The OIE ASF Reference Laboratory Network's
overview of African swine fever diagnostic
tests for field application
Main authors:
Ken Inui(1), Carmina Gallardo(2), Raquel Portugal(3), Linda Dixon(3), Carrie Baton(3) & David Williams(4)
Contributors:
Zhiliang Wang(5), Livio Heath(6) & Jose Manuel Sanchez-Vizcaino(7)
(1)
Food and Agriculture Organization of the United Nations (FAO), Department of Animal Health (DAH), Ministry of
Agriculture and Rural Development (MARD), Hanoi, Vietnam
(2)
Centro de Investigación en Sanidad Animal, CISA, INIA-CSIC, European Union Reference Laboratory for African Swine
Fever (EURL), Valdeolmos, Madrid, Spain
(3)
OIE ASF Reference Laboratory, The Pirbright Institute, Ash Road, Pirbright, Woking, Surrey,
United Kingdom
(4)
IE ASF Reference Laboratory, CSIRO, Australian Centre for Disease Preparedness, Geelong,
O
Victoria, Australia
(5)
IE ASF Reference Laboratory, National Surveillance and Research Center for Exotic Animal Diseases, China Animal
O
Health and Epidemiology Center, Qingdao, People’s Republic of China
(6)
IE ASF Reference Laboratory, Onderstepoort Veterinary Institute, Agricultural Research Council, Onderstepoort, South
O
Africa
(7)
IE ASF Reference Laboratory, Centro de Vigilancia Sanitaria Veterinaria (VISAVET),
O
Universidad Complutense de Madrid, Madrid, Spain
Cover photos:
On-site detection of pathogens in Vietnam / Ken Inui
Pius Clement of the National Agriculture Quarantine & Inspection Authority (NAQIA), conducting an inspection in Papua
New Guinea / David Williams
Disclaimer:
This document summarizes the current knowledge of the OIE ASF Reference Laboratory Network on commercially available Point of
Care (PoC) tests. The mention of specific companies or products of manufacturers, whether or not these have been patented, does not
imply that these have been endorsed or recommended by the OIE in preference to others of a similar nature that are not mentioned.
All commercial kits should be validated according to the OIE international standards. All commercial kits included in the OIE Register
are certified by the OIE as validated and fit for purpose. The PoC tests included in this document are not in the OIE Register nor are
they described in the OIE Manual of Diagnostic Tests and Vaccines for Terrestrial Animals. The OIE Register can be consulted at:
https://www.oie.int/en/what-we-offer/veterinary-products/#ui-id-5
2
The OIE ASF Reference Laboratory Network's
overview of African swine fever
diagnostic tests for
field application
INTRODUCTION
A
frican swine fever (ASF) cannot be Despite not being included in the Register of
differentiated from other febrile Diagnostic Kits certified by the OIE as validated
haemorrhagic syndromes or bacterial as fit for purpose, there are several diagnostic
septicaemias of pigs by either clinical platforms, also known as penside or point-of-need/
or post-mortem examination. Laboratory tests are, point-of-care testing (PoC tests), that are available
therefore, essential for the diagnosis of ASF and are commercially for field testing. These include basic
key to the success of ASF surveillance activities. rapid test kits for detecting antigens or antibodies
using lateral flow devices that are simple to use,
Chapter 3.9.1. of the OIE Manual of Diagnostic require minimal training and can provide a result
Tests and Vaccines for Terrestrial Animals (Terrestrial within approximately 20 minutes.
Manual) describes the recognised international
standards for ASF diagnosis. However, in certain
circumstances, the timely submission, processing
and testing of samples using the diagnostic tests
described in the Terrestrial Manual are not feasible.
3
Rapid antibody tests generally have comparable
levels of sensitivity and specificity to laboratory
enzyme-linked immunosorbent assays (ELISAs), and
lower sensitivity compared with reference tests such
as the immunoperoxidase monolayer assay. These
tests can be used to detect antibodies in pigs that
have survived infection or have survived long enough
to seroconvert.
4
The OIE ASF Reference Laboratory Network's
overview of African swine fever
diagnostic tests for
field application
Table 1. Comparison of four major test platforms for ASF virus detection
* Some tests are designed for use with certain sample types; limited evaluation of sample types for some platforms has been reported.
5
Table 2. Comparison of four major PoC test methods for rapid ASF virus antigen detection
Shenzhen Lvshiyuan
Test Ingenasa Bionote PenCheckTM
Biotechnology Co.
