Electrical methods
Measuring the amount of absorption of microwave energy beamed through the mater-
ial is the most common technique for measuring moisture content and is described in
detail in Anderson (1989), and Thompson (1989). Microwaves at wavelengths between
1 mm and 1 m are absorbed to a much greater extent by water than most other materials.
Wavelengths of 30 mm or 100 mm are commonly used because ‘off-the-shelf’ equip-
ment to produce these is readily available from instrument suppliers. The technique
is suitable for moisture measurement in solids, liquids and gases at moisture-content
levels up to 45% and measurement uncertainties down to š0.3% are possible.
The capacitance moisture meter uses the principle that the dielectric constant of
materials varies according to their water content. Capacitance measurement is there-
fore related to moisture content. The instrument is useful for measuring moisture-
content levels up to 30% in both solids and liquids, and measurement uncertainty
down to š0.3% has been claimed for the technique (Slight, 1989). Drawbacks of
the technique include (a) limited measurement resolution owing to the difficulty in
measuring small changes in a relatively large standing capacitance value and (b) diffi-
culty when the sample has a high electrical conductivity. An alternative capacitance
Measurement and Instrumentation Principles 433
charge transfer technique has been reported (Gimson, 1989) that overcomes these prob-
lems by measuring the charge carrying capacity of the material. In this technique, wet
and dry samples of the material are charged to a fixed voltage and then simultaneously
discharged into charge-measuring circuits.
The electrical conductivity of most materials varies with moisture content and this
therefore provides another means of measurement. Techniques using electrical conduc-
tivity variation are cheap and can measure moisture levels up to 25%. However, the
presence of other conductive substances in the material such as salts or acids affects
the measurement.
A further technique is to measure the frequency change in a quartz crystal that
occurs as it takes in moisture.
Neutron moderation
Neutron moderation measures moisture content using a radioactive source and a neutron
counter. Fast neutrons emitted from the source are slowed down by hydrogen nuclei in
the water, forming a cloud whose density is related to the moisture content. Measure-
ments take a long time because the output density reading may take up to a few
minutes to reach steady state, according to the nature of the materials involved. Also,
the method cannot be used with any materials that contain hydrogen molecules, such
as oils and fats, as these slow down neutrons as well. Specific humidities up to 15%
(š1% error) can be measured.
Optical methods
The refractometer is a well-established instrument that is used for measuring the
water content of liquids. It measures the refractive index of the liquid, which changes
according to the moisture content.
Moisture-related energy absorption of near-infra-red light can be used for measuring
the moisture content of solids, liquids and gases. At a wavelength of 1.94 µm, energy
absorption due to moisture is high, whereas at 1.7 µm, absorption due to moisture is
zero. Therefore, measuring absorption at both 1.94 µm and 1.7 µm allows absorption
due to components in the material other than water to be compensated for, and the
resulting measurement is directly related to energy content. The latest instruments use
multiple-frequency infra-red energy and have an even greater capability for eliminating
434 Summary of other measurements
the effect of components in the material other than water that absorb energy. Such
multi-frequency instruments also cope much better with variations in particle size in
the measured material.
In alternative versions of this technique, energy is either transmitted through the
material or reflected from its surface. In either case, materials that are either very dark
or highly reflective give poor results. The technique is particularly attractive, where
applicable, because it is a non-contact method that can be used to monitor moisture
content continuously at moisture levels up to 50%, with inaccuracy as low as š0.1%
in the measured moisture level. A deeper treatment can be found in Benson (1989).
Ultrasonic methods
The presence of water changes the speed of propagation of ultrasonic waves through
liquids. The moisture content of liquids can therefore be determined by measuring
the transmission speed of ultrasound. This has the inherent advantage of being a non-
invasive technique but temperature compensation is essential because the velocity of
ultrasound is particularly affected by temperature changes. The method is best suited to
measurement of high moisture levels in liquids that are not aerated or of high viscosity.
Typical measurement uncertainty is š1% but measurement resolution is very high, with
changes in moisture level as small as 0.05% being detectable. Further details can be
found in Wiltshire (1989).
Mechanical properties
Density changes in many liquids and slurries can be measured and related to moisture
content, with good measurement resolution up to 0.2% moisture. Moisture content
can also be estimated by measuring the moisture level-dependent viscosity of liquids,
pastes and slurries.
Water separation
Various laboratory techniques are available that enable the moisture content of liquids
to be measured accurately by separating the water from a sample of the host liquid.
Separation is effected either by titration (Karl Fischer technique), distillation (Dean
and Stark technique) or a centrifuge. Any of these methods can measure water content
in a liquid with measurement uncertainty levels down to š0.03%.
Gravimetric methods
Moisture content in solids can be measured accurately by weighing the moist sample,
drying it and then weighing again. Great care must be taken in applying this procedure,
as many samples rapidly take up moisture again if they are removed from the drier
and exposed to the atmosphere before being weighed. Normal procedure is to put the
sample in an open container, dry it in an oven and then screw an airtight top onto the
container before it is removed from the oven.
Measurement and Instrumentation Principles 435
Phase-change methods
The boiling and freezing point of materials is altered by the presence of moisture,
and therefore the moisture level can be determined by measuring the phase-change
temperature. This technique is used for measuring the moisture content in many food
products and in some oil and alcohol products.
Microscope
Sample Sample
in out
Mirror surface
Electrical
cooling
Temperature sensor
(bonded to mirror
surface)
so that the thickness and nature of the condensate can be observed. The instrument is
described in greater detail in Pragnell (1989).
Even small levels of contaminants on the mirror surface can cause large changes in
the dew point and therefore the instrument must be kept very clean. When necessary,
the mirror should be cleaned with deionized or distilled water applied with a lint-free
swab. Any contamination can be detected by a skilled operator, as this makes the
condensate look ‘blotchy’ when viewed through the microscope. The microscope also
shows up other potential problems such as large ice crystals in the condensate that
cause temperature gradients between the condensate and the temperature sensor. When
used carefully, the instrument is very accurate and is often used as a reference standard.