You are on page 1of 99

'LDPRQG

,QVWDOODWLRQ,QVWUXFWLRQV
May 2001 86501-4A
COPYRIGHT  2001 by Instrumentarium Imaging

Documentation, trademark and the software are copyrighted


with all rights reserved. Under the copyright laws the
documentation may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine
readable form in whole or part, without the prior written
permission of Instrumentarium Imaging.

This document is the English-language original. In case of


interpretation disputes the English text applies.

Instrumentarium Imaging reserves the right to revise this


publication from time to time and to make changes in the
content of it without obligation to notify any person of such
revisions or changes.

Windows is trademark of Microsoft Corporation in the United


States of America and other countries. HyperTerminal is a
trademark of Hilgraeve Inc, Monroe, MI, USA.

MANUFACTURED BY: Instrumentarium Imaging


P.O. Box 20
FIN-04301 Tuusula
FINLAND
Tel. +358-10-394 6500
Fax. +358-10-394 6501
E-mail: mammo@fi.instrumentarium.com
Internet: http://www.instrumentarium.com/imaging
Diamond Installation Instructions

7$%/(2)&217(176

 127(6&$87,216$1':$51,1*6  

 35(,167$//$7,21,16758&7,216  
 81,70($685(0(176$1'63$&(5(48,5(0(176  
 (/(&75,&$/6833/<5(48,5(0(176  
 (19,5210(17$/5(48,5(0(176 
 722/61(('(')25,167$//$7,21  
 7(67(48,30(171(('(')25,167$//$7,21  
 75$163257$7,212)$1$66(0%/('81,7  
 (;7(51$/:$51,1*/,*+7&211(&7,21  

 813$&.,1*7+(81,7  
 813$&.,1*7+(&5$7(  
 813$&.,1*7+(75$163257$7,21:+((/6 237,21$/  
 813$&.,1*7+(&2817(5:(,*+77,/(6  

 ,167$//$7,21352&('85(6 
 /,)7,1*837+(&2/8010,1,0803(23/( 
 ,16(57,212)7+(&2817(5:(,*+77,/(6 
 $77$&+0(172)7+(&21752/3$1(/  
 $77$&+0(172)7+(3('$/6  

 7(67,1*7+(&21752/)81&7,216  

 67$5783352&('85(6 
 &+(&.,1*7+(&$66(77(+2/'(56  
 32:(5,1*83  
 $12'(58192/7$*(&+(&.  
 .95(*8/$7,21&+(&. 
 35(+($7$1'0$5(*8/$7,21&+(&.  
6.5.1 Measuring the preheat values ............................................................... 30
6.5.2 Preheat charts........................................................................................ 30
6.5.3 Adjusting the preheat and mA-values .................................................. 31
 $77$&+,1*7+(&29(56  
6.6.1 Installation of the tube head cover........................................................ 32
6.6.2 Left side cover ...................................................................................... 35
6.6.3 Right side cover .................................................................................... 35
6.6.4 Attaching the counterweight protection sheet ...................................... 37

86501 May 2001


Installation Instructions Diamond

 &+(&./,6735,257286(  
 ,1,7,$/6(77,1*6  
7.1.1 Date/Time (normal mode) .................................................................... 39
7.1.2 Using service mode interface................................................................ 39
7.1.3 Printer/Port setup .................................................................................. 40
7.1.4 Connecting film marker (ID II) to Diamond unit ................................. 41
 &+(&.,7(06 
7.2.1 Light field intensity............................................................................... 42
7.2.2 X-ray field alignment............................................................................ 43
7.2.3 Light field alignment ............................................................................ 44
7.2.4 Film processor evaluation..................................................................... 44
7.2.5 C-arm angle display calibration............................................................ 45
7.2.6 Compression force display calibration ................................................. 46
7.2.7 Compression thickness display calibration........................................... 47
7.2.8 kV, mAs and exposure time test ........................................................... 48
7.2.9 Grid alignment ...................................................................................... 48
7.2.10 Voltage check ....................................................................................... 48
 $(&$'-8670(17  
7.3.1 Quick AEC test ..................................................................................... 49
7.3.2 Recording the system settings............................................................... 49
7.3.3 Basic adjustment................................................................................... 50
7.3.4 Automatic AEC detector gain adjustment ............................................ 52
7.3.5 Object thickness compensation............................................................. 53
7.3.6 Thickness compensation....................................................................... 55
7.3.7 kV compensation .................................................................................. 59
7.3.8 Auto-kV programming ......................................................................... 62
7.3.9 Target times .......................................................................................... 63
7.3.10 Initial kV-tables .................................................................................... 65
7.3.11 AEC adjustment quick fix guide .......................................................... 66
7.3.12 Printout of system data ......................................................................... 66
 6<67(0&86720,=$7,217286(535()(5(1&(6  
 ',6$%/,1*;5$<6 
7.5.1 Disabling the radiation.......................................................................... 67
7.5.2 Enabling the radiation.......................................................................... 67

 $33(1',&(6 

May 2001 86501


Diamond Installation Instructions

 127(6&$87,216$1':$51,1*6

127( 2QO\HQJLQHHUVWUDLQHGE\,QVWUXPHQWDULXPDUHDOORZHGWRLQVWDOOWKHXQLWRU
VWDQGE\LWGXULQJWKHLQVWDOODWLRQRUVHUYLFHZRUN3DUWVLQVWDOOHGE\QRQ
WUDLQHGSHUVRQQHOFDUU\QRZDUUDQW\

&$87,21 $OZD\VFKHFNIRUSURSHUVHDWLQJRQDOOWKHFRQQHFWRUVRUH[WHUQDOFRPSRQHQWV
ORFDWHGLQWKHDUHDV\RXKDYHZRUNHG(TXLSPHQWGDPDJHFDQUHVXOWLISRZHU
LVDSSOLHGRUH[SRVXUHVPDGHZLWKLPSURSHUO\VHDWHGFRQQHFWRUVRUH[WHUQDO
FRPSRQHQWV

:$51,1* $OZD\VLQVWDOOWKHSURSHUIXVHVLQWRWKHXQLW)DLOXUHWRGRVRMHRSDUGL]HV
SDWLHQWRSHUDWRUDQGHTXLSPHQWVDIHW\

:$51,1* 7KLVXQLWKDVIORDWLQJJURXQGV$OZD\VPDNHDOOVFRSHPHDVXUHPHQWVGLI
IHUHQWLDOO\DQGDOZD\VEHFDUHIXOZKHQVHUYLFLQJWKHXQLW

&$87,21 :KHQ KDQGOLQJ FLUFXLW ERDUGV DOZD\V XWLOL]H JURXQG ZULVW VWUDSV RU
JURXQGLQJ NLWV WR SUHYHQW GDPDJH WR WKH PRV W\SH FKLSV $OZD\V NHHS
FLUFXLWERDUGVLQDVWDWLFSURRIEDJ

:$51,1* $OZD\VGLVFRQQHFWWKHXQLWIURPOLQHVXSSO\E\UHPRYLQJWKHSOXJIURPZDOO
RXWOHWEHIRUHRSHQLQJWKHFRYHUV

:$51,1* 7KH[UD\WXEHXWLOL]HVDEHU\OOLXPZLQGRZ'RQRWSHUIRUPDQ\RSHUDWLRQV
RQEHU\OOLXPZKLFKSURGXFHGXVWRUIXPHVVXFKDVJULQGLQJJULWEODVWLQJRU
DFLGFOHDQLQJ%HU\OOLXPGXVWRUIXPHVDUHKLJKO\WR[LFDQGEUHDWKLQJWKHP
FDQUHVXOWLQVHULRXVSHUVRQDOLQMXU\RUGHDWK

:$51,1* 7KLV [UD\ XQLW PD\ EH GDQJHURXV WR SDWLHQW DQG RSHUDWRU XQOHVV VDIH
H[SRVXUHIDFWRUVDQGLQVWUXFWLRQVDUHREVHUYHG

:$51,1* 3D\ DWWHQWLRQ GXULQJ WKH LQVWDOODWLRQ QRW GDPDJH RSWRVZLWFKHV FDUP
PRXQWLQJDQGRWKHUSDUWV&KHFNWKHXQLWWKRURXJKO\EHIRUHKDQGLQJLWWRWKH
FOLHQW

86501 1 May 2001


Installation Instructions Diamond

 35(,167$//$7,21,16758&7,216

127( 7ZRSHRSOHDUHUHTXLUHGWKURXJKWKHPHFKDQLFDOLQVWDOODWLRQRQHSHUVRQ
WKHUHDIWHU

127( &DOLEUDWLRQLQVWUXFWLRQVDUHIRUUHIHUHQFHRQO\&RXQWU\VSHFLILFUHJXODWLRQV
PXVWEHREVHUYHGDQGNHSW

:$51,1* 3D\DWWHQWLRQWRWKHVDIHW\RI\RXUILQJHUVZKHQKDQGOLQJWKHFRXQWHUZHLJKW
WLOHSDFNDJH

 81,70($685(0(176$1'63$&(5(48,5(0(176
The unit measurements are shown in figure below. Dimensions are in millimeters
(mm). To convert to inches, multiply given dimension by 0.0394 (1mm
=0.0394in).

Figure 2.1. Unit measurements

The unit must be installed on a level surface in such a way that there is enough
(70cm recommended) space between the c-arm and various objects (walls,
furniture).

May 2001 2 86501


Diamond Installation Instructions

 (/(&75,&$/6833/<5(48,5(0(176
/LQH9ROWDJH: 230 VAC ±10% (nominal)

127( 2SHUDWLQJ5DQJHWR9$&/LQHYROWDJHVPHDVXULQJOHVVWKDQ
9$&PD\UHTXLUHDOLQHERRVWWUDQVIRUPHU

/LQH5HJXODWLRQ ± 2% (full load)


3KDVH single phase (Line, Neutral, Protective Ground or L1, L2,
Protective Ground)
)UHTXHQF\ 50/60Hz
&XUUHQW stand-by 300mA
maximum 20 A
instantaneous peak 30 A
%UDQFK&LUFXLW dedicated Line, Neutral, Protective Ground 2.5mm2
(L1 and L2 10 AWG, stranded wire, continuous run)
circuit breaker 20 A, 2-pole
(outlet per local code requirements or NEMA 6-20)
VDE Wall plug is included
*URXQGLQJ dedicated, wired (not conduit), 0-potential (earth)
ground
6WDQGDUGRXWOHW (230/15A) required for any optional accessories

 (19,5210(17$/5(48,5(0(176
2SHUDWLQJ7HPSHUDWXUH +10...+40°C (60° - 85°F)
+XPLGLW\5DQJH 1585%, non-condensing
:HLJKW 350 kg (772 lbs)
'LPHQVLRQV Height 194 cm (76 1/3")
Width 112 cm (44")
Length 121 cm (47 2/3")

127( 6WRUDJHDQGWUDQVSRUWDWLRQLQGU\FRROSODFH

 722/61(('(')25,167$//$7,21
- Electrical screwdriver (to open the crates)
- Screwdrivers small/medium
- Screwdrivers POZIDRIVE small/medium
- Allen key set (long type 1.5 - 8 mm)
- Wrench set 7 - 19 mm
- Pliers
- patch paint for covers (included in the unit shipment)

86501 3 May 2001


Installation Instructions Diamond

 7(67(48,30(171(('(')25,167$//$7,21
- Digital True RMS Multimeter
- Oscilloscope
- Compression force gauge (bathroom scale)
- Densitometer
- Sensitometer
- Acrylic blocks 1x5mm, 2x10mm, 2x20mm (supplied with Diamond)
- Lux-meter
- 0.1mm Al plates (in USA)
- separate or complete meter for following measurements:
- kV (Unfors, RTI, PMX-I/M or equivalent)
- Radiation meter for mammo to measure dose
- exposure time
- RMI 156 test pahantom for image evaluation
- small round magnet for Hall sensor testing
- portable PC with HyperTerminal® program to retrieve parameter data

 75$163257$7,212)$1$66(0%/('81,7
For transportation in the car, the unit has to be repacked. If the transportation is
short and smooth, the unit may remain in upright position in following conditions:

1) Counterweight tiles remain inside the column

2) The counterweight tiles must rest on the c-arm assembly pin.

3) Run the ECS down on some support piece that rests on the base plate.

4) Leave some slack in cables.

In order to unpack, follow the unpacking instructions in Chapter 3 in reversed


order.

May 2001 4 86501


Diamond Installation Instructions

 (;7(51$/:$51,1*/,*+7&211(&7,21
An external warning light cable can be connected to the Diamond unit (port is
described in Chapter 4.4, Attachment of the pedals). This connections provides
information (relay connection for the lamp) as follows

 [UD\XQLW21

 H[SRVXUH21

In order to hook up the cable, check how far it has to reach and how it affects the
positioning of the unit.

For electrical details, refer to Diamond Service Manual (86502), Chapter 9,


Schematics/Lay-outs, 80520 Power Board Schematics, page 2 of 4. The same
manual includes also a drawing in Chapter 9, 85101 Warning light cable.

86501 5 May 2001


Installation Instructions Diamond

 813$&.,1*7+(81,7
The standard unit delivery comprises of one crate of the following dimensions:

CRATE:
- 2110 x 1320 x 740 mm (83" x 52" x 29 1/8")
- wooden
- weight approximately 470 kg (1 035 lbs.)
- column with carriage and standard accessories

127( :LWKWUDQVSRUWDWLRQZKHHOVWKHXQLWFDQEHSUHLQVWDOOHGDQGFDOLEUDWHGLQ
DQRWKHUURRPDQGWKHQWUDQVSRUWHGWRWKHILQDOSODFHRIXVH

 813$&.,1*7+(&5$7(

127( 7KHFUDWHLQFOXGHVWKHSDFNLQJOLVW0DNHVXUHWKDW\RXKDYHUHFHLYHGDOO
LWHPV

- cut the packing straps


- unscrew the cover
- open up the fastened cover and remove it
- open up and remove the end with the column top so that the package will look
as below

- remove the box(es) that contain accessories and miscellaneous items

May 2001 6 86501


Diamond Installation Instructions

- remove the supporting brackets by loosening the bolts outside the crate
- remove the carriage locking bracket lift out the column and carriage

After removing the covers, the crate looks like this (the transportation wheels in
the image are optional):

Figure 2.2. Column with carriage

86501 7 May 2001


Installation Instructions Diamond

 813$&.,1*7+(75$163257$7,21:+((/6 237,21$/


- unscrew and remove the metal or plastic supports that keep the wheels steady

 813$&.,1*7+(&2817(5:(,*+77,/(6
- remove the board on both sides of the system and remove the counterweight
tiles

127( ,QFDVHWKHXQLWKDVWREHFDUULHGWKURXJKWKHVWHSVRUWLJKWVSDFHVUHPRYH
WKHEDVHSODWHWRPDNHWKHV\VWHPOLJKWHUWRFDUU\

May 2001 8 86501


Diamond Installation Instructions

 ,167$//$7,21352&('85(6

:$51,1* $V FDUP PRYHPHQW LV VWURQJ WKH GHYLFH PXVW EH LQVWDOOHG VR WKDW WKH
RSHUDWRUZLOOQRWJHWFDXJKWEHWZHHQFDUPDQGZDOOWDEOHZKLOHRSHUDWLQJ
UHJXODURUVWHUHRFDUPPRYHPHQW$IWHUWKHLQVWDOODWLRQFKHFNWKDWWKHUHDUH
QRREVWUXFWLRQVWRIUHHFDUPURWDWLRQDOVRZKHQLQSDUNHGSRVLWLRQ7KH
PRYHPHQWFDQDOVREHFDQFHOHGE\SUHVVLQJRQHRIWKHHPHUJHQF\EXWWRQVRU
DQ\EXWWRQRQ'LDPRQGXQLWH[FHSWRQWKH&RQWURO3DQHO

:$51,1* &KHFNWKDWFRXQWHUZHLJKWDVVHPEO\DQGWKHFRXQWHUZHLJKWFDEOHVDUH2.VR
WKDW WKH V\VWHP ZRXOG QRW EHFRPH XQEDODQFHG DQG IDOO RQ RSHUDWRU RU
SDWLHQW

 /,)7,1*837+(&2/8010,1,0803(23/(
1) Check that the c-arm assembly pin is securely attached.

2) Lift the column up smoothly so that the wooden crate base stands up with
the column.

86501 9 May 2001


Installation Instructions Diamond

3) Remove the upper column support. Hold the wooden crate base steady.

4) Remove the lower column support.

5) Remove the base of the crate and lay it somewhere where it does not
disturb the rest of installation process.

6) Open up two screws to remove the tube support.

