shutdown, or if resolution to an alert requires priming.
3 All modules are checked by default. Checked modules are used if the station is used; unchecked modules are not used if the
station is used.
IVD
Chapter 4 | Running Test Requests 35
NOTE: If you are done processing arrays for this assay or will not use fluidics station for
several hours, shut down the fluidics station.
2. Load array cartridges into AutoLoaderDx v.2–start with slot 1 and load contiguously.
3. Click Start.
4. When done, click Complete Step.
IVD
Chapter 4 | Running Test Requests 36
Figure 4.19
Figure 4.20
IVD
Chapter 4 | Running Test Requests 37
Figure 4.21
Communication is restored.
Figure 4.22
NOTE: If the scanner is powered on when you follow this procedure, the appropriate status
shows in the Device Pane. If the scanner is powered off, status is not updated until the
scanner is powered on.
Manual Mode
NOTE: You can scan one array at a time using the manual scan function. Manually scan arrays
that have illegible or missing barcode labels. AMDS attempts to scan the barcode and validate
that the selected test request is correct.
Scan Arrays
1. Click Scanner.
2. Open AutoLoaderDx v.2 door.
3. Insert array into AutoLoaderDx v.2 carousel at slot 1.
4. Close door.
5. Select correct test request on Scanner Worklist.
6. Click Manual Scan (Figure 4.23).
IVD
Chapter 4 | Running Test Requests 38
or
Remove array(s)
NOTE: Scanner does not rescan any previously scanned arrays and always starts the run
with slot 1.
4 To add arrays, remove scanned arrays and consolidate unscanned arrays. Start with slot 1 and load contiguously.
IVD
Chapter 4 | Running Test Requests 39
Data Processing
AMDS processes the scanned data by:
Taking resulting DAT file
Applying a grid
Review Results
1. Click Active Worklist.
2. Click Awaiting Review (Figure 4.24).
A Results window for your particular assay opens.
Non-Active Worklist
View test requests on the Non-Active Worklist:
When test requests were canceled before processing finished
After processing finished but before test requests are archived and purged
Once you archive and purge a test request, you can only view it from CD/DVD or the network location
of the archive.To view non-active test requests, click Non-Active Worklist (Figure 4.25).
IVD
Chapter 4 | Running Test Requests 40
IVD
Chapter 5
Troubleshooting
In this chapter:
Types of Errors
Resolving Alerts
Instrument Troubleshooting
Types of Errors
There are three general types of errors that may occur:
Instrument–For example, communication issues with the hybridization oven, problems with a fluidics
station or module, or problems with the scanner (also refer to the GeneChip® System 3000Dx v.2 User
Guide)
Test Request–For example, the barcode of the array associated with a test request cannot be read, or a
test request gridding failure
Operational–These include data input errors
Affymetrix® Molecular Diagnostic Software (AMDS) tracks and reports on all three error types.
Instrument Errors
If you suspect an instrument issue, view the Device Status pane (Figure 5.1).
Check the device status column to see if there are instrument errors.
NOTE:
Occasionally an instrument issue occurs and no alert is evident in AMDS. One example is a
leaking fluidics module. If you are not sure how to handle an issue and there is no alert with
a recommended resolution, turn off the instrument and contact Affymetrix support.
In some cases you may not be able to resolve an instrument problem. You are directed to
contact Affymetrix support to resolve the problem. Do not attempt to correct this kind of
problem yourself.
IVD
Chapter 5 | Troubleshooting 42
Click on the alert in the Alerts Pane to view the corresponding alert details on the Alert Details pane.
The Alerts Details pane (Figure 5.3) gives a full description and suggestion for corrective action.
Autofocus error
Gridding error
IVD
Chapter 5 | Troubleshooting 43
Operational Errors
Operational are relatively easy to detect and correct because AMDS gives immediate feedback in order
to prevent incorrect operation and/or data input. Some examples are:
Invalid array barcode entry
Array loaded onto fluidics station configured for a different assay type
Resolving Alerts
IMPORTANT: Resolve alerts as soon as they occur. This practice makes it easier to decipher
new alerts as they appear.
NOTE: If the alert concerns a Specimen ID, click Go to test request to select the row in the
worklist that contains the related test request.
3. Click Resolve.
4. At the prompt, enter your password and additional information.
5. Click OK to remove the alert from the list.
NOTE: If after completing the recommended resolution procedure you are unable to
resolve the error, call Affymetrix support.
IVD
Chapter 5 | Troubleshooting 44
Instrument Troubleshooting
Table 5.1
Problem/Error Resolution
Missing fluid Possible cause: Insufficient quantities of fluids or the wrong fluids are provided.