INgezim ASF CROM Ag Anigen ASFV Ag Rapid Rapid Screening Test SLB ASF Antigen
Catalogue no.
(11.ASFV.K.42) Test (RG1407DD) for ASFV (PC-888) Detection RDT
ingenasa.eurofins-
Website www.bionote.co.kr www.penchecktest.com/ lsybt.com/En
technologies.com/home/
5.80 to 10.45
Cost/test (US$) (depending on pack 13.90 2.50 3.50
size)
6
The OIE ASF Reference Laboratory Network's
overview of African swine fever
diagnostic tests for
field application
Table 2 cont’d. Comparison of four major PoC test methods for rapid ASF virus antigen
detection
Shenzhen Lvshiyuan
Test Ingenasa Bionote PenCheckTM
Biotechnology Co.
Rapid (early detection Rapid (early detection Rapid (early detection Rapid (early detection
at PoC) at PoC) at PoC) at PoC)
Easy (anyone can Easy (anyone can Minimal training (e.g. Easy (anyone can
perform) perform) pipette use) perform)
Moderate to high
High specificity
specificity
Evaluated at CISA-
Evaluated at ACDP Evaluated at ACDP
INIA, ACDP, NVLD and
using blood from using blood from
Pirbright; analytical
experimentally infected experimentally infected
sensitivity between 6-7
pigs: 68% of PCR* pigs: 27% of PCR* Peer-reviewed published
Comments log10 TCID50 ASFV in
positive samples were positive samples were journal article
spiked blood and 7.75
positive; 90% of PCR positive; 92% of PCR
log10 TCID50 in blood
negative samples were negative samples were
from experimentally
negative negative
infected pigs (Pirbright)
Peer-reviewed Peer-reviewed
Matsumoto et al.
References Sastre et al. (2016a) publication not yet publication not yet
(2020)
available available
*In-house ASF PCR described by Zsak et al. (2005) used for ACDP for evaluation
7
Table 3. Comparison of three major PoC test methods for rapid ASF virus antibody detection
INGEZIM ASFV-CSFV CROM Ab INGEZIM PPA CROM (11.PPA. GDX70-2 Herdscreen® ASF
Catalogue no.
(11.SFV.K41) K41/25) Antibody
ingenasa.eurofins-technologies. ingenasa.eurofins-technologies.
Website globaldx.com
com/home com/home
Specimen type(s) Whole blood, serum Whole blood, serum, plasma Whole blood, serum, plasma
8
The OIE ASF Reference Laboratory Network's
overview of African swine fever
diagnostic tests for
field application
Table 3 cont’d. Comparison of three major PoC test methods for rapid ASF virus antibody
detection
Rapid (early detection at PoC) Rapid (early detection at PoC) Rapid (early detection at PoC)
Easy (anyone can perform) Easy (anyone can perform) Easy (anyone can perform)
Workers in a backyard pig farm in the Philippines draw samples for testing / Bureau of Animal Industry (BAI), Philippines
9
Table 4. Comparison of Point-of-Care PCR systems for rapid ASF virus DNA detection
Test methods
iiPCR+DNA
System
iiPCR extraction LAMP qPCR qPCR qPCR
(fully automatic)
Manufacturer GeneReach GeneReach OptiGene Tetracore Indical Genesig
www. www. www.
www.indical. www.genesig.