:$51,1* *HWDVVLVWDQFHDQGEHFDUHIXOZKHQHUHFWLQJWKHFROXPQ$VFRXQWHUZHLJKW
WLOHVDUHLQVHUWHGODWHUWKHFROXPQLVQRZRIIWKHEDODQFH

May 2001 10 86501


Diamond Installation Instructions

7) Attach the main switch box on.

&$87,21 5HPHPEHU WR SODFH VWDU ZDVKHUV XQGHUQHDWK WKH EROW WR DWWDLQ SURSHU
JURXQGLQJ

8) Dispose of the wooden crate and other packaging materials. Recycle


according to country-specific regulations.

The next step is to insert the counterweight tiles as described in the following
chapter.

86501 11 May 2001


Installation Instructions Diamond

 ,16(57,212)7+(&2817(5:(,*+77,/(6
&$87,21 &KHFNWKDWWKHDVVHPEO\SLQVDUHLQVHUWHGWKURXJKWKHKROHVDWERWKVLGHVRI
WKHFROXPQDQGIL[HGZLWKDEROW

1) Check that the c-arm assembly pin is inserted. This will hold the c-arm
steady while adding counterweight tiles.

:$51,1* 'R QRW UHOHDVH WKH FDUP DVVHPEO\ SLQ EHIRUH LW LV LQVWUXFWHG WR GR VR
0RYHPHQWRIWKHXQFRYHUHGFRXQWHUZHLJKWFDJHSRVHVULVNWRILQJHUV

Tiles in the counterweight cage.

9LHZIURPDERYH

May 2001 12 86501


Diamond Installation Instructions

2) Slide in counterweight tiles into the counterweight cage one by one. The
longer side of the tile is inserted to the front. Start with the smaller ones
(more space for fingers).

:$51,1* %HFDUHIXOZLWKWKHILQJHUV

In order to avoid the tiles from touching the inside of the column, push them as
much forward as possible so that they are tightly lined up against the back of the
counterweight cage.

9LHZIURPDVLGH
3) When all counterweight tiles have been inserted, attach the support rod
that in order to keep the tiles in place. In order to do this, switch on the
unit and then release the c-arm assembly pin.

127( :KHQWKHSRZHULVVZLWFKHGRQWKHWXEHKHDGUXQVIURPDQXSVLGHGRZQ
SRVLWLRQWRDQXSULJKWSRVLWLRQ

&$87,21 $W WKLV VWDJH WKH FRXQWHUZHLJKW WLOHV DUH QRW \HW VHFXUHG EXW WKH
FRXQWHUZHLJKWFDJHFDQEHGULYHQXSRUGRZQ

:$51,1* 'RQRWPHFKDQLFDOO\IRUFHWKHFDUPDVVHPEO\SDVWWKHYHUWLFDOPRYHPHQW
VHQVRUSLHFHVLQORZHUDQGXSSHUSDUWVRIWKHFROXPQ RQERWKVLGHV 7KH
VHQVRUVKDYHLQWKLVFDVHDYHU\KLJKULVNWRJHWEURNHQ

86501 13 May 2001


Installation Instructions Diamond

4) Drive the the carriage so that the opening underneath the counterweight
tiles and the screw that secures the lower part of the tile support rod are in
level. Then, loosen up the screw a bit.

&$87,21 'RQRWORRVHQXSWKHEROWWRRPXFKVRWKDWLWPLJKWKLQGHUWKHPRYHPHQWRI
WKHFDUULDJHRUGURSGRZQEHWZHHQFROXPQDQGWKHFRXQWHUZHLJKWFDJHDV
LQWRWKHFROXPQ

5) Attach the support rod so that screws align with the slanted side.

May 2001 14 86501


Diamond Installation Instructions

6) Slide the support rod in (A) so that the screw does not let it any further.
Then drive the c-arm so that the slanted side appears (B). Then tighten up
the screw (C) to secure the lower end of the rod.

7) Align the support rod vertically straight. Check that the support rod is
steadily fixed at the lower end and does not move alone, except with the
counterweight cage. Then secure the upper end of the support rod.

8) Run the c-arm to the lowest and highest position and listen to any abrasive
or scratchy sounds. If any, then rearrange some key tiles that seem cause
the sound.

&$87,21 &KHFNWKDWWKHFRXQWHUZHLJKWFDEOHLVQRWZULQNOHGRUDQ\KRZGDPDJHGDQG
WKDWLWFDQRSHUDWHZLWKRXWDEUDVLYHFRQWDFW V WKDWPD\FDXVHVHULRXVGDPDJH
WRFDEOH

:$51,1* 7KHFRXQWHUZHLJKWFRYHUKDWFKLVLQVWUXFWHGWREHLQVHUWHGLQ&KDSWHU
0HDQZKLOHWKHDUHDZLOOUHPDLQRSHQVREHYHU\FDUHIXOWKDWQRRQH KDV
KLVKHUILQJHUVWKHUHZKHQYDULRXVPRYHPHQWVDUHGRQHWRFKHFNIRUQRLVHV

86501 15 May 2001


Installation Instructions Diamond

Transportation wheels can be installed in case the Diamond unit needs to be


moved to another location.

9) Insert the bolt to the base plate and tighten up a bit. Later on, if necessary,
the bolt can be tightened or slackened with a screwdriver (right image).

10) Roll the nut all the way down.

11) Attach the rod and lift up the base plate by tightening up the nut-bolt
connection.

127( /LIWWKHEDVHSODWHHYHQO\XS'RQRWWLJKWHQXSRQHQXWDOOWKHZD\EHIRUH
VWDUWLQJZLWKWKHQH[WRQH

May 2001 16 86501


Diamond Installation Instructions

12) When all four nuts have been fixed, the base plate with Diamond unit is
ready for local transportation. After the final location for the Diamond has
been determined and the unit positioned there, repeat steps 1-15 in reverse
order and slackening instead of tightening.

13) If the floor surface is not level, adjust with a support piece with a
screwdriver. Then attach the plastic cover piece. Diamond base plate can
be adjusted at each corner.

86501 17 May 2001


Installation Instructions Diamond

 $77$&+0(172)7+(&21752/3$1(/
The control panel can be attached to the radiation shield or onto the wall.

1) Check that the unit is switched off.

2) Align the control panel base plate with the groove in the radiation shield
column. Adjust for the height, then secure the base plate with two screws
that force the piece in the middle of the base plate to the radiation shield.
Check that the base plate is steadily fixed to bear the weight of the control
panel.

3) Slide the control panel cord (up from below) through the opening in the
counterpiece. Then, secure the piece with two screws. The piece should
look as below and be steady.

4) Lay the control panel rest on the round end of support rod. Then, take up
a lock piece. Be careful, so that the control panel would not fall off as it
is not yet steady.

May 2001 18 86501


Diamond Installation Instructions

4) Insert the lock piece and tighten up, but leave some slack so that the
control panel can be adjusted for angle. When the required angle has been
found, secure lock piece tightly.

5) Plug in the control panel cord to Diamond (at the back of the Diamond
column).

6) The control panel is now ready for use (after the power switched on after
completion of all other installation work).

Figure 2.3. Control panel attached

86501 19 May 2001


Installation Instructions Diamond

 $77$&+0(172)7+(3('$/6
Attach the cables of the two pedals to the connection box on the back of the
column. See figure below.

Diamond ports:

COM1 for software upload

COM2 for ID II film


marking or label printing

Two ports for two pedals


(paddles)

Diamond Control Panel port

Delta 32 port

Warning light port


Figure 2.4. Connectors for the pedals

May 2001 20 86501


Diamond Installation Instructions

 7(67,1*7+(&21752/)81&7,216
1) In order to test the system, insert the emergency stop button override
(plastic piece with red cable) in place.

:$51,1* 5HPHPEHUWRXQSOXJWKHRYHUULGH SLHFH ZKHQ \RX ILQLVK ZLWK WKH V\VWHP


FKHFNLQJ:KHQWKHRYHUULGHSLHFHLVQRWXVHGLWKDQJVRQDSRQGXLW

:$51,1* %HDULQPLQGWKDWWKHHPHUJHQF\VWRSEXWWRQVDUHQRZLQRSHUDWLRQDO,QFDVH
DQ HPHUJHQF\ VWRS LV QHFHVVDU\ XVH WKH XVHU SRZHU VZLWFK RU WKH PDLQV
VZLWFKDWWKHEDFNRIWKHXQLW DQGLQ)LJXUH'LDPRQG2SHUDWLRQ
,QVWUXFWLRQV  &KDSWHU2SHUDWLQJ&RQWUROV 

2) Check that the circuit breaker in the back of system is in the 21 position.

3) Switch power 21 using the main power switch in the front.When turning
on the system, the tubehead should move forward, the diaphragm blades
are driven their whole traveling distance and the cassette tray inside the
bucky drives to the front and back.

The control panel should give the following indications:

UHDG\ light: ON P*\ display: 


[UD\ light: OFF $XWR3RLQWlight: ON
WLPHU display: 0 0RGH light: Auto-kV
N9 display: 25 )LOWHU light: Mo
P$V display: ---
GHQVLW\ display: +
0

Also, when turning on the system, the Diamond LCD screen displays the current
software version installed into the unit.

&$87,21 'R QRW SUHVV WKH H[SRVXUH EXWWRQ XQWLO DOO QHFHVVDU\ FDOLEUDWLRQV DUH
SHUIRUPHG

86501 21 May 2001


Installation Instructions Diamond

4) Using the N9 and N9buttons , increase kV to 35, decrease


to 23 kV, and then return to the default kV setting.

5) Using the GHQVLW\and GHQVLW\buttons , increase density to


+7, decrease to -7 and then return the setting to 0.

6) Change the filter and notice that the filter does change from molybdenum
to rhodium. Notice that the timer display changes from 0Rto 5K. Select
molybdenum 0R filter.

7) Push the Mode selector key (32) until the PDQXDO PRGH is
entered, the auto exp light and GHQVLW\ display are switched off and the
P$V display will indicate P$V

8) Select 25 kV. Using the P$V and P$V buttons , increase the
mAs value to 500 (300 in New York installations, 600 in G.B
installations), and then decrease the display back down to 80. Return
without saving changes by using the Cancel-key .

127( 3UHVVWKHYHUWLFDOGULYHFRQWUROGRZQDQGUHPRYHWKHDVVHPEO\SLQ

9) Test that the c-arm moves nicely down and up in CC and LAT positions.
Verify also that up and down movements of the carriage start slowly and
then speed up as the switch is held down. If not, check that the
counterweight tiles are inserted correct and causing no friction. Check also
that 4 rollers in carriage running in column groove are correctly adjusted.

:$51,1* 'RQRWPHFKDQLFDOO\IRUFHWKHFDUPDVVHPEO\SDVWWKHYHUWLFDOPRYHPHQW
VHQVRUSLHFHVLQORZHUDQGXSSHUSDUWVRIWKHFROXPQ RQERWKVLGHV 7KH
VHQVRUVKDYHLQWKLVFDVHDYHU\KLJKULVNWRJHWEURNHQ

May 2001 22 86501


Diamond Installation Instructions

10) Press the 5(/-footpedal to drive the (&6-mechanism to its home


position and the compression paddle to its maximum opening. At the
maximum opening the compression force reading is zeroed. Otherwise, if
the reading is non-zero, the system will give an audible warning, an error
message and refuses to turn the c-arm.

11) Test both SHGDOV for compression adjustment. Check that the compression
paddle stops automatically after a pressure of approximately 200N (45 lbs)
is reached. If not, refer to Chapter 7.2.6.1, The mechanical compression
limit RU Chapter 7.2.6.2, Electrical compression force limits and speeds.

12) Check that the breast thickness GLVSOD\ and the compression force changes
when driving the compression down.

13) Check that the ILHOGOLJKW is switched ON when the ’down’ pedal is pushed
and goes OFF automatically after approximately 40 seconds.

86501 23 May 2001


Installation Instructions Diamond

14) Install a small bucky and check that the Bucky 18x24 appears on the LCD
display. If not, check that the sensors are sensitive enough and that the
magnet is located at the right place.

15) Repeat step 14 for large Bucky, magnification tunnel, small and large
paddle.

16) Check with both Diamond buckies that they recognize the customer’s
cassettes. If not, adjust from sense potentiometer.

17) Check that the cassettes move to the front and back correct. If not, check
that there is no friction is cassette tray movement. Try this also in CC and
LAT positions.

18) Turn the c-arm and check that the reading in the angle display on the c-arm
changes.

19) For proper filtration, check that the filter is in place and moves when filter
selection is changed.

May 2001 24 86501


Diamond Installation Instructions

 67$5783352&('85(6
Turn the power 2)) from the main power switch.

&$87,21 7KHSXUSRVHRIWKLVSURFHGXUHLVWRHQVXUHWKDWEHIRUHDSSO\LQJIXOOSRZHUWR
WKHJHQHUDWRUDOOYLWDOVLJQDOVDQGDGMXVWPHQWVDUHFRUUHFW8VHODUJHIRFXV

 &+(&.,1*7+(&$66(77(+2/'(56
Install a cassette holder for normal film/screen imaging. Notice two FDVVHWWH
W\SHV: DIN standard or Darkroom. Diamond has been designed to use DIN
standard cassettes.

Figure 2.5. Patient identification window in a DIN standard cassette allows the
use of Alpha ID II or other identification system to print the
patient data onto the film.

Available upgrade kits for cassette holders:

81044 Cassette holder Bucky 18x24 DIN standard


81043 Cassette holder Bucky 24x30 DIN standard
81048 Cassette holder Bucky 18x24 ROC DIN standard
81047 Cassette holder Bucky 24x30 ROC DIN standard
81050 Cassette holder Magnification DIN standard
81052 Cassette holder MammoSpot DIN standard
81054 Cassette holder Spot 1.6 DIN standard

Darkroom cassettes (242x195x19mm for 18x24 and 305x242x19mm for 24x30)


are not manufactured anymore.

86501 25 May 2001


Installation Instructions Diamond

 32:(5,1*83
Turn the unit 21 from the mains switch (16). The following screen should
appear:

Figure 2.6. Diamond Control Panel

Check that the POWER Board LEDs H1, H4, H6, H10, H13, H16 are lit (6 pcs),
on the GENERATOR Board LEDs H1, H2 and H3 are lit.

May 2001 26 86501


Diamond Installation Instructions

 $12'(58192/7$*(&+(&.
1) Disable the radiation as instructed in section 7.5, Disabling x-rays.

127( $OOWHVWLQJVKRXOGEHSHUIRUPHGLQWKHVHUYLFHPRGH8VHPDQXDOH[SRVXUH
YDOXHVDQGVHUYLFHGLDSKUDJP

:$51,1* $0'%2$5'+$6)/2$7,1**5281'2)92/76

2) Make an exposure. After the exposure check that the anode motor plate
stops running (listen to the sound of the rotating anode) and that no error
messages are displayed on Control Panel LCD-screen.

If this is not the case, perform step 3.

3) Press exposure button to run the anode and make an exposure.

Immediately after pressing the exposure button, the testpoints TP1 and TP2
located at the edge of the AMD board (X4 pin 2 and X5 pin 2) show voltage of
230V. With a WUXH 506 ('igital 9olt 0eter) the readout should be 9DF
(around 20-30Vac with a non-true RMS meter) after 1.5 seconds (run voltage).

If not, adjust the anode running voltage with trimmer potentiometer 5.

Figure 2.7. Anode motor drive board (AMD)

86501 27 May 2001


Installation Instructions Diamond

 .95(*8/$7,21&+(&.

:$51,1*
;UD\V

0HDVXULQJIURPWKH[UD\ SULPDU\PHWKRG

kV is set at the factory. For calibration use calibrate meter, which measures from
x-ray. Unfors, RTI, PMX-I/M or similar can be used in calibration.

2VFLOORVFRSHFRQQHFWLRQV VHFRQGDU\PHWKRG

1) Insert the service diaphragm.

2) Connect the oscilloscope probe on the GENERATOR Board to test point


73(KVFB) and ground of the scope to 73. Refer to Figure 2.7 for
testpoint locations.

&KHFNLQJN9XVLQJ'90RURVFLOORVFRSH

1) Set the unit to 28kV and 100 mAs (To get longer exposure time, increase
mAs.

2) Make sure that the exposure is long enough for the DVM reading to
stabilize). One volt is equal to 10 kV. Testpoint TP20 should have a
reading of 2.8V ±0.15V. If not, adjust from trimmer potentiometer R68.

May 2001 28 86501


Diamond Installation Instructions

Figure 2.8. GENERATOR BOARD

86501 29 May 2001


Installation Instructions Diamond

 35(+($7$1'0$5(*8/$7,21&+(&.

 0HDVXULQJWKHSUHKHDWYDOXHV
1) Switch to service mode

2) Connect the oscilloscope probe on the GENERATOR Board to test point


73(MAFB) and ground of the scope to 73. Refer to Figure 2.7 for
testpoint locations.