Determine the source of the problems on the fluidics station LCD
Follow the resolution procedure in the alert details
Clear the alert, and continue the fluidics run
If you continue to see this error, it may indicate a system failure. Call Affymetrix technical support.
Autofocus Possible cause: Array cartridge glass substrate is smudged or dirty, usually from a thumb print.
Clean the array
Clear the alert
Retry
Monitor/User Interface
Entire AMDS screen not visible; Set back screen resolution to the default using the correct key combination for your style monitor:
cannot interact with complete User Wide-screen Monitor–Ctrl+Alt+W
NOTE: These two keystroke combinations have no effect if used on the wrong monitor (e.g., if
Ctrl+Alt+W is used on the 4:3 monitor).
IVD
Chapter 5 | Troubleshooting 45
Figure 5.5
Figure 5.6
Figure 5.7
Communication is restored.
Figure 5.8
IVD
Chapter 5 | Troubleshooting 46
NOTE: If the scanner is powered on when you follow this procedure, the appropriate status
shows in the Device Pane. If the scanner is powered off, status is not updated until the
scanner is powered on.
IVD
Appendix A
AMDS Administrative Features
This Appendix gives the Administrative features in Affymetrix® Molecular Diagnostic Software
(AMDS).
In this appendix:
View Logs
Manage Assays
Manage Users
Access administrative features in the Administrator pane (Figure A.1). User privileges are defined by user
role–if a button is disabled, you do not have access to that feature.
View Logs
System Logs
System logs contain system-level algorithm and computer messages, as well as recorded parameters of
completed assays, such as date and time, component, type, user, etc. To view the current system log, click
View Logs in the Administrator pane (Figure A.1). The system log is shown in the top position.
To view a system log other than the current one, select one from the System Log list.
Component
Type
User
IVD
Appendix A | AMDS Administrative Features 48
Audit Logs
Audit logs contain information, messages, and parameters for a test request for each step of the workflow.
The Audit Log is found below the System Log in the View Logs window (Figure A.3).
IVD
Appendix A | AMDS Administrative Features 49
4. Click Display.
Workflow step
User
While viewing either a system or audit log, click Export to send the entire log file to the export queue.
The View Logs toolbar shows the total size and number of logs in the queue.
IVD
Appendix A | AMDS Administrative Features 50
Manage Assays
Assay Management enables you to install, repair, or deactivate assays or install SSL certificates.
Install an Assay
Typically, an Affymetrix field service engineer installs assays and certificates. If an assay communicates
with a remote server and the certificate for that server has not been installed in AMDS, you must install
the server certificate first (refer to your ASM user manual to determine if you need to install the
certificate). To install an assay, you must have the assay installation bundle, which is typically provided
on a CD.
Install Assay
1. Click Assay Management (Figure A.6).
2. Place CD with ASM installation bundle into the workstation CD/DVD drive.
3. Click Install Assay.
IVD
Appendix A | AMDS Administrative Features 51
4. In the Assay Installation dialog box (Figure A.8), go to Select the Manifest file and enter or browse
to the manifest file (part of the assay installation bundle).
Deactivate an Assay
You can deactivate an assay if you no longer use this assay and do not want it to appear as a selection in
the AMDS application. There is no uninstall mechanism for an assay.
1. In the Assay Management window, select an assay.
2. Click Deactivate (Figure A.5).
Once you deactivate an assay, you can still complete the workflow for any active test requests for this
assay. However, the assay is no longer available for entering new test requests.
IVD
Appendix A | AMDS Administrative Features 52
Install a Certificate
1. In the Administrator pane, click Assay Management (Figure A.9).
or
Install from local disk drive
4. Navigate to and select a .cer file (the certificate has a .cer extension).
5. Click OK1.
If it is a:
Successful installation–The message “You have successfully added the server certificate ‘x.cer’ to the
AMDS trusted certificate store” opens
IVD
Appendix A | AMDS Administrative Features 53
or
Failed installation–AMDS notifies you with an algorithm message. You are not permitted to access
the server and transfer data.
6. Click OK.
Manage Users
Add a User
Before you add a user, review the AMDS user roles.
If there are no assays installed on the system, you do not have the option of granting assay access to a user.
Add User
1. In the Administrator pane click User Management (Figure A.12).
IVD
Appendix A | AMDS Administrative Features 54
3. In the Add User window, establish new user characteristics (Figure A.14).
Active–Indicates whether or not the user account being created at this time is active
Enter a User ID
Create a Password
Confirm Password
IMPORTANT: Password must comply with the local security policy on the AMDS
workstation. This policy is determined by your IT department.