Website genereach. genereach. optigene. tetracore.com
com com/home
com com co.uk
POCKIT
Micro Duo Indifield Genesig q16
Instrument Nucleic POCKIT Central Genie III T-CORE8 portable PCR qPCR
Acid system instrument
Analyzer
IF-
Catalogue code apmd apcc Gen3-01 T-CORE8 Z-genesig-q16
IN6010093
No. of wells 4 8 8 8 (independent) 9 16
PoC
Testing time 45 min 45 min 30 min 60 min 30-60 min 60 min
instrument
Battery Battery /100- Battery /100-
Power source Battery 100-240 V 100-240 V
/100-240 V 240 V 240 V
Detection
2 2 2 up to 6 2 2
colors
Included in Included in standard standard 200
PCR tube Manufacturer's Manufacturer's
reagent kit reagent kit 200 ul ul
Weight (kg) 0.43 21 1.75 4.5 1.2 2
Cost (US$) 3,000 30,000 18,000 20,000 9,000 9,000
POCKIT Virotype
ASFV ASFV pre-mix ASF w/IC (96 ASFV PCR
Manufacturer's
reagent cartridge; No reactions) Kit /IndiField Z-Path-ASFV
ASF kit
set; lyophilised Wet assay ASFV PCR;
Nucleic acid lyophilised lyophilised
test reagents In-house* No No Yes Yes Yes No
Commercial
No No Yes Yes Yes No
kits**
Cost (US$) / 15 (includes
8 4 to 23 5 to 15 5 to 15 5 to 15
test DNA extraction)
taco Mini
(8 wells);
Included in
Instrument battery- No No No No
PCR
operated;
US$ 6000
Pre-loaded
taco
nucleic (MagMAX™-96 M1 sample Genesig easy
Manufacturer's Included in
acid No Total RNA prep cartridge DNA/RNA
kit PCR
DNA extraction Isolation Kit) kit extraction kit
extraction kits (atc-
pd/rna)
Whole Whole blood,
blood, Whole blood, Serum, Whole blood, serum, Various
Sample type
serum, serum, tissues swabs*** tissues plasma, specimen types
tissues tissue, swabs
95C for
Time 30 min 40 min 30 min 2 min 60 min
2 min
Cost (US$) /
5.00 0.00 0.00 5.00 to 10.00 5.00 to 10.00 5.00 to 10.00
test
10
The OIE ASF Reference Laboratory Network's
overview of African swine fever
diagnostic tests for
field application
Table 4 cont’d. Comparison of Point-of-Care PCR systems for rapid ASF virus DNA detection
Test methods
iiPCR+DNA
System
iiPCR extraction LAMP qPCR qPCR qPCR
(fully automatic)
Manufacturer GeneReach GeneReach OptiGene Tetracore Indical Genesig
High
Sensitivity High High Moderate High High
(LOD<100)
Specificity High High High High High High
Training needs Moderate Low High High High High
Peer-
Level of Evaluated Peer-reviewed reviewed Peer-reviewed Peer-reviewed Evaluated by
Performance assessment by FAO journal article journal journal article journal article FAO
article
Peer-
Daigle et
reviewed Peer-reviewed
Tran et al. Mee et al. al. (2020);
References publication Liu et al. (2019) publication not
(2021) (2020) Elnagar et al.
not yet yet available
(2021)
available
Full automatic,
Low-cost No DNA Same as lab- Same as lab- Same as lab-
just load
equipment extraction based qPCR based qPCR based qPCR
sample and go
Battery-
Advantages
operated
No training
DNA
needed
automatic
extraction
Disadvantages Equipment: high cost
*Validated in-house real-time PCR tests recommended by the OIE are King et al. (2003) and Fernandez-Pinero et al.
(2013).
**PCR Commercial Kits currently validated: INgene q PPA, INGENASA. 11.PPA.K.5TX/Q; Tetracore TC-9017-064;
Virotype ASFV PCR Kit, INDICAL BIOSCIENCE; LSI VetMAXTM Thermo Fisher Scientific; IDEXX RealPCR ASFV Mix,
IDEXX; ID Gene® African Swine Fever Duplex – IDVet; ADIAVET ASFV REAL TIME 100R, BIO-X DIAGNOSTICS.
Commercial LAMP kit available from Geneworks (https://geneworks.com.au/; KIT-ASFV-96P).
***Other sample types such as whole blood and tissues can be tested if DNA extraction is performed prior to LAMP
testing (James et al., 2010)
VetMAXTM African Swine Fever Virus Detection Kit (Taqman® real time PCR) manufactured by Thermo Fisher Scientific LSI S.A.S.
is included in the OIE Register of Diagnostic kits, www.oie.int/en/what-we-offer/veterinary-products/diagnostic-kits/the-register-
of-diagnostic-kits/
Acknowledgements
This initiative was made possible by funding support from the people of Japan through the Ministry of
Agriculture, Forestry and Fisheries, Japan.
11
References
Cappai S., Loi F., Coccollone A., Cocco M., Falconi C., Dettori G., Feliziani F., Sanna M.L., Oggiano A. & Rolesu S.