3) Set the oscilloscope to 1V/div and 20ms/div. Set the trigger to app. +1V
/ positive edge (storage mode).

4) Preheat and compensation values can be changed in service program Pr I


(subprograms 0-5).

5) Turn 21 the unit. Set the unit to 27kV and 16-40 mAs (large focus).

:$51,1*
;UD\V

6) Make an exposure and check the waveform. (refer to Fig. 2.9, 2.10, 2.11).

 3UHKHDWFKDUWV

Figure 2.9. Correct preheat

May 2001 30 86501


Diamond Installation Instructions

Figure 2.10. Preheat rises too fast "Overshoot"

Figure 2.11. Preheat rises too slow

 $GMXVWLQJWKHSUHKHDWDQGP$YDOXHV
For large focus adjustment check that bucky or cassette holder is on. For small
focus adjustment, check that the magnification tunnel is attached.

If the preheat/mA-waveform contains either over- or undershoot at the start-up,


enter the Preheat compensation program PrI04 (large focus) or PrI05 (small focus)
to obtain correct waveform. Adjustment instructions are included in the Diamond
Service Manual section 3.11, Generator.

&$87,21 7RRVORZDULVHWLPHZLOOUHVXOWLQDQHUURUPHVVDJH

86501 31 May 2001


Installation Instructions Diamond

 $77$&+,1*7+(&29(56

 ,QVWDOODWLRQRIWKHWXEHKHDGFRYHU

6.6.1.1 Checking the optoswitches


Check that ParkBack optoswitches are intact. ParkBack movement is
detrimentally affected if any of those is bent.

May 2001 32 86501


Diamond Installation Instructions

6.6.1.2 Attaching top tube head cover


1) Attach the cord. Plug it in so that the metal pin plates face outward,
towards the serviceman. Be careful that the cord would get wrinkled
between the X-ray tube and the tube head cover.

&$87,21 ,IWKHFRQQHFWRULVDWWDFKHGWKHZURQJZD\QRWRQO\WKHEXWWRQVZLOOQRW
ZRUNEXWDOVRWKHV\VWHPLVDWULVNRIJHWWLQJGDPDJHG

2) Lower the top cover (A). Then, lay down the front part (B). Check that
lower tube head cover slides in smoothly into the top cover groove. Then
lower the rear end of top cover down so that the top cover lies all over the
lower top cover (C). Check that the resin covers smoothly the opening
between top and lower tube head cover.

86501 33 May 2001


Installation Instructions Diamond

3) Secure with fingers the front end of the top cover. Do not use any tools or
the paint might scrub off.

4) Secure with a screwdriver the rear part of the top cover.

May 2001 34 86501


Diamond Installation Instructions

 /HIWVLGHFRYHU
1) Align the left side cover with the carriage and attach it with screws.

2) Screw in the frame.

 5LJKWVLGHFRYHU

:$51,1* 5HPHPEHUWRXQSOXJWKHRYHUULGH SLHFH ZKHQ \RX ILQLVK ZLWK WKH V\VWHP


FKHFNLQJ

1) Connect the emergency stop cable coming from the left frame to the one
on the right side frame. Make sure that the cable is well in place and
during movement such as the ECS, does not get caught in any of the
mechanical parts of the system.

86501 35 May 2001


Installation Instructions Diamond

2) Plug the emergency stop cable into the X15 on Power Board. Push it in all
the way until it locks up.

3) Align the right frame openings with the carriage and attach it to the
carriage with screws.

4) Secure the frames at the top.

May 2001 36 86501


Diamond Installation Instructions

 $WWDFKLQJWKHFRXQWHUZHLJKWSURWHFWLRQVKHHW
1) Attach the protective sheet covering the moving counterweight.

The right screw also fixes the upper


part of the cable support bar. Use
only one, left screw to attach the
lower part because the cable support
bar will cover the right screw hole
(circled).

2) Attach the cable support bar as demonstrated in the pictures below. The
same screw fixes the upper part of the cable support bar and the right,
upper part of the counterweight cover sheet.

86501 37 May 2001


Installation Instructions Diamond

The cable support bar covers the hole for right screw of counterweight cover
sheet.

3) Attach the back cover on. Remember to secure the star washers to attain
proper grounding.

127( %HIRUHDWWDFKLQJWKHEDFNFRYHUKDYHWKHSRZHUFRUGWKURXJKWKHUHDUER[
RSHQLQJ

May 2001 38 86501


Diamond Installation Instructions

 &+(&./,6735,257286(

 ,1,7,$/6(77,1*6
Refer to software block diagrams in the Appendix section of this manual.

 'DWH7LPH QRUPDOPRGH


Enter the 6(783-mode in Menu screen:

Then, enter the SETUP-screen:

Choose the 'DWH7LPHoption:

Set the correct time and time, then Exit the screen by pressing the SET -key.

The screen returns to SETUP-screen. Set the hospital information next.

 8VLQJVHUYLFHPRGHLQWHUIDFH
(QWHULQJWKHVHUYLFHPRGH

You can enter into the service mode by pressing kV- and Den-
buttons down simultaneously for 3 seconds until the LCD-display shows the word
"SERVICE".

([LWLQJWKHVHUYLFHPRGH

You can exit the service mode by pressing kV- and Den- buttons
down simultaneously for 3 seconds.

86501 39 May 2001


Installation Instructions Diamond

 3ULQWHU3RUWVHWXS

Enter into the service mode by pressing kV- and Den- buttons down
simultaneously. Then check for and adjust various system settings in service
program $6\VWHP6HWXS&RPPXQLFDWLRQVHWXS (PrA02).

1) Connect a matrix printer or a PC/Laptop to the COM1 serial port.

2) Set the correct parameters.

The serial connection is an ordinary asynchronous receiver/transmitter (RS-232).


The serial port is initialized in Diamond as follows:

- 1200 baud if printer is connected


- 8, N, 2 (8 databits, No parity [always 0], 2 stop bits)

The effect of the ’no parity bit’ and the stop bit is a total of three stop bits, so
receiving end must be configured correspondingly.

May 2001 40 86501


Diamond Installation Instructions

 &RQQHFWLQJILOPPDUNHU ,',, WR'LDPRQGXQLW


ID II film marking or label printing is connected to COM2 serial port. Use the
cable that is supplied with Diamond. For data input, refer to ID II User’s Manual
(1462 B), Chapter 2, Customizing the label.

The printed label displays the following information:

86501 41 May 2001


Installation Instructions Diamond

 &+(&.,7(06
In case of normal service or maintenance:$OOUHTXLUHGFKHFNVDOLJQPHQWVDQG
DGMXVWPHQWV PXVW EH SHUIRUPHG LI QHFHVVDU\ 6HH DGGLWLRQDO LQIRUPDWLRQ
DERXWWKHWHVWVLQ6HUYLFH0DQXDO

 /LJKWILHOGLQWHQVLW\
The light intensity is adjusted in the service program PrF03. If adjustment is
necessary, a lux-meter is recommended for exact adjustment.

Adjust the light intensity at the top of a cassette holder centers of the 4 quadrants
to a minimum average of 160 LUX (ISLC’s). Take note of the effect of the
examination room lighting unless it can be totally switched off.

Outside USA, if a LUX meter is not available, set the collimator lamp voltage
to 11 -11.5 VDC, measured at the Collimator board TP9 to TP12 or at conductor
stands (marked by arrows in the image below).

The halogen lamp lifetime increases when voltage is decreased and vice versa.

Figure 2.12. Collimator board

:$51,1* &KHFNWKDWWKHPLUURURUODPSLQWKHFROOLPDWRUKDVQRWEHHQGDPDJHGGXULQJ
WUDQVSRUWDWLRQDQGFDQEHXVHG

May 2001 42 86501


Diamond Installation Instructions

 ;UD\ILHOGDOLJQPHQW
In the service mode, select the manual mode with 25 kV and 20 mAs.

Place a loaded cassette on top of the cassette holder so that the cassette overhangs
20 mm on the chest wall edge of the holder.

20mm
ruler
acrylic
cass. holder top
cassette
image receptor

Mark the chest wall edge of the film with a ruler placed 20 mm back from the
chest wall edge of the film. (The ruler now shows the actual chest wall edge.)

Make an exposure and develop the film. The exposed area between the ruler and
the chest wall edge of the film must be less than 6 mm and more than 1 mm in
USA, less than 3 mm and more than 0 mm in Europe.

$GMXVWPHQWRIWKH[UD\ILHOG

1) Adjustment screw regulates the front edge of x-ray field

2) Collimator rear plate and side plates regulate side and rear edges of the x-
ray field. Refer to Diamond Service Manual, Chapter 3.2.6, F: Collimator
and readjust the values for step motor movements in service programs
PrF01 and PrF02.

Check separately for each cassette holder (little Bucky, large Bucky,
Magnification, Contact etc.)

86501 43 May 2001


Installation Instructions Diamond

 /LJKWILHOGDOLJQPHQW
The x-ray field adjustment must be checked and aligned prior to the light field
alignment. Place a loaded cassette in the holder and position the corresponding
diaphragm.

Enter the service mode by pressing the N9 and GHQVLW\ buttons simultaneously
and holding them the service screen appears.

Select manual mode at 25 kV and 20 mAs.

1) Place a sheet of white paper on the cassette holder so that the light field
can be seen clearly
2) Place some coins at the corners of the illuminated area on the cassette
holder.
3) Make an exposure and develop the film.

The coins outlined on the film should not be more than 5 mm out of the exposed
field and sum of misalignments on opposite sides not more than 10 mm.

If the light field needs adjustment, refer to Diamond Service Manual (86502),
Chapter 5.4.5, Testing the x-ray field illumination.

 )LOPSURFHVVRUHYDOXDWLRQ
Test equipment requirement: Sensitometer, Densitometer

1) Turn ON the film processor and let it warm up for at least half an hour.

2) Run a clean-up film through the processor. This is just a piece of regular
film, which will pick up dirt, etc., which would appear on your film.

3) Obtain a sheet of the same film type to be used in your Diamond. Select
the color on the sensitometer that matches the color emitted by the
intensifying screen used.

4) Ensure that the emulsion side of the film is towards the light source on the
sensitometer. Expose the two opposite long edges of the film, waiting 10
seconds between exposures, and run it through the processor.

5) Measure the film density with densitometer from the unexposed part of
film. This is the base fog and O.D. should be less than 0.20.

127( $OZD\V UXQ WKH ILOP WKURXJK WKH SURFHVVRU LQ WKH VDPH RULHQWDWLRQ DQG
ORFDWLRQ

6) Find the step, which has the density closest to 1.0. This step number is the
63((',1'(;Record the average density and step number on the film
and in tables below.

May 2001 44 86501


Diamond Installation Instructions

7) Proceed one step above and one step below the step of Speed Index.
Measure and record both densities.

8) Subtract the density value of the step below, from the density value of the
step above the speed index step. This result is the &2175$67,1'(;
Record this on the film and in tables below.

Mark the DATE on this file for future reference.

DATE: TIME:
FILM TYPE: SCREEN TYPE:
WATER TEMP: DEV TEMP:
GROSS FOG: (max 02) SPEED INDEX:
CONTRAST INDEX:

Table 2.1 Film processor evaluation

Notice that the typical values depend on the film/screen combination and consult
the corresponding manufacturer’s instructions.

 &DUPDQJOHGLVSOD\FDOLEUDWLRQ
Run c-arm to various angles and verify if the angles are correct. If not, enter the
service program PrE01 and calibrate the c-arm angle:

1) Drive the C-arm to 180°-angle with manual rotation keys.

2) Press Accept -key.

3) The c-arm drives to 0°-calibration.

4) In case the calibration succeeded, the screen returns to (&DUPURWDWLRQ.

If the calibration is successfully performed, the LCD screen then indicates '3DVV 
In case the calibration was not successful, the LCD screen then indicates )DLO 

5) Repeat the calibration procedure.

6) If the calibration still was not successful, drive the c-arm to from 180° to
0° and then to 180° on the other side. When c-arm passes the 0° it helps
to calibrate.

86501 45 May 2001


Installation Instructions Diamond

 &RPSUHVVLRQIRUFHGLVSOD\FDOLEUDWLRQ
Check with bathroom scale that if the compression force display values are
correct. If not, enter the service program PrG04 and calibrate the compression
force display:

1) With no compression at all, set the 0N force with the Accept -


key.

2) Using the bathroom scale, compress to 200N and press Accept -


key.

3) Drive the C-arm to 90° angle and press Accept -key.

4) In case the calibration succeeded, the screen returns to *&RPSUHVVLRQ


IRUFH.

5) In case the calibration was unsuccessful, the system asks to repeat the
calibration procedure.

7.2.6.1 The mechanical compression limit


In order to prevent excessive compression on the breast, the mechanical
compression limit can be adjusted clutch can adjusted by mechanical clutch
adjustment screw. Check first with a bathroom scale if the mechanical
compression limit conforms to the regulations in your country.

127( 7KHPD[LPXPPDQXDO PHFKDQLFDO FRPSUHVVLRQIRUFHPXVWQRWH[FHHG1


E\,(&VWDQGDUG

If the mechanical compression limit needs adjustment, refer to Diamond Service


Manual (86502), Chapter, 6.4.11, Adjustment of the mechanical compression
limit.

7.2.6.2 Electrical compression force limits and speeds


Diamond incorporates 6 electrical compression force limit steps that may be
adjusted in service program PrG05. These steps are the ones that user can select
at c-arm control panel.

127( 7KHPD[LPXPPRWRUL]HG HOHFWULFDO FRPSUHVVLRQIRUFHLVOLPLWHGWR1E\


,(&VWDQGDUG

Also the compression speeds may be adjusted in service programPrG01.

May 2001 46 86501


Diamond Installation Instructions

 &RPSUHVVLRQWKLFNQHVVGLVSOD\FDOLEUDWLRQ
1) Check the thickness display at 4cm with different compression paddles.

2) When the calibration is performed at some specific force, it forces the


compression arm to move past the horizontal line in relation to the upper
surface of the acrylic (the long dark gray bar).

In order to achieve a proper thickness for clinical use, Diamond allows ± 50mm
extra movement past the horizontal level in relation to the acrylic. This
compression thickness is separately calibrated for each paddle (PrG02) and holder
(PrG03) in service programs. Use proper (70N) compression with the acrylics.

3) Different compression paddles have the surface at different height that


applies the compression to the breast.

Diamond recognizes various paddles, but if necessary, the height differences can
be compensated in service program PrG03.

86501 47 May 2001


Installation Instructions Diamond

 N9P$VDQGH[SRVXUHWLPHWHVW
If it becomes necessary to test kV, mAs and exposure time, refer to Diamond
Service Manual, Chapter 5.1.6, kV, mAs and exposure time test.

 *ULGDOLJQPHQW
The grid is aligned and the speed is adjusted at the factory and should normally
not be adjusted at the site. If adjustment and / or alignment is necessary, however,
follow the instructions in Diamond Service Manual (86502), Chapter 6.4.21, Grid
alignment.

 9ROWDJHFKHFN
The internal voltages can be checked according to the list in Diamond Service
Manual (86502), Chapter 4.13, Fuses and voltages.

May 2001 48 86501


Diamond Installation Instructions

 $(&$'-8670(17
Check the AEC operation before handing the unit over to the customer. Tables
provided within chapters or an AEC calibration table in appendices can be used
to record changes in settings and resulted O.D. values.

Test equipment needed: Sensitometer, Densitometer, Acrylic blocks, Film &


Cassette. ATTACH THE BUCKY DEVICE.

Enter into the service mode by pressing kV- and Den- buttons down
simultaneously for 3 seconds.

 4XLFN$(&WHVW
Testing procedure:

1. Attach the cassette holder.


2. Position, for example, an acrylic plate carefully so that it covers the AEC
detector.
3. Select such a kV that will produce the clinically suitable optical density
at 4cm thickness.
4. Make an exposure using the Auto Time-mode using C1 detector.
5. Develop the film.
6. Measure the film density. If the density is not what you wish it to be,
reprogramming of AEC using densitometer is recommended, but record
system settings before changing anything.

127( &DUU\RXWWKLVWHVWRQFHDZHHNXVLQJWKHVDPHSURFHGXUHDOOWKHWLPH

 5HFRUGLQJWKHV\VWHPVHWWLQJV
The system settings can be exported in hexadecimal format and can be viewed and
stored in Windows Hyperterminal. For HyperTerminal properties setting and
about how to capture, refer to Chapter 3.4, Using HyperTerminal interface with
Diamond.

86501 49 May 2001


Installation Instructions Diamond

 %DVLFDGMXVWPHQW
The gain adjustment affects the operation of the Diamond detectors. One detector
(C1) acts as a benchmark and it is adjusted first. Then the other detectors are
adjusted to equalize the output of all detectors in relation to C1. The graph below
describes the detector numbers. It is useful to know the detector numbers when
the exposure data is examined through hyperterminal.