Under Select Assay Group(s), select assay groups for the user.
Edit a User
You can change any user attribute except User ID.
1. In the User Management window, select a user.
2. Click Edit User.
IVD
Appendix A | AMDS Administrative Features 55
Deactivate a User
To deactivate a user, got to the Edit User window and deselect Active. The account can be reactivated.
Change Windows
1. In Administrator pane, click System Management (Figure A.16).
IVD
Appendix A | AMDS Administrative Features 56
User Performed Service–View system configuration (i.e., device and instrument settings) and manage
Fluidics Station 450Dx v.2, workstation, Scanner 3000Dx v.2, printer, and firewall
Service Logs–View logs of user-performed services
Archive Test Request–Archive and purge test request data.
General Configuration–Change time and time zone settings for your location and configuration flags
for your instruments, and set up network archive location
Manage printer
Manage firewall
5. In Confirm Step dialog box, enter your password, then click OK.
IVD
Appendix A | AMDS Administrative Features 57
6. Give Your ID and Password from this window to the service technician. This enables the Service
Technician to remotely connect to your system to diagnose or resolve your issue.
The service technician connects remotely. The Session list window (Figure A.20) remains on your
computer window while the remote session is in progress.
IVD
Appendix A | AMDS Administrative Features 58
2. In the Stating Support Log Files dialog box (Figure A.22), enter a brief message and/or a text string
to append to the file name.
NOTE: Before(s) proceeding, go to Station and select a fluidics station from the drop-down
list.
IVD
Appendix A | AMDS Administrative Features 59
Check elapsed time–Under Time Elapsed Since, view days elapsed since last bleach or tube change
Bleach protocol–under Procedure, select Bleach, click Perform, then follow bleach procedure.
Change tubing–Under Procedure, select Change Tubing and follow change tubing procedure (refer to
the GeneChip® System 3000Dx v.2 User Guide), then click Perform.
Troubleshoot or test communication–Under Procedure, select Home, then click Perform.
View or change Bleach Interval Warning Limit and Bleach Interval Alert Limit–Go to Configuration to
set limits, then click Save. Note:
Beach reminder schedule is preconfigured by Affymetrix
View or change Change Tubing Warning Limit and Change Tubing Interval Alert Limit–Go to
Configuration to set limits, then click Save. Note:
The change tubing reminder schedule is preconfigured by Affymetrix
Figure A.24 The System Region of the User Performed Services Screen
IMPORTANT: Alerts warn when workstation hard drive nears capacity. Archive and purge
data before system reaches limit. At alert limit, AMDS does not scan arrays and you must
archive and/or purge data to stay below limit.
Restart Scanner–Click if you are experiencing problems connecting to the Scanner 3000Dx v.2
Check Network Archive Storage Free Space–Click to view available space on Network Archive location
(if applicable)
IVD
Appendix A | AMDS Administrative Features 60
Manage Printer
Perform the following functions under Printer in the User Performed Services window (Figure A.25)
Figure A.25 The Printer region of the User Performed Services screen
NOTE: You cannot change the printer. An Affymetrix field service engineer sets up the printer
as part of the AMDS installation process.
Firewall Settings
The AMDS firewall monitors incoming and outgoing messages from the workstation and protects the
workstation from malicious intrusion.
Perform the following functions under Firewall in the User Performed Services window (Figure A.26).
NOTE: Some assays access servers on an intranet or on the internet. When these assay
software modules are installed, a rule is written to the firewall that permits communication
between the AMDS workstation and the remote server. Restoring firewall settings may
overwrite these rules and prevent AMDS from communicating with the server. In this case,
reinstall any ASMs that communicate with remote servers in order to restore communication
between the workstation and server.
IVD
Appendix A | AMDS Administrative Features 61
A User Performed Service Log entry is created for most of the actions you perform on the User Performed
Services window. You can open and view these service logs.
8. In System Management menu, select User Performed Service Logs (Figure A.27).
Date/Time
Procedure
User
Device
Description
IMPORTANT: Once test requests are purged they cannot be restored to AMDS. They are
available only on the CD/DVD containing the archive, any print outs, or similar material.
IVD
Appendix A | AMDS Administrative Features 62
1. In the System Management menu, select Archive Test Requests (Figure A.29).
Note that:
If a test request has a green checkmark in the Archived DVD or Network column, it has been
successfully archived
The System Management tool bar gives the total number of tests that are ready to be archived
2. Click:
AMDS Index to view test requests in AMDS that are ready to be archived
or
DVD Index to view test requests from an archive CD/DVD
Assay name
Array ID
or
Network Archive to archive to a remote drive
NOTE: The network drive must be configured for this feature to be available.