(2017). – Evaluation of a Commercial Field Test to Detect African Swine Fever. Journal of Wildlife Diseases, 53 (3),
602–606. https://doi.org/10.7589/2016-05-112
Daigle J., Onyilagha C., Truong T., Le V.P., Nga B., Nguyen T.L., Clavijo A. & Ambagala, A. (2021). – Rapid and highly
sensitive portable detection of African swine fever virus. Transboundary and Emerging Diseases, 68 (2), 952–959.
https://doi.org/10.1111/tbed.13770
Elnagar A., Pikalo J., Beer M., Blome S. & Hoffmann B. (2021). – Swift and Reliable 'Easy Lab' Methods for the
Sensitive Molecular Detection of African Swine Fever Virus. International Journal of Molecular Sciences, 22 (5),
2307. https://doi.org/10.3390/ijms22052307
Fernández-Pinero J., Gallardo C., Elizalde M., Robles A., Gómez C., Bishop R., Heath L., Couacy-Hymann E., Fasina
F.O., Pelayo V., Soler A. & Arias M. (2013). – Molecular diagnosis of African Swine Fever by a new real-time PCR
using universal probe library. Transboundary and Emerging diseases, 60 (1), 48–58. https://doi.org/10.1111/j.1865-
1682.2012.01317.x
James H.E., Ebert K., McGonigle R., Reid S.M., Boonham N., Tomlinson J.A., Hutchings G.H., Denyer M., Oura C.A.,
Dukes J.P. & King D.P. (2010). – Detection of African swine fever virus by loop-mediated isothermal amplification.
Journal of Virological Methods, 164 (1–2), 68–74. Epub 2009 Dec 4. PMID: 19963011. https://doi.org/10.1016/j.
jviromet.2009.11.034
King D.P., Reid S.M., Hutchings G.H., Grierson S.S., Wilkinson P.J., Dixon L.K., Bastos A.D. & Drew T.W. (2003).
– Development of a TaqMan PCR assay with internal amplification control for the detection of African swine fever
virus. Journal of Virological Methods, 107 (1), 53–61. https://doi.org/10.1016/s0166-0934(02)00189-1
Liu L., Atim S., LeBlanc N., Rauh R., Esau M., Chenais E., Mwebe R., Nelson W.M., Masembe C., Nantima N.,
Ayebazibwe C. & Ståhl K. (2019). – Overcoming the challenges of pen-side molecular diagnosis of African swine
fever to support outbreak investigations under field conditions. Transboundary and Emerging Diseases, 66 (2),
908–914. https://doi.org/10.1111/tbed.13103
Matsumoto N., Siengsanan-Lamont J., Gleeson L.J., Douangngeun B., Theppangna W., Khounsy S., Phommachanh
P., Halasa T., Bush R.D. & Blacksel S.D. (2020). – Evaluation of the diagnostic accuracy of an affordable rapid
diagnostic test for African swine fever antigen detection in Lao People’s Democratic Republic. Journal of Virological
Methods, 286, 113975. https://doi.org/10.1016/j.jviromet.2020.113975
Mee P.T., Wong S., O’Riley K.J., da Conceição F., Bendita da Costa Jong J., Phillips, D.E., Rodoni B.C., Rawlin G.T.
& Lynch S.E. (2020). – Field Verification of an African swine fever Virus Loop-Mediated Isothermal Amplification
(LAMP) Assay During an Outbreak in Timor-Leste. Viruses, 12 (12), 1444. https://doi.org/10.3390/v12121444
Sastre P., Gallardo C., Monedero A., Ruiz T., Arias M., Sanz A., & Rueda P. (2016a). – Development of a novel lateral
flow assay for detection of African swine fever in blood. BMC Veterinary Research, 12, 206. https://doi.org/10.1186/
s12917-016-0831-4
Sastre P., Pérez T., Costa S., Yang X., Räber A., Blome S., Goller K.V., Gallardo C., Tapia I., García J., Sanz A. &
Rueda P. (2016b). – Development of a duplex lateral flow assay for simultaneous detection of antibodies against
African and Classical swine fever viruses. Journal of Veterinary Diagnostic Investigation, 28 (5), 543–549. https://
doi.org/10.1177/1040638716654942
Tran H., Le N.T., Pham B.P., Luu V.Q. & Nguyen V.L. (2021). – Evaluation of an automated insulated isothermal
polymerase chain reaction system for rapid and reliable, on-site detection of African swine fever virus. Journal of the
American Veterinary Medical Association, 259 (6), 662–668. https://doi.org/10.2460/javma.259.6.662
12
12, rue de Prony - 75017 Paris, France
Tel.: +33(0)1 44 15 18 88 | Fax: +33 (0)1 42 67 09 87 | www.oie.int