4=C4

5=C3
7=
L2

6=C2 R2
3=

0=L1 1=C1 2=R1

Figure 2.13. Diamond detectors

 Ask the customer how dark images they want. Light images have less
information and modern viewboxes can display dark images better than
before. Typical O.D. values are for example 1.4-1.6

 Enter the service program PrD02 and check that the overall AEC gain
parameter ’J’ is set to . (In order to start from the very beginning, the
thickness compensation and kV compensation values can be set to 0 and
r1, r2 and detector-specific values to 1.)

 Make an exposure in $XWR7LPH mode using the clinically typical kV (e.g.


28kV) with a 0RO\EGHQXP filter and a PP acrylic. Use a compression
paddle and a Bucky with a loaded cassette in it. Detector must be in center
position &. Also, remember to place the white plastic cover before
exposure, if that has been removed for some reason.

acrylic
cass. holder top
cassette
image receptor

Figure 2.14. Placement of the acrylics (and phantoms!) onto the cassette holder

 Develop the film and mark the optical density of the film (=O.D) to the
graph next page. The higher the O.D., the darker the image and vice versa.

May 2001 50 86501


Diamond Installation Instructions

 Alter the AEC Gain parameter ’J’ in the service program PrD02 to 1.10 (10
% increase), reload the cassette with a new film and make a new exposure.
Develop the film and mark the optical density of the film (=O.D) to the
graph below.

 Draw a line between the two points and read the corresponding GAIN for
the wanted optical density (greater O.D. values correspond to darker
images). Notice that the object is to find the right gain, so the initial gain
value does not have to be 1.00. For example, if 1.00 falls below and 1.10
surpasses the desired O.D., then the correct gain should be some value in-
between. Draw new lines and estimate the correct gain.

 Alter the AEC Gain parameter ’J’ in the service program PrD02) to obtain
the desired O.D. (for example 1.6). Diminishing ’J’ will increase the O.D.
and vice versa.

 Make one control exposure with Auto Time-mode the clinical kV (e.g.
26kV) Molybdenum filter with 40 mm acrylic and measure the O.D. to
check the adjustment. Make note of the exposure time for the thickness
compensation in Chapter 7.3.6.

([SRVXUHWLPH

86501 51 May 2001


Installation Instructions Diamond

 $XWRPDWLF$(&GHWHFWRUJDLQDGMXVWPHQW
In case the values for individual detectors must be adjusted, follow the
instructions in Diamond Service Manual (86502), Chapter, 5.2.3, Automatic AEC
detector gain adjustment.

&KHFNLQJRSWLFDOGHQVLWLHVIRUHDFKILOPW\SH

Every detector should be checked and adjusted to each film. Films can be
different.

1) Make an exposure with a small bucky at clinical kV (for example 27kV)


and 40mm acrylic.

2) Develop the film.

3) Place the film on the AEC detector location figure (included in the
appendices).

4) Draw detector ellipse on the film and mark its name next to it (not inside
of it).

5) Measure a detector-specific optical density in the middle of the ellipse and


write it next to the ellipse name.

For example the detector C4 would have the following notations (x marks the
O.D. reading):

6) If the acquired O.D. is not within the limits, adjust the values as instructed
in Diamond Service Manual (86502), Chapter 5.2.4, Adjustment of an
individual detector.

May 2001 52 86501


Diamond Installation Instructions

 2EMHFWWKLFNQHVVFRPSHQVDWLRQ
Object thickness compensation adjusts linearity that affects the curve form

1) Make three exposures with different phantom thicknesses (for example


20mm, 40mm, 60mm to get short, medium and long exposure time) and
record the corresponding exposure times and O.D:s. Recommended
exposure times would be appr. 500ms, 1sec and 2 second exposures
respectively.

2) Use the clinically typical kV (e.g. 26kV) and Auto Time-mode for all
exposures.

3) Measure the optical density for both molybdenum and rhodium filter. Plot
the values in tables below.

ACR mm O.D. Bucky) O.D. (Contact) O.D. (Magnification)


20 / / /
40 / / /
60 / / /

1.90
1.80
1.70
1.60
1.50
1.40
1.30
1.20
1.10
mm 20 40 60 70

86501 53 May 2001


Installation Instructions Diamond

4) Mark the points to the O.D vs. exposure time graph. Draw a line through
the three points. They should form a horizontal line within “2'.

2'YDOXHVZLWKLQWKHOLPLWV

If the O.D. values for 20mm, 40mm and 60mm thicknesses fall within
these limits, no thickness compensation is required.

For example, if O.D. for 40mm thickness is 1.6, other O.D. values (for
20mm and 60mm thicknesses) should fall between 1.45 and 1.75. If not,
then adjustment is required.

2'YDOXHVRIIWKHOLPLWV

If the optical density with 20mm acrylic or 60mm acrylic are different than
the optical density with than 40 mm acrylic, then thickness compensation
parameters have to be adjusted as instructed in next chapter.

Table 2.2. Object thickness compensation

May 2001 54 86501


Diamond Installation Instructions

 7KLFNQHVVFRPSHQVDWLRQ
The optical density can be adjusted by
thickness specific compensation parameters
which affect the the exposure by increasing
or decreasing the exposure time for a
specific thickness value. Curves can be of
various shapes (graph in the right).

The graph below is a representation of one type curve, which is adjusted to


exemplary target O.D. of 1.6. Adjust the O.D. according to customer’s desires.

1) If the O.D. at thickness of 4cm is at target, then proceed to next step. If


not, adjust the parameter ’J’ (instructions in Chapter 7.3.3, Basic
adjustment) so that you obtain the required O.D. at 4cm.

9DOXHVIURPWKH2'YVH[SRVXUHWLPHJUDSKLQSUHYLRXVFKDSWHU

2) Look for the H[SRVXUHWLPH with which the required O.D. with 40mm
acrylic was obtained (in Chapter 7.3.3). Write this time on the line below
or to the AEC Calibration table in the Appendix section:

Exposure time = This value is ’3’.

3) Look for the O.D. for the shorter exposure time (for example the 20 mm
Acrylic exposure time). Divide the required O.D. with this value. This
value will provide the direction of the adjustment. Test always for exact
values, because the film, processors and processing conditions vary.

This value is ’U’

86501 55 May 2001


Installation Instructions Diamond

4) Look for the O.D. for the longer exposure time (for example the 60 mm
Acrylic exposure time). Divide the required O.D. with this value.

This value is ’U’

5) Enter the service program PrD03 and choose 7KLFNQHVVFRPSHQVDWLRQ.

6) Press the Next -key on until the ’3’ for ’rotation Point’ appears in
screen, then scroll up or down up to set the H[SRVXUHWLPH
which has the correct O.D. from the calculation performed above. This is
the exposure time, which is not affected by thickness compensation (=the
rotation point). When the parameter is set, proceed to set other parameters.

7) Press the Next -key on until the ’U’ appears in screen, then scroll
up or down up to set up the ’U’ parameter. This parameter
gives the amount of rotation at exposure times shorter than the rotation
point ’3’. For example if the short exposure has 12% too dark, a density
enter 0.88 for ’U’.

8) Press the Next -key on until the ’U’ appears in screen, then scroll
up or down up to set up the ’U’ parameter. This parameter
gives the amount of rotation at exposure times longer than the rotation
point ’3’. For example, if the long exposure has 12% too light a density
enter 1.12 for ’U’.

9) Check O.D. after changing the ’U’ and ’U’-parameters. Readjust, if


necessary.

&$87,21 3HUIRUPWKHWKLFNQHVVFRPSHQVDWLRQZLWKERWKILOWHUV 0R5K DQGDOOFDVVHWWH


WXQQHOV %XFN\FDVVHWWHKROGHUPDJQLILFDWLRQWXQQHO VHSDUDWHO\ HVSHFLDOO\
WKHFRPELQDWLRQWKDWWKHFXVWRPHUVWDUWVWRXVH '2127FKDQJHWKHJDLQ
SDUDPHWHUDQ\PRUH$GMXVWWKHGHQVLW\FRUUHFWLRQLQWKHVHUYLFHSURJUDP
3U' WR EULQJ 2' WR VDPH OHYHO ZLWK RWKHU ILOWHUFDVVHWWH KROGHU
FRPELQDWLRQV

10) Accept and exit the screen by pressing the SET -key.

May 2001 56 86501


Diamond Installation Instructions

)LQHWXQLQJRIWKHWKLFNQHVVFRPSHQVDWLRQ

127( )LQHWXQLQJLVRSWLRQDODQGPD\RQO\KDYHPLQRUHIIHFWWRFRPSHQVDWLRQ,W
LV QRW QHFHVVDU\ WR SHUIRUP LI WKH VHQVRU DGMXVWPHQW IRU U DQG U LV
DYDLODEOH

1) Return to the service mode and make two exposures with Auto Time-
mode N9 so that very long exposure time (four times longer than
exposure time for 40 mm Acr = for example 4-5 sec.) and very short
exposure time (3 of exposure time for 40 mm Acr = for example 250-300
ms) are achieved. Write all exposure data to the AEC Calibration table in
the Appendix section.

2) Develop the films and record the optical density of the films (=O.D).

3) Select the adjustment direction of ’6’ (affects the compensation slope at


exposures shorter than rotation point ’3’, when a very thin breast is
examined) and ’6’ (affects the compensation slope at exposures longer
than rotation point ’3’ when a very thick breast is examined) from the table
below.

722/,*+72' 722'$5.2'
U<1 increase S1 decrease S1
U>1 decrease S1 increase S1

U<1 decrease S2 increase S2

U>1 increase S2 decrease S2

Table 2.3 O.D. corrections

’6’ and ’6’ can have values 0, 50, 100, 150...up to 700. The maximum
compensation is 140% or 60% of O.D. in ’3’ (rotation Point).

4) Enter the service program PrD03 and choose 7KLFNQHVVFRPSHQVDWLRQ.

5) Press the Next -key on until the ’6’ appears in screen, then scroll
up or down up to set up the ’6’ parameter. This parameter
affects the compensation slope at exposures much shorter than rotation
point ’P’.

6) Press the Next -key on until the ’6’ appears in screen, then scroll
up or down up to set up the ’6’ parameter. This parameter
affects the compensation slope at exposures much longer than rotation
point ’P’.

86501 57 May 2001


Installation Instructions Diamond

7) Check O.D. after changing the ’6’ and ’6’-parameters. Readjust, if


necessary.
8) Accept and exit the screen by pressing the SET -key.

May 2001 58 86501


Diamond Installation Instructions

 N9FRPSHQVDWLRQ
Optical density can also be adjusted by kV specific compensation parameters,
which affect the exposure by increasing or decreasing the exposure time for a
specific kV value.

1) Check the linearity with a couple of control kVs (e.g. low (24kV), medium
(28kV) and high (32kV), using Molybdenum and Rhodium filter if
available). Use Auto Time-mode and thick enough an acrylic to produce
a clinically typical exposure time at each kV (appr. 1 sec.).

kV O.D. (Bucky) O.D. (Magnification) O.D. (Contact)


23 / / /
24 / / /
25 / / /
26 / / /
27 / / /
28 / / /
29 / / /
30 / / /
31 / / /
32 / / /
33 / / /
34 / / /
35 / / /

86501 59 May 2001


Installation Instructions Diamond

2) Mark the received O.Ds for Bucky to the table below. x = molybdenum
filter, o = rhodium filter

3) Mark the received O.Ds for Magnification to the table below. x = molybdenum
filter, o = rhodium filter

May 2001 60 86501


Diamond Installation Instructions

4) Mark the received O.Ds for Contact to the table below. x = molybdenum
filter, o = rhodium filter

5) Connect the received O.D.:s with a line. If the line shows a regular pattern
close to the required O.D. line in that chart (one of the horizontal lines at
regular intervals) then the O.D. is alright. If not, measure optical density
for each kV, using Molybdenum and Rhodium filter if available.

86501 61 May 2001


Installation Instructions Diamond

6) Plot the received O.D.:s on the chart and check which ones need
adjustment (graph is for reference only). If the line is not linear, check all
kVs.

7) Enter kV Compensation in service program PrD03.

8) Perform necessary changes for all cassette holder and filter combinations.

If the required O.D. is for example 1.6 at 26kV and 1.3 at 23kV, increase in
service program PrD03 (23kV selected), the factor to 1.2. Exposure again and
check the O.D. Test every other kV values (…23, 25, 27, 29…) and then estimate
factors for values in between (…24, 26, 28…).

 $XWRN9SURJUDPPLQJ
Adjust density compensations for each filter and cassette holder combination in
Auto kV mode. Set the new values in the service program PrD03. Usually for the
user the density step change is 10% or 12%, but in the service program the step
is half of this.

May 2001 62 86501


Diamond Installation Instructions

 7DUJHWWLPHV
In Auto-kV mode, Diamond determines whether the calculated exposure time
with the initial kV conforms to target time curve. If the exposure time (function
of compression thickness) is too long or short, Diamond then increases or
decreases the kV value. In order to produce the target exposure time, Diamond
interpolates between two points of the linear target time curve to find the required
kV-value.

If initial kV is badly off, the necessary change in kV might be more than the kV
change limit in service program PrB06 allows. In this case Diamond will use the
kV that is a sum of the initial kV and the maximum allowed change (+/-) in kV.
The initial kV adjustment instructions are in Chapter 7.3.10, Initial kV tables. The
kV change limit can adjusted in service program PrB06.

The curve slope is defined with two thickness points, one at 2cm and another one
at 6cm. Target times can be set in service program PrD05. As the used target time
depends on the country, film and film developer, the parameters have to adjusted
according to local conditions and requirements. For reference only, some users
have used values from 0.4s to 0.5s for lower limit and from 1.0s to 1.6s for upper
limit. If the lower limit is less than 0.2s, the exposure is too quick and grid lines
appear in the image.

The graph below describes how changes in target time limit affect the curve slope.
If the target time limits are set at the maximum values, the result is the steep curve
$. When the lower limit is raised up the and higher limit lowered down, the curve
becomes flatter (% and &) and vice versa.

86501 63 May 2001


Installation Instructions Diamond

May 2001 64 86501


Diamond Installation Instructions

 ,QLWLDON9WDEOHV
The initial kV is the first kV that Diamond uses for a certain thickness. Initial kV
tables can be set in service program PrD06 and the kV change limit in PrB06.
Each cassette holder can have its own Initial kV for various thicknesses. The ideal
case would be that initial kV-values would conform to target time curve values
and that there would no need for kV changes.

1) Expose in Auto kV-mode and check the used kV-values at 2cm, 4cm, 6cm
and 8cm thicknesses.

2) In service program PrD06, select Bucky, Magnification or Contact.

3) Scroll through thicknesses (at 0.5cm intervals) and set initial kV for each
one of them. Set the values from step 1 and make a linear estimate for
other thicknesses.

,1,7,$/.97$%/(
7KLFNQHVV &DVVHWWH &DVVHWWH %XFN\QHZ %XFN\ 0DJQQHZ 0DJQ

GLVSOD\ WXQQHO WXQQHO YDOXHV GHIDXOW YDOXHV 'HIDXOW

QHZYDOXHV GHIDXOW YDOXHV YDOXHV0R

YDOXHV 0RDQG5K DQG5K

0RDQG5K

 23 23 25
 23 23 25
 23 23 25
 23 23 25
 24 24 25
 24 24 25
 24 24 25
 25 25 26
 25 25 26
 25 25 26
 26 26 27
 26 26 27
 26 26 28
 27 27 28
 27 27 28
 27 27 29
 28 28 29
 28 28 30
 29 29 30
 30 30 30

Table 2.4 Initial kV table

86501 65 May 2001


Installation Instructions Diamond

4) Repeat the procedure for each cassette holder and filter combination.

5) Adjust the kV change limit in service program PrB06.

If the initial kV-values are badly off, the target times are not reached because the
required kV change exceeds the adjustment range for kV change limit. Diamond
will make the exposure but the exposure time will be different than what target
time curve suggests. High kV-values tend to reduce contrast, but it is worthwhile
to check if film manufacturer recommends higher initial kV-values.

 $(&DGMXVWPHQWTXLFNIL[JXLGH
Before performing any AEC adjustment work, make sure that kV values are
alright.