IVD
Appendix A | AMDS Administrative Features 63
You can purge a test request record only after it has been archived. Once test requests have been purged,
they are no longer available on the AMDS system.
IVD
Appendix A | AMDS Administrative Features 64
IVD
Appendix A | AMDS Administrative Features 65
Generate Report
1. In the Administrator pane, click System Management.
2. Under In-House, select General Configuration (Figure A.33).
IVD
Appendix A | AMDS Administrative Features 66
IVD
Appendix A | AMDS Administrative Features 67
AMDS Version
Operating System
Workstation name
Fluidics Station 450Dx v.2 and Module Serial Numbers (if present)
The printed version of this report also contains an Affymetrix Technician Service history log.
To print the report, click Print. If the Print button is not enabled (grayed out), contact your Affymetrix
Service Technician to connect a printer to your system.
To export the report, click Export. If the Export button is not enabled (grayed out), see Configure Network
Archive Storage on page 69. Note that the Export button is enabled only after an archive storage location
is established.
IMPORTANT: Do not change the local time or time zone while test requests are in progress
(on any worklist).
Configure
1. In the Time Zone Editable Settings change settings.
A. In Editable Local time, set local time.
B. In Time Zone, set local time zone.
2. Click Save Changes.
IVD
Appendix A | AMDS Administrative Features 68
3. In the General Configuration window, go to the System Configuration Flags section (Figure A.37).
4. Select Turn on Laser to turn on the Scanner 3000Dx v.2 laser automatically when you launch the
AMDS application.
5. Select Disable AutoLoader to disable the Scanner 3000Dx v.2 and manually load/unload arrays from
the Scanner 3000Dx v.2 (refer to Chapter 5 of the GeneChip® System 3000Dx v.2 User Guide).
IMPORTANT: This should be used only if the AutoLoaderDx v.2 has failed and there are
arrays that must be scanned immediately.
NOTE: This section is applicable only to ASMs that require report approval. This section is not
applicable to users of Affymetrix ASMs because they do not require report approvals.
Goto the System Configuration Flags window and select Require User Approval to require the technician
to approve or reject the analysis results as they become available after an assay run (Figure A.37). This
flag is meaningful only if you do not want to approve or reject the results on the AMDS system.
Deselecting this option causes the test requests to immediately move to the Non-Active Worklist upon
completion of the analysis algorithm. You cannot view the results through the Active Worklist.
NOTE: This option is intended for use with a Laboratory Information System (LIS). If you select
this option, you must review the results for each test request. If the box is not selected, the
test requests disappear from the Active Worklist upon completion of the workflow and
reappear on the Non-Active Worklist without user intervention.
IVD
Appendix A | AMDS Administrative Features 69
User Name
Password
3. Click Test Connection to configure the firewall and write a temporary file to share.
4. Check Enable archiving to network.
NOTE: If the test fails, an error message provides assistance troubleshooting the problem.
5. Check Use MD5 checksums to verify transfers to get read-feedback verification of archived files.
6. In the toolbar, lick Save Changes.
IVD
Index
A L
Administrative features 47 Logging off 20
Alerts Logs
resolving 43 viewing 47
AMDS
quick reference guide 19 M
Array Molecular Diagnostic System
cartridge 23 administrative features 47
handling 23
scanning 35
N
Assay
deactivating 51 Normal operational scanning mode 35
installation 50
Assay management panel O
selecting assays 51 Operational and data input errors
resolving 43
B
Barcode reader 21 P
Bleach interval alert limit 59 Peristaltic tubing change 59
Bleach interval warning 59 Printer management 60
Probe array handling 23
C Processing test requests 30
Cancelling
test requests 28 R
Configuration panel 59 Resolving
Creating alerts 43
test request 24
S
D Scanner Status Error 36, 44
Deactivate assay 51 Scanning
normal operational mode 35
E Shutting down 21
Edit user 55 Specimen ID
Error, adding additional information 28
operational and data input 43 Stopping a scan 38
resolving device related 41 System
Scanner Status 36, 44 logoff 20
test requests 42 shutting down 21
F T
Fluidics station Technical support 8
using 34
Test requests
additional information 28
H cancelling 27
Hybridizing 30 creating 24
deleting 27
editing 26
I processing 24, 30
In 51 registration 28
Indications for Use 6 resolving errors 42
washing and staining 33
Install assay 50
Troubleshooting 41
Instrument related errors
instruments 44
resolving 41
Instrument troubleshooting 44
IVD
Index 71
U
User management
add user 53
edit user 55
W
Washing and staining 33
IVD