$(&DGMXVWPHQWTXLFNIL[JXLGH
3UREOHP 3UREDEOH5HDVRQ $FWLRQ &RPPHQWV

No contrast with large Full AEC kVs with large Increase the 7DUJHW WLPH with Results in lower kV
breasts breasts too high. large breast thickness. (PrD05) with large breasts

No contrast with small full AEC kVs with small Increase the 7DUJHW WLPH with Results in lower kV
breasts breasts too high. small breast thickness. (PrD05) with small breasts.

mAs too high, long Full AEC kVs low Decrease the 7DUJHWWLPH values Low kVs yield long
exposure times (PrD05) mAs

Too early termination Start kV too low and/or Change the AEC kV table
of the radiation. maximum allowable kV (PrD03) or increase the kV
change too small. change value

Table 2.5 AEC adjustment quick fix guide

&$87,21 7KH$(&JDLQLVDGMXVWHGIRURQHILOWHUDQGRQHFDVVHWWHKROGHUQRUPDOO\IRU
WKHPRO\EGHQXPILOWHUDQGZLWKWKHEXFN\7KHULJKWRSWLFDOGHQVLWLHVIRUWKH
UHVW RI WKH FDVVHWWH KROGHUV ZLWK 0RO\EGHQXP ILOWHU DQG IRU DOOFDVVHWWH
KROGHUVZLWKUKRGLXPILOWHUDUHGRQHDVGHVFULEHGLQ&KDSWHU

 3ULQWRXWRIV\VWHPGDWD
Diamond settings (Battery RAM) can be downloaded with a HyperTerminal® by
writing EUDP in command prompt (a sample included in the appendices).

For AEC and detector-specific data, write SDUDP DHFGXPS in the command
prompt and make an exposure.

For instructions on how to use the HyperTerminal® refer to Diamond Service


Manual, Chapter 3.4, HyperTerminal interface for Diamond. Settings for
hyperterminal software are 19200 bauds, 8 bits, no parity, 2 stop bits and no
handshake.

May 2001 66 86501


Diamond Installation Instructions

 6<67(0&86720,=$7,217286(535()(5(1&(6
For more customized use of Diamond, refer to Diamond Service Manual (86502)
that has an extensive list of the screen parameters that can be set by the service
technician.

 ',6$%/,1*;5$<6
The emission of x-rays must be disabled in some cases (for example at
exhibition).

127( 7KHIROORZLQJWKUHHLWHPVGLVFRQQHFWVDOOSRZHUIURPWKHJHQHUDWRUEXWWKH
V\VWHPFDQEHVWLOORSHUDWHG7KLVFDQEHGRQHIRUH[DPSOHZKHQLQVWDOOLQJIRU
DQH[KLELWLRQZKHUH;UD\VDUHQRWDOORZHG

 'LVDEOLQJWKHUDGLDWLRQ
1) Turn OFF the unit
2) Unplug connectors ; and ;from the GENERATOR Board (a four-pin
connector).
3) Unplug ; of the Power Board
4) Set the jumpers ---- on the GENERATOR Board to  SRVLWLRQ
[right-hand, non-default position].

 (QDEOLQJWKHUDGLDWLRQ
1) Turn OFF the unit
2) Plug in the connectors ; and ;onto the GENERATOR Board (a four-
pin connector).
3) Plug in ; of the Power Board
4) Set the jumpers ---- on the GENERATOR Board to  SRVLWLRQ
[left-hand, default position]

 $33(1',&(6
The Diamond software block diagrams, factory default settings list, AEC detector
location figure and an AEC calibration list are attached as an appendix at the end
of this document.

86501 67 May 2001


Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

Normal mode: CONTROL PANEL BUTTONS

preprog Bucky 18x24


calibrating.....
CANCEL Auto Auto Auto SET
Dose Filter Rel

Auto kV mode Auto Time mode Manual mode


Button1 Button3
Button2

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
use Modes
se - MENU - 01
in in u
QA Setup

Not
(Quality Assurance)

Not QA MODES SETUP

Button1

Button1

Button3
exposure qa mode

- QA MODE - 10
Place a phantom

SET
CANCEL and expose the film

Button2
sensitometer mode

Scroll and select one


of the following modes Label Text
Date/Time
Abort QA MODE ? 12 Display
- QA MODE - 13 no changes made - SETUP - 30
Pass
CANCEL SET to abort Label Text
EXP

New IQC coefficient CANCEL to return > Date/ Time <


1.05 down Accept up

SET
- QA MODE - 11
Record the O.D
Button3/ 1.45 CANCEL
SET down Accept up Discard changes

use
Record the optical Hospital Name
31 CANCEL
Department
density: 1.xx
Codein
Not
timer display shows
Save changes
iqc coefficient down nextup SET
- QA MODE - 14
Error Scroll and select one
IQC coefficient of the following modes PreProg1
out of range PreProg2 - DATE/ TIME - 32
You have changed
Specimen Hour: 00 Min: 00
system settings Save changes
yy: 00 mm: 01 dd: 29 SET
press SET to accept SET
down next up
CANCEL to discard

- MODES - 20
Specimen Discard changes
CANCEL
> PreProg1 <
up PreProg2 down
- DISPLAY - 33
Brightness: 5 Save changes
Contrast: 5
SET

down next up

Allowed brightness values 0-5 (00-15)


Allowed contrast values 0-5 (15-10)
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

Service mode: CONTROL PANEL BUTTONS

preprog Bucky 18x24


calibrating.....
CANCEL Auto Auto Auto SET
To enter service mode: press kV- and den- simultaneously for 3 secs Auto kV mode Auto Time mode Manual mode Dose Filter Rel

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24 Button1 Button3
Button2
calibrating..... service calibrating..... service calibrating..... service
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults
C: Diagnostics
- GROUPS - 000 D: AEC
Calibration E: C-arm Rotation
> AEC < F: Collimator
up Accept down G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement
L: Motor tests

Scroll and select one of ScrollScroll


and select
and select
one ofone of Scroll
Scroll and
and select
select one
one of of Scroll and select one of Scroll
Scrolland
andselect
selectone
oneofof Scroll
Scroll
and
and
select
select
one
one
of of
the following procedures the following
the following
procedures
procedures the
the following
following procedures
procedures the following procedures the
thefollowing
followingprocedures
procedures the
the
following
following
procedures
procedures

- DIAGNOSTICS -C00 - C-ARM ROTATION-E00 - C-ARM


- COLLIMATOR
ROTATION-500
- F00
- SYSTEM - A00 - SYSTEM
- DEFAULTS
- 300- B00 - DEFAULTS - 300 - AEC - D00 - AEC - 200
Ch. Inputs Angle Calibration Angle
RearCalibration
Plate
Unit Info Unit Thickness
Info Thickness AEC gains AEC gain
> < Pr### > > <Force
Pr###
< Pr### > Ch. <Bucky
> Force Pr###< Pr### > Densities < Pr### > Rotation Speed < Pr##
> Densities < Pr### > Rotation
> Side Plates
Speed <<Pr##
F02
up Accept down up Accept down up up Accept
Acceptdown
down
up Accept down up up
Accept Accept
down down up Accept down up Accept down up Accept down

B01: Start up settings StartCheck


C01: up settings
Inputs Pr301 Check Inputs E01: Angle calibration Pr301
F01:Angle
Rear Plates
calibration
PrA01 Sys params Pr401 Unit info Pr302 PrD01 AECBucky
Check setup Pr201 AEC gainSpeed
B02: Bucky settings AutoCheck
C02: kv settings
Bucky E02: Rotation Pr302
F02:Rotation
Side Plates
Speed

use
PrA02 Comm setup Pr402 Pr303 PrD02 AEC gains Pr202 Densities
B03: Mag settings AutoPrint
C03: time Info
settings E03: Projections Pr303
F03:Projections
Collimator Light
in
Pr403 Thickness compensation PrD03 Densities Pr203
B04: Contact settings Manual
C04: mode
Check settings
Error Codes E04: Auto drive compensation
Thickness

Not
Pr404 Detector compensation PrD04 Detector compensation Pr204 Detector compensation
B05: QA mode settings Stereo
C05: settings
Check Anode
Pr105 Filter movement PrD05 Target Times
B06: PreProg1 settings QA mode settings
Pr106 Tube movement PrD06 initial kV-tables
B07: PreProg2 settings PreProg1 settings
se PrD07 QA- coefficient history
d
B08: PreProg3 settings PreProg2 settings No
tu
PrD08 Detector adjustment
B09: PreProg4 settings

Scroll
Scrolland
andselect
selectone
oneofof Scroll and select one of Scroll and select one of Scroll and select one of Scroll and select one of Scroll and select one of
the
thefollowing
followingprocedures
procedures the following procedures the following procedures the following procedures the following procedures the following procedures

- -COLLIMATOR
COMPRESSION - - ?00
G00 - FILTER - H00 - GENERATOR - I00 - TUBE MOVEMENT- J00 - STEREO MOVEMENT- K00 - MOTOR TESTS -L00
Angle
Auto Calibration
Release Generator Setup TEST OPTION
> >Rotation
Thickn Calibration
Speed < Pr##
< > Filter Calibration < > Preh Adjustment < > Rotation Speed < > Stereo Speed < > CARM <
upup Accept
Accept down
down up Accept down up Accept down up Accept down up Accept down up Accept down

Pr301 Check Inputs


Pr301
F01: Compression
Rear Plate Settings H01: Filter Calibration I01: Generator Setup J01: Speeds K01: Speeds PrL01 CARM

use
Pr302 Check Bucky
Pr302
F02: Thickness
Side PlatesPaddles I02: Preheat Calibration LF K02: Auto drive PrL02 TUBE
Pr303
Pr303
F03: Thickness
Collimator Cassette
Light hold I03: Preheat Calibration SF
in PrL03 COMP

Not
F04: Force Calibration I04: Preheat Compensation LF PrL04 EPS

se
F05: Force Limits I05: Preheat Compensation SF PrL05 STEREO

n u
PrL06 REAR P

i
Not
PrL07 SIDE P
PrL08 FILTER
PrL09 CAS
PrL10 GRID
PrL11 MIRRO
PrL12 LIGHT
PrL13 PLATE
PrL14 TEST OPTION
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

Service program Groups: - System - Group A CANCEL


preprog Bucky 18x24
calibrating.....
Auto Auto Auto SET
Dose Filter Rel

Button1 Button3
Button2

Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults
C: Diagnostics
- GROUPS - 000
D: AEC
Cancel Calibration
> AEC <
E: C-arm Rotation
up Accept down F: Collimator
G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement

Scroll and select one of


the following procedures CANCEL

- SYSTEM - A00
Unit Info
> < Pr###
up Accept down

System Parameters PrA01 Hardware Configuration PrA02


PrA01 Sys params
PrA02 Comm setup
- SYS PARAM - A01 - COMM SETUP - A03
Serial number: 12345 Label out : Com1
Country Code: 123 Label Type: 2
up Accept down up Accept down

Pr301 Check Inputs Pr301 Check Inputs


Parameters:
Pr302 Check Bucky Parameters:
Pr302 Check Bucky

use
Serial n:o: 12345
Pr303 se Label out: None/Com1/Com2/LPT
Pr303
in u
Country Code: 123Not Label Type: 1-3
Effiency Mo:
in
Data out: None/Com1/Com2/LPT

Not
Efficiency Rh: Com1: 1200/9600
Efficiency Al: Com2: 1200/9600
HVL Mo:
HVL Rh:
HVL Al:
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

Service program Groups: - Defaults - Group B CONTROL PANEL BUTTONS

preprog Bucky 18x24


calibrating.....
CANCEL Auto Auto Auto SET
Dose Filter Rel

Auto kV mode Auto Time mode Manual mode


Button1 Button3
Button2
preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24
calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults
C: Diagnostics
Cancel - GROUPS - 000
D: AEC
Calibration
> AEC <
E: C-arm Rotation
up Accept down F: Collimator
G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement

Start Up Mode Settings PrB01


Scroll and select one
of the following modes
- START UP DEF - B01
B01: Start up settings Mode: AutokV/
- DEFAULTS - B00
B02: Bucky settings PreProg: /
PreProg2 B03: Mag settings Button3 up Accept down
> Start up < B04: Contact settings
up Accept down B05: QA mode settings
Pr301 Check Inputs

use
B06: PreProg1 settings Parameters:
Pr302 Check Bucky
B07: PreProg2 settings PreProg: 1/2/3/4
Pr303

in
B08: PreProg3 settings AutoRelease: on/off

Not
B09: PreProg4 settings AutoPark Back: on/off

Bucky Settings
Mode: AutokV/AutoTime/Manual
Default Filter: Mo/Rh/Alu PREPROG4 MODE SETTINGS:

se
Auto Dose: On/Off
Contact Settings PREPROG3 MODE SETTINGS:

u
Autopoint: On/off
Mode: AutokV/AutoTime/Manual

n
Default detector: 0-8 PREPROG2 MODE SETTINGS:

i
Not
Def kV Auto Time Mo:23-35 Default Filter: Mo/Rh/Alu

use
Def kV Auto Time Rh:23-35 Auto Dose: On/Off QA MODE SETTINGS:

se
Def kV Manual Mo :23-35 Autopoint: On/off
n
PREPROG1 MODE SETTINGS: QA- mode: None/Exp/Sensitom

u i
Not
Def kV Manual Rh :23-35 Default detector: 0-8 seDensity: -7 -+7

n in u
Default kV: 23-35

i Not useTarget Time: 1/2


Default mAs:0-33 Def kV Auto Time Mo:23-35

Not
Default Filter: Mo/Rh/QA
Def kV Auto Time Rh:23-35 o t in Auto kV Initial kV: -3- +3
N
Def kV Manual Mo :23-35 se Menu Text
in u
Def kV Manual Rh :23-35 Not kV Change Limit:0-5
Default mAs:0-33

Mag Settings
Mode: AutokV/AutoTime/Manual
Default Filter: Mo/Rh/Alu
Auto Dose: On/Off

use
Autopoint: On/off
Default detector: 0-8

in
Def kV Auto Time Mo:23-35

Not
Def kV Auto Time Rh:23-35
Def kV Manual Mo :23-35
Def kV Manual Rh :23-35
Default mAs:0-33
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

Service program Groups: - AEC - Group D CANCEL


preprog Bucky 18x24
calibrating.....
Auto Auto
Dose Filter
Auto
Rel
SET

Auto kV mode QA-coefficient history Detector Compensation Button1 Button3


Auto Time mode Manual mode
Button2

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24 - DETECTOR COMP -D04
Detector 1 : 0.98

use
calibrating..... calibrating..... calibrating.....
Auto Auto Auto > Detector 2 : 1.01
up Accept down
Rel Rel Rel
in
Pr301 Check Inputs
Not
Pr301 Check Inputs

use
Parameters:
Pr302 Check Bucky
Pr302 Check Bucky Detectors 1-8: 0.5 - 2.0
Pr303
in
Pr303

Not

MENU
Scroll and select one of
the following procedures

- AEC - D00
AEC gain
> Densities < Pr###
Scroll and select one of up Accept down
the following groups A: System setup
B: Defaults Pr301 Check Inputs
C: Diagnostics
- GROUPS - 000 PrD01
Pr302 AECBucky
Check setup
Cancel D: AEC
Calibration PrD02 AEC gains
Pr303
> AEC <
E: C-arm Rotation PrD03 Densities
up Accept down F: Collimator PrD04 Detector compensation
G: Compression PrD05 Target Times
H: Filter PrD06 initial kV-tables
I: Generator se PrD07 QA- coefficient history AEC gains
in u
J: Park Back Not PrD08 Detector adjustment
K: Stereo movement
- AEC gain - D02
AEC gain
AEC Setup g > 1.09
up Accept down

- AEC Setup - D01 Pr301 Check Inputs


Parameters:
Pr302 Check Bucky
> Target O.D 1.50 < AEC gain': 0.5 - 2.0
Pr303
up Accept down Detector L1:1.00
Detector C1:1.00
Pr301 Check Inputs Detector R1:1.00
Parameters: Detector L2:1.00
Pr302
se Check Bucky
in u Target O.D: 1.00 - 2.00 Detector C4:1.00
Not Pr303 Target lim l: 0.20-1.00
- Optical Density - D00 - Optical Density - D00 Detector C3:1.00
Target lim h: 1.00-3.00
Auto kV - Bucky Auto kV mode Detector C2:1.00
Autopoint lim: 50-2500
Mo Rh Alu Detector R2:1.00
-2 0 +1 Bucky Mag Contact

-7.0 Button1 Bucky


.. Button1 Rh
+7.0
Target times
Detector adjustment Initial kV-tables
-7.0 Button2 Mo
Button2 Mag Densities:
..
+7.0
- DENSITY - D03 - TARGET TIMES -D05
Button3 Contact -DETECT ADJUST-D08 - INIT KV-TABLE -D06
-7.0 Button3 Alu Auto kV Calc new gains based Target time2
.. > Auto Time < > Target time1 <
+7.0 on recent exposure?
up Accept down NO YES Bucky Mag Contact up Accept down

DENSITY
STEPS 0.5 Pr301 Check Inputs Pr301 Check Inputs Pr301 Check Inputs
Parameters: Pr301 Check Inputs Parameters:
Pr302 Check Bucky Pr302 Check Bucky Parameters: Pr302 Check Bucky
Pr302 Check Bucky Target time1
- Optical Density - D00 - Optical Density - D00 Auto kV
Pr303 Calc new gains based
Pr303 Bucky / Mag / Contact
Pr303 Pr303 Target time2
ScrollScroll
and select one
kV and of
change kV Compensation - Bucky kV Compensation Auto Time on recent exposure? se
the following procedures Thickness Compensation in u
optical density
Mo Rh Alu Bucky Mag Contact kV Compensation
NO YES Not
- DIAGNOSTICS -300 sTarget
e u
O.D (IQC)
- Optical Density - D00 Not
in
Ch. Inputs
kV Comp
> Ch. Bucky - Bucky - Rh
< Pr###
up kV: 23:
Accept down +0.0 Button1 Rh Button1 Bucky
MENU up next down

Pr301 Check Inputs


Parameters:
Button2 Mo Button2 Mag
Select holder Select holder and Mo/Rh-filter
Pr302 CheckBucky
kV range Bucky:23-39
Pr303 Target times for selected
Mag: :25-39 Button3 Alu Button3 Contact
New adjusted detector Initial kV-table for selected holder
Contact:23-39 holder and Mo or Rh-filter
gains
- TARGET TIMES- D05
NEW DTR GAINS- - INIT KV-TABLE -D06
Target1 - Bucky - Mo
DENSITY
Bucky
STEPS 0.5 DetectorL1 0.98 > Thickness1: 2.0 cm <
Thickness: 0.5: 23 kV
DetectorC1 1.00 < up Next down
Up Down up Next down
Scroll
Scrolland select one
parameter andof
change
the
thefollowing
value procedures Pr301 Check Inputs Pr301 Check Inputs
Parameters:
Pr301 Check Inputs Parameters: Pr302 Check Bucky 2 cm
Parameters: Pr302 Check Bucky Thickness1:
- DIAGNOSTICS -300 Pr302 Check Bucky Thickness: 0.5 - 10: 23-35kV Pr303
Ch.- Inputs
THICKNESS COMP - 200 DetectorL1-R2
Pr303 Pr303 Target1: 0.2-1.0s
Bucky - Rhodium Calculated gains can not Thickness2: 6.0 cm
> Ch. Bucky < Pr###
'P': 0.35 sec be changed here Target2: 1.0-3.0s
up Accept down
up next down

Pr301 Check Inputs


Parameters:
Pr302
'P' Check Bucky
Pr303
'r1'
'r2'
'S1'
'S2'
Service program Groups: - C-arm Rotation - Group E

Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24 Auto Drive PrE04
calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel
- AUTO DRIVE - E04

> Auto drive On <


up Next down

MENU
Pr301 Check Inputs
Parameters:
Pr302 Check Bucky
Autodrive
Pr303 On/Off

Scroll and select one of


the following groups A: System setup
Factory defaults
SET + CANCEL
B: Defaults
C: Diagnostics
- GROUPS - 000
Cancel D: AEC
Calibration
E: C-arm Rotation
> AEC <
CANCEL

up Accept down F: Collimator


G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement

Scroll and select one of Rotaion Speed PrE02 Projections PrE03


the following procedures
C-ARM CALIBRATION
PROCEDURE PrE01 BUTTON2
- C-ARM ROTATION- E00 -ROTATION SPEED-E02 - PROJECTIONS - E02
CANCEL Angle Calibration Slow Speed Mirror Angle: On
- ANGLE CALIBR. - E01 > Rot Speed < StartSpeed > 5 > Lateral angle 045 <
up Accept down up Accept down up Next down
Driven to 180 degrees
Down Accept Up Pr301 Check Inputs Pr301 Check Inputs Pr301 Check Inputs

use
PrE01
Pr302 Angle
Check Calibration
Bucky Parameters:
Pr302 Check Bucky Parameters:
Pr302 Check Bucky
PrE02 Rotation Speed
Pr303 Slow speed: 1-5
Pr303 Lateral angle 0-359 degrees
Pr303

n
BUTTON1/3
BUTTON2

PrE03 Projections Normal speed: 1-5 Oblique angle 0-359 degrees

i
Not
PrE04 Auto drive Ramp up: 1-5 Mirror angle: On/ Off
Ramp down: 1-5 in use
Not

- ANGLE CALIBR. - E01

Fail
- ANGLE CALIBR. - E01

Driven to 180 degr. BUTTON2


up Accept down

- ANGLE CALIBR. - E01

Pass

Speed

Normal speed

Detent speed Slow speed


(Slow speed)

Ramp up Ramp Time


down
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

Service program Groups: - Collimator - Group F CANCEL


preprog Bucky 18x24
calibrating.....
Auto Auto
Dose Filter
Auto
Rel
SET

Button1 Button3
Button2

Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults
C: Diagnostics
- GROUPS - 000
Cancel D: AEC
Calibration
E: C-arm Rotation
> AEC <
up Accept down F: Collimator
G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement

Scroll
Scroll
and
and
select
select
one
one
of of
the
the
following
following
procedures
procedures
CANCEL
- C-ARM
- COLLIMATOR
ROTATION-500 - F00
Angle
Rear Plate
Calibration
> Rotation
> Side
Speed
Plates
< Pr##
<
up up Accept
Acceptdowndown

Pr301
PrF01:
Angle
Rear
calibration
Plate Rear Plate PrF01 Side Plates PrF02 Collimator light Prf03
Pr302
PrF02:
Rotation
Side Plates
Speed
Pr303
PrF03:
Projections
Collimator Light

- REAR PLATE - F01 - SIDE PLATES - F02 - COLLIM. LIGHT -F03


Bucky 18x24: 12345 Bucky 18x24: 12345
Bucky 24x30: 12345 Bucky 24x30: 12345 Intensity > 5
up Accept down up Accept down up Accept down

Pr301 Check Inputs Pr301 Check Inputs Pr301 Check Inputs


Parameters:
Pr302 Check Bucky Pr302 Check Bucky
Parameters: Parameters:
Pr302 Check Bucky
Front Plate : 39002
Pr303 Pr303
Front Plate : 00000 Intensity : 1-5
Pr303
Bucky 18x24: 12345 Bucky 18x24: 12345 Time normal: 20 - 60 secs
Bucky 24x30: 12345 Bucky 24x30: 12345 Time Stereo/Biopsy: 20-90 sec
Mag 1.6 : 12345 Mag 1.6 : 12345
Unknown : 12345 Unknown : 12345
Cont 18x24 : 12345 Cont 18x24 : 12345
Cont 24x30 : 12345 Cont 24x30 : 12345
Mag 1.8 : 12345 Mag 1.8 : 12345
SpotMag 1.6: 12345 SpotMag 1.6: 12345
Stereo dig : 12345 Stereo dig : 12345
Stereo film : 12345 Stereo film : 12345
Mag2.0 : 12345 Mag2.0 : 12345
reserved : 12345 reserved : 12345
reserved : 12345 reserved : 12345
reserved : 12345 reserved : 12345
reserved : 12345 reserved : 12345
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

Service program Groups: - Compression - Group: G CANCEL


preprog Bucky 18x24
calibrating.....
Auto Auto Auto SET
Dose Filter Rel

Button1 Button3
Button2
Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults
C: Diagnostics CANCEL
Cancel - GROUPS - 000
D: AEC
Calibration
> AEC <
E: C-arm Rotation
up Accept down F: Collimator
G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement

Scroll
Scrolland
andselect
selectone
oneofof Compression Setup PrG01 Paddle Calibration PrG02 Cass Holder Calibration PrG03 Compression Force Limits PrG05
COMPRESSION FORCE the
thefollowing
followingprocedures
procedures
BUTTON2
CALIBRATION PROCEDURE - THICKNESS PAD - G02 - THICK. CAS - G03
- COMPR. SETUP- G01 standard small > +00 Bucky 18x24 > +00 - FORCE LIMITS- G05
- -COLLIMATOR
COMPRESSION - - ?00
G00
CANCEL Slow speed standard large > +25 Bucky 24x30 > +25 Position 5 > 175 N
- FORCE CALIBR. - G04 Angle
AngleCalibration
Calibration
> >Rotation
RotationSpeed
Speed< <Pr##
Pr## High Speed > 5 cm up Next down up Accept down Position 6 > 195 N
up Accept down up Accept down
Force to 0 N upup Accept
Accept down
down
Accept Pr301 Check Inputs Pr301 Check Inputs
Pr301 Check Inputs Parameters:
Pr302 Check Bucky Pr302 Check Bucky
Parameters: Pr301 Check Inputs
Parameters:
Pr302 Check Bucky Base point: - 9.5..9.5cm
Pr303 Pr303
no holder : - 9.5..9.5cm Parameters:
Pr302 Check Bucky
Pr301
PrG01Rear
Compression
Plate Setup
Slow Speed:1-5
Pr303 Standard small : - 9.5..9.5cm Bucky 18x24: - 9.5..9.5cm Mode: 0/1
Pr303
Pr302
PrG02Side
Thickness
Plates Paddle
BUTTON2

High Speed:1-5 Standard large : - 9.5..9.5cm Bucky 24x30: - 9.5..9.5cm Position 1 : 0-255N
Pr303
PrG02Collimator
ThicknessLight
Cass holder
Release Speed: 1-5 Mag spot: - 9.5..9.5cm Mag 1.6 : - 9.5..9.5cm Position 2 : 0-255N
PrG04 Force Calibration
PrG05 Force Limits AutoRelease travel: 0 - 10 cm Standard mag: - 9.5..9.5cm Unknown : - 9.5..9.5cm Position 3 : 0-255N
Speed Threshold: 0-150N Implant: - 9.5..9.5cm Cont 18x24 : - 9.5..9.5cm Position 4 : 0-255N
- FORCE CALIBR. - G04 Mag implant: - 9.5..9.5cm Cont 24x30 : - 9.5..9.5cm Position 5 : 0-255N
Delta32: - 9.5..9.5cm Mag 1.8 : - 9.5..9.5cm Position 6 : 0-255N
Fail Spot: - 9.5..9.5cm SpotMag 1.6: - 9.5..9.5cm
- FORCE CALIBR. - G04 Mag spot enhanc: - 9.5..9.5cm Stereo dig : - 9.5..9.5cm
Biopsy 1: - 9.5..9.5cm Stereo film : - 9.5..9.5cm
Force to > 200 N BUTTON2 Biopsy 2: - 9.5..9.5cm Mag2.0 : - 9.5..9.5cm
up Accept down Spot enhanced: - 9.5..9.5cm reserved : - 9.5..9.5cm
Reserved: - 9.5..9.5cm reserved : - 9.5..9.5cm
- FORCE CALIBR. - G04 Reserved: - 9.5..9.5cm reserved : - 9.5..9.5cm
Reserved: - 9.5..9.5cm reserved : - 9.5..9.5cm
Pass
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

Service program Groups: - Filter - Group H CANCEL


preprog Bucky 18x24
calibrating.....
Auto Auto
Dose Filter
Auto
Rel
SET

Button1 Button3
Button2

Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
Scroll and select one of the following procedures
the following groups A: System setup
B: Defaults - FILTER - H 00
C: Diagnostics
- GROUPS - 000
Cancel Calibration
D: AEC > Filter Calibr <
> AEC <
E: C-arm Rotation up Accept down
up Accept down F: Collimator
G: Compression
H: Filter H01 Filter Calibration
I: Generator
J: Park Back CANCEL

K: Stereo movement

Auto Filter Setup PrH01 Filter Position PrH02

-AUTOFILTER SUP-H01 - FILTER CALIBR- H02


Molybden: 12345
Change thickn > 5 cm Rhodium > 12345
up Accept down up Accept down

Pr301 Check Inputs Pr301 Check Inputs


Parameters:
Pr302 Check Bucky Parameters:
Pr302 Check Bucky
Change Thickness: 0- 15 cm
Pr303 Molybden: 12345
Pr303
Rhodium: 12345
Aluminum: 12345
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

Service program Groups: - Generator - Group I CANCEL


preprog Bucky 18x24
calibrating.....
Auto Auto
Dose Filter
Auto
Rel
SET

Button1 Button3
Button2

Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto Auto Auto Auto
Dose Filter Rel Filter Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults
C: Diagnostics
- GROUPS - 000 CANCEL
Cancel D: AEC
Calibration
> AEC <
E: C-arm Rotation
up Accept down F: Collimator
G: Compression
H: Filter
I: Generator
J: Park Back
K: Stereo movement

Scroll and select one of Preheat Compensation SF PrI05


Generator Setup PrI01 Preheat Compensation LF PrI04
the following procedures
PREHEAT ADJUSTMENT
BUTTON2 - GENERATOR - I00
PROCEDURE - GENERATOR SUP-I01 - PREH COMPENS -I04 - PREH COMPENS -I04
Generator Setup LF 26> 170 SF 26> 170
> Preh Calibration LF < SF 27> 180
CANCEL
Anode Brake> On LF 27> 180
- PREH. ADJUST. - I02 up Accept down
up Accept down up Accept down up Accept down
Press EXP button to
start automatic

use
calibration
se
Pr301 Check Inputs Pr301 Check Inputs Pr301 Check Inputs
PrI01 Generator Setup Parameters: Parameters: Parameters:

in u
Pr302 Check Bucky Pr302 Check Bucky Pr302 Check Bucky

in PrI02 Preheat Calibration LF Anode Brake: On/ Off Preheat LF: 0.5 - 1.50 Preheat SF: 0.5 - 1.50

Not
Pr303 Pr303 Pr303

Not
PrI03 Preheat Calibration SF LF 15-39: 0-255 (Large Focus) SF 25-39: 0-255 (Small Focus)
PrI04 Preheat Compensation LF
EX

PrI04 Preheat Compensation SF


PO
SU
RE

- PREH. ADJUST. - I02

Fail

use
in
Not
- PREH. ADJUST. - I02
New LF Preheat: 1.10
Pass
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

CONTROL PANEL BUTTONS

preprog Bucky 18x24

Service program Groups: - Tube Park - Group J CANCEL


calibrating.....
Auto Auto
Dose Filter
Auto
Rel
SET

Button1 Button3
Button2

Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto
Rel Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults Scroll and select one of
C: Diagnostics the following procedures
Cancel - GROUPS - 000
D: AEC
Calibration
> AEC < E: C-arm Rotation - TUBE MOVEMENT - J00
up Accept down F: Collimator
G: Compression > Speeds <
CANCEL
H: Filter up Accept down
I: Generator
J: Park Back
K: Stereo movement J01Speeds

Speeds PrJ01

- TUBE SPEEDS-J01
Fast Speed

se
Slow Speed > 3

u
up Accept down

in
Not
Pr301 Check Inputs
Parameters:
Pr302 Check Bucky
Slow Speed: 1-5
Pr303
Fast Speed: 1-5
Exposure Drive: 1-5
Alpha X Control Panel User Interface Specifications (SW) v1.8 19.3.01 / VEI

Service program Groups: - Stereo Movement - Group K


CONTROL PANEL BUTTONS

preprog Bucky 18x24


calibrating.....
CANCEL Auto Auto Auto SET
Dose Filter Rel

Button1 Button3
Button2
Auto kV mode Auto Time mode Manual mode

preprog Bucky 18x24 preprog Bucky 18x24 preprog Bucky 18x24


calibrating..... calibrating..... calibrating.....
Auto Auto Auto Auto Auto Auto
Dose Filter Rel Filter Rel Rel

MENU
Scroll and select one of
the following groups A: System setup
B: Defaults Scroll and select one of
C: Diagnostics the following procedures
Cancel - GROUPS - 000
D: AEC
Calibration
> AEC <
E: C-arm Rotation - STEREO MOVEMENT - K00
up Accept down F: Collimator Auto drive
G: Compression > Stereo Speeds <
up Accept down
CANCEL
H: Filter
I: Generator
J: Park Back
K: Stereo movement K01Speeds
K02 Auto drive

CANCEL Stereo Speeds PrK01

Auto Drive PrK02 -STEREO SPEEDS-K01


Fast Speed : 5
Slow Speed > 3
up Accept down
- AUTO DRIVE - K04

> Auto drive On < Pr301 Check Inputs


up Next down Parameters:
Pr302 Check Bucky
Low spd limit: 1-4
Pr303
High spd limit: 4-7
Pr301 Check Inputs Speed offset: -2 - 2
Parameters:
Pr302 Check Bucky Breaking spd: 2-4
Autodrive
Pr303 On/Off Ramp in 10ms: 5-50
A sample of Battery RAM contents, valid from software version 1.01.10 and up
(for details, refer to Diamond Service Manual (86502), Chapter 3.2, Service mode
overview):

01.04.17.
16:29

***** System setup *****

Serial#=19999
Country=1
Efficiency Mo=160
Efficiency Rh=120
Efficiency Al=350
HVL Mo=33
HVL Rh=38
HVL Al=47
Label out=0
Label type=1
Data out=0
COM1 Baud=1
COM2 Baud=1

***** Defaults *****

Startup:Preprog=1
Autorel=1
Autoparkback=0
Bucky:mode=0
B filter=0
B autodose=0
B autopoint=1
B detector=1
B autotime kV Mo=28
B autotime kV Rh=28
B manual kV Mo=28
B manual kV Rh=28
B def MAS=5
Mag:mode=0
M filter=0
M autodose=0
M autopoint=1
M detector=1
M autotime kV Mo=28
M autotime kV Rh=28
M manual kV Mo=28
M manual kV Rh=28
M def MAS=5
Contact:mode=0
C filter=0
C autodose=0
C autopoint=1
C detector=1
C autotime kV Mo=28
C autotime kV Rh=28
C manual kV Mo=28
C manual kV Rh=28
C def MAS=5
QA-mode=1
QA def kV=28
QA filter=0
Preprog1:density=0
PR1 targettime=1
PR1 auto kV initial=0 80
PR1 menutext=80 82 79 71 82 32 49 32
PR1 AkV change limit=3
Preprog2:density=0
PR2 targettime=2
PR2 auto kV initial=0
PR2 menutext=80 82 79 71 82 32 50 32
PR2 AkV change limit=3
Preprog3:density=0
PR3 targettime=1
PR3 auto kV initial=6
PR3 menutext=80 82 79 71 82 32 51 32
PR3 AkV change limit=3
Preprog4:density=0
PR4 targettime=2
PR4 auto kV initial=6
PR4 menutext=80 82 79 71 82 32 52 32
PR4 AkV change limit=3

***** Diagnostics *****

Image counter=0 0

***** AEC *****

Target O.D.=160
Target lim l=20
Target lim h=300
AutoPnt lim=1500
AEC gain=115
DetectorL1=100
DetectorC1=100
DetectorR1=100
DetectorL2=100
DetectorC4=100
DetectorC3=100
DetectorC2=100
DetectorR2=100
DEN_AUTOKV_BUCKY_MO=14
DEN_AUTOKV_BUCKY_RH=14
DEN_AUTOKV_BUCKY_AL=14
DEN_AUTOKV_MAG_MO=14
DEN_AUTOKV_MAG_RH=14
DEN_AUTOKV_MAG_AL=14
DEN_AUTOKV_CONTACT_MO=14
DEN_AUTOKV_CONTACT_RH=14
DEN_AUTOKV_CONTACT_AL=14
DEN_AUTOTIME_BUCKY_MO=14
DEN_AUTOTIME_BUCKY_RH=14
DEN_AUTOTIME_BUCKY_AL=14
DEN_AUTOTIME_MAG_MO=14
DEN_AUTOTIME_MAG_RH=14
DEN_AUTOTIME_MAG_AL=14
DEN_AUTOTIME_CONTACT_MO=14
DEN_AUTOTIME_CONTACT_RH=14
DEN_AUTOTIME_CONTACT_AL=14
THC_COMP_P_BUCKY_MO=30
THC_COMP_R1_BUCKY_MO=115
THC_COMP_R2_BUCKY_MO=100
THC_COMP_S1_BUCKY_MO=2500
THC_COMP_S2_BUCKY_MO=2500
THC_COMP_P_BUCKY_RH=30
THC_COMP_R1_BUCKY_RH=115
THC_COMP_R2_BUCKY_RH=100
THC_COMP_S1_BUCKY_RH=2500
THC_COMP_S2_BUCKY_RH=2500
THC_COMP_P_BUCKY_AL=100
THC_COMP_R1_BUCKY_AL=100
THC_COMP_R2_BUCKY_AL=100
THC_COMP_S1_BUCKY_AL=2500
THC_COMP_S2_BUCKY_AL=2500
THC_COMP_P_MAG_MO=30
THC_COMP_R1_MAG_MO=115
THC_COMP_R2_MAG_MO=100
THC_COMP_S1_MAG_MO=2500
THC_COMP_S2_MAG_MO=2500
THC_COMP_P_MAG_RH=100
THC_COMP_R1_MAG_RH=100
THC_COMP_R2_MAG_RH=100
THC_COMP_S1_MAG_RH=2500
THC_COMP_S2_MAG_RH=2500
THC_COMP_P_MAG_AL=100
THC_COMP_R1_MAG_AL=100
THC_COMP_R2_MAG_AL=100
THC_COMP_S1_MAG_AL=2500
THC_COMP_S2_MAG_AL=2500
THC_COMP_P_CONTACT_MO=100
THC_COMP_R1_CONTACT_MO=100
THC_COMP_R2_CONTACT_MO=100
THC_COMP_S1_CONTACT_MO=2500
THC_COMP_S2_CONTACT_MO=2500
THC_COMP_P_CONTACT_RH=100
THC_COMP_R1_CONTACT_RH=100
THC_COMP_R2_CONTACT_RH=100
THC_COMP_S1_CONTACT_RH=2500
THC_COMP_S2_CONTACT_RH=2500
THC_COMP_P_CONTACT_AL=100
THC_COMP_R1_CONTACT_AL=100
THC_COMP_R2_CONTACT_AL=100
THC_COMP_S1_CONTACT_AL=2500
THC_COMP_S2_CONTACT_AL=2500
KV_COMP_BUCKY_MO(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 10
0 100 100
KV_COMP_BUCKY_RH(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 10
0 100 100
KV_COMP_BUCKY_AL(36-39kV)=100 100 100 100
KV_COMP_MAG_MO(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_MAG_RH(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_MAG_AL(36-39kV)=100 100 100 100
KV_COMP_CONT_MO(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_CONT_RH(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_CONT_AL(36-39kV)=100 100 100 100
DETECTOR_COMPENSATION=100 100 100 100 100 100 100 100

***** TARGET TIMES MUST BE SAVED MANUALLY *****


INIT_KV_TBL_BUCKY=23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28 29 30
INIT_KV_TBL_MAG=25 25 25 25 25 25 25 26 26 26 27 27 28 28 28 29 29 30 30 30
INIT_KV_TBL_CONTACT=23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28 29 30

INIT_KV_TBL_DIG=23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28 29 30
QA_COEF_HISTORY=0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0

***** C-arm Rotat *****

ROTATION_SPEED=1 5 1 1
Lateral angle=90
Oblique angle=225
Mirror angle=0
Auto drive=1

***** Collimator *****

***** Rear Plate *****

no holder =39002
Bucky 18x24 =9500
Bucky 24x30 =1000
Mag 1.6 =9500
Unknown =10000
Cont 18x24 =9500
Cont 24x30 =1000
Mag 1.8 =9500
SpotMag 1.6 =9500
Stereo dig =27000
Stereo film =10000
Mag 2.0 =9500
Reserved =10000
Reserved =10000
Reserved =10000
Reserved =10000

***** Side Plates *****

no holder =0
Bucky 18x24 =4800
Bucky 24x30 =300
Mag 1.6 =4800
Unknown =4100
Cont 18x24 =4800
Cont 24x30 =300
Mag 1.8 =4800
SpotMag 1.6 =4800
Stereo dig =17500
Stereo film =4100
Mag 2.0 =4800
Reserved =4100
Reserved =4100
Reserved =4100
Reserved =4100
Intensity=6
Time normal=40
Time St/Bi=80

***** Compression *****


Slow speed=4
High speed=6
Release speed=5
AR travel=15
Sp threshold=50

***** Thick paddle *****

Base point =19


Stand small =19
Stand large =19
Mag spot =19
Mag spt sml =19
Implant =9
Mag implant =9
Delta32 =19
Spot =19
Mag spt enh =9
Biopsy D =19
Biopsy SQ =19
Spot enhanc =9
Biopsy spot =19
Reserved =19
Reserved =19

***** Thick Cas Ho *****

no holder =19
Bucky 18x24 =19
Bucky 24x30 =19
Mag 1.6 =19
Unknown =19
Cont 18x24 =19
Cont 24x30 =19
Mag 1.8 =10
SpotMag 1.6 =19
Stereo dig =7
Stereo film =19
Mag 2.0 =3
Reserved =19
Reserved =19
Reserved =19
Reserved =19
THICKNESS_BASE_POINT_BR=10
Force limit:mode=0
Force limits=70 90 110 130 150 170 200

***** Filter *****

FILTER_MO_CALIBRATION=1950
FILTER_RH_CALIBRATION=150
FILTER_AL_CALIBRATION=3030
Autofilter=0
Thickness=50

***** Generator *****

Anode Brake=0
Preheat=102
LF kV(range:15-39kV)=54 55 55 55 56 56 56 56 57 57 57 56 55 54 54 53 53 53 52 52
51 48 48 48 48
Preheat=101
SF kV(range:15-39kV)=48 48 49 49 49 49 49 49 50 49 49 48 48 47 47 47 46 46 46 45
45 45 45 45 45

***** Park Back *****

Speeds=2 5

***** Stereo move *****

Low spd limit=2


High spd limit=5
Speed offset=10
Breaking spd=3
Ramp in 10ms=25
Auto drive=0
Brightness=5
Contrast=5
Label=32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32
32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 32 3
2 32 32 32 32 32 32 32 32

***** Following not accessible from Service Menu *****

CHECKSUM_BR=13097
ANGLE_CALIBRATION_BR=81
FORCE_CALIBRATION_BR=100 40
CPULSE_BR=100 1005 0
STEREOARMPULSES_BR=24087
STEREOARMPULSEPERANGLE=27
STEREOARMZEROANGLE=1800
STEREOARMLEFTLIMANGLE=1650
STEREOARMRIGHTLIMANGLE=1950
SW_VERSION_BR=17201 0 10 21553 0 0
CARM_TEST_BR=0 2 10 0
TUBE_TEST_BR=0 2 10 0
COMP_TEST_BR=0 2 10 0
EPS_TEST_BR=0 2 10 0
STEREO_TEST_BR=0 2 10 0
COLR_TEST_BR=0 40 10 0
COLS_TEST_BR=0 60 10 0
FILT_TEST_BR=0 60 10 0
CAS_TEST_BR=0 2 10 0
GRID_TEST_BR=0 5 50 0
MIRROR_TEST_BR=0 10 50 0
LIGHT_TEST_BR=0 40 50 0
PLATE_TEST_BR=0 10 50 0
CARM_TEST_LANGLE=900
CARM_TEST_RANGLE=2700
STEREO_TEST_LANGLE=1650
STEREO_TEST_RANGLE=7
MGX / Diamond 27.12.2000 V1.1

Service software setting parameters, defaults and limits


Default min max Comments
Group Name Parameters Value value value
A - System Setup SYSTEM_SETUP_BR
Serial n:o: 19999 0 99999
Country Code: 1 0 999
Effiency Mo: 10.0 0 200 *10 in bram
Efficiency Rh: 10.0 0 200 *10 in bram
Efficiency Al: 10.0 0 1000 *10 in bram
HVL Mo: 0.34 0 60 *10 in bram
HVL Rh: 0.35 0 60 *10 in bram
HVL Al: 0.10 0 60 *10 in bram
Reserved 8 bytes
COMMSETUP_BR
Label out: None/Com1/Com2/LPTNone 0 3 Default: 0
Label Type: 1-3 2 0 2 Default: 1
Data out: None/Com1/Com2/LPT None 0 3 Default: 0
Com1: 1200/9600 9600 0 1 Default: 1
Com2: 1200/9600 9600 0 1 Default: 1
B - Defaults
START_UP_SETTINGS_BR
Preprog: (1/2/3/4) 1 1 4
Autorelease On/Off On 0 1 Default:1
Auto tube Park On/Off Off 0 1 Default:0
Reserved 8 bytes
BUCKY_SETTINGS_BR
Mode: A_kV/A_Time/Manual A_kV 0 2 Default:0
Default Filter: Mo/Rh/Alu Mo 0 2 Default:0
Auto Dose: On/ Off Off 0 1 Default:0
Autopoint: On/Off On 0 1 Default:1
Default detector: 0-8 1 0 8
Def kV Auto Time Mo:23-35 28 23 35
Def kV Auto Time Rh:23-35 28 23 35
Def kV Manual Mo :23-35 28 23 35
Def kV Manual Rh :23-35 28 23 35
Default mAs: 5 0 33
Reserved 8 bytes
MAG_SETTINGS_BR
Mode: A_kV/A_Time/Manual A_kV 0 2 Default:0
Default Filter: Mo/Rh/Auto Mo 0 2 Default:0
Auto Dose: On/ Off Off 0 1 Default:0
Autopoint: On/Off On 0 1 Default:1
Default detector: 0-8 1 0 8
Def kV Auto Time Mo:23-35 28 23 35
Def kV Auto Time Rh:23-35 28 23 35
Def kV Manual Mo :23-35 28 23 35
Def kV Manual Rh :23-35 28 23 35
Default mAs: 5 0 33
Reserved 8 bytes
CONTACT_SETTINGS_BR
Mode: A_kV/A_Time/Manual A_kV 0 2 Default:0
Default Filter: Mo/Rh/Auto Mo 0 2 Default:0
Auto Dose: On/ Off Off 0 1 Default:0
Autopoint: On/Off On 0 1 Default:1
Default detector: 0-8 1 0 8
Def kV Auto Time Mo:23-35 28 23 35
Def kV Auto Time Rh:23-35 28 23 35
Def kV Manual Mo :23-35 28 23 35
Def kV Manual Rh :23-35 28 23 35
Default mAs: TBD 0 33
Reserved 8 bytes
QAMODE_SETTINGS_BR
QA- mode: None/Exp/Sensitom Exp 0 2 Default:1
Default kV: 23-35 28 23 35
Default Filter: Mo/Rh/QA Mo 0 2 Default:0
B - Defaults Reserved 14 bytes
PREPROG1_SETTINGS_BR
PREPROG1_DENSITY_BR Density: -7 -+7 0 0 14 -7.0 to +7.0, step:0.5
PREPROG1_TARGETTIME_BR Target Time: 1/2 1 1 2
PREPROG1_AKVINITIAL_BR Auto kV Initial kV: -3- +3 0 0 6
PREPROG1_MENUTEXT_BR Menu Text PROGR 1 32 90 text string 8
PREPROG1_AKVLIMIT_BR AkV change limit 3 0 5
B - Defaults Reserved 8 bytes
PREPROG2_SETTINGS_BR
PREPROG2_DENSITY_BR Density: -7 -+7 0 0 14 -7.0 to +7.0, step:0.5
PREPROG2_TARGETTIME_BR Target Time: 1/2 1 1 2
PREPROG2_AKVINITIAL_BR Auto kV Initial kV: -3- +3 0 0 6
PREPROG2_MENUTEXT_BR Menu Text PROGR 2 32 90 text string 8
PREPROG2_AKVLIMIT_BR AkV change limit 3 0 5
B - Defaults Reserved 8 bytes
PREPROG3_SETTINGS_BR
PREPROG3_DENSITY_BR Density: -7 -+7 0 0 14 -7.0 to +7.0, step:0.5
PREPROG3_TARGETTIME_BR Target Time: 1/2 1 1 2
PREPROG3_AKVINITIAL_BR Auto kV Initial kV: -3- +3 0 0 6
PREPROG3_MENUTEXT_BR Menu Text PROGR 3 32 90 text string 8
PREPROG3_AKVLIMIT_BR AkV change limit 3 0 5
Reserved 8 bytes
PREPROG4_SETTINGS_BR
PREPROG4_DENSITY_BR Density: -7 -+7 0 0 14 -7.0 to +7.0, step:0.5
PREPROG4_TARGETTIME_BR Target Time: 1/2 1 1 2
PREPROG4_AKVINITIAL_BR Auto kV Initial kV: -3- +3 0 0 6
PREPROG4_MENUTEXT_BR Menu Text PROGR 3 32 90 text string 8
PREPROG4_AKVLIMIT_BR AkV change limit 3 0 5
Reserved 8 bytes
C- Diagnostics
IMAGE_COUNTER_BR Exposure counter 1 0 0 65535 2 words
ERROR_COUNTERS_BR Not used
LAST_ERROR_BR Not used
CHECK_BUCKY_BR Not used
PRINT_INFO_BR Not used
CHECK_ERROR_CODES_BR Not used
CHECK_ANODE_BR Not used
D - AEC
AEC_SETUP_BR
AEC_SETUP_TARGET_OD_BR 160 100 200
AEC_SETUP_TARGET_LIM_L_BR 20 20 100
AEC_SETUP_TARGET_LIM_H_BR 300 100 300
AEC_SETUP_AUTOPNT_LIM_BR 1500 50 2500
Reserved 8 bytes
AEC_GAINS_BR
Base gain 100 40 200
Detector 0 100 40 200
Detector 1 100 40 200
Detector 2 100 40 200
Detector 3 100 40 200
Detector 4 100 40 200
Detector 5 100 40 200
Detector 6 100 40 200
Detector 7 100 40 200

AUTO_KV_BUCKY_BR
Density- Mo (-7.0-+7.0) 0,0 0 28 Default:14
Density - RH (-7.0-+7.0) 0,0 0 28 Default:14
Density- Al (-7.0-+7.0) 0,0 0 28 Default:14
AUTO_KV_MAG_BR
Density- Mo (-7.0-+7.0) 0,0 0 28 Default:14
Density - RH (-7.0-+7.0) 0,0 0 28 Default:14
Density- Al (-7.0-+7.0) 0,0 0 28 Default:14
AUTO_KV_CONTACT_BR
Density- Mo (-7.0-+7.0) 0,0 0 28 Default:14
Density - RH (-7.0-+7.0) 0,0 0 28 Default:14
Density- Al (-7.0-+7.0) 0,0 0 28 Default:14
AUTO_TIME_BUCKY_BR
Density- Mo (-7.0-+7.0) 0,0 0 28 Default:14
Density - RH (-7.0-+7.0) 0,0 0 28 Default:14
Density- Al (-7.0-+7.0) 0,0 0 28 Default:14
AUTO_TIME_MAG_BR
Density- Mo (-7.0-+7.0) 0,0 0 28 Default:14
Density - RH (-7.0-+7.0) 0,0 0 28 Default:14
Density- Al (-7.0-+7.0) 0,0 0 28 Default:14
AUTO_TIME_CONTACT_BR
Density- Mo (-7.0-+7.0) 0,0 0 28 Default:14
Density - RH (-7.0-+7.0) 0,0 0 28 Default:14
Density- Al (-7.0-+7.0) 0,0 0 28 Default:14
Reserved 10 bytes
THICK_COMP_BUCKY_BR
Thick_comp for Mo
Mo-P 0,30 30 200 Default:30
Mo-r1 1,15 50 200 Default:115
Mo-r2 1,00 50 200 Default:100
Mo-S1 2500 0 7000
Mo-S2 2500 0 7000
Thick_comp for Rh
Rh-P 0,30 30 200 Default:30
Rh-r1 1,15 50 200 Default:115
Rh-r2 1,00 50 200 Default:100
Rh-S1 2500 0 7000
Rh-S2 2500 0 7000
Thick_comp for Al
Al-P 1,00 30 200 Default:100
Al-r1 1,00 50 200 Default:100
Al-r2 1,00 50 200 Default:100
Al-S1 2500 0 7000
Al-S2 2500 0 7000
THICK_COMP_MAG_BR
D - AEC Thick_comp for mo
Mo-P 0,30 30 200 Default:30
Mo-r1 1,15 50 200 Default:115
Mo-r2 1,00 50 200 Default:100
Mo-S1 2500 0 7000
Mo-S2 2500 0 7000
Thick_comp for Rh
Rh-P 1,00 30 200 Default:100
Rh-r1 1,00 50 200 Default:100
Rh-r2 1,00 50 200 Default:100
Rh-S1 2500 0 7000
Rh-S2 2500 0 7000
Thick_comp for Al
Al-P 1,00 30 200 Default:100
Al-r1 1,00 50 200 Default:100
Al-r2 1,00 50 200 Default:100
Al-S1 2500 0 7000
Al-S2 2500 0 7000
THICK_COMP_CONTACT_BR
KV_COMP_BUCKY_BR
KV_COMP_MAG_BR
KV_COMP_CONTACT_BR
DETECTOR_COMPENSATION_BR
D - AEC TARGET_TIMES1_BR a*x+s*b, adjusted with end points
TARGET_TIMES2_BR a*x+s*b, adjusted with end points
Reserved 36 bytes
INITIAL_KV_TABLE_BUCKY_BR
INITIAL_KV_TABLE_MAG_BR
INITIAL_KV_TABLE_CONTACT_BR
INITIAL_KV_TABLE_DIG_BR
Reserved 40 bytes
QA_COEF_HISTORY_BR
Reserved 14 bytes
E - C-arm Rotation ROTATION_SPEED_BR
Slow speed: 1-5 TBD Not used
Normal speed: 1-5 TBD Not used
Ramp up: 1-5 TBD Not used
Ramp down: 1-5 TBD Not used
PROJECTIONS_BR
Lateral Angle 90 0 359 Tube up is 180, left straight angle 90
Oblique angle 225 0 359 Tube up is 180, left straight angle 90
Mirror Angle On/off Off 0 1 Default:0
CARM_BUTTONS_BR
CARM_CONTINUOUS_DRV_BR Continuous drive (On/Off)
Reserved 4 bytes
REAR_PLATE_BR
Front plate 39002 0 50000
Bucky 18x24 10000 0 50000
Bucky 24x30 1600 0 50000
Mag 1.6 14000 0 50000
Unknown 10000 0 50000
Cont 18x24 10000 0 50000
Cont 24x30 9000 0 50000
Mag 1.8 14000 0 50000
SpotMag 1.6 10000 0 50000
Stereo dig 27000 0 50000
Stereo film 10000 0 50000
Mag 2.0 14000 0 50000
Reserved 10000 0 50000
Reserved 10000 0 50000
Reserved 10000 0 50000
Reserved 10000 0 50000
F - Collimator SIDE_PLATES_BR
Front plate 0 0 50000
Bucky 18x24 4100 0 50000
Bucky 24x30 0 0 50000
Mag 1.6 3600 0 50000
Unknown 4100 0 50000
Cont 18x24 4100 0 50000
Cont 24x30 0 0 50000
Mag 1.8 3600 0 50000
SpotMag 1.6 4100 0 50000
Stereo dig 17500 0 50000
Stereo film 4100 0 50000
Mag 2.0 3600 0 50000
Reserved 4100 0 50000
Reserved 4100 0 50000
Reserved 4100 0 50000
Reserved 4100 0 50000
F - Collimator COLLIMATOR_LIGHT_BR Light Intensity 3 0 10
COL_LIGHT_TIME_BR Time Normal 20-60sec 40 20 60
time Stero/biopsy 20-90sec 80 20 90
Reserved 10 bytes
COMPRESSION_SETUP_BR
Slow Speed:1-5 3 2 7
High Speed:1-5 5 2 7
Release Speed: 1-5 5 1 7
AutoRelease travel: 0 - 20 cm 15 0 20
Speed Threshold: 0-150N 50 0 150
G - Compression Reserved 6 bytes
THICKNESS_PADDLE_BR
Base point 19 0 40 -9.5 to +9.5, step: 0.5
Stand small 19 0 40 -9.5 to +9.5, step: 0.5
Stand large 19 0 40 -9.5 to +9.5, step: 0.5
Magnification spot 19 0 40 -9.5 to +9.5, step: 0.5
Standard mag 19 0 40 -9.5 to +9.5, step: 0.5
Implant 9 0 40 -9.5 to +9.5, step: 0.5
Magnification implant 9 0 40 -9.5 to +9.5, step: 0.5
Delta32 14 0 40 -9.5 to +9.5, step: 0.5
Spot 19 0 40 -9.5 to +9.5, step: 0.5
Magnification spot enhanced 9 0 40 -9.5 to +9.5, step: 0.5
Biopsy 1 19 0 40 -9.5 to +9.5, step: 0.5
Biopsy 2 19 0 40 -9.5 to +9.5, step: 0.5
Spot enhanced 9 0 40 -9.5 to +9.5, step: 0.5
Reserved 19 0 40 -9.5 to +9.5, step: 0.5
Reserved 19 0 40 -9.5 to +9.5, step: 0.5
Reserved 19 0 40 -9.5 to +9.5, step: 0.5
G - Compression THICKNESS_CASS_HOLDER_BR
No holder 19 0 40 -5.0 to +5.0, step: 0.5
Bucky 18x24 19 0 40 -5.0 to +5.0, step: 0.5
Bucky 24x30 19 0 40 -5.0 to +5.0, step: 0.5
Mag 1.6 19 0 40 -5.0 to +5.0, step: 0.5
Unknown 19 0 40 -5.0 to +5.0, step: 0.5
Cont 18x24 19 0 40 -5.0 to +5.0, step: 0.5
Cont 24x30 19 0 40 -5.0 to +5.0, step: 0.5
Mag 1.8 10 0 40 -5.0 to +5.0, step: 0.5
Mag 1.8 spot 19 0 40 -5.0 to +5.0, step: 0.5
Stereo digital 19 0 40 -5.0 to +5.0, step: 0.5
Stereo film 7 0 40 -5.0 to +5.0, step: 0.5
Mag 2.0 3 0 40 -5.0 to +5.0, step: 0.5
Reserved 19 0 40 -5.0 to +5.0, step: 0.5
Reserved 19 0 40 -5.0 to +5.0, step: 0.5
Reserved 19 0 40 -5.0 to +5.0, step: 0.5
Reserved 19 0 40 -5.0 to +5.0, step: 0.5
G - Compression THICKNESS_BASE_POINT_BR Thickness base point 10 Not used
FORCE_LIMITS_BR
Mode: 0/1 0 0 1 0=Absolute limit, 1=stepwise limit
FORCE_LIMITS_STOP_BR
Position 1 : 0-255N 100 0 255
Position 2 : 0-255N 120 0 255
Position 3 : 0-255N 140 0 255
Position 4 : 0-255N 160 0 255
Position 5 : 0-255N 180 0 255
Position 6 : 0-255N 200 0 255
FORCE_LIMIT_DELTA_BR
Max delta compression 200 Not used
Reserved 2 bytes 50 Not used
H - Filter FILTER_CALIBRATION_BR
Mo 1950 0 20000
Rh 150 0 20000
Al 3030 0 20000
AUTO_FILTER_SET_BR
AUTO_FILTER_MODE_BR On/Off 0 0 1 Default: Off
AUTO_FILTER_THICK_BR Thickness of filter change point 50
Reserved 4 bytes
I - Generator GENERATOR_SETUP_BR
Anode Brake On/Off 0 Default: Off
Use high mAs On/Off 0 Not used
Min kV 23 Not used
Max kV 39 Not used
Reserved 8 bytes
PREHEAT_COMPENSATION_LF_BR
Preheat Calib_LF (-20 - +20) 100 80 120 Default: 0
Preheat_LF 15kV 54 20 235
Preheat_LF 16kV 55 20 235
Preheat_LF 17kV 55 20 235
Preheat_LF 18kV 55 20 235
Preheat_LF 19kV 56 20 235
Preheat_LF 20kV 56 20 235
Preheat_LF 21kV 56 20 235
Preheat_LF 22kV 56 20 235
Preheat_LF 23kV 57 20 235
Preheat_LF 24kV 57 20 235
Preheat_LF 25kV 57 20 235
Preheat_LF 26kV 56 20 235
Preheat_LF 27kV 55 20 235
Preheat_LF 28kV 54 20 235
Preheat_LF 29kV 54 20 235
Preheat_LF 30kV 53 20 235
Preheat_LF 31kV 53 20 235
Preheat_LF 32kV 53 20 235
Preheat_LF 33kV 52 20 235
Preheat_LF 34kV 52 20 235
Preheat_LF 35kV 51 20 235
Preheat_LF 36kV 50 20 235
Preheat_LF 37kV 50 20 235
Preheat_LF 38kV 50 20 235
Preheat_LF 39kV 50 20 235
I - Generator PREHEAT_COMPENSATION_SF_BR
Preheat Calib_SF (0.5-1.50) 100 80 120 Default: 0
Preheat_SF 15kV 48 20 235
Preheat_SF 16kV 48 20 235
Preheat_SF 17kV 49 20 235
Preheat_SF 18kV 49 20 235
Preheat_SF 19kV 49 20 235
Preheat_SF 20kV 49 20 235
Preheat_SF 21kV 49 20 235
Preheat_SF 22kV 49 20 235
Preheat_SF 23kV 50 20 235
Preheat_SF 24kV 49 20 235
Preheat_SF 25kV 49 20 235
Preheat_SF 26kV 48 20 235
Preheat_SF 27kV 48 20 235
Preheat_SF 28kV 47 20 235
Preheat_SF 29kV 47 20 235
Preheat_SF 30kV 47 20 235
Preheat_SF 31kV 46 20 235
Preheat_SF 32kV 46 20 235
Preheat_SF 33kV 46 20 235
Preheat_SF 34kV 45 20 235
Preheat_SF 35kV 45 20 235
Preheat_SF 36kV 45 20 235
Preheat_SF 37kV 45 20 235
Preheat_SF 38kV 45 20 235
Preheat_SF 39kV 45 20 235
Reserved 8 bytes
J - Tube Park SPEEDSTUBE_BR
Fast Speed 1-5 5 1 5 Not used
Slow Speed 2 1 5 Not used
Reserved 12 bytes
K - Stereo SPEEDSSTEREO_BR
STEREO_LOW_SPD_LIMIT_BR Slow Speed 2
STEREO_HIGH_SPD_LIMIT_BR High Speed 5
STEREO_SPD_OFFSET_BR Speeds table offset 10 8 12 -2 - +2
STEREO_BREAK_SPD_BR Breaking speed 3
STEREO_RAMP_DELAY_BR Accel. ramp speed chg.delay 25 in 10 ms
STEREOARM_CONTINUOUS_DRV_BR
Continuous drive (On/Off) 0 0 1 Default: Off
Reserved 16 bytes
Display BRIGHTNESS_BR
Brightness 5 1 5
Contrast 5 1 5
LABEL_BR
Reserved 22 bytes
CHECKSUM_BR
ANGLE_CALIBRATION_BR
FORCE_CALIBRATION_BR
Force slope 100 0 65535
Force offset 40 0 65535
CPULSE_BR
SW_VERSION_BR
STEREOARMPULSES_BR
STEREOARMPULSEPERANGLE_BR
STEREOARMZEROANGLE_BR
$(&&$/,%5$7,213$*( PAGE /

8QLW7\SH 6HULDOQR 'DWH 3URFHVVRUWHPSHUDWXUH

)LOP 6FUHHQ 3URFHVVRUW\SH 3URFHVVRUWLPH

2' 0RGH &DVVHWWH 7KLFNQ )LOWHU N9 N9&RPS P$V WLPH J S U U V V $WWHQWLRQ1RWHV


Diamond AEC detector locations
for detector-specific film O.D. measurement
Film marker text

R1
R2
C4

C3

C2

C1
L2

L1

Instructions:

1) Place a film within the frame

2) Draw the ellipse of the used detector on the film

3) Mark the name of the detector on the film

4) Measure the O.D and write it on the film.


0$18)$&785(5

,QVWUXPHQWDULXP,PDJLQJ
P.O.Box 20, FIN-04301 Tuusula, Finland
Tel. + 358 10 394 6500 • Fax + 358 10 394 6501
info@fi.instrumentarium.com • www.instrumentarium.com/imaging

,167580(17$5,80,0$*,1*2)),&(62876,'(2)),1/$1'

,QVWUXPHQWDULXP,PDJLQJ,QF
300 West Edgerton Avenue, Milwaukee, Wisconsin 53207, USA
Tel. +1 414 747 1030 • Tel. (800) 558 6120 • Fax +1 414 481 8665
info@usa.instrumentarium.com • usa.instrumentarium.com

,QVWUXPHQWDULXP,PDJLQJ)UDQFH
4, Avenue des Roses, 94386 Bonneuil Sur Marne Cedex, France
Tel. +33 1 43 39 51 51 • Fax +33 1 43 39 75 75
instrumentarium.imaging@wanadoo.fr

,QVWUXPHQWDULXP,PDJLQJ,WDOLDVUO
20041 Agrate Brianza (MI) Centro Direzionale Colleoni,
Viale Colleoni, 11, Interno 3, Italia
Tel. +39 039 689 0408 • Fax +39 039 609 1299
instrumentarium@it.instrumentarium.com • www.instrumentarium.it

,QVWUXPHQWDULXP,PDJLQJ'HQWDO*PE+
P.O.Box 2044, 77680 KEHL AM RHEIN, Germany
Tel. +49 7851 932 90 • Fax +49 7851 932 930
kontakt@instrumentarium-imaging.de

,QVWUXPHQWDULXP,PDJLQJ=,(+0*PE+
Isarstrasse 40, D-90451 Nuremberg, Germany
Tel. +49 911 642 070 • Fax +49 911 642 0739
info@ziehm-eu.com • www.ziehm-eu.com

=,(+0,QWHUQDWLRQDO0HGLFDO6\VWHPV
4181 Latham Street Riverside, CA 92501, (USA)
Tel. + 1 909 781 6660 • Fax +1 909 781 6457
mail@ziehm.com • www.ziehm.com

You might also like